May 2010
Tables..................................................................................................................................... vi
1. Introduction ...................................................................................................................... 1
General Description .................................................................................................................. 1
Reference Documents ............................................................................................................... 1
Transmitter Identification ......................................................................................................... 3
Standard Specifications ............................................................................................................. 5
Product Safety Specifications ................................................................................................... 10
ATEX and IECEx Warnings .............................................................................................. 13
ATEX Compliance Documents .......................................................................................... 13
IECEx Compliance Documents ......................................................................................... 14
2. Installation ...................................................................................................................... 15
Transmitter Mounting ............................................................................................................ 15
Process Mounting ............................................................................................................... 15
Manifold Mounted Transmitter ......................................................................................... 17
Transmitter Mounted on a Coplanar Manifold ............................................................. 18
Pipe or Surface Mounting .................................................................................................. 18
Standard Mounting Bracket .......................................................................................... 18
Universal Mounting Bracket .......................................................................................... 20
Venting and Draining ............................................................................................................. 23
Traditional Structure .......................................................................................................... 23
LP1 Low Profile Structure .................................................................................................. 23
LP2 Low Profile Structure .................................................................................................. 24
Installation of Flow Measurement Piping ................................................................................ 24
Filling System with Seal Liquid ............................................................................................... 27
Positioning the Housing .......................................................................................................... 28
Positioning the Display ........................................................................................................... 28
Setting the Write Protect Jumper ............................................................................................ 29
Cover Locks ............................................................................................................................ 29
Wiring .................................................................................................................................... 29
Accessing Transmitter Field Terminals ............................................................................... 30
Wiring the Transmitter to a Control Loop ......................................................................... 30
Multidrop Communication ................................................................................................ 33
Connecting the Transmitter to an I/A Series System .......................................................... 34
Putting a Differential Pressure Transmitter Into Operation ..................................................... 35
Taking a Differential Pressure Transmitter Out of Operation ................................................. 35
iii
MI IDP10-T May 2010 Contents
4. Calibration ...................................................................................................................... 41
General Calibration Notes ...................................................................................................... 41
Calibration Setup .................................................................................................................... 43
Setup of Electronic Equipment ........................................................................................... 44
Field Calibration Setup ...................................................................................................... 44
Bench Calibration Setup .................................................................................................... 45
Calibration Using a PC20 ....................................................................................................... 46
Calibration Using a PC50 ....................................................................................................... 46
Calibration Using a HART Communicator ............................................................................ 46
Calibration Using the Optional Local Display ........................................................................ 46
Zero Adjustment Using External Zero Button .................................................................... 49
Error Messages ...................................................................................................................... 51
5. Configuration.................................................................................................................. 53
Configurable Parameters ......................................................................................................... 53
Configuration Using a PC20 ................................................................................................... 54
Configuration Using a PC50 ................................................................................................... 54
Configuration Using a HART Communicator ........................................................................ 54
Configuration Using the Optional Local Display .................................................................... 54
Character Lists ........................................................................................................................ 64
Error Messages ........................................................................................................................ 65
6. Maintenance.................................................................................................................... 67
Error Messages ........................................................................................................................ 67
Parts Replacement ................................................................................................................... 67
Replacing the Terminal Block Assembly ............................................................................. 67
Replacing the Electronics Module Assembly ....................................................................... 68
Removing and Reinstalling a Housing Assembly ................................................................ 69
Adding the Optional Display ............................................................................................. 70
Replacing the Sensor Assembly ........................................................................................... 70
Rotating Process Covers for Venting ....................................................................................... 72
Index .................................................................................................................................... 75
iv
Figures
1 Transmitter Identification ............................................................................................ 3
2 Top Level Structure Diagram ....................................................................................... 4
3 Minimum Allowable Absolute Pressure vs. Process Temperature
with Fluorinert Fill Fluid ............................................................................................. 7
4 Typical Mounting of an IDP10 Transmitter Supported by Process Piping ................... 16
5 Typical Mounting of an IDP10 Transmitter Supported by a Bypass Manifold ............ 17
6 Typical Mounting of IDP10 Transmitter on Coplanar Manifold ................................. 18
7 Pipe or Surface Mounted Transmitter Using a Standard Bracket .................................. 19
8 Examples of Mounting With a Standard Bracket ......................................................... 19
9 Details of a Universal Bracket ....................................................................................... 20
10 Mounting a Transmitter with Traditional Structure Using a Universal Bracket ........... 21
11 Vertical Pipe Mounting a Transmitter with LP2 Structure Using a Universal Bracket .. 21
12 Horizontal Mounting a Transmitter with LP2 Structure Using a Universal Bracket ..... 22
13 Vertical Mounting - Cavity Draining ........................................................................... 23
14 Vertical Mounting - Cavity Venting ............................................................................. 23
15 Horizontal Mounting - Cavity Venting ........................................................................ 23
16 Vertical Mounting - Cavity Venting ............................................................................. 24
17 Horizontal Mounting - Cavity Venting and Draining .................................................. 24
18 Cavity Venting and Draining ....................................................................................... 24
19 Example of Horizontal Process Line Installation .......................................................... 26
20 Example of Vertical Process Line Installation ............................................................... 27
21 Housing Screw or Clip Location .................................................................................. 28
22 Cover Lock Location .................................................................................................... 29
23 Accessing Field Terminals ............................................................................................ 30
24 Identification of Field Terminals .................................................................................. 30
25 Supply Voltage and Loop Load .................................................................................... 31
26 Loop Wiring Transmitters ........................................................................................... 33
27 Wiring Several Transmitters to a Common Power Supply ........................................... 33
28 Typical Multidrop Network ......................................................................................... 34
29 Local Display Module .................................................................................................. 37
30 Top Level Structure Diagram ....................................................................................... 38
31 Display Test Segment Patterns ..................................................................................... 40
32 4 to 20 mA Output Calibration Setup of Electronic Equipment .................................. 44
33 Field Calibration Setup ................................................................................................ 45
34 Bench Calibration Setup .............................................................................................. 46
35 Calibration Structure Diagram ..................................................................................... 48
36 Calibration Structure Diagram (Continued) ................................................................ 49
37 Configuration Structure Diagram ................................................................................ 57
38 Configuration Structure Diagram (Continued) ............................................................ 58
39 Configuration Structure Diagram (Continued) ............................................................ 59
40 Replacing the Electronics Module Assembly and Display ............................................. 69
41 Replacing the Sensor Assembly ..................................................................................... 71
42 Replacing the Sensor Assembly (pvdf Inserts) ............................................................... 72
43 Sensor Cavity Venting and Draining ............................................................................ 72
v
Tables
1 Reference Documents .................................................................................................. 1
2 Minimum Loop Load and Supply Voltage Requirements ............................................. 9
3 Electrical Safety Specifications ...................................................................................... 12
4 Operation Error Messages ............................................................................................ 40
5 Calibration Menu ........................................................................................................ 47
6 Calibration Error Messages .......................................................................................... 51
7 IDP10-T Configurable Parameters ............................................................................... 53
8 Configuration Menu .................................................................................................... 55
9 Alphanumeric Character List ........................................................................................ 64
10 Numeric Character List ................................................................................................ 64
11 Configuration Error Messages ...................................................................................... 65
vi
1. Introduction
General Description
The IDP10-T intelligent differential pressure transmitters measure the difference between two
pressures applied to opposite sides of a silicon strain gauge microsensor within the sensor
assembly. This microsensor converts differential pressure to a change in resistance. The resistance
change is then converted to a 4 to 20 mA or digital signal proportional to differential pressure or
to the square root of differential pressure. This measurement signal is transmitted to remote
receivers over the same two wires that supply power to the transmitter electronics. These wires
also carry two-way data signals between the transmitter and remote communication devices.
The transmitter allows direct analog connection to common receivers while still providing full
Intelligent Transmitter Digital Communications using a HART Communicator.
