Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
!"#$%&"'(
Hematology Analyzer
1.0 Release
Cat.-No.: 17401/4
!
%&'()'(*
1 INTRODUCTION 3
2 BLOOD 4
2.1 Function of Blood 4
2.2 Composition of Blood 4
2.3 Blood Cell Parameters 5
2.3.1 Red Blood Cells, Hemoglobin 5
2.3.2 White Blood Cells 6
2.3.3 Platelets 7
2.4 Normal Hematology Ranges of Human Blood 7
3 MEASUREMENT METHODS 8
3.1 Volumetric Impedance Method 8
3.1.1 Diluting whole blood 9
3.1.2 Hemolysing blood cells, 3-part WBC differential method 9
3.2 Hemoglobin Determination by Photometry 11
4 ROUTINE UTILIZATION 12
4.1 Sample Handling 12
4.1.1 Anti-coagulant 12
4.1.2 Taking blood 12
4.1.3 Storage of Samples 12
4.1.4 Handling Blood Controls and Calibrators 13
4.2 Reagents and Environmental Conditions 13
4.3 Aperture Blockage - Clogging 14
4.4 Calibration 15
4.5 Quality Control 15
4.6 Measuring Patient Blood 15
4.7 Using Pre-dilution Mode 16
4.7.1 Sample preparation for 1:10 Pre-dilution Mode 16
4.7.2 Sample preparation for 1:3 Pre-dilution Mode 16
4.7.3 Calibrating Pre-dilution Mode 16
5 MEASUREMENT RESULTS, HISTOGRAMS 17
5.1 Typical histogram 17
5.2 Warning Flags 18
6 HUMAN CASES 20
6.1 Neonate 20
6.2 Three year-old Child 21
6.3 WBC clogging 22
6.4 WBC histogram with Nucleated RBCs 23
6.5 Lymphocytosis: WBC out of linearity 24
6.6 Lymphocytosis: measured in 1:10 pre-diluted mode 25
6.7 High MID and MID% 26
6.8 Granulocytosis 27
6.9 Thrombocytopenia 28
6.10 Thrombocytosis, Anisocytosis 29
6.11 Microcytosis (Low MCV) 30
6.12 Macrocytosis (High MCV) 31
6.13 Under-lysed Sample 32
6.14 Over-lysed Sample 33
&'()#$'-.&',.-)/(,.%-/)*+),"-"),./0%*/(,).#)-'./)!""#$%&'$()*+&),&#)"1(2
34) 0%1/)4-'-.$)&.5)6-2&'(#(755)6-')(,.-.*#5)78)8(%,9"#5):;)<.#=%5)>?):".#)@A,.-*1/B5)C.DD.#0*--)E.%F
%."9/)G)E.%=.#/5)HIII
H4) 8#$)$%&#*#&9(.&'(.5*:$&7)(/'$%/5)3/-)(,.-.*#5*:$'J/K)L4)F)M$N)C45)OD1.#P(1)O0.(#0(5)HIII
FOF,%=C@=0;9;=5,=>,M7==:
The three groups of particles in whole blood are:
1. .1:,M7==:,%1770 (erythrocytes, .M%)
2. Q?;91,M7==:,%1770 (leukocytes, QM%)
3. U74917190 (thrombocytesL,UN()
?&9#-*@A*8-##*%(2"(/$'$()*(B*C1(#-*1,2&)*9#((:**@O*$10(2)HB
D$7,.-*@A*E-B$)$'$()*(B*;E=
!1C=<7=8;5,Y,!]M,Y,is the main component of RBCs. It is conjugated protein (with Fe). Its main function is to
transport oxygen from the lungs to tissues and to transport carbon dioxide from the tissues back to the lungs.
Normal HGB concentration of human samples is approximately 14 g/dl = 140 g/l = 87 mmol/l*. The unit is select-
able in the Settings menu.
