Sei sulla pagina 1di 42

SIEMENS

POLYMOBIL III
RX

Service Instructions
from Serial no. 3000

0 Siemens AG 1995
The reproduction, transmission or
us8 of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
granl or registration of a utility
I model or design, are resewed.

Register 5 English
Print No.: RX88-115.061.02.01.02 Date: 10.95
Replaces: n.a.
Chapter 0 Revision
Page 2 of 4
This page intentionally left blank.

POLYMOBIL III Register 5 RXe8-115.061.02 TDRXI Siemens AG


Rev. 01 10.95 Medical Engineering
Contents Chapter 0
Page 3 of 4
Page
1 Prerequisites 1-l
Specially marked text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 1
Documents required. . . _ . . . . . . _ . . . . . . . . . . . . . . . . . . . . . . . .1- 1
Tools and measuring instruments required . . . . . . . . . . . . . . . . _ . . . _ . .1 - 1
Safety notes and protective measures . . . . . . . . . . . . . . . . . . . . . . . . .1- 2
Replacing damaged or missing screws . . . . . . . . . . . . . . . . . . . . . . _ . .1 - 3
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . _ . . . . .l-3

Test with the unit in the operative condition 2-I


High-voltage test with the radiation detector . . . . . . . . . . . . . . . . . . . . . .2 - 1

Error messages 3-1


There are three types of errors: . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1

4 Notes on troubleshooting 49 1
Checking the line voltage, fuses and LEDs ...................... .4 - 1
Checking the supply voltages ............................. .4 - 6
Checking the intermediate circuit voltage ....................... .4 - 8
Checking the maximum main inverter frequency ................... -4 - 9
Oscillating current measurement ........................... 4 - 10
Checking the high-voltage kVNoM and kV,cT ..................... 4 - 12
Checking the filament and tube current ........................ 4 - 13
Setting the maximum filament frequency ....................... 4 - 15
Checking the tube current. .............................. 4 - 16
Checking the kV and tube current (IR) ........................ 4 - 17
Checking the mAs values ............................... 4 - 18
Adjusting the mAs ................................... 4 - 19
Aligning the light field to the radiation field ...................... 4 - 20
Checking and readjusting the counterweight ..................... 4 - 23
Readjusting the counterweight with accessories attached .............. 4 - 23
Checking the control elements on the collimator ................... 4 - 24

5 Replacing important components 5-1


Replacing the single tank ................................ 5 -1
Replacing the collimator. ............................... .5 -2
Replacing the light-beam localizer lamp. ....................... .5 -3
Replacing the caster. ................................. .5 -5

Wmens AG TDRXl RXBEl15.96l~Z Register 5 POLYMOBIL III


Medical Engineenng Rev. 01
Chapter 0 Contents
Page 4 of 4

This page intentionally left blank.

Register 5 TDRXl Siemens AG


POLYMCBIL III Medical Engineering
Prerequisites Chapter 1
Page 1 of 4
Specially marked text

All texts marked with “Warning” contain information regarding


potential hazards to health or life.

11 CAUTION ‘11
1
All texts marked “Caution!” contain information regarding poten-
I’
tial hazards and steps to be taken to avoid hazards.

, -I

1 NOTE 1 All texts marked “Note:” contain additional information regarding


the following procedure step. It is intended to clarify the step or to
indicate ways to avoid predictable problems.

Documents required
l Wiring diagram X0381
l Description of functions RXB&115.641.02...

Tools and measuring instruments required


l Usual service tools
l Digital multimeter:
Fluke 8060 A 97 02 101 Y4290
l 2 channel storage oscilloscope:
Tektronik Type 2232 97 02 234 Y3155
l mAsmeter 81 60 400 RE999
l Protective ground wire and leakage current
tester: BENDER safety tester 97 06 979 Y0526
l Radiation detector 96 60 754 RE999
See SPEED INFO 93/91

During oscilloscope operation the protective ground wire connec-


tion in the power cable must not be interrupted under any circum-
stances.
For measurements where ground loops that may be present
could impair the measuring result, use the TEK amplifier and the
trigger attachment.

