Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Nalco Company
1601 W. Diehl Road
Naperville, IL 60563
Trademarks
PRISM, Vantage, 3D TRASAR, STA-BR-EX, ACTI-BROM, Nalco and the logo are Registered
Trademarks of Nalco Company.
Table of Contents
Section 1: Introduction
1.1 About the 3D TRASAR® 5500 ................................................................................ 1-1
1.2 About this Manual ................................................................................................. 1-1
1.3 3D TRASAR 3500 and 3D TRASAR 5500 Models and Optional Components .. 1-1
1.4 3D TRASAR 3500 and 3D TRASAR 5500 Overview ............................................ 1-3
1.5 How the 3D TRASAR 5500 System Works .......................................................... 1-4
1.5.1 Fluorometer ............................................................................................................... 1-4
1.5.2 3D TRASAR Scale Control, 3D TRASAR Bio-Control, & NCM Corrosion Technology 1-4
1.5.3 Data Storage / Reporting and Communications ......................................................... 1-5
1.6 Safety ..................................................................................................................... 1-6
Section 8: Maintenance
8.1 Maintenance Schedule ......................................................................................... 8-1
8.2 Clean Fluorometer Flow Cell ................................................................................ 8-2
8.3 Check/replace Fluorometer Desiccant ................................................................ 8-2
8.4 Check Sensor Calibration .................................................................................... 8-2
8.5 Probe Replacement Schedule ............................................................................. 8-3
8.6 Cleaning Basket / Wye Strainer ............................................................................ 8-3
8.7 Pellet Feeder Maintenance ................................................................................... 8-3
8.8 Check / Refill Pellet Feeder ................................................................................... 8-3
Section 9: Troubleshooting
9.1 General Troubleshooting ...................................................................................... 9-1
9.2 Alarm Screen Troubleshooting ............................................................................ 9-7
Appendix A: Specifications
Appendix B: Drawings
Appendix C: Replacement Parts/Accessories
Appendix D: Air Purging System
Appendix E: SCADA and LAN
Appendix F: 3D TRASAR Starter Upgrade Procedures
Appendix G: 3D TRASAR Technology for Closed Loop Applications
Appendix H: 3D TRASAR Technology for High Fouling Applications
Appendix I: Halogen Analyzer Control
Appendix J: 3D TRASAR Models for the Pacific
Appendix K: Warranty Policy
Glossary
Section 1: Introduction
1.1 About the TRASAR® 3500 and 5500
The Nalco 3D TRASAR ® 5500 is the most advanced water treatment control system on the market
today. The design is based on years of field research and a broad base of industry experience and
application. The introduction and use of the 3D TRASAR 5500 will dramatically change the standard
for cost-efficient and reliable cooling tower control.
The Nalco 3D TRASAR Starter 3500 makes 3D TRASAR technology more accessible. It is customizable
to meet specific user needs, yet offers the flexibility for future upgrades to utilize all features of 3D
TRASAR.
Both systems are simple to use and feature powerful control algorithms that ensure tighter control,
more optimal performance, and requires a minimum of human oversight and testing. The end result is
a better, safer, more reliable system that runs more efficiently.
Keep in mind, as you begin, that your Nalco Technical Representative has already done some of the
preliminary configuration work for you. Don’t hesitate to contact that person if you have a question or
problem, as they are familiar with your situation and can be very helpful in addressing any specific
needs you may have.
(3D Sc
- Control) a le Con
io trol)
(3D B
Blowdown
Biocides Inhibitor I Inhibitor II Acid Pump
Valve
The 3D TRASAR 5500 utilizes state-of-the-art optical fluorescence sensing to monitor and control
chemical feed. The central component of the system is a rugged 6-channel fluorometer that mea-
sures:
These measured variables are used in control algorithms that anticipate and respond to system
stresses before costly scaling and fouling can occur.
The 3D TRASAR Starter 3500 utilizes proven TRASAR technology to monitor and control chemical
feed. Features of the 3D TRASAR 5500 are available as upgrades to the 3D TRASAR Starter
3500.
1.5.1 Fluorometer
TRASAR chemistries contain a uniquely fluorescing molecule that allows the fluorometer to detect
exactly how much chemical has been fed in order to determine chemical residual. This information
is used to deliver accurate control of chemical levels in your cooling system.
The patented 3D TRASAR on-line fluorometer consists of a “Flow Cell”, an LED light source, filters,
and detectors. The filters and detectors are tuned to specific light frequencies unique to Nalco
products and resultant process reactions. The LED light source is extremely bright at the appropri-
ate frequencies to reliably and repeatably measure the chemical properties of the flowing water . As
a solid-state device, the fluorometer offers trouble free long life.
Water Sample Out
Detector
Excitation
Filter
Emission
Filter
Flow
LED Cell
Light
Source
Water
Water Sample In
Typical cooling water controllers use simple On/Off control schemes and timers to maintain pre-
determined set points in hopes to avoid operational upsets. The problem with this approach is that,
as system stresses change, scale can form, bio-populations can multiply, and corrosion can occur
without ever being detected. Pre-determined set points are only good for one set of stress condi-
tions!!
The 3D TRASAR 5500 Controller takes a revolutionary proactive approach to managing cooling
water systems. Patented state of the art technology is used to prevent scale formation, control bio-
population levels and continuously monitor corrosion rates.
3D TRASAR Scale Control prevents scale formation by continuously monitoring the activity of an
innovative “tagged” high stress dispersant polymer. As scale forming particles increase in the
cooling system, the active polymer is consumed. By monitoring this consumption rate and the active
polymer level, 3D TRASAR Scale Control can take corrective action well before the conditions
needed to produce scale can occur.
NCM Corrosion Monitoring Technology is also incorporated in the 3D TRASAR 5500. This
patented technology automatically recognizes the corrosion rates of up to 2 different metallurgies.
The available metallurgies include mild steel, galvanized steel, copper
, brass, CuNi 90/10, CuNi 70/
30, and aluminum. By monitoring corrosion rates continuously (vs. 90-day snapshots) corrective
action can be taken before the problem becomes serious.
1.6 Safety
Always follow the safety practices listed below:
• Never open the 3D TRASAR controller box or junction box with electrical power engaged.
• Never open the pellet feeder hopper with electrical power engaged.
• Maximum water pressure to the 3D TRASAR skid and pellet feeder should never exceed
90 psi.
• Always wear the appropriate Personal Protective Equipment (PPE) when working on a 3D
TRASAR system (i.e. gloves, protective eyewear, protective shoes, wearing a mask, etc).
• Always observe local and facility safety practices beyond those listed in this manual.
WARNING: Always turn off power before making any electrical connections, which in-
! cludes all interconnecting cables, otherwise permanent damage may occur to system
components.
NOTE: Both the 3D TRASAR controller and the Bio-reporter Pellet Feeder operate well
! below an 85 db noise level.
Introduction
There are 9 steps to installing and starting-up a 3D TRASAR system.
Install Main Control Panel in well-lit, dry Install Bio-Reporter Feeder in well-lit, dry
Unpack all parts and identify them. location convenient to points of use.
location convenient to points of use. Mount
at eye level.
Section Union
0.5
Section Section
2.4
GPM
2.5 2.6
ORP Probe
Plumb utilizing ball valves provided. Inlet Connect discharge back to recirculating Install probes and attach connectors.
water from recirculating system must be at system downstream from supply in basin
5 GPM or greater. where possible.
Details Details
Section Section Details
2.7 2.8 Section
2.9
3D TRASAR 3D TRASAR
Description
Starter Model Model
060-TR3500.88 060-TR5500.88 Wall Mount, S mall Backboard
060-TR3501.88 060-TR5501.88 Frame Mount, Small Backboard
060-TR3510.88 060-TR5510.88 Wall Mount wi th Junction Box
060-TR3511.88 060-TR5511.88 Frame Mount wi th Junction Box
060-TR3520.88 060-TR5520.88 Enclosed Wall Mount wi th Junction Box
060-TR3522.88 060-TR5522.88 Enclosed Wall Mount wi th Junction Box and Pre-Installed Nalco Global Gateway
060-TR3521.88 060-TR5521.88 Enclosed Frame Mount with Junction Box
060-TR3523.88 060-TR5523.88 Enclosed Frame Mount with Junction Box and Pre-Installed Nalco Global Gateway
060-TR3530.88 060-TR5530.88 Air Purged Wall Mount with Junction Box
060-TR3532.88 060-TR5532.88 Air Purged Wall Mount with Junction Box and Pre-Installed Nalco Global Gateway
060-TR3531.88 060-TR5531.88 Air Purged Frame Mount with Junction Box
060-TR3533.88 060-TR5533.88 Air Purged Frame Mount with Junction Box and Pre-Installed Nalco Global Gateway
N/A 060-TR5515.88 Wall Mount for High Fouling Applications
N/A 060-TR5516.88 Frame Mount for High Fouling Applications
N/A 060-TR5525.88 Enclosed Wall Mount for High Fouli ng Applications
N/A 060-TR5526.88 Enclosed Frame Mount for High Fouli ng Applications
Wall Mount: Configured for wall attachment, select when floor space is restricted.
Frame Mount: Configured for floor-mounted installation, select when wall space is
restricted.
Junction Box: Provides convenient access to wiring connections.
Enclosure: Provides weather resistant protection. Select for outdoor installations
and dusty environments.
Air Purge: Depending on site requirements, may satisfy certain industrial explosion
protection ratings. Select for use in hazardous duty applications.
Wireless Gateway: Enables wireless communication of important information such as
alarms and performance indicators.
High Fouling Models: Includes automated fluorometer cleaning system to reduce fouling
from solids, oils and debris found in process water.
Technical Help 521-OM0108.88 2-2
(630) 305-CHEM Copyright 2009 Nalco Company
3D TRASAR 5500 - Installation & Operation Manual
!
Drawings shown with all probes
assembled. pH, ORP, and Cor-
rosion probes are av ailable as
upgrades for 3D TRASAR 3500
systems.
Recommended for
Model Number Description
system sizes:
060-TR5120.88 Pellet Feeder Assembly 12,000 gallons and larger
141-PJ4035.88 Liquid Bio-reporter pump, Iwaki 3,000 to 12,000 gallons
6008505 Liquid Bio-reporter pump, Grundfos Less than 3,000 gallons
Ethernet Crossover Cable (6 ft., orange) Service (Power) Cord 12/3 SJOW Additional Items
Qty. (1) Qty. (1) Replacement Fuses (4), Part # 991-50473718.88
Part # 991-05047661.88 Part # 991-01928722.88 Installation and User Manual, Part # 521-OM0108.88
pH Probe
Qty. (1) Corrosion Probe
Part # 060-TR541 1.88 Probe Restraint Device Qty (2) (Steel and Copper std.)
Qty. (2) Part # 400-NCMP1B.88 (steel)
Part # 400-NCMAC15.88 Part # 400-NCMP2B.88 (copper)
ORP Probe Part # 400-NCM4B.88 - Cu/Ni 90/10 (optional)
Qty. (1) Part # 400-NCMP5B.88 - Cu/Ni 70/30 (optional)
Part # 060-TR5421.88 Part # 400-NCMP7B.88 - 443 Brass (optional)
Part # 6016180 - 6061 Aluminum (optional)
Part # 6016182 - 1 100 Aluminum (optional)
Part # 400-NCMP10.88 Galvanized steel (optional)
PELLET
FEEDER
25'
Pellet Feeder Cable (25 ft.)
Qty. (1)
Part # 060-TR531 1.88
(shipped with pellet feeder only)
Additional Items
Pellet Feeder Brush Part # 500-P2818.88 Unpack and identify all parts.
• Unused cord grips must remain plugged. Install Main Control Panel in well-lit, dry
location convenient to points of use. Mount at
• Enclosure cover must be tightly closed at eye level.
all times.
• Ethernet port cover must also be in place
when this port is not in use.
Exact hole locations for these units are not available due to variations in manufacturing runs from
the enclosure supplier.
For models:
060-TR3501.88, 060-TR5501.88
060-TR3511.88, 060-TR5511.88
060-TR3521.88, 060-TR5521.88
060-TR3523.88, 060-TR5523.88
060-TR3531.88, 060-TR5531.88
060-TR3533.88, 060-TR5533.88
WARNING: Never install the pellet feeder in an explosion rated atmosphere. The
! pellet feeder does not have an explosion proof rating!
WARNING: Always shut-off power before opening the pellet feeder lid and placing your
! hands inside the hopper. NEVER insert your fingers in the pellet drop holes at the
bottom of the hopper!
! WARNING: Never attempt to lift the pellet feeder with a mechanical device, i.e. forklift!
overview.
IMPORTANT: Sample water flow must be a Plumb using ball valves provided. Inlet
! minimum of 5 gpm.
water from recirculating system must be
at 5 GPM or greater.
It is very important to select the appropriate sampling point for your cooling system. Sample points
are system dependent and should be taken from one of the following two locations:
1. Recirculating Pump Discharge - Sample points from the recirculating pump discharge
provide the maximum pressure opportunity for achieving the 5 gpm flow requirement of
the 3D TRASAR system and the Pellet Feeder.
2. Return Header to Cooling Tower - Sample points from the return header give the best
representation of corrosion rate measurements due to system heat loads.
In either case, it is important to understand the impact of the sample point in relation to control lag
times of the chemical feed (lag time is the time it takes the chemical to travel from point of injection
to the controller). In general, the shorter the lag time, the better the control (provided good mixing
has been achieved and recirculation rates are nominal). However, in cooling systems, the effects of
long lag times can be compensated for by decreasing the feed rate of the chemical pump, which
will reduce set-point overshoot.
Consult your Nalco representative for the appropriate sample point location for your system.
Note: If the required 5 gpm flow cannot be achieved through normal system operating
pressures, a booster pump may be needed to increase the flow rate.
Outlet
½”
Union
0.5
GPM
Outlets
IMPORTANT: Do not connect the sensor outlet plumbing to the fluorometer outlet plumb-
! ing – this may impair the flow through the fluorometer when inlet flow pressure varies.
tion.
o Oversized pumps can result in product
overfeeds, which can damage the cooling
system.
o Undersized pumps can result in repeated
timeout alarms, which are triggered when a
relay is energized for longer than a user
specified time. Plumb using ball valves provided. Inlet
water from recirculating system must be
• Do not inject chemicals immediately upstream of the at 5 GPM or greater.
3D TRASAR sampling point. Chemical injection
points should allow for good mixing prior to reaching
the sampling point.
• Avoid feeding biocides near the injection point of the bio-reporter
• Ensure that the suction side of the pump is always flooded.
Note: Acomplete line of chemical feed pumps is available through the Nalco Global Equipment
Solutions Group (NGES). Please call (800)323-8483 or (630) 848-3387 for assistance with
selecting the right pump for your application.
5 GPM
Note: Not all 3D TRASAR installations use the 3D TRASAR
bio-control feature. Pellet Feeder requires 4 to 5 gpm flow
2.5.1 Pellet Feeder Plumbing
Connect discharge back to recirculating
WARNING: Always shut-off power before opening system downstream from supply in basin
where possible.
the pellet feeder lid and placing your hands inside
! the hopper. NEVER insert your fingers in the
pellet drop holes at the bottom of the hopper!
IMPORTANT: The Bio-reporter pellet feeder requires a minimum sample water flow of 4.5
! GPM. The sensor outlet from the 3D TRASAR system or a separate water source can be
used as long as it is returned to the cooling system.
1. Plumb the pellet feeder inlet water to the ¾” ball valve on the left side of the unit.
2. Plumb the pellet feeder outlet to the ¾” ball valve on the right side of the unit.
3. Outlet water must go to a point of good circulation in the cooling system for maximum
and even distribution. Avoid close proximity to any biocide injection points.
4. Connect pellet feeder overflow to tubing and route to drain, or attach theBio-reporter
Liquid Overflow Container (P/N 060-TR5350.88) as shown below. The Overflow
Container must be ordered separately.
3D TRASAR 3D TRASAR
Starter
3D TRASAR Starter Included N/A
Fluorometer
3D TRASAR 6 channel Upgrade Included
Fluorometer
Conductivity Included Included
pH Upgrade Included
ORP Upgrade Included Install
Installprobes andand
probes attach
attach
Corrosion NCM100 Upgrade Included connectors.
connectors as shown.
Tighten swage fitting fully.
IMPORTANT: Corrosion Probe #1 must be mild
! steel if mild steel is one of the two metallurgies.
Refer to the chart below for the order of optimal
corrosion probe locations.
s
as
m
el
el
Ste
nu
Ste
Br
er
mi
pp
/N
m
lv
ld
Alu
Ga
Cu
Co
Ad
Mi
Probe #1 Probe #2
Fluorometer
Corrosion Probe #2
Corrosion Probe #1
WARNING: Always turn off power before making any electrical connections, which in-
! cludes all interconnecting cables, otherwise permanent damage may occur to system
components.
Technical Help 521-OM0108.88 2-25
(630) 305-CHEM Copyright 2009 Nalco Company
3D TRASAR 5500 - Installation & Operation Manual
IMPORTANT: The length of the sample supply tubing may need to be adjusted to attain
proper alignment of the mounting screws with the back panel. DO NOT BEND sample
! tubing. Excessive force may break fluorometer inlet fitting and internals.A broken inlet
fitting will necessitate the purchase of a new fluorometer which will not be covered under
warranty.