#145,%=2@60B6742,!1C=<7=8;5,Y #%!, Y,is the average hemoglobin content of RBCs. Calculated from RBC
and HGB values.
MCH = HGB / RBC in pg or fmol unit*.
#145,%=2@60B6742,!1C=<7=8;5,%=5B159249;=5,Y #%!% is the concentration of HGB in average RBC, cal-
culated from the HGB and HCT values:
MCHC = HGB / HCTabsolute in g/dl, g/l or mmol/l*.
D$7,.-*FA*E-B$)$'$()*(B*=<8*"(",#&'$()/
You should remember this way of operation during the verification/validation process, when you accept the WBC
and 3-part differential results. Do not forget, MID population contains monocytes and some part of the eosino-
phils, sometimes giant lymhocytes, but this is patient dependent.
FOI,'=2C47,!1C49=7=<T,.45<10,=>,!6C45,M7==:
The following table shows a summary of normal ranges of blood cell parameters of healthy adult human patients.
Normal values vary from population to population and can be slightly different in your location.
?&9#-*FA*H(.2&#*6,2&)*6-2&'(#(75*;&)7-/*@O*$10()H4B
D$7,.-*IA*0%1-2&'$%*(B*4(#,2-'.$%*J2"-:&)%-*+-'1(:
The principle of this measurement is that the blood diluted with isotonic solution (diluent) can conduct current by
ionic conduction. A counting chamber made of insulating material (plastics) holds this dilution of the blood, while
there is a small circular hole (aperture) on this chamber, which makes flow of diluted blood possible.
If we place two electrodes on the two sides of this aperture, and we apply constant electric current, the isotonic
solution will begin to conduct electricity, and a voltage can be measured on the aperture.
Applying pressure to the dilution it begins to flow trough the aperture. Once a cell is passing the aperture, a small
change in electric impedance will occur, making the voltage a little bit higher, i.e. a small electric pulse will be
present. The amplitude of the pulse will be proportional to the ratio of the cell volume (size) and the aperture
volume, so the bigger the cell is, the higher pulse we get. The typical aperture size diameter and length is 80 m.
For proper counting differentiation of cells, we need high likelihood (more than 90%) of passing one cell the
aperture at once. Since cell concentration is high in whole blood, we have to :;7691,A?=71,87==: to reduce con-
centration.
The applicable pre-dilution ratio depends on the setting of your analyzer (1:3 or 1:10). See section 4.7 for details
on usage.
HO+O+,/;769;5<,A?=71,87==:
Automated analyzers make the following steps automatically, to reduce the cell concentration to the desired
level.
D$7,.-*OAE$#,'$()*>.(%-//*B(.*<#((:*8-##*8(,)'$)7
3. Then 0.8ml of lyse is added to the MIX dilution (practically while it is moved to the WBC counting chamber
to avoid contamination of MIX chamber with lyse) to form an overall +dF[[,QM%\!]M,:;769;=5. This liquid is
suitable to measure WBC and HGB.
HO+OF,!1C=7T0;5<,87==:,B1770L,HE@429,QM%,:;>>12159;47,C19?=:
Since RBCs are typically 1000 times more concentrated in normal blood, they would interfere with WBC count-
ing, i.e. the high number of RBCs would prevent counting of WBCs, because the aperture containing RBCs
would be always busy.
On the other hand we have to measure HGB, which is inside the RBCs, so we have to call it out somehow from
the cells.
A ?1C=7T0;5<,214<159,`7T01c can be used to dissolve the cell membrane of blood cells, destroying the RBCs,
and creating a complex solution suitable for photometry of HGB.
The membrane of WBCs become selectively permeable so in the slightly hypertonic lyse solution they begin to
shrink to their nucleus. The properly hemolysed liquid contains WBC particles in the 30-400 fl region.
