Sei sulla pagina 1di 5

d e n t a l m a t e r i a l s 2 4 ( 2 0 0 8 ) 13241328

available at www.sciencedirect.com

journal homepage: www.intl.elsevierhealth.com/journals/dema

Optical characterization of bisphenol-A-


glycidyldimethacrylatetriethyleneglycoldimethacrylate
(BisGMA/TEGDMA) monomers and copolymer

Janne Lehtinen a, , Toni Laurila a , Lippo V.J. Lassila b , Pekka K. Vallittu b ,


Jukka Raty c , Rolf Hernberg a
a Tampere University of Technology, Institute of Physics, Optics Laboratory,
Tampere, Finland 33720
b University of Turku, Institute of Dentistry, Department of Prosthetic Dentistry

and Biomaterials Science, Turku, Finland


c University of Oulu, Measurement and Sensor Laboratory, Kajaani, Finland

a r t i c l e i n f o a b s t r a c t

Article history: Objective. The fundamental optical characteristics of dental polymers as a function
Received 16 October 2006 of curing time are exiguously reported in the literature. In this work the refrac-
Received in revised form tive index [1] and the extinction coefcient [2] of bisphenol-A-glycidyldimethacrylate
30 January 2008 triethyleneglycoldimethacrylate (BisGMA/TEGDMA)-based polymers of different volume
Accepted 11 February 2008 fractions have been studied as a function of the curing time. A comparison between the mea-
sured index of refraction and theoretical calculation based on molar refraction is made. Also
the temporal behavior of both the refractive index and extinction coefcient is compared.
Keywords: Methods. Index of refraction: BisGMA/TEGDMA monomer samples were cured with a
BisGMA commercial light-curing unit by using camphorquinone and DMAEMA as the photoini-
TEGDMA tiator system. Curing times of 0, 40, 60, 300, 900 and 1800 s were used. The refractive
Optical properties indices were determined with a refractometer at 470 nm wavelength, where the peak
Optical characterization absorption of the camphorquinone photoinitiator is located. Extinction coefcient: Bis-
Dental composites GMA/TEGDMA samples having different lengths were cured and the relative attenuation
Dental polymer of light at 470 nm was measured as a function of the curing time (0, 40, 60, 300, 900,
Monomer 1800 s).
Refractive index Results. The results show that in general the index of refraction increases during the pho-
Extinction coefcient topolymerization process in a time frame from 0 to 40 s and then approaches a steady-state
value, which depends on the original composition of the admixture. The detailed refractive
index change depends on the volume fractions of the constituent substances. The extinc-
tion coefcient decreases in a time frame from 60 to 900 s and reaches a steady-state value
thereafter.
Signicance. Knowledge of the fundamental optical characteristics of commonly used dental
polymers is essential in the development of advanced light-initiated curing methods.
2008 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.


Corresponding author. Tel.: +358 3 7418883; fax: +358 3 31152090.
E-mail address: janne.lehtinen@tut. (J. Lehtinen).
0109-5641/$ see front matter 2008 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.dental.2008.02.012
d e n t a l m a t e r i a l s 2 4 ( 2 0 0 8 ) 13241328 1325

60, 300, 900 or 1800 s. Cured samples were stored in a freezer


1. Introduction (20 C) for 3 days before the actual measurements.
The degree of monomer conversion (DC%) was monitored
The light curing of dental resins and resin composites in small
with Fourier transform infrared spectroscopy (FT-IR) (Spec-
volumes using halogen lamps or more modern LED (light emit-
trum One, PerkinElmer) having an attenuated total reectance
ting diodes) units is a routine operation. However, efcient
(ATR) sampling accessory. Each sample was powdered and
single-step curing of larger volumes is still a technical chal-
placed on the surface of a ZnSe-ATR crystal. DC% was
lenge. The need for larger resin and resin composite volumes
calculated from the aliphatic C C peak at 1638 cm1 , and
exists in several applications. The current trend is in the use
normalized against the aromatic C C peak at 1608 cm1
of non-metallic materials in implants and bone repairs. More-
according to the formula:
over, in dentistry the understanding of the behavior of the
resin systems optical properties is of importance. However,
 
