Sei sulla pagina 1di 137

MANUAL ON

QUALITY STANDARDS FOR

HIV TESTING LABORATORIES

March 2007

National AIDS Control Organisation


Ministry of Health and Family Welfare

2007@ National AIDS Control Organisation. All rights reserved.


TABLE OF CONTENTS
S.NO. CHAPTERS

Table of contents
Foreword
Acknowledgements
List of abbreviations

1. Basics of HIV 1---5


Structure
Transmission
Immunopathogenesis
Susceptibility

2. Laboratory diagnosis of HIV / AIDS and 6--15


National HIV testing strategies
Introduction
Laboratory investigation for detection
of HIV infection
Causes of false positive and false negative results
Strategies/algorithms of HIV testing
Detection of p24 antigen
Polymerase chain reaction (PCR)
Virus culture
Viral load assay

3. HIV Testing at counseling and testing sites 16--20


(ICTCs/VCTCs and PPTCTC's), etc.
Introduction
Types and Technologies of rapid tests
- Immunoconcentration
- Agglutination
- Immunocomb assay
- Immunochromatography
- Possible outcomes of these rapid tests
Accuracy of rapid tests
Strategies/algorithms for rapid tests
Uses of rapid tests
Quality assurance practices

4. Total quality management 21--25


Management requirements
Technical requirements

5. Laboratory management 26--27


Laboratory configuration
Stock/inventory management of reagents and consumables
Data management

i
6. Quality control 28--46
Introduction
Scientific principles involved
Definition of terms
Essential components
Categories of controls
Dispensation and storage of quality control samples
Determination of acceptable range of quality control to
validate each test run
Variations in the quality control validation guidelines
Importance of 'E' ratios over OD values of external controls to
check validity of runs
Interpretation of aberrant result of QC samples: shift and trend

7. Collection, transport & storage of specimens for HIV 47--50


testing
Performing venipuncture
Transport and storage of samples\

8. Standard operative procedures (SOP) 51--52


The structure of SOP
An example of SOP

9. Documentation and document control 53--54

10. Selection of HIV test kits 55--59


ELISA
Rapid tests
Details of some of the commercially available kits

11. Evaluation of HIV test kits 60--66


Introduction
Preparation of serum panel
Composition of serum panel
Reference test
Testing conditions
Parameters for performance characteristics of a kit
Selection of HIV test kits

12. Equipment maintenance and calibration 67--75


Introduction
Checklist before deciding on purchase of instrument
Maintenance and calibration of different equipment
Calendar of activities for maintenance & calibration of equipment
Function check
Documentation
Preventive maintenance
Operation of ELISA Reader (Multiskan plus)

ii
13. Troubleshooting 76--80
Introduction
Specific assay problems: some examples and their possible solutions
Variations in test results
Common errors in HIV testing and resolution of the same
Vigilance / supervision
Record keeping

14. Implementation of external quality assessment programme at 81--84


different levels of the HIV/AIDS testing network and job definitions
of each level

15. Equipment, supplies and reagents for practicing 85--86


quality control in HIV testing

16. Biosafety practices in HIV testing laboratory 87--98


Introduction
Modes of exposure to blood borne pathogens in the laboratory
Essential steps of biosafety
Biosafety guidelines for laboratory workers
Chemical disinfection commonly used in HIV laboratory
Transport of specimens
Disposal of laboratory waste
Processing of syringes, needles and gloves for reuse
Packing, storage and transport of treated (disinfected) waste
Final disposal of laboratory waste
Container and color coding for disposal of bio medical waste
Spills and accidents
Post exposure prophylaxis against HBV and HIV

17. Selected list of references consulted 99--100

18. Annexures:
Annexure I: Consent form for HIV testing 101
Annexure II: Proforma for requisitioning HIV test 102
Annexure III: Worksheet for ELISA 103
Annexure IV: HIV test report form 104
Annexure V A & B: Formats for reporting to SACS and NACO 105--106
Annexure VI: Proforma for External quality assessment programme for HIV testing 107--111
Annexure VII: Sample EQA Questionnaire for lab supervisors 112--115
Annexure VIII: Sample EQA Questionnaire for lab staff 116-118
Annexure IX: States allotted to the national reference centers 119
Annexure X: Pipetting techniques 120-122

List of Experts For Revised Manual 123

List of Experts involved in Field Testing of the Original Manual 124-125

iii
K. Sujatha Rao
Additional Secretary & Director General
National AIDS Control Organisation, Ministry of Health and Family Welfare, Government of India

FOREWORD

In order to ensure that the result produced by a HIV Testing Laboratory (ICTC/VCTC,
PPTCT, and blood bank laboratory) is reliable, accurate and reproducible, it is essential that
laboratory personnel have the knowledge about all aspects of HIV testing and undertake
stringent quality control measures every day. There is a vast network of HIV testing
laboratories in the country; however there is a need to institute uniform testing and quality
control protocols in these laboratories across the Nation especially in view of the wide
variety of tests kits used by them. The generation of false positive and false negative results
is associated with immense social, ethical and legal implications. It is indeed critical to
assure quality in HIV testing so that accurate results are produced every time the test is
performed.

Manual on "Quality Standards for HIV Testing Laboratories" will provide guidelines for
practicing quality assurance required for HIV testing on a day to day basis; training,
implementation and establishment of quality assurance programs in all the HIV testing
facilities across the country (blood banks, ICTCs/VCTCs, PPTCTs and Blood Bank labs);
evaluation and selection of HIV test kits and standard formats for requisitioning, performing
and reporting the test results.

The guidelines in this manual conform to the recommended quality standards for HIV
testing laboratories as well as the needs of the country based on available infrastructure.

We appreciate the efforts of all the expert who prepared the document; those who field
tested the manual and those who updated the manual.

Director General & Addl. Secy,


National AIDS Control Organization

9th Floor, Chandralok Building, 36 Janpath, New Delhi - 110001 Phone : 011-23325331 Fax : 011-23731746
Email : asdg@nacoindia.org

viuh ,pvkbZoh voLFkk tkusa; fudVre ljdkjh vLirky esa eqr lykg o tkp ik,A
Know your HIV status; go to the nearest Government Hospital for free Voluntary Counselling and Testing.
ACKNOWLEDGEMENTS
The manual entitled Quality Standards in HIV Testing Laboratories is intended for use by laboratory
specialists, laboratory supervisors/in charges and technical staff engaged in HIV testing across the
country .The manual provides background knowledge of HIV; details of laboratory diagnosis of HIV;
step by step practices of quality assurance; HIV testing procedure; evaluation and selection of HIV
kits; troubleshooting and biosafety practices. The document also includes samples of standard
formats used for HIV testing and reporting.

The manual can be referred to for preparing Standard Operative Procedures for any HIV test,
practicing quality assurance and producing accurate results.

We thank Ms. Sujatha Rao, Additional Secretary and Director General, NACO for giving us the
opportunity to revise and update the existing manual.

We are extremely grateful to Dr. Jotna Sokhey, Addl. Project Director NACO, Dr M. Shaukat, Joint
Director, NACO, Dr. D. Bachani, Joint Director, NACO and to all the contributors and experts and the
team at Clinton Foundation who contributed, field tested, and updated this manual to bring the
manual to its final form.

The Contributors

v
LIST OF EXPERTS FOR REVISED MANUAL

1. Dr. Jotna Sokhey, Addl. Project Director, NACO

2. Dr. M. Shaukat, Joint Director, NACO

3. Dr. D. Bachani, Joint Director, NACO

4. Dr. Sandhya Kabra, Deputy Director, NACO

5. Dr. D. Kasana, former Deputy Director ,R & D , NACO

6. Dr. Preeti Mehta, KEM College, Mumbai

7. Dr. Madhuri Thakar, NARI, Pune

8. Dr. R. K. Lodha, AIIMS, Delhi

9. Dr. Mario Tamburini, WHO, SEARO

10. Dr. Usha K. Baveja, former Lab Specialist, Clinton Foundation

11. Dr. Namita Singh , Lab Specialist, Clinton Foundation

12. Dr. Sanjay Sarin, former Lab Specialist, Clinton Foundation

13. Dr. S. N. Misra, Clinton Foundation

vi vi vi 123

vii vii vii


LIST OF EXPERTS INVOLVED IN FIELD TESTING
OF THE ORIGINAL MANUAL

The manual on Quality Assurance practices in HIV Testing Laboratory was field tested on two different

occasions at NIB ,Noida. The following experts contributed:

1. Dr. (Mrs.) Usha K. Baveja Head, HIV/AIDS Division, NICD

22, Shamnath Marg,

New Delhi

2. Dr. PL Joshi Ex-Additional Project Director (Technical)

National AIDS Control Organization

Chandralok Building, 9th Floor


36- Janpath, New Delhi- 110001

3. Dr. Rajesh Bhatia Ex-Director, National Institute of Biologicals,

A-32, Sector- 62, Institutional Area

Phase II, Noida- 201301

4. Dr. Pradeep Seth Ex-Professor & Head Department Of Microbiology,


All India Institute of Medical Sciences
New Delhi: 110029

5. Dr. S.N Misra Consultant, NACO

9th Floor, 36 Janpath,

Chandralok Complex, New Delhi-110001

6. Dr. Shobha Broor Professor


Department Of Microbiology
All India Institute of Medical Sciences
New Delhi: 110029

7. Dr. Shashi Khare Ex-Deputy Director (Quality Control),

National Institute of Biologicals,

A-32, Sector- 62, Institutional Area

Phase II, Noida- 201301

124
8. Dr. V.Ravi Professor & Head
Department Of Neurobiology,
NIMHANS, Bangalore.

9. Dr D.K. Neogi Professor & Head

Department of Virology

School of Tropical Medicine, Calcutta.

10. Dr. N.M Samuel Professor


Department of Experimental Medicine

MGR University, Chennai.

11. Dr. Brajachand Singh Professor


Department of Microbiology,
Regional Institute of Medical science,
Imphal 795004

12. Dr. Charanjeet Sokhey Senior Scientist


Department of Blood products & Microbiology,
National Institute of Biologicals
NOIDA 201301

13. Dr.Ramesh Paranjape Office-in-charge


National AIDS Research Institute,
Plot No. 73, G Block,
MIDC Bhosari, Pune 411026

14. Dr. Arun Risbud Microbiologist


National AIDS Research Institute,
Plot No. 73, G Block,
MIDC Bhosari, Pune 411026

15. Dr. Shameem Banu Professor of Microbiology


Madras Medical College,
Chennai - 600006

125
LIST OF ABBREVIATIONS
BTC : Blood testing centre

CCR-2, 3, 5, CXCR4 : Co-receptors on T cells/ macrophages for HIV

cDNA : complementary DNA

CMV : Cytomegalovirus

CD4 : Cluster determinant 4

CSF : Cerebrospinal fluid

CV : Coefficient of variation

DNA : Deoxyribonucleic acid

EIA : Enzyme Immuno Assay

ELISA : Enzyme Linked Immunosorbent Assay

EQA : External Quality Assessment

E ratio : ELISA ratio

E/R : ELISA / Rapid

gp : Glycoprotein

HBIg : Hepatitis B Immunoglobulin

HBV : Hepatitis B virus

HCW : Health care worker

HCV : Hepatitis C virus

HIV : Human Immunodeficiency Virus

HIV 1 : Human Immunodeficiency Virus Type -1

HIV 2 : Human Immunodeficiency Virus Type -2

HIV -I M : HIV -1 Major group

HIV -I O : HIV -1 Outlier group

vi
HIV I N : HIV -1 New group

HTLV : Human T lymphotropic virus

IB : Immunoblot

NPV : Negative Predictive Value

PPV : Positive Predictive Value

NRL : National Reference Laboratory

SRL : State Reference Laboratory

OD : Optical density

PCR : Polymerase Chain Reaction

PEP : Post Exposure Prophylaxis

PHA : Passive haemagglutination

PI : Protease inhibitors

PPTCTC : Prevention of parent to child transmission centre

QA : Quality Assurance

QAP : Quality Assurance Programme

RIA : Radio Immunoassay

RNA : Ribonucleic acid

SD : Standard deviation

SOP : Standard Operative Procedure

TQM : Total Quality Management

UV : Ultra violet

VCTC : Voluntary Counseling & Testing Centre

ICTC : Integrated Counselling and Testing Centers

WB : Western blot

vii
CHAPTER 1
BASICS OF HIV

Introduction

Human immunodeficiency viruses (HIV) belong to the family Retroviridae and subfamily Lentivirinae and Genus
Lentivirus . Two types of HIV are recognized; HIV-1 and HIV-2. Both differ in geographical distribution, biological and
molecular characteristics and extent of transmissibility. These viruses store their genetic information as ribonucleic
acid (RNA). RNA must be converted to DNA by a special enzyme, reverse transcriptase. HIV-1 has 3 groups, HIV-1
major group (HIV 1-M), outlier (HIV 1-O) and new (HIV 1-N) group. The strains of HIV-1 isolated from people in USA
and Europe are genetically diverse from strains isolated in Africa and Asia. HIV-1 major group can be further classified
into subtypes or clades designated A through K excluding I. Such subtypes have envelope gene sequences that vary
by 20% or more between subtypes. The subtypes also differ in geographical distribution, biological characteristics and
major mode of transmission, etc. HIV-1 subtypes O and N are more distant to all other HIV-1 subtypes but less so
compared to HIV-2. So these are classified under HIV-1 only and have limited distribution in West Africa. HIV-2 has
also been reported from other countries and this also comprises of heterogeneous group of viruses.

Structure

HIV is a 90-120 nm icosahedral, enveloped, RNA virus. HIV comprises of an outer envelop consisting of a lipid bilayer
with uniformly arranged 72 spikes or knobs of glycoprotein (gp) 120 and gp 41 in HIV-1 and gp140 and gp 36 in HIV- 2,
respectively. Glycoprotein 120/140 protrudes out on the surface of the virus and gp 41/36 is embedded in the lipid
matrix. Inside is the p24 protein core surrounding two copies of RNA. This core also contains viral enzymes: reverse
transcriptase, integrase and protease, all essential for viral replication and maturation. Proteins p7 and p9 are bound
to the RNA and are believed to be involved in regulation of gene expression.

Genetic structure

The genetic structure of virus contains both highly conserved and highly variable regions. The high variability of the
virus accounts for drug resistance and evasion from immune response. This also poses problems for development of
a successful vaccine. In an infected individual, quasi-species of a particular viral subtype may be found on account of
constant variability. As a result many recombinant strains like AE, AG and AB may be present in infected individuals.

HIV has structural and regulatory genes coding for structural and regulatory products, respectively. Structural genes
direct the synthesis of physical components of the virus and are also responsible for viral size, shape, structural
integrity and its compartmentalization in host cell. The regulatory genes direct synthesis of proteins that affect the
synthesis of viral components and viral replication. Structural genes are gag, pol and env. HIV also has some
accessory genes vpu, vpr and vif, etc. which increase infectivity and budding.

Replication

Glycoprotein 120/140 of HIV binds to a receptor / receptors on HIV permissive host cell. Predominant receptor is the
CD4 molecule present on T lymphocytes and macrophages, though others such as galactosyl-ceramide (gal C) have

1
also been proposed. Receptors are molecules (proteins and / or glycoproteins) present on the surface of host cells
which facilitate the attachment and entry of viruses into the cell. Entry of virus into the host cell requires certain cellular
co-receptors / factors e.g. CCR-5, CXCR-4, CCR-2 and CCR-3, etc. designated collectively as cell infectivity factors
(CIF). CIF may be a co-receptor or enzyme helping in virus interaction with host cell. Most convincing candidate is the
chemokine receptor related protein, fusin (CXCR-4). Once the gp41/36 of the virus fuses with the host cell membrane,
the capsid is uncoated and a ribonucleoprotein complex capable of reverse transcription is formed. During the
process of reverse transcription cDNA is formed under the effect of viral enzyme, the reverse transcriptase. Reverse
transcription is inefficient in quiescent cells suggesting the involvement of host components in the process. The
nucleoprotein complex formed after transcription comprises of linear double stranded DNA, the gag matrix (MA)
protein, the accessory vpr protein and the viral integrase (IN). This is called pre-integration complex and is transported
into the host cell nucleus. It mediates a complex series of enzymatic steps and integration occurs at cellular loci with
open chromatin structure. Integration probably is an essential step for viral replication. The integrated virus is called
provirus. The virus may not be expressed in many cells and is considered latent. Virus expression can be stimulated
by many viral, cellular and exogenous factors. Other co-existent viral infections e.g. cytomegalovirus, herpes virus
etc. can make the non permissive cells permissive. Maturation of virus also takes place after virus assembly and
budding.

Immunopathogenesis

HIV enters the circulation and targets onto cells expressing CD4 on the surface in various organs including brain and
lymphoid tissues. Infected CD4 lymphocytes migrate to lymph nodes, become activated, proliferate and start
releasing HIV. The antigen driven migration and accumulation of CD4 cells within the lymphoid tissues may account
for dramatic decrease in CD4 cells and generalized lymphadenopathy characteristic of acute retroviral syndrome.
Cellular and humoral responses are established within one week to three months (window period) and host enters the
phase of clinical latency characterized by lack of symptoms and almost initial normal levels of CD4cells.

There is gradual deterioration of the immune system due to depletion of CD4 cells. CD4 cells are destroyed by HIV
mediated cell lysis; formation of syncytia and multinucleated giant cells (infected as well as uninfected cells are
destroyed); virus specific immune responses; superantigen mediated activation of T cells making them permissible to
HIV and programmed cell death (apoptosis). The ability of HIV to replicate in T cells depends on the state of activation
of cells. So, a vicious cycle is established leading to depletion of CD4 cells, destruction of lymphnode architecture and
progressive increase in the viral load.

The progression of HIV disease also depends upon the biological properties of the infecting HIV strain (high vs. low
replicating, syncytium vs. nonsyncytium forming virus).

The HIV disease progresses fast in infants infected in utero because of the immature immune system and rapid
spread of HIV in the body. Most cases vertically infected have a shorter incubation period and a more rapid, fulminant
disease process.

Transmission

Risk factors for transmission of HIV include: unsafe sexual practices with multiple sexual partners, recipient of multiple

2
blood transfusions, use of infected/unsterilised needles or syringes (whether therapeutically or for recreational drug
use), transmission from an infected mother to the foetus/infant before, during, or shortly after delivery and during
breastfeeding by an untreated mother who is infected (perinatal factors that increase transmission include
prematurity, maternal STI, episiotomy, and prolonged rupture of membranes and others). The efficiency of
transmission of HIV is determined by the amount of virus in a body fluid and the extent of contact. High concentration
of free infectious virus and virus infected cells have been reported in blood, genital fluids and cerebrospinal fluid.
Breast milk and saliva yield varying numbers, whereas, other body fluids have a low viral content.

Saliva in adults contains some non-specific inhibitory substances like fibronectins and glycoproteins which could
prevent cell to cell transfer of virus. Thus, saliva is not a likely vehicle of transmission. Urine, sweat, broncho-alveolar
lavage fluid, amniotic fluid, synovial fluid, faeces and tears have been reported to yield zero or few HIV particles.
Hence, these vehicles also do not appear to be important in virus transmission, though standard work precautions
should be undertaken while handling all the body fluids

Breast milk at the time of primary infection in a feeding mother has a high content of virus and may transmit the
infection to the baby. In contrast, cerebrospinal fluid (CSF) also has a high content of virus particularly in individuals
with neurological disease; CSF is not a natural source of virus transmission.

Table1 Efficiency of different routes of HIV transmission and their contribution to total number of cases
reported in India.

Exposure route Percent efficiency * Percentage of total


Worldwide India
Blood transfusion 90-95 5 2.5
Perinatal 20-40 10
Sexual intercourse 0.1 to 10 75 86
Vaginal 0.05-0.1 (60) 85 (heterosexual)
Anal 0.065-0.5 (15) 0.548 (homosexual)
Oral 0.005-0.01 Case reports only
Injecting drugs use 0.67 10 7.3
Needle stick exposure 0.3 0.1
Exposure unspecified 4.2
* see references
Percentage of total in India from NACO website

The most efficient vehicle of HIV transmission is blood (Table 1). However, the risk of infection via blood transfusion is
now extremely low due to strict HIV screening of donated blood. The most common route of transmission is
unprotected, penetrative sexual contact. Different forms of sexual practices carry a variable risk gradient of acquiring
HIV. Cell associated rather than free virus is responsible for disease transmission. Anal intercourse carries a high risk
of transmission because of bowel mucosa which acts as a portal of entry for virus, and also because of a greater
chance of injury to the mucosa. Risk to insertive partner is through infection of lymphocytes and macrophages in the
foreskin or along the urethral canal. In females, HIV transmission occurs when infected cells in the semen gain entry

3
into the female genital tract, and infect the resident lymphocytes, macrophages and probably the uterine epithelial
cells.

The transmission from infected mother to child appears to occur in 20-40% children born to HIV positive mothers
without PPTCT intervention. The source of virus in the new-born is controversial. HIV infection can occur via amniotic
fluid, genital secretions, maternal blood and through the breast milk. Transmission to the baby can occur inutero, and
during or after delivery.

Transmission of HIV infection to Health Care Professionals (HCP) is extremely uncommon. Pooled data from 20
prospective studies suggests that risk associated with needle stick injury from HIV infected blood is approximately 0.3
percent. Further, the risk associated with mucocutaneous contact is too low to be reliably estimated. The risk from
mucosal or non-intact skin is also minimal.

Transmission of HIV through casual contact, sharing utensils, lavatories and through insect bites has not been
documented. Prospective studies offer a conclusive evidence that family members and close household contacts of
HIV infected individuals are not at risk of acquiring HIV infection through casual human contact (shaking hands,
kissing and by sharing of utensils, toilet, bed linen etc.) or by providing routine nursing care. Mosquito bite does not
transmit HIV.

Prevention of HIV Transmission

Transmission of HIV through sexual route can be prevented by creating awareness; promoting mutually faithful sexual
practices, safer sexual behaviors, or celibacy; treating existing sexually transmitted infections. All individuals should
be counseled to take personal responsibility for their sexual health and behavior.

Transmission of HIV through blood and blood products can be prevented by mandatory screening of donated blood
and rational use of blood. It should be emphasized that it is safer to use blood components rather than whole blood.
Whole blood should only be used when so indicated.

Transmission of HIV infection through contaminated needles can be prevented by using sterile needles and syringes;
by reducing the number of injections i.e. prescribing oral drugs in place of injections wherever possible and practice of
standard work precautions at all times while rendering medical services. Post exposure prophylaxis with anti retroviral
drugs as per guidelines should be taken within two hours in case of accidental exposure to HIV-infected blood.

Transmission of HIV infection from infected mother to the new born is prevented currently through intervention with
anti retroviral drug nevirapine under the PPTCT program of NACO.

Susceptibility of HIV
Sterilization
Autoclaving at 121C at 15 lb pressure for 20 minutes
Dry heat 170 C for 1 hr. (holding time)
Boiling for 20 minutes (this kills HBV and HCV also ).

Disinfection minimum 30 minutes contact time


Sodium hypochlorite 0.5% to 1%

4
Ethanol 70%
Povidone iodine (PVI)
Formalin 3-4%
Glutaraldehyde (activated) 2%

References for Table 1:


1. Donegan E, Stuart M, Niland JC, et al. Infection with human immunodeficiency virus type 1 (HIV-1)
among recipients of anti- bodypositive blood donations. Ann Intern Med 1990; 113:7339.
2. Kaplan EH, Heimer R. HIV incidence among New Haven needle exchange participants: updated
estimates from syringe tracking and testing data. J Acquir Immune Defic Syndr 1995; 10:1756.
3. European Study Group on Heterosexual Transmission of HIV. Comparison of female to male and
male to female transmission of HIV in 563 stable couples. BMJ 1992; 304:80913.
4. Varghese B, Maher JE, Peterman TA, Branson BM, Steketee RW. Reducing the risk of sexual HIV
transmission: quantifying the per- act risk for HIV on the basis of choice of partner, sex act, and
condom use. Sex Transm Dis 2002; 29:3843.
5. Bell DM. Occupational risk of human immunodeficiency virus infection in healthcare workers: an
overview. Am J Med 1997; 102:915.
6. Leynaert B, Downs AM, De Vincenzi I; European Study Group on Heterosexual Transmission of HIV.
Heterosexual transmission of HIV: variability of infectivity throughout the course of infection. Am J
Epidemiol 1998; 148:8896.

