- Contents -
1. Purpose
5. Circuit Diagram
- 1 -
HSQI-G0501
1. Purpose
This product is a medical equipment which is steam sterilizer with high pressure.
2.1 Overview
This medical product is a high-pressure steam sterilizer that sterilize the medical
equipments and surgical instruments in hospital and in the clinic with high pressure.
Front
Side Back
2) Exterior Description
- 2 -
HSQI-G0501
- 3 -
HSQI-G0501
- 4 -
HSQI-G0501
500
670 720
2.4 Product Configuration & features
- 5 -
HSQI-G0501
2.4.4 Protection Type & Degree for Electronic Shock: A level device, B type device.
3.1 Performance
1) Rated Voltage and Frequency : AC 220V, 50/60Hz
2) Power Consumption : 2.9Kw
3) Capacity of the Chamber : 50L
4) Sterilizing Pressure : 1.0 kg/㎠ and 2.0 kg/㎠ (+20%)
5) Sterilizing Temperature : 121℃ +4℃ and 132℃ +5℃
6) Protection Type & Degree for Electronic Shock: A level device, B type device
4.1 Preparation
1) Prepare the items for sterilizing for each method of packing & setting.
2) Check the water level in the water Tank.
3) During the sterilization, both items for sterilizing and biological indicator or chemical
indicator should be in the chamber for ensuring well process of sterilization.
- 6 -
HSQI-G0501
5) During the process, user want to stop the process or power off the device, should
press the <Start/Stop> button.
6) Once user press the <Start/Stop> button, vacuum condition is starting. Vacuum
condition of 400-560mmhg is set, water will be supplied to the chamber thru pipeline
of water Tank. Each process state will display on LCD display. Thru control device
by a certain amount of water enters to the chamber. Water valve will be closed and
water supply process is finished.
7) After completing water supply, heating process is started. LCD display, the
sterilization temperature, pressure, and the sterilization time is displayed, display the
current temperature is also displayed and the additional flashes.
8) When the setting temperature is reached, heating process is completed and the
sterilization process is continued. LCD display flashes the sterilization time and the
display unit is seconds.
9) After the sterilization process, the exhausting process will start. LCD display flashes
in the exhaust process, the current temperature and the pressure will be displayed.
Solenoid exhaust valve opens, the steam and residue of water in the chamber will go
to water storage. When LCD display shows 0.0kg/㎠, the exhaust process is
completed.
10) After the exhausting process, drying process is started. Once LCD display shows
0.1kg/㎠, the vacuum pump starts to operate which will dry the moisture of the
items.
11) When the process is completed, the inflow of external air allows the vacuum
condition is turned off and alarm will be announced 3 to 5 seconds for the
completion. Then, the door can be opened.
12) All process is completed, END will display in the LED display. LCD display will show
the applied program.
13) Program Mode Set
Ready Display →
① PACKS SET
- 7 -
HSQI-G0501
Enter →
Temp Up & Down = Enter → Pres Up & Down = Enter →
Menu → END
② INSTRUMENT SET
Up & Down Enter → Temp Up & Down = Enter → Pres Up & Down = Enter →
- 8 -
HSQI-G0501
③ GLOVE SET
Up & Down Enter → Temp Up & Down = Enter → Pres Up & Down = Enter →
④ LIQUIDS SET
Up & Down Enter → Temp Up & Down = Enter → Pres Up & Down = Enter →
- 9 -
HSQI-G0501
⑤ BOWEI-DICK SET
Up & Down Enter → Temp Up & Down = Enter → Pres Up & Down = Enter →
⑥ RE-DRY SET
Up & Down Enter → Temp Up & Down = Enter → Pres Up & Down = Enter →
- 10 -
HSQI-G0501
Up & Down Enter → Packs Key Up & Down Enter → Inst Key Up & Down = Enter →
Glove Key Up & Down Enter → Liquids Key Up & Down Enter → Bowei-D Key Up & Down = Enter →
⑧ PRINTER INTERVALS
⑨ BUZZER VALUME
- 11 -
HSQI-G0501
⑩ LANGUAGE SELECT
Ready Display →
① SYSTEM SETUP
② CHECK COUNT
- 12 -
HSQI-G0501
③ VACUUM SETUP
④ CLOCK SETUP
⑤ LEAKAGE TEST
- 13 -
HSQI-G0501
4.4 Precautions
※ This product maintains the very high temperatures, so that users always try to
secure and must use the procedures listed in the manual.
