Sei sulla pagina 1di 12

The

World
Leader In
3D Surface
Metrology

SEM and AFM: Complementary Techniques


for High Resolution Surface Investigations

There are a wide range of analyti- formation mechanisms are quite History
By
cal techniques which may be used different, resulting in different
P. Russell, The first SEM was constructed in
for materials characterization types of information about the
D. Batchelor, 1938 by von Ardenne by rastering
depending on the type of infor- surface structure. The occurrence
J. Thornton the electron beam of a Transmis-
mation needed. For high resolu- of the SEM and AFM side-by-side
sion Electron Microscope (TEM)
tion surface investigations, two is becoming more common in
to essentially form a Scanning
commonly used techniques are todays analytical laboratories.
Transmission Electron Microscope
Atomic Force Microscopy (AFM) This article will compare and
(STEM) (1, 2). In 1942, Zworkin
and Scanning Electron Micros- contrast the two techniques with
et. al. developed the first SEM for
copy (SEM)Figures 1 and 2, respect to specific types of surface
bulk samples. This configuration
respectively. Each of these measurements, and demonstrate
contains many of the basic
techniques resolves surface how these analytical techniques
principles of todays SEMs (2, 3).
structure down to the nanometer provide information which is
Cambridge Scientific Instruments
scale. However, the image complementary in nature.

