Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Tel:00852-30623001, Fax:00852-31828558
Email: sale@lab-men.com Site: www.lab-men.com
PRICE LIST
Unit Price
No Order ID Product Name Qty
(USD)
1 LDY900 Nanometer particle size analyzer - 1
1
Labmen instrument technology Limited
Tel:00852-30623001, Fax:00852-31828558
Email: sale@lab-men.com Site: www.lab-men.com
Technical standard: GB / T 19627-2005 / ISO 13321: 1996 GB / T 29022-2012 / ISO 22412: 2008
Test range: 1nm-10000nm
Detection Unit: Dynamic Light Scattering (PMT)
Laser Type: Fiber Semiconductor Laser, λ = 532nm, Service Life:> 25000H
High sensitivity and signal-to-noise ratio: The instrument's detector uses a professional
high-performance photomultiplier tube (PMT), photonic signal with high sensitivity and signal to
noise ratio, thus ensuring the accuracy of test results;
2
Labmen instrument technology Limited
Tel:00852-30623001, Fax:00852-31828558
Email: sale@lab-men.com Site: www.lab-men.com
Super computing power: The use of PCS technology to determine nanoscale particle size must be
able to distinguish nanosecond-level signal fluctuations. The core components of the instrument
using ASIC developed high-speed photon correlator, with a very high resolution 6ns recognition
and high signal processing speed can be collected in real-time photon number and calculate the
relevant operations, I set the test determined to respect basis.
Stable optical system: The optical system constructed with short wavelength LD pump laser
source and optical fiber technology makes the photon correlation spectrum detection system not
only small in size, but also has strong anti-interference ability, which lays the foundation for the
accuracy of the test.
High-precision temperature control system: The precision of the temperature control system of
the sample cell and the temperature control system of the laser is up to ± 0.1 ° C, so that the
sample under test and the laser light source are in a constant temperature during the whole
testing process to avoid the influence of the temperature change on the non-test result and
ensure Test accuracy and repeatability;
Super-powerful analysis software: Inversion algorithm in analysis software uses the cumulative
method recommended by international standards and the current more common non-negative
least squares (NNLS) and Contin and other algorithms, the test results and the international
authority of similar products With good consistency
Accurate and stable test: The perfect combination of high-performance hardware and
internationally standardized inversion algorithm, which results in the accuracy and repeatability
of the test results. The test accuracy and repeatability are all higher than those of international
standards.
Technical Parameters
Model LDY900
Implementation standards GB/T 29022-2012/ISO 22412:2008
Test Range 1-10000nm(related with the sample)
Concentration range 0.1mg/L-100mg/L
Accuracy error <1%(National standard sample average particle size)
Repeatability error <1%(National standard sample average particle size)
λ=532nm,LD pump laser(Exclusive with temperature
laser
protection)
detector Photomultiplier tube(PMT)
Scattering angle 90°
Sample cell 10mm*10mm , 4ml(With temperature protection)
Temperature range 8-45℃(Accurate temperature to 0.1℃)
<1Min/tims(not include the sample dispersion time)
Test speed