*D$7,.-*SA
?5"$%&#*IN"&.'*:$BB-.-)'$&#*=<8*1$/'(7.&2*(B*/-#-%'$T-*1-2(#5/$/
D$7,.-*UA*6$/'(7.&2*(B*'('&##5*#5/-:*=<8/
If the software gets over-lysed histogram, it will put flag ) No WBC 3-part differential or flag Q, Y, 3-part
WBC differential is not reliable Y to the results.
?&9#-*IA*81&.&%'-.$/'$%/*(B*:$BB-.-)'*'5"-/*(B*1-2&'(#(75*&)W-./
-9, ;0, ;C@=29459, 9=, D5=A that independently of the manufacturer of the analyzer volumetric impedance
method (HE@429,QM%,:;>>12159;47) can differentiate the type of cells only on the basis of their sizes, but not ac-
cording to their internal, structural differences. Consequently this method is only a 7;C;91:, :;>>12159;49;=5
method for making qualitative hematology report.
HOF,!1C=<7=8;5,/1912C;549;=5,8T,U?=9=C192T
HGB determination is one of the first methods developed and is one of the most important hematology parame-
ters as it relates to the oxygen carrying capacity of blood.
Several HGB determination methods are known, but the cyanmethemoglobin method is commonly used. It is
possible to get a two-component neutralising reagent for the waste of lyses containing cyanide. (There are cya-
nide-free lysing reagents on the market in order to avoid environmental pollution.)
The lyse reagent containing potassium ferrocyanide and potassium cyanide converts the hemoglobin to cy-
anmethemoglobin. This substance absorbs green light of 540 nm.
HGB measurement is made by passing light of 540 nm
wave-length through the WBC dilution, and measuring
the light transmitted with a photo detector.
The light intensity (I) of the sample liquid is logarithmi-
cally proportional to the concentration of HGB:
HGB :#1(;#6-reference / Isample)
D$7,.-*UA*6X<*+-&/,.-2-)'*+-'1(:
The analyzer takes reference HGB measurements on clean diluent in each cycle during the cleaning phase to
ensure stability over time and temperature.
IO+OF,(4D;5<,87==:
After filling the sample tubes to the required level, do not forget to invert the sample tube 51S12,0?4D1 sev-
eral times to activate the anti-coagulant by correct mixing. It can be a real problem especially when taking a
small amount of blood from babies or small animals.
Try to keep the 3 mg/ml maximum EDTA concentration rule as well.
(See section 4.7 for details on pre-dilution mode.)
IO+OH,*9=24<1,=>,*4C@710
Anti-coagulant requires time to take its effect on the sample. On the other hand, it will destroy the cells by time.
The fresh samples require at least JE+[,C;56910 to stabilise after taking, and they should be measured A;9?;5,G
?=620 at standard room temperature (25C).
This rule basically applies for the 3-part WBC differential. The granulocytes are destroyed first, after 12-16 hours
you get a much different histogram with much higher MID, because ageing will result in histogram looking over-
lysed. The correct time of collapsing of white blood cells depends on the patient blood.
If you want to measure later, put the samples into refrigerator to extend their stability. Care must be taken in this
case, because the samples must warm up to room temperature before measuring them. Roll the sample tubes
between your palms to speed up warming. This will help mixing the sample, too. If the sample is not mixed cor-
rectly, PLTs and large WBCs can stuck together forming much bigger particles, which will result in distorted
(lower) cell counts and the appearance of non-existing populations (e.g. in high region of WBC histogram).
IOF,.14<1590,45:,)5S;2=5C15947,%=5:;9;=50
Always 601, 9?1, 21B=CC15:1:, and best quality, 214<1590 possible. The use of poor quality reagent can de-
crease analyzer performance, especially for PLT by increasing blank value and 3-part WBC differential by
modifying hemolysing characteristics.
The analyzer will automatically 265 8745D,C140621C159 to ensure instrument and reagent cleanliness. If blank
values are high, repeat blank measurement to ensure whether it is a stable high background or just some tempo-
rary contamination.
