Voltage 380 to 480 VAC (see note, below)
Capacity 90 KVA @ 85% P.F.
Frequency 50 or 60 Hz +/- 1.0 Hz
Grounding
The ground to the CT system shall originate at the system power source, (i.e., transformer or first
access point of power into a facility) and be continuous to the CT system power disconnect in the
room. A dedicated 1/0 (55mm2) or larger insulated copper ground wire must be run with the
phase wires from the main distribution panel to the PDU. These grounds can be spliced with
“High Compression Fittings” and should be terminated at each distribution panel it passes through.
When the ground is broken for a connection to a panel, it shall be connected into an approved no
insulated
grounding block with the incoming and outgoing ground in this same grounding block,
which is then connected to the steel panel, never using the steel or other material of the panel as
the block.
The resistance between the PDU ground and the facility earth ground must not exceed 0.5
ohm. In addition, the total resistance between the PDU ground and earth must not exceed 2
ohms. Resistance between any two grounded devices must not exceed 0.1 Ohm to ensure equal
potential ground system within the
CT Room.
Mains Supply/
online UPS GANTRY
380 to 480 VAC
90KVA
TABLE
3 pole
MCCB 125 Amps
35 square mm
Copper wire
PDU
35 square mm
Copper wire
CT SCAN
FOR X-RAY DX -525
Power Requirements