Siemens AG TDRXl Register 5 POLYMcf3lL III


Medical Enqneenng
Chapter 1 Prerequisites
Page 2 of 4
Safety notes and protective measures

When performing work and tests, the following must be


j( l
obese& .
- the product-specific safety notes contained in the document,
- the safety notes RAO-000.012.40... in Register 2 of the
POLYMOBIL III file,
- and the general safety notes contained in Register 2 of the TI
binder.

l After the system has been switched off, a de voltage of approx.


300 V is applied for the main inverter.
On D920 the yellow LED V35 lights up.
Within approx. 1.5 minutes the voltage drops to approx. 12 V;
LED V35 goes out.

l Checks and adjustments that must be carried out with radiation


switched on are marked with the symbol 9
When performing work procedures which are marked with this
symbol, radiation protective clothing must be worn.

l Connect the POLY MOENL I II only to a power source (power outlet) meeting the
standards of VDE 0107 or the usual local standards.
l Before working on the open POLYMOBIL III switch OFF the unit with the power switch at
the control panele and unplug the power cable.
l Before removing or inserting PC boards, switch off the generator; in doing so, observe
the ESD regulations.

POLYMOBIL III Register 5 RXBB-115.061.02 TD RX 1 Siemens AG


Rev. 01 10.95 Medial Engineering
Prerequisites Chapter 1
Page 3 of 4
Replacing damaged or missing screws
l Damaged or missing screws must only be replaced by steel screws as specified in the
instaliation drawings that conform to DIN 267 and have the specified tensile strength.

[j[ All Allen screws must have a tensile strength rating of 8.8.

Cleaning
l The unit must always be switched off or disconnected before cleaning.
l Use only water or a luke warm mixture of a household cleaner diluted with water to clean
the unit.
l Do not use an abrasive or organic solvents or cleaning agents containing solvents such
as gasoline used for cleaning purposes, alcohol or stain remover. Do not spray water on
the unit.
l For further information, refer to the Operating Instructions Y2leaningIDisinfecting”.

TD RX 1 RXB8-115.0~; ;o Register 5 POLYMOBIL Ill


Siemens AG
Medical Engineering Rev.01 .
Chapter 1 Prerequisites
Page 4 of 4
This page intentionally left blank.

POLYMOBIL III Register 5 RXB8-115.yl~2 TDRXl Siemens AG


Rev. 01 Medical Engineeting
Test with the unit in the operative condition Chapter 2
Page 1 of 2

Fig. 1

High-voltage test with the radiation detector


1 NOTICE 1 The amplitude can be different for each detector.

. POLYMOBIL ON
l Connect radiation detector to the oscilloscope and place it in the beam path of the
POLYMOBIL III (see Fig. 1).
l Set the slide switch on the radiation detector to measuring range 1 (1 point visible).
l Set an SID of 40 cm and collimate onto the radiation detector with the light localizer
(Fig.1).
l Trigger check exposures using the exposure data specified in the oscillograms (see
Fig.2).

Siemens AG Register 5 POLYMOBIL III


Medial Engineenng
Chapter 2 Test with the unit in the operative condition
Page 2 of 2

CHl: 020mV :!jOms


CH2: 020 mV : 50 ms
90 kViI 0 mAs CH3: 020 mV : 50 ms
CH4: 020 mV : 50 ms
81 kVl10 mAs

70 kVll0 mAs
60 kV/lO mAs

60 kW40 mAs
CHl: 10.0 mV : 200 ms’
CH2: 10.0 mV : 200 ms
60 kW20 mAs
CH3: 10.0 mV : 200 ms

60 kVil0 mAs

Fig. 2

POLYMOBLL III Register5 RXBB-115061.02 TDRXl Siemens


AG
Rev. 01 10.95 Medial Engineering
Error messages Chapter 3
Page 1 of 2
There are three types of errors:
The following error codes are shown in the kV or mAs display on the control panel.

lnitialization error
go... EPROM check sum error
95.. (l-t,,, circuit error
96.. mAno,.,,circuit error
97... kv,,,,., circuit error

Stand-by error:
03... IH, c IHstby
04... IHaCt> IHstby
05.. kV,,,<>O
06... mAad<>
09... High pressure in the tank

Exposure error:
Il... lnverter short circuit
12... kVmax
13... Imax
14... kV,, < kVnom
15.. mAact c mAnom
16... mA, > mAnom
17... maximum exposure time
18... exposure not completed

TDRXl Register 5 POLYMOBIL III


Slemens AG
Medical Engineering
Chapter 3 Error messages
Page 2 of 2
This page intentionally left blank,

POLYMOBIL Ill Register 5 TDRXl Siemem AG


E% 1 Y&F Medii Engineering
Notes on trouble-shooting Chapter 4
Page 1 of 24
In case of errors which impair radiography, perform the following checks.