1) This product uses the AC 220V power cord only.
2) The instructions when you open and close the door pressure gauge (0kg / ㎠) to
check the status.
3) When user load the items in the chamber, items must be dry enough to be loaded
and items have moisture or waterdrop should not be loaded.
4) When user load the items in the chamber, user should load the items from side to
side rather than below the top.
5) When power off state, the door do not allow to open so that user should put power
on when user want to open the door.
6) Users should wear protective gloves due to high temperature during the process and
also after the process for protecting from burns.
7) Sterilizer's inspection, adjustment and maintenance should be conducted safely by
the manufacturer or its representative (or company) for ensuring the their educational
qualifications and experience.
8) Do not disassemble or modify the device.
9) In case of device failure occurs and there is no qualified technician, the device
should be attached Error Display and send to the manufacturer for the service.
10) For the maintenance, except the power cord in advance, then user should wait to be
cooled to ambient temperature.
11) Only someone who have responsible of this device or trained person can use this
sterilizer.
12) When user sterilize pack items, outside of pack item should avoid contact with the
walls of chamber.
13) Should not sterilize explosive, volatile, flammable or toxic substances.
- 14 -
HSQI-G0501
① Around the sterilizer, the explosive, volatile, flammable or toxic substances should
not be loaded.
14) When user use the liquids for sterilization, user should keep following points.
① Use ventilated cover. Screw cap, rubber stopper will not be used.
② Use only Borosilicate glass bottle. Do not use normal glass bottle or container
which are not made for the sterilization.
③ When the process is completed, do not open the door suddenly.
Open the door a little over 10 minutes to wait.
④ Do not shake or move the heated bottle. (Bottle could be exploded.)
15) Do not cook the food (such as chestnut, sweet potato, meat and so on) with
sterilizer.
16) After using external and internal sterilizer should always clean.
17) This device should not be installed in outdoor temperature changes are severe. (Error
occurs will be the cause of).
18) Avoid direct sunlight and do not install with the devices cause high heat.
19) This device installed in the ventilated interior, and should be installed away from the
patients and passengers.
20) Sterilizer must be grounded, and the power switch off when not in use should be.
21) Operating environment conditions: relative humidity range (30 ~ 75%), the ambient
temperature range (10 ~ 40 ℃), pressure range (70 ~ 106 ㎪)
22) When using sterilizer, user should use distilled water.
(General water, groundwater usage may cause the corrosion and scale.)
- 15 -
HSQI-G0501
Error Action
Error Case # 1 (Water level)
Supply water in the water Tank.
Water Supply
Error Case # 2
Door open during the process →
Contact A/S Center of Manufacturer.
Check the door lock &
check the switch.
Error Case # 3
Overheating →
Contact A/S Center of Manufacturer.
over 138℃
Check the temperature sensor.
Error Case # 4
Lack of water supply →
Contact A/S Center of Manufacturer.
provide water over 5mins.
Check the filter.
Error Case # 5
Heating timeout →
Contact A/S Center of Manufacturer.
over 60 mins,
check the heater
Error Case # 6
Pressure sensor error →
Contact A/S Center of Manufacturer.
Check the pressure sensor.
Check the heater.
Error Case # 7
Temperature sensor error →
Contact A/S Center of Manufacturer.
overheating detection
Check the temperature sensor
Error Case # 8
Overpressure →
over 2.8Kg/㎠ Contact A/S Center of Manufacturer.
Check the temperature,
pressure sensor
Error Case # 9
Sterilization temperature range →
Setting temperature+/-4℃ Contact A/S Center of Manufacturer.
Check the heating temperature
sensor.
Error Case #10
Setting temperature, time is over.
Contact A/S Center of Manufacturer.
Check the heater.
Check the water level for 1time.
- 16 -
HSQI-G0501
5. Circuit Diagram
- 17 -