5
CHAPTER 2
LABORATORY DIAGNOSIS OF HIV/AIDS AND NATIONAL HIV
TESTING STRATEGIES

Introduction

Two distinct human immunodeficiency viruses, HIV-1 and HIV-2 are the aetiologic agents of AIDS. Phylogenetically,
HIV-1 is divided into Group M [10 subtypes A through K excluding I]; Group O (9 subtypes) & Group N (new virus).
HIV-2 has been categorized into 5 subtypes (A to E). In Thailand, India and sub-Saharan Africa, ~ 90% of HIV-1
infections are acquired through heterosexual transmission in contrast to 10% in the U.S. and Western Europe.
Subtypes A, C and D predominate in Africa, subtypes E and B are commonly found in Thailand and C is the main
subtype in India; whereas, subtype B predominates in the U.S. and Western Europe.

In 50-93% of cases primary HIV infection is symptomatic with a variety of symptoms ranging from influenza-like or
mononucleosis-like illness to more severe neurological symptoms which can persist from a few days to as long as two
months. Longer acute clinical illness is associated with rapid progression to AIDS.

Laboratory diagnosis is the only method for determining HIV status of such an individual. A number of tests and
diagnostic kits are available to assess the HIV status of individuals. Serological tests are most commonly performed.
The choice of test protocol depends upon:-

Objectives of HIV testing


Sensitivity and specificity of the test used
The prevalence of HIV infection among the population
Resources available
Appropriateness to the strategy of testing
Infrastructural facilities available

Objectives of HIV testing

Blood and blood products safety. This is determined by mandatory screening of all donated blood
Donors of sperms, organs and tissues are tested for HIV to prevent HIV transmission to the recipient
Diagnosis of HIV infection in clinically suspected individuals
Prevention of parent to child transmission after counseling and informed consent
Voluntary testing: high risk groups after counseling and informed consent
Sentinel surveillance to monitor epidemiological trends
Research, surveys, etc.

Sensitivity

It is the accuracy with which a test can confirm the presence of an infection in an infected individual. Tests with high
sensitivity show few false negative results and are meant to be used to screen blood prior to transfusion and ensure
blood safety. It is recommended that 3rd or 4th generation HIV test kits which are 100% sensitive should be preferred
for use at blood banks for screening donated blood.

6
Specificity

It is the accuracy with which a test can confirm the absence of an infection in an uninfected individual. Tests with high
specificity show few false positives and are to be preferred for the diagnosis of HIV infection in an individual.

Prevalence of HIV infection

The probability that a test will accurately determine the true infection status of a person being tested varies with the
prevalence of HIV infection in the population from which the person comes. The higher the prevalence, greater is the
probability that a person testing positive is truly infected i.e. greater is the positive predictive value (PPV) of the test.
The likelihood that a person showing a negative result is truly uninfected i.e. the negative predictive value (NPV)
decreases as the prevalence of HIV infection among the general population increases.

Laboratory investigations for HIV infection

HIV infection can be detected in the laboratory either by detection of antibodies to HIV, or by detection of the virus, its
antigen and its DNA or RNA. Detection of specific antigens, viral nucleic acid & isolation / culture of virus are all tests in
which the presence of the virus is detected. But most of these tests are very laborious and difficult to perform, require
expertise and hence are to be done only in specified laboratories

The indirect predictors of HIV infection (CD4 cell count, 2 microglobulin, etc.) are monitors of immune status of
patients. CD4 cell enumeration is to be done at routine intervals as per the guidelines, to monitor the progression of
disease and response to ART, etc.

The specimens which can be utilized to detect various markers of HIV infection are given below.
(i) Antibody detection
Blood / serum / plasma
3-5 ml. of blood is collected in clean, screw capped plain vial for ELISA and for the
supplemental tests.
Saliva and urine have been used to detect antibodies to HIV but the assays have not been
validated in India.

(ii) Antigen detection


Blood/serum / plasma (most commonly used sample)
Cerebrospinal fluid (used sometimes)
Cell culture supernatant (i.e. the tissue culture fluid-for research only)

(iii) Virus isolation: HIV can be isolated from HIV infected tissues. It can be successfully isolated from
blood (PBMN cells), semen, vaginal / cervical specimen, tissue, CSF and plasma. It is less
successful on other body fluids like saliva, urine, breast milk, tears and amniotic fluid. Virus isolation
is done for research purposes only.

Detection of specific antibodies

This is done by performing initial screening test, which if positive, is followed up by supplemental tests to confirm the
diagnosis.

7
Screening tests

ELISA/EIA: (Enzyme Linked Immunosorbent Assay/Enzyme Immunoassay)


It is the most commonly performed test at blood banks and tertiary labs to detect HIV antibodies.

There are various kinds of ELISA based on the principle of test:


Indirect ELISA
Competitive ELISA
Sandwich ELISA
Immuno Capture ELISA

ELISA is also classified on the basis of the antigens utilized, into:

1st generation: Infected cell lysate is used as the antigen.


2nd generation: Glycopeptides (recombinant antigens) are used as the antigen.
3rd generation: Synthetic peptides are used as the antigen.
4th generation: Antigen and antibodies are detected simultaneously. The assays may use a combination of
recombinant and synthetic peptides as antigens.

When a serum sample tests reactive once by a system of ELISA / Rapid (E/R) test, the test is to be repeated
immediately by a different system in order to confirm the diagnosis at VCTCs and PPTCTCs. If it tests reactive a
second time, the sample is to be taken up further for supplemental tests to confirm the diagnosis. Supplemental tests
are E/R based on different principles of test or different antigen systems. Sometimes, WB, etc. may have to be used to
resolve the sero-discordant results.

ELISA takes up to three hours to yield results. It has a major advantage of being economical. Although rapid tests give
results within minutes, these are relatively expensive per test. Commercial kits are available for both ELISA and rapid
tests. Rapid tests include:
Dot blot assays (immunoconcentration based)
Particle agglutination (gelatin, RBC, latex, microbeads)
Dip stick and comb tests, etc (EL ISA technology based).
Immunochromatography based tests (lateral flow of reagents)

Tests which detect antibody to HIV 1 and 2 and all the subtypes of HIV1 and HIV2, are to be employed

Supplement tests

Second and third ELISA/Rapid


Western blot (WB)

WB is expensive, time consuming and requires expertise to perform. This is mostly done to confirm the diagnosis on
samples which given discordant results in E/R and in legal cases.

Causes of false positive and false negative results

There are many conditions which may give rise to false positive or false negative test results. False positive results
have been reported in cases of haematologic malignant disorders, DNA viral infections, autoimmune disorders,
8
multiple myeloma, primary biliary cirrhosis, alcoholic hepatitis, chronic renal failure, positive RPR (rapid plasma
regain) test and false negative results have been reported in cases of window period prior to seroconversion,
immunosuppressive therapy, malignant disorders, B-cell dysfunction and bone marrow transplantation, etc.

Strategies / Algorithms of HIV testing

Because of the enormous risk involved in transmission of HIV through blood, safety of blood products is of paramount
importance. Since the PPV is low in populations with low HIV prevalence, WHO/GOI have evolved
strategies/algorithms to detect HIV infection in different population groups and to fulfil different objectives. The various
strategies/algorithms, so designed, involve the use of categories of tests in various permutations and combinations.

1. Screening ELISA/ Rapid tests (E/R)


Used in strategy I, II & III
2. Supplemental test: E/R and Western Blot(WB)
WB is used in cases where discordant serological results are obtained; if WB is not available, then
client should be retested all over again after 2-4 weeks.

Strategy I (algorithm I)

Blood/plasma/serum is subjected once to E/R for HIV. If negative, the sample is considered free of HIV and if positive,
the sample is taken as HIV infected for all practical purposes. This strategy is used for ensuring donation safety (blood,
organ, tissues, sperms, etc.). Units of blood testing positive by the screening E/R are destroyed as per guidelines. A
donor who gives consent to know the result of HIV test is informed about the result. In case the test is positive, the
donor is informed about the possible reactive nature of the result and is advised that the same requires confirmation.
The individual is accordingly referred to VCTC for counseling, testing and confirmation of the test result.

Strategy II A (algorithm II A)

A serum sample is considered negative for HIV if the first ELISA or rapid (screening) test is negative, but if reactive, the
same sample is subjected to a second ELISA or rapid test which utilizes a system different from the first one. i.e. either
the principle of test and/or the antigen used is different. It is reported reactive only if the second ELISA/ rapid test
confirms the positive report of the first test. In case the second E/R is non reactive, then the result is taken as negative
for surveillance purposes (II A). Strategy II A is used for sentinel surveillance (anonymous, unlinked testing).

Strategy II B (algorithm II B)

This strategy is used for detection of HIV infection in symtomatic individuals with symptoms of AIDS clinically. The
sample is processed as in Strategy IIA, but a sample reactive with first assay and non-reactive with second assays
subjected to the third tie breaker E/R. If the third E/R is reactive, the samples reported as indeterminate and patient is
called back for repeat testing after 2-4 weeks. If the third test is negative, it is reported as negative. Two to three
different test kits with different antigen system or different principle of the test are required to follow this Strategy IIB of
testing. In this strategy if the first two consecutive tests are positive for presence of HIV antibodies, a positive report
can be given to the patient.

9
Strategy III (algorithm III)

Testing is done as in strategy IIB. However, added confirmation of a third reactive E/R test is required for a sample to
be reported as HIV positive. If the sample gives reactive result with two E/R and non reactive with the third assay, it is
reported as indeterminate and patient is called back for repeat testing after 2-4 weeks. The test used as the
screening test is one with the highest sensitivity (may give high number of false positives) and the supplementary
second and third tests used are with the highest specificity (to minimize false positive reactions). This strategy is used
for diagnosis of HIV infection in asymptomatic individuals. Counseling and informed consent are a must in these
cases. Three different kits with different antigen system and / or different principles of tests are required to follow this
strategy.

Strategy IIB & III are to be used for diagnosis of HIV infection in symptomatic and asymptomatic individuals,
respectively. A sample reactive in first assay and non-reactive in second assay is subjected to a third tie breaker assay.
However, in case the third assay also gives non- reactive result, sample is reported as negative for HIV infection. In
such cases when a negative report has been issued but the client history is suggestive of high risk factors/behaviour,
he may asked to come for follow up testing after 2-4 weeks.

Antibodies to HIV-1 are most commonly and reliably detected by E/R tests and confirmed by different E/R tests.
Antibody testing by ELISA remains the standard method for screening potential blood donors; simultaneous testing
for p24 antigenemia is superfluous because of low sensitivity and expense. Use of improved, third and fourth
generation serological assays demonstrates that seroconversion typically occurs within 3-12 weeks post-infection,
although significant delay can occur in some individuals.

In diagnosing HIV-1 infection, the specificity of ELISA is >99% when properly performed and the sensitivity is >98%. In
low-risk populations, the false-positive rate of combined EIA and WB testing is estimated to be <1 in 100,000. Highly
sensitive and specific agglutination and EIA methods for detection of type-specific antibodies to HIV-2 are also
available.

Strategy I (algorithm I)
(For Transfusion/ transplantation safety)
One test kit required
A1

A1 + A1 -
Consider Consider
Positive2 Negative
(Destroy the unit of blood
as per guidelines refer to VCTC
for confirmation of status after consent)
10
Strategy/Algorithm II A (For surveillance)
2 Test kits required
A11

A1 + A1 -
Report Negative

A21

A1 + A2 + A1 + A2 -
Report positive Report negative

Strategy/Algorithm II B
(Diagnosis of an individual with AIDS indicator disease symptoms.)
3 Test kits required.
A11

A1 + A1 -
Report Negative

A21

A1 + A2 + A1 + A2 -
Report positive
with post test
A31
counselling

A1 + A2 - A3 + A1 + A2 - A3 -
3
Indeterminate Report Negative
11
Strategy/Algorithm III
[To detect HIV infection in asymptomatic individuals (VCTCs, PPTCTs) ]
3 Test kits required
A11

A1 + A1 -
Report Negative

1
A2

A1 + A2 + A1 + A2 -

A31

A1 + A2 + A3 + A1 + A2 + A3 -
Report positive with
Indeterminate3
post test counseling A31

A1 + A2 - A3 + A1 + A2 - A3 -
Indeterminate3 Report Negative
The screening test selected preferably should be fourth generation and the sensitivity of the test should be 100%. The
supplemental/confirmatory tests selected should be >99.8% sensitive and >98.5% specific. Highly sensitive tests
reduce the occurrence of false negative result and fourth generation HIV tests reduce the window period to 2-3 weeks,
thus playing an important part in ensuring blood safety. Highly specific tests reduce the occurrence of false positive
result and thus ensure that a person testing positive as per the strategy II/ III is actually HIV infected.

1
Assays A1, A2, A3 represent 3 different assays
2
Such a result is not adequate for diagnostic purposes: use strategies IIB or III. Whatever the final diagnosis,
donations which were initially reactive should not be used for transfusions or transplants. Refer to VCTC after
informed consent for confirmation of HIV status
3
Testing should be repeated on a second sample taken after 14-28 days. In case the serological results continue to be
indeterminate, then the sample is to be subjected to a Western blot /PCR if facilities are available or refer to the
National Reference Laboratory for further testing.
12
All diagnostic results to be reported after post-test counseling.

Detection of p24 antigen

Detection of p24 viral antigen is expensive and not very reliable. The sensitivity of the test is also limited. Though a
positive test may indicate HIV infection in adults, a negative test does not rule out HIV infection. It is undertaken in the
following situations:

To detect infection in the newborn (not reliable)


To resolve equivocal Western Blot results
To detect infection during early window phase
To diagnose CNS (central nervous system) disease
Late stage of disease (immune collapse)
For research
To monitor response to anti-retroviral therapy ( quantitative assay)

EIA for HIV-1 antigen detects primarily uncomplexed/dissociated p24 antigen, in serum, plasma, CSF or cell culture. It
indicates active infection and allows diagnosis before seroconversion. Quantitative test can predict prognosis and
may be useful for monitoring response to therapy. Disadvantages of antigen detection assays include: poor sensitivity
(only 69% in patients with AIDS and low in neonates < 1 month old); detection is not possible in patients with high titers
of p24 antibody (which complexes with the antigen); and failure to detect HIV-2 antigen.

Polymerase chain reaction (PCR)

PCR can detect proviral DNA during window period, can differentiate latent HIV infection from active viral transcription
and can quantitate the copy number of HIV RNA when used with external standards (e.g. viral load assays). PCR can
successfully differentiate between HIV-1 and HIV-2 infections. Proviral DNA can be detected in peripheral blood
mononuclear cells before seroconversion. Limitations to the diagnostic use of PCR are rare false-negatives, some of
which can be avoided by the use of multiple primer pairs and primers from conserved regions of the genome, and
false-positives due to cross-contamination of the PCR reaction mixture. PCR is not to be used as a screening test.

HIV-1 can be detected by PCR in the CSF of HIV-infected patients independently of disease stage; spread of HIV-1 to
the brain represents an early event during infection which occurs in most asymptomatic individuals. PCR can also be
used to detect HIV infection in neonates born to HIV infected mothers.

Virus culture

Virus culture is another method for identifying HIV infection. Positive culture rates of up to 98% are reported in
confirmed seropositive individuals. The culture method is, however, expensive, labour-intensive, can take weeks for
complete results and potentially exposes laboratory workers to high concentrations of HIV. Virus culture is used for
research (drug sensitivity, vaccine studies, etc.) only.

Viral load assay

Quantitation of HIV RNA in plasma is useful for determining free viral load, assessing the efficacy of anti retroviral

13
therapy and predicting progression and clinical outcome. Because baseline HIV viral load is predictive of survival at 10
years in patients with nearly identical CD4 counts, assessment of baseline viraemia prior to initiation of therapy may
be useful in patient management.

Indirect predictors of HIV infection


Decreased CD4 cells
Increased 2 microglobulin
Increased serum neopterin
Increased IL-2 receptors
AIDS defining illnesses.

Details about enumeration of CD4 cells and the uses of this marker in management of HIV infected individuals are
given in Guidelines on enumeration of CD4 cells using single platform technology.

Unlinked anonymous testing

Such type of screening or testing is not directed to the individual, but has as its objective, the public health surveillance
of HIV infection. It is a method for measuring HIV prevalence in a selected population with the minimum of participation
bias. Unlinked anonymous screening offers a distinct advantage over mandatory or voluntary testing. Unlinked
anonymous testing involves use of blood already collected for other purposes; therefore, the effect of selection bias
will remain, though minimal, and will depend upon time, location and other details of blood collection.

Voluntary confidential testing

Testing is often done for diagnostic purposes. Here it is important that the issues related to confidentiality receive great
attention. Since this method is based on voluntary HIV testing or testing for diagnosis of HIV/AIDS cases, it is
imperative to respect the individual's need to maintain confidentiality. By maintaining confidentiality, it will not only
instill faith in the individual about the health care system in the community but also encourage more and more people
practicing risk behavior to come forward for an HIV test. This testing is done after counseling and informed consent of
the client.

Mandatory testing

When testing is done without the consent of the patient and when the data could be linked to identify the person, it is
called mandatory testing. Mandatory testing is recommended only for screening donors of semen, organs or tissues
(to ensure transplantation safety) in order to prevent transmission of HIV to the recipient of the biological products.

Choice of HIV tests

The choice of tests is also based on the different objectives of HIV testing. The tests that are adopted are the ELISA or
Rapid Test, clubbed together as 'E/R'. One E/R denotes a test done on one single antigen preparation; two is when all
positive samples on first antigen test are repeated on a second antigen preparation / principle and three is when this
test is repeated on the same sample for a third time using a different antigen /principle system. For transfusion safety
purposes, one E/R is used; for surveillance two E/Rs are used; for diagnosis of AIDS cases two or three E/Rs are used,
and for asymptomatic individuals three E/Rs are used.
14
General principles of HIV testing

Testing policy in general should consider the following points:

It should be part of the overall comprehensive preventive programme.


Testing should be technically sound and appropriate to the objective of testing.
Test procedure must be appropriate to the field situation.
Testing procedure must be cost effective.
Laboratory procedures must be monitored for ensuring quality.

HIV testing in health care settings

The fear and apprehension that exists among health care workers in managing HIV infected individuals and AIDS
patients is largely due to the risk of HIV transmission due to a needle stick or other sharp injury. Thus the demand for
mandatory HIV testing of patients admitted in hospitals or undergoing surgery, etc is not rational or appropriate. A
mandatory HIV test is no substitute for Standard Work Precautions that need to be adopted for every patient in a
hospital or any other health care setting. On the other hand testing without explicit consent of the patient has been
proven to be counterproductive in the long run. In the control of the HIV epidemic, such testing can drive the target
people underground and make it more difficult for launching interventions.

The national testing policy reiterates the following:


No individual should be made to undergo mandatory testing for HIV.
Mandatory HIV testing should not be imposed as a precondition for employment or for providing
health care services and facilities.
Any HIV testing except that undertaken for blood safety and transplantation safety must be
accompanied by pretest and post test counseling services and informed consent. Confidentiality of
the result should be maintained.

15
CHAPTER 3
HIV TESTING AT COUNSELLING AND TESTING SITES
( VCTCs AND PPTCTCs) USING RAPID TESTS

Introduction:

Rapid tests are in vitro qualitative tests for the detection of antibodies to Human Immunodeficiency Viruses (HIV)
types 1 and 2 in human serum, plasma, whole blood saliva and urine. Currently HIV testing in India is performed on
serum/whole blood (fingerprick), and plasma. This is because the HIV testing on urine and saliva samples has not
been evaluated and validated in India. In recent times, a large number and wide range of rapid tests of high quality
have become available and are being currently used in laboratories under the following conditions:
Facilities to perform ELISA test are absent,
In emergency cases
Remote blood banks where the collection volume is low and the facilities for ELISA are not available
(ELISA is the test of choice for HIV testing at blood banks) .
Point of care settings like VCTCs, PPTCTCs, and also intervention healthcare sites, etc.

Types and technologies of rapid tests:

Different types of rapid tests are available.


The various technologies on which rapid tests are based include:
Immunoconcentration ( flow through) ( dot blot assays)
Immunochromatography ( lateral flow assays)
Particle agglutination (latex, gelatin, RBCs, etc)
Immunocomb (Dip stick/ comb tests) ( mostly ELISA based)

Immunoconcentration:

This technology is available in the form of dot blot assays. These are multi step tests, very rapid to perform and results
are obtained within a few minutes. The antigens (recombinant / synthetic peptides) are passively blotted on the nitro
cellulose membrane matrix which is bound to the solid support and contains the adsorbent pads to collect the serum
and reagents after their addition. Inbuilt control (internal control) in the form of a spot or line to indicate the validity of
the test is available with each test.

Agglutination:

Agglutination assays are easy to perform and require no wash procedures. The antigens are adsorbed passively on
the carriers (red blood cells, latex particles, gelatin particles and or micro beads). The antibody present in the serum
sample reacts with the antigen adsorbed carriers, resulting in clumping of agglutinated particles. However, if in the
reaction mixture the concentration of antigen and antibody is not optimum, phenomenon of prozone reaction
(antibody excess) occurs which may lead to a false negative reaction. To overcome this initial dilution of the test
sample is to be done as recommended by the manufacturer. Reactions are read visually. The control (uncoated or un
sensitized particles) is run to detect non specific agglutination. If agglutination is detected in both the control and the
test samples the assay needs to be repeated. Examples of these tests are Capillus and Serodia.

16
Capillus agglutination test

Reactive
Latex Aggregation
White Clumping

Non-reactive
No Latex Aggregation
no white clumping

Immunocomb assays:

The immunocomb is a rapid test intended for the qualitative and differential detection of antibodies to HIV-1 and HIV-2
in human serum or plasma. The immunocomb is an indirect solid phase enzyme immunoassay (EIA). The solid phase
is a comb with 12 projections. Each tooth has 3 spots:
Upper spot: Goat antibodies to human immunoglobulin (control to validate the test)
Middle spot: HIV-1 synthetic peptides
Lower spot: HIV-2 synthetic peptides

The developing plate has 6 rows (12 wells each) with each row containing a reagent solution ready for use at different
steps in the assay. Immunoglobulin present in the testing samples is captured by the anti human immunoglobulin
antibody on the upper spot (internal control). Unbound components are washed away. IgG from the sample is
captured on the teeth and reacts with antihuman IgG antibody labeled with alkaline phosphatase which react with
chromogenic components and the results are seen as gray blue spots on the surface of the teeth of the comb

Interpretation of controls:
Appearance of upper spot- Negative control
Appearance of all 3 spots- Positive control
Upper spot does not appear - Invalid result
Interpretation of results:
Appearance of upper spot - Non reactive sample
Appearance of upper and middle spot- Reactive for HIV-1
Appearance of upper and lower spot- Reactive for HIV-2
Appearance of all 3 spots - Reactive for HIV-1 and HIV-2

Post-assay activities:
After the assay, discard the test devices in the box containing 1% sodium hypochlorite
Discard the used filter paper into the bio hazard bags
Swab the workbench and all equipments after use with 1% sodium hypochlorite followed by 70%
alcohol

17
Diagramatic representation immunocomb test

Human Ig
HIV-1
HIV-2

Positive Negative for HIV Reactive Reactive Test not valid


control Control dot seen for HIV-2 for HIV-1 (Human Ig spot control
Reactive for dot not seen)
both HIV-1
and HIV-2
Immunochromatography:

These are lateral flow one step tests. These tests are usually temperature stable and can be stored at the room
temperature. For HIV testing the rapid tests are used to detect the HIV antibodies. Examples of kits based on this
technology are Determine, Unigold and Hemastrip.

Reactive
2 lines of any intensity appear in
both the control and patient areas.

