Sei sulla pagina 1di 29

RPI Field Service Smart Kit

(RPI Part #SCK028)


A Guide to Maintaining
The SciCan STATIM Cassette Autoclave
Replacement Parts Industries, Inc. is pleased to present this valuable work tool that can help
save you and your customers time and money. Included in this booklet you will find Instructions
on how to use each of the tools in this Kit, plus diagrams, exploded views, PM check lists,
error code listings with troubleshooting service tips and a complete list of RPI parts to fit the
SciCan STATIM Cassette Autoclaves.

For use with these models only:


STATIM 2000, model 101102 (with S/N ending in "A")
STATIM 2000, model 121101
STATIM 5000, model 01-201103

CALL (800) 221-9723 or (818) 882-8611 FAX (818) 882-7028


E-MAIL order@rpiparts.com WEBSITE www.rpiparts.com
TABLE OF CONTENTS
Page
1.0 - THEORY OF OPERATION
1.0 - Theory of Operation ..................................................................................................................................................1 The STATIM uses a steam generator (also known as a steam generator/boiler and converted into steam. The
STATIM 2000 Cycle Time, Temperature & Model Listing........................................................................................1 boiler) to create steam, which is then injected into the cas- steam is injected into the cassette. As the steam enters the
STATIM 5000 Cycle Time, Temperature & Model Listing........................................................................................1 sette, which serves the same purpose as a traditional au- cassette, the air is purged out through the vent solenoid
Figure 1 - 2000 & 5000 Keypad Symbols Chart....................................................................................................1 toclave chamber. The temperature is then monitored via valve. The air is continuously expelled into the waste water
2.0 - RPI Field Service Smart Kit (RPI Part #SCK028) .........................................................................................................2 thermocouples. One thermocouple monitors the steam bottle's condensation coil. This allows steam to run through
2.1 - Extension Test Cable (RPI Part #SCT026).........................................................................................................4 generator/boiler temperature and the other monitors the the cassette to maintain proper steam temperature.
2.2 - Start Switch (RPI Part #SCS021) .....................................................................................................................5 cassette temperature, not unlike the traditional steam sen-
2.3 - Pump Tester Bottle (RPI Part #SCK024) ...........................................................................................................6 sor mounted in the chamber. The autoclave does not mon- Most autoclaves reuse the same water over and over, but
itor the pressure with a device such as a transducer or the STATIM uses fresh distilled water each time. Never use
2.4 - Spanner Nut (RPI Part #RPT364) .....................................................................................................................8
gauge. Instead, the pressure reading is a value calculated deionized, demineralized or specialty filtered water. Tap
2.5 - Hex Balldriver (5/64") (RPI Part #RPT297) .......................................................................................................9
by the microprocessor. water must never be used.
2.6 - Cassette Seal Removal Tool (RPI Part #RPT372) ............................................................................................10
2.7 - Liquid Soap (RPI Part #RPS287) ....................................................................................................................10
Instruments are loaded into the cassette and the cassette Once the sterilization stage has finished, the vent solenoid
2.8 - Threadlocker 545 (RPI Part #RPA369) ...........................................................................................................11 valve will open and remain open. The venting stage will
is placed into the cassette bay. Inserting the cassette will
2.9 - Hex Balldriver (9/64") (RPI Part #RPT836) .....................................................................................................11 close the microswitch that activates the vent solenoid expel the steam pressure through the exhaust tubing into
2.10 - Alignment Tool (RPI Part #SCT052)..............................................................................................................12 valve. The user selects a cycle and depresses the start but- the condenser waste bottle. The steam condenses back
2.11 - Water Quality Tester (RPI Part #RPT820)......................................................................................................14 ton. The steam generator/boiler then heats up to the proper into water in the condensation coil. The compressor will
2.12 - Calibration Solution (RPI Part #RPS821) ......................................................................................................15 temperature. The distilled water is then pumped into the run until the drying stage is complete.
3.0 - Exploded View ........................................................................................................................................................16
4.0 - Planned Maintenance (PM) Check List and Installation and Monthly Maintenance for Cassette Seal STATIM 2000 CYCLE TIME, TEMPERATURE & MODEL LISTING
STATIM 2000 (RPI Part #SCK035) .......................................................................................................................18 Unwrapped Wrapped Rubber & Plastics Air Dry
STATIM 5000 (RPI Part #SCK036) .......................................................................................................................20 Model # Temp Time Temp Time Temp Time Time
Installation and Monthly Maintenance for Cassette Seal .....................................................................................22
101102 135C 3.5 135C 10 121C 15 60
5.0 - Service Tips ............................................................................................................................................................24
121101 275F minutes 275F minutes 250F minutes minutes
6.0 - LCD Error Codes .....................................................................................................................................................26
01-122328 135C 4.0 135C 10 121C 20 60
Stage #1 - Warming Up......................................................................................................................................27
275F minutes 275F minutes 250F minutes minutes
Stage #2 - Conditioning .....................................................................................................................................27
Stage #3 - Pressurizing......................................................................................................................................28 STATIM 5000 CYCLE TIME, TEMPERATURE & MODEL LISTING
Stage #4 - Sterilization.......................................................................................................................................29 Rubber Heavy Duty
Stage #5 - Venting .............................................................................................................................................31 Unwrapped Wrapped & Plastics Unwrapped
Stage #6 - Drying...............................................................................................................................................33 Country Model # Temp Time Temp Time Temp Time Temp Time
7.0 - LCD Messages........................................................................................................................................................34
USA 01-201103(-R) 132C 3.5 132C 6 121C 35 132C 6
8.0 - Troubleshooting Tips ...............................................................................................................................................36 01-201104(-R) 270F minutes 270F minutes 250F minutes 270F minutes
9.0 - Reference Blocks....................................................................................................................................................38
Canada All other 135C 3.5 135C 6 121C 15 135C 60
9.1 - Steam Generator/Boiler.................................................................................................................................38 (And all other countries
models 275F minutes 275F minutes 250F minutes 275F minutes
except Czech Republic)
9.2 - Cleaning the Steam Generator/Boiler ............................................................................................................39
9.3 - Steam Generator/ Boiler Calibration Procedures............................................................................................40 FIGURE 1 STATIM 2000 & 5000 KEYPAD SYMBOLS CHART
9.4 - Solenoid Valve ..............................................................................................................................................41
9.5 - Cassette .......................................................................................................................................................42
9.6 - Pump............................................................................................................................................................43
9.7 - Air Compressor .............................................................................................................................................44
Rubber
9.8 - Electrical Diagram ........................................................................................................................................46 Model 2000 Unwrapped Wrapped & Plastics Air Dry Start Stop
9.9 - Plumbing Diagram ........................................................................................................................................48
Rubber Heavy Duty
10.0 - RPI Parts List (for Parts to Fit STATIM 2000 & 5000) ..............................................................................................50 Model 5000 Unwrapped Wrapped & Plastics Unwrapped Start Stop
11.0 - RPI Tools & Accessories List (for Parts to Fit STATIM 2000 & 5000)........................................................................52

1
2.0 - RPI FIELD SERVICE SMART KIT
(RPI PART #SCK028)
Contained in the next several pages are
instructions on how, when, where and why
to use the parts in the RPI Field Service
Smart Kit (RPI Part #SCK028) to service
and repair the SciCan STATIM Cassette
Autoclave models 2000 and 5000.
P 2O T
U
R
H S

E
TE

2.1 Extension Test Cable


RPI Part #SCT026
TI
i ve
LO C
Ad5 45
h es
IS
THSIDE

(see page 4)
UP

IS
TH E

MINIMLINE
SID

FILL
UP

UM

L
VE
LE

OP
ST
T
AR
ST

All Parts Shown Are Included In The Kit!


ive
s

e 5 I
h

4 T
Ad
LO5

C K IN G
LO N N
RU

29
0

RPA369 SCT052

N K
G

U C
IN

R LO
N
I

SCT026 SCS021 O

2.2 Start Switch TH


IS
UP

RPI Part #SCS021


RPT297 (see page 5)
RPT820 RPS821
MINIMUM
FILL LINE

MINIMU
M
FILL LINE

2.3 Pump Tester Bottle


RPT836 SCK024 RPI Part #SCK024
(see pages 6 & 7)

RPT372 RPT364 RPS287 RPC476

2 3
2.1 - EXTENSION TEST CABLE 2.2 - START SWITCH STEP #2 - Once the Start Switch is connected, the next
FIGURE 2.1A step is to connect the Pump Tester Bottle (see 2.3 Pump
(RPI PART #SCT026) LCD Display Board (RPI PART #SCS021) Tester Bottle Instructions, page 6). When testing the
Connector
When testing, troubleshooting or repairing the STATIM au- When testing the pump or decontaminating the steam pump, press and hold the momentary button for
toclave, install the RPI Extension Test Cable. By attaching generator/boiler, it is suggested by the OEM to activate the approximately 25 seconds. When decontaminating the
the Extension Test Cable to the cover and main PC board pump by shorting Pins #18 and #16 together. However, steam generator/boiler, (see Caution note below), press
Keypad Connector the RPI Start Switch offers an easier and safer method. By and slide the momentary button into the hold (i.e. run-
as instructed below, you will be able to extend the dis-
tance between the autoclave and the cover thus allowing connecting the RPI Start Switch to the main PC board con- ning lock) position for approximately 3 minutes. (Caution:
you to place the cover in a horizontal (not vertical) position IS
THSIDEP
nector (P1), you control the activation of the pump via a When decontaminating the steam generator/boiler, the
U

for easier viewing of the LCD display and easier access to hand-held switch. In addition, the RPI Start Switch has a reservoir must be full so that the pump does not run dry
the keypad. hold feature (i.e. running lock) for longer test periods. during the process thus causing severe damage to the au-
IS
TH E
SIDP
U toclave.) See page 39 - Figure 9.2 for complete steam
INSTRUCTIONS INSTRUCTIONS generator/boiler cleaning procedure.
STEP #1 - Disconnect the power to the autoclave. Remove Label STEP #1 - Connect the RPI Start Switch connector to Pins
the cover. Disconnect the two cables from the main PC THIS SIDE UP #18 and #16 on the 20 Pin Connector (P1) located on the
board - one cable from the LCD display board and the main PC board. (See Figure 2.2 and note the following: To
other from the keypad. Leave the cover in the horizontal guide the Start Switch connector to Pins #18 and #16,
position (see Figure 2.1A). start at the pin labeled 20, count down one pin, which
STEP #2 - Connect the Extension Test Cable mating con- FIGURE 2.1B is Pin #18, then count down one more pin, which is Pin
nector to the LCD display board cable. LCD Display #16. Gently guide the Start Switch connector to both Pins
STEP #3 - If the OEM keypad is installed, connect the Ex- Board Connector #18 and #16, leaving all other pins exposed.) (Caution: If
tension Test Cable mating connector to the keypad cable the Start Switch is connected to any other pins except #18
connector (see Figure 2.1B). and #16, severe damage may occur to the autoclave.) See
If the RPI keypad is installed, connect the Extension Test IS
Device Activation Chart, page 47.
THSIDEP
Cable mating connector to the RPI keypad cable connec- Extension U

tor - making sure that both connectors have the This Side Test Cable
Up label facing in the upward direction (see Figure 2.1C). FIGURE 2.2
(Note: The RPI keypad cable connector does not have an
TIP - To guide the Start Switch
alignment key, so it is very important to connect as per in- OEM Keypad Connector to Pins #18 and
structions and Figure 2.1C). Cable Connector IS
THSIDEP

(with alignment key)


U
#16, start at the pin labeled
STEP #4 - Connect the Extension Test Cable mating con- 20 20, count down one pin,
nectors to the main PC board - making sure that the
Press and hold for which is Pin #18, then count
Extension Test Cable connector has the This Side Up Momentary feature down one more pin, which is
label facing in the upward direction (see Figure 2.1A). FIGURE 2.1C Pin #16. Gently guide the Start
(Caution: If the Extension Test Cable is connected incor-
LCD Display Start Switch Switch Connector to both Pin
rectly, severe damage may occur to the autoclave.) (RPI Part #SCS021)
Board Connector #18 and #16, leaving all other
pins exposed.

IS
THSIDEP
Extension
ck)
U

N K
G

U C
IN
the ng Lo

R LO
Test Cable

N
r
fo nni
s lide (Ru
e
and tur
RPI Keypad IS ress fea
Cable Connector THSIDEP P old
(without alignment key) U H

4 5
2.3 - PUMP TESTER BOTTLE 7/16 wrench. Connect the metering tube to the Pump STEP #4 - FILTERS: removing filtered fittings). Clean the mesh filter of both
(RPI PART #SCK024) Tester Bottle male connector and tighten. If the autoclave has been upgraded with Drain Kit the inlet and the outlet fittings by placing the fittings
STEP #2 - PRIME PUMP: Fill reservoir with distilled water with Pump Filter (RPI Part #SCK037), be sure to re- under running water, then blow both ends out with an air
The RPI bottle has a swivel fitting on the lid that allows
only and activate the pump for 2-3 seconds to purge any move the mesh filter from both the inlet and outlet fit- hose. Reinstall the filtered fittings using Teflon tape
you to control the position and balance of the bottle. In ad-
trapped air from the tubing (until water passes through tings (if they have not already been removed). The ULKA (see Figure 2.3B). Be sure to prime the pump in order
dition, RPI includes 8 of tubing already connected to the
the pump) (see page 5, 2.2 - Start Switch). Empty the brand pump does not have the mesh filters. Replace the to purge trapped air before running a complete cycle. If
swivel fitting which allows you to move the bottle a safe
bottle of any water that may have entered it during the purge Water Filter (In-Line) (RPI Part #SCF034). Repeat Step the water level still does not reach the MINIMUM FILL
distance away from the autoclave during testing.
cycle. The bottle must be completely empty before proceed- #2 - Prime Pump, then if the water level still does not LINE within 25 seconds, replace the pump (see Tip in
The OEM now recommends the use of the Reservoir Cap & ing with the test. reach the MINIMUM FILL LINE within 25 seconds, re- Figure 2.3B). For additional information regarding the
Filter (RPI Part #SCK059) along with the In-Line Water Filter STEP #3 - TEST CYCLE: Before activating the pump for place the pump. pump see page 43 - Figure 9.6.
(RPI Part #SCF034) on all units. The mesh filters should be re- the Test Cycle, have a stopwatch or another timing device If the unit does not have the Drain Kit with Pump
moved and the autoclave should be upgraded to the In-Line available for use during the Test Cycle. Activate the pump Filter (RPI Part #SCK037) installed, then clamp off the RELATED RPI PARTS:
Water Filter (Drain Kit with Pump Filter RPI Part #SCK037). and note the amount of time that it takes for the water pump inlet tubing from the reservoir using hemostats or Drain Kit with Pump Filter (RPI Part #SCK037)
level in the bottle to reach the MINIMUM FILL LINE. a tubing clamp (Note: It is important to clamp the tubing Metering Tube (RPI Part #SCT053)
INSTRUCTIONS If the water level reaches the MINIMUM FILL LINE within so that the reservoir does not drain). Disconnect the me- Reservoir Cap & Filter (RPI Part #SCK059)
STEP #1 - TO BEGIN: Disconnect the pump outlet meter- 25 seconds, pump testing is complete and no further tering tube from the pump. Remove the filtered elbow Teflon Tape (RPI Part #RPT579)
ing tube (see Figure 2.3A) from the steam generator/ steps are required. outlet fitting and the filtered barb inlet fitting from the Water Filter (In-Line) (RPI Part #SCF034)
boiler by using a 3/8 wrench on the compression nut If the water level does not reach the MINIMUM FILL LINE pump. (Caution: Use a wrench on the pump body while
while supporting the steam generator/boiler fitting with a within the specified time, proceed to the next step.

