E007 Filament Fault Power supply has detected 1. Check for open
filament current <2 amps. filament in X-ray
tube.
2. Check for poor
connections in the
cathode cable.
3. Check fuses on
filament board(s).
E009 Power Supply Not Ready Power supply is not ready Retry
to exposure.
make an exposure.
E011 MA During Exposure Too High Generator CPU detected mA Recalibrate X-ray tube.
greater than allowed tolerance.
5.3.3 Error Messages (Cont)
ERROR
E012 CODE MA DuringMESSAGE
Exposure Too Low PROBLEM
Generator CPU detected mA ACTION Recalibrate X-ray tube.
E015 AEC mAs Exceeded - Exposure AEC less exposure
than allowed tolerance.
exceeded allowed 1. Check exposure
Terminated back up mAs. technique settings.
E013 Manually Terminated Exposure Operator released exposure Re-take exposure
2. Check that correct if
switch during exposure. necessary.
AEC chamber is
Check for faulty switch
energized.
contacts or wiring.
E016 Tomo Back-up Timer - Tomo exposure exceeded 1. Check exposure
E014 AEC Back-up
Exposure Timer - Exposure
Terminated AEC exposure
back up time. exceeded allowed 1. Check exposure
technique
Terminated back up time. technique
settings.settings.
2. Increase that
Check tomo correct
back up AEC
time
chamber is
if necessary. energized.
E020 Thermal Switch Interlock #1 X-ray tube # 1 too hot and Wait for X-ray tube #
Error its thermal switch has opened.1 to cool.
E021 Thermal Switch Interlock #2 X-ray tube # 2 too hot and Wait for X-ray tube #
Error its thermal switch has opened.2 to cool.
E029 X-Ray Input Active During X-ray input active during power Check X-ray switch and
Initialization Phase on initialization phase. input for short circuit.
E030 Fluoro Input Active During Fluoro input active during Check fluoro switch
Initialization Phase power on initialization phase. and input for short circuit.
5.3.3 Error Messages (Cont)
E049
E044 Not Enabled
Invalid Requestedcommunication
Communication Message Received function not ResetReprogram
error. to enable
programmed
message not to be and
valid enabled.
ignored. function.
E050 Generator Limit Data Corrupt Generator detects corrupt Reprogram generator
generator limit data. limit data or set factory
defaults.
E051 AEC Feedback Error (No Generator has detected no 1. Check that X-ray tube
Feedback Signal Detected) feedback signal from AEC is pointing at correct
device. AEC device.
2. Check AEC cable for
damage and proper
connection.
E052 High Small focus filament Generator detects small focus Check small focus
current error in standby filament current greater than filament board.
limits in standby mode.
E053 High Large focus filament Generator detects large focus Check large focus
current error in standby filament current greater than filament board.
limits in standby mode.
E054 AEC Reference out of range AEC reference has reached a Re-adjust AEC
maximum or minimum limit. calibration including
density to operate within
AEC range (O to
1OVDC).
E055 No fields selected in AEC AEC enabled but no fields are Select AEC field(s).
selected.
E056 Receptor Disabled All Receptors have no X-ray Program receptor(s) with
tube programmed. tube number.
5.3.3 Error Messages (Cont)
E058 Console Back-up Timer Console has detected exposure Call product support.
exceeded backup time and
terminated exposure.
E059 Housing Heat Limit Exceeded X-ray tube housing has Wait for tube to cool.
exceeded limit.
E060 High kV Error KV exceeds high KV tolerance 1. Check the output of
level. the kV reference
DAC on the
generator CPU
board.
2. Measure the output
of the generator with
a dynalyzer or a non-
invasive kVp meter.
E062 EXP_SW Signal Active in The EXP_SW signal on the Call product support.
Standby State generator Interface and
generator CPU board is enabled
when it should be disabled.
E063 Factory Defaults Set SW1 switch 8 on the generator Set SW1 switch 8 to its
CPU board is set to default the non default position. The
generator CPU NVRAM with generator will not exit the
factory defaults. initialization phase until
this switch is set.
5.3.3 Error Messages (Cont)
E067 Power Supply Duty Cycle Limit Power supply duty cycle limit Re-evaluate technique
exceeded. factors. Allow generator to
cool if possible. If
exposures are continued,
serious generator damage
may result due to
overheating.
E070 Generator software key error Defective or missing GAL U29 Call product support for
on generator CPU board new GAL U29.
734573.
E071 DAP overflow error The accumulated DAP value Reset the DAP.
exceeds the display limit.
E072 DAP device error The DAP device is not 1. Check the DAP
functional. wiring.
2. Check the DAP
interface board.
E073 DAP data error The DAP configuration data is Reset factory defaults.
corrupted.
E100 Calibration Error - Maximum mA Maximum mA has been exceed Repeat auto calibration
Exceeded during auto calibration. and/or decrease standby
current.
E101 Calibration Error - Calibration Auto calibration has exceeded 1. Check to see if the
Data Table Exceeded data table length due to an filament standby
excessive number of exposures. current is too low.
2. Retry auto
calibration.
E102 Calibration Error - Maximum Maximum filament current for the 1. Check to see if the
Filament Current Exceeded selected focus has been maximum filament
exceeded. current limit can be
increased.
2. Retry auto
calibration.
E103 Calibration Error - Manually Operator released exposure Retry auto Calibration.
Terminated button during auto calibration.
E105 Calibration Error - Minimum mA Minimum generator mA was Reduce filament standby
Not Calibrated exceeded at start of calibration. current on primary
This is usually caused by too and/or secondary
high a filament standby current filament.
on the primary and or secondary
filament. (Primary is the current
filament being calibrated,
secondary is the other filament
and applies only to generators
with two independent filament
supplies).