Non-reactive
1 line appears in the control area
and no line in the patient area.

Invalid
No line appears in the control area.
Do not report invalid results. Repeat test
with a new test device even if a line
appears in the patient area.

Possible outcomes of these rapid tests( Immunoconcentration, Immunofiltration, Dipsticks and Immunocombs) :
Reactive or positive : Appearance of test band and control band
Non reactive or negative : Appearance of control band only
Invalid : Absence of control band. Test has failed. Repeat the Test with a new
device

Advantages:
No equipment is required to perform the test. Micropipettes may be needed to dilute the sample for
some assays eg. agglutination assays
Limited infrastructure is needed. Some of the rapid tests can be stored at room temperature. These
tests have wide temperature range stability.
Rapid assays can be used in remote peripheral labs and for circumstances where same day results
are required e g VCTCs and PPTCTs
18
The rapid assays are versatile as far as sample stability is concerned. These can be used on whole
blood, serum, plasma, saliva or urine. Different kits are available for use on these different samples
Most of the rapid tests are one to five step procedures
Reading is subjective and visual with naked eye examination. However now rapid test readers are
available for some tests, allowing us to perform semiquantitative analysis. Rapid immunoassay
results can now be read objectively using CCD chips (as in digital cameras).
Rapid tests are now more robust
Management of waste generated by performing rapid tests is easier to manage as compared to that
generated by performing ELISA
Rapid tests work out more cost effective overall particularly in set ups like that in VCT and PPTCTCs
as they are more robust, there is lesser cost of the requisite lab infrastructure and lesser cost for
sample collection

Disadvantages
More expensive because of single test packaging and performance.
Test performance may vary by the product.
Refrigeration required for some of the products.
Each test card cannot be quality controlled with an external quality control sample.
There may be issues of sensitivity, specificity, negative predictive value and positive predictive value
in relation to the reference gold standard. These should be carefully considered while procuring the
product.

Accuracy of rapid tests

Many rapid tests available in the market are as accurate as the classical assays, e.g., ELISA. Some of the rapid tests
may be less accurate. This happens when manufacturers do not follow total quality management and good
manufacturing practices and cut the costs. In order to ensure quality of rapid tests, the WHO carries out and publishes
evaluations as part of its Bulk Procurement Scheme. In addition, rapid tests can be validated vis--vis the classical
assays in the specific locations before use. This was done by NACO before the rapid test kits for use at VCTC's were
recommended i.e., the available rapid test kits were evaluated in a blinded multicentric study vis--vis available
ELISA's by NICD on behest of NACO and the rapid test kits were recommended for use at VCTS's and later at
PPTCTC on the basis of evaluation results obtained. HIV rapid tests were found to perform as reliably as ELISA's in
the field.

Strategies for rapid test use

Rapid tests are used following the algorithm based on serial testing following the national strategies of testing
appropriate to the objective of testing. The rapid test selected for screening should be 100% sensitive. Highly specific
(> 98.5%) rapid tests should be selected for supplementary and confirmatory testing.

Uses of rapid tests

These tests can be performed on patients at point of care facilities like peripheral health care centers, pharmacies,
emergency vehicles, out patient clinics, VCTC's, blood banks, PPTCT centers, hospital wards, homes, in the field and
hospital wards.

19
Quality assurance practices

Most of the rapid test kits have inbuilt IgG / sample / reagent control spot/line which monitors the quality of kit as well as
the quality of performance of the test. The IgG spot/line always shows up as positive provided the kit reagents are
functioning optimally and the test has been performed as per the manufacturer's instructions. However these are the
kit controls. In order to ensure the validity of the test result, and quality performance of test, a known positive external
control must be tested along with the test samples using the same kit (same lot number and expiry, etc.) on day of
performance of the assay. This will ensure quality of the kit as well as the procedure.

20
CHAPTER 4
TOTAL QUALITY MANAGEMENT

Introduction

Medical science is advancing rapidly and everyday new technologies, which are more sensitive and specific, are
being developed. The more sensitive the technology, the more precise and stringent the processes have to be, to
produce accurate results. This can only be achieved by practicing total quality management.

A variety of highly sensitive tests have been developed and are being used to detect HIV infection in different
healthcare settings. Minor error in performance of HIV tests can lead to repercussions of far reaching significance. A
false positive result may have catastrophic implications for an individual and a false negative result may have
disastrous implications for transfusion safety.

Efficient performance of tests can only be ensured by rigorously conforming to the norms, elements and practices of
total quality management.

Elements of total quality management

Management requirements
Organization and management
Quality system
Document control
Procurement of equipment, reagents and supplies
Maintenance contracts
Internal audit
Complaints
Preventive action
Corrective action

Technical requirements
Personnel
Civil Infrastructure
Environment
Equipment
Sampling
Handling of test and calibration items
Assuring quality of test and the results
Reporting of results
Biosafety and safe waste disposal

Management requirements
Organization

The organization under which the laboratory functions, is legally responsible for the activities undertaken in the
laboratory. These activities (HIV testing, etc.) should meet the requirements of the clients and regulatory authorities.
21
The organization provides the managerial and technical personnel who have the resources and authority needed to
carry out the assigned functions i.e. HIV testing in this case.

The organization ensures that policies and procedures are in place right from the collection of the sample to the
reporting of results. The quality system practiced is able to identify the occurrence of departure from standard
procedures and errors and takes corrective actions as and when required.

The organization is responsible for assigning responsibilities, ensuring smooth inter relationship of personnel who
manage, perform and verify the laboratory activities. Organization also provides supervision and has overall
responsibility for technical operations and provision of resources (equipment, space and supplies) and manpower to
carry out quality HIV testing.

Quality system

Quality System also dictates that there will be a quality manual which gives details of all processes and procedures to
be followed for HIV testing; a biosafety manual which will give details of standard work precautions to carry out the
procedures and tests in the lab and guidelines on safe disposal of waste.

Document control

The documents form a part of the quality system. There are a number of documents generated in the laboratory
ranging from the policy documents, procedure documents, worksheets, instruction sheets, equipment documents,
reports and records, etc. A system has to be in place to control all these documents. The documents should be
numbered and available with the organization. All the documents are available and accessible to the laboratory
personnel. Documents with controlled access should be clearly listed.

Authorized, current documents should be available at all locations where operations are performed. The old
documents should be archived as per the laboratory norms for historical reference. Documents on procedures,
equipment, etc. should be revised from time to time as required. Details on documentation are given in the relevant
chapters .

Procurement of equipment, reagents and supplies

Procurement of equipment and supplies is an integral part of the quality system. Mostly it is done through the standard
tendering process and contracts. The laboratory should ensure that the equipment, kits, reagents, disposables and
other supplies meet the quality standards, are inspected and checked for quality before actual use. The procedure to
procure equipment and kits is discussed in detail in the relevant chapters.

Maintenance contracts

All equipment in the laboratory must be under the annual maintenance contract (AMC) so that there are no
interruptions in the functionality due to sudden breakdowns. AMC should be inbuilt in the procurement process.

Internal audit

Internal audits should be conducted periodically and in accordance with a predetermined schedule and procedure to
verify and ensure that the laboratory procedures fulfill the requirements of the quality system. Internal audit serves to
identify any lapses/errors in the system which can be corrected by corrective actions. The internal audit report should
be tabled and discussed with the lab administration/ management so as to ensure adequate addressal of the non-
conformances raised during the audit.
22
Complaints

There should be a procedure and policy in place to resolve the complaints received from a client. These may include
delay in reporting the result, rude behavior of personnel, break in confidentiality and wrong result, etc. There should be
an investigation following the complaints, the responsibility should be fixed and corrective action taken to resolve the
complaint. A note on corrective action taken should be made in the records.

Preventive action

Preventive Action is a pro-active process to identify opportunities for improvement to prevent breakdown in
equipment, work and quality system.

Corrective action

A problem with the quality system or with technical operations of the laboratory may be identified during internal/
external audit, management reviews, feedback from clients or staff observations. The corrective action has to be
taken to resolve the problem. For this, an investigation to identify the cause of problem should be carried out by
analysis of all potential sources of the problem. Many a times this may be the most difficult task. However once the root
cause/ causes of the problem are identified corrective action should be taken (such as repair of an equipment, training
of staff), documented and effect of corrective action must be monitored

Technical requirements

Personnel

It is the responsibility of the management to ensure that staff posted in the laboratory has the requisite competence to
carry out testing. The personnel working in the HIV testing laboratory should know how to operate and manage the
equipment; how to interpret, record and report test results and how to implement all components of quality
management system .The staff must be adequately trained and continue to undergo refresher trainings to maintain
competence. All materials and supplies required to practice standard work precautions must be provided by the
management. Immunization against HBV must be given to the staff working in the laboratory.

Civil infrastructure

The laboratory should have adequate space to carry out the procedures, house the equipment and space for
personnel. There should be enough lighting and all power requirements should be met. Surfaces should be smooth
and disinfectable. There should be a pest control system in place. Working surfaces preferably should be acid
resistant.

Environment

A clean environment should be provided for the laboratory. The temperature and humidity should be maintained as per
the requirements of the testing procedure and the equipment maintenance etc .The environmental conditions should
not adversely affect the required quality of test or invalidate the test. Smoking, eating and drinking should be
prohibited in the laboratory and signs to this effect must be displayed.

23
Equipment

It must be ensured that the equipment needed for the required testing is available and functional. The list of equipment
required for the HIV testing is given in the relevant chapter . The equipment should be maintained as per the
instructions of the manufacturer and calibrated as recommended. The equipment should be operated by authorized
staff and the documents on equipment should be a part of document control.

The details to be documented about equipment include name of equipment, name of manufacturer, serial number or
any other unique identification, record of checks as per specifications, manufacturer's instruction manual, and the
contract for maintenance, malfunction modification/ updating, etc.

The procedure for calibrating the measuring equipment should be available. Other details on equipment procurement,
maintenance and use are given in the chapter on equipment maintenance and calibration.

Sampling

Sampling is a defined procedure whereby part of a substance, material or product is taken to provide for testing or
calibration of a representative sample of the whole. The sampling procedure should describe the selection, sampling
plan, withdrawal and preparation of a sample from a substance, material or product to yield the required information.
For example before the HIV test kits are imported in the country ,the Drug Controller General of India lifts a sample of
kits for evaluation by the statutary institute. .

Handling of test and calibration items

The procedures for identification, transportation, receipt, handling, protection, storage, retention and disposal of test /
or calibration items , including all provisions necessary to protect the integrity of the test or calibration items should be
established and made available to the staff. There should be no deterioration of test and of calibration items during
storage, handling and preparation. The instructions for transportation, etc. if any accompanying the test or calibration
items should be provided and followed.

Assuring the quality of test results

Quality control procedures for monitoring the validity of the test should be undertaken each time the test is performed.
Appropriate statistical techniques should be employed to review the validity of test results. All such techniques should
be listed in the quality manual/SOP for the benefit of the lab staff. The standards can be achieved by:

- Regular use of certified reference material.


- Participation in external quality assessment/proficiency testing program.
- Replicate the test using same or different methods.
- Retesting of sample if required

Reporting the results

The results of the test carried out should be reported accurately, clearly, unambiguously and to the right person in the
right manner i.e. after post test counseling. The result should be written clearly for interpretation by the clinician.
24
The report should include patients ID, name and address of the laboratory, mention of the sample, name of the test
carried out, name of the test performed, name of the person who performed the test and the name of the person who
reviewed the test result. All results should be reported with appropriate units and with applicable reference range.

Biosafety

It is essential to have the guidelines on standard work precautions in the laboratory. The bio safety practices are
extremely important to prevent transmission of blood borne viruses to the technologist. These precautions should be
practiced diligently at all times while working in the laboratory and/or providing other services in the health care facility.
The bio safety practices are detailed in a separate chapter.

Safe disposal of waste

Much of the waste generated in the laboratory may be infectious. The procedure for disinfection and disposal of waste
must be laid down and practiced strictly. Waste should be disposed as per the guidelines given elsewhere in this
manual.

25
CHAPTER 5
Laboratory Management

Introduction

There is always pressure on the laboratory to produce quality results, using current technology, while keeping up with
increasing demand to aid clinicians and program managers. This can be ensured by instituting a proper lab
management system that looks after all the various aspects of HIV testing. A well developed lab management system
should work to increase staff efficiency and reduce reagent wastage. It should therefore cover pre-analytical,
analytical as well as post-analytical stages of laboratory testing.

A well designed lab management system should consist of the following:

Laboratory Configuration

The key to a reliable HIV testing lab is for it to be optimally configured. This should involve the following:
Equipment layout: The lay out of the equipment should be in line with the testing requirements of the lab.
This may vary for a lab with high or a low testing throughput. This would include layouts for refrigerator, water
bath, incubator, centrifuge, ELISA reader, etc.
Workflow staging and direction: Work areas should be arranged to allow uni-directional sample flow and
defined space for each test step. Defined areas are needed for:
Specimen collection
Specimen receiving and storage
Specimen preparation
Specimen testing-instrument
Result production, validation and release
Reagent and consumable storage
Electrical requirements: These should be in line with the testing requirements of the laboratory.
Staffing requirements: Decisions on staffing should give due consideration to the required level of expertise
in terms of the type of laboratory and the tests to be performed. It should also take into account the number of
trained staff required on the basis of the operational volumes.

Stock / inventory management of reagents and consumables

The laboratory should have a well defined inventory management system. The inventory or the stock management
system could be manual or electronic depending upon the available resources. The system should be designed to
ensure the following:
Uninterrupted reagent supply to prevent reagent stock-outs.
Clearly define the buffer stocks as well as the re-order levels
Availability of other consumables, (e.g. pipette tips, gloves, needles, syringes, vacutainers, and other
tubes).
Stocks are in line with the testing demand/needs of the lab to prevent a situation where there is an
excess stock of reagents / consumables. Any such situation would lead to wastage of precious
resources on account of expiry of these goods.

26
Data management

Proper record keeping of patient results is vital for providing optimal patient care and gaining knowledge from patient
data collected. A well defined data management system should:

Ensure reliable and rapid delivery of results to clients / clinic sites


Ensure clinics have systems for receiving and processing result data
Ensure the laboratory maintains records of result data for defined periods
Use standard reporting formats
Ensure that dedicated human and other resources for data management are assigned

In addition, the data management system should work to ensure that:

Laboratories examine all specimens accepted to ascertain that they meet the proper criteria for data
entry and processing
Laboratories verify that the results are for the intended patient
Laboratories enter and process data correctly, to ensure that the results are given out in a correct and
efficient manner. This can be achieved in the following manner :
Results should be entered on both the worksheet and patient result form
Worksheets should be filed by date in the laboratory for easy result retrieval
Patient report form results should be entered and processed.
All patient reports should be signed by the concerned technologist / technician and the same
should be verified by the lab head.

The release of results in the correct form is as important as conducting the test. The results should only be handed
over to the concerned individual after confirming all the antecedents / particulars of the concerned individual & post-
test counseling. Confidentiality should be maintained in all respects.

Another important aspect of data management system is the archiving of results. Archives can be be electronic or
paper-based. It should use a consistent system for data storage based on one or more variables (collection date, clinic
site, patient ID) that allows easy reference and retrieval of records.

27
CHAPTER 6
QUALITY CONTROL

Introduction

Quality control practice aims to assess the measurement error. It is defined as the set of systematic procedures
required to be implemented in the laboratory to evaluate and monitor the accuracy and precision of any analytical
process. Quality control indicates that the assay is valid, all test conditions for that run have been met, and all test
results for that run are reliable.

Quality control however does not ensure that the results are accurate since the accuracy of the result may be
influenced by factors like characteristics and limitations of the test employed. Moreover a run that passes the quality
control does not ensure that the results have been reported properly and/or reported to the right individual in the right
manner.

In order to assure that the results produced by a HIV testing laboratory (BTC, surveillance centre, VCTC and
PPTCTC) are accurate, it is essential that laboratory personnel have the knowledge about each and every aspect of
HIV testing. There is a vast network of HIV testing laboratories in the country. Though the monitoring of the proficiency
program has been started (EQAP) there is still a lot to be done to make the program a success. The laboratories are
not practicing internal quality control on a day-to-day basis. The problem is further compounded by a wide variety of
test kits used by these laboratories. As the generation of false positive and false negative results is associated with
social, ethical and legal implications, it is extremely important to practice quality assurance for HIV testing to avoid the
occurrence of inaccurate results. The standard operative procedures for practicing quality assurance are available
with few HIV testing laboratories across the country. Moreover, the simple measures that could be undertaken to
ensure quality of testing in the labs is often viewed as a complicated task due to lack of background knowledge.

This chapter has been designed as an attempt to help the laboratory personnel at HIV testing laboratories to
understand, adopt and practice quality control for HIV testing in their respective facilities. The chapter describes the
procedures for quality control in HIV testing presuming that infrastructure, manpower and equipments required for the
purpose are in place. This can also be used as a model to prepare standard operating procedures for practicing quality
assurance for other transfusion transmitted infections.

Scientific principles involved

Routine procedures for separation, handling and storage of serum / plasma.


Preparation of serial dilutions of serum / plasma.
Running Enzyme Immunoassay (EIA) for HIV, employing HIV specific antigen coated plates.
Calculation of statistical values like mean, standard deviation and coefficient of variation.
Plotting of quality control graphs.
Calculation of performance characteristics of an EIA kit in terms of sensitivity, specificity, predictive
values and delta value.
Ensuring quality assurance of rapid HIV testing procedures

28
Definition of terms

Total quality management

Total quality management (TQM) refers to a comprehensive organizational approach that is focused on continually
improving the quality and efficiency with which the laboratory operates. TQM is a component of quality assurance
(QA) that implements and increases overall quality of the programme.

Quality assurance

Quality assurance is defined as the means by which the laboratory ensures that the final results reported are correct
and are achieved in a standard, reproducible and traceable manner. Quality assurance helps ensure avoidance of
mistakes; consistency of performance; data integrity; efficiency and cost effectiveness; customer satisfaction and
laboratory esteem and credibility. Quality assurance practices are important to acquire standards to national and
international accreditation.

Quality assessment

Quality assessment also known as proficiency testing and external quality assessment (EQA) is a means to
determine the quality of results generated by the laboratory, which involves incorporation of proficiency panels
prepared by an external agency into routine testing and analysis of the results produced after the testing of this panel.
The process is known as the External Quality Assessment Programme. These results are then assessed and
compared with the standard results, a performance score is developed and feed back is given to the participating
laboratories.

The aim is to assess the performance of the laboratory for specific tests, identify the factors responsible for incorrect
results (material, processes, performance) and try to correct these through corrective actions and training.

Specimens of EQAP (external quality assessment programme) must be handled in the same way as a specimen is
handled for diagnostic purposes in the laboratory.

The EQA process involves the following elements:-


- Gathering information from the participant laboratories regarding the kits/assays being used.
- Preparation of serum panel.
- Distribution of panel
- Collection of results from the participating laboratories
- Collation and analysis of results
- Sending the analysis report to the participating laboratories
- Taking corrective actions for the participating laboratory performing poorly on EQA panel through
locating the problem and resolving the same.
- Issuing participation certificate to all the participating laboratories.

Quality control (QC)

Quality control refers to the continual measures which must be undertaken for each assay to ensure that the test is
working accurately as per the limits of the test so as to produce valid and acceptable results.
29
Essential components
The following essential components of quality control must be performed during every assay:
Each test run must include one full set of controls that should yield results within the limits of the
standard for acceptability and validity of the run. The values of controls obtained for the first run of the
day cannot be used for subsequent runs. The controls have to be included in each run.
Any run not having at least the minimum recommended number of controls falling within the
acceptable range is invalid and should be repeated.
All test kits should be used before the expiry date.
Physical parameters of the test such as incubation time and temperature must be followed as per
manufacturer's instructions.

Categories of controls Internal controls/kit controls


Refers to the negative and positive controls provided in the kits. These controls are useful but have the following
limitations:
They can be used only in those batches of kits from which they originate.
Most of the controls in commercial kits are established artificially in a manner that minor deterioration
of the kit may not be detected by the readings of kit controls.

For the above reasons, the quality control programme relies more on the use of external controls for internal quality
control.
External controls
These are a set of controls included from outside i.e. not belonging to the kit. They are applied for each run/assay to
continually monitor assay performance (validity).

A model Quality Control Programme that incorporates the external controls has the following main components:

Making an external control with desirable anti HIV titer available

Determination of acceptable ranges

Inclusion of external controls in each run

Data collection

Analysis

Reporting
Sources of external controls:
National Reference Laboratories.
Commercial control panels (e.g. BBI panel from Boston Biomedical Inc.). However, these controls
are expensive and thus may not be affordable for routine use by most laboratories.
Producing in-house external controls from the sera subjected routinely for testing.
Samples collected from another laboratory. However, the samples should be recharacterized and
well standardized before being employed as external QC sample.
Pooled test kit controls - ideal economically.

30
Use of freeze-dried controls

Freeze dried control serum samples can be stored at 2-8C for a very long time. In addition, the shipment of freeze
dried samples is very simple and straight forward. However, during the process of freeze drying, antibody activity,
especially of the borderline samples may be lost. Borderline control samples should not be freeze dried.

Preparation of external controls with desired titer

The categories of external controls to be used with each run comprise of a positive, a borderline reactive and a
negative control. However, the ideal external control (or if one has to employ only one category) is the borderline
reactive sample since this control is capable of detecting any minor error in assay performance. This is important in
order not to generate false results among test samples having an OD near the cut off value.

The steps for preparing a borderline reactive external quality control sample are:
Selection of a high titer HIV positive plasma/serum.
Serial dilution of the high titer HIV positive sample using normal HIV negative human serum.
Selection of a suitable sample dilution to achieve the desired titer for use as external control with
borderline reactivity.
Batch production: preparation of bulk external (low positive) control sample.
Batch validation.
Dispensation and storage.
Validation after storage.

Selection of a high titer HIV positive sample


Sources could be (i) either an HIV infected individual or (ii) a unit of blood found to be positive for HIV. In the former
case, adequate quantity of serum could be obtained from blood collected from the individual and allowed to clot while
in the latter case, the plasma is separated and recalcified to yield serum. Both plasma and serum are heat inactivated
at 56C for 30 minutes. The source is subjected to an appropriate EIA (preferably more than one variety) to confirm
high titer of anti HIV antibody before collecting bulk specimen.

Procedure for obtaining serum from plasma

Ideal material for preparation of external positive controls is serum. This is because plasma tends to be unstable on
long storage and may clot spontaneously during transport. Under many circumstances blood collected from a blood
bank may be the only source for HIV positive and HIV negative samples and thus plasma must be recalcified to
produce serum. However, recalcification should be done with care because excessive recalcification may affect some
assays like gelatin particle agglutination, adversely. Only citrated blood should be recalcified.

Recalcification of plasma to obtain serum


The procedure for obtaining serum from plasma is given below:
A 10X recalcification solution (0.25M CaCI2. 6H2O, 55g and 0.08M MgCl2. 6H2O, 16g dissolved in 100
ml distilled water) is prepared and autoclaved at 121C for 20 min at 15 lbs pressure.
1.5 ml of recalcification solution is added to 1 unit of blood or 250 ml of plasma which has been
brought to room temperature.
Incubation is carried out at 37C for 30-60 min (until clot formation) followed by overnight storage at
4C.
31
Centrifugation is carried out at 1500 rpm for 20 min.
The serum is separated aseptically.
A small volume of serum is tested for HIV antibodies by screening and confirmatory tests (as well as
for HBV and HCV). The bulk (i.e. remaining quantity of serum) can be stored in smaller and
convenient volumes at -20C, after labeling.
If plasma is more than 7 days old, thrombin should be added to the plasma before re-calcification.

Serial dilution of the high titer HIV positive sample

This procedure is intended to prepare a sample with low reactivity to be included as external QC sample in each run.
Many commercial organizations supply low positive QC panel sera which do not require this exercise unless there is a
reason to recheck the titer.