Figure 1. Schematic of the major components of an AFM Figure 2. Schematic of the primary components of a
showing the feedback loop for TappingModeTM operation. typical SEM
produced the first commercial used form of SPM, many other
instrument in 1965. A number of SPM techniques have been
improvements have occurred since developed which provide informa-
this time, resulting in an increase tion on differences in friction,
in resolution from 50nm in 1942 adhesion, elasticity, hardness,
to ~0.7nm today. Besides the electric fields, magnetic fields,
development of morphological carrier concentration, temperature
imaging, the SEM has been distribution, spreading resistance,
developed to detect signals which and conductivity.
are used to determine composi-
tional information, such as X-rays,
Imaging Mechanisms
backscattered electrons,
cathodoluminesce, Auger elec-
Scanning Electron Microscopy
trons, and specimen current.
The operation of the SEM
The development of the AFM was consists of applying a voltage
preceded by the development of between a conductive sample and
Figure 3. SEM image of an integrated single crystal the Scanning Tunneling Micro- filament, resulting in electron
silicon cantilever and tip which has an end radius of 2 scope (STM) in 1981 at IBM emission from the filament to the
to 10nm. Tips for AFM are typically made of silicon or Zurich Research Laboratory by sample. This occurs in a vacuum
silicon nitride. Bar=100m.
Binnig and Rohrer (4). Its ability environment ranging from 10-4 to
to view the atomic lattice of a 10-10 Torr. The electrons are
sample surface earned the inven- guided to the sample by a series of
tors the Nobel Prize in Physics in electromagnetic lenses in the
1986. Although the STM pro- electron column. A schematic of a
vides subangstrom resolution in all typical SEM is shown in Figure 2.
three dimensions, it is limited to The resolution and depth of field
conductive and semiconductive of the image are determined by
samples. To image insulators as the beam current and the final
well as conductors, the Atomic spot size, which are adjusted with
Force Microscope (AFM) was one or more condenser lenses and
developed in 1986 (5), and the the final, probe-forming objective
first commercial AFMs were lenses. The lenses are also used to
produced in 1989 by Digital shape the beam to minimize the
Instruments. effects of spherical aberration,
chromatic aberration, diffraction,
AFM provides three-dimensional and astigmatism.
surface topography at nanometer
lateral and subangstrom vertical The electrons interact with the
Figure 4: TappingMode AFM image of 1.4 resolution on insulators and sample within a few nanometers
monoatomic steps on epitaxial silicon deposited on
conductors. From this beginning, to several microns of the surface,
(100) Si. 1m scan.
the field of Scanning Probe depending on beam parameters
Microscopy (SPM) was born and sample type. Electrons are
which consists of a family of emitted from the sample primarily
techniques that involves scanning a as either backscattered electrons or
sharp tip across the sample surface secondary electrons. Secondary
while monitoring the tip-sample electrons are the most common
interaction to form a high resolu- signal used for investigations of
tion image. Although the AFM surface morphology. They are
has become the most commonly produced as a result of interac-
2
tions between the beam electrons conducted by a piezoelectric
and weakly bound electrons in the tube scanner which scans the tip (a)
conduction band of the sample. in a raster pattern with respect
Some energy from the beam to the sample (or scans to the
electrons is transferred to the sample with respect to the tip).
conduction band electrons in the The tip-sample interaction is
sample, providing enough energy monitored by reflecting a laser
for their escape from the sample off the back of the cantilever
surface as secondary electrons. into a split photodiode detector.
Secondary electrons are low By detecting the difference in
energy electrons (<50eV), so only the photodetector output
those formed within the first few voltages, changes in the cantile-
nanometers of the sample surface ver deflection or oscillation
have enough energy to escape and amplitude are determined. A
be detected. High energy beam schematic of this can be seen in (b)
electrons which are scattered back Figure 1.
out of the sample (backscattered
electrons) can also form secondary The two most commonly used
electrons when they leave the modes of operation are contact
surface. Since these electrons mode AFM and
travel farther into the sample than TappingModeTM AFM, which
the secondary electrons, they can are conducted in air or liquid
emerge from the sample at a much environments. Contact mode
larger distance away from the AFM consists of scanning the
impact of the incident beam probe across a sample surface
which makes their spatial distribu- while monitoring the change in
tion larger. Once these electrons cantilever deflection with the
escape from the sample surface, split photodiode detector. A Figure 5: (a) SEM image of rugged polysilicon thin
they are typically detected by an feedback loop maintains a film. 100,000x, Bar=0.1m; (b) TappingMode AFM
Everhart-Thornley scintillator- constant cantilever deflection by image of the same with roughness measurement.
photomultiplier detector. The vertically moving the scanner to 1m scan.
SEM image formed is the result of maintain a constant photodetec-
the intensity of the secondary tor difference signal. The
electron emission from the sample distance the scanner moves
at each x,y data point during the vertically at each x,y data point
rastering of the electron beam is stored by the computer to
across the surface. form the topographic image of
the sample surface. This
feedback loop maintains a
Atomic Force Microscopy
constant force during imaging,
AFM consists of scanning a sharp which typically ranges between
tip on the end of a flexible 0.1 to 100nN.
cantilever across a sample surface
while maintaining a small, TappingMode AFM consists of
constant force. An integrated oscillating the cantilever at its
silicon tip and cantilever can be resonance frequency (typically
seen in Figure 3. The tips ~300kHz) and lightly tapping
typically have an end radius of on the surface during scanning.
2nm to 20nm, depending on tip The laser deflection method is
type. The scanning motion is used to detect the root-mean-
3
square (RMS) amplitude of will discuss measurements of
cantilever oscillation. A feedback different vertical scales of topogra-