FIGURE 2.3A FIGURE 2.3B


Metering Tube (RPI Part #SCT053)
to Steam Generator/Boiler
290

Outlet Fitting Mesh Filter


Flow

Flow
290
Reservoir Cap & Filter
Metering Tube (RPI Part #SCK059)
(RPI Part #SCT053)
O
I Reservoir Inlet Fitting
Drain Kit with Pump Filter
Bottle Male Connector (RPI Part #SCK037);
Filter only (RPI Part #SCF034) TIP - If the unit does not have the
Drain Kit with Pump Filter (RPI
Bottle Inlet Tube (8)
Part #SCK037) and the Reservoir
TIP - The OEM now recommends the use Cap & Filter (RPI Part #SCK059),

Flow
of the Reservoir Cap & Filter (RPI Part now is a good time to upgrade. If
#SCK059) along with the In-Line Water time does not permit an upgrade,
MINIMUM Filter on all units. The mesh filters should remove and clean mesh filters
Pump Tester Bottle
FILL LINE

(RPI Part #SCK024) be removed and the autoclave should be inside inlet and outlet fittings as
upgraded to the In-Line Water Filter (Drain described above and anticipate
Kit with Pump Filter RPI Part #SCK037). an upgrade at the next service.

6 7
2.4 - SPANNER NUT 2.5 - HEX BALLDRIVER (5/64") and (2) #2 Split Lockwashers (RPI Part #RPH253). The
(RPI PART #RPT364) FIGURE 2.4 (RPI PART #RPT297) screws and lockwashers are also available separately and
can be used with the OEM switch.
When removing the solenoid flute/bonnet, it is suggested TIP - First remove the nut, yoke and - FOR USE WHEN REPLACING
coil to access the flute/bonnet.The
by the OEM to use a pair of pliers on the welded area of the
Spanner Nut (RPI Part #RPT364) offers THE MICROSWITCH Apply Silicone Gasket Maker (RPI Part #RPS639) to side
flute/bonnet. If done incorrectly, the pliers could collapse of switch and mount to probe bracket angle. Apply a seal-
the hollow area of the flute/bonnet, thus preventing the an easier and safer way to remove The OEM microswitch is held in place with two small slot- ing bead of gasket maker around switch housing to close
plunger from moving freely. and install the flute/bonnet. ted pan head screws which are difficult to access. The off any leak paths between the switch and bracket. Watch
compressor must be removed in order to remove these for steam leaks from the cassette bay as they can cause
However, using the Spanner Nut (RPI Part #RPT364) offers screws. The RPI microswitch comes with two socket head damage to the microswitch and other internal compo-
an easier and safer way to remove and install the cap screws which are easy to install or remove when nents. Verify integrity of cassette seal steam ports and
flute/bonnet. Nut using the Hex Balldriver. The Hex Balldriver is designed to probe bracket alignment (see page 12, 2.10 - Alignment
reach far into the autoclave to easily remove the screws, Tool (RPI Part #SCT052).
Coil even at an angle, and without having to remove the com-
INSTRUCTIONS (RPI Part #SCC005)
pressor (see Figure 2.5). Caution: Disconnect the power RELATED RPI PARTS:
STEP #1 - Disconnect the power to the autoclave. Remove Coil Bracket to the autoclave before servicing and be sure the steam Microswitch Kit (RPI Part #SCK007)
the nut, coil bracket, and coil. generator/boiler has completely cooled. Silicone Gasket Maker (RPI Part #RPS639)
STEP #2 - Place the Spanner Nut over the flute/bonnet
Socket Head Cap Screw (RPI Part #ADS120)
and align the Spanner Nut pins with the holes in the base NOTE - Microswitch (RPI Part #SCK007) includes (2) #2- Lockwasher (RPI Part #RPH253)
of the flute/bonnet (see Figure 2.4). 56 x 1/2" lg. Socket Head Cap Screws (RPI Part #ADS120)
STEP #3 - Using a wrench on the Spanner Nut; loosen the Spanner Nut
flute/bonnet to access the internal parts of the solenoid (RPI Part #RPT364)
valve. FIGURE 2.5
STEP #4 - Service as needed. Flute/Bonnet
TIP - Apply Silicone Gasket Maker (RPI
RELATED RPI PARTS: Probe Bracket Angle Part #RPS639) to side of switch and
Double Ended Spanner Wrench (RPI Part #RPT501) mount to probe bracket angle. Apply a
Male Connector (RPI Part #RPF227) sealing bead of Gasket Maker around
Push-In Elbow Fitting (RPI Part #RPF363) switch housing to close off any leak
Hex Balldriver (5/64") paths between switch and bracket.
Solenoid Valve Repair Kit (RPI Part #SCK003) Solenoid - Plunger Kit (RPI Part #RPT297)
Solenoid Coil (RPI Part #SCC005) (RPI Part #SCK003)
Teflon Tape (RPI Part #RPT579) Male Connector
Valve Body
For additional information and tips see page 41 - Figure (RPI Part #RPF227)
9.4.
Port #2 Socket Head Cap Screw
(RPI Part #ADS120)
Lockwasher
Port #1 (RPI Part #RPH253)

Microswitch Kit
(RPI Part #SCK007)

Push-in Elbow Fitting Probe Bracket Kit


(RPI Part #RPF363) 2000 (RPI Part #SCK038); 5000 (RPI Part #SCK042)
Probe Bracket
2000 (RPI Part #SCB039); 5000 (RPI Part #SCB043)

8 9
2.6 - CASSETTE SEAL REMOVAL 2.7 - LIQUID SOAP (4/PKG) 2.8 - THREADLOCKER 545 2.9 - HEX BALLDRIVER (9/64")
TOOL (RPI PART #RPT372) (RPI PART #RPS287) (RPI PART #RPA369) (RPI PART #RPT836)
The Cassette Seal Removal Tool was designed to help The Liquid Soap comes in a convenient single-use foil Threadlocker 545 comes in a tube with a slender neck al- The Hex Balldriver is designed to reach far into the auto-
remove the cassette seal for either maintenance or pouch thats easy to store and use. It is recommended to lowing it to reach difficult areas, and the tube has a reuse- clave to easily access the probe bracket mounting screws
replacement without scratching the cassette lid or dam- use the soap on the cassette seal for installation and reg- able lid. It is for use on all pneumatic and hydraulic fittings for alignment or replacement without removing other com-
aging the seal. ularly scheduled maintenance every 30 days. requiring a high temperature sealant and threadlocker. ponents. This job, without this long reach Balldriver, may
have required removal of the pump, steam generator/
INSTRUCTIONS INSTRUCTIONS INSTRUCTIONS boiler, and/or solenoid valve. This Balldriver will fit the OEM
Slip the tip of the tool behind the seal at one of the corners Open one package of Liquid Soap and lightly lubricate the Clean threads, shake well and use 2-3 drops on the or RPI probe bracket mounting hardware. See Figure 2.9.
of the cassette and pry the seal loose (see Figure 2.6). cassette seal using entire contents (see Figure 2.7). Dis- threaded portion of a fitting to lock and seal. (Note: Caution: Disconnect the power to the autoclave before
Once the first corner is free, take hold of the cassette seal pose of packet after use. The SDS is available on the RPI Accommodates operating pressures up to 10,000 PSI and servicing and be sure the steam generator/boiler has
and pull down and out in order to remove it from the cas- website (www.rpiparts.com). temperatures ranging from -65F to 300F.) The SDS is completely cooled.
sette groove. The cassette seal and groove can now be in- available on the RPI website (www.rpiparts.com).
spected and cleaned. For additional information and tips, RELATED RPI PARTS: RELATED RPI PARTS:
see Installation and Monthly Maintenance Guide Cassette Seal 2000 (RPI Part #SCS001) USE WITH RPI PARTS: Probe Bracket Kit 2000 (RPI Part #SCK038)
that is supplied with RPI's Cassette Seals (RPI Part #'s Cassette Seal 5000 (RPI Part #SCS029) Check Valve (RPI Part #SCK011) Probe Bracket Kit 5000 (RPI Part #SCK042)
SCS001 & SCS029) or see page 22. Sci-Dry (2 oz. Spray Bottle) (RPI Part #SCA054) Male Connector (RPI Part #RPF227) Probe Bracket 2000 (RPI Part #SCB039)
Push-in Elbow Fitting (RPI Part #RPF363) Probe Bracket 5000 (RPI Part #SCB043)
For additional information and tips, see Installation and Push-in Elbow Fitting (RPI Part #RPF377) Probe Bracket Gasket (RPI Part #SCG010)
FIGURE 2.6
Monthly Maintenance Guide that is supplied with RPI's Safety Valve (38 PSI) (RPI Part #SCV027)
TIP - Slip the tip of the Cassette Cassette Seals (RPI Part #'s SCS001 & SCS029) or see Safety Valve (70 PSI) (RPI Part #SCV004) FIGURE 2.9
Removal Tool behind the seal at page 22.
one of the corners of the cassette TIP - If realigning existing probe bracket,
and pry the seal loose with ease. loosen (DO NOT REMOVE) 4 mounting screws
FIGURE 2.7 FIGURE 2.8 and compression fitting connections to allow
repositioning of probe bracket steam ports.
TIP - lightly lubricate the cassette seal TIP - Use 2-3 drops of Thread-
using liquid soap at time of installation locker 545 on the threaded portion
Cassette Seal Removal Tool as well as during regularly scheduled of a fitting to lock and seal.
(RPI Part #RPT372) maintenance every 30 days.
Microswitch Kit
(RPI Part #SCK007)
ive Installed
I es 5

h
CT

Ad
4
Thermocouple

LO5
Liquid Soap
Cassette Seal (RPI Part #RPS287) (RPI Part #SCT030)
STATIM 2000: (RPI Part #SCS001) Probe
STATIM 5000: (RPI Part #SCS029) Bracket Gasket
(RPI Part #SCG010)

Threadlocker 545
(RPI Part #RPA369) Probe Bracket Kit
2000 (RPI Part #SCK038)
5000 (RPI Part #SCK042)
Reuseable Lid Hex Balldriver (9/64") Probe Bracket
(RPI Part #RPT836) 2000 (RPI Part #SCB039)
5000 (RPI Part #SCB043)

10 11
2.10 ALIGNMENT TOOL latched and closed. See Figure 2.10A. STEP #6 - Lightly snug the probe bracket mounting screws. STEP #9 - Coat entire inside of cassette with Sci-Dry
(Caution - Ensure that the Alignment Tool is properly in- Move the cassette in and out several times to ensure proper (RPI Part #SCA054).
(RPI PART #SCT052) stalled before closing the cassette lid or damage can occur.) alignment. Leave the cassette in the engaged position. Now
When removing or replacing the probe bracket assembly, STEP #2 - Carefully attach the bottom tray to the cassette tighten mounting screws to retain bracket position. The RELATED RPI PARTS:
the steam ports must be properly aligned. Use this Align- lid and close to properly retain the Alignment Tool. The cas- probe bracket assembly is now aligned and ready for use. Probe Bracket Kit 2000 (RPI Part #SCK038)
ment Tool to easily and accurately set the position of the sette lid should close normally and retain the Alignment STEP #7 - Remove Alignment Tool. Reinstall or replace Cas- Probe Bracket Kit 5000 (RPI Part #SCK042)
steam ports to the cassette. Tool. DO NOT FORCE IT CLOSED! sette Seal (RPI Part #SCS001 fits STATIM 2000 & RPI Part Probe Bracket 2000 (RPI Part #SCB039)
STEP #3 - The cassette is now ready to be used to align #SCS029 fits STATIM 5000) using Liquid Soap (RPI Part Probe Bracket 5000 (RPI Part #SCB043)
INSTRUCTIONS probe bracket steam ports. #RPS287) included with the probe brackets. See page 10 - Probe Bracket Gasket (RPI Part #SCG010)
Separate the cassette tray from the cassette lid. Place STEP #4 - If realigning an existing probe bracket, loosen Figure 2.7. For additional information regarding the care and Cassette Seal 2000 (RPI Part #SCS001)
the lid on counter top and remove the cassette seal. The (DO NOT REMOVE) 4 mounting screws (using 9/64" Hex maintenance of the cassette seal, see Installation and Cassette Seal 5000 (RPI Part #SCS029)
cassette seal must be removed before continuing with Balldriver - RPI Part #RPT836) and compression fitting con- Monthly Maintenence Guide that is supplied with RPI's Cas- Microswitch Kit (RPI Part #SCK007)
Alignment Tool installation (see page 10 - Figure 2.6). nections to allow repositioning of probe bracket steam ports. sette Seals (RPI Part #'s SCS001 & SCS029) or see page 22. Thermocouple (RPI Part #SCT030)
STEP #5 - Slowly insert the cassette into the cassette bay STEP #8 - After successful installation or realignment of Liquid Soap (RPI Part #RPS287)
STEP #1 - Tilt the cassette lid and install Alignment Tool (armature). As the cassette approaches the probe bracket probe bracket, install or realign thermocouple. See RPI Part Hex Balldriver (9/64") (RPI Part #RPT836)
into steam ports from the inside of the cassette lid as (still loosely installed), position bracket assembly up and #SCT030 Installation Instructions that are available on the
shown. The square tabs on the tool must engage the down and sideways until probe bracket steam ports and RPI website (www.rpiparts.com).
square notches in the cassette lid, and the tool must lay cassette engage easily. Once properly inserted, the spring
flat within the cassette seal groove. The Alignment Tool bar should engage and lock the cassette into position. A FIGURE 2.10B
must be seated correctly to avoid damaging the tray when "click" will be heard and felt. See Figure 2.10B.
TIP - If realigning existing
probe bracket, loosen (DO NOT
FIGURE 2.10A REMOVE) 4 mounting screws
TIP - Angle cassette lid as shown, using and compression fitting con-
gravity to aid in tool positioning while nections to allow repositioning
Alignment Tool of probe bracket steam ports.
(RPI Part #SCT052) assembling and closing the cassette.