Stable ?;<?,84BD<2=65: can be caused by several 2140=50d
1. %=594C;5491:,214<159 if the reagent caps are open on the tanks, and dust or bacteria could get into
the reagent. Replace the diluent, if high PLT blank does not disappear even after several cleaning cy-
cles. Replace lyse, if high WBC blank is present, but PLT blank is good.
2. U12;=:;B,5=;01, can appear on the WBC channel due to protein build-up on the WBC draining tube.
Ask for service to decrease automatic cleaning cycle counter.
3. )71B92;B47,5=;01 can interfere with measurement by causing false pulses. Do not forget, our aperture
generates very small electrical signals in the several hundred micro-Volts (100 V) region. If there is a
device generating high-energy signals near the analyzer, it can harm counting by generating electrical
noise.
The main protection against electromagnetic noise is <==:,<2=65:;5< and shielding. It is therefore very
important that the power supply has a solidly grounded earth lead. If this is not sufficient to overcome the
problem, separate ground wire connected to the rear grounding point (marked) may be needed.
If you face to a high background problem:
! Replace the reagents to check their condition.
! Try to find out what could change in the environment of the analyzer causing interference.
! Try to operate the instrument from another power outlet, or relocate it completely (into another
room, if possible), then try again.
! Ask for the help of qualified service personnel, if necessary.
You should run blank measurement if you replace either reagent during the daily work to ensure cleanliness.
You have to know, that the analyzer will store these values, and e.g. PLT result will be compensated by the
blank value: UN(210679,j,UN(C140621:,Y,UN(8745D.
D$7,.-*YA*LBB-%'*(B*>&.'$&#*8#(77$)7*'(*+84
IOJ,k647;9T,%=592=7
Quality control (kO%O) is a feature to monitor the operation and stability of the instrument.
The HUMACOUNT offers 6 levels of separate Q.C. measuring capability, i.e. you can use 6 different blood con-
trols in the same time to check stability.
The manufacturers of blood control usually offer at least 3 different levels: +, 7=A abnormal, +,5=2C47 (suitable
for calibration as well), and +,?;<? abnormal blood controls.
If you program them to levels 1, 2 and 3, you can measure them every day before starting daily routine meas-
urements to ensure stability of the system. In some laboratories working under quality assurance system
there is a regulation to do so.
You can use the N1STEl155;5<0 chart to get visual information about the everyday results.
If 0=C1,=>,9?1,2106790,421,=69,=>,9?1,@12C;00;871,245<1, try another vial of blood control to see if the results
are outside too on this control. If the second control confirms that the results are outside the permissible range,
you can 21B47;82491,9?1,4547TR12.
Do not forget, the 0948;7;9T,=>,45,=@15,S;47,=>,87==:,B=592=7,depends on the manufacturer, but it is usually,+I
:4T0 only. The closed vials can be used longer (typically for 3 months).
IOG,#14062;5<,U49;159,M7==:
m11@,9?1 04C@7;5<,@210B2;@9;=50 of the Users Manual.
You can make big mistakes if the sample is not correctly homogenised (all counted parameters will be higher),
contains micro-bubbles (lower effective sample value, lower counted parameters), or cold (PLT count will be
bad). At least 30 minutes is required for a sample tube to reach the room temperature if it was in refrigerator
before.
It is recommended to use a 2=942T,C;Z12 to keep samples homogenous during the measuring session. If you do
not have sample mixer, ;5S129,9?1,96810,6@0;:1,:=A5,01S1247, 9;C10 before sampling. '1S12, 0?4D1 them to
avoid formation of micro-bubbles inside, because 0C477, 8688710, @21S159, B=221B9, 04C@7;5< (air is sampled
instead of blood).
After measurement, B?1BD,9?1,2106790,45:,?;09=<24C0 to identify any sort of disorder. .1@149, 9?1, C140621E
C159 in case of any doubt, you can adjust the lyse setting if necessary.
