P R O D U C T
Installation Qualification and
w w w. p e r k i n e l m e r. c o m / i n s t r u m e n t s

Operation Qualification (IQ/OQ)

N O T E
Be confident your results selected key activities performed Key Capabilities
will be accurate and and verified during installation,
Our IQ/OQ program produces
compliant everyday. before the instrument is put into
the thorough audit trail needed
service for actual analysis. The to meet internal and external
It is not enough that your labs IQ/OQ documentation is specifically validation requirements
instruments perform to manufac- designed to your instrument
Documented tests confirm proper
turers standards. More importantly, including your company name,
installation and operation
they need to meet the regulatory address, instrument serial number,
requirements of your lab everyday. purchase order number and an Log book for recording instrument
PerkinElmers team of service individualized protocol number. history ensures your instrument
follows good GLP and GMP guidelines
experts understand the increasing This is more than a simple instru-
need for validation and regulatory ment performance verification More factory-trained service people
compliance and has designed a our IQ/OQ program produces the than anyone else in the industry,
available globally
complete Installation and Operational thorough audit trail of system
Qualification (IQ/OQ) program performance you need to meet
including an extensive Log Book validation requirements imposed
for complete instrument history. by internal and external regulatory
This complete program provides groups giving you the confidence
a permanent record identifying in your instrument everyday.
Ensure youre ready to go Compare equipment as received Install hardware including all
right from the start with purchase order including soft- tubing, columns, cables for power,
ware, accessories, spare parts, and data flow and instrument control
Installation Qualification (IQ) provides
documentation such as tutorials,
formal checks and documentation Install or confirm factory installation of
manuals and operating procedures
to confirm that the instrument and its software on the computer following
components have been supplied as Check equipment for any damage the installation instructions
ordered and properly installed in during shipping
Configure peripherals, e.g., printers
your lab environment (Figure 1).
Check that the documentation and equipment modules
Upon delivery, the shipment will be
supplied is of the correct issue
visually inspected and reviewed for Document equipment tutorials,
completeness followed by installation of Establish that the laboratory manuals and operating procedures
all hardware, software and peripherals. environmental requirements are
Once completely installed, the system Perform installation functionality
appropriate for the instrument (power,
will be turned on to confirm all check and prepare installation report
bench, lighting, accessibility etc.)
modules and systems work correctly. Activate utilities and pre-condition
Finally, all information determined is Review all hazards and safety
instrument (e.g. warm-up, leak-test,
then documented within the protocol. precautions, including materials
flow controls etc.)
that may come in contact with
For the Installation Qualification the instrument Check that the response to the
PerkinElmers Customer Support initial application of power is as
Engineers will: expected and record deviations

Record all relevant activities


and actions and sign all required
documents
Installation Qualification and Operational Qualification
of the PerkinElmer Series 200 LC Autosampler Locate the certificates of conformity
and place them in the appropriate
protocol appendices
Certification
Confidence your instrument
The Installation Qualification and Operational Qualification final report for the is ready to run
PerkinElmer
Once the instrument has been installed,
Series 200 LC Autosampler Operation Qualification (OQ) will
Serial number 123N4567890
demonstrate that the instrument will
has been reviewed and was found to meet the requirements necessary for the function according to its operational
instrument to be used in the customers laboratory. specifications. The Service Engineer
will ensure the system is functioning
to acceptable limits by testing relevant
FINAL REPORT instrument specifications, parameters
and ranges as determined by
Reviewed by (Name and Title) Signature Date
PerkinElmers Instruments Performance
Authorized PerkinElmer Representative Verification (IPV). At the completion
of the OQ all of the operation tests
are documented, the results are
noted, and all relevant certificates
Authorized Customer Representative are completed and signed. Once the
instrument is validated, the Service
Engineer will provide instrument
familiarization training to confirm
Customer QA Representative (if applicable) that you are ready to run your
instrument that very day. The
Service Engineer will also review
and explain the contents of the
Figure 1. IQ/OQ Certification confirms all formal checks and documentation have been completed.
Instrument Log Book so you are familiar Life Sciences Optoelectronics Instruments Fluid Sciences

with the information provided.


For Operation Qualification
PerkinElmer Service Engineers will:
Perform basic instrument control
functions from both the instruments
keyboard and from the computer
(if applicable)

Perform basic functions of the


application software

Test the equipment hardware for


proper functionality

Perform Instrument calibration


as documented in the supplied instrument system log book
configuration and Test Certificate

Document all the operation


test results

Perform familiarization training

Explain the purpose and contents


of the Instrument Log Book and
associated documentation Figure 3. Instrument System Log book ensures a complete record of activities
is maintained.

Complete Instrument History Expert Support Services


Model
Serial No.
for your lab
The Instrument System Log Book
Installation/Operational Qualification is an integral part of the IQ/OQ Not only are we prepared to assist
Protocol No. Date program as it ensures a complete with IQ/OQ, PerkinElmers uptime
C.S.E. Signature record of activities and is maintained experts are ready to assist you to
Customer Signature for the entire working life of your maximize your productivity and
instrument (Figure 3). With this Log optimize your labs performance.
Instrument Performance Verification
Book, users are able to keep records Backed by 60 years of leading edge
Next IPV Due Date
of all changes, modifications, service, instrumentation experience,
Figure 2. The IQ/OQ sticker affixed to the maintenance, parts, etc. All actions PerkinElmers uptime experts are
instrument helps you keep your instrument with regard to the PerkinElmer system passionate about exceeding customer
testing up-to-date. are traceable, a requirement for GLP expectations, every time. We offer
and GLAP. The instrument Log Book more factory-trained service people
Ensure youre in compliance has more than a dozen sections ranging than anyone else in the industry,
from who-to-reach at PerkinElmer available globally. Our team has
To help ensure your instrument to what reference material was used expertise across a broad range of
stays in compliance, the PerkinElmer the last time the unit was serviced. technologies so we can support your
Service Engineer will affix an IQ/OQ Maintaining an up-to-date Log Book entire lab. Through our extensive
sticker to the validated instrument of the overall history of the instrument internal global programs such as Six
(Figure 2). This sticker will help you or system will satisfy GLP and Sigma and ISO certification, we can
keep your instrument testing up-to- ISO-9000 requirements. If for any ensure consistent and reliable support
date by noting the compliance date reason you need a copy of the on a global basis. So no matter what
of the instrument and the date the PerkinElmer factory generated protocol, your laboratory size or budget, we can
next IPV should be performed. these documents are archived assist you in improving every aspect
electronically to ensure easy access. of your labs performance, from
instrument care and repair, training,
validation, consulting and more.
PerkinElmer Instruments
710 Bridgeport Avenue
Shelton, CT 06484-4794 USA
Phone: 800-762-4000 or
(+1) 203-925-4600
www.perkinelmer.com/instruments

2002 PerkinElmer, Inc. All rights reserved. PerkinElmer is a trademark of PerkinElmer, Inc.
All trademarks depicted are the property of their respective holders or owners.
PerkinElmer reserves the right to change this brochure at any time and disclaims liability for editorial, pictorial, or typographical errors.

D-6589 KG040205 Printed in USA

Potrebbero piacerti anche