The transmitter is often used for measuring fluid flow rates across a primary device such as an
orifice plate, but can also be used for other types of differential pressure measurements such as
liquid level, interface level, or density measurements. The IDP10 can also be supplied with direct
connected or remote pressure seals to isolate the measuring element from corrosive or viscous
fluids.
For more detailed information on the principle of operation of the transmitter, refer to document
TI 037-096, available from Invensys.
Reference Documents
Table 1. Reference Documents
Document Description
Dimensional Prints
DP 020-342 Dimensional Print PSFLT Pressure Seals
DP 020-343 Dimensional Print PSFPS and PSFES Pressure Seals
DP 020-345 Dimensional Print PSFAR Pressure Seals
DP 020-347 Dimensional Print PSTAR Pressure Seals
DP 020-349 Dimensional Print PSISR Pressure Seals
DP 020-351 Dimensional Print PSSCR Pressure Seals
DP 020-353 Dimensional Print PSSCT Pressure Seals
DP 020-354 Dimensional Print PSSSR Pressure Seals
DP 020-355 Dimensional Print PSSST Pressure Seals
DP 020-446 Dimensional Print IDP10, IDP25, and IDP50 Differential Pressure Transmitters
DP 022-335 Dimensional Print Model CO Compact Orifice
Parts Lists
PL 006-172 Parts List Model CO Compact Orifice
1
MI IDP10-T May 2010 1. Introduction
Document Description
PL 009-005 Parts List IDP10 Differential Pressure Transmitter
Instructions
MI 020-328 Instruction Bubble Type Installation for Liquid Level
MI 020-329 Instruction High Accuracy Flow Measurement
MI 020-366 Instruction I/A Series Intelligent Pressure Transmitters Operation, Configuration,
and Calibration Using a HART Communicator
MI 020-369 Instruction Pressure Seals
MI 020-427 Instruction Intrinsic Safety Connection Diagrams and Nonincendive Circuits
MI 020-495 Instruction PC20 Intelligent Transmitter Configurator
MI 020-501 Instruction PC50 Intelligent Field Device Tool (Installation and Parts List)
MI 020-505 Instruction PC50 Intelligent Field Device Tool (Operation Using HART
Protocol)
MI 020-520 Instruction PC50 Intelligent Field Device Tool with Advanced DTM Library
(Operation Using HART Protocol)
MI 022-138 Instruction Bypass Manifolds - Installation and Maintenance
MI 022-335 Instruction Model CO Compact Orifice
Technical Information
TI 1-50a Technical Information Liquid Density Measurement
TI 001-051 Technical Information Liquid Interface Measurement
TI 001-052 Technical Information Liquid Level Measurement
TI 37-75b Technical Information Transmitter Material Selection Guide
TI 037-097 Technical Information Process Sealing of I/A Series Pressure Transmitters for use
in Class 1, Zone 0, 1, and 2 Hazardous Locations
2
1. Introduction MI IDP10-T May 2010
Transmitter Identification
See Figure 1 for transmitter data plate contents. For a complete explanation of the Model
Number code, see the parts list. The firmware version is identified on the top line of the display
when VIEW DB (View Database) is selected in the top level structure (see Figure 2).
3
MI IDP10-T May 2010 1. Introduction
E E
(measurement M1 or M2) (measurement M2 or M1)
N
N
E
RERANGE LOCAL MODE, GO TO RERANGE MENU
N
E
CONFIG LOCAL MODE, GO TO CONFIGURATION MENU
N
E
CALIB OFF-LINE, GO TO CALIBRATION MENU
N
E
VIEW DB ON-LINE MODE
N N STEP THROUGH DATABASE DISPLAY
E
E
TST DSP ON-LINE MODE
N
N STEP THROUGH DISPLAY TEST PATTERN
E
E
CANCEL EXIT MODE SELECT MENU, RETURN TO ON-LINE MODE
N
N = NEXT BUTTON E = ENTER BUTTON
4
1. Introduction MI IDP10-T May 2010
Standard Specifications
Operative Limits
Influence Operative Limits
Sensor Body Temperature(a)
Silicone Fill Fluid -46 and +121C (-50 and +250F)
Fluorinert Fill Fluid -29 and +121C (-20 and +250F)
pvdf Inserts -7 and +82C (20 and 180F)
Electronics Temperature -40 and +85C (-40 and +185F)
With LCD Display -40 and +85C (-40 and +185F)(b)
Relative Humidity 0 and 100%
Supply Voltage 11.5 and 42 V dc
Output Load(c) 0 and 1450 ohms
Mounting Position No Limit
Vibration 6.3 mm (0.25 in) double amplitude from 5 to 15 Hz with aluminum
housing and from 5 to 9 Hz with 316 ss housing.
0 to 30 m/s (0 to 3 g) from 15 to 500 Hz with aluminum housing
and
0 to 10 m/s (0 to 1 g) from 9 to 500 Hz with 316 ss housing.
(a) Refer to MI 020-369 for temperature limits with pressure seals.
(b) Display updates are slowed and readability decreased at temperatures below -20C (-4F).
(c) 250 minimum load is required for communication with a HART Communicator.
5
MI IDP10-T May 2010 1. Introduction
NOTE
Static pressure zero shift for all calibrated spans can be eliminated by readjusting the
zero output at nominal operating static pressure.
! CAUTION
1. Exceeding the maximum overrange pressure can cause damage to the transmitter
degrading its performance.
2. The transmitter could be nonfunctional after application of the proof pressure.
6
1. Introduction MI IDP10-T May 2010
Temperature C
-80 0 30 60 90 120
140
120
100
(operating area above curve)
80
60
40
20
Mounting Position
The transmitter can be mounted in any orientation. It can be supported by the process
piping. It can also be mounted directly to a vertical or horizontal pipe or surface mounted
using an optional mounting bracket. The housing can be rotated up to one full turn to
any desired position for access to adjustments, display, or conduit connections. See
Positioning the Housing on page 28. The display (if present) can also be rotated in the
housing to any of four different positions at 90 increments. See Positioning the
Display on page 28.
NOTE
Position effect zero shift for all calibrated spans can be eliminated by readjusting
zero output after installation.
Approximate Mass
Without Process Connectors 3.5 kg (7.8 lb)
With Process Connectors 4.2 kg (9.2 lb)
With Optional 316 ss Housing Add 1.1 kg (2.4 lb)
With Pressure Seals Varies with seal used
Process Connections
IDP10 transmitters are connected to the process via a 1/4 NPT thread or any one of a
number of optional process connectors.
7
MI IDP10-T May 2010 1. Introduction
Electrical Connections
Field wires enter through 1/2 NPT, PG 13.5, or M20 threaded entrances on either side of
the electronics housing. Leads terminate under screw terminals and washers on the
terminal block in the field terminal compartment. To maintain RFI/EMI, environmental,
and explosionproof ratings, unused conduit connection must be plugged with metal plug
(provided), inserted to five full threads for 1/2 NPT connections; seven full threads for
M20 and PG 13.5 connections.
Adjustable Damping
The transmitter response time is normally 1.0 second or the electronically adjustable
setting of 0.00 (none), 0.25, 0.50, 1, 2, 4, 8, 16, or 32 seconds, whichever is greater, for a
90% recovery from an 80% input step as defined in ANSI/ISA S51.1.
Output Signal
4 to 20 mA dc linear or 4 to 20 mA dc square root; software selectable. The output is
remotely configurable from the HART Communicator and locally configurable with the
pushbuttons on the display.
NOTE
Only 4 to 20 mA linear output on absolute pressure, gauge pressure, and flange
level transmitters.
Power-up Time
Less than 2.0 seconds for output to reach the first valid measurement, then at the
electronic damping rate to reach the final measured variable value.