STEP 6 - Connect the flow switch shuttle (pre-wired to the controller) to the flow switch hous-
ing. Adjust the position of the ball valve as needed to maintain proper alignment with
the flow switch assembly (slotted holes are provided on the three-way valve bracket
for adjustments.)
STEP 7 - Connect the pre-installed fluorometer cable to the fluorometer.
probe.
Install
Installprobes andand
probes attach
attach
connectors.
connectors as shown.
Tighten swage fitting fully.
Install
Installprobes andand
probes attach
attach
Install the corrosion probes as follows: connectors.
connectors as shown.
Tighten swage fitting fully.
STEP 1 - Carefully unpack both corrosion probes and
locate the two probe restraints.
STEP 2 - Attach the probe restraints to each probe. The ring portion of the restraint is installed
between the hex nut and the threaded portion of the swage fitting as shown below .
! WARNING: The probe restraint must be installed on all 3D TRASAR units at all times.
Install
Installprobes andand
probes attach
attach
connectors.
connectors as shown.
Tighten swage fitting fully.
STEP 4 - Insert the copper probe in location #2 as indicated in Section 2.6. Probe tips must be
parallel to the ground. Two raised dots on the probe end cap can be used to orient
the tips.
STEP 5 - Tighten the swage fittings fully (finger tight plus 1¼ turn)
STEP 6 - Connect the corrosion probe cables to either of the two controller inputs (there is one
input connection on each side of the controller). The 3D TRASAR controller
automatically recognizes the probe metallurgy – no calibration is required.
2.7.2
Controller
Box Layout
Notes:
1. Relay voltage jumpers (terminals 4 & 5) -
Remove for non-powered output. Refer 2.7.8
and 2.7.9 for control relay details.
2. Control relays fused at 2.5 amps.
3. Alarm relay fused at 1.0 amps.
5. Board traces and contacts rated at 5 amps
(85-250 VAC).
6. For motor driven pumps a separate motor
starter must be used.
There are 7 pre-defined control relays in the 3D TRASAR controller that can be wired for powered
or non-powered operation. Follow the wiring diagrams below to connect metering pumps and
valves for powered control relay outputs (120 VAC, fused @ 2.5 amps).
In order to power the relay with 120 VAC, a jumper must be placed across
! control relay terminals 4 & 5 in the controller box, or the Line & COM terminals in
the junction box.
Fuse
Fuse
Contact
Contact
1 2 3 4 5
1 2 3 4 5
NO N L COM
NO N L COM
WHT
GRN
BLK
Controller
Controller Box
Box
Jumper
There are 7 pre-defined control relays in the 3D TRASAR controller that can be wired for powered
or non-powered operation. Follow the diagrams below to connect motor starters and/or solenoids
for non-powered control relay outputs (fused @ 2.5 amps).
The factory wired jumper must be removed from control relay terminals 4 & 5
! in the controller box, or the Line & COM terminals in the junction box, in order
to remove power from the relay output.
Fuse
Fuse
Fuse
Fuse Contact 2.5 amp
Contact
Contact 2.5 amp
1 2 3 4 5
1 2 3 4 5
1 2 3 4 5 NO N L COM
NO N L COM
NO N L COM
Controller
Box
Controller
Box
Controller
Box Ear
th
LINE COM NO
Earth Ground Bus
N Gr
ou
nd
Motor
Bu
Starter s
Motor
Starter
Probe wiring for the Fluorometer and Conductivity probe is done at the factory
.
The pH, ORP, and corrosion probes must be installed and connected using the quick disconnects.
IMPORTANT: Make sure the pH and ORP probes are connected to the appropriate
! terminals as they look very similar and will display a reading even if connected improperly.
The analog inputs must be externally powered. They can be used to control conductivity
, pH, and
ORP. Refer toAppendix I for information on the halogen analyzer
.
The analog outputs are internally powered for a max. load of 600 ohms. The analog outputs can be
used to transmit any of the following variables:TRASAR product dosage, Tag Polymer product
dosage, Cell Fouling, Nalco Scale Index, Nalco Bio-Index, Conductivity, pH, ORP, Corrosion
#1, Corrosion #2, and Sample Water Temperature.
There are two factory wired digital inputs. The first is the
fluorometer flow switch, which is used to suspend control of any relay associated with the fluorom-
eter upon loss of sample flow.
The second digital input is the system interlock, which is intended for use on cooling systems that
operate intermittently. Afactory installed jumper wire is placed across the terminals of this input
(leave in place under normal operation). Removal of this jumper will suspend operation ofALL
control relays and initiate an alarm.
Controller Box
Connection Controller Box Terminal
Junction Box Terminal
Description Board Reference (Removable
Connector)
Digital Input 1
Fluorometer Flow TB10 Flow (+) N/A
Switch
TB10 Flow (-) N/A
Digital Input 2
TB10 INTERLOCK (+) INT (+)
System Interlock
TB10 INTERLOCK (-) INT (-)
There are two user wired digital inputs that can be used to log and/or alarm the state of an
unpowered contact or totalize the counts from a low frequency pulse water meter. Note: Please call
630-305-CHEM for wiring assistance with “hall-effect sensor” transistor output water meters.
For water meter inputs, each pulse represents a pre-defined amount of water usage (i.e., 10 gal-
lons/pulse, 100 gallons/pulse, etc.). The gallon scaling factor is user-defined in the 3D TRASAR
Configurator. Note: The minimum on-time and off-time pulse width is 5 milliseconds for a valid
pulse. The maximum number of pulse counts is 10 per second.
The second alarm output is a float switch alarm that is only used with 3D TRASAR models that are
fully enclosed. This alarm is initiated if water starts accumulating in the enclosure. This alarm
should be sent to a location where it can be responded to as quickly as possible to avoid damaging
the electronics.
The phone line connection is located inside the controller box and can be connected through the
RJ11 analog phone jack or hardwired to terminals 1 and 2.
Connection Controller Box Controller Box Terminal Junction Box
Description Board Reference (or Phone Jack) Terminal
Modem
TB-1 1 AP 1
(Phone Line)
TB-2 2 AP 2
RJ11 Analog
Phone Jack
Note: Connect the analog “Tip” and “Ring” phone wires from the
telephone company to pins 1 and 2. The modem input is not polarity
sensitive, so it does not matter which order the wires are connected.
2. Double check all cables to ensure correct and tight electrical Calibrate probes following
instruction on screen.
connections.
5. Verify that a minimum of 5 gpm is flowing through the unit as indicated by the rotometer
.
6. Verify that the circuit breaker in the appropriate electrical service box is switched on.
9. Adjust the water flow through the fluorometer to a minimum of 0.5 gpm as indicated by the
flowswitch.
10. Verify that a “No FlowAlarm” does not exist on the 3D TRASAR controller display
. No flow is
indicated by a flashing “F”.
11. Verify that a “System Interlock” alarm does not exist on the 3D TRASAR controller display
.
System Interlock alarms are indicated by a flashing “I”.
IMPORTANT: Make sure all relays are in “Auto” mode after checkout. Pumps or valves
left in “Man-On” mode will result in excessive waste of chemical and water which may
! cause damage to the cooling system.
• Through the controller keypad, verify that all of the new settings from the “Configurator”
have been accepted into the controller.
• The “System Start-up” mode allows you to temporarily suppress alarms for a user defined
period of time. This prevents false alarming and relay time-outs from occurring until the
cooling system has reached chemical equilibrium.
NOTE: Although many settings can be changed through the 3D TRASAR controller
! keypad, the configurator must be used for initial controller setup and to make
changes to control methods.
NOTE: The configurator software operates optimally when Lotus Notes is installed on
your PC. Lotus Notes is used as a means of obtaining the most recent data for use in the
! configurator. The configurator also has the capability of generating email messages (with
the user’s consent), which are sent via the user’s Notes mailbox at the next replication.
However, the program has been designed to operate in the absence of Notes with a more
limited capability.
For Nalco Windows XP users, the configurator program should be pre-installed. To open
click Start, Programs, Nalco Software, 3D TRASAR, 3DT Configurator.
If the program is not pre-installed, it can be downloaded from SMS. To download, click
Start, Programs, System Manager. Scroll through programs, locate Nalco 3D TRASAR
Configurator.
A second program, 3Cdaemon, will also be installed. This is a TFTP server program used
only for controller software upgrades.
Display Panel
Menu
Analog System
Blowdown Contrast Plot Data
Alarms Startup
SCADA Control
pH Test Email
Address Methods
SCADA Service
Mode
Timer Status
Response
Email
SCADA External
Baud Inputs
Product
Factors
Diagnostic
Data
Menu
This submenu is used to access the system control settings. Only items that have been
configured as “controlling” will display.
Menu
TRA Background Shows the TRASAR background fluorescence value. Note: Do not use
TRASAR Control or 3D TRASAR scale control if TRABackground is > 15%.
TAG Background Shows the tagged polymer background fluorescence value.Note: Do not
use TAG Polymer Control if TAG Background is > 75%. Do not use 3D
TRASAR scale control if TAG Background is > 30%.
TAG Ratio Shows the ratio for tagged polymer background fluorescence correction.
TAG Method Toggles between Fixed offset correction and Online background correction
for tagged polymer control. Note: The preferred Online TAG Method can only
be used if TAG Ratio is between 1.41 and 1.71. Fixed TAG Method can be
used for any TAG Ratio value.
IMPORTANT: Periodically check the Tagged Polymer level on the controller against the
level determined by the Nalco DR/2800 or DR/890 Active Polymer T est (Reagent P/N
! 460-S0811.75). The results from the Active Polymer Test must be multiplied by (Product
Factor divided by 10) to convert to ppm of product for accurate comparisons.
Select Fixed for TAG Method and enter 25% for TAG Bkgd
Note: Tagged polymer products must be fed for a minimum of 5 HTI’s beforeActive Polymer Test
results can be used for comparison.
Menu
4.4.6 pH Screen
Displays the pH set point and control range.
NOTE: Control ranges will only be displayed if applicable. Control range is not
! displayed if 3D Scale Index schemes are used. In these cases, the ranges are
determined by the control algorithm.
Menu
NOTE: The duty cycle percentages for high and low alarms
! can be set to different values if needed via the Configurator
.
Startup Factor allows you to extend the relay time-out period during
startup. Chemical residuals may be extremely low at startup and
require an extended period to equilibrate. The Startup Factor (select-
able between 1 and 99) multiplies this initial time out period for all
relays (i.e. if the Relay Timeout is set for 60 minutes and the Startup
Factor is set for 5, the timeout period will be 300 minutes for startup).
Pump timeout alarms will be cleared automatically if the related parameter concentration is within
the defined control range or can be cleared manually at the controller via the keypad.
Relay Timeout - Maximum time for continuous relay operation before an alarm is initiated and the
Failsafe Duty mode begins.
Failsafe Duty - When any alarm condition exists for the parameter associated with a relay
, the relay
reverts to the Failsafe Duty mode. Failsafe Duty can be set as “Fixed” or “Smart”.
The Failsafe Duty set as "Fixed" is the percentage of a 15-minute period that the relay is
activated. For example, if the failsafe duty is set at 20%, the relay will be activated for 3
minutes and deactivated for 12 minutes. This pattern repeats as long as the alarm condition
exists. Set these carefully as improperly set failsafe modes can cause under or overfeed
conditions during "alarm" conditions.
Note: “Fixed” Failsafe Duty can be set via both the Configurator and the Key-
pad, which is used for Pump Timeout and NoFlow alarm conditions. Level alarm
! Failsafe Duty can be set to different values for high or low level alarm conditions
ONLY through the Configurator .
The Failsafe Duty set as "Smart" is the average run time of that relay for the last 7 days.
Selecting this mode allows the system to continue to operate under average conditions
based on past performance during failsafe operation. TheFailsafe Duty median is used for
all alarm conditions with the “Smart” setting
Menu
This submenu is used to display and edit general settings. The arrow keys can be used to navigate
between fields with more than one setting (i.e., the time field has both hours and minutes).
Date…………………………………Shows the date setting in the format configured through the PC.
SCADA Address…………………..The Modbus address for the controller. This information needs
to be supplied by the customer only if SCADA is used.
SCADA Baud………………………Selects the SCADABaud rate (1200, 2400, 4800, 9600, 19200).
The Network submenu screen accesses the network settings of the ethernet port.
Before making changes to the network settings check with your local Network Administrator or ISP
provider. Any changes made must also be made in the configuration program!
DHCP……………………….. Enables/disables the assignment of the controller IP address for
use on a network. In most cases, this should be disabled.
This submenu is used to display the start-up system parameters as entered by theuser. This is the
starting point for the water and energy calculations.
Menu
The information key accesses general performance information. There are no adjustments or
changeable settings accessible through these screens.
Nalco Scale Index - Reports the status of the Nalco Scale Index calculations. This is a
unit-less measure of the scaling tendencies of the system water. It is used for trending
and is not to be viewed as a direct indication of scaling as each system will trend
uniquely.
Nalco Bio-Index - Reports the status of the Nalco Bio-Index calculations. This is a unit-
less measure of biological activity. It is used for trending as each system will read
uniquely based on operating characteristics.
Corrosion Probe #1 - Reports the corrosion rate for Corrosion probe # 1 (Note: Corro-
sion shown in screen display above indicates a mild steel (stl) probe is installed).
Corrosion Probe #2 - Reports the corrosion rate for Corrosion probe # 2 (Note: Corro-
sion shown in screen display above indicates a Copper (Cu) probe is installed).
Consumption - Reports the actual consumption of the Tag Polymer. It indicates the
difference between the quantity fed and the active quantity remaining in the system. (In
the screen display above, 18% consumption means 82% of active polymer remains.)
Menu
Temp................................ Real time measurement from the RTD in the conductivity probe.
Menu
Blowdown - Calculated blowdown rate in gallons per day based on mass balanceequations.
Makeup - Calculated make-up rate to the system based on mass balance equations.
Heat Rejection - Calculated heat rejection in BTU based on mass balance equations.
Recirculation Rate - Calculated water recirculating rate based on mass balance equations.
System Volume - Calculated system volume based on feed rates and residuals.
Menu
The plots can show a time scale of either 1, 7, or 30 days. Time scale is changed by pressing the
soft key labeled X Zoom.
The plots can also be changed to show 1X or 2X by pressing the soft key labeled
Y Zoom.
For Example:
In the 30-day view, the graph spans 1-week time periods. The average value for the week is dis-
played next to the y axis. Likewise, in the 7-day view, the graph spans 1-day. The average value
for the day is displayed.
In the 1-day view, the cursor is a line and corresponds to a single point whose value is displayed.
In each view, the cursor can be moved left or right using the arrow keys. Scrolling off the screen
will display data from the next or previous period of time. The previous 90-days of data may be
displayed.
Menu
4-20 mA inputs shows scaled values. Digital Input 1 shows the System Sample Flow On/Off status.
Digital Input 2 shows the System Interlock On/Off status. Digital Inputs 3 and 4 may be configured
to show On/Off status of a non-powered contact (see DI3 in figure below) or as low frequency water
meter counter to show current and previous day’s total usage (see DI4 in figure below).
External inputs can be used as a control signal for pH, ORP, and Conductivity. Refer to page 2-45.
Main Menu
Clean and Calibrate - Accesses the calibration for the fluorometer, pH, conductivity, and ORP.
Manual Relay Control - Selects control relay mode (Manual - Off -Auto).
Pumping Rates - Enter pump feed rates based on pump calibration information.
Slug Feed Biocide - Use this submenu to manually slug feed biocides.
System Startup - Activates the Controller "Startup" mode which delays alarms.
Test Email - Sends a test data e-mail and alarm e-mail to the 3D TRASAR Web.
Service Response Email - Sends one of ten service e-mails to the 3D TRASAR Web.
Menu
Menu
STEP 1
Close fluorometer inlet valve and turn 3-way fluorometer discharge valve to
sample position.
STEP 2
Clean the fluorometer by injecting acid with a syringe into the flow cell. Fill the
syringe with 60 ml of dilute acid. 10% SulfuricAcid (P/N 460-S0800.75) is rec-
ommended. If iron fouling exists, use 1:1 HCl(P/N 460-S0726.75) for cleaning.
Screw the syringe onto the fitting, inject the dilute acid at a slow
, steady rate into
the flow cell, and allow it to stand for 2 minutes. Rinse well with water. Using the
flow cell brush (P/N 500-P2817.88), gently push the brush down through the
opening and scrub the full length of the flowswitch and fluorometer tube.
IMPORTANT: If using 1:1 HCl, do not leave the acid in the flow cell for more than
five minutes.
STEP 3
Following the screen prompts, use a second clean 60 ml syringe and slowly
flush the flow cell with 180 ml of blank solution. Then slowly inject another 180 ml
blank solution and allow it to remain in the flow cell by leaving the syringe at-
tached. Press Continue.
STEP 4
Following the screen prompts, use a third clean 60 ml syringe and slowly inject
the flow cell with 180 ml of calibration solution(P/N 460-S0940.75) and allow it
to remain in the flow cell by leaving the syringe attached. PressContinue.
