Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Info Codes
Info 1 Compensator out of range
Cause: The instrument is tilted too much. The dual-axis
compensator can not compensate for the inclination.
Action: Leve the instrument or disconnect the dual-axis
compensator.
Info 2 Wrong face
Cause: The operation was carried out while the instrument was
in an illegal mode. E.g: Trying to measure in the wrong
face.
Action: Change to face 1, showing angles in the display and
retry.
Info 3 Distance already recorded
Cause: The distance to the current object has already been
registered.
Action: If a new registration is required a new measurement
must be carried out.
Info 4 - Invalid measurement
Cause: - The measurement is invalid, e.g. several
measurements towards the same point or the measured
points lies 200 grads from each other, P20, Free Station.
- Trying to perform a calculation which is dependent from
a distance without having measured any distance, P20
Free Station and Z/IZ.
Action: - Check that the circumstances above does not occur
and redo the measurement.
Info 5 - Undefined mode or table
Cause: Tries to use a display or output-table that does not
exist.
Action: Choose another table or create a new.
Info 6 - Vertical angle less than 15grads from horizontal angle
Cause: The vertical angle is less than 15grads from the
horizontal angle when performing a Tilt Axis Calibration.
Action: Redo the calibration with an increased horizontal angle.