8
1. Introduction MI IDP10-T May 2010
Supply Voltage
Power supply must be capable of providing 22 mA when the transmitter is configured for
4 to 20 mA output. Ripple of up to 2 V pp (50/60/100/120 Hz) is tolerable, but
instantaneous voltage must remain within specified range.
The supply voltage and loop load must be within specified limits. This is explained in
detail in Wiring on page 29. A summary of the minimum requirements is listed in
Table 2.
Table 2. Minimum Loop Load and Supply Voltage Requirements
HART No HART
Communication Communication
Minimum Resistance 250 0
Minimum Supply Voltage 17 V 11.5 V
Test Points
The two lower banana plug receptacles (designated CAL) can be used to check transmitter
output when configured for 4 to 20 mA. Measurements should be 100-500 mV dc for
0-100% transmitter output.
Remote Communications
The transmitter communicates bidirectionally over the 2-wire field wiring to a HART
Communicator. The information that can be continuously displayed is:
Process Measurement (expressed in one or two types of units)
Transmitter Temperature (sensor and electronics)
mA Output (equivalent)
The information that can be remotely displayed and reconfigured includes:
Output in Percent Flow (square root) or Pressure Units (linear). Percent Display in
Linear mode on local display is also supported.
Zero and Span, including reranging
Zero Elevation or Suppression
9
MI IDP10-T May 2010 1. Introduction
Communications Format
Communication is based upon the FSK (Frequency Shift Keying) technique. The
frequencies are superimposed on the transmitter power/signal leads.
4 to 20 mA Output
The transmitter sends its differential pressure measurement to the loop as a continuous
4 to 20 mA dc signal. It also communicates digitally with the HART Communicator at
distances up to 3000 m (10 000 ft). Communication between the remote configurator
and the transmitter does not disturb the 4 to 20 mA output signal. Other specifications
are:
10
1. Introduction MI IDP10-T May 2010
! WARNING
To maintain IEC IP66 and NEMA Type 4X protection, the unused conduit opening
must be plugged with the metal plug provided. Use a suitable thread sealant on both
conduit connections. In addition, the threaded housing covers must be installed.
Turn covers to seat the O-ring into the housing and then continue to hand tighten
until the cover contacts the housing metal-to-metal.
NOTE
1. These transmitters have been designed to meet the electrical safety description
listed in Table 3. For detailed information or status of testing laboratory
approvals/certifications, contact Invensys.
2. Wiring restrictions required to maintain electrical certification of the transmitter
are provided in Wiring on page 29.
11
MI IDP10-T May 2010 1. Introduction
12
1. Introduction MI IDP10-T May 2010
13
MI IDP10-T May 2010 1. Introduction
EN 50281-1-1: 1998
EN 60079-15: 2004
14
2. Installation
! CAUTION
To avoid damage to the transmitter sensor, do not use any impact devices, such as an
impact wrench or stamping device, on the transmitter.
NOTE
1. The transmitter should be mounted so that any moisture condensing or draining
into the field wiring compartment can exit through one of the two threaded
conduit connections.
2. Use a suitable thread sealant on all connections.
Transmitter Mounting
The IDP Series differential pressure transmitter can be supported by the process piping or
mounted to a vertical or horizontal pipe or surface using the optional mounting bracket. See
figures below. For dimensional information, refer to DP 020-446.
NOTE
1. If the transmitter is not installed in the vertical position, readjust the zero output
to eliminate the position zero effect.
2. When structure codes 78/79 are used (pvdf inserts) with the IDP10 transmitters,
the process connection must be made directly to the pvdf inserts in the high and
low side process covers.
Process Mounting
With process mounting, the transmitter mounted to and supported by the process piping.
15
MI IDP10-T May 2010 2. Installation
SEE
SEE NOTE
SEE NOTE
NOTE
16
2. Installation MI IDP10-T May 2010
17
MI IDP10-T May 2010 2. Installation
ADAPTER PLATE
AND GASKETS
18
2. Installation MI IDP10-T May 2010
APPROXIMATELY 3 IN
CLEARANCE REQUIRED
FOR ACCESS TO MOUNTING
BOLTS AND VENT SCREW.
VERTICAL DN 50 OR 2 IN PIPE
BRACKET SHOWN. ROTATE U-BOLT 90
FOR MOUNTING TO HORIZONTAL
PIPE
19
MI IDP10-T May 2010 2. Installation
U-BOLT ASSEMBLY
FOR DN 50 OR 2 in PIPE
HOLES FOR
U-BOLT AND
SURFACE BOLTS TO MOUNT
MOUNTING TRANSMITTER
ON FOUR TO BRACKET
SIDES OF THIS
BRACKET LEG,
BOLTS TO MOUNT
TRANSMITTER TO
BRACKET
20
2. Installation MI IDP10-T May 2010
Figure 10. Mounting a Transmitter with Traditional Structure Using a Universal Bracket
Figure 11. Vertical Pipe Mounting a Transmitter with LP2 Structure Using a Universal Bracket
21
MI IDP10-T May 2010 2. Installation
Figure 12. Horizontal Mounting a Transmitter with LP2 Structure Using a Universal Bracket
22
2. Installation MI IDP10-T May 2010
PROCESS
DRAIN SCREW
COVER
OPTIONAL
SIDE VENT
SHOWN
PLUG
VENT SCREW
23
MI IDP10-T May 2010 2. Installation
process covers. If the transmitter is connected with a length of impulse piping, such piping should
slope up to the transmitter for gas applications and down for liquid applications.
VENT
SCREW
IN-LINE
PROCESS
CONNECTION
VENT &
DRAIN
SCREWS
24
2. Installation MI IDP10-T May 2010
The transmitters are shown below the level of the pressure connections at the pipe (usual
arrangement, except for gas flow without a seal liquid), and with filling tees in the lines to the
transmitter (for a seal liquid).
If the process fluid being measured must not come in contact with the transmitter, the transmitter
lines must be filled with a suitable seal liquid (see procedure in next section). In such a case, the
transmitter must be mounted below the level of the pressure connections at the pipe. With steam
flow, the lines are filled with water to protect the transmitter from the hot steam. The seal liquid
(or water) is added to the lines through the filling tees. To prevent unequal heads on the
transmitter, the tees must be at the same elevation and the transmitter must be mounted vertically
(as shown). If a seal liquid is not required, elbows can be used in place of the tees.
Tighten drain plugs and optional vent screws to 20 Nm (15 lbft). Tighten the four process
connector bolts to a torque of 61 Nm (45 lbft).
Note that the low and high pressure sides of the transmitter are identified by an L-H marking on
the side of the sensor above the warning label.
With medium viscosity seal liquids and/or long transmitter lines, larger valve sizes should be used.
NOTE
1. With a horizontal line, pressure connections at the pipe should be at the side of
the line. However, with gas flow without a seal liquid, connections should be at
top of line.
2. With a vertical line, flow should be upwards.
3. For liquid or steam flow, the transmitter should be mounted lower than the
pressure connections at the pipe.
4. For gas flow without a seal liquid, the transmitter should be mounted above the
pressure connections at the pipe; for gas flow with a seal liquid, the transmitter
should be mounted below the pressure connections.
5. Invensys recommends the use of snubbers in installations prone to high levels of
fluid pulsations.
25
MI IDP10-T May 2010 2. Installation
HIGH
PRESSURE
SIDE
FILLING TEES
PIPE OR TUBING
OPTIONAL
3-VALVE
MANIFOLD
26
2. Installation MI IDP10-T May 2010
PROCESS DIRECTION OF
SHUT OFF VALVES PRESSURE FLOW
LOW TRANSMITTER
FILLING TEES PRESSURE
SIDE
HIGH
PRESSURE
SIDE
PIPE OR TUBING
OPTIONAL
3-VALVE
MANIFOLD
27
MI IDP10-T May 2010 2. Installation
! CAUTION
To prevent loss of seal liquid and contamination of process fluid, never open both
process shutoff valves and manifold shutoff valves if the bypass valve is open.