STEP 5
Successful calibration will be indicated and must be “accepted”. Controller will
automatically re-boot after accepting calibration. Turn valves back to operating
positions.
Menu
STEP 2 - Close the sample inlet, lower sensor outlet, and 3-way valves.
STEP 5 - After prompt, rinse probe well in clean DI water and wipe clean.
Menu
STEP 2 - Close the sample inlet, lower sensor outlet, and 3-way valves.
STEP 4 - Swirl the probe in a beaker of clean, fresh ORP 200 mV buffer
solution (P/N 460-S0932A.75). Press Continue.
STEP 5 - After prompt, rinse probe well in clean DI water and wipe clean.
Menu
A 1-point calibration utilizes the system water as the calibration point. tI is best
used to fine-tune your calibration to compensate for minor drifting. Leave the
probe in the process water. The 1-point calibration range is +/-10% of the
second point value of the previous 2-point calibration.
Tips
√ Conduct the 2-point conductivity calibration after completing other cali-
brations (pH, ORP, etc.).
√ A calibrated conductivity meter with correct temperature coefficient for
the process water is required.
√ Measure the water sample and obtain an accurate conductivity value.
√ Follow the instructions on the screen.
√ Standard solution for the probe calibration is not required. If using a
standard solution for the calibration, choose a solution close to the
value of the process water conductivity value.
Menu
Tip
√ A 1-point conductivity calibration can be performed when conductivity
has drifted, or changed <+/- 10% of the second point value of the previ-
ous 2-point calibration
STEP 1 - Leave probe in the Tee with water flowing (probe should have been
installed at least 15 minutes for equilibrium.) Select 1-point
calibration.
Menu
C a lib r a tio n T y p ic a l T y p ic a l
R e p la c e m e n t M a in te n a n c e
P a r a m e te r S ta n d a r d s M a in te n a n c e P ro b e
P ro b e P a rt # In d ic a tio n
R e q u ir e d S c h e d u le * L ife
U s e s ys te m w a te r C le a n a n d
C o n d u c ti v i t y
C o n d u c ti v i t y 0 6 0 - T R 5 4 4 1 .8 8 a n d p e rfo rm a "1 " c a li b ra te In d e fi n i te
d ri f t
p o i n t c a li b ra ti o n m o n t h ly
C le a n a n d
C e ll fo u li n g C a lib ra t i o n S o lu ti o n
F lu o ro m e te r 0 6 0 - T R 5 2 2 0 .8 8 c a li b ra te In d e fi n i te
a la rm (4 6 0 -S 0 9 4 0 .7 5 )
m o n t h ly
p H 7 .0 B uffe r
(4 6 0 -S 0 4 0 7 .8 8 ) C le a n a n d
pH 0 6 0 - T R 5 4 1 1 .8 8 p H d rift c a li b ra te U p to 1 yr.
p H 1 0 .0 B uffe r m o n t h ly
(4 6 0 -S 0 4 0 8 .8 8 )
200 mV
(4 6 0 -S 0 9 3 2 A .7 5 ) C le a n a n d
ORP 0 6 0 - T R 5 4 2 1 .8 8 O R P d rift c a li b ra te U p to 1 yr.
600 mV m o n t h ly
(4 6 0 -S 0 9 3 3 A .7 5 )
4 0 0 - N C M P 1 B .8 8
S te e l
4 0 0 - N C M P 2 B .8 8
C opper
4 0 0 - N C M P 4 B .8 8
C u N i 9 0 /1 0
4 0 0 - N C M P 5 B .8 8
C u N i 7 0 /3 0
S tu d y In s p e c t p ro b e U p to 6
C o rro s i o n N o t re q u i r e d
c o m p le te d m o n t h ly m o n th s
4 0 0 - N C M P 7 B .8 8
B ra s s 4 4 3
6016180
6 0 6 1 A lu m i n u m
6016182
1 1 0 0 A lu m i n u m
4 0 0 - N C M P 1 0 .8 8
G a lva n i z e d S te e l
Menu
All configured control relays can be placed in Manual On, Manual Off, or
Auto through these screens.
STEP 1 - After selecting desired relay, highlight Current Mode and press Edit.
STEP 2 - Use arrow keys to select desired mode (Manual Off, Manual On, Auto) and press
Accept.
STEP 3 - After selecting desired mode, highlight Auto After and press Edit (Auto After is
the amount of time the relay will remain in the selected mode before reverting back
to auto).
STEP 4 - Enter the Auto After value in hours and minutes using the keypad (0-4-2-5 equals 4
hours 25 minutes). Entering 00:00 will put the relay in the selected mode continu-
ously.
STEP 1 STEP 3
STEP 2 STEP 4
Menu
Chemical feed pumping rates are used to monitor and report on chemical usage and several calcu-
lations of water and energy use. The inert TRASAR product pumping rate is used to determine
system volume. The accuracy of this reporting is dependant on the accuracy of the pumping rates
input here.
For specific directions on proper calibration techniques, refer to the pump manufacturers instruc-
tions supplied with the pumps. The calibrated cylinder draw-down method is suggested for greatest
accuracy.
Menu
There are three options for slug feeding biocides. Each choice will follow the pre-bleed and lockout
conditions setup in the configuration.
1. Use Hand Feed Nonox to physically add a non-oxidizer by an operator to the cooling system.
Note: If feeding Glutaraldehyde the configured oxidizing feed will be suspended for 8 hours.
2. Use Hand Feed Oxidizer to physically add an oxidizer by an operator to the cooling system.
Menu
Menu
NOTE: pH and ORP are high impedance measurements, and it is difficult to detect
missing or miswired probes. Therefore, the pH and ORP Sensor Errors are not
! always triggered. Inaccurate or fluctuating readings are another indication of a
miswired or faulty probe.
• Turbidity High
• Temperature High
• Temperature Low
• Cell Fouling High
• No Flow
• Interlock
Interlock - The Interlock digital input provides the ability to use a signal, for example, from the
recirculation pumps to enable or disable the 3D TRASAR control relays. If the recirculation pumps
are turned off, the contact is opened, and the Interlock alarm occurs immediately , disabling all
control relays. Control is restored when the recirculation pumps are turned back on and the contact
is closed, which clears the Interlock alarm. If this feature is not needed, the Interlock input must be
shorted with a wire jumper so the contact is always closed.
Menu
Control Alarms in this category are related to Failsafe Duty and can be set to different values for High
and Low alarms via the configurator. In monitoring mode, the parameter monitored may be sepa-
rated from the relay that is on timer control for a different product.
Menu
WARNING: DO NOT use 3D TRASAR Scale Control if the TRASAR Background is > 15%
or TAG Background is > 30%. DO NOT use Tagged Polymer Control if TAG Background is
! > 75%. Refer to Section 4.4.1 for more details.
3D TRASAR Scale Control uses NSI to make setpoint adjustments in response to both high and
low stress conditions. There are two different control methods to choose from, either 3D TRASAR
Scale Blowdown Control, or 3D TRASAR Scale Tagged Polymer Control.
5.1.1 3D TRASAR Scale Blowdown Control – Uses NSI to Adjust the Blowdown
Setpoint
• If NSI < 0: the blowdown setpoint is increased.
(e.g. NSI = – 20; blowdown setpoint adjusted by + 2.0%)
• If NSI = 0: no blowdown setpoint adjustment is made.
• If NSI > 0: the blowdown setpoint is decreased.
(e.g. NSI = + 20; blowdown setpoint adjusted by - 2.0%)
5.1.2 3D TRASAR Scale Tagged Polymer Control – Uses NSI to Adjust the Tagged
Polymer Setpoint
• If NSI < 0: the Tagged Polymer setpoint is decreased.
(e.g. NSI = – 20; Tagged Polymer setpoint adjusted by - 2.0%)
• If NSI = 0: no Tagged Polymer setpoint adjustment is made.
• If NSI > 0: the Tagged Polymer setpoint is increased.
(e.g. NSI = + 20; Tagged Polymer setpoint adjusted by + 2.0%)
Conductivity
TRASAR
100
Tagged Polymer
0
Before Step 1 Step 2 Steps 3 & 4
Time
Before - Make sure to baseline your system performance data before initiating the 4-Step process. By
doing this you will be able to clearly see the added value provided by 3D TRASAR Scale Control.
Step 1 TRASAR Control Mode – The goal of this step is to determine the Tagged Polymer setpoint
to be used in Step 2. The 1-Drum product is controlled by traditional TRASAR while the aTgged
Polymer level is monitored. Blowdown is controlled based on conductivity during this period. It will
take approximately 5 HTI’s to complete Step 1.
Step 2 3D TRASAR Scale Blowdown Monitoring Mode – In this step the 3D TRASAR controller
learns the typical rate of polymer consumption needed to calculate NSI. The 1-Drum product
control is switched to Tagged Polymer Control using the setpoint determined in Step 1. The
TRASAR level will now only be monitored and blowdown will still be controlled by conductivity
. After
the learning period is complete you can move to Step 3.
Step 3 3D TRASAR Scale Blowdown Control Mode – This step switches the blowdown control
from conductivity-based to 3D TRASAR Scale Control. The controller will now respond to high
system scale stresses by reducing the blowdown setpoint.
Step 4 Advanced 3D TRASAR Scale Blowdown Control Mode – This step adds to 3D TRASAR
Scale Blowdown Control the feature of increasing the blowdown setpoint during low scale stress
conditions to maximize system water and chemical efficiencies.
Note: On the controller LCD, “Product-i” reading indicates the inert TRASAR level and “Product-a”
reading indicates the active Tagged Polymer level, as ppm product.
!STOP! The “-i” (TRASAR) reading should be greater than the “-a” (TAG) reading.
IF NOT – call 630-305-CHEM before proceeding and remain in TRASAR control.
• Change the product feed pump electrical connections to TAG POLYMER control Relay # 2.
• Leave the blowdown valve connected to BLOWDOWN control Relay # 6.
• In the Configurator Control Settings window:
1. Click the Blowdown/pH tab, set the Blowdown Control Method to 3D TRASAR Scale
Control.
2. Under 3D TRASAR Scale Control Settings, check the Monitor Only (Setpoint does not
change) box. By checking this box, no blowdown setpoint adjustments will be made.
3. Click the Inhibitor tab, change the Product Control Method to Tagged Polymer Control,
enter the last month average Tagged Polymer level as the Inhibitor 1 Set Point, and accept
the default control range.
• Do NOT change any other settings.
5.2.2 Start-Up Procedure for 1-Drum 3D TRASAR Scale Blowdown Control (con’t)
Upload the new configuration to the 3D TRASAR controller.
• The Configurator program should request an automatic “REBOOT” of the controller.
If not – reboot manually either by using the power switch or through the Configurator: go to “Con-
troller” —> “Tools” —> “Reboot Controller”.
• After the controller reboots – go to the Configurator Control Settings window, under the
Blowdown/pH tab, click the RELEARN NOW button to perform a Relearn. This will clear the
memory of learned scaling stresses and start the new Learning Period.
• The number of cycles in the “Learning Period” is automatically calculated and varies based on the
user input HTI.
• The system is now under 3D TRASAR Scale Blowdown Monitoring Mode. Note: The blowdown
set point will NOT be adjusted based on NSI until Step 3.
Note: The controller LCD will show dashes (- - -) for NSI, and the data log will show a value of–999
for NSI until the required “Learning Period” is completed. The “Learning Period” will last between 7
to 8 HTI’s if the system HTI is < 72 hours, and 22 days if the HTI is > 72 hours. The “Learning
Period” is over when NSI values start to appear.
IMPORTANT: The blowdown set point will now be decreased by as much as 30% in
response to high system stress but will NOT be allowed to go above the user-entered set
! point since the “Set Point Increase Limit” is 0%.
Conductivity
100 TRASAR
Tagged Polymer
Time
Before - Make sure to baseline your system performance data before initiating the 4-Step process.
By doing this you will be able to clearly see the added value provided by 3D TRASAR Scale
Control.
Step 1 TRASAR Control Mode - The purpose of this step is to gradually replace any old
UNTAGGED polymer product with the new, TAGGED polymer product. The TRASAR Product is
controlled based on TRASAR while the Tagged Polymer Product is slaved to the TRASAR Product.
Blowdown is controlled based on conductivity during this period.
Step 2 3D TRASAR Scale Blowdown Monitoring Mode - In this step the 3D TRASAR controller
learns the typical rate of polymer consumption needed to calculate NSI. The T
agged Polymer
product will now be controlled based on the setpoint recommended by the 3D TRASAR Optimizer .
The TRASAR product will still be under TRASAR control mode. Blowdown is controlled at the same
conductivity setpoint, while in Scale Monitoring mode with no blowdown setpoint adjustments.
Step 3 3D TRASAR Scale Blowdown Control Mode - This step switches the blowdown control
from conductivity-based to 3D TRASAR Scale Control. The controller will now respond to high
system scale stresses by reducing the blowdown setpoint.
Step 4 Advanced 3D TRASAR Scale Blowdown Control Mode - This step adds to 3D TRASAR
Scale Blowdown Control the feature of increasing the blowdown setpoint during low scale stress
conditions to maximize system water and chemical efficiencies.
This example shows that the Tagged Polymer product Pumping Rate should be 5.53 times the
TRASAR product Pumping Rate (i.e. TRASAR Pumping Rate = 1 gpm; Tagged Polymer Pumping
Rate = 5.53 gpm).
5. Click OK.
5.2.4 Start-up Procedure for 2-Drum 3D TRASAR Scale Blowdown Control (con’t)
• Upload the new configuration to the 3D TRASAR controller.
• The Configurator program should request an automatic “REBOOT” of the controller.
If not – reboot manually either by using the power switch or through the Configurator: go to “Con-
troller” —> “Tools” —> “Reboot Controller”.
• Operate the cooling system under TRASAR Control and slaved Tagged Polymer Control for
approximately 5 HTI’s. The Tagged Polymer product level should reach steady state after 5 HTI’s,
while any old untagged polymer will be reduced to zero.
• Once the Tagged Polymer product level reaches steady state move to Step 2.
NOTE: If the steady state Tagged polymer product level needs to be achieved in a shorter
! period of time, it is OK to slug feed the product. It is still necessary to run for approxi-
mately 5 HTI’s to replace untagged polymer with Tagged polymer.
Note: The controller LCD will show dashes (- - -) for NSI, and the data log will show a value of–999
for NSI until the required “Learning Period” is completed. The “Learning Period” will last between 7
to 8 HTI’s if the system HTI is < 72 hours, and 22 days if the HTI is > 72 hours. The “Learning
Period” is over when NSI values start to appear.
5.2.4 Start-up Procedure for 2-Drum 3D TRASAR Scale Blowdown Control (con’t)
STEP 3 - 3D TRASAR Scale Blowdown Control Mode
• In the Configurator Control Settings window:
1. Click the Blowdown/pH tab, leave the Blowdown Control Method in 3D TRASAR Scale
Control.
2. Under 3D TRASAR Scale Control Settings, click the Setpoint Limits button.
3. Enter 0% for Setpoint Increase Limit, 30% for Setpoint Decrease Limit, and click OK.
4. After returning to the Blowdown/pH page, uncheck the Monitor Only (Setpoint does not
change) box.
• Do NOT change any other settings.
• Upload the modifications to the 3D TRASAR controller.
• The system is now under 3D TRASAR Scale Blowdown Control.
IMPORTANT: The blowdown set point will now be decreased by as much as 30% in
response to high system stress but will NOT be allowed to go above the user-entered set
! point since the “Set Point Increase Limit” is 0%.
IMPORTANT: The blowdown set point will now be decreased by as much as 30% in
response to high system stress AND will be allowed to go above the user-entered set point
! by as much as 10% in response to low system stress.
Conductivity
Time
Before - Make sure to baseline your system performance data before initiating the 4-Step process. By
doing this you will be able to clearly see the added value provided by 3D TRASAR Scale Control.
Step 1 TRASAR Control Mode - The purpose of this step is to gradually replace any old
UNTAGGED polymer product with the new, TAGGED polymer product. The TRASAR Product is
controlled based on TRASAR while the Tagged Polymer Product is slaved to the TRASAR Product.
Blowdown is controlled based on conductivity during this period.
Step 2 3D TRASAR Scale Tagged Polymer Monitoring Mode - In this step the 3D TRASAR control-
ler learns the typical rate of polymer consumption needed to calculate NSI. The T agged Polymer
product will now be controlled based on the setpoint recommended by the 3D TRASAR Optimizer .
The TRASAR product will still be under TRASAR control mode. Blowdown is controlled at the same
conductivity setpoint. No Tagged Polymer setpoint adjustments will be made until Step 3.
Step 3 3D TRASAR Scale Tagged Polymer Control Mode - This step switches the Tagged
Polymer control to Scale Index Control. The controller will now respond to high system scale
stresses by increasing the Tagged Polymer setpoint.
Step 4 Advanced 3D TRASAR Scale Tagged Polymer Control Mode - This step adds to 3D
TRASAR Scale Tagged Polymer Control the feature of decreasing the Tagged Polymer setpoint
during low scale stress conditions to maximize chemical efficiency.
5.3.2 Start-up Procedure for 2-Drum 3D TRASAR Scale Tagged Polymer Control
STEP 1 - TRASAR Control Mode
• Connect the TRASAR product feed pump to TRASAR control Relay # 1.