X47;:49;=5 (acceptance of the results) is a very important process. During validation the hematologist or clinician
can sort out those reports, which require confirmation by manual counting under microscope. Do not forget, use
IOK,"0;5<,U21E:;769;=5,#=:1
Pre-dilution mode is used in two cases:
1. if B4@;7742T,9681 (e.g. 25 l of volume) is used for taking blood from the patient,
2. in case of 7;5142;9T,122=2 due to high cell concentration.
In both cases the sample must be prepared before measurement. Create the necessary dilution of whole blood
(or capillary blood) sample according to the set-up pre-dilution mode (you can check it by selecting pre-dilution
mode in Measurement menu, the ratio appears on the screen +dH or +d+[)
U21@421,49,71409,+[[,n7,=>,@21E:;7691:,04C@71, because in pre-dilution mode the analyzer samples double vol-
ume (approx. 60 l).
IOKO+,*4C@71,@21@4249;=5,>=2,+d+[,U21E:;769;=5,#=:1
1. Put,FJ[n7o of clean diluent or isotonic solution into a clean cup or test tube.
2. Add FJn7o,of whole or capillary blood and homogenise it by rotating the cup circularly
The above sample is suitable for measurement in +d+[,@21E:;769;=5,C=:1.
IOKOF,*4C@71,@21@4249;=5,>=2,+dH,U21E:;769;=5,#=:1
1. Put,+J[n7o of clean diluent or isotonic solution into a clean cup or test tube.
2. Add,J[n7o,of whole or capillary blood and homogenise it by rotating the cup circularly
The above sample is suitable for measurement in +dH,@21E:;769;=5,C=:1.
IOKOH,%47;8249;5<,U21E:;769;=5,#=:1
1. %47;82491 the analyzer in 5=2C47,C=:1 using blood control.
2. %21491 the desired @21E:;769;=5 of blood control (1:10 or 1:3) with the method you wish to use for patient
samples as well.
3. $B9;S491,@21E:;769;=5,C=:1 in Calibration menu.
4. U12>=2C,B47;8249;=5 with the prepared sample.
5. After accepting the pre-dilution factors :14B9;S491,@21E:;769;=5,C=:1 in Calibration menu to avoid mix-
up of calibration factors.
6. Make sure to 601,9?1,04C1,C19?=:,45:,@2=B1:621 (tools, settings of pipettes, dispenser volume, etc.)
9=,B2149;5<,04C@710,>=2,C140621C159, too. By doing so you will be able to compensate the inaccura-
cies of your dilution procedure.
oExact volumes may vary. Keep the ratio: e.g. +,65;9 of sample + H,65;90 of diluent (for 1:3 mode) constant.
Evaluation:
WBC three part Possibly lyse problem. Try reducing the lyse amount and repeat the
warning or WBC three measurement. In extreme Lymphocytosis case you can get this flag.
Q part diff. unsuccessful Check the discriminators in the WBC histogram. If the discriminators are in the
0--*8&/-*OZ*@I*&): proper place (the populations can be separated even by eye) then the results
E(7*%&/-/ are correct and acceptable.
Possibly lyse problem, but in some pathological samples (too high
) No WBC three part
lymphocytes), it can happen.
Repeat the blank measurement. If HGB blank is not stable there are probably
HGB blank is high, or
! bubbles in the WBC chamber. Run a cleaning and try blank again. Close the
no HGB blank
side door if open during measurement.
WBC blank is high, or Repeat blank measurement, or run prime lyse and try blank again.
M
no WBC blank Possibly lyse contamination, or noise problem.
Check the 1. RBC-LYM discriminator. If it is in the minimum point (or close to
it), accept the results. Otherwise repeat the measurement.
WBC/RBC limit
N If the repeated measurement gives similar results and discriminator 1. is in
warning
wrong place then the MID and GRA results are OK, but the WBC and LYM
results can be higher because of the RBCs.