Selection of diluents

The recommended diluent is normal human serum rather than normal saline or other sample diluents. This is to keep
the antibodies in natural serum protein environment. Adequate quantity of serum for use as diluent can be obtained
from a pool of normal healthy individuals (HIV, HBV and HCV negative) or from the unit of blood negative for HIV, HBV
and HCV. In the later instance, plasma has to be recalcified to obtain serum as per the procedure mentioned earlier.

Protocol for serial dilution

Two fold serial dilutions(doubling dilutions) are prepared where a fixed volume of HIV positive serum is mixed and
transferred into successive tubes containing an identical volume of diluent (i.e. normal human serum). When one unit
volume of positive serum is mixed with an equal volume of normal human serum (diluent) the final dilution of the
positive control becomes two fold or 1 in 2 (i.e. tube No. 2 in the figure 1). Accordingly when unit volume of material
from tube number 2 (already diluted to 1 in 2) is mixed with unit volume of diluent in the next tube (tube no 3 in the
figure) the dilution of positive serum will be again two fold of that of tube No 2 i.e. 1 in 4 of original and so on. This
exercise of doubling dilutions will be helpful in subsequent steps (vide below) when ultimately the requisite dilution of
positive serum, that would represent the low positive serum as external control, is to be worked out. For example, if it is
found that dilution 1 in 256 of the positive control serum gives value acceptable to be used as external control, then
100l (or 0.1 ml) of this positive serum can be added to 25.50 ml of normal human serum to get 1: 256 dilution.

Fig. 1 Scheme for preparation of doubling dilution of HIV positive serum

Discard

1 2 3 4 5 6 7 8 9 10
Tube No
11

- - - - - - - - - - -
Positive
serum
Diluent 200ml 100ml 100ml 100ml 100ml 100ml 100ml 100ml 100ml 100ml 100ml
Final
dilution (reciprocal) N 2 4 8 16 32 64 128 256 512 1024
Log2 dilution 0 1 2 3 4 5 6 7 8 9 10

32
Performance of EIA with serially diluted HIV positive sample

Each dilution is considered as a separate sample.


An EIA kit is selected that is quality control approved and is well within the valid expiry date.
Each dilution made above is charged in triplicate in the EIA plate (triplicates are shown in figure 2 as
A1, A2 and A3 for each dilution).
The triplicates of any particular dilution should not be charged in adjacent wells, but should be spaced
apart randomly all over the plate to minimize well to well variation (intra run variation) as shown in
Figure - 2.
The EIA is performed as per the manufacturer's protocol.

Fig. 2 Conducting EIA with the dilutions of (in triplicates) HIV positive serum.

1 2 3 4 5 6 7 8 9 10 11 12

A CP 1 2 1
B CP 3
C CP
D CN 1
E CN 2 2
F CN 3 3
G
H
CP = positive control (kit controls); CN = negative control (kit controls)
1, 2, 3...12 = Indicate sample (pre-diluted) numbers
Note: Controls in the plate map are displayed as per the protocol of Innotest ELISA.

Recording the results

The OD values of the samples (i.e. triplicates of each dilution as well as the controls supplied with the kit) are recorded
as per the validity of the run. Cut off value is calculated on the basis of the reading of the controls supplied with the kit
(internal controls) as per the kit protocol. The mean of the triplicate OD values of each dilution is calculated. The
dilutions are plotted in X axis while the E ratios are plotted on the Y axis (vide below for the significance of E ratio).

Expression of antibody level in terms of 'E' ratio

Cut off values in EIA may differ depending on the principle of test, manufacturer and the recommended protocol for its
calculation. Following are the few examples of acceptable ranges of cut off values recommended by different
manufacturers. (Table 1).

Even with the day to day use of kits from same manufacturer with identical batch numbers, some degree of variations
in the internal controls (supplied with the kits) are encountered, that result in turn in the variation of the calculated cut
off value, that is calculated on the basis of OD values of the internal controls. This is due to variation in factors like
incubation conditions, preparation of the reagents, plate to plate as well as well to well variation in the amount of
33
coated antigens, etc. However, such factors influencing the OD values of the controls would also expectedly influence
the OD values of the test samples in similar direction. Thus the relative reactivity of a given test sample in relation to cut
off value would not vary much. This relative reactivity of a test sample in relation to cut off value in a particular run is
expressed and termed as E ratio. This is the ratio between the sample OD and cut off OD. Hence it is more appropriate
to express the degree of ELISA reactivity in terms of E ratio rather than individual OD values as explained. Following
are the examples of two unknown samples, designated X and Y, run on different dates using kits from same
manufacturer and same batch (Table 1 and 2).

Table 1 Ranges of cut off values recommended by different manufactures

Name of the kit Principle of Manufacturer Range of cut off OD values (in
10 consecutive runs employing
same batch of kit )

1. Innotest HIV 1+2 Indirect EIA Innogenetics NV, 0.176 - 0.313


Belgium

2. Detect HIV Indirect EIA Biochem Immuno 0.160 - 0.240


System Inc, Canada

3. UBI HIV 1/2 EIA Indirect EIA United Biomedical 0.125 - 0.205
Inc., USA

4. Recombigen HIV-1 / Indirect EIA Cambridge Biotech 0.393 - 0.518


HIV - 2 EIA Ltd., Ireland

Table - 2 Variations in the EIA values of two samples X and Y run in different EIA kits

Sample number Run day Sample OD Cut off OD E ratio

X 1.262 3.476
1
Y 0.029 0.363 0.066

X 1.512 3.625
2 0.417
Y 0.032 0.076

X 1.778 3.432
3 0.518
Y 0.037 0.071

As evident from above example, the variation of OD values expressed as 'E' ratio is much less compared to OD value
above.

Selection of suitable sample dilution from the dilution curve

Results usually show a sigmoidal curve.


As mentioned earlier, dilutions suitable for using as low positive controls are generally selected at 'E'
ratio of 1.5-2.0 and around 0.70 (i.e. 1.5-2.0 times and about 0.70 times the cut off OD value) for an
indirect and competitive EIA respectively (Figure 3.).
34
Fig. 3 EIA reactivity (E ratio) of HIV-1 positive serum by indirect EIA Innotest HIV-1/HIV-2,
(Innogenetics kit) in relation to serial dilutions in normal serum.

10
9
8
7
6
o
5
4
3
2
1
0
1 2 3 4 5 6 7 8 9 10 11

Log2 dilution of HIV positive serum

Solid line indicates the cut-off limit for interpretation of result


Arrow indicates suitable dilutions for preparation of external control

Batch production: Preparation of bulk external control

Volume requirement

The requisite volume of the low positive control to be prepared for use as external control will depend on:
How long the external control of the current batch needs to be used (e.g. for one year or 6 months).
The sample volume required for the assay (e.g. 10 - 20l for most indirect EIAs).
How often the assay is performed (i.e. number of specimens tested per day).
This will indicate the number of times the external control would be needed over a specified period.
Number of laboratories to which QC sample is to be supplied for use as external control.

Procedure
HIV positive sample and diluent are filtered to remove bacteria and fungi. However, filtration of
recalcified serum is often difficult as filters tend to clog by fibrin clots. Hence the positive sample
should be centrifuged, prefiltered with a 0.8m biological filter to remove aggregated proteins or
debris and then filtered through a 0.22 m biological filter to remove contaminating bacteria, although
there will be inevitable reduction in volume of the valuable specimens selected as controls.
The positive sample is centrifuged and filtered (0.22 m) followed by heat inactivation at 56C for 20
minutes.
Positive sample and volume of diluent i.e. normal human serum containing preservative (e.g.
Bronidox) is mixed in the volumes required to yield the required titer (as described above) and kept
for 16-20 hours at 4C on a magnetic stirrer.
The required titer is ensured by running EIA for 10 runs for 10 days.
The batch thus produced is aliquoted into large polypropylene storage containers (e.g. 250 ml).
Polyethylene containers are not recommended as these absorb antibodies and thus may affect the
titer of the QC sample.

35
An I.D number is assigned to the lot with date of production mentioned.
All containers to be used to produce the batch should be autoclaved.

Batch validation: Determination of degree of inter-aliquot variation


Principle

This process is to check that the batch has been sufficiently mixed and is homogenous so as to minimize the inter-
aliquot as well as inter run variation. Higher the dilution required to produce QC sample of required titer, more is the
importance of such validation.
Aliquots (25 shown in figure) are dispensed from each storage bottle. This means 125 aliquots from 5
storage bottles.
These aliquots are tested in the assay or assays in which the sample is to be used as a QC sample
subsequently. The batch should be accepted if (i) the reconstituted batch after aliquoting yields the
targeted titer (ii) the samples show minimal variation following aliquoting (i.e. inter aliquot variation).

The aliquots of samples as selected above are subjected to EIA by the EIA kit of the brand that is going to be employed
in daily use. The internal controls supplied with the kit are included to check validity of the run. The kit selected for this
purpose should be licensed and quality checked. Preferably, an external control from the previous lot should also be
included in the run.

The mean (X), standard deviation (SD) and coefficient of variation (CV) of the 'E' ratios of a total of 25 aliquots are
calculated as per example given below in Table - 3.

Table - 3 An example of values obtained while testing batch variation of 25 batches of


external control

36
Mean

The mean (expressed as X) is calculated by adding individual observations (X1+X2+X 3------Xn or X) and then
dividing by the number of observations (expressed as n).

Thus mean is calculated as follows:

In the data shown in table, the mean 'E' ratio calculated as the sum of individual 'E' ratios divided by number of aliquots
is found to be as follows:

Standard deviation

Standard deviation (SD) is a measure of the variation or dispersion of observations about the mean and defines the
expected range of a control in relation to the mean value.

Standard deviation is calculated as below:


_
Each of the individual values ('E' ratios) are compared with the mean (X) to find out the deviation from the mean. Thus
for sample with 'E' ratio as X1 the deviation will be X1 ~ X, for the next sample with 'E' ratio as X2, the deviation will be
X2 ~ X and so on. The deviation is also expressed as 'd' conventionally.

37
Coefficient of variation

The coefficient of variation (CV) is expressed as percentage and the following formula is used.

Fig. 4 Batch validation (inter-aliquot variation) of 25 aliquots of external control sample;


coefficient of variation (CV) is calculated to be 6.82%.

A Levy-Jennings chart in which E ratios are plotted against the Y axis and the individual samples (aliquots) plotted
against the X axis are shown in Figure - 5.

38
Fig. 5 Levy-Jennings chart to plot a QC graph on the basis of values ('E' ratios) obtained in
20 consecutive days run of external control.

Coefficient of variation less than 10% is considered to be an indication of little inter aliquot (batch) variation and thus
consistency can be expected in the performance of various aliquots as external quality control in daily run.

Dispensation and storage of quality control samples

Even though, one may start with sterile serum samples, subsequent use and manipulation in the laboratory may easily
contaminate these. Some of the preservatives that could be used are (i) Bronidox (5-bromo-5-nitro-1, 3-dioxane) in
propylene glycol to a final concentration of 0.05% in serum, may be used conveniently. (ii)Thiomersal (mercuric
chloride) can be used in a final concentration of 0.01%, but is effective only for a few weeks as it loses its activity when
exposed to light. (iii) Sodium azide is not recommended as it inactivates peroxidase conjugate.

Following validation, all the batches of reconstructed low positive controls are stored at -20C or below in non self
defrosting freezer upto 1 year. These batches (already validated) are to be utilised one at a time for preparation of
large number of aliquots, with volume per aliquot sufficient to last for one week in routine testing (e.g. 200l/ aliquot). It
is recommended that the aliquots are stored divided in two freezers in different locations keeping in mind the
possibility of electrical failure. One aliquot is thawed at the beginning of the week for use for that week only, following
which it is discarded. It is stored at 2 to 8C in between, during the week. Use of quality control samples is to validate
each test run.

Determination of acceptable ranges of quality control to validate each test run

Range for accepting individual runs of the external control as valid is an important parameter that needs to be
established. This is carried out by testing the external controls for a sufficient time to establish the limits before the
actual use of external controls to validate the runs. This is done by employing statistical parameters like mean,
standard deviation and coefficient of variation. Subsequently, the external controls, in conjunction with the internal kit
controls, can be used to validate all test runs.

Procedure
The external QC sample is tested in at least 20 runs (e.g. in 20 consecutive days) to make statistically
significant observations.

39
The mean and standard deviation of 'E' ratio of the external controls are calculated (as per method
mentioned above) in each of the 20 runs separately.
The coefficient of variation (CV) of the external controls is calculated on the basis of mean and
standard deviation of the 'E' ratios.

It is important that coefficient of variation (CV) of the 'E' ratios observed on different dates is minimal (i.e. <20%). An
example of values obtained in 20 consecutive runs of external controls is shown in Table - 4.

Table -4 Examples for calculations of acceptable range of external control

Data analysis

As per the standard methods of calculations the following observations are made about the above mentioned runs of
EIA:
Mean ( X) = 1.90
Standard deviation (SD) = 0.19

Limit of Mean 1SD = 1.71-2.09


Limit of Mean 2SD = 1.52-2.28

Coefficient of variation (%) = 11.6%

40
Having assured the minimal batch variation (inter aliquot variation) i.e. CV < 20%, as tested in at least 20 consecutive
runs, the external control is considered to be suitable for inclusion as external control in subsequent daily runs of EIA.

Preparation of a QC graph (Quality Control Chart)

Following validation of minimal variation of the external control values (i.e. CV < 20%) the 'E' ratios are plotted on the Y
axis in the Levy-Jennings chart against consecutive dates of runs plotted on the X axis. On the basis of standard
deviation (SD), calculated as above, a limit of 2 SD is drawn on either side of the mean This limit of mean 2 SD is
accepted as the limit for assessing the function of external controls in day to day monitoring of the run of EIA to check
validity Figure 5.

Incorporation of external control in daily test run

The external controls are treated as any other serum sample being tested in the EIA run.
EIA is performed as per manufacturer's guidelines with the conditions identical for the external
control as well as test samples. The observed OD values of the internal (kit) controls are checked and
matched with the limits prescribed by manufacturer as acceptable to ensure that they are within
acceptable limits.
The OD value of the external control is recorded.
The 'E' ratio for the external control is calculated on the basis of OD value of the external control and
cut off value obtained in the same EIA run.

OD value of external control


'E' ratio =
Cut off OD value

The observed 'E' ratio is matched with the Levy-Jennings chart prepared (as described above).

Interpretation

If the 'E' ratio of the external control falls beyond 2 SD limit of mean in Levy-Jennings chart, the run of EIA is treated
as quality control failed or invalid even though the internal kit control values are within acceptable limits. On the other
hand, if it falls within the limit mentioned above, the run is considered as valid.

Variations in the quality control validation guidelines

Some authorities recommend inclusion of the borderline external control in duplicate in each run. A run is considered
as invalid if (i) both external control values are beyond the Mean 2SD limit or (ii) one of external control values is
beyond 3 SD limit or (iii) the external control value exceeds the standard deviations in two consecutive runs. While
adding the duplicate external controls, it is recommended that one is charged into the well in the first vertical row
immediately after dispensing the internal kit controls, while the other is charged in the last vertical row randomly. Care
is taken to add the duplicate controls as per the normal sequence of charging the specimens in the wells i.e. the first of
the duplicate control is charged in the beginning while the second one is charged at the end along with the test
specimens which are being charged in the last row. This procedure is likely to take care of the variation of values due to
time delay of operation of various steps between the Ist row and last row.

41
According to some authors, setting up the acceptability limit as Mean 2 SD is likely to lead to
perceptible degree of rejection due to random error. Hence, they recommend the extension of the
limit of SD to + 2.5 instead of 2 with the assumption that 5% or less of any set of control values will fall
outside the range due to chance.

A relatively more stringent measure of quality control recommends incorporation of external controls
(or even internal controls) in triplicate at 3 random wells in the ELISA plate in each run to check intra-
run variation.

Importance of 'E' ratio over OD value of external controls to check validity of runs

There are two methods by which the reference QC graph may be represented.

The first one is to plot the OD values of the external control on the Y axis and the consecutive run members (i.e. run
dates in case one plate is run per day) on the X axis. The second option is to plot the 'E' ratios i.e. ratio between the OD
value of external control and the cut off value on the Y axis instead of the OD value of the external control alone, the
determinants on the X and Y axis remaining as in the first method.

The second method is considered to be more accurate alternative since any general change in the conditions during
the run of EIA on a specific day/run will affect the OD value of the external control as well as kit controls towards a
similar direction (explained in details previously in section 2). This is exemplified in the following example of six EIA
runs on consecutive dates using permissible limit of (i.e. mean 2SD) in the values of 20 runs mentioned in example
for setting the limit of acceptability of external controls for validity of runs.

The acceptability limits of external controls for validation of test results calculated as below (based on OD values of 20
runs above).

Mean (X) of OD = 0.428


SD of OD = 0.055

Limit of Mean 1SD = 0.373 - 0.483


Limit of Mean 2SD = 0.318 - 0.538

On the other hand, as shown in Table -4, the following values are obtained if the 'E' ratios in the same runs of EIA are
taken into consideration to validate the run.

Mean (X) of 'E' ratio = 1.90


SD of 'E' ratio = 0.19
Limit of Mean 1SD = 1.71 - 2.09
Limit of Mean 2SD = 1.52 - 2.28

Now, let it be presumed that the following six consecutive runs of EIA are evaluated for their validity based on the
performance of external control Table 5.

42
Table - 5 Recording of OD values in six consecutive days of test runs.

It can be observed from Figure - 6 and 7 that when the OD values of the external controls on different days of test run
are matched against the acceptability limit (mean 2SD) drawn on the basis of OD values of the external control
alone, the run on day 2 appears to be invalid since the observed OD value on that particular day (i.e. day 2) lies beyond
acceptability limit i.e. more than mean 2 SD limit or 0.538). On the other hand, if the corresponding 'E' ratios are
matched against the acceptability limit (i.e. mean 2SD ) drawn on the basis of 'E' ratios of the external control, the run
on day 2 appears to be within the acceptability limit, while the run on day 4 appears to be invalid (Figures - 6 and 7.).

Fig. 6 Validity of runs evaluated on the basis of OD values of the external control alone.
Cut off OD values are shown at the bottom as broken line

Fig. 7 Validity of runs evaluated on the basis of ELISA ratios (E ratio i.e. external control OD/
cut off OD) of the external control.

If the OD values are further examined for the runs in question, it is clear that, on day 2 although the OD value of the
external control went up, there was some degree of rise in cut off OD value as well. Thus the net change in the ratio
between the external control sample OD and cut off OD (i.e. 'E ratio) did not fall outside the acceptability range when

43
matched in terms of 'E' ratio. On the other hand, on day 4 of run, there was some degree of elevation of external control
OD value while cut off value did not show any such change rather it showed change on reverse side which is not as per
expectation (as explained in earlier section). Thus there is a rational ground to consider the run on day 4 as invalid
Table - 3 & 4, Figure - 6 and 7

Fig. 8 Interpretation of aberrant results of QC: shift and trend


Illustrations of shift and trends

Interpretation of aberrant results of QC: shift and trends (Figure - 8)


Control values of six consecutive runs fall on one side of the mean line.
Can be due to the following:
Switching to a new lot of kits
New reagents
Changes in incubation temperature
New technician

Trend

Six successive points distributed in one general direction (towards either higher result or lower result) within the
acceptable range.

Can be due to the following:


Deteriorations of reagents
Slowly faltering equipments e.g. pipette

Actions taken for aberrant QC readings include the following checks:

The date the aliquot of external QC sample in use was opened is checked. If aberrant results are
enumerated in 3 consequent runs the aliquot in use is discarded and a new vial of QC sample is
opened. Both aliquots (new and discarded aliquot) are tested on the next run to verify the QC
sample's deterioration.
The storage conditions of the QC sample: The aliquot in use can be stored at 4 C for 1 week, all other
aliquots must be stored at -20C. No QC sample should be freeze/thawed and refrozen.

44
The date of expiry of kit/reagents and/or any indications of deterioration/or contamination.
The performance of the operator of the test to ensure the assay is being performed correctly.
Whether or not the aberrant results were produced with a particular batch of kit.

Further actions to be taken may include

Checking equipment (e.g. washer, incubator, reader).


Checking the kit and reagents more thoroughly for deterioration or contamination.

Flow CHART FOR PERFORMING HIV TESTING USING QUALITY CONTROL ALIQUOT
Performance of EIA with external control (HIV)

45
FLOW CHART FOR PERFORMING RAPID HIV TESTING USING QUALITY
CONTROL ALIQUOT

Selection of Rapid test * Licensed

* Quality checked

Performance of Rapid HIV test * As per manufacturers guidelines

* Internal controls are checked

Inclusion of external control in * Include one positive external control


daily run daily

Examine cartridge, comb, etc. * Examine for either coloured


for result dot/dots, or coloured line/lines
and/or agglutination

Validation of rapid test result * Ig spot should always give positive


result.
Positive control should give positive
result (colored dot, line and/or
agglutination

Test result invalid * -Ig spot not seen


-Positive control result does not
show as positive

46
CHAPTER 7
COLLECTION, TRANSPORT AND STORAGE OF
SPECIMENS FOR HIV TESTING

Introduction

Almost all laboratory procedures for HIV testing are performed on patients' blood; serum or plasma; hence the
collection of blood is described below.

Performing venipuncture:

Gloves should be worn and sterilised /disposable syringes and needles should be used.

For avoiding soiling, a piece of linen with a layer of dressing pad ( a sheet of absorbent cotton
between two layers of gauze piece) or simply a big piece of absorbent cotton may be placed below
the forearm before commencing veni-puncture.

After collecting 3- 5 ml of blood aseptically, it should be carefully transferred from the syringe without
squirting into a sterile plastic leak proof specimen container preferably screw capped. The containers
should be labelled before commencement of venipuncture. The cap may be tightly screwed after the
blood has been transferred to the vial.

After blood is collected, the tourniquet is removed and the needle is withdrawn. The patient is given a
dry sterile cotton swab to press over the site of venipuncture. Elbow may be flexed to keep the cotton
swab in place till the blood stops. Any blood spill is carefully wiped with 70% ethanol/10% bleach
solution.

All the swabs and cotton pieces are placed in plastic bags for disposal. If the outside of the vial is
visibly contaminated with blood, it should be cleaned with 10% freshly prepared sodium hypochlorite
solution.

The blood is allowed to clot for 30 minutes (not more than 2 hours) at room temperature. The clot may
be gently broken if necessary using sterile pasteur pipettes.

Alternatively vacuum based blood collection systems (vacutainers) can also be used. These vacuum
based collections are relatively safe for usage and harbour minimum risk of unwanted exposure to
infected blood.

Separation of sera samples:

- The vial / vacutainer should be centrifuged at 3000 rpm for 10 minutes to separate serum to avoid
haemolysis. If no centrifuge is available, the blood with clot may be left overnight in the refrigerator at
4C.The clot will retract and get separated from serum.

47
The specimen vial is un-stoppered, the serum is drawn off by sterile Pasteur pipette and transferred
to a sterile plastic screw capped leak proof tube.

Addition of preservative:

The usual preservative should not be added since it inactivates conjugates and gives rise to false
serological results.

If necessary, 5 bromo, 5 nitro, 1-3 dioxane in propylene glycol at a final concentration of 0.05% is
recommended as preservative.

Thiomersal at a final concentration of 0.01% is effective only for a few weeks as it loses activity when
exposed to light.

Storage of serum specimens:


The sera samples are placed in leak proof plastic containers in the refrigerator at 2-8C, for upto
48 hours for storage. The storage of samples should be as per test requirements.
The outside of the container is checked for visible contamination with blood which should be cleaned.
All the specimen vials must be adequately labelled with patient details.
Then the specimen vials are packed in a second tightly capped unbreakable container
surrounded by adequate packing material (see figure 1 ahead).

For storage for a longer time, specimens must be frozen at - 200C or deep-freezing at-70C is advised,
in labs where deep freezers are available.