Caliphatic /Caromatic
large volume polymerization techniques may be limited by DC% = 1 100% (1)
Ualiphatic /Uaromatic
the dimensional changes due to the polymerization shrinkage
[35].
where Caliphatic = absorption peak at 1638 cm1 of the cured
Dental resin composites based on BisGMA (2,2-bis[4-(2-
sample, Caromatic = absorption peak at 1608 cm1 of the cured
hydroxy-3-methacrylyloxypropoxy)phenyl]propane) are com-
sample, Ualiphatic = absorption peak at 1638 cm1 of the
monly used in dental restorations and in some bone cements
uncured sample, Uaromatic = absorption peak at 1608 cm1 of
[68]. However, despite the wide use of various light curable
the uncured sample.
BisGMA-based resin mixtures in dentistry, their fundamental
The fraction of the remaining double bonds for each spec-
optical properties at the curing wavelength during the curing
trum was determined by using a standard baseline technique,
process are relatively weakly covered.
in which the heights of aliphatic and aromatic peaks are com-
The refractive index of bulk material describes how light
pared.
is refracted at the material boundaries. On the other hand,
The index of refraction was measured in the Measurement
a materials extinction coefcient describes the attenua-
and Sensor Laboratory (Kajaani, Finland) using a commercial
tion of light per unit length as it propagates through the
ABBE-refractometer (ABBE 60/ED, Bellingham & Stanley Ltd.)
material. These two wavelength-dependent coefcients are
[11]. As a light source a broadband xenon lamp (Oriel Ltd.,
essential in the characterization of the materials optical prop-
66002) was used. The lamp light was coupled to the refrac-
erties, e.g., absorbance, transmittance and reectance. In this
tometer through a monochromator (Oriel Ltd., 77250) in order
work the fundamental optical quantities, refractive index
to select the desired blue 470 nm spectral range. The band-
and extinction coefcient, were experimentally determined
width of the blue light was 5 nm. In the refractometer the
as a function of the curing time for the BisGMA/TEGDMA
blue light beam was directed to the front end of the sample
monomer/polymer system. The measurements were made
volume and the critical angle of total internal reection was
using blue light of blue 470 nm wavelength. The peak absorp-
measured [12]. The refractive index can be calculated from the
tion wavelength of the commonly used camphorquinone (CQ)
angle of the total internal reection. To enhance diffuse scat-
photoinitiator is located at 470 nm spectral region, and there-
tering of the cured solid polymer samples, the front surface of
fore it is the most relevant wavelength region in CQ-based
each sample plate was roughened with abrasive paper. Diffuse
curing systems [9,10].
scattering at the front surface was necessary to improve the
accuracy of the measurement.
The index of refraction of an amorphous polymer can be
2. Materials and methods calculated approximately by using the molar properties of
a single-polymer structural unit. For every chemical bond
Four types of BisGMA/TEGDMA monomer admixtures were that contributes to the structural unit, bond refraction can be
used in the measurement of the index of refraction. In dened. Summing of these properties gives the molar refrac-
all the admixtures CQ and DMAEMA (dimethylaminoethyl tion of a molecule, from which an approximate value of the
methacrylate) were used as the photoinitiator system and bulk index of refraction can be obtained with several meth-
TEGDMA (triethylene glycol dimethacrylate) as a diluent. Bis- ods. One of these methods is the molar refraction according