! WARNING
If the electronics housing is removed for maintenance, it must be hand tightened to
the bottom of the threads, but not over-tightened upon reassembly. See Removing
and Reinstalling a Housing Assembly on page 69.
RETENTION CLIP
HOUSING
ANTI-ROTATION SCREW
OR RETENTION CLIP
CUP
CLIP
! CAUTION
Do not turn the display more than 180 in any direction. Doing so could damage its
connecting cable.
28
2. Installation MI IDP10-T May 2010
Cover Locks
Electronic housing cover locks, shown in Figure 22, are provided as standard with certain agency
certifications and as part of the Custody Transfer Lock and Seal option. To lock the covers,
unscrew the locking pin until approximately 6 mm (0.25 in) shows, lining up the hole in the pin
with the hole in the housing. Insert the seal wire through the two holes, slide the seal onto the
wire ends and crimp the seal
Wiring
The installation and wiring of your transmitter must conform to local code requirements.
! WARNING
ATEX requires that when the equipment is intended to be used in an explosive
atmosphere caused by the presence of combustible dust, cable entry devices and
blanking elements shall provide a degree of ingress protection of at least IP6X. They
shall be suitable for the conditions of use and correctly installed.
NOTE
Invensys recommends the use of transient/surge protection in installations prone to
high levels of electrical transients and surges.
29
MI IDP10-T May 2010 2. Installation
EXTERNAL EARTH
(GROUND)
(+) (+)
HHT
TRANSMITTER BANANA PLUG RECEPTACLES FOR
SIGNAL CALIBRATION CONNECTIONS. TO READ
CONNECTIONS TRANSMITTER OUTPUT, ATTACH METER
() (-) LEADS HERE (100 TO 500 MV REPRE-
CAL SENTING 4 TO 20 mA CURRENT).
OPTIONAL SHORTING BAR (SB-11) TO
REDUCE MINIMUM VOLTAGE FROM
11.5 V dc TO 11 V dc ALSO PLUGS IN HERE.
30
2. Installation MI IDP10-T May 2010
Any combination of supply voltage and loop load resistance in the shaded area can be used. To
determine the loop load resistance (transmitter output load), add the series resistance of each
component in the loop, excluding the transmitter. The power supply must be capable of
supplying 22 mA of loop current.
1450
1400 TYPICAL SUPPLY VOLTAGE
AND LOAD LIMITS
1300
900 NOTES:
1. THE MINIMUM LOAD FOR THE HART COMMUNICATOR
IS 250 .
OUTPUT LOAD,
800
2. THE TRANSMITTER CAN FUNCTION WITH AN OUTPUT L
LESS THAN THE MINIMUM, PROVIDED THAT A REMOTE
700
CONFIGURATOR IS NOT CONNECTED TO IT. CONNECTIN
A REMOTE CONFIGURATOR WHILE OPERATING IN THIS
600 COULD CAUSE OUTPUT DISTURBANCES AND/OR COMM
CATION PROBLEMS.
500
MINIMUM LOAD
400 (SEE NOTE) OPERATING AREA
300
200
100
0
0 10 20 30 40
11.5 42
SUPPLY VOLTAGE, V dc
Examples:
1. For a loop load resistance of 880 , the supply voltage can be any value from 30 to
42 V dc.
2. For a supply voltage of 24 V dc, the loop load resistance can be any value from 250 to
594 (zero to 594 without a HART Communicator connected to the transmitter).
To wire one or more transmitters to a power supply, proceed with the following steps.
1. Remove the cover from the transmitter field terminals compartment.
2. Run signal wires (0.50 mm2 or 20 AWG, typical) through one of the transmitter
conduit connections. Use twisted single pair to protect the 4 to 20 mA output and/or
remote communications from electrical noise. Maximum recommended length for
signal wires is:
31
MI IDP10-T May 2010 2. Installation
3050 m (10,000 ft) using single pair cable and adhering to requirements of HART
physical layer implementation defined in HART Document HCF_SPEC-53. Use
CN=1 when calculating max. lengths.
1525 m (5000 ft) in a multidrop (15 devices maximum) mode.
Screened (shielded) cable could be required in some locations.
NOTE
Do not run transmitter wires in same conduit as mains (ac power) wires.
3. If shielded cable is used, earth (ground) the shield at the power supply only. Do not
ground the shield at the transmitter.
4. Plug unused conduit connection with the 1/2 NPT, PG 13.5 or M20 metal plug
provided (or equivalent). To maintain specified explosionproof and dust-
ignitionproof protection, plug must engage a minimum of five full threads for 1/2
NPT connections; seven full threads for M20 and PG 13.5 connections.
5. Connect an earth (ground) wire to the earth terminal in accordance with local
practice.
! CAUTION
If the signal circuit must be earthed (grounded), it is preferable to do so at the
negative terminal of the dc power supply. To avoid errors resulting from earth loops
or the possibility of short-circuiting groups of instruments in a loop, there should be
only one earth in a loop.
6. Connect the power supply and receiver loop wires to the + and terminal
connections.
7. Connect receivers (such as controllers, recorders, indicators) in series with power
supply and transmitter as shown in Figure 26.
8. Reinstall the cover onto the housing by rotating it clockwise to seat the O-ring into
the housing and then continue to hand tighten until the cover contacts the housing
metal-to-metal. If cover locks are present, lock the cover per the procedure described
in Cover Locks on page 29.
9. If wiring additional transmitters to the same power supply, repeat Steps 1 through 8
for each additional transmitter. The setup with multiple transmitters connected to a
single power supply is shown in Figure 27.
10. A HART Communicator or PC-Based Configurator can be connected in the loop
between the transmitter and the power supply as shown in Figures 26 and 27. Note
that a minimum of 250 must separate the power supply from the HART
Communicator or PC-Based Configurator.
32
2. Installation MI IDP10-T May 2010
(a)
CONDUIT INDICATOR
+
+
POWER
PLUG SUPPLY
UNUSED _ _
CONDUIT + _
CONNECTION
CONTROLLER
HART COMMUNICATOR(b) OR RECORDER
(a) RUN CONDUIT DOWN TO AVOID MOISTURE BUILDUP IN TERMINALS COMPARTMENT.
(b) THERE MUST BE AT LEAST 250 TOTAL RESISTANCE BETWEEN THE HART COMMUNICATOR
OR PC-BASED CONFIGURATOR AND THE POWER SUPPLY.
POWER
SUPPLY
250
min.(a) 250 250
min.(a) min.(a)
+ + +
Multidrop Communication
Multidropping refers to the connection of several transmitters to a single communications
transmission line. Communications between the host computer and the transmitters takes place
digitally with the analog output of the transmitter deactivated. With the HART communications
33
MI IDP10-T May 2010 2. Installation
protocol, up to 15 transmitters can be connected on a single twisted pair of wires or over leased
telephone lines.
The application of a multidrop installation requires consideration of the update rate necessary
from each transmitter, the combination of transmitter models, and the length of the transmission
line. Multidrop installations are not recommended where Intrinsic Safety is a requirement.
Communication with the transmitters can be accomplished with any HART compatible modem
and a host implementing the HART protocol. Each transmitter is identified by a unique address
(1-15) and responds to the commands defined in the HART protocol.
Figure 28 shows a typical multidrop network. Do not use this figure as an installation diagram.
Contact the HART Communications Foundation, (512) 794-0369, with specific requirements
for multidrop applications.
HOST MODEM
LOAD
POWER
SUPPLY IDP10-T IDP10-T IDP10-T
The HART Communicator can operate, configure, and calibrate IASPT transmitters with HART
communication protocol in the same way as it can in a standard point-to-point installation.