Microswitch Kit
(RPI Part #SCK007)
Installed

Cassette Lid
square notches
Thermocouple Internal View of
Exhaust Steam Port (RPI Part #SCT030) Cassette Bay
(Armature)
Probe Bracket Gasket Spring Bar
(RPI Part #SCG010)
Injection Steam Port Hex Balldriver (9/64")
(RPI Part #RPT836)
Probe Bracket Kit
2000 (RPI Part #SCK038); 5000 (RPI Part #SCK042)
Grounding Wire Probe Bracket
2000 (RPI Part #SCB039); 5000 (RPI Part #SCB043)

12 13
2.11 WATER QUALITY TESTER INSTRUCTIONS 2.12 CALIBRATION SOLUTION The Calibration Solution is certified traceable to NIST Stan-
dard Reference Material Potassium Chloride (7447-40-7)
(RPI PART #RPT820) STEP #1 - Remove cap. (RPI PART #RPS821) and will provide a reading on the Tester of 800 ppm when
STEP #2 - Turn Water Quality Tester on (slide switch on
The Water Quality Tester (RPI Part #RPT820) is a meter de- top of meter to the ON position). See Figure 2.11. CALIBRATION Check calibration of the Water Quality tested at room temperature (77F). Consult temperature
signed to measure water quality. This meter determines STEP #3 - Rinse sensing probes (see Figure 2.11) in dis- Tester (RPI Part #RPT820) approximately every 4 months chart on Calibration Solution pouch for ppm values vs.
the total amount of dissolved solids (TDS) present in the tilled water by submersing and stirring gently. using Calibration Solution (RPI Part #RPS821). Check cal- temperature condition. SDS sheet available on the RPI
water being used for sterilization. SciCan states the total STEP #4 - Immerse Tester in a water sample without ex- ibration by immersing the Tester in a container of Calibra- website (www.rpiparts.com).
dissolved solids (TDS) allowable in the water reservoir to ceeding the max level line and without bottoming out the tion Solution (the solution is supplied in a single use pouch
be a maximum of 5 Parts Per Million (ppm) or a conduc- meter. Stir gently and wait for the reading to stabilize. and can serve as the container - see Figure 2.12) without If calibration reading is outside of the solutions rated ppm
tivity of less than 10 Micro Siemens (s/cm). RPIs Water STEP #5 - The reading should be 5 ppm. exceeding the max level line and without bottoming out value, adjust Calibration Trimmer by inserting screwdriver
Quality Tester displays its readings in Parts Per Million the meter. Stir gently and wait for the reading to stabilize. (supplied with Tester) through hole in the side of Tester and
(ppm) and will automatically compensate for temperature BATTERIES - Expected battery life should be approxi- adjust potentiometer until the value matches the specified
variations. mately 150 hours. Batteries should be replaced when the ppm value from the chart on the pouch (see Figure 2.12).
display will not turn on or fades. The Tester uses (4) 1.5V
coin cells (Type LR44 or 357).

FIGURE 2.11 FIGURE 2.12


Water Quality Tester (RPI Part #RPT820) TIP - Check calibration by TIP - If calibration reading is outside of
immersing the Tester in a the solutions rated ppm value, adjust
FRONT BACK container of Calibration Solu- Calibration Trimmer by inserting screw-
tion (the solution is supplied driver (supplied with Tester) through hole
ON/OFF Battery in a single use pouch and in the side of Tester and adjust poten-
Switch Compartment can serve as the container). tiometer until the value matches the
specified ppm value from the chart on
Water Quality Tester the pouch.
(RPI Part #RPT820)
Liquid Pocket
Crystal Clip
Display Calibration
Trimmer Calibration Solution
(RPI Part #RPS821)

TD
Water Quality Tester 1
S
Calibration
Maximum (RPI Part #RPT820) Trimmer
Immersion
MAX
Level LEVEL Screwdriver
Minimum Temperature Chart
Immersion
Level

Calibration Solution
Sensing (RPI Part #RPS821)
Probes

14 15
STATIM STATIM
3.0 EXPLODED VIEW Cassette Seal Item Description 2000 5000

L
VE
LE
STATIM 2000 Shown (Internal) 1 35

P
O
ST
2 11 Compressor Filter SCF002

T
AR
ST
12 Air Compressor Kit SCK020
13 Safety Valve (70 PSI) SCV004 SCV004
14 Metering Tube (290) SCT053 SCT053
Cassette Bay
(Armature) 15 Double Thermal Fuse Assembly SCF006 SCF006
34 16 Screw (#6-32 X 1/4) ADS054 ADS054
17 Biological Filter SCF0324 SCF0324
4
18 Push-In Elbow Fitting RPF363 RPF363
3
19 Leveler Leg RPF362 RPF362
5 20 Solenoid - Coil SCC005 SCC005
6 21 Solenoid - Plunger Kit SCK003 SCK003
22 Screw (#8-32 X 1/2 - SEMS) RPH817 RPH817
7
23 Power Switch SCS023 SCS023
24
24 Power Cord Replacement SCK0258
8
24 Power Cord RPC291 RPC291
25
25 Condenser Waste Bottle Kit SCK016 SCK016
Drain 33
Tube 25 Condenser Waste Bottle with Lid SCB0182 SCB0182
26 Rubber Foot CAF011 CAF011

FLOW
36 27 Pump Kit (INVENSYS) SCP050 SCP050
9 27
27 Pump Kit (ULKA) SCP051 SCP051

REMOVE THIS LABEL


SERIES STELIZER
FOR USE WITH 5000
LEAVE THIS LABEL ON
SERIES STERILIZER
FOR USE WITH 2000
37 28 Fuse (15A, 250V) - Time Delay RPF0711
10 13 32 28 Fuse (2A, 250V) RPF367 RPF367
14 29 28 Fuse (1/4A, 250V) - Time Delay RPF368 RPF368
11 22 28 Fuse (1/2A, 250V) - Time Delay RPF060
20 31 30 28 Fuse (1A, 250V) - Time Delay RPF0611
21 29 Drain Kit with Pump Filter SCK037
29 Cable Holder RPC702 RPC702
12 29 Female Quick Connect Fitting (White) RPF428 RPF428
29
30 Drain Tube Assembly Kit RPK432 RPK432
15 30
30 Male Quick Connect Fitting (White) RPF429 RPF429
16 31 Male Connector RPF227 RPF227
STATIM STATIM 32 Water Filter (In-Line) SCF034 SCF034
Item Description 2000 5000 33 Reservoir Cap & Filter SCK059 SCK059
17 1 Cassette Seal SCS001 SCS029 33 Reservoir Cap SCC060 SCC060
18 22 2 Bubble Level SCL012 SCL012 33 Reservoir Filter SCF061 SCF061
3 Cord Clip ADC058 ADC058 34 Fascia Gasket SCG009 SCG044
I

4 Probe Bracket Gasket SCG010 SCG010 35 Keypad (Old Style) SCK0155


U I
C K
C O
N
A US
E
28 35 Keypad SCK0406
H

5 Probe Bracket SCB039 SCB042


AR
Q
IF

M
A U T IO N !

!C
AU T I O N !

cp
c

19
!C

n
W
L

A EN

35 Keypad (New Style) SCK0417


S E

6 Probe Bracket Kit SCK038 SCK043


P E R
LO
PR O C K
ING

23 26 27
enser
Waste
Bottle 7 Microswitch Kit SCK007 SCK007 36 LCD Kit SCK0623 SCK0623
Cond
IMU
M
ERA
TIO
N

22 8 Thermocouple SCT030 SCT030 36 LCD Decal SCD0633 SCD0633


25
MAXLINE OP
FILL ILY a
DA for
, listen
in place
ect
k conn
quic .
ng ... e harm
re
n locki ensu caus
Whe " to may r ...
"click ...and n wate.
whe Line

36 Display Assembly SCA022 SCA022


lizer Fill

9 Copper Tubing Kit SCK048 SCK049


steri mum r ...
ate Mini wate
r oper...the lled
Neve h diste
widt
voir
reser
l the le.
Refil bott
the
d on t ...
icate le mus
...ind
e bott !
r wastlizer. N
ense steri TIO
cond ...the ...
The CAU ediately
!

37 Mounting Pad RPP819 RPP819


, imm

10 Check Valve Kit SCK011 SCK011


steam
leaks
ect
k conn
quic
the ce
If servi
your
tact
...con

24
al.
ssion
profe
M
IMU
MINLINE
FILL

FOOTNOTES: 1. Older styles only 2. Models with RPI's Condenser Waste Bottle Kit (RPI Part #SCK016) 3. Model #01-121101 4. Serial #2101DK0274 and above
5. Model #101102 (Old Type, One Piece Style with S/N's ending with an 'A') 6. Model #01-201103 7. Model #01-121101 (New Style, Two Piece Style)
8. Includes 90 Liquid Tight Strain Relief with Locking Nut (RPI Part #RPB365).

16 17
4.0 - PLANNED MAINTENANCE (PM) CHECKLIST
Sterilizer PM Kit (RPI Part #SCK035) Sterilizer PM Kit (RPI Part #SCK035)
For parts to fit the SciCan STATIM 2000 For parts to fit the SciCan STATIM 2000 - continued

PART RPI PART # STERILIZER PM TIP PART RPI PART # STERILIZER PM TIP
Cassette Seal SCS001 Refer to the Installation and Monthly Maintenance Guide included with the Water Filter (In-Line) SCF034 Debris is the #1 reason for pump failure, so before replacing this filter, be
Cassette Seal, or see page 22. or Pump Mesh Filters sure to check the reservoir for dirt and debris. Inspect and clean or replace
the Reservoir Cap & Filter (RPI Part #SCK059). Drain, clean and flush the reser-
Remove the old Cassette Seal. Clean the cassette groove and inspect the voir. Refill reservoir with distilled water after filter replacement.
cassette tray and lid for damage. Lubricate and install the new Cassette Seal.
If your autoclave comes equipped with an In-Line Water Filter, cut the cable
The OEM recommends lubrication of Cassette Seal every 30 days and ties and remove the old filter. Note the flow arrow direction on the filter body
replacement every 6 months or 500 cycles whichever comes first. Clean the (flow direction is from the reservoir to the pump). Install the new filter using
outer edge of cassette tray and the exposed area of the Cassette Seal weekly. the new cable ties included. Be sure to prime the pump in order to purge
Apply Sci-Dry (RPI Part #SCA054) every 10 cycles and after cleaning. trapped air before running a complete cycle. If your machine does not have
an In-Line Water Filter, then be sure to clean the pump mesh filters (located
Compressor Filter SCF002 Remove the pump filter cover. Remove filter and clean out any dust or debris. inside the pumps input and output port fittings). See page 6 and 7 - Figure
Replace filter and cover. If you find that the old Air Compressor Filter is wet 2.3 A and B.
or shows signs of water staining, then the check valve is leaking and must be
replaced to protect the air compressor from damage. RPI offers a Check Valve The OEM recommends replacing the Water Filter (In-Line) every 6 months or
Kit (RPI Part #SCK011) which includes translucent tubing making it easy to see 500 cycles whichever comes first.
a leaky check valve.
If the autoclave youre working on has the mesh filters, we recommend you
The OEM recommends replacing the Air Compressor Filter every 6 months or upgrade to the Drain Kit with Pump Filter (RPI Part #SCK037), making future
500 cycles whichever comes first. service much easier and providing a much better filtering system to protect
the pump (along with providing a quick & easy way to drain the reservoir
Biological Filter SCF032 This Biological Filter fits STATIM 2000 models with serial numbers through the front of the autoclave).
2101DK0274 and up, or if the autoclave has been upgraded. (Note: If your
autoclave is not equipped with this filter set-up, then you will not need this Other Routine Apply Sci-Dry (RPI Part #SCA054) every 10 cycles and after cleaning. Coat
filter). Maintenance all internal surfaces of cassette. Sci-Dry minimizes spotting and allows in-
struments to dry more efficiently.
Note the flow arrow direction on the filter body (flow direction is from the air
compressor to the check valve). If you find that the old Biological Filter is wet The OEM recommends the replacement of the Check Valve Kit (RPI Part
or shows signs of water staining, then the check valve is leaking and must be #SCK011) and the steam generator/boilers Safety Valve (70PSI) (RPI Part
replaced to protect the air compressor from damage. RPI offers a Check Valve #SCV004) every 2 years, and the use of the Reservoir Cap & Filter (RPI Part
Kit (RPI Part #SCK011) which includes translucent tubing making it easy to see #SCK059) on all versions of the STATIM autoclave.
a leaky check valve.
To aid in proper cleaning and maintenance of autoclaves, RPI also offers a
The OEM recommends replacing the Biological Filter every 6 months or 500 Cleaning Kit (RPI Part #RPK791) which includes a variety of brushes, sponges
cycles whichever comes first. and cleaning pads assembled in a convenient carrying case.

18 19
4.0 - PLANNED MAINTENANCE (PM) CHECKLIST
Sterilizer PM Kit (RPI Part #SCK036) Sterilizer PM Kit (RPI Part #SCK036)
For parts to fit the SciCan STATIM 5000 For parts to fit the SciCan STATIM 5000 - continued

PART RPI PART # STERILIZER PM TIP PART RPI PART # STERILIZER PM TIP
Cassette Seal SCS029 Refer to the Installation and Monthly Maintenance Guide included with the Water Filter (In-Line) SCF034 Debris is the #1 reason for pump failure, so before replacing this filter, be
cassette seal, or see page 22. or Pump Filters sure to check the reservoir for dirt and debris. Inspect and clean or replace
the Reservoir Cap & Filter (RPI Part #SCK059). Drain, clean and flush the reser-
Remove the old Cassette Seal. Clean the cassette groove and inspect the voir. Refill reservoir with distilled water after filter replacement.
cassette tray and lid for damage. Lubricate and install the new Cassette Seal.
If your autoclave comes equipped with an In-Line Water Filter, cut the cable
The OEM recommends lubrication of Cassette Seal every 30 days and ties and remove the old filter. Note the flow arrow direction on the filter body
replacement every 6 months or 500 cycles whichever comes first. Clean the (flow direction is from the reservoir to the pump). Install the new filter using
outer edge of cassette tray and the exposed area of the Cassette Seal weekly. the new cable ties included. Be sure to prime the pump in order to purge
Apply Sci-Dry (RPI Part #SCA054) every 10 cycles and after cleaning. trapped air before running a complete cycle. If your machine does not have
an In-Line Water Filter, then be sure to clean the pump mesh filters (located
Air Compressor Filter SCF031 The Air Compressor Filter is used with the Medo air compressor only. (Note: inside the pumps input and output port fittings). See page 6 and 7 - Figure
If your autoclave is not equipped with this type of air compressor, then you will 2.3 A and B.
not need this filter).
The OEM recommends replacing the Water Filter (In-Line) every 6 months or
If you find that the old Air Compressor Filter is wet or shows signs of water 500 cycles whichever comes first.
staining, then the check valve is leaking and must be replaced to protect the
air compressor from damage. RPI offers a Check Valve Kit (RPI Part #SCK011) If the autoclave youre working on has the mesh filters, we recommend you
which includes translucent tubing making it easy to see a leaky check valve. upgrade to the Drain Kit with Pump Filter (RPI Part #SCK037), making future
service much easier and providing a much better filtering system to protect
The OEM recommends replacing the Air Compressor Filter every 6 months or the pump (along with providing a quick & easy way to drain the reservoir
500 cycles whichever comes first. through the front of the autoclave).