loop maintains a constant oscilla- phy, beginning with very smooth
tion amplitude by moving the surfaces and working up to very
scanner vertically at every x,y data rough surfaces to determine how
point. Recording this movement the surface topography affects the
forms the topographical image. ability of each technique to
The advantage of TappingMode perform the measurement.
with respect to contact mode is
that it eliminates the lateral, shear Atomically Smooth Surfaces
forces present in contact mode.
This enables TappingMode to Atomically smooth surfaces can
image soft, fragile, and adhesive occur either naturally, such as on
(a) surfaces without damaging them, mineral surfaces, or by processing,
which can be a drawback of such as polishing and epitaxial
contact mode AFM. growth on semiconductor, data
storage, and optical surfaces. A
TappingMode AFM image of an
Comparison of
epitaxial silicon surface is shown
Techniques
in Figure 4. Note that, unlike
There are a number of different SEM, the AFM can measure in all
ways to compare and contrast three dimensions (x, y, and z) with
these two techniques with respect a single scan. Since the AFM has
to each other. Although investi- a vertical resolution of <0.5, it
gations that use both SEM and can resolve the 1.4 monoatomic
AFM to characterize a material are silicon steps on the surface as well
(b) common, there are just a few as calculate an RMS roughness of
studies that directly discuss the 0.7 (14). On a sample this
complementary nature of the smooth, the SEM has difficulty
techniques (6-13). A comparison resolving these features due to the
of these techniques will be subtle variations in height.
conducted with respect to 3
factors: (1) Surface Structure, (2) Thin Films
Composition, and (3) Environ-
ment. The comparisons are On most thin films, the SEM and
presented for typical equipment AFM produce a similar represen-
configurations and operating tation of the sample surface. A
(c) procedures. common application of surface
investigations of thin films
consists of determining changes in
Surface Structure morphology with variations of
Figure 6. (a) SEM image of partially GaP-covered Si
after chemical beam epitaxy deposition for 10 minutes. Although both SEM and AFM are deposition parameters, such as
30,000x, Bar=1m; (b) AFM image of the same similar in lateral resolution, there temperature, pressure, time, etc.
sample as in figure 6a showing the presence of
are situations in which one Figure 5 shows SEM and AFM
nodules during the growth of GaP by chemical beam
epitaxy. 10m scan; (c) Cross-sectional measurement technique can provide a more images of a polysilicon thin film at
with AFM across the image in Figure 6b showing 3 complete representation of the approximately the same lateral
nodules which have a height of approximately 70nm. sample surface, depending on the magnification. The two images
(16) information desired. One show similar surface structure,
principle difference is in how the however, they differ in the other
two techniques process vertical types of information that can be
changes in topography. Below we
4
acquired on this sample. The tedious and time consuming.
three-dimensional nature of the Since the AFM data contains the
AFM can be used to calculate height information, determining
changes in roughness and surface whether a feature is a bump or pit
area variations due to differences is straightforward. As can be seen
in deposition parameters. For the in Figures 6b and 6c, the features
SEM, a large area view of the on this sample are bumps. This
variations in surface structure can information was used in the study
be acquired all at once (such as of the growth mechanisms of GaP
several mms), whereas a 100m x on Si during chemical beam
100m area is typically the largest epitaxy deposition (16). Determi-
area viewed by an AFM. These nation of whether these features
(a)
images are an example of rugged were small bumps or depressions
polysilicon films which are used as would have changed how the
capacitors in memory devices. By deposition process was altered to
making these films rough, the produce an epitaxial GaP film.
surface area is increased which
makes it possible to hold more High Aspect Ratio Structures
charge without increasing the
lateral dimensions of the capaci- Semiconductor processing
tors on the chips. By adjusting commonly requires measurements
the deposition parameters and of high aspect ratio structures such
using the AFM to analyze the as trenches and via holes. In a
surface area of the films, the SEM, these structures are typically
deposition parameters needed to measured in cross section by
produce a film with the maximum cleaving the wafer and imaging (b)
surface area were determined (15). the sample on end to obtain the
dimensions of the structure. A
Figure 7. (a) Cross-sectional SEM image of
Another example of the difference common example of this is seen in polysilicon lines which shows undercutting due to
between the two techniques is in Figure 7a. In contrast, the AFM reactive ion etching. Scale bar=100nm; (b)Cross-
interpreting subtle differences in image of a trench or via is made sectional measurement of developed and incompletely
height. In the SEM image, by scanning over the sample developed vias in photoresist acquired by
changes in slope can result in an surface. The ability of the AFM TappingMode AFM. In order to image the high aspect
ratio structures on the sample, a silicon tip machined
increase in electron emission from to measure these structures with a focused ion beam was used to scan the vias.
the sample surface, producing a nondestructively makes it possible 6.2m scan.
higher intensity in the image. for the wafer to be returned to the
However, it can sometimes be production line after the measure-
difficult to determine whether the ment is acquired. An AFM image
feature is sloping up or down. For of vias in photoresist is shown in
instance, in the SEM image in Figure 7b. To image some higher
Figure 6a it is very difficult to aspect ratio structures, the proper
determine whether the small tip shape is needed for the AFM
round structures are bumps or to scan narrow openings and steep
pits, even when tilting the sample sidewalls. Although the SEM
stage in the SEM. The only other measurement is destructive to the
option would be to cleave the sample, the ability to image the
sample through one of these undercuts of these lines is a useful
features and look at the sample in application that AFMs are not
cross-section, which would be typically designed to perform.