NOTE
IASPT transmitters with HART communication protocol are set to poll address 0
(POLLADR 0) at the factory, allowing them to operate in the standard point-to-
point manner with a 4 to 20 mA output signal. To activate multidrop
communication, the transmitter address must be changed to a number from 1 to
15. Each transmitter must be assigned a unique number on each multidrop
network. This change deactivates the 4 to 20 mA analog output.
34
2. Installation MI IDP10-T May 2010
1. Make sure that both upstream and downstream manifold valves are closed.
2. Make sure that the bypass valve is open.
3. Slowly open the upstream manifold valve.
4. Close the bypass valve.
5. Slowly open the downstream manifold valve.
! WARNING
When venting pressure from the transmitter, wear suitable protective equipment to
prevent possible injury from process material, temperature, or pressure.
35
MI IDP10-T May 2010 2. Installation
36
3. Operation Via Local Display
A local display, as shown in Figure 29, has two lines of information. The upper line is a 5-digit
numeric display (4-digit when a minus sign is needed); the lower line is a 7-digit alphanumeric
display. The display provides local indication of measurement information.
The display can be configured to meet your specific needs. If configured Show 1, M1 is
displayed. If configured Show 2, M2 is displayed. To temporarily view the alternate
measurement, press the Enter button. After showing this measurement for a brief period, the
display reverts to the configured display. If configured Toggle, the display toggles between M1
and M2. When M2 is displayed, an M2 message blinks in the lower right of the display.
The display also provides a means for reranging, calibration, and configuration, viewing the
database, testing the display, and reranging the transmitter via the 2-button keypad. You can
access these operations by means of a multi-level menu system. Entry to the Mode Select menu is
made (from normal operating mode) by pressing the Next button. You can exit this menu, restore
your prior calibration or configuration, and return to the normal operating mode at any time by
going to Cancel and pressing the Enter button.
The top level structure diagram is shown in Figure 30.
34.5
INH2O
NEXT ENTER
NEXT ENTER
PUSHBUTTON PUSHBUTTON
37
MI IDP10-T May 2010 3. Operation Via Local Display
E E
(measurement M1 or M2) (measurement M2 or M1)
N
N
E
RERANGE LOCAL MODE, GO TO RERANGE MENU
N
E
CONFIG LOCAL MODE, GO TO CONFIGURATION MENU
N
E
CALIB OFF-LINE, GO TO CALIBRATION MENU
N
E
VIEW DB ON-LINE MODE
N N STEP THROUGH DATABASE DISPLAY
E
E
TST DSP ON-LINE MODE
N
N STEP THROUGH DISPLAY TEST PATTERN
E
E
CANCEL EXIT MODE SELECT MENU, RETURN TO ON-LINE MODE
N
N = NEXT BUTTON E = ENTER BUTTON
38
3. Operation Via Local Display MI IDP10-T May 2010
Reranging
You can access the Rerange mode in the top level menu (see Figure 30). Entry to the Mode Select
menu is made (from normal operating mode) by pressing the Next button. The display reads
RERANGE. You can then adjust M1 URV and/or M1 LRV in the following two submenus.
NOTE
If M1 MODE is in a square root mode, regardless of engineering units selected,
RERANGE is automatically done in the following default pressure units:
inH2O, if M2 MODE is a type of square root.
M2 EGU units, if M2 MODE is linear.
The bottom line of the display indicates the default units during RERANGE.
Following RERANGE, the display automatically switches back to the configured
engineering units.
M1 URV:
To edit the upper range value, press Enter at the prompt M1 URV. Use the procedure
Entering Numerical Values on page 38 section to edit this parameter.
M1 LRV:
Similar to M1URV immediately above.
NOTE
M1 LRV is bypassed if M1 MODE is configured as square root since M1 LRV must
be zero.
39
MI IDP10-T May 2010 3. Operation Via Local Display
ALL SEGMENTS ON
Error Messages
Table 4. Operation Error Messages
Error
Parameter Condition Tested Message Action
Normal Write Protection WR PROT Displays periodically to notify user
Operation Enabled that unit is in Write Protect.
Any non-On-line OFFLINE Notifies user of a non-On-line
Condition condition.
Startup Database OK or INITERR User should perform SET GDB
corrupted procedure. See SET GDB: on
page 63.
40
4. Calibration
NOTE
1. For best results in applications where high accuracy is required, rezero the
transmitter output once it has stabilized at the final operating temperature.
2. Zero shifts resulting from position effects and/or static pressure effects can be
eliminated by rezeroing the transmitter output.
3. When checking the zero reading of a transmitter operating in the square root
mode, return the output to the linear mode. This eliminates an apparent
instability in the output signal. Return the transmitter output to the square root
mode after the zero check is complete.
4. After calibrating transmitters operating with a 4 to 20 mA (or 1 to 5 V dc)
output signal, check the underrange and overrange output values to ensure that
they extend beyond 4 and 20 mA (or 1 and 5 V dc) respectively.
41
MI IDP10-T May 2010 4. Calibration
42
4. Calibration MI IDP10-T May 2010
When the transmitter is zeroed to compensate for installed position effect, the
transmitter can have either LRV differential pressure applied (CAL LRV) or zero
differential pressure applied (CAL AT0). If using a zero-based range, either method
produces the same result. However, if the range is not zero-based, it is advantageous to
have both methods available.
For example, consider a differential pressure transmitter having a range of 50 to 100
psig. If it is not feasible to vent the transmitter to atmosphere for zeroing (or to bypass
the high and low sides for zeroing), it can be zeroed while the LRV differential
pressure of 50 psi is applied by using the CAL LRV function. On the other hand, if
the transmitter has been installed but there is no pressure in the process line yet (or
the high and low sides can be connected by a bypass valve), it can be zeroed while
open to atmosphere (or bypassed) by using the CAL AT0 function.
a. Zeroing with LRV Pressure Applied (CAL LRV):
Before using this zeroing function, apply a differential pressure to the transmitter
equal to the value of LRV stored in the transmitter database.
When you zero the transmitter, the internal digital value of differential pressure is
trimmed to be equal to the value of LRV stored in the database and the mA
output set to 4 mA.
If zeroing is done when the applied differential pressure is different from the LRV
value in the database, the internal digital value of differential pressure is biased by
the difference in the values but the output is still set at 4 mA.
The CAL LRV and CAL URV function should be used when calibrating a
transmitter for a specific range with known input differential pressures applied for
the LRV and URV.
b. Zeroing with Zero Pressure Applied (CAL AT0):
Make sure that the applied differential pressure is at zero. This means venting the
transmitter to atmosphere or opening a bypass valve to connect high and low
sides.
When you zero the transmitter, the internal digital value of the differential pressure is
trimmed to be equal to zero and the mA output set to an appropriate value such that
the mA output is a nominal 4 mA when the LRV pressure is applied later.
Calibration Setup
The following sections show setups for field or bench calibration. Use test equipment that is at
least three times as accurate as the desired accuracy of the transmitter.
NOTE
It is not necessary to set up calibration equipment to rerange the transmitter to a
different range. The transmitter can be accurately reranged by simply changing the
lower range value and the upper range value, which are stored in the transmitter
database.
43
MI IDP10-T May 2010 4. Calibration
(+)
() () (+) () (+)
! CAUTION
With liquid service, drain both sides of transmitter to avoid calibration errors.
2. If a calibration screw is being used, remove the vent screw and replace it with the
calibration screw. Connect the pressure source to the calibration screw using
6 x 1 mm or 0.250 inch tubing.
If a calibration screw is not being used, remove the entire vent screw assembly or drain
44
4. Calibration MI IDP10-T May 2010
plug (as applicable) from the high pressure side of the transmitter. Connect
calibration tubing using a suitable thread sealant.
3. Close the bypass valve opened in Step 1.
4. Complete the setup shown in Figure 33.
NOTE
For vacuum applications, connect the calibrating pressure source to the low pressure
side of the transmitter.
5. If calibrating the output signal, also connect equipment as shown in Figure 32.
If calibrating the output signal, also connect equipment as shown in Figure 32.