Biological Filter SCF032 Note the flow arrow direction on the filter body (flow direction is from the air Other Routine Apply Sci-Dry (RPI Part #SCA054) every 10 cycles and after cleaning. Coat
compressor to the check valve). If you find that the old Biological Filter is wet Maintenance all internal surfaces of cassette. Sci-Dry minimizes spotting and allows in-
or shows signs of water staining, then the check valve is leaking and must be struments to dry more efficiently.
replaced to protect the air compressor from damage. RPI offers a Check Valve
Kit (RPI Part #SCK011) which includes translucent tubing making it easy to see The OEM recommends the replacement of the Check Valve Kit (RPI Part
a leaky check valve. #SCK011) and the steam generator/boilers Safety Valve (70PSI) (RPI Part
#SCV004) every 2 years, and the use of the Reservoir Cap & Filter (RPI Part
The OEM recommends replacing the Biological Filter every 6 months or 500 #SCK059) on all versions of the STATIM autoclave.
cycles whichever comes first.
To aid in proper cleaning and maintenance of autoclaves, RPI also offers a
Cleaning Kit (RPI Part #RPK791) which includes a variety of brushes, sponges
and cleaning pads assembled in a convenient carrying case.

20 21
4.0 - INSTALLATION AND MONTHLY MAINTENANCE FOR CASSETTE SEAL
FIGURE A FIGURE B FIGURE C
Cassette Seal (RPI Part #SCS001) to fit the SciCan STATIM 2000
Cassette Seal (RPI Part #SCS029) to fit the SciCan STATIM 5000

STEP #1 - The cassette must be opened and cooled before removing the cassette seal. See Figure A.

STEP #2 - Start with any of the (4) corners and begin prying the cassette seal loose using the RPI Cassette Seal Removal
Tool (RPI Part #RPT372) and by pulling down and out at the same time. See Figures B and C.

STEP #3 - Once the first corner is free, take hold of the cassette seal and pull down and out in order to remove it from the
cassette groove. Be careful not to tear the (2) locating tabs in each corner of the cassette seal. See Figures B
FIGURE D FIGURE E FIGURE F
and C.
Locating
Tabs
STEP #4 - Once the cassette seal has been removed, it is important to inspect, clean and remove any residue and/or debris
from the cassette groove and outer edge. Check for dings and dents on both halves of the cassette; if present, a
proper seal may not occur. See Figure D. Residue
and/or Port
Debris
STEP #5 - Using (1) package of RPI Liquid Soap (RPI Part #RPS287), lube the cassette seal completely. Note: It is not nec- Holes
essary to lube the inside of the cassette seal. See Figure E.

STEP #6 - Now you are ready to install the cassette seal. FIGURE G FIGURE H FIGURE I
STEP #7 - When installing the cassette seal, it is best to start in the back corner, nearest the port holes of the cassette.
Slip the cassette seal under and into the corner, placing the locating tabs at the edge of the open corner. Now
align the port holes and their locating tabs with the cassette. The locating tabs must protrude through the
square notches of the cassette lid. See Figure F.

STEP #8 - Install the other corners in the same manner. Now insert the cassette seal into the groove of the cassette one
side at a time. See Figures G and H.

STEP #9 - As you move along installing the cassette seal, check the corners to ensure that the locating tabs stay in place.
USAGE TIPS
Note: The sides should be inserted until smooth without waves or bumps. See Figure I.
Always note the position of the cassette when inserting
it into the unit if the cassette is installed upside down,
the thermocouple will bend and the sterilizer will become
Sci-Dry (RPI Part #SCA054) should be applied after
every 10 cycles and after cassette cleaning. Coat all in-
ternal surfaces of cassette. Sci-Dry minimizes spotting
STEP #10 - After you have completely installed the cassette seal, make a final inspection of all corners and all locating inoperative. and allows instruments to dry more efficiently.
tabs. Remove excess soap from the port holes as the soap can cause a build up on the steam ports and in
turn cause a bad seal.
The OEM recommends lubrication of Cassette Seal every
30 days and replacement every 6 months or 500 cycles
whichever comes first. Clean the outer edge of cas-
Watch for steam leaks from the cassette bay as they can
cause damage to the microswitch and other internal com-
ponents, thus rendering the sterilizer inoperative.
STEP #11 - Run a few test cycles to check for steam leaks, then apply Sci-Dry (RPI Part #SCA054). Note: During the sette tray and the exposed area of the Cassette Seal with
test cycles, it is normal to hear a hissing sound at first. The sound is from the excess liquid soap expelling
and the cassette seal seating in place.
RPI Liquid Soap (RPI Part #RPS287) weekly. Apply Sci-
Dry (RPI Part #SCA054) every 10 cycles and after Refill the reservoir with distilled water only. At the same
time that the reservoir is refilled, the condenser waste
bottle should be emptied and refilled. Refill the condenser
cleaning.
waste bottle with tap water to the Minimum Fill Level as
STEP #12 - After running the test cycles, if steam still leaks, reinstall the cassette seal and run additional test cycles. If
Always keep the inside of the cassette clean. Remove shown on the bottle. (Note: Never operate sterilizer when
steam leaks still occur, it is possible that the problem was not a faulty cassette seal, but something else, so residue and/or debris from the metal edge of the lower water level is above the Maximum Fill Line or below the
refer to the operators manual or contact your service professional. half of the cassette on a regular basis. Minimum Fill Line.)

22 23
5.0 SERVICE TIPS NOTES

1. CASSETTE DIAGNOSTICS move the tubing from the fitting, cut off approximately 7. STEAM GENERATOR/BOILER DIAGNOS- __________________________________________
If water is leaking at the drain from the cassette bay (un- from the end of the tubing, then reinsert the tubing TICS
derside of autoclave), then do the following: into the fitting that will usually remedy the problem. For older units with an aluminum steam generator/boiler, __________________________________________
Inspect and clean cassette seal and cassette. IMPORTANT: The RPI Push-In Elbow Fitting will not run a new cycle and view the LCD display. After comple-
Confirm that the cassette seal port holes are not dam- properly grip and retain the OEMs Teflon tubing. Only tion of the "Pressurization Stage", The "Sterilization __________________________________________
aged and that they are properly aligned within the cas- use RPIs Nylon Tubing (RPI Part #RPT370) with the Stage" should begin within 10 seconds. If it takes longer
sette lid groove. RPI Push-In Elbow Fitting (RPI Part #RPF363). than 10 seconds, the steam generator/boiler needs clean- __________________________________________
Ensure that there is no debris or residue on outer edges The Condenser Waste Bottle (RPI Part #SCK016) is a ing - see page 39 - Figure 9.2 "Cleaning the Steam
__________________________________________
of cassette tray. major component in the exhaust path and must not be Generator/Boiler". For Calibration Procedure see page
Verify that outer edges of cassette tray are straight (not allowed to fill up past the maximum line marked on the 40 - Figure 9.3. __________________________________________
nicked, dented, or damaged). bottle. The entire exhaust path (from the venturi plate in
Check the integrity of the hinge. the cassette through the solenoid valve to the waste bot- 8. CONTINUOUSLY RUNNING MAJOR __________________________________________
Inspect and clean steam ports on probe bracket. If you tle) must be free of kinks, pinch points or any restriction COMPONENT
observe excess debris that is not easily removed or if the or blockages. These conditions can cause malfunctions If the air compressor, solenoid valve, pump or steam gen- __________________________________________
steam ports appear to be worn or damaged, remove probe and create error codes to be displayed. erator/boiler activates, but does not turn off at its appro-
bracket for proper evaluation. Replace probe bracket if See associated error codes: Cycle Fault 6, 7, 8, 10, 11, 15 priate time, the main PC board is bad and must be __________________________________________
necessary. See pages 12 and 13 - Figures 2.10 A and B. (see pages 29 thru 31) and LCD messages SERVICE replaced. See pages 46 and 47 - Figure 9.8 "Electrical
NEEDED and CHECK CASSETTE (see page 34). Also Diagram". __________________________________________
2. CHECK VALVE consult Troubleshooting tips wraps/instruments remain
__________________________________________
A leaky check valve can cause moisture or steam to seep wet after drying (see page 37). 9. TEMPERATURE ADAPTER BOARD UP-
inside the air compressor thus causing the compressor to GRADE (ALEX BOILER KIT) __________________________________________
fail. The OEM recommends replacing the check valve and 4. GFI (GROUND FAULT INTERRUPTER) (AVAILABLE FROM THE OEM):
safety valve on the steam generator/boiler every two If Ground Fault Interrupter trips when the STATIM is turned This upgrade kit adds a third verification thermocouple __________________________________________
years. The RPI tubing in the Check Valve Kit (RPI Part on or plugged into an AC outlet, then check for: A leaky (which monitors the cassette exhaust temperature be-
#SCK011) is translucent - making it easy to see a leaky check valve (from the steam generator/boiler) which would tween the cassette and the solenoid valve). This kit also in- __________________________________________
check valve. For your convenience, RPI includes a new allow steam/water to enter the air compressor and could cludes an upgraded steam generator/boiler (also available
Check Valve at no extra cost in the Air Compressor Kit (RPI create a short circuit (see Service Tip #2 Check Valve). separately from the OEM), temperature adaptor board, __________________________________________
Part #SCK020). Its the RPI Advantage! microprocessor kit and miscellaneous plumbing parts to
5. KEYPAD complete the upgrade. See page 33 - 6.0 Error Codes: __________________________________________
3. CONDENSER WASTE BOTTLE Verify that the keypad cable is properly seated in the Cycle Fault 72 and 98
If excessive steam is escaping through the vent holes in main PC board connector. __________________________________________
the lid, verify that there is adequate water in the bottle If an RPI Keypad (RPI Part #SCK015, SCK040 or SCK041)
__________________________________________
(must be filled to the minimum fill line). has been installed, be sure the small label on the cable
If an exhaust leak is detected at the fitting junction on that reads THIS SIDE UP is facing up and away from __________________________________________
top of the lid, inspect and replace O-ring (RPI Part the PC board (see page 4 - Figure 2.1 A and C).
#RPO343) on the Panel Mount Quick Disconnect Fitting Replace keypad if necessary. __________________________________________
(RPI Part #RPF370).
Inspect Tubing and replace if discolored or brittle (RPI 6. LEVELING __________________________________________
Part #RPT380). Also inspect the connection at the In- For proper operation (before servicing the autoclave), ver-
Line Quick Disconnect Fitting (RPI Part #RPF371). ify that the autoclave has been properly leveled. Note the __________________________________________
When engaging the Quick Disconnect fittings, listen for position of the bubble in the bulls-eye Level (RPI Part
a click to ensure proper engagement & sealing. #SCL012) that is located on the top right of the autoclave. __________________________________________
If steam is leaking from the Push-In Elbow Fitting (RPI A properly leveled autoclave would show the bubble in the
__________________________________________
Part #RPF363) located at the rear of the autoclave, in- 4 to 5 oclock position. If leveling is improper, adjust the
spect the fitting and tubing connection. Service Tip: Re- feet until proper level is achieved. __________________________________________

24 25
6.0 LCD ERROR CODES At the start up of the sterilization process, the earlier mod- 6.0 LCD ERROR CODES (continued)
els display the operation message Select Program in the
(Note: Error Codes that are not currently used by Sci- LCD display panel, while the later models display Select STAGE #1 WARMING UP
Can are listed as "Not Currently In Use" or not listed A Cycle. Both messages, while different, actually mean During this stage, the distilled water heats up from room temperature to 95C in the steam generator/ boiler. The water is ac-
in this booklet.) the same thing the user is being asked to select which tually pumped from the reservoir and then heated in the steam generator/boiler.
of the three cycles to begin.
The key to STATIM error codes is to first identify which
model you are servicing because the terms used in the Error codes and operation messages indicated on the LCD LCD MESSAGE LCD MESSAGE
error codes and operation messages may differ between panel differ on the earlier models versus the later models. CHECK CASSETTE #1 (CYCLE FAULT #1) CHECK CASSETTE #2 (CYCLE FAULT #2)
the earlier and later models. The earlier models display the phrase Check Cassette The steam generator/boiler did not heat up. The cassette Cassette temperature (~95-100C) was not reached within
followed by a number 1-8, while the later models display temperature (~95-102C) was not reached within ~3 min- ~1 minute and 20 seconds (or within the time out period).
Cycle Fault followed by a number 1-27. With both the utes (or within the time out period).
Earlier STATIM 2000 Model 101102
earlier and later models, each number represents a dif- THINGS TO CHECK
Models Serial Number ends
with an "A" ferent type of error. (Note: Not all numbers are used, so THINGS TO CHECK Large steam leak is present.
only those numbers that are used will be addressed in this Check thermal fuse for continuity. Replace if necessary Over-loaded cassette tray.
Later STATIM 2000 Model 121101 booklet.) In addition to error codes, an operation message Thermal Fuse (RPI Part #SCF006) see page 38, Clean and lubricate cassette seal see page 10, 2.7 -
Models may also appear on the LCD display for both the earlier Figure 9.1 - Steam Generator/Boiler. Liquid Soap (RPI Part #RPS287). Replace the Cassette
STATIM 5000 Model 01-201103 Large steam leak is present. Seal if necessary (RPI Part #SCS001 fits model 2000),
and later models. Error codes and messages help to trou-
bleshoot the exact cause of the problem. In addition, the Over-loaded cassette tray. and (RPI Part #SCS029 fits model 5000).
It is important to know the software revision your STATIM system on both the earlier and later models is designed to Clean and lubricate cassette seal see page 10, 2.7 - Inspect and clean steam ports on probe bracket.
is running in order to properly troubleshoot and use these abort when an error occurs. Liquid Soap (RPI Part #RPS287). Replace the Cassette
error codes. Seal if necessary (RPI Part #SCS001 fits model 2000) or
During the sterilization process, there are six stages that (RPI Part #SCS029 fits model 5000).
At power ON, the display will read: the machine goes through from start to finish and the LCD Inspect and clean steam ports on probe bracket.
STATIM 2000 S201R501 displays each stage number as it occurs during the Check pump for proper operation see page 6, 2.3 -
process. The following sections describe each LCD error Pump Tester Bottle (RPI Part #SCK024).
This display will report whether you have a model 2000 or code or message that might appear during any of the six Check steam generator/boiler see page 38, Figure
5000 and shows the autoclave software revision level. The sterilization stages. In addition, this booklet includes trou- 9.1 - Steam Generator/Boiler.
revision level in this example shows REV. 501. This will bleshooting tips and "Things to Check" for each error code
only be displayed for 5 seconds. Then it will switch to se- and message.
lect a cycle.

