Sei sulla pagina 1di 4

SCOPE

MTS 101
MTS 301
Meter Test System

Check the energy meter accuracy...


... on MTS Series
MTS 101, MTS 301
The Product
MTS 101 / MTS 301 Meter Test System
Energy meter is one of the most important component in the electricity generation & supply system.
The revenue of the generation, transmission and distribution organizations is highly depended on the
accuracy of the energy meters. If the the energy meter malfunctions or becomes inaccurate, there can
be a huge loss to these organizations.
Energy Meters are susceptible to malfunctioning because of many factors, which directly aects its
accuracy. Thereby directly inuencing the revenue of service provider/ supplier.
Errors in the Energy meter due to many reasons, such as:
Dirt (on the disk, in the air gaps).
Magnetic/metal particles(in the permanent magnet air gap)
Gummy oil and/or dirt in bearings
Damaged mechanical components.
Disk rubbing
Improper mesh of gears or dirty gearing.
Improperly adjusted bearings
Vibration of the meter
External magnetic elds which may add to, or subtract from, the normal meter magnetic ux.
Short-circuited turns in meter coils.
Tampering
Harmonics
Voltage Unbalance
Frequency variations
Temperature, as most of the meters are outdoors
Imperfections in the embedded software or sensitivities in the electronic circuitry
Voltage Transient Susceptibility
These factors majorly aect the accuracy of energy meter, hence directly aecting the revenue.
Therefore, we need to test the energy meter for its accuracy and other related factors. So that we get
the specied accuracy and desired output.
SCOPEs meter testing system is a Handheld Unit used for online testing of both Three Phase meters as
well as Single Phase meters and dierent metering systems.
Again, it is used for both Static meters and Electromagnetic meters etc.
It records KW,KWH,KVAR, KVARH, Power factor, Phase angle, etc
With these functions, it is widely used in Power Utility, Substations,Telecom machine room, Power con-
sumption examination, Circuit breakage inspection, Power abnormality analysing, Asset Management
and Industrial Production Management.

MTS consists of handheld Main Unit, Current clamps and Voltage testing leads and scanning head.

MTS Main Unit Current Clamp Voltage Test Lead Scanning Head

www.scopetnm.com
SCOPE
Features
Smart, portable and easy touch-screen (TFT) operation
Combination of electricity meter tester and power quality analyzer
Online monitor for real-time data testing and comparison
Safe and quick measurement for electricity anti-theft
Test result playback on screen and transferable to PC for report printing
User-friendly interface and vivid screen display, simplify on-site testing power meters

User can choose from two models based on the application:

Function MTS 101 MTS 301


Three phase metering unit
Three phase power meter
Meter Single phase metering unit
Single phase metering unit
measurement Single phase energy meter
Single phase power meter
High-volt power meter
Measure CT ratio and angle dierence
CT ratio Test CT ratio and phase deviation
(optional)
1. Test active energy, reactive energy, Power fac-
Active energy,
tor in low-volt with load
Online monitor Reactive energy,
2. Active energy, reactive energy, Power factor of
Power factor
power meter in dierent voltage rate.
Wire connection - Visual indication for incorrect wiring connection

Connection diagram for 3-Phase Meter Menu Settings and Result

Connection diagram for Single-Phase Meter Menu Settings and Result

www.scopetnm.com
MTS 101, MTS 301
Specications

Parameter MTS 101 MTS 301


Voltage Signal Input: Direct Input
Input Impedance >2M, 20pF
Testing Voltage Range 10 to 500Vrms 0.5% 10 to 700Vrms 0.5%
Peak Voltage 700V 1000V
Power consumption <0.5VA/phase
Current Signal Input: Indirect input, with dierent current clamps
Testing Current Range 20Arms, 60Arms, 100Arms 0.5% 20Arms, 500Arms, 1000Arms 0.3%
Power consumption <2.0VA/phase head
Frequency range 40 Hz to 70Hz 0.01Hz
Measurement accuracy
Active power 0.5%
Reactive power 1.0%
Active energy 0.5%
Reactive energy 1.0%
Power factor 0.005
Screen and memory
Screen 320 x 240, 3.5 TFT LCD touch-screen 640 x 480, 3.5 TFT LCD touch-screen
Internal Memory 16 MB Flash
External Memory 256MB
PC Connectivity USB port available
Power supply Rechargeable Li-ion battery workable for 5 to 6 hours
Power adaptor 100 to 240VAC/16.8VDC-1A adaptor
Weight and Dimension 0.7kg, 277 169 36mm mm
Environment 0 to 50 C max, 95% RH (non-condensing)
Insulation >100M, 2kVAC/50Hz, 1 min >2M,1500VAC/50Hz, 1 min
Impulse constant FL = 3600(5/Ie)P/kWh, Where Ie is CT range

www.scopetnm.com