5
Rough Surfaces give a more complete "picture" of
the sample.
One of the key advantages of the
SEM with respect to other types
Composition
of microscopy is its large depth of
field. This ability makes it Both SEM and SPM provide
possible to image very rough compositional information
surfaces with millimeters of through a variety of techniques.
vertical information within a SEM is the only one of the two
single image. A SEM image of techniques which provides
non-woven polyethylene oxide elemental analysis, however, both
fibers can be seen in Figure 8a. SEM and AFM are associated with
(a) The depth of field and small beam techniques which can provide
size makes it possible to image the compositional information
fibers far below the top layer. This through analyzing materials and
ability also makes it possible to physical properties of the sample.
measure very rough surfaces over Some of the most common of
larger lateral areas as well. Al- these methods are described
though the AFM can measure below.
vertical surface variations below
0.5, its ability to measure a tall SEM
structure comes from how far the
scanner can move vertically. Along with the secondary electron
Standard scanners typically have 5 emission which is used to form a
to 6m of vertical range, however, morphological image of the
(b) in some configurations the vertical surface in the SEM, a number of
range approaches 10m or larger. other signals are emitted as a result
Figure 8. (a) SEM image of a non-woven textile For scanning areas that have of the electron beam impinging
sample of polyethylene oxide fibers. The large depth heights of greater than 5 to 10ms on the surface, as shown in Figure
of field of the SEM makes it possible to image fibers of variation, the SEM would be 9. Each of these signals carries
which are 10s of ms below the upper layer of fibers. better suited for the analysis. information about the sample
Bar=10m; (b) SEM image of Y2O3 crystal. Bar=1m.
which provide clues to its compo-
Another example of a complex sition.
three-dimensional surface struc-
ture which shows how the SEM Two of the most commonly used
and AFM can complement each signals for investigating composi-
other can be seen in Figure 8b. tion are x-rays and backscattered
The convoluted three dimensional electrons. X-ray signals are
Y2O3 oxide crystal shown growing commonly used to provide
out of a relatively flat Y2O3 thin elemental analysis by the attach-
film on a Si substrate is easily ment of an Energy-Dispersive
imaged in the SEM (Figure 8b). Spectrometer (EDS) or Wave-
Although the AFM would have length-Dispersive Spectrometer
probelms imaging the obtuse (WDS) to the SEM system. X-
angels and enclosed areas of this ray emission results from inelastic
surface, the roughness of the Y2O3 scattering between the beam
film can be measured whereas in electrons and the electrons of the
the SEM image the surface sample atoms. This interaction
roughness is not evident. There- results in the ejection of an inner
fore, the two techniques together shell electron from the atom,
6
creating a vacancy that is filled by electrostatic fields, carrier concen-
an outer shell electron. This jump tration, temperature distribution,
from an outer to inner shell results spreading resistance, and conduc-
in a change in energy that pro- tivity. Many of these techniques
duces either a x-ray or Auger consist of looking simultaneously
electron. The emitted x-ray has at another signal while performing
energy equal to this change. The standard AFM imaging. One of
x-rays are then detected by either a the most common techniques for
lithium-drifted silicon detector for mapping differences in materials
an EDS system, or a gas propor- properties is PhaseImagingTM.
tional counter detector for a WDS PhaseImaging is conducted during
system. A typical x-ray spectrum TappingMode AFM operation by
collected with an EDS system is monitoring the phase lag between
shown in Figure 10. the oscillating drive signal used to
Figure 9. Signals emitted from a sample surface
drive the cantilever and the
after interaction with an electron beam.
Backscattered electrons are the oscillating detection signal from
result of beam electrons being the photodiode detector. This
scattered back out of the sample. signal will indicate differences in
In this case, the incident beam viscoelasticity and/or adhesion
electrons undergo a number of across the imaged area. This
scattering events within the technique is commonly applied to
specimen in which very little mapping the distribution of
energy is lost, allowing these polymers in a heterogeneous
electrons to go much deeper into system, or mapping the distribu-
the sample than secondary tion of filler, such as silica or
electrons and still emerge from the carbon black, in a polymer matrix.
sample surface to be detected. An example of PhaseImaging on a
The percentage of beam electrons polyethylene film is shown in
that become backscattered Figure 12. Other ways to get
electrons has been found to be similar information are by Force
dependent on the atomic number Modulation AFM, which maps Figure 10. EDS X-ray spectrum of an AlGaN thin
of the material, which makes it a differences in elasticity across the film on SiC substrate showing the presence of N,
Ga, and Al.
useful signal for analyzing the sample surface, and Lateral Force
material composition. Once these Microscopy (LFM), which maps
electrons escape from the surface differences in friction across the
they are detected by either the sample surface.
Everhart-Thornley detector or a
solid state detector. An example There are also techniques that can
of a backscattered image of a PbSn be used to investigate long range
alloy is shown in Figure 11. forces across the imaged area.
Magnetic Force Microscopy
AFM/SPM (MFM) and Electric Force
Microscopy (EFM) map the
Although an AFM does not magnetic and electrostatic field
provide elemental analysis, it can gradients, respectively, which
supply compositional information extend from the sample surface.
by differentiating materials based These techniques are performed Figure 11. Backscattered SEM image of an PbSn
on physical properties, such as by using either a magnetic or alloy showing contrast based on the atomic number
stiffness, elasticity, compliance, conductive probe to map the of the two components. The brighter areas are Pb-
friction, adhesion, magnetic and attractive and repulsive forces rich. 5,000x, Scale bar=1m.