STATIM autoclaves manufactured from 1990 to 1994


may display select a program and the software revi- STAGE #2 CONDITIONING
sion may not be shown.
After initial warming stage, the steam is injected into the cassette; the steam temperature continues to rise from 95C. The vent
solenoid will open and close several times to maintain proper cassette steam temperature, expelling steam/air into the con-
denser waste bottle.

For LCD Messages and Things to Check, refer to Stage #1 - Warming Up: Check Cassette #1 (Cycle Fault #1) and Check Cas-
sette #2 (Cycle Fault #2).

26 27
6.0 LCD ERROR CODES (continued) 6.0 LCD ERROR CODES (continued)
STAGE #3 PRESSURIZING STAGE #4 STERILIZATION
The pressure increases as the cassette temperature rises to the factory set temperatures for the Unwrapped, Wrapped and The temperature and pressure is maintained for the chosen cycle. Air/steam is being purged from the cassette as the vent
Rubber & Plastic cycles. The vent solenoid will open and close several times to maintain proper cassette steam temperature, solenoid valve opens and closes several times to maintain optimum temperature. The pump continues to pump distilled water
expelling steam/air into the condenser waste bottle. into the steam generator/boiler which in turn pumps saturated steam into the cassette. The condenser waste bottle collects
and allows the steam to cool down safely.
LCD MESSAGE (For Model 2000 only) LCD MESSAGE (For Model 2000 only)
CHECK CASSETTE #3 (CYCLE FAULT #3) CHECK CASSETTE #5 LCD MESSAGE LCD MESSAGE
The cassette fails to pressurize or reach temperature The steam generator/boiler may not be getting enough CHECK CASSETTE #6 CYCLE FAULT #6
(~95 - 110C) within ~70 seconds (or within the time out water or steam generator/boiler is suspect. Applies to the STATIM 2000, Model 101102 only (with Applies to STATIM 2000, model 121101, and STATIM
period). serial number that end with an "A" without a water 5000, model 01-201103
THINGS TO CHECK quality sensor).
THINGS TO CHECK A weak pump due to dirty mesh filters in the pump This Cycle Fault #6 will appear only with a water qual-
Clean and lubricate the cassette seal see page 10, fittings or dirty Water Filter (RPI Part #SCF034). Check The steam generator/boiler may be contaminated. The ity sensor installed.
2.7 - Liquid Soap (RPI Part #RPS287). Replace the pumps operation by using the Pump Tester Bottle Kit steam generator/boiler's temperature should be at least
Cassette Seal if necessary (RPI Part #SCS001 fits model (RPI Part #SCK024) see pages 6 and 7, 2.3 - Pump 5C higher than the cassette's temperature during the Steam generator/boiler calibration is needed see page
2000), and (RPI Part #SCS029 fits model 5000). Tester Bottle (RPI Part #SCK024). If pump passes test, Sterilization Stage. 40, Figure 9.3 - Steam Generator/Boiler Calibration
Inspect and clean steam ports on probe bracket. boiler must be replaced. Procedure. If problem (Error code) repeats, replace steam
Solenoid valve plunger may be compromised by debris Clean or calibrate steam generator/boiler see page 39 THINGS TO CHECK generator/boiler.
or may be worn. Check the plunger inside the flute/bon- Figure 9.2 or page 40 Figure 9.3. For earlier models only: The steam generator/boiler may
net for proper movement. Replace Solenoid Plunger Kit be contaminated by mineral deposits, calcium, etc., that THINGS TO CHECK
(RPI Part #SCK003) if necessary. See page 41, Figure have accumulated over time. These deposits are caused Water level in the waste bottle may be full, or the ex-
9.4 - Solenoid Valve. by using low quality water. haust path may have a restriction or blockage.
Check exhaust tubing for kinks or pinch points. Note: RPI
LCD MESSAGE Drain the reservoir. Clean the steam generator/boiler using offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
CHECK CASSETTE #4 (CYCLE FAULT #4) Jelmar CLR Cleaner (removes mineral deposits, calcium, haust Tubing sold by the foot (RPI Part #RPT380).
The "Sterilization Stage" fails to begin within specified etc). Refill the reservoir using high quality distilled water. Check the cassette tray venturi plate for blockage - see
time noted in the chart below. See page 14, 2.11 - Water Quality Tester and page 39, page 42, Figure 9.5.
Figure 9.2 - Cleaning the Steam Generator/Boiler. Check for steam leaks.
STATIM Model # Should begin "Sterilization Stage"
Solenoid valve plunger may be compromised by debris
2000 101102 Within ~ 10 minutes of reaching
S/N "A" temperature (~ 95-110C )
For later models only: Calibrate the steam generator/boiler or may be worn. Check the plunger inside the flute/bon-
2000 121101 Within ~ 6 minutes of reaching see page 40, Figure 9.3 - Steam Generator/Boiler net for proper movement. Replace Solenoid-Plunger Kit
temperature (~ 95-110C ) Calibration Procedure. (RPI Part #SCK003) if necessary see page 41, Figure
5000 01-201103 Within ~ 6 minutes of reaching 9.4 - Solenoid Valve.
temperature (~ 95-110C ) Important Note: The later STATIM models have a built-in
water quality sensor. Do not try to clean the later STATIMs'
THINGS TO CHECK steam generator/boiler. If you use a lesser quality water,
Refer to Stage #3 - Pressurizing: "Check Cassette #3 the LCD on the later STATIM will display "Water Quality is
(Cycle Fault #3)". not Acceptable" or "Distilled H20 only". The sensor de-
Check Safety Valve (70 PSI) (RPI Part #SCV004) for leaks tects unacceptable levels of dissolved solids in the reser-
and replace if necessary. voir water. See page 14, 2.11 - Water Quality Tester.
Leaky check valve see page 24, Service Tip #2. In-
spect the compressor air filter. If it is wet, replace the
Check Valve Kit (RPI Part #SCK011), the Air Filter (RPI Part
#SCF002) and/or the Biological Filter (RPI Part #SCF032).

28 29
6.0 LCD ERROR CODES (continued) 6.0 LCD ERROR CODES (continued)
STAGE #4 STERILIZATION (continued) STAGE #5 VENTING
LCD MESSAGE CHECK CASSETTE #8 (CYCLE FAULT #8) (continued) The sterilization cycle is completed and the vent solenoid valve opens to: 1) allow the pressure in the cassette to depressurize
CHECK CASSETTE #7 (CYCLE FAULT #7) THINGS TO CHECK (continued) via the exhaust path into the condenser waste bottle; and, 2) allow the steam to cool down safely.
The steam generator/boiler or cassette temperature has Solenoid valve plunger may be compromised by debris
LCD MESSAGE LCD MESSAGE
dropped below its set point due to a steam leak. or may be worn. Check the plunger inside the flute/bon-
CYCLE FAULT #11 CYCLE FAULT #14
net for proper movement. Replace Solenoid-Plunger Kit
Cassette temperature failed to drop within 60 seconds of Applies to STATIM 2xx software only.
THINGS TO CHECK (RPI Part #SCK003) if necessary. See page 41, Figure
the end of a cycle during venting.
Clean and lubricate the cassette seal see page 10, 9.4 - Solenoid Valve.
Steam generator/boiler temperature "ran away" (above
Figure 2.7 - Liquid Soap (RPI Part #RPS287). Replace If problem repeats, clean or calibrate steam generator/
THINGS TO CHECK 171C) in the Sterilization Stage.
the Cassette Seal if necessary, (RPI Part #SCS001 fits boiler - see page 39, Figure 9.2 or page 40, Figure
Check exhaust tubing for kinks or pinch points. Note: RPI
model 2000), and (RPI Part #SCS029 fits model 5000). 9.3.
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex- THINGS TO CHECK
Inspect and clean steam ports on probe bracket.
haust Tubing sold by the foot (RPI Part #RPT380). A weak pump due to dirty mesh filters in the pump fit-
If the cassette is hard to remove after venting takes LCD MESSAGE
Check the cassette tray venturi plate for blockage see tings or dirty Water Filter (RPI Part #SCF034). Check
place: Solenoid valve plunger may be compromised by CYCLE FAULT #9Not Currently In Use
page 42, Figure 9.5 - Cassette. pumps operation see pages 6 and 7, 2.3 - Pump
debris or may be worn. Check the plunger inside the
Check the solenoid valve coils electrical performance - Tester Bottle (RPI Part #SCK024). If pump passes test,
flute/bonnet for proper movement. Replace Solenoid- LCD MESSAGE
see page 41, Figure 9.4 - Solenoid Valve. boiler must be replaced.
Plunger Kit (RPI Part #SCK003), if necessary. See page CYCLE FAULT #10
Check for debris in the orifice and ports holes of the so- Steam generator/boiler calibration is needed see page
41, Figure 9.4 - Solenoid Valve. Temperature and pressure failing to decrease.
lenoid valve. 40, Figure 9.3 - Steam Generator/Boiler Calibration
Water level in the waste bottle may be full, or the ex-
Solenoid valve plunger may be compromised by debris Procedure.
haust path may have a restriction or blockage. THINGS TO CHECK
or may be worn. Check the plunger inside the flute/bon-
Check exhaust tubing for kinks or pinch points. Note: RPI Water level in the bottle may be full, or the exhaust path
net for proper movement. Replace Solenoid-Plunger Kit LCD MESSAGE
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex- may have a restriction or blockage.
(RPI Part #SCK003) if necessary. See page 41, Figure CYCLE FAULT #15
haust Tubing sold by the foot (RPI Part #RPT380). Check exhaust tubing for kinks or pinch points. Note: RPI
9.4 - Solenoid Valve. The cassette temperature climbed ~3C above the set
Check the cassette tray venturi plate for blockage - see offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
point limit in the Sterilization Stage.
page 42, Figure 9.5 - Cassette. haust Tubing sold by the foot (RPI Part #RPT380).
LCD MESSAGE
Check steam generator/boiler see page 38, Figure Check the cassette tray venturi plate for blockage
CYCLE FAULT #12 THINGS TO CHECK
9.1 - Steam Generator/Boiler. see page 42, Figure 9.5 - Cassette.
Failure to measure temperature. NOTE: This can happen at Water level in the waste bottle may be full, or the ex-
Check the solenoid valve coils electrical performance -
any stage. haust path may have a restriction or blockage.
LCD MESSAGE see page 41, Figure 9.4 - Solenoid Valve.
Check exhaust tubing for kinks or pinch points. Note: RPI
CHECK CASSETTE #8 (CYCLE FAULT #8) Check for debris in the orifice and port holes of the so-
THINGS TO CHECK offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
There is a 5C temperature difference between the steam lenoid valve.
Check for open thermocouple(s). At room temperature, haust Tubing sold by the foot (RPI Part #RPT380).
generator/boiler and the cassette. Solenoid valve plunger may be compromised by debris
it should read ~10 . Replace if necessary Thermo- Check the cassette tray venturi plate for blockage see
or may be worn. Check the plunger inside the flute/bon-
couple (RPI Part #SCT030). page 42, Figure 9.5 - Cassette.
THINGS TO CHECK net for proper movement. Replace Solenoid-Plunger Kit
Check for debris in the orifice and port holes of the
Water level in the waste bottle may be full, or the ex- (RPI Part #SCK003) if necessary. See page 41, Figure
LCD MESSAGE solenoid valve.
haust path may have a restriction or blockage. 9.4 - Solenoid Valve.
CYCLE FAULT #13Not Currently In Use Solenoid valve plunger may be compromised by debris
Check exhaust tubing for kinks or pinch points. Note: RPI
or may be worn. Check the plunger inside the flute/bon-
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex- For older units with an aluminum steam generator/boiler,
net for proper movement. Replace Solenoid-Plunger Kit
haust Tubing sold by the foot (RPI Part #RPT380). run a new cycle and view the LCD display. After comple-
(RPI Part #SCK003) if necessary. See page 41, Figure
Check the cassette tray venturi plate for blockage see tion of the "Pressurization Stage", The "Sterilization
9.4 - Solenoid Valve.
page 42, Figure 9.5 - Cassette. Stage" should begin within 10 seconds. If it takes longer
Steam generator/boiler calibration is needed see page
Check the solenoid valve coils electrical performance - than 10 seconds, the steam generator/boiler needs clean-
40, Figure 9.3 - Steam Generator/Boiler Calibration
see page 41, Figure 9.4 - Solenoid Valve. ing - see page 39 - Figure 9.2 "Cleaning the Steam
Procedure.
Check for debris in the orifice and port holes of the Generator/Boiler".
solenoid valve.