Checking the line voltage, fuses and LEDs


a) Checking the line voltage
l Measure the supply voltage on site at the safety socket using the digital muttimeter.

For voltages N 240 V AC see Startup Instructions


RXBS115.034.02... chapter 3.

l Plug the power plug of the POLYMOBIL into the safety socket.
l POLYMOBIL ON

If the POLYMOBIL cannot be switched on although the power


plug and power voltage have been connected, the overcurrent cir-
cuit breaker Fl may have responded (automatic circuit breaker
l/Fig. 1)

. POLYMOBIL OFF.

Siemens AG TD RX 1 RXBB-115.061.02 R3gster 5 POLYMOBIL III


Medical Engineenng Rev. 01 10.95
Chapter 4 Notes on trouble-shooting
Page 2 of 24

Fig. 1

b) Checking the fuses


l Checking the overcurrent circuit breaker:
- Loosen the hand screw and the protective ground wire terminal and take the switch
box off the cassette compartment.
- Check if the overcurrent circuit breaker Fl (l/Fig. 1) has responded and, if so, unlock it.

Fuse value: 15 A

POLYMOBIL III Register 5 RXB&ll5.0l6;~.J11 TDRXl Siamens AG


Rev. 01 Medical Engineering
Notes on trouble-shooting Chapter 4
Page 3 of 24

09’20 (X0381-92;)

l Checking the fuses on 0920

Loosen the 4 Allen screws (ZFig.1, 4 mm key size) and remove the cover.

Check the following fuses:

Fl 2 AT Closing circuit
F2 2 AT Closing circuit
F3 0.25 AT Tl secondary
F5 4 AT Filament circuit
F6 3.2 AT 24 V / Main inverter control
F7 10 AT Light-beam localizerr

Stemens AG TDRXl Register 5 POLYMOBIL III


Medical Engineermg
Chapter 4 Notes on trouble-shooting
Page 4 of 24

0920 (X0381-92D)

c) Checking the LEDs on D910 and D920


. POLYMOBIL ON

After the initialization, the default data 60 kV and 10 mAs are displayed.

With a supply voltage of 110 V, LED Vl2 (green) is illuminated on


I ,,g,O
.

Stand-by operation
l On D920 the following LEDs are illuminated:
V40 (yellow) +24V
V39 (yellow) hament
V41 (yellow) I / 0 (ON/OFF)
V35 (yellow) + 300 V intermediate circuit voltage

POLYMOBIL III Register 5 RXBEll5.0,6&3’ TD RX 1 Siemens AG


Rev. 01 Medical Engineering
Notes on trouble-shooting Chapter 4
Page 5 of 24

KV mAa

-I

D910 (X0381-91E)

Preparation ON
l In addition, VI3 (green) lights up on D910.

Trigger an exposure
l In addition, the following LEDs light up on D910:
V15 (green) exposure
VI8 (green) SS
V28 (yellow) radiation display

Error message
l V16 (red) lights up on D910.

Register 5 POLYMOBtL Ill


Siemens AG TDRXl
Medical Engfneermg
Chapter 4 Notes on trouble-shooting
Page 6 of 24

_1
I3920 (X0381-92D)

Checking the supply voltages


l When the system is switched on, relays HR and Kl on D920 pull in and supply
transformer 11, power supply M6 and the intermediate circuit with line voltage.
l Measure the 24 V supply voltage with the digital multimeter on D920 at MP I’+ 24 V” and
“OV”..
Measured voltage approx. 28.2 V Tolerance + 5 V

l Measure the filament circuit voltage supply on the electrolyte capacitor


C46 at +/ - on D920.