7
between the tip and the sample. hardness from indentations made
MFM is commonly used to detect at the same forces, producing
the domain structure of magnetic different sized indents. Scratching
bits written on magnetic media, to and wear testing may also be
evaluate the performance of conducted with this configuration
magnetic heads, and to investigate to investigate adhesion and
the magnetic structure of experi- delamination of films under a
mental materials. This is con- small applied force.
ducted by a routine called
LiftModeTM in which a Tapping-
Figure 12. Phase image of two components which are Environment
used to form a polyethelene (PE) film. The phase
Mode topographic image and a
image (right) clearly shows the distribution of the two magnetic image are acquired over One of the primary differences
polymers due to differences in stiffness which is not the same area. LiftMode consists between these two types of
evident from the topographic image (left). 2m scan. of first collecting a line scan in microscopy is the environment in
TappingMode of the surface which they are performed, i.e.,
morphology. The tip is then lifted SEM is conducted in a vacuum
above the surface and a second environment, and AFM is
scan is made over the same line conducted in an ambient or fluid
using the saved topographic scan environment. There are several
to maintain a constant tip-sample issues which make environment
separation. The long-range an important issue. First, there is
magnetic forces shift the reso- a frequent need in fields such as
nance frequency of the oscillating biology and biomaterials to study
cantilever, which is detected to hydrated samples. These two
produce the magnetic image. An techniques compensate for this
example of bits written on a need by different means: an
textured hard disk is shown in environmental chamber for a
Figure 13. Magnetic Force Microscopy (MFM) image
Figure 13. SEM, and a fluid cell for the
of overwritten tracks on a textured hard disk. The
topography (left) was imaged using TappingMode; the AFM. Second, the SEM is
magnetic force image of the same area (right) was Although the AFM is applied so required to work in a vacuum
captured with LiftMode (lift height 35 nm) by mapping that it is nondestructive to the environment due to the nature of
shifts in cantilever resonant frequency. 25m scan. sample surface, it can be used to the technique which brings up the
(17)
study differences in mechanical issues of vacuum compatibility of
properties by performing the sample, the conductivity of
nanoindention to investigate the surface, and vacuum mainte-
hardness differences between nance. To image poorly conduc-
materials. This technique uses a tive surfaces without sample
diamond tip mounted on a stiff, charging may require conductive
stainless steel cantilever. A coatings or staining, which may
TappingMode AFM image is alter or obscure the features of
collected with the probe to interest; or it may require low
determine the area of interest for voltage operation, or an environ-
indentation, the nanoindention is mental chamber, which may
then made at a specified force, and sacrifice resolution.
an image is then collected of the
indented area. An example of For SEM, hydrated samples are
comparing the difference between addressed by placing a specimen
diamond like carbon films on a in an environmental chamber with
hard disk is shown in Figure 14. either an electron transparent
In this example, the two films window or a small aperture for the
demonstrate a difference in beam to enter the chamber. The
8
chamber is typically flushed with accessories, AFM can also be used
an inert gas saturated with water in varied gaseous environments
vapor. Common applications are and at elevated temperature. The
to either investigate hydrated latter is particularly important for
surfaces to preserve their surface research and development of
structure when hydrated, or to polymers.
reduce charging on insulating
samples. An example of imaging
Further Discussion
of a pesticide film on skin can be
seen in Figure 15. For the One thing to keep in mind when Figure 14: Indentations on two different diamond-like
electron beam to interact with the comparing these two techniques is carbon thin films using three different forces (23, 34,
surface in this configuration, it that although SEM and AFM and 45N) with four indents made at each force to
must go though an environment appear very different, they actually compare differences in hardness. 500nm scans.
of gas and water vapor. One share a number of similarities.
drawback of this configuration is Both techniques raster a probe
that it will result in an increase in across the surface to detect some
scattering of the electron beam on interaction with the surface to
the way to and from the surface, form an image. Both have a
which may result in the sacrifice lateral resolution which is similar
of image quality and resolution. in scale (although under certain
conditions AFM is superior). And
One of the primary attractions to both techniques have image
the AFM is its ability to image artifacts that the operator is