Sample transportation
These instructions are recommended for specimen transportation:

The shipment of infectious agents is regulated by the Transportation of Dangerous Goods Act and the International Air
Transport Association (IATA) dangerous goods regulations. HIV infected specimens are classified as infectious class
6.2 substances under the United Nations (UN) no. 2814. The packaging must adhere to UN class 6.2 specifications.
Packaging requires a 3 layer system as described below (see Fig. 1 for a diagrammatic representation):

The specimen tube, in which serum is to be transported, should not have cracks / leakage. It should
preferably be made of plastic and be screw capped. The outside of the container should be checked
for any visible contamination with blood which should be disinfected.

Place the tube containing the specimen in a leak-proof container (e.g. a sealed plastic bag with a zip
lock or alternatively the bag may be stapled and taped) and pack this container inside a cardboard
canister / box containing sufficient material (cotton gauze) to absorb all the blood should the tube
break or leak.

Cap the canister/box tightly.

Fasten the request slip securely to the outside of this canister. This request slip should have all details
i.e. name, age , sex , risk factors, history of previous testing, etc. and should accompany the
specimen. The request slip should be placed in a plastic ziplock bag to prevent smudging on
accountsof spillage.
48
For mailing, this canister/box should be placed inside another box containing the mailing label and
biohazard sign.

The diagram ( fig 1) depicts the method of sample transport for a single/ few ( 2-3) samples that could fit into
the secondary container shown in the diagram. The size of the primary sample container will vary with the
number of samples being transported. For a larger number of samples ,a tube rack (or some such container )
may be used wherein the samples can be transported in the upright position and at appropriate temperature.
The packaging instructions for the transport of a larger number of samples are given below:
The specimen should be carefully packaged to protect it from breakage and insulated from extreme
temperature
Label appropriately and mention the test/s being requested for that sample. The collection site should make
use of a unique identification number as sample identity. Names of the patients should be avoided to prevent
confusion on account of duplication of names as well as to maintain confidentiality.
Secure the vacutainer cap carefully and seal it further with sticking tape ( placed so that it covers the lower part
of the cap and some part of the tube stem.
During packaging, the tubes containing specimens should be placed in a tube rack and packed inside a cool
box (plastic or thermocol) with cool/ refrigerated / frozen gel packs ( as appropriate to keep the sample at the
recommended temperature for the test ) placed below and on the sides of the tube rack. Place some cotton or
other packaging material between the tubes to ensure that they do not move or rattle while in transit.. Cool
box required for transportation could be a plastic bread box or a vaccine carrier. Seal/secure the lid of the cool
box .
This cool box should then be placed in a secure transport bag for purposes of shipping to the testing facility.
The request slips should be placed in a plastic zip lock bag and fastened securely to the outside of the cool
box with a rubber band and sticking tape.
A biohazard label should be pasted on the visible outer surface of the package containing the samples. The
package must be marked with arrows indicating the 'up' and 'down'side of the package
Samples should be transported to the receiving laboratory by commercial courier or be hand delivered by a
trained delivery person.
The collection site must have prior knowledge of the designated testing days of the laboratory to which the
samples are being sent.
No transport should be done during weekends and holidays or non-testing days of the testing laboratory
unless prior arrangement has been made with the receiving laboratory.

Note: Use overnight carriers with an established record of consistent overnight delivery to ensure arrival of specimen
within the specified time.

Figure 1 Packaging of specimen for transport to the laboratory.

49
Safe Handling and disposal of sharps:

Extreme care should be used to avoid auto-inoculation.

All chipped or cracked glassware should be discarded in appropriate containers.

Broken glass should be picked up with a brush and pan. Bare hands must never be used.

The disposable needles should never be manipulated, bent, broken, recapped or removed from
syringes.

The used sharps should never be passed directly from one person to another.

One should always dispose of his/her own sharps.

Used needles should be discarded in puncture-proof rigid containers (plastic or cardboard boxes)
after disinfection in 0.5-1% freshly prepared sodium hypochlorite solution (common bleach) and
never in other waste containers. If a needle shredder/destroyer is available, only the needles or the
needles along with syringe nozzle may be shredded depending upon the type of the shredder

Sharp disposable containers should be located close to the point of use.

Sharp disposal containers should be sent for disposal when three-fourth full.

In case suitable means of disposal of syringe / needles are not available, these disposable syringes
should be heated in dry ovens and be allowed to mutilate to prevent recycling of plastic syringes.
Needles can be incinerated.

IN CASE OF DAMAGE OR LEAKAGE


IMMEDIATELY NOTIFY
PUBLIC HEALTH
AUTHORITY

50
CHAPTER 8
STANDARD OPERATIVE PROCEDURES

Standard operative procedure (SOP) must be prepared for each activity and all tests being conducted in the
laboratory. The SOP should give detailed instructions on the following:
Name and scope of the test
Name of the kit being used
Name of the procedure (ELISA, Rapid Test)
Collection, transport and management of the specimen.
Storage and inventory (logging in the specimen and storing for reference or retesting).
Test requisition: the information required to be collected on the sample which may be relevant and
have implications for interpretation of the test.
Details of environmental requirements (temperature, humidity etc.)
Details of performance of test-each step of sample processing to arrive at the result.
Instructions for quality control practices, types of internal (kit) and external controls required to be
processed for validation of the test run.
Instructions regarding the interpretation of the test.
Instructions on use of the algorithm as per the policy.
Instructions on recording the test result.
Instructions on reporting the test result, whom to report and how to report.

SOP should be available at testing site and should be followed every time the test is performed.
The SOP must be simple to follow and should be in the form of flow diagram / chart, one step leading into the next and
easy to interpret.
SOP must also include all the pre-analytic, analytic and post-analytic requirements for performing the test keeping the
quality issue in view.

A sample SOP
Name of the test procedure.
Intended use: detection of HIV-1 and HIV-2 antibodies.
Principle of the test procedure: details as per the manufacturer's kit insert have to be given.
Details of the kit contents like conjugates, negative and positive controls, plate / cartridge / comb etc.
List the materials which are required for performance of the test but are not provided in the kit.
Storage instructions.

51
The precautions to be taken while performing the test, standard work precautions and any other
precautions specific to the test.
Specimen collection and storage of whole blood, serum or plasma; How the serum / plasma should
be separated; How the specimen is to be used and how it should be stored.
Quality controls required to be put to validate the tests.
Details of the test procedure as per the instructions in the kit insert.
Details about how to validate and interpret the results as per instructions of the kit insert.
Limitations of the test in detecting infection.
Standard work precautions.
Details about disposal of left over samples and other wastes generated.
Details about recording the result and reporting.

52
CHAPTER 9
DOCUMENTATION AND DOCUMENT CONTROL

Introduction

Quality management system documents are extremely important to ensure quality practices in the laboratory. All
instructions, processes, procedures must be documented; the staff must be aware about the documents and
understand how the documents work. Documents must be available to the staff whenever they are needed. The staff
must follow the instructions provided in the documents. Since, the functions, processes and procedures of
laboratories are constantly changing, so the documents should be revised constantly as required.

Verbal instructions are very unreliable as a means of communication as these may not be heard, may be
misunderstood, may be ignored or may be quickly forgotten particularly so if it is complex technical information.

The documents required for smooth functioning and accreditation of a laboratory include:

Quality manual: This documents and gives details of the quality policy, the quality system structure
and the key areas of activity. It gives an overview of all the functions carried out by the laboratory and
provides link to all other documentation.
Procedure documents: gives details of processes and each procedure being performed in the
laboratory documented in the form of standard operative procedures (SOP). SOP must be easily
accessible and available to the staff. It also gives details of assignment of various procedures to the
staff.
Work instructions: These are to be specified for each staff member and should be clear and
understandable by the staff - what to do and how to do it? It also describes how specific tasks
referenced in procedures are to be carried out. The jobs for each staff are defined and documented
for optimal and quality functioning and to avoid conflicts.
Forms: This should include all types of forms being used by an organization / laboratory to perform
the required jobs, collect the relevant information important for performing a test / job, analysis and
interpretation of results, etc. Sample collection forms, report forms, indent forms, are some of the
examples of the forms.
Records: A number of records are generated in the course of activities of the organization /
laboratory. These records must be maintained for future reference, audit and legal purposes, etc.
These records include details of personnel, training, results, etc.

Documents need to be numbered and filed. Documents need to be revised/changed as and when required e.g. when
a procedure is changed, the documents i.e. work instructions and work-sheets need to be changed accordingly. The
documents should be designed in such a way that they are easy to follow, legible and serve the required purpose. The
aim is that the document should work well for communicating information.

The current documents should be available & accessible for use. The old documents should be archived and not
destroyed. This is important to provide an accurate historical record. For example, if a modification is made to a
laboratory method, it is important to know what was the modification, who made the modification and when was the
modification made. The revised/changed document must be shared/ communicated to the staff.
53
Types of documents

Customer complaint document.


Corrective and preventive action document.
Equipment maintenance, calibration, and monitoring document (details in chapter on equipment
maintenance and calibration).
Equipment inventory document.
Reagent inventory document.
Safety manual.
List of suppliers of equipment, reagents and other materials used in the laboratory.
Documents detailing the contracts to perform special activities (e.g. research projects, etc). Samples
of contracts to be stored.
Documents on standards being used by the organization.

For example, corrective and prevention action document will detail what preventive action needs to be taken to avoid
breakdown of an equipment. Preventive actions are defined as the proactive processes to identify the potential
sources of non-conformance. This is discussed in details in the chapter on equipment maintenance and calibration.
The corrective actions taken to repair equipment, improve a procedure, improve performance, etc. must also be
documented.

Example
Inventory on equipment

Record the unique identification facts of each equipment (Manufacturers serial no. or lab number).
Maintain record of name, number, date received and condition upon installation, maintenance
history, future maintenance policy - all are documented in the equipment inventory document.
Maintain records of service of equipment and calibration of equipment.
There must be a document giving the instructions for proper use, maintenance, and frequency of
servicing required.
Maintain details of calibration of equipment like name of equipment, date of calibration, name of
person who calibrated.

Document control

Certain documents need to be controlled i.e. such documents may not be accessible to everyone in the lab. All such
documents which need to be controlled should be clearly specified and the accessibility must also be defined. These
may include financial records of the lab and all those documents which the management feels are essential to be
controlled. At the same time, control of documents should be exercised in a manner that it does not interfere with the
quality management system of the lab and should not affect the quality of the result.

54
CHAPTER 10
SELECTION OF HIV TESTING KITS

Introduction

Selection of the appropriate HIV test kits is an essential prerequisite for accuracy and reproducibility of test results and
for quality management. A large number of assays are available commercially, offering essential as well as attractive
performance characteristics namely sensitivity, specificity, efficiency and predictive values. However, the
performance of the test will depend upon the conditions in the laboratory, technical expertise, population being tested,
etc.

An endeavour should be made to selects kits which are evaluated and approved by WHO and DCGI. The
characteristics of the selected kits should fulfil the objectives of testing. The kits as far as possible should be user
friendly.

The performance characteristics to be considered for the selection of kits are as under:
Sensitivity
Specificity
Efficiency
Positive predictive value
Negative predictive value
Stability
Ease of performance of test.

The parameter's values should be within those recommended by the technical expert committee, G.O.I. on this
subject. The final approval of the test kit is given by DCGI on the basis of the evaluation reports submitted by the
identified evaluation centres

Sensitivity: Sensitivity of a test is defined as its ability to detect truly infected individuals as also its ability
to detect very small amounts of analyte.
Specificity: This is the ability of a test to correctly identify all the uninfected individuals i.e. there should be
no false positives.
Efficiency: It is the overall ability of a test to correctly identify all positives as positive and all negatives as
negative.
Predictive values: This is the measure of value of a test in relation to the prevalence of the disease in the
population. The value of a test in addition to the parameters described above depends on the
population being tested.

Positive predictive value (PPV) is the ability of the test to identify actually infected individuals among all persons giving
a positive result with the kit being evaluated.

Negative predictive value (NPV) is the ability of the test to correctly identify the really non infected individuals from
among all the persons giving negative result.

55
If the prevalence is high, the PPV of a test will be high i.e. an individual testing positive will mean actual infection.
Whereas, in low prevalence area the chance that an individual testing positive is actually infected is lower.

RECOMMENDATIONS FOR SELECTION OF TEST KITS

The testing centre should procure and have available at all times, HIV test kits based on three different principles and
/or antigen systems to enable them to perform HIV testing as per NACOs HIV testing strategies at VCTCs and
PPTCTCs and one highly sensitive test kit (100%) to screen blood at blood banks to ensure blood safety.

HIV (ELISA) test kits - general specifications


1. Should be solid phase microplate ELISA using HIV I & II recombinant and/or synthetic peptide
antigens.
2. The assay should detect HIV I and II antibodies.
3. The assay should detect antibodies to all sub-types of HIV I.
4. The assay should be able to detect antibodies of HIV I and II during early seroconversion period.
5. The assay should have reactive and non-reactive controls with each kit.
6. The kit should have a shelf life of minimal 12 months at the port of discharge or consignees end which
ever is applicable.
7. Adequate literature detailing the components, methodologies, validity criteria, performance
characteristics, storage conditions, manufacturing and expiry dates should be provided with each kit.
8. The assay should have sensitivity level at 99.8% and above and specificity level at 98% and above.
9. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and
transport at 20C - 80C.

HIV test kits (sandwich ELISA)


1. Should be solid phase microplate sandwich ELISA using HIV I & II recombinant and/or synthetic
peptide antigens

The selection criteria 2-9 are identical to those applicable as above for HIV (ELISA) test its general specifications

HIV (ELISA) capture/competitive test kits


1. Solid phase microplate capture ELISA in which solid phase is coated with anti-immunoglobin (anti
IgG, IgM, IgA).
2. If assay is competitive ELISA, the solid phase should be coated with recombinant and/or synthetic
peptide antigen.

The selection criteria 3-9 are identical to those applicable as above for General Specifications

HIV Rapid test kits-general specifications


1. Should be solid phase/particle coated with synthetic/recombinant HIV I & HIV-II antigens.
2. The assay should detect antibodies to HIV I & HIV II by Enzyme Immuno Assay / Agglutination/
Immunochromatography/any other principle.
3. The product should be able to detect antibodies to HIV I and II during early sero-conversion period.
4. The product should have positive and negative controls.
5. The kit should have a shelf life of minimal 12 (twelve) months at the port of discharge or consignees
end which ever is applicable.
56
6. Adequate literature dealing the components, methodologies, validity criteria and performance
characteristics of the product should be provided with each kit.
7. Should have sensitivity level at 99.8% and above and specificity level at 98% and above.
8. The supplier/local agent should have facility for storage of kits at 2C - 8C
9. The total procedure time should not be more than 30 minutes.
10. Provision shall be made for conducting single test at a time.
11. The tests should be easy to perform and the kits should preferably be stable at room temperature (22
25C.

HIV Rapid test kits- principle of Enzyme Immuno Assay


1. Should be a Solid phase/particle coated with synthetic/recombinant HIV I & HIV II antigens.
2. The assay should detect HIV I and II antibodies by principle of Enzyme Immuno Assay.

The selection criteria 3-11 are identical to those applicable as above for HIV Rapid test kits (general specifications)

HIV Rapid test kits -principle of agglutination


1. Should be solid phase/particle coated with synthetic/recombinant HIV I & HIV II antigens.
2. The assay should detect HIV I and II antibodies by agglutination.

The selection criteria 3-11 are identical to those applicable as above for HIV Rapid test kits (general considerations)

HIV Rapid test kits - any other principle excluding Enzyme Immuno Assay and
Agglutination
1. Should be solid phase/particle coated with synthetic/recombinant HIV I & HIV-II antigens.
2. The assay should detect antibodies to HIV I & HIV II by any other principle excluding Enzyme Immuno
Assay and Agglutination.

The selection criteria 3-10 are identical to those applicable as above for HIV Rapid test kits-general specifications

DETAILS OF SOME OF THE COMMERCIALLY AVAILABLE HIV RAPID TEST KITS

Immunofiltration*

Immunoassay

57
Agglutination

Different manufacturers use different raw materials as matrix for adsorbing the HIV peptides. The immunofiltration
involves the vertical flow of the reagents. As the reagents pass through the matrix the reactants get adsorbed and
concentrated (immunoconcentration) on the viral peptides on the matrix in case of HIV positive sample to produce
a colored spot /line

ELISA TESTS

58
Combinations as per principle

I. Immunofiltration + Immunoassay + Agglutination


II. Immunofiltration + Immunoassay + Lateral Flow/ Immuno chromatography
III. Immunofiltration + Agglutination + Lateral Flow/ Immuno chromatography
IV. Immunoassay + Agglutination+ Lateral Flow/ Immuno chromatography

Examples

I. Tridot/Pareekshak/ HIV Spot + Immunocomb HIV 1& 2Bispot/HIV EIA Comb/CombAIDS RS +


Capillus/NEVA
II. Tridot/Pareekshak/ HIV Spot + Immunocomb HIV 1& 2Bispot/HIV EIA Comb/CombAIDS RS + SD
Bioline HIV 3.0/ Retrocheck HIV/ Precise
III. Tridot/Pareekshak/ HIV Spot + Capillus/NEVA + SD Bioline HIV 3.0/ Retrocheck HIV/ Precise
IV. Immunocomb HIV 1& 2Bispot/HIV EIA Comb/CombAIDS RS + Capillus/NEVA + SD Bioline HIV
3.0/ Retrocheck HIV/ Precise

Care has to be taken that atleast two kits selected should be able to differentiate between HIV 1 & 2
For example: Do not select Pareekshak + Comb AIDS RS + Capillus as none of these kits can differentiate between
HIV 1 & 2, instead Tridot + HIV EIA Comb + Capillus can be selected

Combinations as per antigen

All the kits have recombinant antigens except Immunocomb & HIV 1& 2 Bispot, which have synthetic antigen.
CombAIDS RS has a combination of recombinant & synthetic peptides. To avoid confusion it will be better to use kits
with different principles rather than different antigens. But in dire situations, a combination of two immunoassays
having different antigens can be used along with one more test of other principle.

59
CHAPTER 11
EVALUATION OF HIV KITS

Introduction

Periodic evaluation of HIV test kits is a prerequisite of good quality management. A large number of assays are
available commercially each offering essential as well as attractive performance characteristics namely sensitivity,
specificity, efficiency and predictive values. However, the performance of the test will depend upon the conditions in
the laboratory, technical expertise, population being tested, etc. So, wherever possible, evaluation of kits should be
done under the same conditions and in the same locale using the local panel of sera to assess the performance
characteristics and efficiency of a test kit. New and improved kits are also being developed continuously and these
also need evaluation in the same way as the established kits. Reports of evaluation of kits provide a feed back to the
manufacturers to improve their products so that the kits may be fit for use under different situations and in different
geographical areas. To have an idea about the performance characteristics of these HIV kits, prior evaluation by some
notified agency should be made and is mandatory. Situations under which these kits may have to be evaluated can be
as follows:
Kits under import; evaluation with local serum panel is very important.
New kits developed/manufactured indigenously; to assess efficacy.
To resolve controversies regarding discordant results or deterioration of components and reagents.
This may be due to break in cold chain during transit and/or storage of in-use kits

As far as the first category is concerned DCGI asks the importer to get the representative sample of kits evaluated,
using a locally made and well characterized panel by the Institutes identified for this purpose. Currently NIB is the
statutory institute which perform evaluations. If found suitable, only then the kits should be allowed to be imported for
distribution, both in Government as well as private sector.

In second case, when the kits are manufactured indigenously, the performance characteristics of the kits should be
ascertained by evaluating different batches of kits (the final product) with the standard serum panel for evaluation of
kits at least at two to three different geographical sites and consistent results should be obtained. Only the kits
performing to the standards (sensitivity and specificity, etc.) as laid down should be licensed for marketing.

In the third case, the licensed, and/or approved kit's performance may be questioned particularly when different
centers report inconsistent results. This can happen due to break in cold chain during transport and storage of HIV kits.
To resolve this problem the kits have to be evaluated by the identified centers to ensure optimal performance.
This chapter gives the basic guidelines to be followed to ensure proper evaluation of HIV test kits.

Kit evaluation process

A process must be followed to perform evaluation of kits.


The process comprises of:
Gathering information on the kit to be evaluated from the manufacturer
Examining the available data on the kit.
Writing an evaluation protocol.
Selecting an evaluation panel.

60
Performing the testing
Collating and analyzing the data.

Information about the performance characteristics (sensitivity, specificity, predictive values, etc.) of the kit is
ascertained from the manufacturer. The information helps to select the right type of evaluation panel. The evaluations
can be done by a single site or multiple sites. In case of multi-site evaluations, a QC sample should be included to
assess inter-laboratory variations.

Evaluation panel

A serum panel is defined as a collection of well characterized serum samples for a specific reference/purpose.
Serum panel can be prepared and used for different purposes. These are:
Panel for evaluation of kits i.e. evaluation panel including seroconversion panel
Panel for proficiency testing (external quality assurance)
Panel for quality control in the laboratory (internal quality control)

Volume of panel samples

The volume of serum being processed to prepare panel should be sufficient to meet the requirements for at least 2-3
years for the purpose for which the panel is prepared. For example 6 ml. of serum will be sufficient to evaluate
approximately 100 kits.

Collection of specimen for panel

Ideally material used for panel should be serum although plasma can also be used. However, plasma tends to be
unsuitable as it clots spontaneously on storage. Therefore plasma should be defibrinised with thrombin or recalcified
and then processed for making the panel.

Aliquoting of serum

Repeat freezing and thawing of specimens affects the quality of the panel. To avoid this, samples should be aliquoted
in volumes that may meet the particular panel requirement at one time. Suitable anti-bacterial agents e.g. Bronidox
(0.05%) should be added to panel samples to prevent contamination.

Characterization of serum panel

Serum samples should be well characterized so that there should not be any doubt about the final outcome of the test
results. Specimens should be tested using high quality HIV antibody ELISA or any other screening assay followed by
confirmatory Western Blot assay to determine their true status and should also distinguish between HIV-1 and HIV - 2
types. As characterization of each unit in the panel is quite expensive, use of large volume of each specimen will be
more cost effective.

Storage

Aliquots must be stored at or below -70C. Glass vials should not be used as they may eventually crack with long term
storage in frozen condition.
61
Transport of panel

To provide safety for laboratory staff, couriers and community, specimens should be transported under International
Air Transport Association (IATA) regulation, for this purpose. Specimens should be packed, labeled and documented
as classed under 6.2 UN 2814 to minimize the risk of infection to anyone during transit.

Composition of serum panel

These are well characterized serum samples, those which have given consistent results with all the available HIV test
kits. These sera should be from individuals in whom the true HIV status (infected or not infected) is known with high
degree of certainity. Such samples are called pedigreed sera. Panel must include reactive, nonreactive, weakly
reactive (grey zone) and indeterminate classified sera. Both HIV-1 and HIV-2 reactive sera should be included in the
panel. The panel should be representative of the population/area where the kits are going to be used.

To prepare a local panel, sera should be tested with all the available kits.

At least 50 specimens should be included in each panel of reactive, weakly reactive, indeterminate
seroconversion and non reactive. The panel size can be increased as more the number of sera in
each panel, more statistically valid is the comparison.
Non reactive panel should also include representative sera from individuals suffering from local
endemic diseases (e.g. malaria, tuberculosis, hypergammaglobulinaemia, etc.) and sera with high
IgM levels.
Panel specimens should be appropriately processed/characterized and stored as aliquots in frozen
condition. All specimens should be in the same condition as far as storage is concerned when test for
evaluation of kits is performed.
Specimens should not be haemolysed, lipaemic and turbid etc.

Reference test

A well established, reference gold standard test must be available to the laboratory for comparison, the results of
which should correlate with the actual status of the patient. Currently, such a test in serologic testing for HIV infection is
considered to be Western blot (WB) test. However, sometimes WB may give indeterminate results in which case more
sophisticated tests like synthetic peptide based line immunoassay, polymerase chain reaction (PCR) and virus culture
can be undertaken. The tests are expensive, labor intensive and require higher level of technical expertise. Thus for
the moment WB usually is regarded as the Reference/Gold standard test. When a number of kits are being evaluated
the tie breaker for false positive/false negative is usually a WB test, sometimes supported by more sophisticated tests
as mentioned above.