Too many RBC cut
Repeat the measurement with increased lyse amount.
from WBC
. If the WBC measuring time is too high (more than 8 sec. in ) it could be
0--*8&/-*@I*&):*8&'
aperture clogging. In this case perform cleaning and repeat the measurement.
%&/-/
WBC coincidence is
The results are out of the linearity range. Make a dilution with an external
too high. Linearity
dilutor with the set-up (1:10 or 1:3) dilution rate.
# error.
Do not forget to calibrate pre-dilution mode before using it for measuring
0--*8&/-*OZ*@I*&):
samples.
8&'*%&/-/
?&9#-*KA*0,22&.5*(B*C&.)$)7*B#&7/*.-#&'-:*'(*=<8[6X<
RBC/PLT valley is too high, or the PLT peak is not high enough.
RBC/PLT limit flag
7 If the discriminator is in a wrong place (in the PLT or RBC histogram) then
0--*8&/-*Y
repeat the measurement for a correct PLT result.
Perform cleaning and repeat the measurement (probably clogging).
D RBC peak warning
If it becomes a general problem, change the RBC aperture.
RBC/PLT coincidence
C is too high. Same action as in case of warning flag,#.
Linearity error.
RBC/PLT data
: Same action as in case of warning flag,/.
package errors
0 RBC/PLT time error Same action as in case of warning flag,/.
B RBC/PLT clogging Same action as in case of warning flag,%.
?&9#-*OA*0,22&.5*(B*C&.)$)7*B#&7/*.-#&'-:*'(*;<8[>\?
Warning flags can be grouped according to measurement conditions and according to the problems relating to
the blood sample.
Measurement conditions: when the flags are related to clogging (BL, %L, :L, /L, D), probably hemolysing problems
()L,8L,ML,@), time out (0L,*), and pressure problems (Fatal pressure error). In this case we suggest repeating the
measurement.
Problems related to blood sample: few or overlapped thrombocytes (7), possible presence of nucleated RBCs
(N), 3-part WBC differential problems (Q). In this case we suggest to change lyse setting and repeat measure-
ment (you can use the arrow keys before measuring to adjust the default lyse setting in 0.1 ml steps). If it gives
the same result, you can accept the results, keeping in mind that parameters marked by asterisk (o) are not
100% reliable.
For PLT (7), lyse setting is not applicable, so if the PLT/RBC discriminator is in correct place, accept PLT result.
The third case is when repeating is required in pre-dilution mode. There is linearity error cell concentration is
too high causing coincidence (CL,#L,.) (see section 3.1).
The asterisk flag (o) near a parameter shows some doubt suspected during the evaluation of that parameter. The
reasons can be: a high PLT blank (PLT value will be marked), a case of indefinite discriminator setting (default
location is used for some reasons, related parameters will be marked), etc.
Another flagging method is evaluation against the normal ranges. If some of the parameters is out of range it
gets a (E) flag if under the range, or gets (f) if over the range. You can customise ranges for all kind of patients
by setting the corresponding lower and upper ranges. If you set 0 for a range limit, it will be not verified (see
Users Manual: Limits).
GO+,'1=5491
This sample comes from a sick new-born baby.
$BB1@9,9?1,2106790,(there are no flags, and all discriminators are in correct position).
Evaluation:
Evaluation:
Evaluation:
possibly N-RBCs
Evaluation:
Evaluation:
You can see the higher WBC result. All asterisks disappeared from differential parameters.
Another effect comes from the accuracy of pre-dilution.
Since we did not calibrate pre-dilution mode before measurement of pre-diluted sample, there are significant
differences in RBC and PLT results. It is due to the non-calibrated pre-dilution mode. _=6,0?=67:,B47;82491,@21E
:;769;=5,C=:1,4>912,B47;8249;5<,5=2C47,C=:1 in order to compensate the differences between the ideal (cal-
culated) and real (created) dilution (see section 4.7.3).