Measured voltage
in stand-by mode approx. 18.5 V Tolerance + 4 V
with preparation approx. 16.4 V Tolerance + 3 V

POLYMOBIL III TD RX 1 Siemens AG


Medical Engineering
Notes on trouble-shooting Chapter 4
Page 7 of 24

‘I,. - 7 - - - - .
I
I L ___---A

.I
j/ gl !Il
LP
KV mh

II91 0 (X0381-91E)

l Measure the supply voltages on D91O.X3.


+ 5 V at X3.4 and X3.1 (0 V) Tolerance + 0.5 V
+15 V at X3.5 and X3.6 (0 V) Tolerance + 1 V
- 15 V at X3.7 and X3.6 (0 V) Tolerance + 1 V

TD RX 1 Register 5 POLYMOBIL Ill


Siemens AG
Medical Engimermg
Chapter 4 Notes on trouble-shooting
Page 8 of 24
Checking the intermediate circuit voltage
On DIM an intermediate circuit voltage of .. . up to 350 V is
m applied!

During oscilloscope operation the protective ground wire connec-


tion in the power cable must not be interrupted under any circum-
stances.
For measurements where ground loops that may be present
could impair the measuring result, use the TEK amplifier and the
trigger attachment.

l Connect oscilloscope or digital multimeter to DI 11 .Rl and R2.

L Oszilloskop d
DVM
Dill (X0381-1116)
l Depending on the supply voltage, a value between 250... 350 V DC must be measured
with the digital multimeter.

Charging and discharging procedure

I
314 v
050 v :!is

t t
HR “’ t
relay relay :
ON OFF

POLYMOBIL III Register 5 RXBall5.o~yl~ TDRXI Siemens AG


Rev. 01 Medical Engineering
Notes on trouble-shootina ChaDter 4
Page 9 of 24
Checking the maximum main inverter frequency
. POLYMOBIL OFF
l Open bridge”S1” on D920 (solder).
l Connect oscilloscope on D910 to MP “REG” and “GND”.
. POLYMOBIL ON
l Trigger exposure.

- Err 15 is displayed.
- No radiation.

05.0 v : lops

F =lOkHz
TLr= 100 ps Tolerance: + 0 ps / - 5 ps

l Adjust the maximum main inverter frequency with potentiometer P2 to I391 0


(X0281-91E).
l POLYMOBIL OFF
l Connect jumper W ‘I on 0920 (solder).

Siemens AG TDRXl Register 5 POL’fMOBlL III


Medical Engineering
Chapter 4 Notes on trouble-shooting
Page 10 of 24
Oscillating current measurement
l POLYMOBIL OFF
l Connect oscilloscope on D920 to MP “IS”and “GND”.

Integrated current transformer.


- iV150A

. POLYMOBIL ON
l Trigger exposure with default values 60 kV, 10 mAs.

Main inverter ok
l During the kV rise
TeK Run: 1 .COMS/s Hi&s Trig?
-r r 1
L 1
I
I

l After the kV rise


Tsk Run: 1 .ook~S/s Hi Res Trig?
I:T : I
1

lVg5OA

/ I 1
.,us 41)omV

POLYMOBIL III Register 5 TD RX 1 Sjemens AG


Medical Engineering
Notes on trouble-shooting Chapter 4
Page 11 of 24
Error case
Only one diagonal is fired.
Tek stop: Single Seq SOP3

TD RX 1 Register 5 POLYMOBIL III


Siemens AG
Medical Engineering
Chapter 4 Notes on trouble-shooting
Page 12 of 24
Checking the high-voltage kVNoM and kVAcT
Connect oscilfoscope on D910 to MP “kVS” (1 V 2 20 kV nominal value) and “kV1’(1 V 2
10 kV actual value) and “GND” (X0381-6).
POLYMOBILON
Trigger an exposure with the default values 60 kV; 10 mAs.

1.00 v : 500 ms
2 20 kV

CH4: 02.0 v : 500 ms


1V 2 10kV

KV

PCLYMOBIL III Register 5 FM?; 15.061.02 TDRXl S&ens AG


10.95 Medical Engineering
Notes on trouble-shooting Chapter 4
Page 13 of 24
Checking the filament and tube current
Filament current
l Connect oscilloscope on D910 to MP “IHS” (NOMINAL), “IH” (ACTUAL) , “11’
(ACTUAL)
And “GND”.
. POLYMOBIL ON
After approx. 7 seconds the stand-by filament current comes on.
l Trigger an exposure with the default values 60 kV, 10 mAs.