Testing condition

All evaluations must be performed under identical conditions. The tests must be conducted in the same laboratory
using same equipment (pipettes, instruments, etc.), by the same technician, using kits which have not expired and
have been stored properly and using the exact protocols provided in the package insert.

62
The technicians must not know the results obtained on the panel with other kits i.e. the evaluation testing is done
blindly.

Parameters for performance characteristics of a kit

The parameters to be calculated to assess the performance characteristics of the kits include:
Sensitivity
Specificity
Efficiency
Delta values
Positive predictive value
Negative predictive value

Sensitivity

Sensitivity of a test is defined as its ability to detect truly infected individuals as also its ability to detect very small
amounts of analyte. Sensitivity can be calculated by the following formula:
True positives
Sensitivity = ----------------------------------------------------- X 100
True positives + false negatives

Or

Number of samples detected as positives by the kit under evaluation


Sensitivity = ----------------------------------------------------------------------------------------------X 100
Total number of positives by the reference test employed

Example
Total sera tested= 100
HIV infected = 10 Results with the reference test
HIV noninfected = 90

Results with kit under evaluation: 9 out 10 HIV infected detected as positive
9
Sensitivity of the kit under evaluation will equal to --------- X 100 = 90%
9+1
Specificity

This is the ability of an assay to correctly identify all the uninfected individuals i.e. there should be no false positives
(reactive result in the absence of disease).

The specificity can be calculated by the following formula:

True negatives
Specificity = ------------------------------------------------ X 100
True negatives + False positives

63
Or
Number detected as negatives by the kit under evaluation
Specificity = -------------------------------------------------------------------------------------- X 100
Total number of negatives detected by reference test employed

Example
Total sera tested= 100
HIV infected = 10 Results with the reference test
HIV noninfected = 90

The results obtained with kit under evaluation, 9 positive out of 10 HIV infected and 1 positive out of 90 HIV
noninfected.
89
Specificity of the kit under evaluation will equal to -------------X 100 = 98.9 % (approximately)
89+1
Efficiency

It is the overall ability of a test to correctly identify all positives as positive and all negatives as negative. Efficiency is
determined as below:

True positives + True negatives


Efficiency = ----------------------------------------------------------------------X 100
True positives + False negatives + True negatives + False positive

Or

Total number detected as positives + negatives by test kit


Efficiency = ---------------------------------------------------------------------------------------------------- X 100
Total number of positives + negatives detected by reference test employed

Example
Total sera tested= 100
HIV infected = 10 Results with the reference test
HIV noninfected = 90

The kit being evaluated detected 9 out of 10 HIV infected as positive and one out 90 non-infected also as positive.
9+ 89
So efficiency of the kit will equal to ------------------- X 100 = 98%
9+1+89+1
Delta value

This is a statistical measure of sensitivity and specificity and is calculated from the actual OD values generated by the
test. This also helps to determine how close the O.D. value of false positive and or false negative is to the real positive
and or real negative value, respectively. If these values are too close the test is flawed.

Delta values are defined as the distance of the mean O.D. ratio of the sample population from the cut off (C.O.)
measured in standard deviation units.

64
Mean O.D. / Cut off ratio (log 10)
Delta values = -------------------------------------------------------- X 100
Standard deviation

Positive delta value is calculated on mean of ODs of positive samples and negative delta value is calculated on mean
ODs of negative serum samples. When two tests are being compared, the test that has a higher positive delta value
would characterize positive samples more accurately; similarly the test with higher negative delta value will better
classify the negative samples.

Example

Positive predictive values

This is the measure of value of a test in relation to the prevalence of the disease in the population. The value of a test in
addition to the parameters described above depends on the population being tested.

Positive predictive value (PPV) is the ability of the test to identify actually infected individuals among all persons giving
a positive result with the kit being evaluated.

True positives
PPV = -------------------------------------------------- X 100
True positives + False positives

Negative predictive value (NPV) is the ability of the test to correctly identify the actually non infected individuals from
among all the persons giving negative result.

True negatives
NPV = -------------------------------------------------- X 100
True negatives + False negatives

If the prevalence is high the PPV of a test will be high i.e. an individual testing positive will mean actual infection.
Whereas, in low prevalence area the chance that an individual testing positive is actually infected is lower.

Example

Low prevalence area 1%


Total sera tested = 1000
HIV infected = 10 with the reference test
HIV noninfected = 990

65
Results obtained with kit under evaluation: Positive results obtained in 8 individuals, two from HIV infected group and
6 false positives from HIV non infected group.

Negative results obtained in 992 individuals, 984 from the non infected group and 6 false negatives from HIV infected
group.

2
The PPV = ------- X 100 = 25%
8

984
The NPV = ------------- X 100 = 99.3%
992
Example

High Prevalence area 10%


Total sera tested = 1000
HIV infected = 100 with the reference test
HIV noninfected = 900

Results obtained with HIV kit under evaluation; 105 tested positive, 95 from HIV infected group and 10 from HIV non
infected group i.e. gave false positive result. Negative result was obtained in a total of 900 cases. 892 of 900 were from
HIV non infected group and 8 were from HIV infected group.

95
PPV = --------------- X 100 = 90.5% (approx.)
105

892
NPV = --------------- X 100 = 99.12%
900

All the above parameters are ascertained by the laboratory carrying out the evaluation of kits. The parameter values
should be within those recommended by the Technical Expert Committee, G.O.I. on this subject. The final approval of
the test kit is given on the basis of the evaluation reports submitted by the identified evaluation centers. For further
information on test kit evaluation, see also: Guidelines for Assuring the Accuracy and Reliability of HIV Rapid Testing:
Applying a Quality System Approach. 2005. Centers for Disease Control/WHO.

66
CHAPTER 12
OPERATION, MAINTENANCE
AND CALIBRATION OF EQUIPMENT

Introduction

Preventive maintenance is the best measure to ensure that the equipment will function properly. Preventive
maintenance schedules should be followed regularly for optimal performance of equipment. Ordinarily, maintenance
schedules are set up as tasks to perform daily, weekly, monthly, every six months, and yearly. Depending on such
factors as work load, types of instrumentation and number of employees, a record or check-sheet of such activities
can be adopted for any given laboratory to help the laboratory manager keep track of necessary tasks.

Checklist before deciding on the purchase of an instrument


The instrument specifications should fit the purpose.
The specifications should conform to local conditions such as power supply, humidity and climate.
The advice of other laboratories using a similar instrument should be sought.
The prices of different brands with similar specifications should be compared, not forgetting prices of
spare parts.
Running costs should be compared.
Availability of after-sale services, including maintenance, should be assured.
An operations and maintenance manual must be supplied.
On delivery, an extra supply of commonly needed replacement parts (e.g. carbon brushes, fuses,
bulbs, electrodes, heating elements) should be obtained.
The equipment must be assessed for technical safety.
The need for an operator from the laboratory to receive on-the-job training in operation, maintenance
and minor repairs must be recognized by the supplier or a competent national agency.

Operation, Maintenance and calibration of different equipment


Centrifuges

Two types of centrifuges are currently used, mechanical and electrical. The major aspects in maintenance of different
types of centrifuges used in laboratory routine are as follows:

The centrifuges must be positioned exactly horizontally to prevent the instrument moving away from
its place when out of balance during centrifugation. Check if the rubber buffers are in the buckets.
It is critically important that the centrifuge load is balanced at all times. Therefore buckets loaded in
matched pans and tubes should be arranged so that balanced tubes oppose each other in the
centrifuge head. This is necessary to maintain the same forces of gravity in the opposite positions of
the tubes. This arrangement involves placing a dummy, i.e., a tube filled with the appropriate
volume of water corresponding to the weight of the volume, in the oppositely positioned test tube,
when an odd number of specimens must be centrifuged.
Haematocrit centrifuges need not be balanced before use, as the capillary tube samples are small.
Capillaries should be plugged at one end to avoid spilling and loss of blood. The plugged end should
always be placed against the sealing gasket.
67
Turn the speed control slowly up and down.
Stop the centrifuge immediately if it makes an abnormal noise.
After use, the buckets should be inverted to drain dry.
After any sample spillage, wipe and disinfect immediately.
Clean the centrifuge at short intervals (preferably daily) because it is one of the most frequently used
instruments.
Check mounting and replace if necessary.
Check brushes and bearings every 3 months. Replace if necessary.
Check for corrosion and clean and repaint if necessary.
Calibration: use a pre-calibrated tachometer to check centrifuge speed.

Refrigerators

The following general advice may be helpful for maintenance:


Refrigerators must be so placed that sufficient air can pass the condenser (at the back of the
refrigerator) for exchange of heat and also to facilitate cleaning of the condenser.
The refrigerator door must seal perfectly to prevent warm outside air from entering the cool chamber.
Calibration: Use a pre-calibrated thermometer to ensure accuracy of temperature check.

Daily Checks

Check temperature daily. It should not exceed 12C. Application of battery driven mobile or stationary thermometers is
recommended, preferably those including continuous printing or plotting of temperature measured when heat-
sensitive reagents are stored for long periods.

Monthly Checks
Clean cool chamber and defrost the evaporator monthly.
Clean refrigerator from outside.
Clean condenser of dust.
Clean door gasket.

Hot air oven

Hot air ovens are mainly used for drying laboratory equipment and medical devices in dry air. Some hot ovens are also
used for sterilization. Sterilization in dry air is only effective when the material is exposed for 60 minutes to 160C or for
40 minutes to 180C. It is important to remember that the timing of sterilization is sufficient when the holding period
begins after the air in the oven has reached its expected temperature.

Use of hot air ovens

Set up the thermostat at the required temperature prior to sterilization.


If there is a fan, check if it is working.
Allow to continue heating for an additional 60 minutes after the temperature reaches the pre-set
degree.
Switch off the heat when the time is up.
Wait until the temperature falls to 40C before opening the door.
Calibration: use a pre-calibrated thermometer to ensure accuracy of temperature checks.
68
Autoclave

Bacteria cannot survive in an autoclave environment; however, sterilization does not necessarily kill viruses. It should
be borne in mind that autoclaves need careful handling and must be regularly inspected. They can be hazardous and
can seriously injure a person with hot steam accidentally escaping from the instrument.

The main factors influencing perfect steam sterilization are


saturated steam
temperature
time

Use of Autoclaves
Prepare the material for autoclaving with indicator paper.
Fill the bottom of the autoclave with demineralized water up to the support.
Before placing the material in the autoclave, ensure that there is an adequate amount of water. Close
the lid and make sure that the rubber washer is in its groove. Screw down the clamps firmly.
Open the air outlet valve.
Turn on the heating (electric element). Do not leave the autoclave unattended.
Close the outlet valve when a constant jet of steam starts to escape. Reduce the heat as not to heat
too quickly.
Once the expected temperature is reached, reduce the heat to maintain the temperature.
Do not touch drainage tap, outlet or safety valve while heating under pressure.
When the time required for sterilization has passed, turn off the heat completely.
When the temperature falls below 1000C, open the outlet valve slowly. Do not leave the outlet valve
unopened for too long.
Never unscrew the lid clamps or open the lid except after the whistling sound stops.
Leave the sterilized material to cool before its removal from the autoclave.
Check the autoclave tape (used in the preparation of the material to be sterilized) which must have
turned black and the covering paper brown (not yellow or black).
Calibration: use a pre-calibrated pressure gauge to check accuracy of pressure in the autoclave.

Adequate sterilization by autoclaves and dry ovens should be monitored by the weekly use of a biological (spore
suspension) or chemical indicator.

69
Incubators

Incubators should be subjected to continuous recording of temperature. However, if it is not possible, the temperature
must be recorded every day and before opening of the incubator.
The incubator should have a fan to maintain uniform temperature.
Incubators are used for bacterial culture by laboratories working in microbiology.
The incubator must maintain a constant temperature of 37 1 C.
Temperature should be daily recorded in incubators.
Like all laboratory instruments, incubators must be cleaned and disinfected routinely at short
intervals (at least every fortnight) and after spilling of infectious material.
The actual temperature must correspond to the thermometer control when the instrument is used.
Calibration: use a pre-calibrated thermometer to ensure accuracy of temperature checks.

Water bath

It is important that the water bath maintains a constant temperature within a narrow range ( 0.5C) when used for
kinetic measurements such as determination of enzyme activities or clotting assays. Inadequate adjustment and
insufficient stabilization of the temperature will strongly affect the results of kinetic measurements.

Use of water baths


The level of water in the water bath must be above the level of the solution to be incubated.
Open containers, vials or tubes must not be incubated in a water bath with an open lid to avoid
contamination and dilution of the incubated material by condensed water.
The water bath must be refilled regularly to prevent growth of algae and bacteria.

Pipettes
Each pipette, whether manual, semi-automated or automated, must be tested periodically to
determine if it is delivering the correct volume.
Pipettes are used for volumetric measurements of liquids. They may be made of glass or plastic and
may have different volumes, ranging from 2 mL to 100 mL. They may be calibrated or non-calibrated.
Glass pipettes are either calibrated to contain (In) or to deliver (Ex).
The variety of designs for these devices is so numerous that it is difficult to discuss the subject
systematically. It must be noted, however, that the replacement of broken conventional glass pipettes
can be very costly, and that there may be other economic advantages in using mechanical pipettes.
Mechanical pipettes are preferably used to pipette small volumes (4L to 1000L); they have a higher
precision and are less subject to errors in pipetting than glass pipettes. Volumes less than 500L
should be pipetted with mechanical pipettes. However, mechanical micropipettes can only be
recommended where a reliable supply of standardized disposable tips can be guaranteed. They are
usually of air displacement (indirect), or direct displacement design. To avoid contamination during
pipetting samples, most designs use a disposable tip which is discarded after each delivery.
The practice of washing and reusing disposable tips is to be discouraged, as any cleaning procedure
will change the wettability of the plastic. In addition, drying even to only slightly elevated
temperatures may distort the tip. This will prevent a good pneumatic seal with the pipette body and
change the volume of liquid to be pipetted.

70
After each use, the mechanical pipettes must be kept in an upright position and thoroughly cleaned at periodic
intervals. Pipettes are precise and important basic instruments of the laboratory, and as such, need to be calibrated at
least every 3-6 months. Several methods of pipette calibration are available (see annexure X)

pH meters

A pH meter needs to be standardized before each run with a standard buffer of pH 7.0. However, in instances when the
work is related to a pH range of less then 6.0, it is advisable to use a standard buffer of pH 4.0. The buffer solution
should be checked monthly with another pH meter and discarded if the pH deviates by more than + 0.4 or if the buffer is
contaminated with microorganisms.

Glass electrodes must be stored in buffer solutions at pH 4 to pH 8. The buffer solution must be regularly changed at
short intervals. Glass electrodes which have been stored for longer periods must be soaked in 0.1 molar HCI for at
least four hours. Thereafter they must be carefully washed with distilled water. The same procedure must be applied
for dried-up electrodes. The life-span of properly maintained glass electrodes is about two years. Thereafter they
should be replaced. Aging of an electrode is indicated when the constant electrode potential does not develop 20
seconds after insertion of the electrodes into the ion solution. Glass electrodes are sensitive to mechanical damage.

Calibration of pH Meters

To obtain a precise measurement of pH, the pH meter must be calibrated with two different buffers at pH 4 and pH 7
every day (two-point calibration). For the calibration of a pH meter special buffer solutions must be used, the pH of
which should be near the pH of the solution to be measured. Phosphate buffers and acetate buffers are preferable.
Calibration measurements should be done using plastic containers.

Use of the pH Meter


Switch off the measuring circuit
Wash the electrode with deionized water
Standardize the electrode at the start of the day with two standard buffers (pH 7.0, pH 4.0)
Transfer the electrode to a small beaker containing the standard buffer, pH 7.0
Measure the pH of the standard buffer and adjust the buffer control knob to reading
Wash the electrode and gently wipe with fibre-free material.
Transfer the electrode to another beaker containing the standard buffer, pH 4.0
Read the pH of the standard buffer. It should be very close to expected pH value (pH 4.0)
Use fresh standard buffer every day. Discard standard buffer if it is contaminated or cloudy

Measuring the pH of the test solution

The following steps must be taken:

Wash the electrode with deionized water and transfer it to the test solution. With proper electrodes
the potential establishes 5 to 20 seconds after insertion of the electrodes into the sample solution. Air
bubbles at the electrodes must be avoided since they cause a drift of the electrode potential.
Compensate for the temperature at each run. Measure the temperature and adjust the temperature
dial to the reading.
71
Do not read the pH before the reading is stable.
Record the pH reading of the pH solution.
Wash the electrode and store it in a container with storage solutions.

Balances

Balances are used to measure weight and mass. The principle of weighing is based on attracting forces between two
separate masses. In daily life this is the attracting force between mass and the earth. In laboratories, weighing is an
essential step in preparing defined quantities of reagents and reaction mixtures.

There are two main categories of balance

Mechanical balances
Electromagnetic balances (battery driven or connected to the mains)

The following factors influence weighing and can cause errors in measurement

Temperature
Moisture (atmospheric humidity)
Electrostatic effects
Magnetism
Gravitational forces
Air

Maintenance of balances
The following guidelines are worth remembering:
The balance should be placed on a solid vibration-free surface, free from dust and at even
temperature, away from sunlight.
The instrument must be placed in an exactly horizontal position.
The balance should be zeroed prior to each use.
Use the smallest possible vessel for weighing. Avoid weighing in vessels made of plastic, because
they can become electrostatically charged. Use instead glass vessels or weighing paper, as
applicable. The weighing vessel and the sample to be weighed should be at the ambient
temperature. Never put your hand into the weighing chamber so as not to warm it up.
Place the sample to be weighed in a weighing vessel in the middle of the weighing pan, to avoid
corner-load error. Liquids or powders should never be directly weighed on the pan. The weight of the
weighing vessel needs to be determined prior to placing the substance to be weighed. A tweezer is
useful as a substitute for hands in the weighing chamber.
Use clean weighing vessels. Keep the working place, weighing chamber and weighing pan clean. To
avoid any possible corroding effect of chemicals, any spillage must be cleaned immediately.
Biological materials may be a source of infection. Disinfection can be done with 70% alcohol.
After completing weighing return the balance to zero weight.
Keep the working place at the balance as clean as possible.
Weight of the material to be weighed should be within the range of the balance.

72
Photometer instruments (ELISA reader etc.)

Most photometric instruments require calibration to ensure accuracy and linearity of their readings. This is usually
accomplished using special calibration plates, available from the manufacturer. The plates consist of different wells,
each capable of producing a different O.D. reading. Observed readings are compared to theoretical values and
evaluated using confidence limits. Likewise, automated diluters must be calibrated. This is usually accomplished by
diluting a standard colour solution and reading O.D. spectrophotometrically. Results must be within 10% of the
expected limits. The manufacturer of any automated instrument can be contacted for details of these procedures. The
annual maintenance contract should include the above mentioned check-ups and laboratory incharges should ensure
the implementation.

Calendar of activities for maintenance and calibration of equipment


Daily

Controls or calibrators added to each run.


Recalibration of instruments if necessary; For ELISA readers, the proper filter is put in place.
Waste containers, rinse sample ports, etc. are emptied with distilled H2O.
All spills cleaned up.
Reagent levels are checked.
ELISA washers flushed.
Biohazardous waste disposed.

Weekly

Optical components etc. are kept free from dust.


Surfaces of instruments are cleaned. Fresh batches of reagents are prepared as needed.

Monthly

Electronic or optical checks are performed on all components. Many automated and semiautomated
instruments have built-in programs for calibration.

Every 6 months

Filters are cleaned or changed, fluid lines and tubing for signs of deterioration are checked and
replaced as needed. All pipettes are regressed as needed.

Yearly

All fluid lines and tubing of major instruments are changed. A service call for factory representative if
possible is done. Pipettes are recalibrated.

Function checks

It is essential that laboratory personnel know and document that all equipment is in good condition each day of use.
This can be accomplished by undertaking function checks, often referred to as calibration and validation.

73
Calibration

That process which is applied to quantitative measuring or metering of equipment to assure its accurate operation
throughout its measuring limits.

Validation

The steps taken to confirm and record the proper operation of equipment at a given point of time in the range in which
tests are performed.

Documentation

The assurance that a piece of equipment is operating properly can best be judged by examining its performance over
time. Records of performance parameters, therefore, are a vital element in the proper operation of laboratory
equipment. Some suggested information is provided below:

Name and serial number of instrument


Elements to be checked and kind of data to be collected
Frequency of checking
Record of data
Changes made to restore accuracy and precision, if any
Signature with date of the person performing these tasks.

Preventive maintenance

Maintenance of equipment is an extremely important function in the microbiology laboratory. Unfortunately, this is
often grossly neglected because of indifference on the part of laboratory workers and on the erroneous belief that it is
too costly. The expense of such maintenance policies as inspection, lubrication and adjustment of instruments is
insignificant when compared with the cost of emergency repairs, rebuilding or overhauling equipment, and the
additional personnel time and materials involved in producing test results when equipment is down.

Preventive maintenance is defined as a programme of scheduled inspections of equipment and instruments resulting
in minor adjustments or repairs for the purpose of delaying or avoiding major repairs and emergency or premature
replacements. It provides the following advantages over breakdown maintenance.

Better quality results


Identification of components showing excessive wear
Greater safety
Fewer interruptions in services
Lower repair costs
Less stand by equipment requirements

ELISA Reader

Regular maintenance of the ELISA machine should be carried out according to the manufacturers' instructions and
may vary between brands. The machine itself should not be opened.

74
Note: The filters must be protected from moisture and fungal growth. Keep Silica gel packet in the filter box.

Plate washer

Plate washers are critical in ELISA assay performance. The washer works on a simple principle and comprises of
wash fluid, waste fluid reservoirs, pressure and vaccum pumps, a dispense manifold and a plate carrier.

The wash fluid is pressurized and a valve opens and allows the fluid through a manifold and into assay microwells. The
waste fluid under vaccum is aspirated back through the manifold to the waste container. A number of cycles of
dispense and aspiration comprises the washing of a plate. At the end of the wash procedure the wells are empty of the
fluid.

After use care

Fill the rinse bottle with about 500 ml of distilled water.


Dispose off the unused wash buffer. Rinse with distilled water, a couple of times and leave about 500
ml in the wash bottle. Fix the cap tightly.
After using the washer switch off power.

75
CHAPTER 13
TROUBLESHOOTING

Introduction

In spite of being vigilant and following the SOP, many a times a test run fails. The reason may be apparent in some
cases but not in others. Troubleshooting refers to measures undertaken to determine why a run has failed.

General approach to trouble shooting

Factors responsible for some common problems / errors

Specimen: lysed blood, lipaemia, deterioration, volume not sufficient.


Clerical/transcription: wrong name/ID no/wrong result transcribed, labeling, etc.
Kit related: conjugate gone bad (deterioration of reagents), storage not proper and inter lot
variations, etc.
Laboratory variations: due to use of different kit, different testing procedure (SOP), different
technique. This kind of variation may be inter-laboratory as well as intra-laboratory. So, results are
non-reproducible.
Environmental conditions: higher temperature, humidity.
Equipment problems: ELISA reader, washer, calibration of pipettes, water bath and refrigerator,
etc.
Technical errors: carelessness, not following SOP, pipetting errors, worksheets not maintained.
Borderline reactor/grey zone reactors.
Laboratory does not practice quality control procedures.

Methods for identifications of some common problems


Review protocol with the technician who ran the test, carefully checking for procedural omissions or
changes.
Double check component expiration dates.
Verify that all physical parameters of the assay were followed and met (e.g. times and temperatures).
Confirm that the support equipment such as pipettes, plate washing and reading systems, etc. are
working properly. Verify that the preventive maintenance and servicing procedures have been
performed.
Check the wavelength of the ELISA plate reader selected for the test.
Verify the calibration of the pipettes. If the assay calls for 10L of specimen and the pipette delivers
9L, this constitutes a 10% error and may be sufficient to cause the controls to fall outside acceptable
limits. Proper calibration of pipettes is essential for accurate results.
Check the quality of the distilled water (pH).
Check that reagents are not contaminated and were prepared and stored properly. There are several
situations in which characteristic appearances in a test system may indicate problems. Examples
include :
1) Change in the colour of the substrate e.g. if tablets appear orange when they should appear yellow
to white - indicating that the tablets have been contaminated or have deteriorated, and thus cannot
be used.

