Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Article in Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with
Materials and Atoms January 2004
DOI: 10.1016/S0168-583X(03)01582-9
CITATIONS READS
4 37
3 authors, including:
Walid A. Metwally
University of Sharjah
24 PUBLICATIONS 129 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Walid A. Metwally on 11 May 2015.
The user has requested enhancement of the downloaded file. All in-text references underlined in blue are added to the original document
and are linked to publications on ResearchGate, letting you access and read them immediately.
Nuclear Instruments and Methods in Physics Research B 213 (2004) 357363
www.elsevier.com/locate/nimb
Abstract
A semi-empirical model is derived and tested for a 90 Sr beta-particle transmission thickness gauge for aluminum
alloys. Monte Carlo simulation is also used to verify the forward scatter part of the model. The model accurately
accounts for thickness as well as forward scatter, gamma-ray background, and the measurement interference of sample
density and composition. The model parameters are obtained and the model is benchmarked with industrial experi-
mental data from a 90 Sr(90 Y) gauge.
2003 Elsevier B.V. All rights reserved.
Beta-particle transmission gauges have long The semi-empirical model proposed and tested
been used and are industrially very important [2,3] here is based on the use of fundamental macro-
for measuring and controlling the thickness of scopic cross-sections for the basic interactions of
rolled products such as paper, plastic, and metal. beta-particles, combined with the assumption that
Beta-particle sources of industrial importance in the transmission of beta-particles is exponential up
this application include 90 Sr(90 Y), 204 T1, 85 Kr and to some characteristic maximum range.
147
Pm. While several exponential and linearized The fundamental macroscopic cross-sections
models have been proposed and used for calibra- for two basic interactions are used. They are: (1)
tion purposes, the authors are not aware of any scattering and (2) attenuation. The form of the
models that explicitly account for composition and scattering interaction XS is taken as
the forward scatter inherent to beta-particle Xn
0168-583X/$ - see front matter 2003 Elsevier B.V. All rights reserved.
doi:10.1016/S0168-583X(03)01582-9
358 R.P. Gardner et al. / Nucl. Instr. and Meth. in Phys. Res. B 213 (2004) 357363
The XS parameter accounts primarily for large where kA and kS are constants for a given beta-
angle scatter while the XA parameter accounts particle transmission gauge. The model then be-
primarily for small energy losses along a beta- comes
particle path that does not result in signicant
changes of direction. Rt R0 expkA XA qt kS XS qt B: 5
2.1. Thin-beam or good geometry Note that no energy dependence is given in this
model. It is reasoned that one eective energy can
For a beta-particle transmission gauge with be considered and, therefore, energy dependence in
thin-beam geometry, the basic gauge response Rt the model can be omitted. This could be modied
is taken as if found necessary to include two or more eective
Rt R0 expl=qqt B; 3 energies by using the sum of multiples of the pre-
3 sent model. However, the present model already
where q is sample density in g/cm , t is the sample
contains six unknown parameters (R0, kA , XA , kS ,
thickness in cm, l is the semi-empirical sample
XS and B) that must be obtained from experi-
attenuation coecient in cm1 and B is the back-
mental data and this must be increased to account
ground response. This model is modied to include
for the two beta-particle spectra of the 90 Sr(90 Y)
sample composition explicitly by substituting for
sources.
l=q:
When the two beta-particle spectra present in a
l=q kA XA kS XS ; 4 90
Sr(90 Y) source are considered, one must use the
1000
Experiment
Transmission Model
900
700 +18.0Exp(0.0414XAt)
600
RESPONSE (CDS)
500
400
300
200
100
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
2
SAMPLE DENSITY X THICKNESS (g/cm )
sum of two transmission terms. The model then where kc and lc are parameters associated with the
becomes gamma-ray response. The linear attenuation co-
ecient lc for pure aluminum is 0.0414 cm1 . Now
Rt R0expk1A XA qt k1S XS qt
the model has two additional parameters for a
expk2A XA qt k2S XS qt B; 6 total of 10. If we assume that lc is constant and
where the subscripts 1 and 2 refer to the rst and equal to the value for pure aluminum, and that XA
second beta-particle spectra, respectively. This and XS are known for a given known sample, we
model contains two additional parameters for a still have seven parameters in the model that must
total of eight. be determined by the least-squares analysis of data
In the case of the 90 Sr (90 Y) source, there is a from standard known samples.
