WL 373
123.4 °C
123.4 °C Wärmeleitung von Gasen
und Flüssigkeiten
Thermal Conductivity of
Gases and Fluids
Aus / Off
123
Heizer / Heater Leistung / Power
Leistung / Power Watt
Experiment Instructions
i
02/2012
Table of Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Structure of the Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Unit Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
4 Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5 Experiments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.1 Calibrating the Heat Exchanger. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.1.1 Preparing for the Experiment . . . . . . . . . . . . . . . . . . . . . . . . 10
5.1.2 Performing the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.1.3 Evaluating the Experiment: Calibration Curve . . . . . . . . . . . 11
5.2 Determination of Coefficients of Thermal Conduction . . . . . . . . . . . . 14
5.2.1 Preparing for the Experiment . . . . . . . . . . . . . . . . . . . . . . . . 14
5.2.2 Performing the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5.2.3 Evaluating the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.1 Working Sheet: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.2 Symbols and Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.3 Table with Coefficients of Thermal Conduction . . . . . . . . . . . . . . . . . 20
ii
02/2012
iii
02/2012
1 Introduction
1 Introduction 1
02/2012
2 Safety
Symbol Explanation
Electrical Voltage
Hot Surfaces
Notice
2 Safety 2
02/2012
WARNING
Reaching into the open measurement and
control unit can result in electric shocks.
• Disconnect from the mains supply before open-
ing.
• Work should only be performed by qualified
electricians.
• Protect the measurement and control unit
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
against moisture.
WARNING
Risk of burns.
The knurled bolt and the heat exchanger are very
hot.
• Leave the unit cool down.
2 Safety 3
02/2012
3 Unit Description
1 2 3 4
WL 373
123.4 °C
123.4 °C W ärmeleitung von Gasen
und Flüssigkeiten
Thermal Conductivity of
Gases and Fluids
Aus / Off
123
Heizer / Heater Leistung / Power
Leistung / Power Watt
2 Base
3 Unit Description 4
02/2012
3.2 Function
Cooling water
2 4 1 3 connection
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
Cooling water
connection
3 Unit Description 5
02/2012
NOTICE
Ensure that the sealing ring is correctly seated!
Fig. 3.3 Dismantling the
Heat exchanger
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
Aus / Off
123 max. 140...150W using the potentiator (4).
Heizer / Heater Leistung / Power
Leistung / Power Watt
The heater can be switched on and off using the
power switch (5).
5 4 3
3 Unit Description 6
02/2012
NOTICE
Do not reverse the cables for the thermocouples!
3 Unit Description 7
02/2012
4 Principles
Q = A
T1 ----------- (4.1)
·
Q
T2 ·
This yields the relationship for the heat flow Q :
·
Q
Q =
· A T-
----------------------- (4.2)
with
·
Q Head flow through the wall under
investigation
Fig. 4.1 Thermal conduction in a plane A Wall surface area
wall
T Temperature gradient in the wall
Wall thickness
Coefficient of the thermal conduction
is temperature dependant; in general the mean
temperature in the wall is used.
(in W/Km) can be thought of as the energy that
is conducted every second through a 1m thick
wall over an area of 1m2 when there is a temper-
ature difference of 1K between the opposite sides
of the wall.
4 Principles 8
02/2012
5 Experiments
5 Experiments 9
02/2012
·
V = V
---- (5.1)
Fig. 5.1 Shut off valves t
V Capacity water [ltr]
t Time [s]
5 Experiments 10
02/2012
2
A = 0,0396m 0,126m = 0,0157m (5.2)
Calibration values are marked with .
= 0,0242 1 + 0,003 T (5.3)
where
T - Mean air temperature in °C
T1 + T2
T = ------------------
- (5.4)
2
T1 Temperature heater
T2 Temperature cooling water
5 Experiments 11
02/2012
QA = A T-
·
------------------------ (5.5)
where
Temperature difference
T = T1 – T2 (5.6)
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
· ·
QB = P–QA (5.7)
·
P T1 T2 T Air Q B
in W in °C in °C in °C in W/Km in W
5 Experiments 12
02/2012
·
Q B in W
70
60
50
40
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
30
20
10
0
0 10 20 30 40 50 60 70
T in °C
5 Experiments 13
02/2012
5 Experiments 14
02/2012
· ·
QA = P – QB (5.9)
· ·
P T1 T2 T QB QA Water Literature Var.
in W in °C in °C in °C in W in W in W/Km in W/Km in %
Tab. 5.2 Example measurement results when using water as the measured medium
5 Experiments 15
02/2012
· ·
P T1 T2 T QB QA Air Literature Var.
in W in °C in °C in °C in W in W in W/Km in W/Km in %
Tab. 5.3 Example measurement results when using oil as the measured medium
5 Experiments 16
02/2012
6 Appendix
· ·
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
P T1 T2 T QB QA Medium Literature
in W in °C in °C in °C in W in W in W/Km in W/Km
· ·
P T1 T2 T QB QA Medium Literature
in W in °C in °C in °C in W in W in W/Km in W/Km
6 Appendix 17
02/2012
·
Q B in W
70
60
50
40
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
30
20
10
0
0 10 20 30 40 50 60 70
T in °C
6 Appendix 18
02/2012
6 Appendix 19
02/2012
6 Appendix 20
02/2012
Dimensions:
Base plate:
(L x W x H) 250 x 400 x 150 mm
Measurement amplifier:
(W x H x D) 350 x 130 x 250 mm
Weight: approx. 8 kg
Heat exchanger:
Effec. transfer area 0,0157 m2
Mean Effec. Diameter 39,6 mm
Slot width 0,4 mm
Heater
Max. Power Consumption 160 W
Thermocouples 2 x Type K
6 Appendix 21
02/2012
6.5 Index
C
Calibration curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Coefficient of thermal conduction . . . . . . . . . . . . . . . . . . . . . . . 1, 8, 9, 20
Convection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cooling water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
F
Fourier law . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
H
All rights reserved, G.U.N.T. Gerätebau, Barsbüttel, Germany 02/2012
Heat sink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Heat source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Heater power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
I
Impurities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
M
Measurement and control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
P
Plane wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Power consumption, heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
R
Rapid action hose coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
S
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Thermal conduction in a plane wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Thermocouple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Topics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
22