insulating surfaces at high resolu- trained to identify. The SEM has
tion in fluid. Imaging samples in had a much longer time to mature
a hydrated state with an AFM is as a technique and to develop an
commonly performed by enclos- understanding of how to identify
ing the sample and probe in a and avoid artifacts, but the rapid
fluid environment, as shown in adoption and implementation of
Figure 16. Since AFM is not AFM has resulted in a similar
based on conductivity, the image understanding of artifacts. This
Figure 15. Environmental SEM image of a pesticide
and scanning mechanism is not article has avoided discussing such film on skin. A hydrated environment was needed in
disturbed by the presence of the artifacts unless they are relevant to order to maintain the integrity of the pesticide layer
fluid. Common applications for the comparison. Furthermore, by and to reduce charging. Bar=1mm.
AFM investigations in fluid are in using two techniques which are
the biological sciences, biomateri- complementary, one technique
als, crystal growth, force interac- will often compensate for the
tion studies, and for investigating imaging artifact of the other
processes in situ (Figures 17, 18). technique.
The resolution of the image will
be determined by the radius of the However, one should be wary of
tip, the applied force, and the combined systems in which an
noise floor of the instrument. AFM is placed inside the SEM
Because of these factors, this chamber. One of the true
configuration allows the study of advantages of the AFM is its
hydrated specimens at a lateral ability to perform high resolution
resolution of 1 to 5nm and a measurements outside of a
vertical resolution down to 0.5 vacuum environment. Placing it
without sample damage, as seen in inside a vacuum environment
the image of the GroES chaperon reduces its flexibility and increases
(Figure 17). With the appropriate its operating time. Combined
9
Laser Diode AFM/SEM systems often have References
Position
Sensitive reduced capabilities and typically
1. Van Ardenne, M. (1938) Z.
Detector compromise the performance of
Phys. 109, 407
Oscillating
Glass Cell both instruments.
Cantilever/Tip
2. Goldstein, J. I., Newbury,
Summary D.E., Echlin, P., Joy, D.C., Fiori,
C., Lifshin, E., Scanning Electron
Flow-thru SEM and AFM are complemen-
Fluid Microscopy and X-ray Mi-
Chamber tary techniques that provide a
O-ring Seal croanalysis, 1981, Plenum
more complete representation of a
Publishing Corp., New York, p. 3.
surface when used together than if
Cells or Molecules
each were the only technique
3. Zworykin, V.K., Hillier, J.,
available. These techniques
Snyder, R.L., (1942), ASTM
Sample or Specimen overlap in their capabilities to
Support
Bulletin 117, p. 15.
provide nanometer scale lateral
information. However, they
4. Binning G, Roher H, Gerber
Figure 16. Fluid cell for an AFM which allows imaging deviate in the fact that the AFM
C, Weibel E, Surface Studies by
in an enclosed, liquid environment. can provide measurements in all
Scanning Tunneling Microscopy,
three dimensions, including
Phys. Rev. Lett. 49 (1982) 57.
height information with a vertical
resolution of <0.5, whereas the
5. Binnig G B, Quate C F, and
SEM has the ability to image very
Gerber Ch., Atomic Force
rough samples due to its large
Microscope, Phys, Rev. Lett., 12
depth of field and large lateral
(1986) 930
field of view.
6. Neves, B.R.A., Salmon, M.E.,
The SEM can provide elemental
Russell, P.E., Troughton, E.B.
analysis using X-ray detection,
Comparitive Study of Field
whereas the AFM can provide
Emission-Scanning Electron
compositional information based
Microscopy and Atomic Force
on physical properties. The fact
Microscopy to Access Self-
that the two techniques operate in
Assembled Monolayer Coverage
different environments can be a
on Any Type of Substrate,
strength when used together since
Microscopy and Microanalysis 5
the AFM does not encounter
(1999) 413.
Figure 17. Image of two GroES molecules positioned
vacuum issues (difficult sample
side-by-side in physiologic fluid, demonstrating 10 preparation, sample modification,
7. Castle, J.E., Zhdan, P.A.,
lateral resolution and 1 vertical resolution. The etc.) and may image samples in an
Characterization of Surface
entire molecule measures 84 across, and a distinct enclosed fluid or other environ-
45 heptameric crown structure protrudes 8 above Topography by SEM and SFM:
ment. The vacuum environment
the remaining GroES surface and surrounds a central Problems and Solutions, J. Phys.
depression. 18nm scan. Image courtesy of Z. Shao,
of the SEM makes it possible to
D: App Phys, 30 (1997) 722.
University of Virginia. (18) conduct a number of techniques
that require vacuum, such as X-ray
8. Lemoine, P., Lamberton, R.W.,
analysis. By having both tech-
Ogwu, A.A., Complementary
niques side-by-side in an analytical
Analysis Techniques for the
facility, the overall scope of
Morphological Study of Ultrathin
analytical capabilities is broad-
Amorphous Carbon Films, J.
ened, adding to the flexibility of
App. Phys., 86 (1999) 6564.
the facility.