45
MI IDP10-T May 2010 4. Calibration
SHUTOFF VALVES
CALIBRATING
PRESSURE
SOURCE
BLEEDER VALVES
(NEEDLE TYPE)
46
4. Calibration MI IDP10-T May 2010
NOTE
It is not necessary to use the ADJ4mA or ADJ20mA menu selections (commonly
known as mA Trim) unless there is a plant requirement to make the 4 and 20 mA
output values exactly match readings on certain plant calibration equipment and
the calibration operations done result in a small but unacceptable difference
between the transmitter mA output and the test equipment mA readout values.
Proceed to calibrate your transmitter by using the Next key to select your item and the Enter key
to specify your selection per Figures 35 and 36. At any point in the calibration you can Cancel,
restore your prior calibration and return to the on-line mode or Save your new calibration.
47
MI IDP10-T May 2010 4. Calibration
E E
CAL AT0 AT0 DONE CAL AT0: To set or reset the zero point at zero
N N pressure, apply zero differential pressure to the
E E transmitter and, at display of CAL AT0, press
CAL LRV LRV DONE
Enter. This can be done whether LRV is zero
N N
or not. Completion is indicated by the display
E E AT0 Done.
CAL URV URV DONE
E = ENTER
N N
N = NEXT
CAL LRV: To set or reset 0% of range input,
ADJ 4mA
E apply differential pressure to the transmitter
equal to the lower range value (LRV) in the
E
A 4mA A 4mA transmitter database and, at display of
N N
E CAL LRV, press Enter. Completion is
E indicated by the display LRV Done.
A 4mA A 4mA
N N N E
E
CAL URV: To set or reset 100% of range
A 4mA A 4mA input, apply differential pressure to the
N N E transmitter equal to the upper range value
E (URV) in the transmitter database and, at
A 4mA A 4mA
display of CAL URV, press Enter.
N N E
Completion is indicated by the display URV
Done.
ADJ20mA
E
ADJ4mA: If you configured your transmitter
E operating mode as 4 to 20 mA, you can adjust
A 20mA A 20mA
N N E the 4 mA output by going to ADJ4mA using
E
the Next button and press Enter. This menu
A 20mA A 20mA item is bypassed if you configured your
N N E transmitter for multidrop mode (poll address
N other than zero).
E
A 20mA A 20mA
N N E
CALDATE
(continued on next
figure)
48
4. Calibration MI IDP10-T May 2010
E
CALDATE
N
Display Day Increment Day
E
N
N Display Month Increment Month
E
N
Display Year Increment Year
E
E
CANCEL Discard all changes, return to ONLINE.
N
E
SAVE Save database changes, return to ONLINE.
N
*If character is not the last position on the display line, advances to next character.
**If character is the last position on the display line, advances to next menu item.
NOTE: Commentary about this diagram immediately follows.
Commentary on Figure 36
CALDATE:
This is not a required entry but can be used for recordkeeping or plant maintenance
purposes. To edit the calibration date, go to CALDATE with the Next button and press
Enter. You then can change the day, month, and year. The display shows the last date
with the day flashing. Use the Next button to step through the menu of digits to select
the desired day, then press Enter. Repeat this process for the month and year.
49
MI IDP10-T May 2010 4. Calibration
50
4. Calibration MI IDP10-T May 2010
Error Messages
Table 6. Calibration Error Messages
Error
Parameter Condition Tested Message User Action
Password Password BAD PWD Bad password entered, use another.
Protection
Write Write protection REJECT Displays when user attempts an action that
Protection enabled is write protected.
ZERO Internal offset too BADZERO Check applied pressure, configured
large M1 LRV and configured M1 EOFF.
SPAN Slope too large or too BADSPAN Check applied pressure, configured
small M1 LRV and configured M1 EFAC.
M1 URV M1URV > max URV>FMX Entered pressure is greater than maximum
pressure in EGU rated pressure of transmitter. Check entry.
Verify EGUs.
M1URV < min URV<FMN Entered pressure is less than minimum
pressure in EGU rated pressure of transmitter. Check entry.
Verify EGUs.
M1 URV = M1 LRV LRV=URV Cannot set span to 0. Check entry. Check
M1 LRV.
M1 turndown exceeds BADTDWN Check entry. Check M1 LRV.
limit
URV < 0 with M1 or URV<LRV Square root mode with nonzero LRV is
M2 SqRt not valid. Change LRV to 0.
M1 LRV M1LRV > max LRV>FMX Entered pressure is greater than maximum
pressure in EGU rated pressure of transmitter. Check entry.
Verify EGUs.
M1LRV < min LRV<FMN Entered pressure is less than minimum
pressure in EGU rated pressure of transmitter. Check entry.
Verify EGUs.
M1 URV = M1 LRV LRV=URV Cannot set span to 0. check entry. Check
M1 URV.
M1 turndown exceeds BADTDWN Check entry. Check M1 URV.
limit
51
MI IDP10-T May 2010 4. Calibration
52
5. Configuration
Configurable Parameters
Table 7 lists all the configurable parameters and the factory default for the IDP10-T
Transmitters. The factory default values have been customized if the transmitter was ordered with
optional feature -C2. The table also shows which parameters are configurable with the integral vs.
remote configurators.
Configurable
with
Factory Integ. Remote Application
Parameter Capability Default Indic. Config. Requirement
Descriptors
Tag Number 8 characters max Tag Number No Yes
Descriptor 16 characters max Tag Name No Yes
Message 32 characters max Inst Location No Yes
Input
Calibrated Range LRV to URV in units See (b) below Yes Yes
listed in (a) below when not
specified per S.O.
Output
Measurement #1 Output 4 to 20 mA or Fixed 4 to 20 mA Yes Yes
(PV) Current. Specify Poll
Address (1-15) for Fixed
Current.
Measurement #1 Mode Linear or type of square Linear Yes Yes
root in (d) below
Measurement #1 EGUs If linear, select from Units of Yes Yes
units listed in (a) below; Calibrated Range
If Sq.Rt., select from
units listed in (c) below
Measurement #2 Mode Linear or type of square Linear Yes Yes
(SV) root in (d) below
Measurement #2 EGUs If linear, select from Units of Yes Yes
units listed in (a) below; Calibrated Range
If Sq.Rt., select from
units listed in (c) below
Temp. Sensor Fail Strategy Normal oper. or failsafe Fail-safe Yes
Fail-safe High or Low High Yes Yes
External Zero Enabled or Disabled Enabled Yes Yes
Damping 0 to 32 seconds. None Yes Yes
53
MI IDP10-T May 2010 5. Configuration
Configurable
with
Factory Integ. Remote Application
Parameter Capability Default Indic. Config. Requirement
Poll Address 0 - 15 0 Yes Yes
LCD Indicator (e) Meas #1 EGU or % Lin Meas #1 EGU Yes No
(a) psi, inHg, ftH2O, inH2O, atm, bar, mbar, MPa, kPa, Pa, kg/cm2, g/cm2, mmHg, torr, mmH2O.
(b) Span Code A: 0 to 30 inH2O; Span Code B: 0 to 200 inH2O; Span Code C: 0 to 30 psi;
Span Code D: 0 to 300 psi; Span Code E: 0 to 3000 psi.
(c) gal/s, gal/m, gal/h, gal/d, Mgal/d, ft3/s, ft3/m, ft3/h, ft3/d, Igal/s, Igal/m, Igal/h, Igal/d, l/s, l/m, l/h, Ml/d, m3/s,
m3/m,
m3/h, m3/d, bbl/s, bbl/m, bbl/h, bbl/d, %flow.
(d) Square root with cutoff below 1% of calibrated pressure range or with linear below 4% of calibrated pressure
range.
(e) Measurement #2 can be displayed at any time by pressing the Enter button regardless of the local display
configuration. This reverts to Measurement #1 or % Lin (as configured) when power is cycled off and on.