TaK Stop: Siwie Sea 25.0 S/s

IHS 2 ..,. I ,,,,

IH

POLYMOBIL ON+t 1
Nominal
value output
Stand by
filament circuit
f
C. ON

Lb ON
t
+
e / b OFF

Slemens AG TDRXl Register 5 POLYMOBIL III


Medical Engineering
Chapter 4 Notes on trouble-shooting
Page 14 of 24

TeK Stop:SingleSea 1005/s

IHS
I I II Ill III I

TeKStop:Single~q 2.SOMWs

Stand by filament
current
I

T I
/
G-
L _
Exposure filament
I ‘L L! current
bl LO.OYS U-l‘J f 2.40”
ET1 2*30v z.oc v 20 .OMS

POLYMCBIL III Register 5 ;~~~15.0$.0~ TDRXl Siemens AG


” Medical Engineering
Notes on trouble-shooting Chapter 4
Page 15 of 24
Setting the maximum filament frequency
. POLYMOBIL OFF
l Remove fuse F5 on D920.
l Connect oscilloscope on D910 to MP “CAL” and “GND”.
. POLYMOBIL ON

jl Err 3 is displayed.

l Using P3, set the maximum filament frequency on D910 (100 kHz 1 10 ps)

CH2:

Siemens AG TD RX 1 Register 5 POLYMOBIL III


Medical Engmeering

,
Chapter 4 Notes on trouble-shooting
Page 16 of 24
Checking the tube current
l Connect oscilloscope on D910 to MP WS” (NOMINAL), “IR” (ACTUAL) and GND”.
POLYMOBIL ON
A&4 l
0 Trigger an exposure with the default values 60 kV, 10 mAs.

: 500 ms

: 500 ms

Register 5 TDRXl Siemens AG


POLYMOBIL Ill Medical Engineering
Notes on trouble-shooting Chapter 4
Page 17 of 24
Checking the kV and tube current (IR)
l Connect oscilloscope on D910 to MP “kV”, W ’ and “GND”.
l POLYMOBIL ON
l Trigger exposure with the default values 60 kV, 10 mAs.

kV CH2: 1 .OO V : 50 ms
CH3: 1.00 V : 50 ms

IR

kV CH2: 1.00 V : 5 ms
CH3: 1.00 V : 5 ms

Prolonged transient response

) NOTICE 1 The transient response of the tube current can be set with Pl on
- D910 .

Siemens AG TD RX 1 Register 5 POLYMOBIL III


Medical Engineering
Chapter 4 Notes on trouble-shooting
Page 18 of 24

Fig. 2

Checking the mAs values


l Remove the 4 Allen screws ( 3 mm key size) from the lid of the single tank and take off
the lid (Fig. 2).
l Remove jumper “mA+ / mA - ” in the single tank.
l Connect the mAs meter to the “mA + / mA -” sockets (X0381-5/5).
l Trigger the following exposures:

I /
Setting at control panel ) valid mAs values
I 40 kV, 200 mAs I 195 . . . 205 mAs I
66 kV, 100 mAs 97 .. . 103 mAs
90 kV, 1 mAs 1 ,O . . . 1,3 mAs

l Remove the mAs meter and reinsert the jumper in the single tank at “mA+ / mA -“.
l Reinstall the lid of the single tank.

POLYMOBIL III Register 5 RXBB-115061.02 TD RX 1 Siemns AG


Rev. 01 10.95 Medical Engineering
Notes on trouble-shooting Chapter 4
Page 19 of 24
Adjusting the mAs
Unplug plug X2 on D910.
Establish connection (jumper) from plug contact X2.5 (side of the board) toMP “P5 VR”
on D910.
Establish connection (jumper) from plug contact X2.6 (side of the board) to MP llGND1’
on D910.
Connect oscilloscope on D910 to MP “Fl” and “GND” (1 V c 3,2 kHz).
POLYMOBIL ON