30 31
6.0 LCD ERROR CODES (continued) 6.0 LCD ERROR CODES (continued)
STAGE #5 VENTING (continued) STAGE #5 VENTING (continued)
LCD MESSAGE CYCLE FAULT #20 (continued) LCD MESSAGE LCD MESSAGE
CYCLE FAULT #16 THINGS TO CHECK (continued) CYCLE FAULT #27 CYCLE FAULT #72 and 98
Steam generator/boiler temperature "ran away" (above Steam generator/boiler calibration is needed see page The steam generator/boiler temperature has exceeded Communication error between the microprocessor and the
171C) in the autoclave. 40, Figure 9.3 - Steam Generator/Boiler Calibration 150C for 25 seconds (or a time out period). temperature adapter board.
THINGS TO CHECK Procedure.
A weak pump due to dirty mesh filters in the pump fit- THINGS TO CHECK If your autoclave is equipped with the temperature adapter
LCD MESSAGE A weak pump due to dirty mesh filters in the pump fit- board (Alex Upgrade), the following two Cycle Faults may
tings or dirty Water Filter (RPI Part #SCF034). Check
CYCLE FAULT #21 thru #24Not Currently In Use tings or dirty Water Filter (RPI Part #SCF034). Check apply see page 25, Service Tip #9.
pumps operation see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test, pumps operation see pages 6 and 7, 2.3 - Pump
LCD MESSAGE Tester Bottle (RPI Part #SCK024). If pump passes test, THINGS TO CHECK
boiler must be replaced. CYCLE FAULT #25
Check for continuous power (non-pulsing) to steam gen- boiler must be replaced. Check connection between the square microproces-
Unable to power up the steam generator/boiler. Steam generator/boiler calibration is needed see page sor and the main PC board. Also check the connection
erator/boiler. If continuous power is observed, replace
main PC board. THINGS TO CHECK 40, Figure 9.3 - Steam Generator/Boiler Calibration between the temperature adapter board and the main
Steam generator/boiler calibration is needed see page Check thermal fuse for continuity. If needed, replace the Procedure. PC board. Ensure there are no bent or shorted pins.
40, Figure 9.3 - Steam Generator/Boiler Calibration Thermal Fuse (RPI Part #SCF006) see page 38,
Procedure. Figure 9.1 - Steam generator/boiler. Validation thermocouple supplied with this Alex Boiler
A weak pump due to dirty mesh filters in the pump fit- Kit may require calibration, which is not covered in this
LCD MESSAGE tings or dirty Water Filter (RPI Part #SCF034). Check booklet. Consult OEM for this calibration procedure.
CYCLE FAULT #17 and #18Not Currently In Use pumps operation see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
LCD MESSAGE boiler must be replaced.
CYCLE FAULT #19 Steam generator/boiler calibration is needed see page
Steam generator/boiler calibration invalid. 40, Figure 9.3 - Steam Generator/Boiler Calibration
THINGS TO CHECK Procedure.
Steam generator/boiler calibration is needed see page
LCD MESSAGE
40, Figure 9.3 - Steam Generator/Boiler Calibration
CYCLE FAULT #26
Procedure.
The sterilization stage has failed ~3 minutes after the cas-
STAGE #6 DRYING
If problem repeats, replace the microprocessor/eprom
sette reaching the sterilization temperature. The air compressor forces air through the heated steam generator/boiler into the cassette, therefore drying the instruments.
(matched set) or main PC board, then recalibrate the
For 4xx and up software only: Cycle Interrupted is dis- The drying cycle will run for 60 minutes or until interrupted by pressing the Stop button. Keep in mind that when the Stop but-
steam generator/boiler see page 40, Figure 9.3 -
played 2 times, and Cycle Fault #26 is displayed upon ton is pressed, the following message is displayed: Please Wait - Cycle Complete. Remove cassette only when the message
Steam Generator/Boiler Calibration Procedure.
the 3rd failure. displays: Remove Cassette Cycle Complete".
LCD MESSAGE THINGS TO CHECK
CYCLE FAULT #20 Check for steam leaks. There are no error codes applicable to this stage.
Pump failure. Solenoid valve plunger may be compromised by debris
THINGS TO CHECK or may be worn. Check the plunger inside the flute/bon-
A weak pump due to dirty mesh filters in the pump fit- net for proper movement. Replace Solenoid-Plunger Kit
tings or dirty Water Filter (RPI Part #SCF034). Check (RPI Part #SCK003) if necessary. See page 41, Figure
pumps operation see pages 6 and 7, 2.3 - Pump 9.4 - Solenoid Valve.
Tester Bottle (RPI Part #SCK024). If pump passes test, Steam generator/boiler calibration is needed see page
boiler must be replaced. 40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.

32 33
7.0 LCD MESSAGES 7.0 LCD MESSAGES (continued)
APPLIES TO ANY STAGE APPLIES TO ANY STAGE (continued)
LCD MESSAGE SERVICE NEEDED (continued) LCD MESSAGE LCD MESSAGE
GARBLED OR NO READ-OUT ON THE LCD THINGS TO CHECK (continued) SELECT A CYCLE/PROGRAM OR INSERT CASSETTE REFILL RESERVOIR EMPTY WASTE BOTTLE
Check the cassette tray venturi plate for blockage see This can be displayed at any time before or during a Reservoir water level may be below minimum level. Cycle
THINGS TO CHECK page 42, Figure 9.5 Cassette. cycle/program. will not start.
Make sure that the cable from the LCD board (RPI Part Check for debris in the orifice and port holes of the so-
#SCA022 or Kit #SCK062) is properly seated at the main lenoid valve. THINGS TO CHECK THINGS TO CHECK
PC board connector. Solenoid valve plunger may be compromised by debris The cassette thermocouple may be bent and stopping Check reservoir water level and refill if necessary.
If using the Extension Cable (RPI Part #SCT026), make or may be worn. Check the plunger inside the flute/bon- the cassette from fully engaging the microswitch. For Check condenser waste bottle water level and empty if
sure that the cables are connected properly see page net for proper movement. Replace Solenoid-Plunger Kit additional information see the Installation instructions necessary.
4, 2.1 - Extension Test Cable. (RPI Part #SCK003) if necessary. See page 41, Figure for RPI Part #SCT030 available on RPIs website, See page 36, 8.0 Troubleshooting Tips - "Water Qual-
The microprocessor may need to be reseated in its 9.4 - Solenoid Valve. www.rpiparts.com. ity Sensor Troubleshooting".
socket. CAUTION: Disconnect power to the autoclave The steam generator/boiler may need to be cleaned The microswitch may have failed. The microswitch is a
and discharge the static electricity from your body see page 39, Figure 9.2. N.O. type switch. Replace the Microswitch Kit (RPI Part LCD MESSAGE
before touching the microprocessor (by touching the Steam generator/boiler calibration is needed see page #SCK007) if necessary. See page 9, Figure 2.5. CYCLE INTERRUPTED/NOT STERILE
autoclaves metal chassis with both hands). Failure to 40, Figure 9.3. Applies to all later STATIMs.
properly discharge static electricity can damage the LCD MESSAGE
microprocessor and make it unusable. Press firmly on LCD MESSAGE PRESS STOP TO RESET THINGS TO CHECK
the microprocessor and retest. CHECK CASSETTE Applies to all later STATIMs. A power failure or a time out condition occurred during a
NOTE: If none of the above solves the problem, replace (DOES NOT DISPLAY NUMBER ON LCD) cycle. The error must be cleared by pressing the Stop but-
either the LCD PC Board (RPI Part #SCA022 or Kit Applies to the original STATIM 2000 model 101102 only. THINGS TO CHECK ton.
#SCK062) or the main PC board. After an error fault occurs, the error must be cleared be-
The temperature in the cassette has not dropped below fore shutting down by pressing the Stop button. NOTE: If this message appears in three consecutive cycle
LCD MESSAGE 103C after the venting cycle. failures, Cycle Fault #26 will appear on the display. See
SERVICE NEEDED LCD MESSAGE page 32 Cycle Fault #26.
Applies to the original STATIM 2000 model 101102 only. THINGS TO CHECK WATER QUALITY IS NOT ACCEPTABLE OR DISTILLED
Steam generator/boiler has exceeded 171C or cassette Water level in the bottle may be full, or the exhaust path H2O ONLY
temperature has exceeded 147C. may have a restriction or blockage. The water placed in the reservoir is of low quality and has
Check exhaust tubing for kinks or pinch points. Note: RPI been detected by reservoirs water quality sensor.
THINGS TO CHECK offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex- Applies to the STATIM models with water quality sensor
If this message appears BEFORE depressing the Start haust Tubing sold by the foot (RPI Part #RPT380). only.
button, then check continuity of thermocouples. Ther- Check the cassette tray venturi plate for blockage see
mocouples should read ~ 10 at room temperature or page 42, Figure 9.5 - Cassette. THINGS TO CHECK
main PC board may be faulty. Check the solenoid valve coils electrical performance - Test the quality of the water being used, See page 14,
If this message appears AFTER depressing the Start see page 41, Figure 9.4 - Solenoid Valve. 2.11 - Water Quality Tester.
button, check pumps operation see pages 6 and 7, Check for debris in the orifice and port holes of the so- DO NOT TRY TO CLEAN THE STEAM GENERATOR/
2.3 - Pump Tester Bottle (RPI Part #SCK024). If pump lenoid valve. BOILER WITH ANY TYPE OF CLEANERS.
passes test, boiler must be replaced. Solenoid valve plunger may be compromised by debris Simply drain the reservoir and refill using distilled water
Water level in the waste bottle may be full, or the ex- or may be worn. Check the plunger inside the flute/bon- only. Do not use deionized, demineralized , or specialty
haust path may have a restriction or blockage. net for proper movement. Replace Solenoid-Plunger Kit filtered water. Tap water must never be used.
Check exhaust tubing for kinks or pinch points. Note: RPI (RPI Part #SCK003) if necessary. See page 41, Figure See page 36, 8.0 Troubleshooting Tips - "Water Qual-
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex- 9.4 - Solenoid Valve. ity Sensor Troubleshooting".
haust Tubing sold by the foot (RPI Part #RPT380).

34 35
8.0 TROUBLESHOOTING TIPS 8.0 TROUBLESHOOTING TIPS (continued)

CONDITION: CONDITION: CONDITION: CONDITION:


WATER QUALITY SENSOR TROUBLESHOOTING WILL NOT POWER UP WRAPS/INSTRUMENTS REMAIN WET AFTER DRYING IF "BUZZING NOISES" ARE HEARD
Applies to the STATIM models with water quality sensor Sounds like the solenoid valve is buzzing with the cassette
only. THINGS TO CHECK THINGS TO CHECK inserted or at improper times during a cycle.
Check power source to power cord, AC receptacle to Check air filters, and replace if necessary (RPI Part
THINGS TO CHECK power switch, and wiring to the main PCB. #SCF002, SCF031 and SCF032). THINGS TO CHECK
Troubleshooting Procedure: Check 15 amp ceramic fuse. Check leveling of the STATIM Note the position of the Solenoid valve plunger may be compromised by debris
STEP #1 - Verify that the main PC board is operational. Inspect, clean and adjust contacts of fuse holder if nec- bubble in the Bulls-Eye Level (RPI Part #SCL012) that is or may be worn. Check operation of plunger inside the
Measure the voltage between P1-1 and P1-3. The reading essary. located on the top right of the autoclave. A properly lev- flute/bonnet for proper movement. Replace Solenoid-
must be between 8.4 and 9.7 VDC. If not, the main Main PC board is bad and must be replaced. eled autoclave would show the bubble in the 4 to 5 Plunger Kit (RPI Part #SCK003) if necessary. See page
PC board is bad, see pages 46 and 47, Figure 9.8 - For all of the above see pages 46 and 47, Figure 9.8 - oclock position. If leveling is improper, adjust the feet 41, Figure 9.4 - Solenoid Valve.
Electrical Diagram. Electrical Diagram. until proper level is achieved.
Water level in the waste bottle may be full, or the ex- CONDITION:
STEP #2 - Remove power from autoclave, disconnect the CONDITION: haust path may have a restriction or blockage. IF "RATTLING NOISES" CAN BE HEARD
water quality sensor from J4-3 and J4-4. Restore power WILL NOT PRINT Check exhaust tubing for kinks or pinch points. Note: RPI Sounds coming from air compressor during the drying
and begin an unwrapped cycle. If cycle starts with sen- After an error fault occurs, the error must be cleared be- offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex- cycle or at improper times during a cycle.
sor disconnected, then the main PC board is bad. If RE- fore shutting down by pressing the Stop button. haust Tubing sold by the foot (RPI Part #RPT380).
FILL RESERVOIR EMPTY WASTE BOTTLE error appears Check the cassette tray venturi plate for blockage - see THINGS TO CHECK
on LCD proceed to the next step. THINGS TO CHECK page 42, Figure 9.5 - Cassette. Check the air compressor filter. If stains or moisture are
If the time and date are not displayed on the printer's Check air compressor for proper operation - see pages present, replace the Air Compressor Filter (RPI Part
STEP #3 - Short probe terminals J4-3 and J4-4 together LCD, check power cable connections. 44 and 45, Figure 9.7 - Air Compressor. #SCF002) and the Check Valve (RPI Part #SCK011).
and start a cycle. If WATER QUALITY NOT ACCEPTABLE Check the loading of printer paper, paper must roll out Check solenoid valve for proper operation - see page
displays on LCD proceed to Step #4; if not the main PC from the top. Verify that the paper is not jammed. 41, Figure 9.4 - Solenoid Valve. Note: Once the air compressor filter shows signs of mois-
board is bad. Replace printer if necessary. ture, the check valve has failed and must be replaced.
CONDITION: Damage will occur with the air compressor if unattended.
STEP #4 - Remove the short from J4-3 and J4-4. Short CONDITION: IF "CLICKING NOISES" ARE HEARD This can lead to a costly repair.
the float terminals together on the main PC board (J4-5 & LEAKING STEAM: Sounds like the solenoid valve is being activated without
J4-6). If the cycle starts, the main PC board is good and the PRV (PRESSURE RELIEF VALVE) the cassette inserted or at improper times during a cycle. CONDITION:
water quality sensor is bad. If the LCD reads WATER (Also known as Safety Valve) IF MAJOR COMPONENT CONTINUOUSLY RUNS
QUALITY NOT ACCEPTABLE, the main PC board is bad. THINGS TO CHECK Air compressor, solenoid valve, pump or steam genera-
THINGS TO CHECK The microswitch may have failed. One cause for this fail- tor/boiler activates and does not turn off at it's appropri-
STEP #5 - To verify the water quality sensor, short the two Check cassette seal port holes and steam ports on probe ure could be a cassette steam leak. Repair leak and ate time.
pins coming from the front of the sensor together. The sen- bracket (look for obstruction). Clean and lubricate the retest. If condition persists, replace the Microswitch Kit
sor should read less than 1 . cassette seal see page 10, Figure 2.7 - Liquid Soap (RPI Part #SCK007). See page 9, 2.5 - Hex Balldriver THINGS TO CHECK
(RPI Part #RPS287). Replace the Cassette Seal (RPI Part (5/64") (RPI Part #RPT297), For Use When Replacing Main PCB board is bad and must be replaced.
Upon verification or replacement of water quality sensor, #SCS001 fits model 2000) or (RPI Part #SCS029 fits the Microswitch.
test the reservoirs water quality using the Water Quality model 5000) if necessary.
Tester (RPI Part #RPT820). See page 14, Figure 2.11 - NOTE: This can also happen with the cassette installed
Water Quality Tester. Check the plumbing from the steam generator/boiler to before or just after starting a cycle. The steam works its
the probe bracket fittings (look for obstruction). See way past the switchs seal into the switch causing the N.O.
page 38, Figure 9.1 - Steam Generator/Boiler. contacts to short out and become closed (or possibly an in-
If the problem repeats, replace the Safety Valve (70 psi) termittent signal). The autoclave would then proceed as if
(RPI Part #SCV004). the microswitch had been activated.