GMA/TEGDMA volume fractions were 50/50, 70/30, 80/20 and to Vogel, which is dened as RV = R = nM, where RV is
i V,i
99/1. A TRIAD 2000TM Visible Light Curing Unit (Dentsply Inter- the total molar refraction, RV,i the bond refraction of the con-
national Ltd.), with 275 W halogen light source, was used for stituent, n the index of refraction and M the molar mass of
the curing. the particular polymer molecule. The indices of refraction for
Samples for the refractive index measurements BisGMA and TEGDMA were calculated based on the molar
were fabricated using a stainless steel mold having masses and tabulated values for bond refractions of the con-
10 mm 40 mm 2 mm sample volume. The surface of stituent bonds [13]. The bond refraction values determined
the steel mold was coated with mold wax to facilitate at the standard yellow Fraunhofer line of sodium (589 nm)
the detachment of the cured samples. After injecting the were used. The refractive index of the heterogeneous Bis-
BisGMA/TEGDMA monomer admixture into the mold, a GMA and TEGDMA polymer admixtures were calculated by
borosilicate glass plate (5-mm thick) was placed on top of weighing the two compounds indices of refraction with their
the mold to ensure the smoothness and consistency of the volumetric fractions [14]. With this method, the heteroge-
volume of the samples. Samples were then cured for 0, 40, neous index of refraction is calculated with the following lower
1326 d e n t a l m a t e r i a l s 2 4 ( 2 0 0 8 ) 13241328

Table 1 Theoretical refractive indices with their lower


and upper boundaries for the BisGMA/TEGDMA polymer
admixtures studied in this work
BisGMA/TEGDMA (%) n

50/50 1.5451.547
70/30 1.5611.563
80/20 1.5691.571
99/1 1.585


and upper limits for the refractive index: 1/n2low = (v /n2 ),
 2 i i i
n2high = v n
i i i
, where nlow and n high are the lower and upper
limit for the index of refraction, i is the volumetric fraction of
each compound in a heterogeneous polymer and ni the refrac-
tive index of each compound.
For the measurement of the extinction coefcients half-
cylindrical Teon molds with borosilicate glass plate ends
were made. The sample lengths were 13, 5, 7 and 10 mm. Bis-
GMA/TEGDMA admixtures (50/50, 70/30, 99/1) were injected
into the molds and cured with the TRIAD 2000TM light-curing
unit. Curing times were 0, 40, 60, 300, 900 and 1800 s and the
extinction coefcients were measured immediately after each
curing. In the measurement of the extinction coefcient a con-
ventional 55 W broadband halogen lamp was used as a light
source. A monochromator (Lomo MDR-2) was used to separate
a 4-nm spectral band around the central 470 nm wavelength.
After traversing the sample the light beam was focused to a
photodiode (Ophir Ltd., PD300-UV). For each cured sample the
relative attenuation of the blue light intensity was measured
as a function of the sample length. The extinction coefcient
was then calculated for each sample from the BeerLambert
law: Fig. 1 (a) Index of refraction as a function of curing time
I  1
with different BisGMA/TEGDMA admixtures. Standard
1
 = ln deviation caused by the ABBE-refractometer was negligible
Ii Li L1
so it is not indicated in the gure. (b) The difference
where  is the extinction coefcient, I1 is the measured between the index of refraction for the unpolymerized and
intensity from the 1-mm long sample, which was used as a 1800 s cured resin as a function of the BisGMA
reference, Ii is the measured intensity from the samples with concentration.
lengths i = 2, 3, 5, 7 and 10 mm.

3. Results on the molar ratio of the monomer system. Fig. 1b shows that
an increase in the BisGMA content decreases the refractive
The dependence of the indices of refraction on curing time index difference between the non-polymerized and polymer-
and on BisGMA concentration are shown in Fig. 1a and b, ized samples. Table 1 shows the theoretical refractive indices
respectively. From Fig. 1a it can be observed that for all the of different admixtures based on the molar refraction approx-
admixtures a similar change in the refractive index appears imation with lower and upper limits. The theoretical and
during the curing from 0 to 40 s. During this time interval measured values are in good agreement.
the index of refraction increases rapidly and then stabilizes The dependence of the extinction coefcients of differ-
to a nal value thereafter. Refraction indices are dependent ent admixtures as on curing time is summarized in Table 2.