1 NOTICE 1 Err 6 is displayed.

1 V 2 3.2 kHz

.CH2: 02.0 V : 10 p.s

5V e 16kHz

I I t
! ! 1
j-- 62-5 P ---+

l Adjustment with potentiometer P7 on D9lO (adjust to 16 kHz)

Siemens AG TD RX 1 Register 5 POLYMOBIL III


Medcal Engineerrng
Chapter 4 Notes on trouble-shootina
Page 20 of 24

Length

Marking

Side marking

Fig. 3

Aligning the light field to the radiation field


Operating sequence:
l Load a 24 cm x 30 cm or 10”~ 12”cassette with film and place it on a table or a similar
base.
l Using a tape measure, set a vertical SID of 100 cm or 40”to the upper edge of the
cassette.
l Using the control buttons, set a format of 18 cm x 24 cm or 8”~ IO.
l Switch on the light localizer and align the cassette.
l Attach radiopaque markings (e.g. washers, coins) to the cassette as shown in Fig. 3.
Attach a washer as lateral marking.
l Trigger an exposure (60 kV, 1OmAs) and develop the film.
. Using a waterproof felt pen, write the following data on the developed fitm.
- Set SID
A Film size
- Radiation field size

POLYMOBIL Ill Register 5 RXBB-115.0,60’.;&2 TDRXl Siemens AG


Rev. 01 Medical Engineering
Notes on trouble-shooting Chapter 4
Page 21 of 24

- Film

- Light field

- Radiation field

I y2
Fig. 4

Evaluation:
l Measure the deviations between light field edges and radiation field edges on all four
sides (XI ,X2,YI, Y2) as shown in Fig. 4.
l Determine the total deviations in the X and Y direction (ignore the +/- signs).
l Both the length deviation (Z Y) and the width deviation (C X) must be less than 1.6 cm.

Siemens AG TD RX 1 Register 5 POLYMOBIL III


Medcal Engineering
Chapter 4 Notes on trouble-shooting
Page 22 of 24

Fig. 5 Fig. 6

l If the deviation is higher, loosen the 4 Allen screws slightly (arrows/Fig. S/Fig. 6) and
move the collimator accordingly.
Then, tighten the screws at the collimator (arrows/Fig. 5/Fig. 6) again.
l Repeat the check and, if necessary, adjust the collimator again until the deviation
between the light field and the radiation field is within the admissible tolerance
(<1,6cm).

POLYMOBIL III Register 5 RXBB-115061.02 TDRXl Siemens AG


Rev. 01 10.95 Medical Engineering
Notes on trouble-shooting Chapter 4
Page 23 of 24

Fig. 7 Fig. 8

Checking and readjusting the counterweight


l Unlock the support arm (l/Fig. 7) and release the turning knob (2/Fig. 7).
Without any accessories attached, the support arm should be easy to move across the
entire movement range and stop in any position.

Readjusting the countennreight with accessories attached


If the support arm does not stop in any position with accessories attached, readjust the
spring tension for the counterweight.
l Take the cover (S/Fig. 7) off the stand.
l Completely release the additional brake on the support arm (2IFig. 7).
l Move the support arm into the horizontal position.
l Tighten the screw by means of a 17mm fork wrench (4/Fig. 8).
Adjust the spring tension so that the support arm is balanced in the horizontal position,
i.e. the forces needed for lifting and iowering the tube unit are equal.

Siemens AG TD RX 1 ;;$-;15.061.02 Register 5 POLYMOBIL III


Medical Engineering 10.95
Chapter 4 Notes on trouble-shooting
Page 24 of 24

06
Fig. 9 Fig. 10

Checking the control elements on the collimator


l Check the light-beam localizer operated both at the control panel and at the collimator.
To do so, press the q button (G/Fig. 10 or 7/Fig. 9),
the light-beam localizer lamp lights up for 20 s f 1 s.
l Check if the double-slot diaphragm can be rotated * 45”.
l Check that format adjustment (blades) is easy and smooth by turning the collimator
adjustment knobs (8/Fig. 9).