• Connect the Tagged Polymer product feed pump to TAG POLYMER control Relay # 2.
• Connect the blowdown valve to BLOWDOWN control Relay # 6.
• At the controller, activate the “System Start-up” mode (Refer to Section 4.10.11).
• Launch the Configurator on your laptop PC and select the configuration for this system.
• In the Configurator Alarm Settings window:
1. Click the Failsafe Relay Settings tab, set the Blowdown Valve Timeout Limit to a mini-
mum of 240 minutes (If the blowdown valve is not being controlled by the 3D TRASAR
controller, set the Blowdown Valve Timeout Limit to zero).
2. Click OK.
• In the Configurator Control Settings window:
1. Under Installed Sensors set the Conductivity Probe to Controlling.
2. Click the Blowdown/pH tab, set the Blowdown Control Method to Conductivity
Setpoint, enter the desired Blowdown Set Point, and accept the default control range.
3. Click the Inhibitor tab, set the Number of Inhibitor Products to 2 or more, choose the
Product Names, set the Control Method of the TRASAR Product to TRASAR Control,
enter the TRASAR Product Set Point value recommended by the 3D TRASAR Optimizer,
accept the default control range, and set the Control Method of the Tagged Polymer Prod-
uct to Slaved.
4. Still under the Inhibitor tab, enter the TRASAR product Pumping Rate (R1) and the TAG
Polymer product Pumping Rate (R2) according to the following equation. Note: Make sure
to match the calibration of the TRASAR product pump and the TAG polymer product pump
to the feed rates entered.
This example shows that the Tagged Polymer product Pumping Rate should be 5.53 times the
TRASAR product Pumping Rate (i.e. TRASAR Pumping Rate = 1 gpm; Tagged Polymer Pumping
Rate = 5.53 gpm).
5. Click OK.
5.3.2 Start-up Procedure for 2-Drum 3D TRASAR Scale Tagged Polymer Control
(con’t)
• Upload the new configuration to the 3D TRASAR controller.
• The Configurator program should request an automatic “REBOOT” of the controller.
If not – reboot manually either by using the power switch or through the Configurator: go to “Con-
troller” —> “Tools” —> “Reboot Controller”.
• Operate the cooling system under TRASAR Control and slaved Tagged Polymer Control for
approximately 5 HTI’s. The Tagged Polymer product level should reach steady state after 5 HTI’s,
while any old untagged polymer will be reduced to zero.
• Once the Tagged Polymer product level reaches steady state move to Step 2.
NOTE: If the steady state Tagged polymer product level needs to be achieved in a shorter
! period of time, it is OK to slug feed the product. It is still necessary to run for approxi-
mately 5 HTI’s to replace untagged polymer with Tagged polymer.
Note: The controller LCD will show dashes (- - -) for NSI, and the data log will show a value of–999
for NSI until the required “Learning Period” is completed. The “Learning Period” will last between 7
to 8 HTI’s if the system HTI is < 72 hours, and 22 days if the HTI is > 72 hours. The “Learning
Period” is over when NSI values start to appear.
5.3.2 Start-up Procedure for 2-Drum 3D TRASAR Scale Tagged Polymer Control
(con’t)
STEP 3 - 3D TRASAR Scale Tagged Polymer Control Mode
• In the Configurator Control Settings window:
1. Click the Inhibitor tab, leave the Tagged Polymer Control Method in Scale Index Control.
2. Under 3D TRASAR Scale Control Settings, click the Setpoint Limits button.
3. Enter 30% for Setpoint Increase Limit, 0% for Setpoint Decrease Limit, and click OK.
4. After returning to the Inhibitor page, uncheck the Monitor Only (Setpoint does not
change) box.
• Do NOT change any other settings.
• Upload the modifications to the 3D TRASAR controller.
• The system is now under 3D TRASAR Scale Tagged Polymer Control.
IMPORTANT: The Tagged Polymer set point will now be increased by as much as 30% in
! response to high system stress but will NOT be allowed to go below the user-entered set
point since the “Set Point Decrease Limit” is 0%.
IMPORTANT: The Tagged Polymer set point will now be increased by as much as 30% in
response to high system stress AND will be allowed to go below the user-entered set point
! by as much as 10% in response to low system stress.
2. How does 3D TRASAR Scale Control measure the Scaling Stress in a cooling
system?
Inert TRASAR
TRASAR Decay:
Hydraulic Loss
Tagged Polymer
After the product(s) is(are) fed to the system and the pump(s) is(are) turned off, the concentrations
of inert TRASAR (also known as PTSA) and tagged polymer will decrease due to hydraulic loss,
such as blowdown and system leaks. The hydraulic loss impacts the inert TRASAR level and
tagged polymer level equally. When scaling stress increases in a cooling system, the polymer
decay rate will be faster than the inert TRASAR decay rate. Therefore, the Relative Rate of Poly-
mer Consumption, or the rate of polymer decay compared to the inert TRASAR decay, will be
increased. 3D TRASAR Scale Control measures the Relative Rate of Polymer Consumption, K, as
an indication of the scaling stress.
It should be noted that the cooling system is usually operated under some level of scaling stress.
The key is to know when the scaling stress is changed, the magnitude of the change in scaling
stress (upsets) compared to the normal operating conditions and the timely corrective actions that
need to be taken to prevent scaling.
3. I'm already using TRASAR, why should I switch to 3D TRASAR Scale Control?
3D TRASAR Scale Control not only monitors product feed but also provides a superior level of
control by monitoring the active polymer level in the system and determining the scaling tendency of
the cooling tower. With 3D TRASAR Scale Control, scaling stress is automatically reduced by
maintaining the desired active polymer level in the system or by automatically blowing down the
system when needed.
• TRASAR provides only product feed set point control. It does not measure any scaling tendency
or system performance of the cooling tower.
• 3D TRASAR Scale Control employs a new improved Fluorometer with better accuracy for better
product feed control.
• 3D TRASAR Scale Control can maintain the desired residual tagged polymer level to minimize/
avoid scaling.
• 3D TRASAR Scale Control also allows one to determine if deviations from the learned normal
non-scaling operating condition is occurring in real time, so that corrective actions can be taken
automatically (blowdown or feed more scale inhibitor when needed) to minimize and/or prevent
scaling of critical heat exchanger surfaces.
4. What differences will I see when I run 3D TRASAR Scale Control compared to what
I do now?
• Polymer feed level will be constant while the TRASAR level may vary.
• Conductivity set point may automatically be adjusted in response to changes in scaling stress
(one-drum programs).
• Conductivity set point or tagged polymer product set point may automatically be adjusted in
response to changes in scaling stress (two-drum programs).
System Parameters
Turbidity (NTU) <150
Background Fluorescence Correction 1
Fixed offset 30%
Online real-time 30%
HTI (hrs) Minimum of 14 2
Volume (gal) No limit
Blowdown Rate (gpm) No limit
System Volume/Recirc Rate Ratio (min) <20
Biocide Compatibility
Bleach Compatible
STA-BR-EX Compatible
ACTI-BROM Compatible
Towerbrom 960 See Bromine limits
Towerbrom 991 See Bromine limits
CI2 Gas Compatible
Cl02 Not Compatible3
BCDMH with EASIBROM Feeder See Bromine limits
Stabilized Bleach Compatible
Isothiazolinone Compatible
Glutaraldehyde Compatible
90005 Compatible at 160 ppm 4
DBNPA Compatible
Upper limit of halogen concentration-
impact on polymer/TRASAR fluorescence
Chlorine (FRC)5 <1.5
Bromine (TRO)5 <5
Other Variables
Iron (ppm) <10
SO2 None
Sulfides (ppm) None
Copper (ppm) <40
1
Applies to operation with 10 ppm active polymer- operation with less polymer requires
commensurately lower background level limits.
2
Inhibitor chemical stability should be considered as a potential limiting factor in long HTI
(e.g. >~200 hr) operation.
3
Status may change; Look in Optimizer or call 630-305-CHEM.
4
If foaming occurs on application, loss of TRASAR 2 (PTSA) or polymer may be observed.
Follow 90005 use recommendations to avoid foaming.
5
Corresponds to a 10% loss of fluorescence when exposed to halogen for 24 hrs.
8. What 3D TRASAR Products are required for 3D TRASAR Scale Control to work?
3D TRASAR Scale control programs require both an inert TRASAR product and a T
agged Polymer
product. Two different programs are available:
• One-drum programs:
The product contains both TRASAR (also known as PTSA) and Tagged polymer.
• Two-drum programs:
Product 1 contains TRASAR (also known as PTSA).
Product 2 contains Tagged Polymer
For a complete list of available 3D TRASAR products, please refer to the 3D TRASAR product
selection guide.
10. How does the 3D TRASAR 5500 controller detect "active" polymer?
A new component of 3D TRASAR products is a "tagged" polymer, either Nalco's tagged PRISM
polymer or tagged High Stress Polymer (HSP). The 3D TRASAR 5500 controller measures the
fluorescent intensity of the "tag" to determine the concentration of residual dispersant polymer
.
12. I don't feel comfortable allowing the 3D TRASAR controller to adjust the
conductivity setpoint. Must I allow 3D TRASAR to do this, or can I simply establish a
set point as I always have?
It is acceptable to control system Conductivity based on a user-defined Conductivity setpoint.
Setpoints can be selected during configuration and changed at any time. However , to realize the
benefits of 3D TRASAR Scale Control, the controller should be allowed to automatically adjust the
setpoint in order to effectively maximize cycles of concentration to minimize operating cost, or
reduce the cycles as necessary to prevent scale, corrosion and fouling.
13. Can 3D TRASAR Scale Control be operated in Monitoring Mode only (monitoring
Nalco Scale Index, but without Setpoint adjustments)?
Yes. 3D TRASAR Scale Control can be operated in the Monitoring Mode to monitor the Nalco
Scale Index (NSI) without setpoint adjustments.
For 3D TRASAR Scale Monitoring Mode, the NSI will be calculated upon the completion of the
Learning Period. The setpoint adjustment limits are set to zero.
14. Can I compare the Relative Rate of Polymer Consumption with other commonly
used indices, such as Calcite Saturation Ratio?
Yes. The scaling stress or the Relative Rate of Polymer Consumption, K, can be compared to other
commonly used indices, such as Langelier Saturation Index (LSI), the Ryznar Stability Index (RSI)
and the Pukorius (Practical) Scale Index (PSI) as practical predictors of calcium carbonate forma-
tion. The scaling stress or the Relative Rate of Polymer Consumption, K, can also be compared to
computer saturation models such as the French Creek Water Cycle or the Calcite Saturation Ratio.
In addition, the scaling stress or the Relative Rate of Polymer Consumption, K, correlates very
closely with the difference between the magnesium and calcium cycles (or Mg-Ca cycle split).
However, unlike the "snap shots" provided by grab sample analyses for obtaining LSI, RSI, PSI,
Calcite Saturation Ratio, and Mg-Ca cycle split, 3D TRASAR Scale Control evaluates the scale-
forming tendency of a cooling system continuously by measuring the scaling stress or the Relative
Rate of Polymer Consumption and takes corrective actions in a timely manner before scale can foul
the system.
180
Relative Rate of Polymer Consumption or Scaling Stress
160
140
120
100
80
60
40
20
0
20 30 40 50 60 70 80 90 100 110 120 130 140
17. 3D TRASAR Scale Control has been operational for a while. Why is there no NSI
value appearing on the 3D TRASAR controller display?
• 3D TRASAR Scale Control may still be in the Learning Period.
-After a Power Cycle, both the Polymer Consumption and NSI will display as "-----".
-After a Power Cycle, the Polymer consumption reading will be redisplayed at the end of the
next cycle whether or not the Learning Period has been completed.
-After a Power Cycle, the NSI reading will be redisplayed at the end of the next cycle only if
the Learning Period has been completed.
18. The LCD was showing polymer consumption. Why is it now showing "-----" on
LCD?
• After a Power Cycle, the Polymer Consumption (and NSI) will display as "-----". The Polymer
Consumption reading will be redisplayed at the end of next cycle.
• If you just performed a Relearn, during the very first cycle the polymer consumption will be dis-
played as "-----". The polymer consumption will be displayed/recorded at the end of each cycle.
For Two-Drum, the polymer consumption, TP%, is calculated by integrating the Relative Rate of
Polymer Consumption with respect to time.
20. I was feeding 23265 and now feed 3DT265. What are 3DT265-i and 3DT265-a
readings on the LCD display ?
When feeding one-drum 3D TRASAR Product, in this example, 3DT265:
• 3DT265-i is the concentration of the inert component or TRASAR, as ppm of product.
• 3DT265-a is the concentration of the active polymer, as ppm of product.
21. For one-drum programs, why does the -a reading not vary with polymer
consumption during tagged polymer control?
The TRASAR concentration will vary with time depending on the stress level. As the polymer
consumption changes, the product feed compensates to maintain the tagged polymer at the
setpoint (Product-a). Only the polymer consumption number, and the TRASAR level will vary with
system stress changes when the product feed is controlled by tagged polymer (Product-a).
22. (For one-drum program) My "-a reading" is higher than my "-i reading". Is this
correct?
The cooling system is usually operated under some level of scaling stress. Asa result, there will be
some polymer consumption (TP% > 0) and the "-a reading" should be lower than the "-i reading".
However, when the system stress is very low, sometimes the "-a reading" can be slightly greater
than the "-i reading".
If the "-a reading" is significantly greater than the "-i reading", it is also possible that:
22. (For one-drum program) My "-a reading" is higher than my "-i reading". Is this
correct? (con’t)
To determine the root cause, perform the following S0940 Check - it verifies fluorometer operation
and helps determine if recalibration is required:
• Rz should equal 20
• Rs should equal 40
6. If recorded values are close to the results obtained above, the Fluorometer is working
correctly and the calibration is good. Otherwise - recalibrate.
7. Use only DI water for calibration. Do not use make-up or other water source for calibration,
as this will result in an erroneous calibration curve.
If Calibration check is okay, check if the entered background fluorescence is correct. Collect/send
water samples for background analyses if necessary.
• After a power cycle or after the power is turned on, if Product level(s) > lower deadband, there is a
wait interval during which the blowdown is shut off.
• In case of running NSI control, the set point may have been adjusted to a higher value.
Note: Stabilized Chlorine, aka N-2818 and N-7345 CANNOT be used. If using Chlorine Dioxide,
call 630-305-CHEM.
NOTE: 3D TRASAR Bio-Control cannot be run in ‘Monitoring Mode’. For NBI numbers to
be meaningful, 3D TRASAR Bio-Control MUST be in control of the oxidant feed pump (e.g.
! no external ‘slug feed’ or external control of oxidant feed). During controller based slug
feed of oxidizing biocide the NBI reverts to –999 since the bio-reporter is not in control of
the Oxidant during that period.
Slug feeds can be set to occur on regular schedule or they can be individually initiated.
Scheduled Slug Feeds: Use the 3D TRASAR Configurator slug feed screen to define the
schedule for feed. We do not recommend more than one slug feed per week. If more
frequent slugs are desired, use the 3D TRASAR Bio-Shock mode.
Manual Slug Feeds: There are two methods of manually initiating slug feeds to the
system:
1. Controller Configurator: Initiate slug from the OxBio Slug Feed screen after
connecting to the controller either directly or via a phone line.
2. Controller Keypad: To initiate a manual slug feed from the controller keypad
press <Actions> and then <Slug Feed Biocide>. Choose:
<Hand Feed Ox> If biocide is applied with pumps or means not controlled by
the controller
<Relay Feed Oxidizing Biocide> In the relay feed option, the controller
applies the oxidizing biocide for the slug.
Important! 3DTBR10 is formulated with concentrated sulfuric acid. Use chemical feed
! equipment compatible for feeding concentrated sulfuric acid into water streams.
• Connect feed pump to control relay 3 (see Section 2 for wiring details).
• Plumb pump suction tube to bio-reporter liquid container and ensure pump is primed.
• Do not feed bio-reporter directly upstream of skid/fluorometer sample stream inlet.
• Do not feed bio-reporter into a low flow/mix or dead-leg portion of system.
• Do not feed bio-reporter into a makeup water stream if flow is intermittent.
• Return liquid pump discharge directly back to tower basin or tower return header.
• Manually turn on pump from the keypad to verify operation (see Section 4.10.8 for manual
relay control).
• Set pump feed rate between 0.1 and 0.2 gpd. Make sure pump does not lose prime at this
low rate.
NOTE: Verify oxidant residual is below 0.1 ppm TRH before proceeding any further! Use
! DPD Total Colorimetric Test (P/N 460-S0461.87) with the Nalco DR/890 or DR/2800.
STEP 3 - Using the Configurator, set up the controller for 3D TRASAR Bio-Control or 3D TRASAR
Bio-Shock mode.
• In the System Information window, enter the system volume (system volume needs to be
as accurate as possible).
• In the Control Settings window, set the oxidizing biocide control method to “3D TRASAR
Bio-Control” or “3D TRASAR Bio-Shock Mode”.
• If using scheduled Ox slugs with 3D TRASAR Bio-Control or 3D TRASAR Bio-Shock
mode, set up the slug parameters (prebleed, feed, post feed and recovery) in the Ox slug
window.