Table 2 Extinction coefcients as a function of time for different BisGMA/TEGDMA admixtures


 (cm1 ) Curing time (s)

0 40 60 300 900 1800

50/50 3.3 0.5 3.4 0.4 3.3 0.5 0.7 0.1 0.4 0.1 0.5 0.1
70/30 3.3 0.1 3.4 0.6 3.4 0.6 1.0 0.5 0.6 0.3 0.8 0.3
99/1 2.7 0.5 2.6 0.4 2.5 0.4 0.8 0.3 0.4 0.1 0.5 0.2

Measurement error is expressed as standard deviation.


d e n t a l m a t e r i a l s 2 4 ( 2 0 0 8 ) 13241328 1327

tion is, in many cases, a rapid process. According to the results,


polymerization increases the refractive index.
On the other hand, the absorption of light at 470 nm
depends mainly on the amount of photoinitiator (CQ). The
curing radicalizes the photoinitiator molecules, which then
bond with the monomer molecules in order to enable free rad-
ical chain polymerization [16]. Bonding with the monomers
reduces the amount of the photoinitiator that can absorb the
curing light and therefore the extinction coefcient at 470 nm
decreases during curing. The extinction coefcient decreases
until the extinction is dominated by the BisGMA/TEGDMA
mixture. To verify the reduction of CQ as a function of
the curing time the transmittance was monitored with a
UVvis recording spectrophotometer (Shimadzu Corp., UV-
2501PC). The transmittance of 2-mm thick slabs of 50/50
BisGMA/TEGDMA polymer admixture was measured with dif-
Fig. 2 Extinction coefcients of BisGMA/TEGDMA 50/50 ferent curing times. From Fig. 3 it can be observed that the
and 99/1 admixtures as a function of the curing time. Error absorption due to CQ decreases as a function of curing time.
bars indicate standard deviation. This decrease is pronounced at longer curing times. The
authors suggest that this explains the difference in the tem-
poral behavior between the extinction coefcient and the
refractive index.
Fig. 2 shows the dependence of the extinction coefcient of
It needs to be emphasized that the present study deter-
the 70/30 BisGMA/TEGDMA admixture on curing time. Unlike
mined optical properties of resins only. In many dental
the index of refraction, the extinction coefcient remains con-
applications the resins are used as a part of compos-
stant at the beginning of the curing process. However, from 60
ite materials, where particulates or bers are used as
to 900 s a decrease in the extinction coefcient occurs. Further
llers and structural reinforcements. The detailed optical
curing after 900 s does not have any signicant effect on the
behavior of isotropic particulate ller composite resins and
materials extinction coefcient.
iso/ortho/anisotropic ber-reinforced composite resins [17]
needs to be investigated further.
4. Discussion The results obtained in this work aid in the development of
novel and optimized curing methods for larger volume resins,
The present study revealed that the index of refraction for instance in root canal applications, where light transmis-
increases with polymerization of the resin system. At the time sion to the apical part of the root is not adequate [18]. There
point of 40 s a plateau in the index of refraction was reached are also interests to develop novel single-step curing methods
for the resin mixtures containing substantial quantities of for larger volume medical polymer composites [19]. If low-
TEGDMA resin, whereas the resin mixture predominantly of shrinking polymer systems investigated here, or those based
BisGMA showed a slight ongoing increase in the index even on ring-opening polymerization [20] mature to clinical use,
after longer curing times up to 300 s. Thus the highly viscous the characterization of the optical properties described in the
BisGMA resin takes more time to polymerize by light activation present study could be of great importance.
than those monomer mixtures which have been diluted with,
e.g., monomer of TEGDMA. A comparison between the theo-
retical and measured values shows good agreement despite
the difference between the wavelengths of the measured and
calculated values. The values of theoretical calculations differ
at most by 1% from the measured ones. In BisGMA/TEGDMA
admixtures with the 50/50 and 70/30 volume fractions, this dif-
ference can be explained by the wavelength dependence of the
refractive index. At higher volume fractions, 80/20 and 99/1,
the theoretical molar refraction approximation yields slightly
too high refractive index values. The extinction coefcients of
the resin mixtures decreased with curing time. After 60 s of
curing, the coefcient of extinction drops considerably up to
about 900 s.
The difference in the temporal behavior of the extinc-
tion coefcient and the refractive index can be explained by
exploring the fundamental reasons for these two phenomena.
Refractive index is basically related to the molecular proper- Fig. 3 The transmission of BisGMA/TEGDMA admixture
ties (polarizability) and the density of the substance [15]. The with 50/50 volume fraction as a function of the wavelength
reorganization of the molecular structure during polymeriza- for different curing times.
1328 d e n t a l m a t e r i a l s 2 4 ( 2 0 0 8 ) 13241328

