1) Please contact with Chison Service Dept get the newest upgrade package, including
two files: fpga.ZIP and upgrade.ZIP
2) Unzip them, you will get two folders as below; upgrade folder and fpga folder, copy
them to the root of U disk.
3) Pls confirm when you open the folders, there should be subfile in the folder;
6) Click the SW upgrade item firstly and confirm YES , the system will upgrade the
software automatically.
Note: The whole procedure is about 5minutes.
7) The system will restart when you confirm OK after upgrade successed.
Note: The upgrade procedure is not finshed after you upgrade the software, you
have to upgrade the FPGA to work with the software.
8) Please go to System interface again and click the item FW upgrade to upgrade,
then the system will upgrade automatically after you confirm YES.
Note: The whole procedure is about 16 minutes.
9) The unit will turn off when you confirm OK after upgrade successed.
10) Please check the system information after the whole upgrade procedure.