POLYMOBIL Ill Register 5 RXB&ll5.0,6$ TDRXl Siemeos AG


Rev. 01 Medical Engineering
Replacing important components Chapter 5
Page 1 of 6

Fig. 1

Replacing the single tank


To replace the single tank, remove it from the mobile stand as follows:
l Move the arm system in the lowest position and check if the safety locking (pin l/Fig. 1)
has engaged.

The stand cannot be moved upwards any longer.


l Tighten the turning knob(3/Fig. 1).
l Loosen the threaded ring and pull out the connector of the single tank.
l Using an 8 mm Allen key, loosen the two screws on the bracket (2/Fig. 1) approx. two
turns.
l Lift the single tank off vertically and place it on a soft base.
l Detach the collimator (see chapter 5, page 2).
l Install the new single tank in the reverse order.
l Check alignment of light field to radiation field and adjust if necessary. (see “Aligning the
light field to the radiation field”).

TD RX 1 Register 5 POLYMOBIL Ill


Siemens AG
Medical Engmeering
Chapter 5 Replacing important components
Page 2 of 6

Fig. 2 Fig. 3

Replacing the collimator


In the case of damage, the collimator has to be replaced completely.
Proceed as described in the following:
l Loosen two screws @ /Fig. 3) and remove the rear panel.
l Mark the 5 connecting cables of the collimator and disconnect them.
l Remove the two screws on both sides (G/Fig. 3) of the upper cover.
l Turn the two knobs (8/Fig. 4) in their final position and remove them.
l Remove the front panel after having loosened the two screws (7/Fig. 4).
Caution: Front panel is connected to collimator via cables (g/Fig. 5).
l Remove the collimator after loosening the four Allen screws (arrows /Fig. 2/Fig. 3).

(/] Hold the collimator.


Lift upper cover and pull out connecting cable through the case.

l Attach the new collimator, fasten it with the four screws (arrows /Fig. 2/Fig. 3) and center
it.
l Connect the connecting cables and screw down the rear panel.
l Check agreement of light and radiation field and make adjustment if required
(see “Agreement of light and radiation field”).

POLYMOBIL Ill Register 5 RXBB-11 S.C$UJ TDRX 1 Siemens AG


Rev.01 Medical Engineering
Replacing important components Chapter 5
Page 3 of 6

Fig. 4 Fig. 5

Replacing the light-beam localizer lamp


l The replacement light-beam localizer lamp is located below the cassette tray.
l Turn both knobs (8/Fig. 4) to an end position and remove them.
l Remove front panel after loosening both screws (7/Fig. 4) .
Caution: The frontpanel is connected to the collimator via cables (S/Fig. 5).
l Loosen both screws (1 O/Fig. 5) and unscrew the bracket carrier of the temperature
switch).
l Take out the defective lamp (1 l/Fig. 5) and replace it by a new lamp.

(II Do not touch the glass envelope with your bare fingers.

l Screw down the bracket and the front panel.


l Adjust both knobs and fasten them.

Siemens AG TD RX 1 Register 5 POLYMOBIL 111


Medical Engineering
Chapter 5 Replacing important components
Page 4 of 6

Fig. 6

The light-beam localizer lamp voltage can be adjusted at resis-


tance R6 (S/Fig. 6).
The voltage at the lamp should be II .5 V 3- 0.2 V.

l Check function of the light-beam localizer lamp.

Register 5 Fg$l5.960’ &L? TDRXI Siamens AG


POLYMOBIL II\ Medical Engineering
Replacing important components Chapter 5
Page 5 of 6

Fig. 7

Replacing the caster


Herefore:
1 Only loosen the Allen screws at the right and left (key width 2.5 mm)!
2 Move hexagonal shaft backwards
3 Remove pedal
4 Remove allen screws (key size 6 mm)
5 Remove caster.
6 Bring the threaded holes of the new caster into position
l Install the new caster in the reverse order.
l Check pedal positions (see operating instructions)

TD RX 61 Friedrich
TD RX 1 / Kern FL
TDU 3 / Hay

Siemens AG TD RX 1 RX&115061.02 Register 5 POLYMOBIL III


Medical Engineering Rev. 01 10.95
Chapter 5 Replacing important components
Page 6 of 6
This page intentionally left blank.

Register 5 ~~m~ymo~ TO RX 1 Siemens AG


POLYMOBIL III Medical Engineering