[9] Nomoto R. Effect of light wavelength on polymerization of


Acknowledgements light-cured resins. Dent Mater 1997;23:6073.
[10] Cook WD. Kinetics and properties of a photopolymerized
The research was nancially supported by the Technology dinmethacrylate oligomer. J Appl Polym Sci 1991;42:
Agency of Finland (Tekes). The authors thank Emilia Anttila 220922.
[11] Rheims J, Koser J, Wriedt T. Refractive-index measurements
and Outi Krintila for their kind help with the transmission
in the near-IR using an Abbe refractometer. Meas Sci
measurements.
Technol 1997;8:6015.
[12] Hecht E. Optics. 4th ed. Addison Wesley; 2002. p. 123.
references [13] van Krevelen D. Properties of polymers. 3rd ed. Elsevier
Science; 1997.
[14] Brandrup J. Polymer handbook. 4th ed. John Wiley & Sons
Inc; 1998.
[1] Hecht E. Optics. 4th ed. Addison Wesley; 2002. p. 66.
[15] Atkins PW, Friedman RS. Molecular quantum mechanics. 3rd
[2] Bohren CF, Huffman DR. Absorption and scattering of light
ed. Oxford University Press; 2000. p. 397.
by small particles. John Wiley & Sons Inc.; 1998. p. 69.
[16] Hart H, Crane L, Hart D. Organic chemistry. 11th ed.
[3] Cook WD, Beech DR, Tyas MJ. Structure and properties of
Houghton Mifin Company; 2003. p. 402.
methacrylate based dental restorative materials.
[17] Tezvergil A, Lassila LVJ, Vallittu PK. The effect of ber
Biomaterials 1985;6:3628.
orientation on the thermal expansion coefcients of the
[4] Ferracane JL. Developing a more complete understanding of
ber reinforced composites. Dent Mater 2003;19:
stresses produced in dental composites during
4717.
polymerization. Dent Mater 1984;21:3642.
[18] Le Bell A-M, Tanner J, Lassila LVJ, Kangasniemi I, Vallittu PK.
[5] Versluis A, Douglas WH, Cross M, Sakaguchi RL. Does an
Depth of light initiated polymerization of glass ber
incremental lling technique reduce polymerization
reinforced composite in a simulated root canal. Int J
shrinkage stresses? J Dent Res 1996;75:8718.
Prosthodont 2003;16:4038.
[6] Peutzfeldt A. Resin composites in dentistry: the monomer
[19] Tuusa SM-R, Puska MA, Lassila LVJ, Vallittu PK. Residual
systems. Eur J Oral Sci 1997;105:97116.
monomers released from glass bre-reinforced composite
[7] Lassila LVJ, Nohrhstron T, Vallittu PK. The inuence of short
photopolymerized in contact with bone and blood. J Mater
term water storage on the exural properties of
Sci Mater M 2005;16:1520.
unidirectional glass ber-reinforced composite. Biomaterials
[20] Eick JD, Kostoryz EL, Rozzi SM, Jacobs DW, Oxman JD,
2002;23:22219.
Chappelow CC, et al. In vitro biocompatibility of
[8] Belkoff SM, Mathis JM, Erbe EM, Fenton DC. Biomechanical
oxirane/polyol dental composites with promising physical
evaluation of a new bone cement for use in verteplasty.
properties. Dent Mater 2000;18:41321.
Spine 2000;25:10614.

Potrebbero piacerti anche