54
5. Configuration MI IDP10-T May 2010
NOTE
1. You can configure most parameters using the local display. However, for more
complete configuration capability, use a HART Communicator or PC-Based
configurator.
2. During configuration, a single change can affect several parameters. For this
reason, if an entry is entered in error, re-examine the entire database or use the
Cancel feature to restore the transmitter to its starting configuration and begin
again.
55
MI IDP10-T May 2010 5. Configuration
Proceed to configure your transmitter by using the Next button to select your item and the Enter
button to specify your selection per the following three figures. At any point in the configuration
you can Cancel your changes and return to the on-line mode, or Save your changes.
56
5. Configuration MI IDP10-T May 2010
E N N N
POLLADR 0 1 2 15
N E E E E
E N
EX ZERO EXZ DIS EXZ ENA
N E E NOTES:
1. OUTFAIL sets mA output to go
either High or Low under certain
E N failure conditions, such as a sensor
S2 FAIL S2 FATAL S2 NOFTL
failure.
N E E 2. Square root functions should not
be selected on absolute and gauge
pressure or level transmitters.
E N
OUT DIR(a) OUT FWD OUT REV
3. LIN PCT provides percent output
N E E on LCD indicator only (linear mode).
Percent flow in square root is selectable
under MI EGU.
E N
OUTFAIL FAIL LO FAIL HI
N E E
E N E Display N Implement
OFFL MA LAST MA USER MA
Digit Digit
N E E
E N N N
DAMPING NO DAMP DAMP 1/4 DAMP 1/2 DAMP 32
N E E E E
N
E N N N
M1 MODE M1 LIN M1SQ<1C M1SQ<4L M1SQ<nC
N E E E E
N
Display Digit Increment Digit
E
** *
57
MI IDP10-T May 2010 5. Configuration
E N
M1 DISP(a) M1 EGU LIN PCT
E E
N
N
E N N N
M1 EGU gal/s gal/m gal/h %flow
E E E E
N
N
or
N N
inH20 inHg atm
E E E
E N
M1EOFF Display Digit Increment Digit
N E
** *
E N
M1EFAC Display Digit Increment Digit
E
N
** *
E N
M1 URV Display Digit Increment Digit
N E
** *
N
E N
M1 LRV Display Digit Increment Digit
N E
** *
E
M2 MODE Similar to M1 MODE
N
E (a) Linear Mode only.
M2 EGU Similar to M1 EGU
N
DISPLAY
*If character is not the last position on the display line, advances to next character.
(continued on next figure)
**If character is the last position on the display line, advances to next menu item.
58
5. Configuration MI IDP10-T May 2010
E N N
DISPLAY SHOW M1 SHOW M2 TOGGLE
E E E
E
CALDATE
N
Display Day Increment Day
E
N
N Display Month Increment Month
E
N
Display Year Increment Year
E
E N
ENA PWD NO PWDS CFGONLY CFG+CAL
E E E
N or E N or E
Display N Increment
Character Character Display N Increment
E Character Character
** *
E
CFG PWD
N or E
E E Display N Increment
SET GDB CLEAR DB Performs Reset Character Character
and returns
N N E
to ONLINE ** *
N
*If character is not the last position on the display line, advances to next character.
**If character is the last position on the display line, advances to next menu item.
59
MI IDP10-T May 2010 5. Configuration
60
5. Configuration MI IDP10-T May 2010
DAMPING:
To configure additional damping, go to DAMPING with the Next button and press
Enter. Use the NEXT button to select NO DAMP, DAMP 1/4, DAMP 1/2, DAMP 1,
DAMP 2, DAMP 4, DAMP 8, DAMP 16, or DAMP 32 and press Enter.
M1 MODE:
To configure the mode of the primary output (PV), go to M1 MODE with the Next
button and press Enter. Use the Next button to select M1 LIN (linear), M1SQ<1C
(square root with cutoff below 1% dp (M1 URV in linear units of dp), M1SQ<4L (square
root with linear below 4% of dp (M1 URV in linear units of dp), or M1SQ<nC (square
root with user configured cutoff specified between 0 and 20% of the flow upper range
value, M1 EFAC) and press Enter. You cannot configure this parameter as square root if
OUT DIR was configured as OUT REV or if M1LRV is not zero.
NOTE
Cutoff in M1SQ<1C and M1SQ<4L is in percent of differential pressure but cutoff
in M1SQ<nC is in percent of flow rate.
M1 DISP:
To configure the optional local indicator to show engineering units or percent in linear
mode, go to M1 DISP with the Next button and press Enter. Use the Next button to
select M1 EGU or LIN PCT and press Enter. LIN PCT only provides percent readings on
the local display. The M1 engineering unit is used for remote communication of
Measurement #1, even if LIN PCT is selected.
M1 EGU:
To configure pressure or flow engineering units for your display and transmission, go to
M1 EGU with the Next button and press Enter. If M1 MODE is configured as M1 LIN,
you are asked to specify one of the following pressure labels: psi, inHg, ftH2O, inH2O,
atm, bar, mbar, MPa, Pa, kPa, kg/cm2, g/cm2, mmHg, torr, mmH2O, mH2O. or
hW60 (inH2O at 60F). Your transmitter then automatically adjusts M1EFAC
(engineering factor), M1 URV (upper range value), and M1 LRV (lower range value).
M1EOFF is set to zero.
If M1 MODE is configured as M1 SQ<1C, M1SQ<4L, or M1 SQ<nC, you are asked to
specify one of the following flow labels: %flow, gal/s, gal/m, gal/h, gal/d, Mgal/d, ft3/s,
ft3/m, ft3/h, ft3/d, Igal/s, Igal/m, Igal/h, Igal/d, l/s, l/m, l/h, Ml/d, m3/s, m3/m, m3/h,
m3/d, bbl/s, bbl/m, bbl/h, bbl/d, T/h, lb/h, kg/h, Nm3/h (normal m3/h), Sm3/h
(standard m3/h), Am3/h (actual m3/h), or MMSCFD (million scfd). Check the M1EFAC
(engineering factor) and, if necessary, adjust it as follows:
M1EFAC:
This parameter is used to input the numerical relationship between the measured span in
pressure units and the displayed (and transmitted) span in flow units. It is the displayed
URV in flow units (which is also the span in flow units since flow ranges must be zero-
based).
Example:
61
MI IDP10-T May 2010 5. Configuration
For a 200 inH2O transmitter with a measured range of 0 to 100 inH2O and displayed
range of 0 to 500 gal/m, M1EFAC = 500.
To edit the span in your configured flow units, press Enter at the prompt M1EFAC. Use
the procedure Entering Numerical Values on page 38 to edit this parameter.
M1 URV:
To edit the upper range value, press Enter at the prompt M1 URV. Use the procedure
Entering Numerical Values on page 38 section to edit this parameter.
M1 LRV:
Similar to M1URV immediately above.
NOTE
M1 LRV is bypassed if M1 MODE is configured as square root since M1 LRV must
be zero.
M2 MODE:
M2 is a secondary measurement (SV) that is read by the HART Communicator and can
be displayed on the optional display. You might use this feature to display M1 in flow
units and M2 in comparable pressure units. To configure this parameter, go to M2
MODE with the Next button and press Enter. Use the Next button to select M2 LIN
(linear), M2SQ<1C (square root with cutoff below 1% dp (M2 URV in linear units of
dp), M2SQ<4L (square root with linear below 4% of dp (M2 URV in linear units of dp),
or M2SQ<nC (square root with cutoff specified between 0 and 20% of the flow upper
range value, M2 EFAC) and press Enter.
M2 EGU:
Similar to M1 EGU.
DISPLAY:
To display M1, M2, or to toggle between M1 and M2, go to DISPLAY with the Next
button and press Enter. Use the Next button to select SHOW M1, SHOW M2, or
TOGGLE and press Enter.