small amount of gamma radiation (a 2.1862-MeV
gamma-ray) associated with the source, so an ex- 2.2. Broad-beam or poor geometry
ponential attenuation term must be added for it as
well. Now the model becomes In the case of broad-beam geometry, one also
has an added forward scatter component. The
form of this component is taken as
Rt R0expk1A XA qt k1S XS qt
a
expk2A XA qt k2S XS qt F t CS XS qt R0 expcA XA qt cS XS qt:
kc explc XA qt B; 7 8
250
150
RESIDUALS (CDS)
100
50
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
2
SAMPLE DENSITY X THICKNESS (g/cm )
With this modication the nal total model for the 3. Determination of model parameters
broad-beam 90 Sr beta-particle transmission gauge
becomes The model parameters were determined using
the code CURMOD developed at CEAR, which is
Rt R0expk1A XA qt k1S XS qt a code that is capable of determining nonlinear
expk2A XA qt k2S XS qt parameters by a nonlinear (Marquardt) search
while simultaneously determining the linear pa-
CS XS qta R0 expcA XA qt cS XS qt rameters by a linear method. This code is a mod-
kc explc XA qt kN : 9 ication and extension of the code CURFIT
described in the text by Bevington [1] which was
So the nal model contains 11 parameters that based on the algorithm suggested by Marquardt
must be determined from a least-squares t to [4] for optimal nonlinear searching by optimum
experimental data seven nonlinear and four lin- combined use of gradient and linearized methods.
ear. One of the nonlinear parameters might be Other codes familiar to the authors that search
taken as zero and omitted the coecient CS for on multiple model parameters when some of the
the scatter attenuation in the forward scatter re- parameters are nonlinear, must do so by using the
sponse. One might also make the parameter a nonlinear search method for all parameters. For
unity or some other xed integer. If these two example, the sample problem in Bevington [1] does
things are done (cS 0 and a 1), the number of this. This approach has the disadvantage that each
parameters is reduced to nine, six nonlinear and new nonlinear parameter that is introduced to a
three linear. model increases the degree of diculty by almost
an order of magnitude.
4. Model t results
ported for 90 Sr in standard texts (like [3]) and a tration distance into the aluminum alloy sheet
trial-and-error stripping method. (The model used before a large angle scatter was chosen from the
is Eq. (7) with R0 491:0, k1A 18:0, k1S 1:32, pertinent exponential distribution with the ap-
k2A 3:85, k2S 0:283, kc 18:0, lc 0:0414 and propriate linear attenuation coecient, the large
B 0.) With this approach the model t shown in angle scatter was chosen from an isotropic distri-
Fig. 1 with the residuals shown in Fig. 2 were bution, the scatter energy loss was calculated from
obtained. It then occurred to the authors that the the appropriate energy balance equation, and at-
missing feature was probably caused by forward tenuation to and from the large angle scatter to-
scatter. ward the detector within the aluminum sheet was
calculated from the appropriate simplied power
4.2. Monte Carlo simulation of forward scatter law range equation. A schematic diagram of this
process is shown in Fig. 3.
To further substantiate this idea, a simple The results of this Monte Carlo simulation are
Monte Carlo simulation of the forward scatter was shown compared to the values obtained experi-
attempted. In this simulation the beta-particle mentally by the previously described approach in
emission angle was assumed to be isotropic from Fig. 4. The Monte Carlo results were normalized,
0 to 90, the beta-particle energy was chosen from but the shape of the predicted forward scatter
the pertinent distribution by rejection, the pene- for both approaches matches well, verifying that
250
Residuals
Monte Carlo Simulation
200
150
RESIDUALS (CDS)
100
50
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
2
SAMPLE DENSITY X THICKNESS (g/cm )
Fig. 4. Residuals of experimental data compared to Monte Carlo simulation of forward scattering.
362 R.P. Gardner et al. / Nucl. Instr. and Meth. in Phys. Res. B 213 (2004) 357363
250
Experiment
Model
A
RS=A4(XSt) 3 Exp(A1XAtA2XSt)
200
A1=3.0944576, A2=0.4092344
A =0.5853393, A =408.0195316
3 4
ChiSquare=0.116
150
RESIDUALS (CDS)
100
50
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
2
SAMPLE DENSITY X THICKNESS (g/cm )
1000
Experiment
Model
Differences X 10
800 RC=491.4Exp(17.0XAt1.22XSt)
+491.4Exp(4.22XAt0.302XSt)
+408.88(X t)0.636Exp(3.04X t0.400X t)
S A S
+17.4Exp(0.0414XAt)
600
RESPONSE (CDS)
400
200
200
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
2
SAMPLE DENSITY X THICKNESS (g/cm )