76
2) If the reagent indicator cells or carrier particles in an agglutination assay appear clumped or
otherwise heterogeneous - this reagent may have been contaminated or may have deteriorated.
Most package inserts describe the appearance of the reagents contained within the kit, and thus
ensure comparison of the actual appearance with what is expected.

Specific assay problems: some examples and their possible solutions

All wells in an ELISA are of the same colour in each row. This could be either due to conjugate going
bad, not working or substrate going bad. Try another kit from the same lot to pinpoint the cause of
error.
The plate visually reads O.K. i.e. different wells have different colours but the result on
spectrophotometer shows almost identical ELISA values. This is due to error in detection and/or
reference wavelength of the spectrophotometer/ELISA reader. Correct the wavelength and then re-
read the results.
Negative control strip has an unexpected band. Due to contamination between wells and/or reaction
has been allowed to develop too long. Repeat the entire procedure with same controls plus negative
control from another lot number to identify error.

Variations in test results

The HIV test results produced by the same laboratory (reproducibility) and or by a different laboratory on the same
specimen may vary. This can be due to error at any step right from collection of specimen to testing and final reporting
of the result.

The reasons for inaccurate results may be on account of:

Specimen problems
Clerical errors
Kit dependent problems
Technologists dependent errors
Equipment problems
Environmental problems and their influence on the test result.
Non repeatable (reproducible) results.

Specimen problems
Error Resolution
1. Insufficient volume for repeating the test in case Collect appropriate volume (3-5 ml) of blood depending
of reactive result. upon the objective of testing and strategy of testing.

2. Haemolysis can cause false positive or false Repeat sample. In case it is not possible record "sample
negative result in ELISA and background noise haemolysed" in the result.
on spot test.

3. Lipaemia causes pipetting error, otherwise it Buy widebore pipettes and pipette tips. Record "sample
causes no error in the test result. lipaemic" in the result.

4. Bacterial contamination degrades antibody and Refrigerate the samples for short storage (one week) and
will affect the borderline reactive result. store at -20C for long storage(>one week).

77
5. Freeze/thaw cycle may affect the borderline Avoid freeze-thawing as far as possible till the test are
positive samples. performed.

6. Aliquoting errors All aliquots should be labeled appropriately. A uniform


method of storing samples/aliquots should be followed in
the laboratory. Organise the specimen in racks and leave
a tube space empty between aliquots made from different
specimens.

Clinical/clerical errors

Error Resolution

1. Logging in specimen : there may be mixing of Set up the log books/records properly ; collect second
names, wrong history, wrong number, etc. sample from the same patient and perform testing.

2. Result print out : the well number on worksheet Transcribe the result from the print out sheet to the
and the print out not matched properly worksheet carefully. Repeat the test for resolution of the
(transcription error). result.

3. Errors during translating of results from the Supervisory review/vigilance of results by a second
worksheet to the report form. person can correct the error.

4. Reporting to the wrong person or result Report result to the correct person after post test
communicated without post test counselling counselling and maintain the confidentiality. Reporting
and confidentiality. should be as per the "National HIV Testing Policy".

Kit dependent errors

Error Resolution

1. Used after expiration date. Expiry dates should be displayed in bold letters on the kit
packing. Regular checks should be made to ensure that
no kits remain in the lab after the expiry dates. Stocks
should be rotated to ensure that kits with lesser expiry are
used before the kits with longer expiry dates.

2. Mixing of reagents from different kits/lot Use reagents for the kits for which they are made. Do not
numbers. mix reagents from different kits.

3. Performance characteristics (sensitivity, Do not use unsatisfactory kits i. e. those which do not
specificity and delta values) not satisfactory. meet the quality standards.

4. Deterioration/contamination of one or more Do not use contaminated/deteriorated (control values


component or reagents of the kit. (faulty different from those mentioned in package insert) kits.
transport/storage). Store at the optimal temperature and maintain cold chain
during transport of kits.

5. Intra-lot and inter-lot variation in kit performance Random monitoring of performance characteristics of the
due to faulty manufacturing practices. kits to ensure the quality.

78
Common errors in HIV testing and resolution of the same

Procedural error / technologist dependent errors Resolution

1. Dilution errors. Carefully calculate the volumes of all components


required for the test and perform dilutions of reconstituted
components accordingly to get the desired volumes.

2. Scratching off the coated antigen from the well. Do not allow pipette tip to touch the plate well while adding
the sample.

3. Inconsistent technique for test and controls Test each sample as well as the quality control sample in
particularly quality control samples. exactly the same way.

4. Mixing reagents from different lots of kits. Use reagents for the kits for which they are prepared . Do
not mix reagents.

5. Mixing of samples. The worksheets should be meticulously prepared


indicating which specimen goes into which well. The
internal quality control samples should be randomly
placed for each run.

6. SOP not followed to perform the test. Each step of SOP must be diligently followed.

7. Wrong pipetting due to carelessness. Supervision vigilance and training for accurate pipetting.

Equipment based errors


Error Resolution

1. Use of inappropriate pipette tips. Use appropriate volume pipette tips to deliver accurate
volumes.

2. Equipment not maintained as per requirement. Maintain and calibrate the equipments as per the
requirement (Refer to chapter on equipment maintenance
and calibration).

3. ELISA washer not working satisfactory. The washer should be placed under annual maintenance
contract for optimal working.

4. Use of incorrect U.V. filters. Use the correct U.V. filters for reading results (as per the
directions in package insert).

5. Improper reader ELISA reader must be placed under maintenance for


accurate results.

6. Refrigerator/deep freeze not working to Calibrate refrigerator to ensure that right temperatures
optimum level. are maintained in refrigerator/deep freeze.

Environment dependent errors


Error Resolution

1. Improper temperature can affect the enzyme Calibration of incubators to maintain the right
reaction and the result. temperatures is very important.

2. Drying up of ELISA plates Maintain humidity so that the plates do not dry.

79
Non repeatable/non-reproducible results

Causes Resolution

1. Mislabelling of specimen. Repeat testing on 2nd sample collected from the same
patient.

2. Specimen deterioration. Parallel testing of the sample with stored sample. Repeat
testing.

3. Borderline reactors. Follow up and repeat testing.

4. Carelessness of the technologist. Vigilance in the laboratory. Training of personnel.

5. SOP not changed with the change in kit. Review SOP and change if required particularly if new
types of kits are purchased.

Vigilance / Supervision

Periodic monitoring of the performance of laboratory workers is essential for ensuring quality of testing. This can be
done by officer incharge of laboratory by resubmitting an already tested specimen as a test sample without the
knowledge of the laboratory worker. If the results of the resubmitted sample are consistent with the previous results
that means quality is being maintained. This should be done in a healthy way and under no circumstances the
laboratory worker should be intimidated/reprimanded/ticked.

Vigilance means scrutiny of laboratory and every step of the test to ensure accurate results every time. Scrutiny has to
be random and of each step of the test right from collection, transport of specimen to reporting of result to the right
person in the right manner. Every step of quality control has to be practised every day to ensure that the procedures
and results are accurate and no false positive or false negative results are generated by the laboratory. This may be
done by regular internal audit of the laboratory.

Record keeping

Monitor the records from collection and logging in of the specimen in the laboratory to reporting of the results. This will
include the following details:

Name of individual or specimen identification number.


Date of collection of specimen and date received in laboratory.
Name of requesting physician.
The test requested.
Worksheet.
Result of the test.

The worksheet records the date of test, type of kit (Lot no. etc.) used, the test samples (nos.), quality control samples
(nos.), name of laboratory worker, O.D. values obtained and cut-off values, etc. The worksheets are an important
document for quality performance of tests. A model of worksheet is given in annexure III.

80
CHAPTER 14
IMPLEMENTATION OF EXTERNAL QUALITY ASSEMENT PROGRAMME AT
DIFFERENT LEVELS OF THE HIV TESTING NETWORK AND JOB
DEFINITIONS OF EACH LEVEL

A wide network of HIV testing laboratories operating at different levels exist in the country. The network includes
National and State Reference Laboratories (NRLs / SRLs,), Voluntary Counseling and Testing Centers (VCTCs),
PPTCTCs and blood bank HIV testing laboratories. These centers and laboratories are uniformly distributed across
the country.

Each and every facility / laboratory engaged in HIV testing has to practice quality assurance to ensure that every time
the testing is performed the results produced are accurate, within the limits of the procedure and the reagents / kits
used.

The network of laboratories all over India is huge and scattered. So, the Quality Assurance Programme (QAP) is being
implemented in three phases.

Phase I

All the National Reference Laboratories participate in this phase of QAP. Apex centre i.e. NIB distributes the
characterized serum panel (10) to the National Reference Lab along with the procedural proforma. The participating
labs process the External Quality Assessment serum panel exactly the same way as the routine samples and send the
results to NIB within 15 days for analysis and feedback. This exercise is repeated once every 6 months. NIB also
conducts EQA training program for the NRLs.

Those National Reference Laboratories which are already participating in any of the External Quality Assessment
Programmes being conducted by WHO Reference / collaborative laboratories may continue to do so.

Phase II

All the NRLs conduct training programme on EQAP and Internal Quality control as well as aspects of HIV Testing for
the identified State Reference Laboratories (SRL) and distribute the EQA panel. Composition of serum panel is the
same as that from apex reference laboratory. The SRLs test the EQA panel in the same way, under same conditions
and using same kits which are used for testing client / patient sample within two to four weeks. The reports are sent
back to the respective NRLs for analysis and feedback.

Phase III

Each National Reference Laboratory has been allotted states for implementation of EQA for all the HIV testing
facilities existing in the respective states. For this purpose the National Reference Labs and or SRLs prepare the
External Quality Assessment (EQA) serum panel of ten well characterized sera as per details already given. This
panel comprising of 10 sera is sent to each and every HIV testing facility namely VCTC, PPTCTC & blood banks by the
NRL/SRL in the respective allotted states.
National AIDS Control Organization, Government of India facilitates the conduction of EQAP at various levels by
81
providing funds to NRLs/SRLs and issuing directives to SACS, NRL and participating VCTC, PPTCT and blood bank
labs, etc for compliance, emphasizing the importance of EQAP. The participating laboratories in the states test the
EQA panel sent by NRL/SRL as routine test samples and send the results to the assigned NRL/SRL within 2 weeks.
NRL/SRL analyse the data and provide feed back to the participating labs. In case of discordance of result the experts
from NRL visit the defaulting centre / laboratory for trouble shooting. NRL/SRL also conduct trainings for VCTCs and
other HIV testing labs in the state.

The National Reference Labs/SRL liase with the Project Director, State AIDS Control Society to train the staff from
VCTC, PPTCTC and Blood Banks to participate in the External Quality Assessment Programme. The training
programme is funded by respective Project Directors. It is a five days hands on training programme.

The modus operandi for implementation of EQAP in states at grass route level can be decided by the Officer In
charges, NRLs/SRLs corresponding to the size of states, etc.

The panel is sent to the participating laboratories maintaining the cold chain (2-80C). The serum panel comprises of a
total of 10 samples (4 positive, 4 negative and 2 borderline positive). The NRL sends all the relevant papers,
proformas (Annexure VI) for performing the test and reporting back of the results obtained by the participating
laboratories to the SRL/ NRL.

The networking of HIV testing facilities for EQAP is given in annexure IX. Each National Reference Laboratory has
been allotted SRLs/HIV testing cnentres in various states as mentioned in the annexure.

The NRLs/SRLs cater to the ever expanding HIV testing facilities in their respective regions for EQAP.

Job definitions for the network of labs engaged in implementing EQAP


The levels of labs:
Apex Reference Laboratory
National Reference Laboratories (NRLs) and State Reference Laboratories (SRLs).
Blood bank laboratory, VCTC, PPTCTC

Apex Laboratory
Job definition
Preparation of serum panels for conducting external quality assurance.
Evaluation of panels prepared by the National Reference Laboratories
Participation in the EQA program conducted by the WHO collaboration Centre.
Trouble shooting the problems faced by the National Reference Laboratories.

National Reference Laboratories and State Reference Laboratories


Job definition
To send panel to apex centre for quality control and evaluation of the panel
Distribution of panels to SRLs, VCTCs, PPTCTCs and blood banks.
Analysis of data and feedback to participating SRLs, VCTCs, PPTCTCs and blood banks.
Solving problems faced by SRLs and other HIV testing centers with support from apex laboratory as
and when required.
Testing representative samples received from SRLs and other HIV testing labs (20% of all positives
and borderline positive sera and 5% of all negative sera).

82
Training of SRLs and HIV testing laboratories.
Visiting SRLs and HIV testing labs for trouble shooting

VCTCs, PPTCTCs and blood bank laboratories,


Job definition
Testing of external quality control panel sent by NRL/SRL.
Generation of data from blood bank labs, VCTCs and PPTCTCs.
Referring lab related problems and problem sera (sero-status not resolved) to respective SRL
Sending representative samples to SRL for quality assurance (20% of all positives and borderline
positive sera and 5% of all negative sera).
Should possess an independent HIV testing laboratory with adequate floor space, personnel for
carrying out quality assurance work.

Recommendation on development of serum panels and evaluation of kits


Development of panels

Panels will be prepared by the NRLs and SRLs. The serum panel should comprise of 40% positive, 40% negative and
20% border line reactive sera.

Flow of sample panel from Apex Laboratory /NRL/SRL

Apex Reference Laboratory

National Reference Laboratories

State Reference Laboratories

VCTCs, PPTCTs, Blood Banks

Transportation of panel
All the panels are to be transported from one laboratory to another in the temperature range of 2-8C. The panel
should consist of 10 serum samples and each individual sample should have a volume of at least 100 microlitres.

Turn around time for reporting


The participating laboratory will send back the result of the panel within four weeks of the receipt of the panel to the
next level of laboratory linked with the centre and the higher laboratory will provide feed back to the respective
laboratories within 8 weeks.

Frequency of external assessment exercise


It will be conducted at least once a year to begin with and preferably twice a year, during months of February/March
and August/ September preferably.

Panel evaluation
Apex laboratory as well as the National Reference labs will carry out the evaluation of panel and analysis of the results.
NACO will inform the apex laboratory and NRLs about the details of HIV test kits that are likely to be purchased and
distributed to the blood banks and blood testing centers.

83
Kit evaluation

The apex laboratory and the National Reference Laboratories will have the responsibility of evaluation of kits as per
requirement of NACO and other national bodies. It is also suggested that each fresh batch of kits purchased by NACO
should be evaluated for sensitivity, specificity and delta values before it is distributed to the blood banks and other
laboratories so that the quality of the kits supplied is assured.

The apex laboratory and the NRL should put together efforts to prepare panel of sera to be used for evaluation of kits.
This panel should include (i) sera from different geographical locations in the country (ii) sera against prevalent types
and subtypes (iii) sera from patients with chronic diseases such as tuberculosis, leprosy, kala azar, auto immune
diseases, malaria, etc. and (iv) sera from HIV sero-negative individual with risk behaviour for acquiring HIV infection.

The panel established by the Apex Laboratory and National Reference laboratories will be tested for thermostability
and reproducibility in the Apex Laboratory and NRLs. It is also recommended that regular surveillance should be
maintained at each level. Five percent of negative samples, 20% of positive samples and all border line reactive
samples may be monitored at the respective higher levels; grey zone samples should be repeated.

The networking of HIV testing facilities for EQAP:

84
CHAPTER 15
EQUIPMENT, SUPPLIES AND REAGENTS FOR PRACTICING
QUALITY CONTROL IN HIV TESTING*

Biosafety " Laboratory coats/gowns (front closed).


" Chappals, slippers (top covered).
" Rubber gloves.
" Sodium hypochlorite / ethanol / glutaraldehyde / any other
appropriate disinfectant.
" Discarding jar.
" Disposable paper sheets.
" Disposable bags.
" Autoclave, preferably incinerator.
" Lysol/labolene (for disinfecting floor) .
" Formalin (for disinfecting rooms).
" Needle destroyer

Collection of serum samples from " Disposable syringes and needles.


individual/plasma from blood bags (from " Screwcapped plastic tubes.
blood banks). " Centrifuge for separation of serum / plasma.
" Vacutainers (vacuum based blood collection systems)

Recalcification of plasma to serum. " CaCl 2. 6H2O; MgCl2. 6H2O.


" Centrifuge bottles (250 ml, presterilised, screwcapped).
" Water bath.
" Vacuum pump, seitz filter with filter pads, biological filters (
0.8m, 0.6m).
" Magnetic stirrer.

Preparation of external control " Water bath (35-44C).


" Incubator BOD ( 35-44C).
" Balance (physical / electrical).
" Rocker (40 - 60 rockings / minute for running Western Blot).
" ELISA reader filters (405nm, 492nm, 540nm and paper
role).
" ELISA washing system.
" ELISA / rapid kits for HIV testing.
" Glass pipettes - 2ml, 5 ml, 10 ml.
" Microtips ( capacity 25l, 50l, 100l).
" Multichannel pipettes ( 50-200l).
" Micro pipettes ( 5-50l, 50-200l, 200-1000l).
" Microtip container.
" Pasteur pipette, rubber teats.
" Plastic forceps, plastic troughs.
" Tube racks.
" Graph paper.

85
" Calculator (scientific).

Sterilisation " Hot air oven.


" Autoclave.

Storage / preservations " Refrigerator (domestic with freezer compartment).


" Deep freeze (-30C, -70C).
" Polypropylene storage bottles ( 250 ml.).
" Plastic storage vials .
" Syringes for filtration through biological filters (50 ml.).
" Racks for holding storage vials.
" Biological filters ( 0.8m, 0.2 m).
" Bronidox (5 - bromo - 5 - nitro - 1, 3dioxane) and Propylene
glycol (solvent for Bronidox)/ Thiomersal.

Referral " Sample carrier ( e.g. Thermocol box).

Expected existing back up " Water distillation plant.


" Glass ware (e.g. Beakers, flasks, measuring cylinders, etc).
" Markers, parafilm (for sealing storage vials).
" Air conditioner.
" UPS (uninterrupted power supply).
" Biological safety cabinet.
" ELISA washer (automated / semi-automated).

*Uses of these components have been included in relevant sections.

86
CHAPTER 16
BIOSAFETY FOR HEALTH CARE WORKERS WITH SPECIAL
REFERENCE TO HIV

Introduction

Ever since AIDS has been recognized in 1981, the concern about bio safety in medical science has attained a new
dimension. However, any measure regarding safety, needs assessment of the extent of risk, failing which, there is
possibility of undue panic or carelessness due to over-estimation or under-estimation of the risk respectively.

Fortunately, the threat of HIV infection to health care workers (HCW) has never assumed so much of problem. This
has been mainly due to the fact that HIV is known for its fragile nature, having very low power of surviving in natural
conditions like drying, temperature, etc. as well as due to its relatively lower concentration in blood and body fluids.

Blood is the single most important source of HIV, HBV, HCV and other blood borne infections for HCWs. The risk of
infection varies with the following factors:
Type of exposure e.g. exposure of intact skin, nonintact skin, mucous membrane or needle stick
injury.
The amount of blood involved in the exposure; hollow bore needles, cannula, etc. carry more blood
than IM (intramuscular) needles.
The amount of virus in patient's blood at the time of exposure (maximum during acute viraemia).
Prevalence of infection in the population.
Number of exposures in case of needlestick injuries from individuals with unknown serostatus.
Timely availability of post exposure prophylaxis (PEP).

Table1. Modes of exposure to blood borne pathogens in the laboratory

87
Table2. Essential steps of biosafety

The essential biosafety measures in relation to various steps of activities in a health care setting are as follows.

Bio safety guidelines for laboratory workers

General precautions for laboratory workers


Wear gloves for all manipulations of infectious materials or where there is possibility of exposure to
blood or body fluids containing blood. Ensure generous supply of good-quality gloves.
Discard gloves whenever they are thought to have become contaminated; wash your hands, and put
on new gloves.
Do not touch your eyes, nose, or other exposed membranes or skin with gloved hands.
Do not leave the workplace or walk around the laboratory wearing gloves.
Wash your hands with soap and water immediately after any contamination and after work is
completed. If gloves are torn, wash your hands with soap and water after removing the gloves.
Wear a laboratory gown, smock, or uniform when in the laboratory, wrap-around gowns are
preferable. Remove this protective clothing and leave it in the laboratory when leaving the laboratory.
When work with material potentially infected with HIV is in progress, close the laboratory door and
restrict access to the laboratory. The door should have a biohazard symbol and No Admittance sign
posted.
Keep the laboratory clean, neat, and free of extraneous materials and equipment.
Disinfect work surfaces when procedures are completed at the end of each working day. An effective
all-purpose disinfectant is a 1% hypochlorite solution.
Place used needles, syringes, and other sharp instruments and objects in a puncture-resistant
container. Do not recap used needles and do not remove needles from syringes.
Never use mouth pipetting.

88
Perform all technical procedures in a way that minimizes the creation of aerosols, droplets, splashes,
or spills.
Do not eat, drink, smoke, store food or personal items, or apply cosmetics in the laboratory.
Have an effective insect and rodent control programme as per standard recommendation.

Guidelines for those who collect blood samples

Major hazards

Blood contamination of the hands while drawing blood.


Penetrating injuries caused by needles and other sharp instruments or objects.

Guidelines

Inspect your hands for cuts, scratches, or other breaks in the skin.
If the skin is broken, cover with occlusive bandage and wear gloves.
Take care to avoid contaminating your hands during the collection of blood.
If blood gets on gloves, these should be discarded.
Place used needles and syringes in a puncture-resistant container. Do not recap used needles. Do
not remove needles from syringes. Do not use needle clippers.
Seal specimen containers securely. Wipe the exterior of the container free of any blood
contamination with a disinfectant.
Wash your hands with soap and water immediately after any blood contamination and after work is
completed.
Wear the laboratory gown.
In the event of needlestick or other skin puncture or wound, wash the wound thoroughly with soap
and water. Encourage bleeding.
Report any contamination of the hands or body with blood, any puncture wound, or cut to the
supervisor and the health service.

Table 3. Safety guidelines for laboratories engaged in serological testing of HIV

Chemical disinfectants commonly used in HIV laboratory

The common disinfectants that are used for bio safety against HIV are mentioned below:

89
Chlorine - sodium hypochlorite
For wiping work benches and specimen containers, cleaning gloves, etc. 1% freshly prepared
sodium hypochlorite solution is used.
For heavily soiled equipment, blood spills, etc. 10% sodium hypochlorite solution is needed.
Sodium hypochlorite solutions gradually lose their strength, requiring daily preparation of fresh
solutions for use.
The effectiveness of hypochlorite solutions is neutralized by blood, serum, and other proteinaceous
material. Daily replacement is required.

Chlorine - calcium hypochlorite


Available in powder, granule, or tablet form.
Decomposes at a slower rate than sodium hypochlorite.
Availability of chlorine - (70%)
0.1% solution is obtained by dissolving 1.4 g dry chemical in 1 litre of water.
1.0% solution is obtained by dissolving 14 g dry. chemical in 1 litre of water.

Chlorine - sodium dichloroisocyanurate (NaDCC )


Available as tablet.
60% available chlorine.
0.1% available chlorine solution is obtained by a dilution of 1.7 g stock in 1 litre of water.
1.0% solution is obtained by a dilution of 17 g in 1 litre of water.
More stable than sodium and calcium hypochlorite.

Chloramine
Available in powder or tablet form
25% available chlorine
Releases chlorine at a slower rate than other chlorine compounds.
Higher concentrations are required for effective disinfection.
General purpose - 20 g stock in 1 litre of water.
40 g stock in 1 litre of water for disinfecting heavily soiled equipment, blood spills, etc.
Chloramine is more stable than sodium and calcium hypochlorite.

Alcohol-ethanol (ethyl alcohol) and 2-propanol (isopropyl alcohol)


Ethanol and 2-propanol are effective disinfectants, particularly for cleaning surfaces such as the
exterior of specimen containers and bench tops.
For maximum effectiveness they should be used in a 70% concentration (70% alcohol, 30% water).