10
9. Van Helleputte, H.R.J.R., Meeting of the ECS, vol. 94-1, gallium phosphide/silicon
Haddeman, T.B.J., Verheijen, Pennington, NJ (1994) 461. heterostrucutres, Materials Science
M.J., Baalbergen, J-J., Compara- and Engineering, B22 (1993) 97.
tive Study of 3D Measurement 15. Strausser, Y.E., Schroth, M.,
Techniques (SPM, SEM, TEM) Sweeney, J.J., Characterization of 17. Babcock, K., Magnetic Force
for Submicron Structures, the low-pressure chemical vapor Microscopy: High resolution
Microelectronic Engineering, 27 deposition grown rugged imaging for data storage, Data
(1995) 547. polysilicon surface using atomic Storage, September 1994.
force microscopy, J. Vas. Sci.
10. Nessler, R., Scanning Technol. A 15, 1997, 1007. 18. Mou, J., Czajkowsky, D.M.,
Microscopy Technologies: Scan- Sheng, S., Ho, R., Shao, Z.,
ning Electron Microscopy and 16. Kelliher, J.T., Thornton, J., High resolution surface structure
Scanning Probe Microscopy, Dietz, N., Lucovsky, G., of E. coli GroeS Oligomer by
Scanning 21 (1999) 137. Bachmann, K.J., Low tempera- Atomic Force Microscopy, FEBS
ture chemical beam epitaxy of Letters 381 (1996) 161.
11. Fischer, A., Jentoft, F.C.,
Weinberg, G., Schlogl, R., Niesen,
T.P., Bill, J., Aldinger, F., De
Guire, M.R., Rhle, M., Charac-
terization of Thin Films Contain-
ing Zirconium, Oxygen, and
Sulfer by Scanning Electron and
Atomic Force Microscopy, J.
Mat. Res. 14 (1999) 3725.

