FOR CT SCAN

Rating
Voltage 380 to 480 VAC (see note, below)
Capacity 90 KVA @ 85% P.F.
Frequency 50 or 60 Hz +/- 1.0 Hz

Grounding
The ground to the CT system shall originate at the system power source, (i.e., transformer or first
access point of power into a facility) and be continuous to the CT system power disconnect in the
room. A dedicated 1/0 (55mm2) or larger insulated copper ground wire must be run with the
phase wires from the main distribution panel to the PDU. These grounds can be spliced with
“High Compression Fittings” and should be terminated at each distribution panel it passes through.
When the ground is broken for a connection to a panel, it shall be connected into an approved no
insulated
grounding block with the incoming and outgoing ground in this same grounding block,
which is then connected to the steel panel, never using the steel or other material of the panel as
the block.
The resistance between the PDU ground and the facility earth ground must not exceed 0.5
ohm. In addition, the total resistance between the PDU ground and earth must not exceed 2
ohms. Resistance between any two grounded devices must not exceed 0.1 Ohm to ensure equal
potential ground system within the
CT Room.

Mains Supply/
online UPS GANTRY
380 to 480 VAC
90KVA

TABLE

3 pole
MCCB 125 Amps
35 square mm
Copper wire

PDU

35 square mm
Copper wire

CT SCAN
FOR X-RAY DX -525

Power Requirements

Parameters For Three Phase Power Supply


Voltage Range Manual ( 415 Volts +10% /-15%)
Frequency 50 Hz
Maximum Momentary Power Demand 50 KVA
Long Time power Demand 4 KVA
Load Power Factor 95%
Max.line Resistance Allowed For Full Rating 0.2 Ohms
With Reduced Rating 0.50 Ohms
Main Switch And Fuse Rating 100 Amps
Mains Cable 35 Square mm

FOR MAMMOMAT 1000


NOTE :- ALL ABOVE EQUIPMENTS MUST HAVE PROPER SEPARATE GROUNDS.