• If using Ox slugs or 3D TRASAR Bio-Shock mode, make sure the ORP probe is clean and
calibrated.
• Select the desired bio-reporter and indicate if the oxidant feed pump is optimally sized,
oversized or undersized for the system.
• In the Alarm Settings window, select the appropriate Alarm Pump Duty for the oxidizing
biocide (see Section 4.5.4).
• Upload the modified configuration to the controller.
• Check to make sure the Configurator program asked for a ‘Reinitialization or Reboot’ of
the Controller. If not, either power cycle the Controller using the on/off switch or Reboot
from Configurator program.
STEP 4 - If using a pellet feeder and the system volume exceeds 160,000 gal, deliver the required
number of bio-reporter Quick Dose Pouches to the cooling system (see Section 6.7).
! WARNING: ALWAYS TURN ‘OFF’ POWER BEFORE OPENING PELLET FEEDER LID!
NOTE: Wear appropriate safety equipment when handling the pellets or filling the feeder
.
! Refer to MSDS supplied with pellets for appropriate handling.
NOTE: It is very important that the feed hopper remain dry. Moisture or water in the hopper
! can cause pellet degradation and possible jamming.
If pouches are unavailable and your system size exceeds 160,000 gal use the following table to
calculate startup dosages. It is imperative that the startup pellets be placed in a region of good flow
so that they dissolve quickly (do not dump into a slow flowing area in the tower basin).
Periodically check the inside of the pellet feeder. If excessively dirty it will require a cleaning. If kept
dry, cleaning is rarely required.
1. Manually remove the pellets from the feeder and store them temporarily in a clean DRY
location.
2. Use a soft cloth to wipe clean any debris from the vessel sides, bottom to top.
3. Do not use a damp rag or introduce any moisture to the vessel.
When clean refill with the old pellets and manually feed a pellet to test for correct operation.
Refer to the bio-reporter MSDS sheet for safe handling procedures and required safety
! gear.
Contact with any moisture will cause the pellets to impart an intense blue coloring. This is
! water-soluble and can be washed away with water.
Replace Pellet Shield - Section 6.9.3 Indicated by alarm, pellet feeder jamming, or
visual inspection.
STEP 1 - Turn pellet feeder power “Off”. See Figure PF2 for location of pellet feeder power
switch.
STEP 2 - Locate the clear tube assembly that connects the pellet feeder body to the
injection valve. See Figure PF1.
Exit Tube
Clear
Tube
Exit Tube
Power
Switch
STEP 3 - Directly above the clear tube assembly is the exit tube. Remove the clear tube
from the coupling immediately above the ball valve. No tools are required, it is a
snug but unglued connection.
STEP 4 - Insert the tube brush (provided with the pellet feeder)and clean the exit tube.
DO NOT USE WATER. Use the brush ONLY. The LED is 1" to 2" up from the
bottom of the tube.
STEP 7 - Manually feed a pellet at the 3D TRASAR controller from the pellet feeder diagnostic
screen. Press ‘0’ to feed a pellet.
STEP 1 - Turn off power to the pellet feeder (switch located below hopper see
Figure PF2).
STEP 2 - Remove the upper plumbing assembly to access the top of the ball valve. This is
done by removing the upper union fitting and the pipe strap (see Figure PF3).
Pipe Upper
Strap Plumbing
Assembly
Plumb
Upper
to drain
Union
Figure PF3
Note: Under normal operation, the side drain port should be plumbed to drain, or connected to the
bio-reporter Liquid Overflow Container (P/N 060-TR5350.88).
Adjustment
Nut
Figure PF4
STEP 4 - The unit can be manually activated to determine if excessive leaking occurs. Small
“burps” of water during rotation are normal. If excessive leaking persists, tighten
further.
STEP 6 - Manually feed a pellet at the 3D TRASAR controller from the pellet feeder diagnostic
screen. Press ‘0’ to feed a pellet.
STEP 7 - At the pellet feeder diagnostic screen, verify that B‘ all’ value is in the range of
400 to 600. If not, repeat steps until satisfactory.
Figure PF7
Pellet Feeder Shield Kit
Wearing a minimum of rubber gloves and dust mask, clean out any product from the Pellet
Feeder hopper. Do not use water! Contact with the product will cause a blue “stain”.Avoid
! contact. The stain is not permanent and will wash away.
STEP 4 - Loosen the four screws and washers holding the tabs of the pellet
shield. Note the orientation to the ‘Exit Tube’ (shown in drawings).
STEP 5 - Remove the pellet shield, rotor retainer, and screws/washers as a unit.
Figure PF8
Figure PF9
STEP 6 - Remove the screws and washers from the tabs of the removed shield.
STEP 7 - Remove the used pellet shield from the rotor retainer and replace it with a new one.
STEP 9 - If the area around the rotor is full of powder, clean it out to allow the rotor retainer to
fit properly.
STEP 10 - Insert the entire assembly into the feeder. Be sure to properly align the pellet shield
with the exit tube - the pellet shield should cover the exit tube as shown in the
previous figures. Be sure the washers are in place.
STEP 11 - Secure the assembly with the screws. Do not over-tighten as the plastic can be
damaged.
Planktonic counts measured by dipslides or DMAshould be between 103- 105 cfu/ml. Since
3D TRASAR Bio-Control responds to the total system microbial activity, it does not necessarily feed
oxidizing biocide in response to the floating population measured by water samples and dip-slides.
Under 3D TRASAR Bio-Control, coupon swabs give population counts of108 cfu/sq cm or less.
• The system is NOT an open recirculating cooling water system (e.g. once through, closed
loop, etc.).
• The system has highly halogenated make-up water which is added to the tower in incre-
ments greater than 5% of the system volume.
• Bleach (as opposed to bromine) is used with frequent or persistent ammonia contamina-
tion.
• System has high background fluorescence (often associated with yellow tower water).
Before bio-reporter is added to the system, check diagnostic screen and see if Rz or Rf
reading (background) is greater then 8 ppb. To check the diagnostic data press Informa-
tion key, select Diagnostic Data, select Fluorometer Data, read Rz and Rf (number is in
ppb).
• The amount/timing of oxidizing biocide fed to the system is constrained to be below the
amount necessary for adequate control. An example: power plants where the time of
oxidizing biocide feed is limited by regulatory constraints. Another example: the amount of
halogen fed is limited by how much money is allocated towards microbial control.
System Parameters
Turbidity (NTU) <150
HTI (hrs) 6-400
Volume (gal) No limit1
Blowdown Rate (gpm) No limit1
System Volume/Recirc Rate Ratio (min) <20
Biocide Compatibility
Bleach Compatible
Stabrex Compatible
Actibrom Compatible
Towerbrom 960 Compatible2
Towerbrom 991 Compatible
CI2 Gas Compatible3
Cl02 Not Compatible3
BCDMH with E ASIBROM Feeder Compatible3
Stabilized Bleach Not Compatible
Isothiazolinone Compatible
Glutaraldehyde Compatible
DBNPA Compatible
90005 Compatible
Ammonia Levels
with B leach (ppm) <20
with Stabrex (ppm) <100
with A ctibrom (ppm) <300
Other Variables
Iron (ppm) N/A 4
SO2 None
Sulfides (ppm) None
Copper (ppm) <300
1
Systems with volume greater than 2,000,000 gallons and blowdown rate greater than 1,200 gpm need to use
3DTBR10.
2
Compatible only with 3D TRASAR Bio-Shock mode.
3
Field tests in progress or trial sites needed. Call 630-305-CHEM.
4
0.5 ppm as Fe2+, but cooling towers are oxidizing environments and Fe2+ is not generally found.
Does the Make-Up water come in big slugs (system water volume changes by >10% due to Make-
Up valve control dead band)? If so, this system may not work with 3D TRASAR Bio-Control. Use of
3D TRASAR Bio-Control in these conditions can cause blue coloration in the water
.
We have run systems with Make Up water that have Total Residual Halogen of 3 ppm successfully
with 3D TRASAR Bio-Control, where the above two conditions did not apply
.
If oxidant feed is controlled externally, it is likely that oxidant feed is not in “micro slugs”. This
means the ratio drops precipitously when excess oxidant is fed. The 3D TRASAR controller then
tries to compensate for bio-reporter loss by feeding more bio-reporter which a) lowers the ratio b)
can turn the tower blue because of reaction with excess oxidant.
The value of the Nalco Bio-Index itself is meaningless in this case because the ratio is not a func-
tion of microbial activity, but of oxidant levels in the tower. Another way of saying it is that for NBI
numbers to be meaningful, 3D TRASAR Bio-Control MUST be in control of when the oxidant pump
comes on (ratio has exceeded pre-set limit) and when it shuts off (oxidant feed is enough to kill the
microbes).
If, prior to 3D TRASAR Bio-Control, the customer was underdosing biocide, 3D TRASAR Bio-
Control will increase the dosage. On the other hand, in systems with excessive biocide usage, it
may significantly reduce biocide usage.
Systems that maintained oxidizing biocide at a set halogen residual either by ORP control, or by
Hach Cl-17 control or even by manually-adjusted continuous biocide feed, have typically reduced
their oxidizing biocide usage under 3D TRASAR Bio-Control, sometimes dramatically.
Systems that were previously slug-dosed once per day to once per week typically record an in-
crease in oxidizing biocide usage under 3D TRASAR Bio-Control.
Technical Help 521-OM0108.88 6-22
(630) 305-CHEM Copyright 2009 Nalco Company
3D TRASAR 5500 - Installation & Operation Manual
There are no Biocides that are specific for Legionella. 3D TRASAR Bio-Control applies conven-
tional oxidizing biocides
Legionella are only a small fraction of microbial population in cooling tower. 3D TRASAR Bio-
Control monitors the general planktonic and sessile microbial activity in the tower and applies
biocide as a response. 3D TRASAR Bio-Control does NOT apply Biocides specifically for
Legionella.
For systems where control of Legionella and health related risks is a water treatment program goal,
Nalco Environmental Hygiene Services should be contacted for risk assessment and management.
Numerous case studies have shown that 3D TRASAR Bio-Control is effective in controlling biofilm
and general microbial populations. SRB’s are typically found under deposits and biofilm. Control of
biofilm can help control SRB proliferation in the system.
SRB’s are only a small fraction of microbial population in cooling tower. 3D TRASAR Bio-Control
monitors the general planktonic and sessile microbial activity in the tower and applies oxidizing
biocide as a response. 3D TRASAR Bio-Control does NOT apply Biocides specifically for SRB’s.
SRB’s are typically found under deposits, which are inaccessible to the bio-reporter and oxidizing
biocide. It is effective to use a biodetergent and a non-oxidizing biocide to better disperse and treat
the deposits in the system.
If high SRB counts are encountered in the system consider supplementing 3D TRASAR Bio-Control
with regular Oxidizing biocide slugs. You may also consider using 3D TRASAR Bio-Shock mode for
a period in order to bring the system under good microbial control.
Even though the bio-reporter reacts with and responds to algae, the microbial activity attributable to
algae growth usually represents only a fraction of the total system microbial activity. For this
reason, 3D TRASAR Bio-Control may not apply oxidizing biocide specifically in response to, or at a
dosage sufficient to control, algae growth.
3D TRASAR Bio-Control does not maintain high residuals of halogen in a tower. If algae is a known
problem and over-halogenation has been the historical method of algal control, it is crucial that an
algaecide be routinely used in conjunction with the oxidizing biocide under bio-reporter control.
If algaecide use is not feasible, another option is to use regular scheduled oxidizing biocide slugs
overlaid on 3D TRASAR Bio-Control.
3D TRASAR Bio-Shock mode, the all slug mode of oxidizing biocide application with bio-reporter
control may be also used in systems where algae is a problem
Algae grows in semi-wetted areas exposed to light. Chlorine and bromine are not particularly
effective under these conditions. Using these halogens at high levels in the recirculating water
results in a vapor pressure of halogen in the tower decks resulting in algae kill. Since 3D TRASAR
Bio-Control only maintains very small TRH levels in the water, algae control by this mechanism is
not active.
15. What key issues I should be aware of before I decide to use 3D TRASAR Bio-
Control?
• 3D TRASAR Bio-Control should be used in conjunction with an appropriate oxidizing
biocide, a non-oxidizer and a bio-dispersant. Additionally, if presence of algae has been a
problem in the past and was handled through high TRH levels in the tower, an algaecide such as
N-90005 must be used.
• 3D TRASAR Bio-Control will impart a slight blue color to the tower water. The effect will be ampli-
fied as the depth of the water in the basin is increased. This is normal.
• 3D TRASAR Bio-Control may or may not reduce the oxidant usage in the system. It depends
upon whether the system was in good control or not before 3D TRASAR Bio-Control.
• 3D TRASAR Bio-Control is not the same as maintaining a fixed halogen residual. The measured
TRH levels depend upon the system microbiological activity and demand. In fact, measured TRH
levels may be zero during a significant portion.
15. What key issues I should be aware of before I decide to use 3D TRASAR Bio-
Control? (con’t)
• 3D TRASAR Bio-Control applies oxidizing biocides based on the measured total microbiological
activity — planktonic and sessile. The bulk of the microbiological activity in the system is sessile
and 3D TRASAR Bio-Control responds to this measured activity with oxidizing biocide. 3D
TRASAR Bio-Control will maintain planktonic counts around 104 cfu/ml for the most part. But
injured and inhibited microbes (inactive) will be nursed back to health on dip slides or petrifilm and
will show up in the planktonic counts. Planktonic population fluctuations between 10 3
-105 cfu/ml
should be expected.
• Note that there will always be systems where microbiological control is not a chemical treatment
issue. There are mechanical, operational and chemical parameters (e.g. dead legs, severely
fouled tower basins, continuous process leaks etc.) that can make bio-control difficult. Good
microbiological control may not be possible where these issues are not separately addressed.
Bio-Reporter Control: Bio-Reporter is fed into the system as a liquid or a pellet based on the mea-
sured concentration of the Rz signal. The liquid bio-reporter is fed based on dead band control. For
pellets, the decision is made every 20 minutes. The number of pellets that will be fed at a time is a
function of the pellet concentration, entered system volume and historical system response.
Fluorometer signals: Rz and Rf intensities are collected every 5 sec and a ratio value (Rf/Rz) is
calculated for each data point. The data collected is validated to ensure that the fluorescence data
is above a lower threshold value and below saturation values for the fluorometer .
Data Processing: A quadratic fit to the collected ratio values is calculated at the end of each 20 min
cycle. A goodness of fit criteria (chi-square) is used to evaluate if the data is too noisy or the fit is
unsuitable. The oxidizing biocide pump is turned on/off depending on the value of the calculated
ratio and the slope of the fit.
Control Band (3D TRASAR Bio-Shock): When the ratio reaches a set point, bio-reporter feed into
the system is suspended. The system slugs oxidant into the tower. The duration of the slug feed is
set by the user. During the Ox-slug the oxidizing biocide is maintained under ORP control with a
preset setpoint and range. The oxidant slug can also be simple timer based.After the slug is com-
pleted the system waits for the ORP readings to dissipate to preslug values. Bio-Reporter is then
again added to the system and the ratio is monitored.
Control Loop: After 20min, the response of the system is again evaluated by processing the ratio
and a decision is made again on whether or not to feed oxidant and for how long.
Ox Slugs: Oxidizing biocide slugs may be applied to the system when it is under normal 3D
TRASAR Bio-Control. These slugs may be scheduled or be manually triggered from the controller.
During the Ox slug bio-reporter feed is suspended. The duration of the slug feed is set by the user
.
During the Ox-slug the oxidizing biocide is maintained under ORP control with a preset setpoint and
range. The oxidant slug can also be simple timer based.After the slug is completed the system
waits for the ORP readings to dissipate to preslug values. Bio-Reporter is then again added to the
system and normal 3D TRASAR Bio-Control resumes. We recommend using the slug feed feature
once a week in order to vary the system stresses for good microbial control
Since the oxidant slug condition is defined by the user, the actual oxidant applied can be adjusted.
If an increased oxidant dosage is desired, the ORP set point and/or the slug duration can be in-
creased. Please note that there is NO oxidant feed between slugs. Under classic 3D TRASAR Bio-
Control, the oxidant dosage cannot be adjusted.
3D TRASAR Bio-Shock Mode requires a functional ORP probe to be in place and recalibrated.
The ORP probe is used to control the oxidant slug and monitor the system in the recovery period
following the slug. Once the ORP signal has dropped to within 5% of the pre-slug levels, the
bio-reporter is redosed. If the ORP signal does not fall back to within 5% of the pre slug level within
24 hours (default value), the system will trigger a “slug recovery failure” alarm, and the oxidant will
be dosed based on the failsafe settings. When the alarm is cleared (either by acknowledging the
alarm on the 3D TRASAR controller screen or through a power cycle), the system will return to
normal operation and resume bio-reporter feed.
The figure below shows the bio-reporter ratio value during 3D TRASAR Bio-Shock mode. When it
reaches the setpoint the Oxidant relay is triggered (to 129). The Oxidant relay remains under ORP
control for the preset duration. Thereafter when the ORP signals returns to preslug values and the
recovery period is complete, bio-reporter is again dosed into the system and the ratio is monitored.