12. Campbell, A.L, Bunning,


T.J., Stone, M.O., Church, D.,
Grace, M.S., Surface Ultrastruc-
ture of Pit Organ, Spectacle, and
Non Pit Organ Epidermis of
Infrared Imaging Boid Snakes: A
Scanning Probe and Scanning
Electron Microscopy Study J.
Struct. Biol. 126 (1999) 105.

13. Reynolds, G.W., Taylor, J.W.,


Correlation of Atomic Force
Microscopy Sidewall Roughness
Measurements with Scanning
Electron Microscopy Line-Edge
Roughness Measurements on
Chemically Amplified Resists
Exposed by X-ray Lithography, J.
Vac. Sci. Technol. B 17 (1999)
2723.
Figure 18. Living human vascular endothelial cells imaged in culture media by
TappingMode AFM. These images, collected at 30 minute intervals, reveal the
14. Strausser, Y.E., Doris, B., movement of living cells which were incubated with 200ng/ml vascular endothelia
Diebold, A.C., Huff, H.R., growth factor (VEGF). With TappingMode, the nucleus as well as other
Measurement of Silicon Surface submembraneous structures are visible. These cells appear flatter and more
Microroughness with AFM, elongated compared to control (untreated) cells. 65m scans. Sample courtesy of
Georges Primbs, Miravant Inc.
Extended Abstracts of the 185th
11
To Contact the Author:

Phil Russell or Dale Batchelor


Analytical Instrumentation Facility
College of Engineering
North Carolina University
Box 7531
Raleigh, NC 27695-7531
Email: prussell@ncsu.edu,
dale_batchelor@ncsu.edu

John Thornton
Digital Instruments/Veeco Metrology
Email: john@di.com

112 Robin Hill Road


Santa Barbara, California 93117
T: (800) 873-9750
T: (805) 967-1400
F: (805) 967-7717
Email: info@di.com
www.veeco.com
TappingMode, LiftMode, and PhaseImaging are
registered trademarks of Veeco Instruments, Inc. Distributors World Wide AN46 2/01

Potrebbero piacerti anche