36 37
9.0 REFERENCE BLOCKS 9.0 REFERENCE BLOCKS (continued)
FIGURE 9.1 - STEAM GENERATOR/BOILER FIGURE 9.2 - CLEANING THE STEAM GENERATOR/BOILER
To check the thermal fuse continuity, disconnect fuse To check steam generator/boiler heater, disconnect For later models only: Calibrate the steam generator/ STEP #4 - Now start an "Unwrapped Cycle". As soon
assembly at main PC board J1-3. Measure resistance fuse assembly at main PC board J1-3. Measure re- boiler see page 40, Figure 9.3 Steam Generator/ as you hear the pump begin running, quit the cycle by
of thermal fuse assembly. An open reading signifies sistance between the two steam generator/boiler Boiler Calibration Procedure. IMPORTANT NOTE: The pressing the Stop button. This step is to add water to
the fuse is bad and must be replaced. Reconnect wire terminals, resistance should read 11 . If not, re- later STATIM models have a built-in water quality sen- the steam generator/boiler and warm up the solution
at main PC board J1-3. IMPORTANT NOTE: To avoid place boiler. Reconnect wire at main PC board J1-3. sor. Do not try to clean the later STATIMs steam gen- inside the steam generator/boiler. Let the cleaner go to
damaging the new thermal fuse assembly try to iden- Check line voltage at main PC board J1-3 and J1-4 erator/boiler. If you use a lesser quality water, the LCD work for about 15 minutes.
tify the cause of the failure. There are a number of during warm up cycle. If reading is correct, verify on the later STATIM will display Water Quality is not
causes for thermal fuse failure which include a weak voltage at steam generator/boiler terminals. Acceptable or Distilled H20 only. The sensor detects STEP #5 - Using the RPI Start Switch (RPI Part
water pump, steam leaks, faulty thermocouple, and/or Steam generator/boiler thermocouple should read ~ unacceptable dissolved solids over the limits in the #SCS021). Connect the smart switch to pin #'s 16 and
malfunctioning steam generator/boiler (which may re- 10 at main PC board screw terminals (J6 Boiler). water. Verify water quality, See page 14, 2.11 - Water 18 on the 20 pin connector (P1) on the main PC board.
quire cleaning or calibration). Quality Tester. See page 5, Figure 2.2.

Tubing (only) (Translucent) Safety Valve (70 PSI) (RPI Part #SCV004) For earlier models only: The steam generator/boiler STEP #6 - Be sure the water reservoir is full and the
(RPI Part #RPT373) OEM recommends replacing the Check may be contaminated by mineral deposits, calcium, condenser waste bottle is at its minimum fill level.
If moisture or steam is present Valve and Safety Valve every two years etc., that have accumulated over time. These deposits Select a new unwrapped cycle then press and lock
in this tube, replace Check Valve Check Valve Kit are caused by using low quality water. Verify water the RPI Start Switch (RPI Part #SCS021). This will man-
(RPI Part #SCK011) quality, See page 14, 2.11 - Water Quality Tester. ually run the pump. Continue running the pump for ~3
To Air Compressor To Steam Injection Port
minutes (to purge/flush all debris from the steam gen-
on Probe Bracket The cleaning steps below apply to the STATIM 2000, erator/boiler).
Model 101102 only (with serial number that ends with
To Metering Tube on Pump an "A"), without a water quality sensor. STEP #7 - Remove the start switch, drain the con-
Add CLR here when cleaning denser waste bottle, refill the reservoir, and thoroughly
steam generator/boiler. Steam Generator/Boiler
See page 39, Figure 9.2 STEP #1 - Allow the steam generator/boiler to cool clean the cassette (removing debris purged/flushed
down, then remove the safety valve from the top of the from steam generator/boiler cleaning). Run a new un-
steam generator/boiler, See page 38, Figure 9.1. wrapped cycle and view the LCD display. After com-
pletion of the "Pressurization Stage", The "Sterilization
Steam Generator/Boiler To Main PC Board J1-4 STEP #2 - Measure ~1 oz. of CLR cleaner and pour it Stage" should begin within 10 seconds. If it takes
Thermocouple into the steam generator. Clean up any spills or drips longer than 10 seconds, repeat the cleaning procedure.
from the steam generator/boiler immediately.
NOTE: If cleaning the steam generator/boiler repeat-
Screw (#6-32 x 1/4) STEP #3 - Reinstall the safety valve into the steam edly does not eliminate the problem, replace the steam
(RPI Part #ADS054)
Screw (#8-32 x 1/2-SEMS) generator/boiler (use a thread sealer on the threads of generator/boiler.
(RPI Part #RPH817) the safety valve).
3 places
Double Thermal
Fuse Assembly
(RPI Part #SCF006)

DATA CHART
Steam Generator/Boiler Heater: ~11 (J1-3 & J1-4)
Thermocouple: ~10 @ Room Temp. (J6 Boiler) To Main PC Board J1-3

38 39
9.0 REFERENCE BLOCKS (continued) 9.0 REFERENCE BLOCKS (continued)
FIGURE 9.3 - STEAM GENERATOR/BOILER CALIBRATION PROCEDURE FIGURE 9.4 - SOLENOID VALVE
Applies to STATIM 2000 and 5000 STEP #8 - The calibration continues and it takes dif- SERVICE TIP
STEP #1 - Turn the autoclave off and remove the power ferent paths depending on whether it is a new style or Avoid crushing the flute/bonnet Solenoid Valve Electrical Data
cord from the AC outlet. Remove the cover, refill the reser- old style main PC board. (which would prevent the plunger 1) Coil Resistance: 3.25- 3.96 Meg.
voir (with distilled water) and install the RPI Extension Test from moving up and down freely) NOTE: This Coil is a FWR Coil and the
Old Style: Calibration will take ~ 5 mins. A long beep resistance must be checked on both
Cable (RPI Part #SCT026). See page 4, Figure 2.1. by using a Spanner Nut (RPI Part
will be heard upon completion. The 00 on the display leads. Then reverse the leads and
#RPT364) or Spanner Wrench (RPI Nut
STEP #2 - Connect your RPI Start Switch (RPI Part will change similar to Figure 3 below. recheck. The resistance must be the
Part #RPT501).
#SCS021) to the pins marked W1 on the main PC board. same in both directions or the coil is
Now press the STOP button allowing the system to Inspect and clean plunger rubber
See pages 46 and 47, Figure 9.8 - Electrical Diagram. Coil bad.
vent. Display will read PRESS STOP TO RESET when seat, replace if necessary. Uneven (RPI Part #SCC005)
STEP #3 - Push and slide the button on the Start the venting is complete. Pressing the STOP button wear could indicate a problem.
Switch to lock the switch in the ON position. again will reset the autoclave and the display will now Lubricate the plunger (metal body 2) Coil AC Line Voltage: 120V/60 Hz, 9W.
read SELECT A CYCLE (SELECT A PROGRAM). Cali- only) with a drop of 3-IN-ONE oil. While in the Warming Stage, Line
STEP #4 - Plug in the power cord and turn the power
bration is complete and the autoclave is ready for nor- This can also help reduce valve Voltage should be present at termi-
switch to "ON".
mal use. noise. nals J1-7 and J1-8 on main PC
STEP #5 - The LCD display should appear similar to Inspect and clean valve body orifice board. If not, main PC board is bad.
New Style: When cycle reaches STERILIZATION
Figure 1 below. and ports, looking for damage and
PHASE unit will beep twice. Autoclave will automati-
STEP #6 - Insert cassette. cally vent after 3-1/2 minutes. The boiler calibration is debris.
Coil Bracket
now complete and the LCD display should change sim- Check that the coil has been rein-
STEP #7 - Begin calibration procedure by holding down stalled in the proper orientation. Flute/Bonnet
the unwrapped button and simultaneously pressing ilar to Figure 3 below.
the start button. The display should appear similar to After venting, press the STOP button to end calibra- Shading Disk Spanner Nut
Figure 2 below. tion cycle. Autoclave is now ready for normal use. (RPI Part #RPT364)
Spring

Plunger Solenoid Plunger Kit


(RPI Part #SCK003)
FIGURE 1 FIGURE 2 FIGURE 3 O-ring
Orifice
OLD STYLE OLD STYLE OLD STYLE
Valve Body Copper Tubing Kit
F9> CALIBRATION <C8 00> CALIBRATION* <C8 F6> CALIBRATION <C8 (RPI Part #SCK048) &
(RPI Part #SCK049)
+32.00C 65.00C +32.00C 65.00C +2.00C 135.00C Push-in Elbow Fitting
(RPI Part #RPF362 Male Connector
Old Style: Upper Left corner will Port #2 Port #1 (RPI Part #RPF227)
change to "00" and an asterisk I
ive
es 5

h
CT

Ad
4
will appear after CALIBRATION*

LO5
NEW STYLE NEW STYLE NEW STYLE
Thread Locker
25.5 C8 24.1 F9 25.5 00* 24.1 F9 145.5 F3* 135.9 F9 (RPI Part #RPA369)
1.4 1.4 9.6
Mounting Bracket
New Style: Unit will beep 3 times. Screw (#8-32 x 1/2 -SEMS) (not included)
C8 will change to 00* (RPI Part #RPH817)
2 places

40 41
9.0 REFERENCE BLOCKS (continued) 9.0 REFERENCE BLOCKS (continued)
FIGURE 9.5 - CASSETTE FIGURE 9.6 - PUMP
CASSETTE DIAGNOSTICS: Verify that outer edges of cassette tray are straight
If water is leaking at the drain from the cassette bay (not nicked, dented, or damaged). ULKA VERSION PUMP Use this clear protective tubing
(underside of sterilizer), then do the following: Check the integrity of the hinge. with Autoclave Software R604 and above. to avoid water leakage onto AC
input wiring
Remove and clean cassette seal and groove. Rein- Check the cassette tray and lid for warpage.

Flow
stall cassette seal. Run a cycle and reinspect for leak- Straighten or replace as necessary. ULKA Pump was introduced in conjunction
age. See page 10, Figure 2.6. Inspect and clean steam ports on probe bracket. If with Main PC Board Rev. 7 (Software ver-
Confirm that the cassette seal port holes are not dam- you observe excess debris that is not easily removed, sion R604) and was never supplied with
aged and that they are properly aligned within the or if the steam ports appear to be worn or damaged, mesh filter installed into inlet and outlet
cassette groove. remove probe bracket for proper evaluation. Replace fittings. NOTE: Ulka Pump can be used on
Ensure that there is no debris or residue on outer probe bracket if necessary. See page 13, Figure Main PCB Rev. 6.4 with upgraded software

Flow
edges of cassette tray. 2.10B. to Rev. R604 and above.

1
2
Black (Line)
BACK REAR VIEW
Cassette Lid Square Notches
Injection Steam Port
Exhaust Steam Port } Located on
Probe Bracket
To Main PCB J1-5
White (Neutral)

1
(for cassette seal alignment)

2
To Main PCB J1-6
Spring Clip

INVENSYS VERSION PUMP


Steam Port
with Autoclave Software R603 and older.

Flow
sealing
Clean/Remove Mesh Filter
surfaces Effective March 2006, all pumps were
supplied with the spring clip (grounding)
Cassette Lid RPI Cassette Seal Part Number Over-Temp Switch
to improve water conductivity circuitry.
Cassette Seal Port Holes is located in this approximate
location (inside Seal Channel) In early 2006, the OEM introduced the
Locating Tabs Clean/Remove
and Port Holes Reservoir Cap & Filter (RPI Part #SCK059). Mesh Filter
Lid Effective June 2007, the OEM stopped
using the mesh filter (installed inlet and
Venturi Plate outlet fittings) and offered the In-Line

Flow
Water Filter (RPI Part #SCF034) to be in-
stalled between the reservoir and pump.

1
Tray The OEM now recommends the use of

2
Black (Line)
the Reservoir Cap & Filter (RPI Part To Main PCB J1-5
#SCK059) along with the In-Line Water White (Neutral)

1
Filter on all units. The mesh filters should

2
To Main PCB J1-6
be removed and the autoclave should be
Check these
upgraded to the In-Line Water Filter (Drain
Kit with Pump Filter RPI Part #SCK037). Spring Clip Use this clear protective tubing
areas for debris (Ground Contact) to avoid water leakage onto AC
build up input wiring

42 43
9.0 REFERENCE BLOCKS (continued) 9.0 REFERENCE BLOCKS (continued)
FIGURE 9.7 - AIR COMPRESSOR FIGURE 9.7 - AIR COMPRESSOR (continued)
The OEM recommends replacing the compressor filters every 6 months or every 500 cycles (whichever comes first). The OEM recommends replacing the compressor filters every 6 months or every 500 cycles (whichever comes first).