ATP measurements cannot be done on-line. The bio-reporter monitors the microbiological activity
continuously, on-line and in real time.
ATP measurements cannot be used to control the biocide pump. The bio-reporter technology is
used for continuous measurement and control of microbial activity.
ATP can survive for a long time after the micro-organism itself is dead. Therefore the readings
obtained from ATP analysis can give an incorrect picture of even the planktonic microbial activity of
a system.
21. How does 3D TRASAR Bio-Control differ from using dip slides or plate counts to
measure bio-activity?
3D TRASAR Bio-Control responds to the total microbial activity in the system, the floaters
(planktonics) and the slime (sessile, microbes attached to surfaces). The sessile population consti-
tutes >98% of the total microbial population and cause the problems associated with microbial
proliferation (under-deposit corrosion, impedance of heat transfer, odor, environments for pathogen
growth, slime). Dip slides and plate counts only measure the floating microbial population and miss
the bulk of the microbes.
3D TRASAR Bio-Control responds to microbial activity, not population. On the dip slides all mi-
crobes flourish and form colonies. In the system, injured or inhibited microbes are not active and
do not need biocide.
3D TRASAR Bio-Control responds to microbial activity instantaneously. Dip slides and plates need
24-48 hrs to develop. During that period 10’s of generations of microbes may flourish and repro-
duce in the system.
3D TRASAR Bio-Control measures and responds to microbial activity at the correct time with the
correct dosage of oxidizing biocide. This results in good control with efficient use of biocide.
ORP control is based on an electrochemical measurement and maintains the oxidation reduction
potential in the system between preset bounds. The ORP setpoint suitable for a particular system
is water chemistry and probe dependent. It’s a good idea to monitor the halogen residual in the
tower at a particular ORP and fine tune the settings. Maintenance and frequent calibration is critical
for good control.
Technical Help 521-OM0108.88 6-29
(630) 305-CHEM Copyright 2009 Nalco Company
3D TRASAR 5500 - Installation & Operation Manual
ORP is an electrochemical measurement and the readings from a probe can drift significantly in
short periods. It is important to keep the ORP probe clean and well-calibrated. The bio-reporter is
an optical measurement, and maintenance and drift are not problematic.
For ORP control the user chooses the control setpoint. For bio-reporter control, the control band is
preset depending on the oxidizing biocide chosen.
ORP control applies oxidizing biocide so that the system maintains a set ORP. In bio-reporter
control, oxidizing biocide is applied based on microbial demand and therefore biocide usage is
efficient.
Since oxidizing biocide is dosed based on demand, corrosion is often reduced significantly under
Bio-Control. Therefore is typically a reduction in delignification of wood structures, steel stress
corrosion and cracking, AOX formation under this control method.
When sufficient oxidizing biocide has been added to the system and the microbial activity is
brought into control, the excess oxidant interacts with the reacted bio-reporter. The controller stops
oxidizing biocide application.
Because both the microbial activity and the biocide dosing is continuously monitored it is no longer
necessary to maintain a measurable excess oxidizing biocide in the system (residual halogen).
Under 3D TRASAR Bio-Control, the halogen residual may be between 0.8 ppm to 0.02 ppm as
measured by a total halogen DPD test.
It is difficult to ensure that the bleach is consistently overfed especially during process leaks or
when the make-up water becomes contaminated (e.g. with ammonia). 3D TRASAR Bio-Control
ensures that the correct amount of bleach is applied at the correct time.
3D TRASAR will alarm if operational conditions such as an air-locked oxidizing biocide pump or
empty oxidizing biocide porta-feeds occur. Thus issues can be dealt with in a more timely manner
and the system does not develop severe fouling problems in the interim.
Continuous overfeed of bleach can result in high levels ofAOX formation in the tower water leading
to issues with discharge of the blowdown.
Bleach overfeed can result in SS pitting and stress corrosion cracking of equipment.
Finally, even though bleach has a low $/lb cost, over feed can result in unnecessary high costs.
1,000 3.0
Before 3D Bio-Control After 3D Bio-Control
900
2.5
800
500 1.5
400
1.0
300
200
0.5
100
- -
Time (each division = 1 day)
Figure: ORP and copper corrosion rates are measured by 3D TRASAR. 3D TRASAR
Bio-Control is superior to ORP control because it 1) reduces the variation in oxidant
concentration, 2) reduces the average oxidant concentration and 3) reduces copper
corrosion rates by eliminating high ORP levels.
25. Can you give me an example where 3D TRASAR Bio-Control varied the Oxidant
feed based on system demand?
In this tower oxidizing biocide is STA-BR-EX. Tower is under 3D TRASAR Bio-Control. On 8/23/04
the tower is slugged with 6 ppm of N73550 (biodetergent). Entire system (deadlegs, unused
towers, unused chillers) is opened up to detergent. 100 ppm of N7330 is added to the tower 1 hr
later. As untreated and sloughed up biofilm is exposed, the Nalco Bio-Index increases and the
duration and frequency of the oxidizing biocide feed (Ox pump On time) is increased. The Nalco
Bio-Index has a double-humped profile (second peak on 08/24/04) due to the detergent continuing
to slough off deposits while the N7330 concentration is reduced by consumption and blowdown.
System is finally brought back to normal levels of STA-BR-EX feed by 08/25/04 at which time it is
treated to 50 ppm of N90005 (quaternary amine algaecide). The algaecide is also a surfactant and
a slow acting bactericide. There is an initial increase in Nalco Bio-Index followed by a 21 hr period
during which no STA-BR-EX is fed into the tower. Compare this to normal operation period (08/20/
04-08/22/04). The consecutive application of two non-oxidizers led to low microbiological activity
and consequently a 21 hr period of no oxidant feed into the system. Planktonic counts taken on
08/26/04 show good microbial control.
Ox_pump_ on_time
300 BF_ratio_ index 30
N90005 slug 25
Ox Pump on sec/600s
250
No Ox feed20
3D Bio Index
200 N73550 & for 21 hrs 15
N7330 slug 10
150
5
100 0
-5
50
-10
0 -15
8/20/04 8/22/04 8/24/04 8/26/04 8/28/04
For a complete list of available 3D TRASAR products, please contact your Nalco Sales Represen-
tative.
30. Should I slug feed my tower with oxidizing Biocide before starting 3D TRASAR
Bio-Control?
No. You should wait till the halogen levels in the tower dissipate to less than 0.1 ppm total residual
as measured by the Hach DPD test before starting 3D TRASAR Bio-Control, otherwise your tower
water will turn blue.
31. What will happen if I initiate 3D TRASAR Bio-Control without waiting for TRH
levels to be <0.1 ppm?
If you initiate 3D TRASAR Bio-Control without waiting for the halogen levels to dissipate to <0.1
ppm total residual halogen as measured by the Hach DPD test, there is a potential for the tower
water to turn intensely blue. Large quantities of the bio-reporter product may be wasted.
Some towers will always register a total residual >0.1 ppm total halogen, even when the oxidizing
biocide feed into the tower has been discontinued for 24 hours. In these towers, it may be OK to
start 3D TRASAR Bio-Control without waiting for thetotal halogen residuals to dissipate to
<0.1ppm provided at least 24 hours have elapsed since discontinuation of oxidizing biocide feed.
Ensure there is no free halogen residual in the tower water and call630-305-CHEM before startup.
A minimum of one hour of data has to be acquired before a calculated value is shown. If the sys-
tem has just been dosed with bio-reporter, it will take between 4-24 hours (depending on microbial
activity) for the Ratio_Value to reach the control Ratio set point. During this period, oxidizing bio-
cide will not be applied to the system. If after 24 hours oxidizing biocide has not been fed into the
system please call 630-305-CHEM.
When you wish to follow the 3D TRASAR Bio-Control algorithm and verify correct operation the
easiest thing to do is to download the 3D TRASAR Bio-Control data file (through data download
screen). The file can be opened in Excel and the Nalco Bio-Index can be reviewed along with the
‘ox pump on time’. The ox pump on time is the length of time (in seconds) the 3D TRASAR Bio-
Control turned the oxidant pump on each 20 minute decision interval.
You can also verify that the oxidizing biocide relay has been feeding biocide by checking the [Calcs]
portion of the ‘General Data’ log in column Z [Ox Bio Relay Duty %] . You will see a number (per-
centage) listed if the ox_biocide relay has come on in the previous 24-hour period.
The Nalco Bio-Index value should be between -10 and +10 during normal operation of 3D TRASAR
Bio-Control after the initial 4-24 hour period has elapsed.
33. How do I determine if the Rz concentration in the system has reached 20 ppb?
• Check the 3D TRASAR controller display.
Press Information key, select Diagnostic Data, select Fluorometer Data, select Rz
(ppb).
• Check the 3D TRASAR configurator.
Connect to the controller. Once connection is established, the Controller menu will appear.
Select View Current Readings, Data Tab. Under System Performance box locate the
Bio-reporter value in ppb.
34. Can I trim-feed oxidizing biocide with a separate pump at low level to
continuously maintain a halogen residual while on 3D TRASAR Bio-Control?
Only if the oxidizing biocide is being applied to the Make-Up waterAND the conditions that prevent
3D TRASAR Bio-Control use in systems with high halogen levels in Make Up water do not apply .
If oxidant is fed by external (not 3D TRASAR Bio-Control) control into the tower water, it is likely
that there will be overfeed during times of low microbial activity resulting in degradation of the bio-
reporter and BioProduct signals. To compensate, more dye will be added to the system and the
water will become colored.
34. Can I trim-feed oxidizing biocide with a separate pump at low level to
continuously maintain a halogen residual while on 3D TRASAR Bio-Control? (con’t)
Continuous addition of Oxidizing biocide and bio-reporter into the system can result in the Ratio
staying artificially low (below set point) and no true measure of microbial activity (Nalco Bio-Index)
will be obtained.
Nalco Bio-Index can only work by applying mini slugs of oxidizing biocide. The extent and timing of
the slug has to be controlled by the controller.
35. Can you make 3D TRASAR Bio-Control feed more oxidant into the system?
No. The amount of oxidizing biocide added depends upon the measured microbial activity in the
system. This cannot be externally modulated or adjusted. However, the scheduled slug option
can be used to periodically supplement 3D TRASAR Bio-Control.
3D TRASAR Bio-Shock mode, an all slug mode of Biocide feed is an alternative method of applying
oxidizing biocide to the system. The bio-reporter monitors the microbial activity and when it reaches
a preset value, an Ox slug is triggered. By setting the duration and ORP level of the slug, the user
can modify the amount of biocide entering the system.
36. Can you make 3D TRASAR Bio-Control feed less oxidant into the system?
No. The amount of oxidizing biocide added depends upon the measured microbial activity in the
system. This cannot be externally modulated or adjusted.
37. How responsive is 3D TRASAR Bio-Control to changes in the bio-population?
3D TRASAR Bio-Control monitors the fluorescence signal from the reacted and unreacted bio-
reporter every 5 seconds. It assesses the data-stream every 20 minutes and makes a decision
regarding whether to turn the oxidizing pump on or not.
If there is a process leak into the system that either increases the amount of nutrient available to
the microbial populations or decreases the efficacy of the biocide (sour leak, ammonia etc) 3D
TRASAR Bio-Control will respond with increased biocide feed within 20 minutes of the measured
event. There may be a delay in response in systems with long recirculation times. The Nalco Bio-
Index will take longer to show an increase because it is a 6 hour average measurement.
Typically process leaks affecting the microbiological population have been responded to at or
before the time that other effects have been observed in the system such as pH changes, turbidity
changes, conductivity changes or ORP changes.
38. When using 3D TRASAR Bio-Control, what halogen residuals can be expected?
The amount of halogen residual measured in the system water will depend on microbial activity in
the system. This in turn can depend upon factors such as ambient temperature, process leaks,
system design etc.
The highest level of halogen (total residual) observed in a cooling tower under 3D TRASAR Bio-
Control is 0.8 ppm. The lowest level observed is 0.0 (below detection limit of Hach DPD test).
During periods of moderate demand, total halogen residuals in a cooling tower can be expected to
be 0.1 - 0.4 ppm TRH. ORP levels at which control is maintained depends upon the oxidizing
biocide used (STA-BR-EX: 150-340 mV, Bleach or Actibrom: 270-340mV).
Variation in ORP (mV) and total halogen levels in a cooling tower under 3D TRASAR
Bio-Control.
In this graph, STA-BR-EX was the oxidizing biocide. On-line halogen was monitored with
a Hach Cl-17. Using 3D TRASAR Bio-Control, ORP varied between 180 - 320 mV and
the halogen levels varied between 0.0 - 0.4 ppm (total). 3D TRASAR Bio-Control adjusts
halogen dosage based on demand. There is thus seasonal and diurnal variations in the
concentration.
40. My dip slides show >10,000 cfu/ml. Does this mean 3D TRASAR Bio-Control does
not work in my system?
No. 3D TRASAR Bio-Control may be working well in your system. The bio-reporter monitors the
total system microbial activity: planktonic (floating) and sessile (biofilm). The biofilm constitutes the
vast majority of the microbial population in the system (>98%) and is unevenly distributed over the
system surfaces. Dip slides and normal DMA only monitor the floating population, which may not
represent the bulk of the microbial activity.
Planktonic counts measured by dip slides or DMAshould be between 103-105 cfu/ml. Swabs of
coupons in systems with 3D TRASAR Bio-Control give population counts of 107 cfu/sq cm or less.
Excursions outside this range may happen occasionally, especially after a detergent feed or the
opening of a dead-leg. If the problem persists, please call 630-305-CHEM.
It is important to implement a full microbial treatment plan for the system including use of a
biodetergent (to allow penetration of the bio-reporter and also the oxidizing biocide into the depos-
its) and a non-oxidizing biocide. High planktonic counts may be indicative of continuous inoculation
of the system from dispersing biofilm.
Since the system has just been dosed with the bio-reporter, the Ratio is low and it is below the
control band. As the microbes convert some of the bio-reporter the Ratio will come up to the set
point. The Nalco Bio-Index is calculated from the averaged value of the Ratio and is not a good
indicator of the Bioactivity in the system at system startup.
42. My Nalco Bio-Index has been reading -80 for the past week. Is there a problem?
There are several causes for a persistently low Nalco Bio-Index:
• Is your Makeup water highly halogenated? Are there other sources of oxidant feed into the
tower? Persistent very negative Nalco Bio-Index readings indicates that the bio-reporter is being
degraded by a separate oxidant feed. Plot the ORP(mV) data from the standard log and see how
it varies over time. Other sources of oxidant can be detected by the ORP being maintained or
going up with makeup water ingress.
• Your HTI may be too low for 3D TRASAR Bio-Control.
• Is there any other alarm condition on your controller? Other conditions may be causing the low
Nalco Bio-Index.
NBI represents the average value of the Rf/Rz Ratio with respect to the control band over the past
6 hours. Because it is an averaged value, the response time of this function is slow
.
The standard data log saves a snapshot of the state of the relay every 15minutes. The relay may
have been in between these logs. The calc (once-a-day) log correctly records the Oxidizing biocide
relay duty cycle for the past 24 hours. Check this number to determine if biocide has been applied
or not. The 3D Bio data also logs the Ox relay on times accurately.
If the system has just been dosed with bio-reporter, it will take between 4-24 hours (depending on
microbial activity) for the ratio to reach the control ratio set point. During this period, oxidizing
biocide will not be applied to the system. If, after 24 hours, oxidizing biocide has not been fed into
the system, please call the help-desk.
Do you have highly halogenated makeup water or other sources of oxidant feed into your system?
The tower can turn bluer than normal if any one of the following conditions is true:
48. Do I have to adjust the Pellet Feeder ball valve for leakage?
Over time, especially in systems where a large number of tablets are being fed (> 50 per day), the
seals may need to be adjusted. This is detailed in the manual. If the drain line is leaking when no
pellets are being fed, or if large amounts of water are being expelled during a feed, the seals
should be adjusted.
49. Why does my Pellet Feeder ball valve keep jamming with pellets?
This has happened in systems with no, low, or intermittent flow through the pellet feeder. The outlet
line from the pellet feeder should have no blockage. If the flow through the flow switch on the
controller goes below 0.5 gpm, the system will register a no-flow alarm and pellets will stop being
fed. However, if the feeder is plumbed to a separate line that becomes plugged, or if the flow is
throttled to less than 5 gpm but greater than 0.5 gpm, the feeder will continue to feed. This may
result in undissolved pellets below the ball valve. Please check the flow through the feeder prior to
beginning pellet feeding. Strainers before and in the pellet feeder may clog resulting in no-flow or low
flow conditions. A separate flow switch on the feeder line (with a 3 to 5 gpm cut-off) could be added
in series with the controller flow switch if intermittent flow is a problem.
Downloaded data is compatible with Vantage V100. Data is sent to the 3D TRASAR Web every
hour from the 3D TRASAR controller via e-mail.
7.5.1 3D TRASAR Web Set Up
3D TRASAR Web is accessed through Nalco's website (www.nalco.com). You must have a Nalco
Extranet Account to enter the site. Contact your Nalco representative for information on how to
obtain an extranet account.
The “Performance
Summary” page
provides a high-level
view of the state of
every controller
assigned to a user.
• Select File to Load - click Browse to enter the directory location and name of the 3D
TRASAR Data file to import.