CALDATE:
This is not a required entry but can be used for record-keeping or plant maintenance
purposes. To edit the calibration date, go to CALDATE with the Next button and press
Enter. You then can change the day, month, and year. The display shows the last date
with the day flashing. Use the Next button to step through the library of digits to select
the desired day, then press Enter. Repeat this process for the month and year.
ENA PWD:
To enable or disable the password feature, go to ENA PWD with the Next button and
press Enter. Use the Next button to select NO PWDS (password not required for either
calibration or configuration), CFGONLY (password required to configure but not to
calibrate), or CFG+CAL (passwords required to both configure and calibrate) and press
Enter.
If you selected CFG ONLY, the display changes to CFG PWD. Press either the Next or
Enter button. Use the Next button to step through the library of characters to select the
62
5. Configuration MI IDP10-T May 2010
desired first character, then press Enter. Your selection is entered and the second
character flashes. Repeat this procedure until you have created your password. If the
password has less than six characters, use blanks for the remaining spaces. When you have
configured the sixth space, the display advances to the next menu item.
If you selected CFG+CAL, the display changes to CAL PWD. To create the Calibration
password, press either the Next or Enter button. Use the Next button to step through the
library of characters to select the desired first character, then press Enter. Your selection is
entered and the second character flashes. Repeat this procedure until you have created
your password. If the password has less than six characters, use blanks for the remaining
spaces. When you have configured the sixth space, the display advances to CFG PWD.
Use the same procedure to create the configuration password.
NOTE
In normal operation, the CAL PWD allows access to only calibration mode. The
CFG PWD allows access to both configuration and calibration.
! CAUTION
Record your new password before saving changes to the database.
SET GDB:
If your transmitter database becomes corrupted and you receive an INITERR message
upon startup, this function enables you to rewrite all calibration and configuration values
with default values.
! CAUTION
Any calibration and configuration values that you have entered will be lost. Therefore,
SET GDB should not be selected if your transmitter is functioning normally.
63
MI IDP10-T May 2010 5. Configuration
Character Lists
Table 9. Alphanumeric Character List
Character List*
@
, (comma) (
A-Z (uppercase) )
[ *
\ +
] -
^ .
_ (underscore) /
space 0-9
! :
;
# <
$ >
% =
& ?
*List only applies to Model 275 HART Communicator not to optional local display.
Character List
. (decimal point)
0 through 9
64
5. Configuration MI IDP10-T May 2010
Error Messages
Table 11. Configuration Error Messages
65
MI IDP10-T May 2010 5. Configuration
66
6. Maintenance
! DANGER
For nonintrinsically safe installations, to prevent a potential explosion in a Division 1
hazardous area, de-energize transmitters before you remove threaded housing covers.
Failure to comply with this warning could result in an explosion resulting in severe
injury or death.
Error Messages
For error messages displayed on the HART Communicator refer to MI 020-366.
Parts Replacement
Parts replacement is generally limited to the electronics module assembly, housing assembly,
sensor assembly, terminal block assembly, cover O-rings, and optional display. For part numbers
relating to the transmitter and its options, see PL 009-005.
67
MI IDP10-T May 2010 6. Maintenance
! CAUTION
The electronics module is one assembly at this point and is electrically and
mechanically connected to topworks with a flexible ribbon signal cable, a 2-wire
power cable, and in some cases, a cable for an external zero pushbutton. Do not
exceed the slack available in these cables when removing the assembled module.
5. Unplug all cable connectors from the rear of the electronics module and place the
module on a clean surface.
6. Predetermine connector orientation, then insert the cable connectors into the
replacement module. Replace the module in the housing using care not to pinch the
cables between the module and the housing. Tighten the two screws that secure the
module to the housing.
7. Connect the cable from the digital display to the electronics module. Ensure that the
O-ring is fully seated in the display housing. Then, holding the digital display by the
tabs at the sides of the display, insert it into the housing. Secure the display to the
housing by aligning the tabs on the sides of the assembly and rotating it about 10 in
a clockwise direction.
8. Reinstall the cover onto the housing by rotating it clockwise to seat the O-ring into
the housing and then continue to hand tighten until the cover contacts the housing
metal-to-metal. If cover locks are present, lock the cover per the procedure described
in Cover Locks on page 29.
9. Turn on transmitter power source.
The module replacement procedure is now complete.
68
6. Maintenance MI IDP10-T May 2010
HOUSING ASSEMBLY
TO REMOVE ELECTRONICS
MODULE, REMOVE TWO
CROSS RECESS SCREWS.
! WARNING
Failure to reuse or install the proper O-ring for a CSA labeled product violates
ANSI / ISA 12.27.01.
6. Feed the sensor cable through the housing neck into the electronics compartment.
7. Screw the housing onto the sensor neck until it bottoms. Do not over tighten. Be
careful not to damage the sensor cable or dislodge the neck O-ring.
69
MI IDP10-T May 2010 6. Maintenance
8. If your housing has an anti-rotation screw, engage the screw until it touches the sensor
neck and back it off 1/8th turn. It is important that the screw is not touching the
sensor. Fill the screw recess with red lacquer (Foxboro Part Number X0180GS or
equivalent). The housing can then be rotated up to one full turn counterclockwise for
optimum access.
9. If your housing has a retention clip, insert the clip over the boss in the housing neck so
that the hole in the clip is aligned with the hole in the boss. Install the screw but do
not tighten. Rotate the housing up to one full turn counterclockwise for optimum
access. Tighten the retention clip screw and fill the screw recess with red lacquer
(Foxboro Part Number X0180GS or equivalent). The housing can still be rotated for
optimum access.
10. Reinstall the electronics module per Steps 6 through 9 in the previous procedure.
70
6. Maintenance MI IDP10-T May 2010
pressure for one minute. There should be no leakage of the test fluid through the
gaskets. If leakage occurs, retighten the cover bolts per Step 5 (or replace the gaskets)
and retest.
! CAUTION
Perform hydrostatic test with a liquid and follow proper hydrostatic test procedures.
PROCESS COVER
SENSOR
PROCESS COVER
GASKETS
71
MI IDP10-T May 2010 6. Maintenance
BOTTOMWORKS WITH
PROCESS CONNECTOR
CODE 7
PVDF INSERTS
PROCESS CONNECTIONS
STANDARD
ORIENTATION
PROCESS COVERS
CONDENSED
LIQUID FREELY
DRAINS
72
6. Maintenance MI IDP10-T May 2010
! CAUTION
Perform hydrostatic test with a liquid and follow proper hydrostatic test procedures.
73
MI IDP10-T May 2010 6. Maintenance
74
Index
C
Calibration 41
Using a HART Communicator 46
Using a PC20 46
Using a PC50 Configurator 46
Using the Local Display 46
Calibration Notes 41
Calibration Setup 43
Configuration 53
Using a HART Communicator 54
Using a PC20 Configurator 54
Using a PC50 Configurator 54
Using the Local Display 54
Cover Locks 29
D
Display, Positioning the 28
E
Error Messages
Calibration 51
Configuration 65
Operation 40
H
Housing, Positioning the 28
I
Identification 3
Installation 15
M
Maintenance 67
Mounting 15
O
Operation Via Local Display 37
75
MI IDP10-T May 2010 Index
P
Parts Replacement 67
Piping, Installation of Flow Measurement 24
R
Reference Documents 1
Reranging 39
S
Seal Liquid, Filling the System with 27
Specifications
Product Safety 10
Standard 5
W
Wiring 29
Write Protect Jumper, Setting the 29
Z
Zero Adjustment Using External Zero Button 49
ISSUE DATES
DEC 2001
OCT 2003
APR 2004
FEB 2005
JUN 2005
FEB 2006
AUG 2006
OCT 2007
JUL 2008
MAY 2010
Vertical lines to the right of text or illustrations indicate areas changed at last issue date.