MEDO PUMP Output: 2.84 PSI CAUTION: THOMAS PUMP Output: ~ 4.28 PSI CAUTION:
STATIM 2000 (1990 to present ) NEVER OIL STATIM 5000 (2003 thru 2011) NEVER OIL
Air Compressor Kit (RPI Part
UNITS WITHOUT BIOLOGICAL FILTER The THOMAS Pump draws unfil-
#SCK020) - includes Check Valve
tered air into sterilizer. Air is then
Kit (RPI Part #SCK011).
Steam Generator/Boiler Fitting pushed through the Biological
- To avoid possible valve failure, Filter (RPI Part #SCF032) and Flo
The MEDO Pump draws air w
debris must be completely re- through the check valve into the
through the Compressor Filter
moved from the fitting before steam generator/boiler
(RPI Part #SCF002) located in the
installing new Check Valve (RPI
front of the compressor. Air is
part #SCK011) using supplied
pushed through the Biological
threadlocker on fitting threads.
Filter (RPI Part #SCF032), if the
unit has one. Then air passes
Input
through the check valve into the
steam generator/boiler. Biological Filter
(RPI Part #SCF032)
White (#10)
UNITS WITH BIOLOGICAL FILTER To Main PCB J1-10
Black (#9)
Steam Generator/Boiler Fitting To Main PCB J1-9
- To avoid possible valve failure,
debris must be completely re-
moved from the fitting before MEDO PUMP Output: 4.28 PSI CAUTION:
installing new Check Valve (RPI STATIM 5000 (1995 thru 2003 & NEVER OIL
part #SCK011) using supplied 2011 to present)
threadlocker on fitting threads.
The MEDO Pump draws air from
Flo the Air Compressor Filter (RPI
w Part #SCF031) mounted to the Flo
w
outside of the rear chassis of the
sterilizer. Air is then pushed Input
through the Biological Filter (RPI
Part #SCF032) and through the
check valve into the steam gen-
erator/boiler.
Input w
Biological Filter Flo
(RPI Part #SCF032) White (#10)
If present Air Compressor Filter White (#10)
To Main PCB J1-10 (RPI Part #SCF031) To Main PCB J1-10
Compressor Filter
(RPI Part #SCF002) Black (#9) Biological Filter Black (#9)
To Main PCB J1-9 (RPI Part #SCF032) To Main PCB J1-9

44 45
9.0 REFERENCE BLOCKS (continued) FIGURE 9.8 - ELECTRICAL DIAGRAM
Data Chart
Steam Generator/Boiler Heater: -11 (J1-3 & J1-4)
Water Solenoid Coil (FWR): 3.25 - 3.96 Meg
Quality Thermocouples: ~10 @ Room Temp. J5 Cassette; J6 Boiler
Sensor Thermal Fuses: Open @ 216C
Thermocouple Wires used in this
terminal block need Diagnostic LED Reference
to be solder tinned, Black Reservoir
CR5 Steam Generator/Boiler
Microswitch or an insulation
White Diagnostic LED's will be illumi-
ferrule be used. CR6 Air Compressor

Yellow
Blue nated when their corresponding
device is switched ON CR7 Solenoid Valve
Red

Shield
CR10 Pump

Red
Red
Float (Not Shown) Blue

Yellow

Shield
If the Air Compressor, Solenoid Valve, Pump or Steam Generator/Boiler
Blue
Green/Yellow activates, but does not turn off at the appropriate time, the Main PCB Board
Line is bad and must be replaced.
White J1-10 F4
Neut. Black CR5

J3-1

J4-1
GND
J3-6

J4-6
J1-9
White J1-8
CR7 +- +- Fuse Chart
Chassis White J1-7 CR10 J5 J6 BOILER Earlier Style Later Style
Gnd White J1-6 CR6 CHAMBER
Replacement Fuse 2000 & 5000 2000 & 5000
Black P2
J1-5
#14 White P1 P3 RPI Part #RPF060 (1/2A)
AWG J1-4 Cycle 20 19
Black 18 17 1/4" x 1-1/4" Glass F3 & F4
F2 F1
Power Line Black 16
14
15
13
2000 Only
White W1 RPI Part #RPF061 (1A) In-Line Pump
Switch P4 Power
12
10
11
9 1/4" x 1-1/4" Glass (Not Shown)
Neut. J1-1 8 7 2000 Only
#14 F3 6 5
4 3 RPI Part #RPF071 (15A) F1 & F2 F2 & F3
AC Receptical

AWG 2 1 1/4" x 1-1/4" Ceramic Line Voltage Line Voltage

Line
P1 RPI Part #RPF367 (2A)
#14 AWG Steam
Green/Yellow

5mm x 20mm Glass F1


#14 AWG Neut.
Generator/Boiler

Double Thermal Fuse


Line RPI Part #RPF368
5mm x 20mm Glass F2
Neut.
Thermocouple
Over-Temp. Heater Device Activation Chart (Jumpering)
Switch Pump P1-16 to J3-5 (GND)
Solenoid P1-18 to J3-5 (GND)
FWR Air Compressor P1-20 to J3-5 (GND)

Keypad Chart Connect to P4


Air Solenoid Invensys Pin # 1 2 3 4 5 6 7
Compressor Valve Pump Function Stop Start Hemostat Hand Closed Open Common
Scissors Scissors

Unmarked wires are #22 AWG

46 47
9.0 REFERENCE BLOCKS (continued)
FIGURE 9.9 - PLUMBING DIAGRAM
Female Quick Connect
Fitting (Drain Line)
Condenser Waste Bottle Kit
(RPI Part #SCK016)
Exhaust Tubing Kit
(RPI Part #SCK017)
Kit includes 8' of Tubing
(RPI Part #RPT380) and RPT380
(5 pieces) Cord Clip

Injection Port
(RPI Part #ADC058)

Exhaust Port
In-Line
Quick Disconnect

Steam

Steam
(RPI Part #RPF371)
ADC058
(x5)

Panel Mount
Quick Disconnect
(RPI Part #RPF370)
Includes O-Ring
(RPI Part #RPO343)

Condensation Coil
(RPI Part #SCC019)

Reservoir
Thermocouple
Condenser Waste
Bottle With lid
(RPI Part #SCB018)

Pump

TIP - Upgrade to Solenoid Check


Air Valve Valve
RPI's "User-Friendly" Compressor
Condenser Waste Bottle
Kit (RPI Part #SCK016).
Steam In-Line
Our one-of-kind design Generator/Boiler Water Filter
allows your customer to re-
move the Condenser Waste Bottle
Biological
from the exhaust tubing in one easy step. Filter
Simply push and release the quick con-
nect on the lid and the bottle is free from
the exhaust tube for easy maintenance. Condenser Waste Bottle

48 49
10.0 - RPI PARTS LIST 10.0 - RPI PARTS LIST

00

00
00

00
(FOR PARTS TO FIT STATIM 2000 & 5000) (FOR PARTS TO FIT STATIM 2000 & 5000) (continued)

20

50
20

50

IM

IM
IM

IM
RPI RPI

AT

AT
AT

AT

ST

ST
ST

ST
Part # OEM Part # Description Part # OEM Part # Description
ADC058 No OEM Part # Available CORD CLIP - 10/PKG SCK003 01-100998S SOLENOID-PLUNGER KIT
CAF011 No OEM Part # Available RUBBER FOOT - 4/PKG SCK007 01-100571S (2000) / MICROSWITCH KIT
RPB365 53-100518A 90 LIQUID TIGHT STRAIN RELIEF WITH LOCKING NUT 01-101621S (5000)
RPB703 01-108152S (Kit) BARB TEE - 5/PKG SCK011 01-100574S / 01-1016275 CHECK VALVE KIT
RPC288 No OEM Part # Available POWER CORD SCK015 01-101707S KEYPAD 5
RPC291 01-101647S POWER CORD SCK016 01-100812S CONDENSER WASTE BOTTLE KIT
RPC701 01-108152S (Kit) CABLE CLAMP - 5/PKG SCK017 01-100204S EXHAUST TUBING KIT
RPC702 01-108152S (Kit) CABLE HOLDER - 5/PKG SCK020 01-100560S AIR COMPRESSOR KIT
RPF060 No OEM Part # Available FUSE (1/2A, 250V) - TIME DELAY - 5/PKG SCK025 01-100572S POWER CORD REPLACEMENT
RPF061 No OEM Part # Available FUSE (1A, 250V) - TIME DELAY - 5/PKG 1 SCK035 01-100028S (Gasket only) STERILIZER PM KIT
RPF071 01-103472S FUSE (15A, 250V) - TIME DELAY - 5/PKG 1 SCK036 01-101649S (Gasket only) STERILIZER PM KIT
RPF362 01-101553S LEVELER LEG - 4/PKG SCK037 01-108152S (kit) & DRAIN KIT WITH PUMP FILTER
RPF363 01-100782S PUSH-IN ELBOW FITTING 01-106637S (kit)
RPF367 01-103498S FUSE (2A, 250V) - 5/PKG SCK038 01-100570S PROBE BRACKET KIT
RPF368 01-103497S FUSE (1/4A, 250V) - TIME DELAY - 5/PKG SCK040 01-109671 KEYPAD 6
RPF370 No OEM Part # Available PANEL MOUNT QUICK DISCONNECT 2 2 SCK041 01-109680S KEYPAD 7
RPF371 No OEM Part # Available IN-LINE QUICK DISCONNECT 2 2 SCK042 01-101618S PROBE BRACKET KIT
RPF377 No OEM Part # Available PUSH-IN ELBOW FITTING SCK048 01-109128S COPPER TUBING KIT
RPF379 No OEM Part # Available UNION FITTING (COMPRESS-ALIGN 1/8") SCK049 01-109129S COPPER TUBING KIT
RPF428 No OEM Part # Available FEMALE QUICK CONNECT FITTING (WHITE) SCK059 01-101783S / RESERVOIR CAP & FILTER
RPF429 No OEM Part # Available MALE QUICK CONNECT FITTING (WHITE) 01-109300S (Filter Only)
RPH111 No OEM Part # Available HEX NUT (#8-32) - 50/PKG SCK062 01-104381S / 01-109494S LCD KIT 3
RPH817 No OEM Part # Available SCREW (#8-32 x .500" - SEMS) - 25/PKG SCL012 01-100684S (Kit) BUBBLE LEVEL
RPK432 No OEM Part # Available DRAIN TUBE ASSEMBLY KIT (SMALL) SCP050 01-100568S (2000) / PUMP KIT (INVENSYS PUMP)
RPO244 No OEM Part # Available O-RING - 6/PKG 01-101761S (5000)
RPO385 No OEM Part # Available O-RING - 12/PKG 2 2 SCP051 01-110445S (2000) / PUMP KIT (ULKA PUMP)
SCA022 01-104382S / DISPLAY ASSEMBLY 01-110446S (5000)
01-1043815 (Display only) SCS001 01-100028S CASSETTE SEAL
01-1094945 (Display only) SCS023 01-100573S POWER SWITCH
SCB018 01-100724S CONDENSER WASTE BOTTLE WITH LID 2 2 SCS029 01-101649S CASSETTE SEAL
SCB039 01-100570S (Complete Assy.) PROBE BRACKET SCT030 No OEM Part # Available THERMOCOUPLE
SCB043 01-101618S (Complete Assy.) PROBE BRACKET SCT053 01-108816S METERING TUBE (290)
SCC005 01-104303S SOLENOID - COIL SCV004 01-106787S / 74-100294A SAFETY VALVE (70 PSI)
SCC019 No OEM Part # Available CONDENSATION COIL SCV027 No OEM Part # Available SAFETY VALVE (38 PSI)
SCC060 01-101783S RESERVOIR CAP FOOTNOTES: 1. Older styles only 2. Models with RPI's Condenser Waste Bottle Kit (RPI Part #SCK016) 3. Model #01-121101
(Reservoir Cap & Filter) 4. Serial #2101DK0274 and above 5. Model #101102 (Old Type, One Piece Style with S/N's ending with an 'A') 6. Model #01-201103
SCD063 01-106950S LCD DECAL 3 7. Model #01-121101 (New Style, Two Piece Style)
SCF002 01-100207S COMPRESSOR FILTER
SCF006 01-104461S / 10-106462S DOUBLE THERMAL FUSE ASSEMBLY
SCF031 01-101652S AIR COMPRESSOR FILTER
SCF032 01-102119S BIOLOGICAL FILTER 4
SCF034 01-106637S WATER FILTER (IN-LINE)
SCF061 01-109300S FILTER
SCG009 01-100842S FACIA GASKET
SCG010 01-100865S PROBE BRACKET GASKET
SCG044 01-101650S FACIA GASKET
FOOTNOTES: 1. Older styles only 2. Models with RPI's Condenser Waste Bottle Kit (RPI Part #SCK016) 3. Model #01-121101
4. Serial #2101DK0274 and above 5. Model #101102 (Old Type, One Piece Style with S/N's ending with an 'A') 6. Model #01-201103
7. Model #01-121101 (New Style, Two Piece Style)
continued on next page

50 51
11.0 - RPI TOOLS & ACCESSORIES LIST NOTES

00

00
(FOR PARTS TO FIT STATIM 2000 & 5000)

20

50
IM

IM
RPI

AT

AT
ST

ST
Part # OEM Part # Description
RPA369 No OEM Part # Available THREADLOCKER 545
RPB792 No OEM Part # Available LARGE DIA. BRUSH (1-3/4")
RPB793 No OEM Part # Available SMALL DIA. BRUSH (3/8")
RPB794 No OEM Part # Available SCRUB BRUSH
RPB795 No OEM Part # Available HANDLE BRUSH
RPK791 No OEM Part # Available CLEANING KIT
RPP798 No OEM Part # Available CLEANSING PAD - 3/PKG
RPP819 No OEM Part # Available MOUNTING PAD - 10/PKG
RPS287 No OEM Part # Available LIQUID SOAP - 4/PKG
RPS381 No OEM Part # Available HIGH TEMPERATURE SLEEVING
RPS639 No OEM Part # Available SILICONE GASKET MAKER 598
RPS797 No OEM Part # Available SPONGE (4-1/4" x 6")
RPS821 No OEM Part # Available CALIBRATION SOLUTION - 4/PKG
RPT083 No OEM Part # Available CABLE TIE (4" WHITE) - 25/PKG
RPT297 No OEM Part # Available HEX BALLDRIVER (5/64")
RPT364 01-103471S SPANNER NUT
RPT372 No OEM Part # Available CASSETTE SEAL REMOVAL TOOL
RPT373 No OEM Part # Available SILICONE TUBING
RPT378 No OEM Part # Available 1/8 OD TEFLON TUBING
RPT380 No OEM Part # Available EXHAUST TUBING
RPT394 No OEM Part # Available INSULATED FERRULE
RPT480 No OEM Part # Available CABLE TIE (HIGH TEMP) - 25/PKG
RPT482 No OEM Part # Available TERMINAL CRIMP TOOL
RPT501 No OEM Part # Available DOUBLE ENDED SPANNER WRENCH
RPT502 No OEM Part # Available FUSE PULLER
RPT647 No OEM Part # Available TUBING (CLEAR 1/2" - ID PVC)
RPT820 01-103139S WATER QUALITY TESTER
RPT836 No OEM Part # Available HEX BALLDRIVER (9/64")
RPW382 No OEM Part # Available HOOK UP WIRE (BLACK)
RPW805 No OEM Part # Available HOOK UP WIRE (WHITE)
RPW806 No OEM Part # Available HOOK UP WIRE (GRN/YEL)
RPW813 No OEM Part # Available HOOK UP WIRE (BLACK)
RPW814 No OEM Part # Available HOOK UP WIRE (WHITE)
RPW815 No OEM Part # Available HOOK UP WIRE (GRN/YEL)
SCA046 8OZPLUS (Stat-Dri Plus) / SCI-DRY (16 oz. REFILL)
32OZPLUS (Stat-Dri Plus)
SCA047 No OEM Part # Available SCI-DRY (16 oz. REFILL - CASE)
SCA054 2OZPLUS (Stat-Dri Plus) / SCI-DRY (2 oz.)
8OZPLUST (Stat-Dri PlusT)
SCA055 No OEM Part # Available SCI-DRY (2 oz. CASE)
SCK024 01-100713S PUMP TESTER BOTTLE KIT
SCK028 No OEM Part # Available FIELD SERVICE SMART KIT
SCS021 No OEM Part # Available START SWITCH
SCT026 No OEM Part # Available EXTENSION TEST CABLE
SCT052 01-106776S ALIGNMENT TOOL
(Probe Alignment Jig)

52 53
Parts mentioned in this booklet are manufactured by Replacement Parts Industries, Inc. to fit SciCan STATIM Cassette Autoclave Models 2000 and 5000.
All product names used in this document are trademarks or registered trademarks of their respective holders. All rights reserved RPI 2013. SCK028INS REV A 09/15

Potrebbero piacerti anche