• V100 Target Database - click Browse to select an existing Vantage V100 database
(*.mdb) or to specify a directory location for a new Vantage V100 database file (*.mdb) to
be created. The database name will default to be the customer name and location
assigned in the controller.
• Select System - there may be multiple systems in one database, select the appropriate
system for the data to be stored.
• Data Source Name - displays the Vantage V100 Data Source Name.
• Click Transfer Data
Note: Use Vantage V100 for more permanent data storage. T ransfered data will be ap-
! pended to the Vantage V100 database, and can be archived for long term storage.
The 3D TRASAR controller will only store up to 65 days worth of data.
STEP 1 – On the Reports form, choose the 3D TRASAR EZ Report then select the parameters and
the date (with optional statistics range) for data that you wish to include in your report.
STEP 2 – Select the graphs you wish to include in your report. Each row in the grid represents one
graph in the report.
STEP 2 – Select the appropriate 3D TRASAR Data Source and double-click the measurement in
the grid.
The Nalco Global Gateway provides advanced security including NA T, plus a built in firewall to
protect the controller from unauthorized access via the Internet. W
hen properly installed and config-
ured, the functionality of both the modem and local Ethernet port remain unchanged.
It is important to verify that the Nalco Global Gateway is right for your application before purchasing.
Verifying the site for the Nalco Global Gateway is similar to finding a place in an industrial environ-
ment to make a cell phone call. Even the best cell phone on the market is of little use without an
adequate, interference free, stable signal. Use the guidelines below to determine signal availability in
your area.
• If you have access to a GSM/GPRS cell phone, and can make a web/data connection from
the location where the controller will be installed, there is a very high probability that the
Nalco Global Gateway will work from that location.
• If you have a non-data GSM phone and can make a voice call from the location that the
controller will be installed, there is a good probability that the Nalco Global Gateway will work
from that location.
• If you cannot make a data connection or phone call, you will probably need theindoor/
outdoor high gain antenna option (P/N 060-TR5284.88, cable not included*). This usually
occurs when the controller is installed in basements, metal buildings, or next to electrical
equipment.
Detailed installation and configuration instructions are included with the unit. Contact your local
Nalco Representative for ordering information.
Section 8: Maintenance
The 3D TRASAR system was designed for high reliability utilizing state-of-the-art electronics,
elaborate fail-safes and reliable components. Maintenance activities are generally required on
components which interact with the cooling system water directly (i.e. Probes, strainers and the
like). Performing periodic maintenance according to the schedule listed below will significantly
improve reliability and minimize downtime.
Maintenance Task* 1 2 3 4 5 6 7 8 9 10 11 12
Fluorometer-clean flow cell X X X X X X X X X X X X
Fluorometer-calibrate X X X X X X X X X X X X
Fluorometer-check dessicant X X X X X X X X X X X X
pH probe-clean/calibrate X X X X X X X X X X X X
ORP probe-clean/calibrate X X X X X X X X X X X X
Corrosion probes-check X X X X
Conductivity probe-clean/calibrate X X
Strainers-check/clean X X X X X X X X X X X X
Pellet feeder-check/refill Perform weekly
Pellet feeder-clean Prior to refi ll
Pellet Feeder Desiccant-check/replace Prior to refi ll
Pellet feeder ball valve-tighten X X X X
Pellet feeder rotor shield-replace X
*Assuming component is installed.
Using the flow cell brush (P/N 500-P2817.88) provided with the Start-up kit (P/N 500-TSRKIT.88 or
500-TSRKITSTR.88), gently push the brush down through the opening and scrub the full length of
the flowswitch and fluorometer tube.
Monthly, check the color of the desiccant indicator on the front of the 3D TRASAR fluorometer
.
If the indicator is blue, desiccant does not need to be replaced. If the indicator is pink, unscrew the
desiccant cover from fluorometer. Replace both the indicator (P/N 060-TR5223.88) and desiccant
canister (P/N 060-TR5222.88).
# Calibration
Sensor Reason for Calibration
Points
2 (blank and
Fluorometer Monthly, to verify accuracy
S0940)
Monthly, to fine-tune calibration to
1
compensate for mi nor drifting.
pH After installation of new probe, or
if probe readings become erratic 2
due to fouling/degradation.
Monthly, to fine-tune calibration to
1
compensate for mi nor drifting.
ORP After installation of new probe, or
if probe readings become erratic 2
due to fouling/degradation.
Monthly, to fine-tune calibration to
1
compensate for mi nor drifting.
Conductivity After installation of new probe, or
if probe readings become erratic 2
due to fouling/degradation.
Technical Help 521-OM0108.88 8-2
(630) 305-CHEM Copyright 2009 Nalco Company
3D TRASAR 5500 - Installation & Operation Manual
The wye strainer assembly on the pellet feeder serves as a dissolving chamber for the bio-reporter
pellets. If any residue from bio-reporter pellets remains in the strainer wear rubber gloves when
removing and cleaning this strainer mesh.
Close the inlet and outlet valves and unscrew the plug at the end of the wye. Remove the mesh and
rinse with clean water.
Section 9: Troubleshooting
Alarm
Indication Probable Cause Corrective Action
Type
TRASAR, Tagged Polymer, ORP, pH (Caustic Control)
1) Sensor probe is out of calibration. 1) Check calibration and recalibrate if needed.
Measurement 2) High Alarm threshold set too low. 2) Readjust alarm threshold.
High exceeds High 3) Chemical feed rate too high causing "overshoot". 3) Reduce chemical feed rate.
Alarm threshold 4) Chemical pump siphoning causing overfeed. 4) Elminate condition causing siphoning.
5) Process leak or other contaminant in system. 5) Confirm (with other data) and eliminate source.
1) Sensor probe is out of calibration. 1) Check calibration and recalibrate if needed.
Measurement 2) Low Alarm threshold set too high. 2) Readjust alarm threshold.
Low below Low 3) Chemical feed rate too low or interrupted. 3) Check feed line or increase chemical feed rate.
Alarm threshold 4) Product container empty. 4) Refill/replace product inventory.
5) Process leak or other contaminant in system. 5) Confirm (with other data) and eliminate source.
pH (Acid Control)
1) Sensor probe is out of calibration. 1) Check calibration and recalibrate if needed.
Measurement 2) High Alarm threshold set too low. 2) Readjust alarm threshold.
High exceeds High 3) Chemical feed rate too low or interrupted. 3) Check feed line or increase chemical feed rate.
Alarm threshold 4) Product container empty. 4) Refill/replace product inventory.
5) Process leak or other contaminant in system. 5) Confirm (with other data) and eliminate source.
1) Sensor probe is out of calibration. 1) Check calibration and recalibrate if needed.
Measurement 2) Low Alarm threshold set too high. 2) Readjust alarm threshold.
Low below Low 3) Chemical feed rate too high causing "overshoot". 3) Reduce chemical feed rate.
Alarm threshold 4) Chemical pump siphoning causing overfeed. 4) Eliminate condition causing siphoning.
5) Process leak or other contaminant in system. 5) Confirm (with other data) and eliminate source.
Conductivity
1) Sensor probe is out of calibration. 1) Check calibration and recalibrate if necessary.
Measurement
2) High Alarm threshold set too low. 2) Readjust alarm threshold.
High exceeds High
3) Blowdown valve throttled too low or plugged. 3) Check blowdown valve for proper operation.
Alarm threshold
4) Process leak or other contaminant in sy stem. 4) Confirm (with other data) and eliminate source.
1) Sensor probe is out of calibration. 1) Check calibration and recalibrate if necessary.
Measurement
2) Low Alarm threshold set too high. 2) Readjust alarm threshold.
Low below Low
3) Blowdown valve throttled too high or stuck open. 3) Check blowdown valve for proper operation.
Alarm threshold
4) Process leak or other contaminant in sy stem. 4) Confirm (with other data) and eliminate source.
Temperature
1) RTD probe is out of calibration 1) Verify temperature using a thermometer .
Measurement
2) High Alarm threshold set too low. 2) Readjust alarm threshold.
High exceeds High
3) Cooling system problem. Note: PVC piping is rated 3) Diagnose and resolve cause of excessive bulk
Alarm threshold
for 140 F max. water temperature.
1) RTD probe is out of calibration. 1) Verify temperature using a thermometer .
Measurement
2) Low Alarm threshold set too high. 2 ) Readjust alarm threshold.
Low below Low
3) Cooling system problem. Note: Freezing will burst 3) Diagnose and resolve cause of low bulk water
Alarm threshold
piping. temperature.
Alarm
Indication Probable Cause Corrective Action
Type
[XX] Corrosion ([XX] = Probe metallurgy)
Mesurement 1) Insufficient flow rate past probe. 1) Verify minimum of 5 gpm flow is present.
High exceeds High 2) High Alarm threshold set too low. 2) Readjust alarm threshold.
Alarm threshold 3) Corrosion treatment program needs adjustment. 3) Consult your Nalco Representative.
Cell Fouling
Mesurement 1) Fluorometer flow cell is fouled. 1) Clean flow cell with supplied brush and cleani ng
High exceeds High solution. Recalibrate as needed.
Alarm threshold 2) High Alarm threshold set too low. 2) Readjust alarm threshold.
Control Relays (TRASAR, Tagged Polymer, Bio-reporter, Oxidizing Biocide, Blowdown, pH)
Relay has been 1) Tank empty/pump lost prime/faulty pump or valve. 1) Check tank/pump/valve/feed line. Repair as
Relay activated for needed.
Timeout longer than relay 2) Relay Timeout Limit set too low. 2) Readjust alarm threshold.
timeout limit 3) Feed/bleed rate too low for demand. 3) Increase feed/bleed rate.
General Sensors
1) Wiring or other connecti on is loose. 1) Retighten wires or other connections. Be sure wire
Error detected
insulation is not preventing good contact.
Sensor with i ndicated
2) Sensor probe failure. 2) Replace probe. Probe replacement may be
sensor probe.
included in purchased service plan.
1) Flow rate too low or back pressure too high. Flow 1) Investigate and correct.
Sample Insufficient water switch requires 0.5 gpm.
Flow flow through 2) Flow switch fouled. 2) Clean flow switch.
Low fluorometer cell 3) Flow switch wiring is loose or not making good 3) Replace flow switch
connection.
1) The interlocking device is not active. 1) Check. This may be expected operation.
The interlock
Interlock 2) Interlock is not used and jumper is not in place. 2) Replace jumper (refer to wiring diagram for
input has been
Absent 3) Interlock wiring is loose or not making good location and di rections.
deactivated
connection. 3) Check and correct wiring connections.
1) Cable connector on Fluorometer is loose. 1) Disconnect and inspect for bent pins or signs of
Controller is damage, then reconnect.
unable to 2) Wiring from fluorometer is loose or miswired in 2) Check and correct if needed (see wiring diagram
Fluor.
communicate controller. for details).
Comm.
with the 3) Component failure. 3) Reboot system by momentarily cycling power to
fluorometer initiate recovery. If unsuccessful contact Help D esk at
630-305-CHEM for assistance.
The sensor 1) Sensor board is loose preventing good con nection. 1) Gently press on "Red" circuit board inside
Sensor board is not controller to ensure snug fi t.
Board communicating 2) Component failure. 2) Reboot sytem by momentarily cycling power to
Comm. with the initiate recovery. If unsuccessful contact Help D esk at
controller 630-305-CHEM for assistance.
3D Bio Control
1) Pellet Feeder cable loose or con nections not made 1) Disconnect and inspect both Pellet Feeder
Controller is not correctly. connector and wi ring in controller, then reconnect.
Pellet
communicating 2) Pellet Feeder power switch is not turned on. 2) Turn power switch on.
Feeder
with P ellet 3) Component failure. 3) Reboot system by momentarily cycling power to
Comm.
Feeder initiate recovery. If unsuccessful contact Help D esk at
630-305-CHEM for assistance.
The P ellet 1) Pellet supply is low. 1) Check and fill Pellet Feeder, or redistribute
Feeder has not remaining pellets.
Pellet completed a
Feeder pellet feed
Busy before the next
pellet is called
for
The P ellet 1) Pellet Feeder is empty. 1) Check and refill Pellet Feeder.
Pellet
Feeder has not
Supply
fed a pellet
Low
when called for
Pellet The P ellet 1) The clear pellet feed tube is blocked or dirty 1) Check and clean feed tube usi ng supplied brush.
Sensor Sensor LED is preventing light transmission.
Error blocked.
Alarm
Indication Probable Cause Corrective Action
Type
1) Ball valve separated from motor shaft. 1) Contact Help Desk at 630-305-CHEM for
Ball Valve Pellet Feeder ball assistance.
Error valve not rotati ng
Alarm
Indication Probable Cause Corrective Action
Type
1) Pellet Feeder unable to bring bio-reporter level up 1) Add additional Quick Dose pouch and wai t for
to setpoint in a reasonable time. mixing. If alarm persists, repeat. Check bio-reporter
Bio- Bio-reporter
2) Excessive Oxidizing Biocide degrading bio- level at controller di splay in Information>Diagnostic
reporter level below
reporter. Source of Oxidizing Biocide may be makeup, Data>Fluorometer Data (Rz).
Low control threshold
manual feed, process or other leak, or oversized 2) Remove source of excess biocide.
biocide pump.
Bio-reporter 1) Same causes as Bio-reporter Low. 1) Same. Alarm must be cleared at the display.
level below
Bio- control threshold
reporter for an extended
Low period, and
Failsafe oxidizing
biocide enters
failsafe mode.
1) Bubbles in flow cell. 1) Reduce flow rate or remove source of entrained air
Bio-Data
Excessive noise (i.e., air-locked pump).
Noise
in data 2) Bioreporter feed point is too close to TRASAR 2) Change feed point to a location further
High
5000. downstream of the intake.
Excessive noise 1) Same causes as Bio Data Noise High. 1) Same. Alarm must be cleared at the display.
in data for an
Bio-Data extended
Noise period, and
High oxidizing
Failsafe biocide has
entered failsafe
mode.
1) ORP probe is fouled and needs calibration 1) Clean and calibrate ORP probe.
2) The system takes a long ti me to dissipate the 2) In the Slug Feed screen <3D BioControl Recovery
After a Slug the Oxidizing biocide fed during the slug. Settings> <Show Advanced Recovery Options>
Slug ORP did not 3) Too much Oxidizing biocide was applied during the Increase the ti me limit.
Recovery return to pre-slug slug. 3)Decrease the oxidizing Biocide applied by
Failure values within the changing the ORP setpoint for the Oxidant slug. If the
time limit. slug is applied by timer control, decrease the time for
the Oxi dant feed. If TB 960 is the Oxidizing biocide,
decrease the weight of the Biocide applied per feed.
1) The Oxidant pump is airlocked or failed to apply the 1) Fix pump or feeder.
biocide. 2) Increase the ORP Rise Limit Time in the <Slug
The ORP did not
2) The recirculati on time in the tower is long and the Oxidizing Biocide> <Slug Feed Settings.>
Ox Shock increase the
applied Biocide is not seen by the ORP probe within 3) Decrease the min ORP rise <Slug Oxidizing
Failed minimum value
the time limit. Biocide> <Slug Feed Settings>
during a slug
3) The mi nimum ORP rise setting is too large. 4) Calibrate/clean probe.
4) The ORP probe is fouled/miscalibrated.
Relay has been 1) Tank empty/Pump lost prime/faulty pump or valve. 1) Check tank/pump/valve/feed line. Repair as
activated for 2) Relay timeout li mit set too low. needed.
longer than relay 3) Pump rate is too low for demand. 2) Readjust alarm threshold.
timeout limit. 4) The Oxidizing biocide is n ot being applied. 3) Increase pump rate.
Oxidant In 3D Bio-Shock 5) The time limi t between feeds is set too long for the 4) Check the pump/feeder.
Relay mode the system. 5) Decrease the minimum time between feeds.
Timeout BioReporter is
calling for too
frequent
oxidizing
biocide feed
Bio Relay has been 1) Tank empty/Pump lost prime/faulty pump or valve. 1) Check tank/pump/valve/feed line. Repair as
Reporter activated for 2) Relay timeout li mit set too low. needed.
Relay longer than relay 3) Pump rate is too low for demand. 2) Readjust alarm threshold.
Timeout timeout limit 3) Increase pump rate.
3D Scale Control
1) System upset such as a hi gh pH event, process 1) Conductivity setpoint has been decreased
System Rate of Polymer leak, high heat load event, etc. automatically to induce blowdown.
Stress Consumption is 2) If coincident with or subsequent to a Conducti vity 2) Enter lower conductivity setpoint and relearn
High high adjustment MIN alarm, the controller is unable to system stress.
respond to the high stress condition.
Conductivity 1) System stress is low compared to learned stress. 1) If higher cycles are required, enter higher
Cond. Adj. Setpoint conductivity setpoint and relearn system stress.
MAX adjusted to the Otherwise, no acti on is required.
maximum
Conductivity 1) System stress is high compared to learned stress. 1) If lo wer cycles are required, enter lower
Cond.
Setpoint conductivity setpoint and relearn system stress.
Adj.
adjusted to the Otherwise, no acti on is required.