In situations like this, you can accept WBC values from this measurement, while RBC and PLT values are used
from the previous measurement (case 5).
Evaluation:
Evaluation:
Evaluation:
! normal WBC ! warning flag 7: due to the relatively low peak of PLT,
the analyzer is not sure about the correct placement of
! high MID: microcytic RBC PLT/RBC discriminator, consequently it will put
suspect of asterisks near PLT related results
monocytosis
! if you can see that the PLT/RBC discriminator is in
right place (like now) we can accept the marked results
Evaluation:
Evaluation:
Evaluation:
High MCV usually incorporated with low RBC. This is typical in situations where the patient is bleeding. The
bone-marrow tries to compensate low HGB (HCT) by releasing bigger RBCs.
Evaluation:
Evaluation:
KO+,'=2C47,!1C49=7=<T,.45<10,=>,$5;C470
*@1B;10 QM% .M% !]M !%( #%X #%! #%!% UN(
/&] 6-17 5,5-8,5 12-18 37-55 60-77 19,5-24,5 32-36 200-500
%$( 5.5-19.5 5-10 8-15 24-45 39-55 13-17 31-35 300-800
!&.*) 5,5-14,3 6,8-12,9 11-19 32-52 37-58 10-20 31-38 100-600
M&X-') 4-12 5-10 8-15 24-46 40-60 11-17 30-36 100-800
#&"*) 9-21 4,8-6,3 8-15 30-44 50-90 12-13 30-36 200-900
.$( 6,4-26,2 7-10 10,8-17,5 35-51 57-65 15-22 30-35 190-900
.$MM-( 2-15 4-8,6 9-18 30-53 57-90 16-31 32-39 120-800
U-] 11-16 5-8 10-16 32-50 50-68 17-21 30-34 320-720
*!))U 4-12 9-15 9-15 25-50 28-40 8-12 31-34 110-800
]&$( 4-10 8-13 8-12 20-38 16-25 5-7 30-36 300-600
Y @F Y
GHJ?0 @] [# @] [# 7[:# ^ B# "7 7[:# @] [#
?&9#-*PA*0,22&.5*(B*H(.2&#*4-'-.$)&.5*;&)7-/)@O*$10(2)34B
M*P)E7?)'./-*P1"9)Q.-')I4R)9%)*+)%S/()@%.--%()T.-)$#,(1F%S/(,B
M*P)E7?)'./-*P1"9)Q.-')I4U)9%)*+)%S/()@%.--%()T.-)*V(1F%S/(,B
M*P)E7?)'./-*P1"9)Q.-')I4W)9%)*+)%S/()@*V(1F%S/(,B
If you compare the WBC histograms above, you can come to the conclusion that the amount of lyse greatly de-
termines the WBC 3-part differential parameters.
Total WBC decreases as lyse goes higher, because lysing of RBCs is better with more lyse.
On the other hand, the GRA population gets overlapped with LYM making impossible their correct separation by
discriminators.
As a summary we can declare that the 8109,7T01,4C=659,?121,;0,4@@2=ZO,[OI,C7,(probably 0.45 ml),>=2,HE@429
:;>>12159;47,@424C19120, while WBC is acceptable in all cases. This is the reason why you should repeat meas-
urement in some cases with different lyse volume.
Evaluation:
Evaluation:
Evaluation:
! flag / means WBC data error. normal (see RBC histogram: the peak of
the curve is higher than in normal
! the reason of data error is similar to case)
clogging
Evaluation:
KOIO+,%49d,'=2C47,*4C@71
This sample is not typical, rather ideal case of cat samples.
Evaluation:
Evaluation:
Evaluation:
Evaluation:
Evaluation:
Evaluation:
Evaluation:
Evaluation:
Evaluation:
Evaluation:
***
eMail: human@human.de
Internet: http://www.human.de
01/2004-04