MIN
minimum
Appendix A: Specifications
Monitoring Functions
Corrosion Monitoring
Continuous on-line measuring (with Auto-Probe Recognition) mild steel and copper (standard)
range 0 to 99 mils per year.
Additional Monitored Variables
• Cell Fouling
• Turbidity
• Temperature
• Background Fluorescence
• Relay Status
• Pulse counter with water meter
Additional Calculated Variables
• Holding Time Index
• System Volume
• Blowdown Rate
• Heat Rejection (Btus)
• Makeup Water Rate
• Recirculation Rate
• Pump Duty
• Product Usage
Specifications
Plumbing Requirements
Control Panel 3/4" FNPT inlet, (2) outlets (1) 3/4" FNPT & (1) 1/2" FNPT (once inlet and outlet ball
valves are installed.) Total flow required 5 gpm. Outlet piping must be run separately.
Pellet Feeder 3/4" FNPT inlet, 3/4" FNPT outlet, 5 gpm required, 20 psig differential across both
units.
Electrical
85 to 250 VAC @ 20 amps service
Sample requirements
minimum Conductivity of 10 uS/cm required for Corrosion measurement.
Inputs
• (2) unassigned 4-20 mAunpowered non-isolated inputs (used for logging and alarm level call out.)
• (1) Flow input (contact closure)
• (1) Interlock input (contact closure)
• (2) Spare digital inputs (contact closure) used to log input states or low frequency water meter
counts
• Modbus RS-232 or RS-485 (choose one) RTU SCADAport
Outputs
• (7) control relay outputs rated 2.5 amps each (larger loads require external motor starters)
• (8) 4-20 mA powered non-isolated outputs
Alarms
• (1) Alarm output relay (rated 1.0 amp max unpowered)
• E-mail notification of alarm conditions (via modem to ISP or Customer’s LAN-based mail server.)
Enclosure
Nema 4X
Ambient temperature
40-120 °F (4-50 °C)
Process water temperature
40-120 °F (4-50 °C)
Relative Humidity
0-95% non-condensing
Maximum water pressure
90 psi
Display/Keypad
Numeric keypad plus specialty keys, graphical display with plotting functions
Security
Password protected (user assignable)
Data management
All variables are data logged in spreadsheet format
Communications
• Ethernet port for direct PC connection
• Built-in modem for direct remote connection or access to 3D TRASAR web via Internet (requires
ISP)
• Optional Nalco Global Gateway with monthly service package.
Approvals
UL, CUL, CE
Software
• 3D TRASAR configurator for system configuration and data transfer
• Optional Vantage® 100 SPC software for comprehensive data management and analysis.
Appendix B: Drawings
Please see the attached pages for drawings of the 3D TRASAR models. Please note that drawings
show all probes installed. pH, ORP and Corrosion probes are available accessories for the
3D TRASAR Starter (3500 series) models.
Drawings of the Pellet Feeder PlumbingAssembly and High Pressure Water Reinjection system are
also included.
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
FLUOROMETER
OUTLET
SAMPLE
OUTLET
SAMPLE
INLET
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
FLUOROMETER
OUTLET
SAMPLE
OUTLET
SAMPLE
INLET
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
FLUOROMETER
OUTLET
MiniPurge
INSTRUMENT
AIR
SAMPLE
OUTLET
SAMPLE
INLET
Actions
Alarms Back
7 8 9
4 5 6
1 2 3
MENU HELP
0
FLUOROMETER
OUTLET
MiniPurge
INSTRUMENT
AIR
SAMPLE
OUTLET
SAMPLE
INLET
060-TR5120.88 Bio-reporter, Pellet Feeder Assembly, for system sizes > 12,000 gallons
141-PJ4035.88 Bio-reporter, Liquid Feed pump, for system sizes 3,000 to 12,000 gallons
6008505 Bio-reporter, Liquid Feed pump, for system sizes up to 3,000 gallons
* This part is available as an optional accessory, but is not included with either the Start-up
Kit or the Maintenance Value Package.
A copy of the manual shown below is included with models 060-TR3530.88, 060-TR3531.88,
060-TR3532.88, 060-TR3533.88, 060-TR5530.88, 060-TR5531.88, 060-TR5532.88 and
060-TR5533.88. Use this manual for start-up of theAir Purge System. Call 630-305-CHEM for more
assistance.
SCADA - General
The 3D TRASAR Controller provides a Modbus RTU Slave connection. Users can select either an
RS-232 or RS-485 hardware link (choose one). Asingle 3-pin SCADA connector is provided in the
Controller (plug-in to selected communications port).
Note 2: The “System Volume Average”, “Heat Rejected”, and “Recirculation Rate” registers can be
read as Long or Word values (2 words per Long). The values of each Long are stored in two con-
secutive words with the most significant value stored in the lower address.
Note 3: In some data acquisition systems, the two 16-bit registers comprising the Floating Point
may need to be switched. This is usually a configuration setting within the polling set-up routine.
Note 4: The scaled floating point registers (Integers and Longs) were added as a convenience for
control systems that do not support reading Floating Point Input registers. The Integer and Long
Input registers contain the same information as the Floating Point Input registers. ables
T 3, 4, 5 & 6
document all registers currently supported by the 3D TRASAR Controller .
Note 6: If the controller is configured to display SI units, the corresponding engineering units will
appear on the screen and be translated to it’s corresponding Modbus register . As example, “Sample
Temperature” will be converted from Deg F to Deg C. The corresponding Modbus registers (Input
Float and Input Integer) will hold the value in Deg C units.
RS-232
Controller
(DB9, PC
SCADA 232
Connection)
GND GND (Pin 5)
TX RX (Pin 2)
RX TX (Pin 3)
RS-485 Connection
RS-232
Controller
(DB9, PC
SCADA 485
Connection)
GND GND
B -
A +
Control Relay 7 – pH
Actions à Manual Relay Control à “Product Name” State
Alarm Relay
Information à Operating Data à (Alarm Bell in upper left corner of display)
Digital In 1 – No Flow
Information à External Inputs à DI1 State
Digital In 2 – Interlock
Information à External Inputs à DI2 State
Digital In 3 – Spare
Information à External Inputs à DI3 State
Digital In 4 – Spare
Information à External Inputs à DI4 State
Rz (Bioreporter)
Information à Diagnostic Data à Fluorometer Data à Rz
Rs (Bioproduct)
Information à Diagnostic Data à Fluorometer Data à Rs
Turbidity
Information à Operating Data àTurbidity
OR, Information à Diagnostic Data à Fluorometer Data à Turbid
Background
Information à Diagnostic Data à Fluorometer Data à Bkrnd
Cell Fouling
Information à Diagnostic Data à Fluorometer Data à Fouling
Conductivity
Information à Operating Data à Cond
OR, Information à Diagnostic Data à Sensor Data à C
pH
Information à Operating Data à pH
OR, Information à Diagnostic Data à Sensor Data à pH
ORP
Information à Operating Data à ORP
OR, Information à Diagnostic Data à Sensor Data à ORP
Sample Temperature
Information à Operating Data à Temp
OR, Information à Diagnostic Data à Sensor Data à ORP
Nalco Bio-Index
Information à System Performance à N Bio Index*
*Controller must be configured for 3D Bio Control to be visible
Polymer Consumption
Information à System Performance à Consumption
HTI
Information à Operating Data à HTI
Blowdown Rate
Information à Water & Energy à Blowdown*
*Controller must be configured for Blowdown Control to be visible
Makeup Rate
Information à Water & Energy à Makeup
Analog Input 1
Information à External Inputs à 4-20 Input 1
Analog Input 2
Information à External Inputs à 4-20 Input 2
Heat Rejected
Information à Water & Energy à Heat Rej.
Recirculation Rate
Information à Water & Energy à Recirc Rate
Note: Some variables may not appear on the display if the controller is not configured to use them.
These variable(s) should be ignored when processing the Modbus block register response.
This diagram shows the recommended network configuration. Note that the use of a router (P/N
060-TR5270.88) is required on LAN’s with more than 25 computers on the same segment as the
3D TRASAR controller. If this is ignored the controller’s Ethernet stack may overflow and communi-
cation to the controller may not be possible.
The 3D TRASAR Controller sends data and alarms to Nalco via email.All that is required is that the
customer’s SMTP server grants access to the 3D TRASAR Controller and allows it to send out-
bound email to the Nalco mail server (3dtrasar.com).
There are four different upgrades available for the 3D TRASAR Starter system:
1. Order Upgrade
Order the 3D TRASAR Starter Fluorometer Upgrade, P/N 060-TR3200.88. Order through Nalco
customer service, or through the Order Wizard.
UltraSand Plus
Technical Help 521-OM0108.88 G-1
(630) 305-CHEM Copyright 2009 Nalco Company
3D TRASAR 5500 - Installation & Operation Manual
Refer to operation manual shipped with high pressure water reinjection system for more details.
This appendix supercedes information found in the main portion of the 3D TRASAR manual. Read all
sections prior to beginning any installation.
1. Identify Hardware Model
Wall Mount Frame Mount
Description
Model Model
2. Fluorometer Installation
The fluorometer is shipped unattached to prevent damage. Follow installation instructions from sec-
tion 2.6.1 in the manual with the following exception: a t-fitting is installed on the outlet side of the
Fluorometer, instead of a flowswitch.
3. Wiring
A junction box is not available with any of the 3D TRASAR models for High FoulingApplications.
Analog output 1 is pre-wired (jumpered) toAnalog input 1 for the purpose of monitoring and
logging flow cell fouling data (every 15 minutes.)
The following steps must be completed as part of the configuration forall high fouling applications
in order to monitor cell fouling. Select task button from 3D TRASAR Configurator main window , then
follow list of steps.
To attach the fluorometer cleaning system, connect the pump outlet to the
t-fitting located between the solenoid valve and Fluorometer . Pump ships with
a 3/8" check valve which threads directly to the t-fitting. Connect the pump and
solenoid valve power cords to the pigtails pre-wired to relay 5.
Plumb outlet of fluorometer to drain or lower pressure point in system (minimum differential 20 psig.)
IMPORTANT: Prior to physical cleaning of the fluorometer , close ball valve beneath
fluorometer and set relay #5 to manual off (through keypad pressActions>Manual
! Relay Control>Relay #5 (S0340) >Current State>Edit>Manual Off>Accept.)
Press Auto After then Edit to select amount of time the relay will remain in selected
mode prior to reverting back to auto. Open ball valve when finished with cleaning.
7. Maintenance
In addition to the maintenance schedule recommended for standard 3D TRASAR units, perform the
following activities:
• After start up, check basket strainer, magnet and rotometers weekly. Determine appropriate
maintenance schedule based on degree of fouling.
• Check pump tubing of automated fluorometer cleaning system for wear monthly,
replace if necessary.
8. Replacement parts
Part Number Description
001-H04066.88 Containment basin, PE, for 5 gallon pail, 7.5" H x 21.5" dia
6010754 Reducing bushing, ¾" MNPT x ½" FNPT (included with S0340)
Estimated Estimated
Part Number Description Equipment Shipping Dimensions (LxWxH)
Weight (lbs) Weight (lbs)
• 991-05059581.88 Solution pump, Grundfos, DME2, Maximum capacity 0.5 gal/hr, PVDF
pump head, PTFE gaskets and ball valve, 110-240VAC, 50-60Hz.
• 991-05059502.88 Solenoid valve, Stainless Steel body with PTFE seals, 1/4” FNPT,
240VAC/60Hz, 220VAC/50Hz.
Different wire colors are also used in the control box. Specifications are shown below
.
Connection Controller Box Controller Box
Wire Color
Description Board Reference Terminal
Ground TB9-4 Gnd Not used
Green with
Ground TB9-3 Gnd
Yellow Stripe
Neutral TB9-2 N Blue
Line Power TB9-1 L Brown
Select the appropriate option above (Monitor or Control) for theAnalog Input that the halogen ana-
lyzer is connected to. If you are using the halogen analyzer to control oxidant feed, select the Bio-
cides tab in Control Settings and enter the set point and control range. The default range of the
analog output from the halogen analyzer is 0 ppm at 4mA and 5 ppm at 20mA. If this is suitable for
the application, enter these values into the corresponding fields in configurator
. If you alter the output
settings of the halogen analyzer make the adjustments in these fields so the input is scaled cor-
rectly.
After the halogen analyzer is chosen, the FRH or TRH concentration will replace ORP on the sys-
tem operation screen on the OBC.
Limitations
Warranty does not cover damage caused by misuse, neglect, accident (including a force majeure
event), improper installation or improper maintenance or repair. Nalco’s liability under this warranty
is limited to repair or replacement of defective items, or, refund of or credit for the product price
excluding shipping.
• Credit: Products purchased from Equipment Solutions may be returned for credit within
three (3) months of purchase if they have never been placed in service and they are re-
turned in the original packaging complete with all accessories, component with all accesso-
ries, component parts, and manuals.
• Repairs: Products eligible for billable repair include standard items that were purchased
within the past five (5) years.
Glossary
3D TRASAR Bio-Control
3D TRASAR Bio-Control is a proprietary Nalco technology for measurement and control of micro-
biological activity in open recirculating cooling water systems. 3D TRASAR Bio-Control uses a
fluorescent bio-reporter to probe the entire system, measuring both planktonic and sessile microbial
activity and feeding oxidizing biocides in response.
3D TRASAR System
The Nalco 3D TRASAR system is the most advanced water treatment control system on the market
today. It represents a total system approach, encompassing service, innovative chemistry, ad-
vanced software and sophisticated equipment, that allows you to manage stresses in cooling tower
systems on a continuous, ongoing basis.
3D TRASAR Configurator
A PC-based tool which interfaces with the 3D TRASAR controller. The 3D TRASAR Configurator
allows Nalco sales representatives to configure the controller, establish remote communications,
download data, update controller software, and change control methods.
3D TRASAR Optimizer
A PC-based tool which allows Nalco sales representatives to design basic cooling water programs,
evaluate program alternatives, and analyze the impact of controller or makeup water variables on
system performance.
Order Wizard
A PC-based tool which assists Nalco representatives with ordering 3D TRASAR chemicals, equip-
ment and service.
Bio-reporter
Bio-reporter is a fluorescent molecule, which is enzymatically changed upon interaction with active,
respiring microbes (both planktonic and sessile). The fluorescence of the bio-reporter and the
product of its interaction with bacteria is monitored continuously by an in-line fluorometer. The rate
of change of the fluorescence with time gives a measure of the microbiological activity in the sys-
tem.
Conductivity measurement
The ability of a sample to conduct heat or electricity. Electrical conductivity is usually expressed in
microsiemens (uS.)
Corrosion monitoring
A measurement of the rate at which corrosion is occurring in a system. Corrosion is the deteriora-
tion of metal caused by a reaction with its surrounding environment. Corrosion is measured in units
of mils per year (mpy.)
Nalco Bio-Index
Nalco Bio-Index is a measure of microbiological control in the system based on the bio-reporter
response. As the Nalco Bio-Index increases the relative microbial control decreases.
ORP measurement
Oxidation Reduction Potential. A measure of the oxidation-reduction potential of a sample. ORP is
measured in millivolts (mV.)
pH measurement
A measure of the acidity or alkalinity of a sample, numerically equal to 7 for neutral solutions,
increasing with increasing alkalinity and decreasing with increasing acidity. The pH scale commonly
in use ranges from 0 to 14 pH units.
Start up factor
Extends the relay timeout function during start up. The start up factor is multiplied by the relay
timeout period (ie. if the “relay timeout” is set to 60 minutes, and the start up factor is set to 5, the
total timeout period will be 300 minutes.)
Relay timeout
Maximum allowable time a relay is allowed to be energized before an alarm is triggered, and the
failsafe duty mode begins (user set.)
Tagged Polymer
A new component of TRASAR cooling water programs (ie. Nalco’s Prism polymer or High Stress
Polymer (HSP.)) The concentration of the “tag” equals the concentration of available, un-reacted
dispersant polymer. The 3D TRASAR algorithm compares the inert TRASAR signal to that of the
tagged signal, and then makes appropriate control decisions, either to increase or decrease the
treatment concentration or increase or decrease the conductivity set-point.
Turbidity measurement
A measure of how cloudy or dirty the water is, obtained by quantifying the degree to which light
traveling through a water column is scattered by the suspended particles. (The greater the scatter-
ing, the higher the turbidity.)
Vantage V100
Vantage V100 is Nalco’s field data management software program. 3D TRASAR data files can be
imported into Vantage V100 for the purpose of trending and reporting operational data.
Nalco Company Operations
North America: Headquarters – 1601 West Diehl Road, Naperville, IL 60563-1198 ٠ USA
Energy Services – 7705 Highway 90-A, Sugarland, Texas 77478 ٠ USA
Europe: Ir.GF.Tjalmaweg 1 ٠ 232 BV Oegstgeest ٠ The Netherlands
Asia Pacific: 2 International Business Park ٠ #02-20 The Strategy Tower 2 ٠ Singapore 609930
Latin America: Av das Nações Unidas 17.891 ٠ 6° Andar 04795-100 ٠ São Paulo ٠ SP ٠ Brazil
www.nalco.com
Nalco, the logo, PRISM, Vantage, 3D TRASAR, STA٠BR٠EX, ACTI-BROM, are Registered Trade-
marks of Nalco Company.