Iodophore - povidone iodine (PVI)


Disinfectant activity is similar to that of hypochlorite solutions.
More stable and less corrosive (however, do not use on aluminium or copper).

Formaldehyde -formalin
Excellent disinfectant
Formalin containing 35%-40% formaldehyde, 10% methanol and water.
Should be diluted 1:10 (a solution containing 3.5%-4% formaldehyde) for disinfection.
Rinse the equipment before reuse.

90
Glutaraldehyde
High-level disinfectant.
Extremely useful for the disinfection of non-disposable heat-sensitive equipment and instruments.
Usually available as a 2% aqueous solution.
Immersion in the solution destroys vegetative bacteria, fungi, and viruses in less than 30 minutes.
Ten hours' immersion is required for the destruction of spores.
Treated equipment must be thoroughly rinsed.
Prepared solutions should not be kept more than 2 weeks, or according to the manufacturer's
instructions.

The selection of an appropriate disinfectant will also be weighed in terms of contact time, physical characteristics like
irritant and toxic properties etc. In a developing country like India cost is another important consideration.

Transport of specimens

Transport of specimens (e.g. serum, blood etc.) is a frequently performed activity for a laboratory engaged in HIV
diagnosis. The broad guidelines are as follows:

General guidelines
Specimen containers should be screwcapped, leak-proof resistant plastic or glass.
After the container is closed and sealed it should be wiped with a disinfectant - e.g. 1% sodium
hypochlorite solution. On receipt of a specimen, before the container is opened it should be wiped
with a disinfectant as above.
Within the health care facility and laboratory, the specimen containers should be placed in racks to
maintain them in an upright position. The racks are to be transported in leak-proof plastic, metal or
rigid thermocol containers.
From field collection sites or between laboratories, leak-proof plastic or metal boxes with secure,
tight-fitting covers should be sought.

Disposal of laboratory waste


Waste should be segregated into different categories at the site of generation i.e. in the laboratory.
Non infectious waste is collected in black bags and disposed as general waste.
Infectious waste should be decontaminated prior to its storage, transport and disposal.

Solid infectious waste is disposed as follows


Sharps needles and syringe nozzles are shredded in needle-destroyer (if available); if not available
decontaminated as in below.
Scalpel blades/ lancets / broken glass should be put in an impermeable container with bleach that
should be filled to three fourth only, transferred to plastic/ cardboard boxes, sealed to prevent spillage
and transported to incinerator.
Glass ware should be disinfected and sterilised by autoclaving followed by cleaning; never attempt
cleaning before autoclaving.
Swabs are chemically disinfected followed by incineration.

Disposable items include single use products (syringes, gloves, sharps, etc.). As these items are often recycled and
have the risk of being reused illegally, these should be disinfected by dipping in freshly prepared 1% sodium
91
hypochlorite for 30 minutes to 1 hour. Bins which can be used for this purpose are a set of twin bins, one inside the
other with the inner one being perforated and easily extractable. This minimizes contact when the contents are being
removed.

Disposable items like gloves, syringes, etc. should be shredded, cut or mutilated before disposal followed by deep
burial or properly accounted before disposal. Extreme care should be taken while handling the needles.

Liquid wastes generated by the laboratory are either pathological or chemical in nature and are disposed of as follows:
Non infectious chemical waste should first be neutralised with reagents and then flushed into
conventional sewer system.
The liquid infectious waste should be treated with a chemical disinfectant for decontamination then
neutralized and flushed into the sewer.
Solid wastes are collected in leak-resistant single heavy duty bags or double bags may be used.
Bags having different colour with red labels mentioning date and details of waste are recommended.
The bags are tied tightly after they are three-fourths full.

Processing of glass syringes, needles and gloves for reuse

These are the items most frequently handled in any laboratory handling blood or other infectious fluids/materials.
Hence procedures for their reuse are mentioned below.

Reuse of needles and glass syringes


Gloves must be worn.
Extreme care must be exercised to prevent needle stick injuries and/or cuts.
Leave the needle attached to the syringe.
Aspirate hypochlorite or another suitable disinfectant into the syringe.
Immerse the syringe and the attached needle in disinfectant solution, horizontally in a flat tray.
Leave them immersed in disinfectant solution for 20 minutes.
Discharge the disinfectant solution from the syringe and needle.
Rinse the syringe and needle, filling and emptying several times.
Examine needles and syringes for needle barbs, the rubber ring, the needle hub fit to the syringe,
readable syringe markings, etc.
Disinfect or sterilize the syringe and needle by autoclaving (in a steam steriliser) or boiling in water
prior to reuse.

Reuse of gloves
Generally recommended for laboratory workers who handle blood specimens or other suspect fluids potentially
infected with HIV. Gloves reduce the risk of contamination of the hands with blood, but will not prevent the occurrence
of penetrating injuries or cuts caused by needles, other sharp instruments, or broken glass or plastic.

It is important to remember that gloves are meant to supplement, not replace good infection control practices,
including the hygienic practice of proper hand washing.
Test the gloves for peel, cracks, punctures, etc.
Thoroughly rinse your gloved hands in a hypochlorite disinfectant solution.
Rinse your gloved hands in clear water.
Wash your gloved hands with soap and water and rinse.
Remove the gloves and hang them up to dry.
Wash your hands.
92
Packing, storage and transport of treated (disinfected) wastes

All segregated and disinfected waste should be packed in proper containers and colour-coded bags with red label
biohazard signs (Table - 4). All containers used for storage of such waste shall be provided with a properly covered lid
to avoid spillage during handling and transit of such waste.

Such containers should be inaccessible to scavengers and protected against insects, birds, animals and rain. The
waste should be transported in vehicles authorised for this purpose only. No such waste should be stored in the place
where it is generated for a period of more than two days.

Disposal of laboratory wastes

Waste which cannot be incinerated (e.g. plastics) should be pre-treated by disinfection and disposed of in an
environmentally sound manner.

Waste should not be dumped, discharged or disposed in any place other than a site identified for the purpose.

All treatment and disposal facilities including burial pits shall be located at a specified area away from the general
service area of the hospital, public places and residential areas.

All precautions and personal safety measures should be taken (including provision of protective clothing, masks,
gloves, gumboots, goggles, etc. as may be necessary). Hepatitis B vaccine is recommended for affording protection
to all personnel engaged in handling biomedical waste, or being exposed to such wastes against infection from
handling or exposure.

All liquid waste should be disinfected and flushed in the sinks at the point of generation.

Table 4. Container and colour coding for disposal of bio-medical wastes

ks

93
ks

Spills and accidents


Spills of blood /infectious material can be quite frequent in a busy laboratory. The essential steps and
requirements in case of spills are as follows:

Essential steps
Wear gloves

Cover with paper towel/absorbent material

Pour hypochlorite 10%

Leave for 30 minutes

Clean the area

Finally, wipe the surface with disinfectant again & mop up to leave a clean area.

Essential requirements

Gloves to be worn, avoid direct contact of gloved hand with spill.


Sweep broken glass/fractured plastic with dustpan and brush.
Needle stick or other puncture wounds, cuts and skin contaminated with spill should be washed with
soap and water; encourage bleeding.
Report immediately.
Maintain record.

Occupational exposure to blood/body fluids and contaminated sharps, etc.


An "exposure" that may place a Health Care Provider (HCP) at risk of bloodborne infection is defined as a
percutaneous injury (e.g. needle-stick or cut with a sharp instrument), contact with the mucous membranes of the eye
or mouth, contact with non-intact skin (particularly when the exposed skin is chapped, abraded, or afflicted with
dermatitis), or contact with intact skin when the duration of contact is prolonged (e.g. several minutes or more) with
blood or other potentially infectious body fluids.

Body fluids that are potentially infectious include - blood, semen, vaginal secretions, cerebrospinal fluid, synovial,
pleural, peritoneal, pericardial and amniotic fluids or other body fluids contaminated with visible blood. Exposure to
tears, sweat, urine, faeces, saliva of an infected person is normally not considered as an "exposure" unless these
secretions contain visible blood.
94
The exposure must be reported to the appropriate authority and condition must be treated as an emergency. Prompt
reporting is absolutely essential.

Post exposure prophylaxis against HBV, HCV & HIV

HBV

In recent years immunization of HCWs against HBV has been successfully employed widely. However some of the
potential limitations are lack of consensus regarding the ideal regime of vaccination, the required level of protective
antibody and poor understanding about the variables that might hamper expected response.

Table 5. Recommendations for Hepatitis B prophylaxis following percutaneous or


permucosal exposure

* HBIG dose 0.06 ml/kg IM


@ Adequate anti-HBs Ag is > 10 SRU by RIA or positive by EIA.

Hepatitis C (HCV)

Hepatitis C is most commonly transmitted by parenteral route, i.e., through infected blood transfusion or sharing of
needles and syringes in IV drug users. It is a common agent of co-infection with HIV.
Like the other bloodborne infections, the mainstay of preventing occupational infection is through prevention of
exposure. Risk of transmission after needlestick injury is 2-3%. There is no vaccination at this time, and no
recommended chemoprophylaxis. For HCW exposed to Hepatitis C, follow up care is recommended.

95
HIV
Management of Exposure :

Steps to be taken on accidental exposure to blood (or body fluid containing blood) are:
Wash wound immediately with running water and soap
Inform the lab /hospital management and document occupational accident
Consult with nearest ART centre/ resource for Post-exposure prophylaxis, evaluation, and follow-up (as per
the National guidelines on PEP)
Counselling and collection of blood for testing from the exposed HCW with written informed consent must be
done.
Whenever possible confidential counselling and testing of source for Hepatitis, HIV etc must be done. A
history should be taken as well to ascertain likely risk of the source. (PEP should be provided to the exposed
HCW until report of source is available and confirmed negative).
Risk of infection and transmission must be evaluated
Never delay start of therapy due to debate over regimen. Begin with basic 2-drug regimen, and change if
warranted, once expert advice is obtained
Reevaluation of the exposed person should be considered within 72 hours post exposure, especially as
additional information about the exposure or source person becomes available. The exposed person is
advised to seek medical evaluation for any febrile illness that occurs within 12 weeks of exposure.
Administer PEP for 4 weeks .PEP should be provided until result of the source's
test is available and confirmed negative or until course completed ,if source positive or unknown
A repeat HIV test of the exposed individual should be performed at 6 weeks, 12 weeks and 6 months post-
exposure, regardless of whether or not PEP was taken

Ideally, prophylaxis should be begun within 2 hours of exposure. Late PEP (within 72 hours) may still be
useful as early treatment of HIV infection, in case infection has occurred.

Donts:
Do not panic!
Do not reflexively place pricked finger into mouth
Do not squeeze blood from wound, this causes trauma and inflammation, increasing risk of transmission
Do not use bleach, alcohol, betadine, or iodine, which may be caustic, also causing trauma

Dos:
Remove gloves, if appropriate
Wash site thoroughly with running water. Irrigate thoroughly with water or saline if splashes have gone
into the eye or mouth.
Seroconversion after occupational exposure

If transmission of a bloodborne infection occurs after occupational exposure which has been documented, the HCP
has a right to receive treatment and care for this illness. It is also the right of all persons infected thusly, to freedom
from discrimination regarding their working conditions. Such persons are entitled to have all of their human rights
respected, beginning firstly with the right to confidentiality regarding their health care.

96
Table 6. HIV Postexposure prophylaxis evaluation

Table7. Post-exposure chemoprophylaxis when source patient is drug nave

of

Nelfinavir 750 mg three times daily or


any other boosted protease inhibitor.
(for higher risk categories - consult
expert).

Drugs not recommended for PEP:


l ddI and d4t; together
l Nevirapine

In case the source patient is on ART and considered to be having drug resistance (clinically) then the ART
97

124 124

125 125
physician should be consulted and appropriate ARVs should be given for PEP.

Schedule of necessary measures regarding therapy and follow up in case of PEP

The person should be provided with pre test counseling and PEP be started as discussed above. Before starting PEP,
3-5 ml of person's reference blood sample is to be taken and sent to the laboratory for testing and storage. It is
important that a serum sample is collected from the HCW as soon as possible (zero hour) after exposure for HIV
testing, failing which it may be difficult to attribute the acquired infection due to exposure in occupational setting. This
may have bearing on the claims of compensation from health authorities. The first sample for HIV testing is collected
immediately after exposure and 2nd at 6 weeks, 3rd at 12 weeks and last at 6 months after the exposure. During the
follow-up period, especially the first 6-12 weeks when most infected persons are expected to show signs of infection,
the recommendations for preventing transmission of HIV are to be followed by the HCW. These include refraining from
blood, semen, organ donation and abstaining from sexual intercourse. In case sexual intercourse is undertaken a
latex condom must be used correctly and consistently. This reduces the risk of HIV transmission. In addition, women
should not breast-feed their infants during the follow-up period after exposure to prevent exposing their infants to HIV
in breast milk.

The antiretroviral drugs for PEP are to be given for four weeks. Government of India has already made the resources
available with various State AIDS Societies to meet with the expense of PEP for HCWs. The drugs for PEP must be
available round the clock.

The report of exposure and PEP has to be sent to Addl. Director (Technical) National AIDS Control Organisation, GOI.

Pregnancy and PEP

If the HCW is pregnant at the time of occupational exposure to HIV, the designated authority/physician must be
consulted about the use of anti-viral drugs for post exposure treatment.

Facts known about the safety and side effects of these drugs

Most of the information known about the safety and side effects of these drugs is based on studies of their use in HIV-
infected individuals. For these individuals, ZDV and 3TC have usually been tolerated well except for nausea, vomiting,
diarrhoea, tiredness, or headache for people taking ZDV.

Steps to be undertaken by the infection control officer on receiving information about occupational exposure
All the needle stick injuries should be reported to the State AIDS Society giving the details of the type
of exposure. and the measures taken to manage the same.The State AIDS Societies should in turn
inform NACO about the cases periodically.
A registry is available with NACO for follow-up of all such cases.
Infection control officers in all hospital have been directed to ensure that anti retroviral drugs for PEP
are available in casualty at all the times.

98
CHAPTER 17
SELECTED LIST OF REFERENCES CONSULTED

1. WHO publication Safe blood and blood products introductory Modules No. 1 to 3, 1995.

2. WHO publication, Eastern Mediterranean series No. 14, 1995.


Quality System for Medical Laboratories-guidelines for implementation and monitoring.

3. Retroviral testing - Essentials for quality control and laboratory diagnosis. Constantine. NT. Callahan JD and
Watts DM, (Eds) CRC Press, London, 1992.

4. Gradwohl's Clinical Laboratory Methods & Diagnosis. Sonnenwinth CA & Jarett. C.V. (Eds) Mosby
Company, St. Louis, Missouri, 1980.

5. Quality Control in Microbiology, Prier JE, Bartoia J and Friedman H. (Eds) University Park Press. Baltimore,
1975.

6. WHO publication, Blood Transfusion - A guide to the formulation and operation of a transfusion service.

7. Talib VH and Dutta AB ; A text book of Blood Banking and Transfusion Medicine, CBS publishers and
distributions, New Delhi, 1st Edition, 1995.

8. Pavri K. Standard bio safety guidelines for blood banks; Transfusion bulletin, 1991, 4: 6 - 11.

9. Saran RK and Makroo RN. Transfusion Medicine Technical Manual, DGHS, Govt. of India, 1991.

10. Ministry of Health & Family Welfare, Government of India publication. Standards for Blood Banks and
Transfusion Services, DGHS, New Delhi 1987.

11. American Association of Blood Banks. Standards for Blood Banks and Transfusion Services, 12 th edition,
1987.

12. WHO/UNAIDS publication Guidelines for Organising National External Quality Assessment Schemes for HIV
serological testing; 1996.

13. Report of Quality Assurance Workshop - South East Asian & Western Pacific Region. Bangkok. Thailand, 23
rd to 31 st January, 1997.

14. Workshop Document : SEAWP Region Laboratory Quality Assurance Workshop 7-11 February, 2000,
National Serological Reference Laboratory, Australia in collaboration with Bureau of Laboratory Quality
Standards, Department of Medical Sciences, Ministry of Public Health, Thailand.

15. HIV testing policies and guidelines WHO 1994.

16. AIDS HIV: Reference Guide for Medical Professional 4th Ed. Fahey, John L, and Flamming D.S., CIRID,
UCLA Williams and Wilkins, London, 1996.
99
17. AIDS Etiology Diagnosis Treatment and Prevention 4th Ed. Curran J., Esses M., and Fauci A.S, Pippincott -
Raven, 1997.

18. HIV Testing Manual: Laboratory Diagnosis, Biosafety and Quality Control. NACO document.

19 NABL documents on accreditation.

20 ISO 17025 and other ISO documents.

100
Annexure I

Consent form for HIV Testing

This is to state that I have been counselled about the HIV test to be conducted on me and have been explained about
the implications of the test result- positive, negative or indeterminate. All the details pertaining to HIV, its transmission,
and testing procedure have been explained to me. Its limitations and interpretation of results have been explained to
me in a manner that I can understand.

I, hereby, give my consent for the test to be conducted on me in order to ascertain my HIV sero-status.

Signature:

Date:

Note:

1. It may be noted that general consent obtained for carrying out procedures in hospital does not include HIV
consent.

2. In case of minor, the consent should be obtained from the parents.

3. In case of unconscious patients, where there is a need for diagnosis of HIV for management of the patient,
consent should be obtained from the parents, spouse/closest relative available at that time.

4. In case no attendant is available, the test, if necessary for management of the patient, may be carried out on
recommendations of two attending doctors.

101
Annexure II

102
Annexure III

103
Annexure IV

104
105
Additional Project Director (Technical) N.A.C.O., Ministry of Health & Family Welfare
106
Additional Project Director
Annexure VI

107
Annexure VI contd.

HIV ANTIBODY TEST RESULTS


ELISA RESULTS (SRL/BLOOD BANKS/OTHER LABS PERFORMING ELISA)
Please record the reading for the sample (column A) and the appropriate cutoff value/rate (column B) and calculate
the Sample Ratio/Cut off Ratio (A-B) for each panel sample. Please record the test interpretation given to the sample
after the testing.

108
Annexure VI contd.
HIV ANTIBODY TEST RESULTS
RAPID (e.g. particle agglutination results)
Please record if a reaction is observed with unsensitised particles for each sample (see control column). If your
laboratory titrates positive samples please record the titre (note that this is not essential, but optional). Please list
under 'Remark' if you performed an absorption step with any samples showing agglutination activity with unsensitised
particles.

109
Annexure VI contd.
HIV ANTIBODY TEST RESULTS
EXTRA TEST RESULT FORM

110
Annexure VI contd.
HIV ANTIBODY TEST RESULTS

HIV ANTIBODY STATUS REPORT


Please give a final anti-HIV status for each of the panel samples.

Example: Report the status of each panel sample as Negative, Indeterminate /


Equivocal or Positive and place a tick in the right hand column of this
form if you consider any panel sample for further testing.

111
Annexure VII
Name of the NRL / SRL Conducting the Training Programme
(Sample Questionnaire)
Workshop on External Quality Assessment Programme for HIV Testing

Participant Group: Laboratory supportive personnel

Pre-Test / Post-Test Evaluation


Time :30 min.

Name of the Participant:

Name of the Institute:

TICK ( ) THE CORRECT ANSWER

1. The choice of protocol for serological testing does not depend on the following:
(i) Objective of HIV testing (ii) Sensitivity and specificity of the test
(iii) Prevalence of HIV infection in (iv) Time limit for performance of test
the population
(v) Serum or plasma as specimen

2. In multiple testing strategy for HIV infection, the strategy II B is employed for
(i) Blood bank (ii) Diagnosis of individual with AIDS defining symptoms
(iii) Detection of HIV status in VCTC

3. The most important limitation of p24 antigen for diagnosis of HIV infection is:
(i) Poor specificity (ii) Poor sensitivity
(iii) Poor correlation with therapy (iv) Inability to diagnose HIV-2 infection

4. External quality assessment scheme is usually organized by


(i) Local officer-in-charge of the (ii) Local supportive laboratory personnel
laboratory
(iii) External agency / National /
Regional reference laboratory

5. Recommended method for conversion of plasma to serum is


(i) Calcification (ii) Thrombinisation
(iii) Heating the plasma at 56C in (iv) Incubating at 37C
water bath

6. Recommended diluent for making serial dilution of strong reactive serum for preparing level control is
following expert
(i) Normal saline (ii) Phosphate buffer saline at pH 7.2
(iii) Serum negative for infectious (iv) Distilled water
markers (like HIV, HBV, HCV etc.)
112
Annexure VII.....contd.
7. In order to find out the extent of variation of reactivity of a sample over 7 consecutive days in ELISA kits of
same batch number, it is preferable to draw inference from
(i) OD values of the samples (ii) Positive control values
(iii) Negative control values (iv) 'E' ratio of the sample

8. Any given HIV positive serum when subjected to 4 kits of different brands
following things are likely to be observed
(i) 'E' ratio differs (ii) 'E' ratio does not differ

9. Recommended range of 'E' ratios of an external control in indirect ELISA is


(i) 10-20 (ii) 1.5-2
(iii) 0.01-0.1 (iv) 0.1-1

10. Levy Jenning's chart for validation of test result usually employees the range of
(i) Mean 1SD (ii) Mean 2SD
(iii) Mean 3 SD (iv) Mean 4 SD

11. Six consecutive points of external control values gradually moving in one direction up or below the mean line
in the Levy Jenning's chart possibly indicate:
(i) Sudden change in the batch of kit (ii) Change in the technician Performing the test
(iii) Change of incubator with (iv) Gradually deteriorating reagents
different temperature set point

12. For storage of the aliquot of external control sample for daily use over 7 days
Temperature required is
(i) 0C (ii) 2-8C
(iii) <-30C (iv) <-70C

13. The recommended preservative for long term preservation of external control serum is
(i) Sodium azide (ii) Thiomersal
(iii) Bronidox

14. Standard quality control practice requires


(i) Inclusion of external control in (ii) Quality check of every run equipments
(iii) Changing the working hand
(performing the test frequently)

15. Two populations X and Y were subjected to a screening test (p) for HIV infection with sensitivity and specificity
of 99.5% and 99% respectively. Prevalence of HIV infection in population X is 20 per 1000 which the same in
population Y is 5 / 1000. The positive predictive value of the screening test (p) in population X will be:

(i) Lower than that in population Y (ii) Higher than that in population Y
(iii) Identical compare to that in
population Y

113
Annexure VII.....contd.
16. While choosing a kit for diagnosis of HIV infection, three kits were evaluated with a panel of 200 sera
comprising of 100 true positive and 100 true negatives the following characteristics were obtained. Mark the
one with maximum positive Delta value.

17. The most important limitation of a rapid test for diagnosis of HIV infection is
(i) Poor sensitivity (ii) Poor specificity

18. You want to buy one micropipette for your laboratory for HIV testing. Placed below is a checklist of major
(basic) specifications that you need to consider before the purchase. Specify (mark ) if any of these is
considered unnecessary by you.

(i) Name sticker to avoid mixing (ii) Devise for comfortable


among personnel resting on hand
(iii) Colour of the micropipette black (iv) Easy tip purging
or grey
(v) Slim design to negotiate through (vi) Autoclavable
narrow neck containers upto a
volume of 50 ml.
(vii) Simple for cleaning and servicing (viii) Colour coded caps, \
matching stickers for selection of appropriate tips.

19. Some common methods for calibration of micropipette are the following with the exception of (Mark ).
(i) Pipetting of coloured solutions (ii) Weighing distilled water or
followed by spectrophotometric mercury
analysis
(iii) Pipetting volumes of radioisotopes (iv) Commercially available
spectrophotometric calibration kit
(i) Weighing the micropipette at weekly
interval for 4 week

20. The 99% Sensitivity of an HIV test means that:


(A)
(i) The test will correctly identify (ii) The test will be negative in
99 to 100 infected individuals 99% of uninfected persons
(iii) The technician must be careful in (iv) The accuracy of the test
running the test divided by its cost
(v) The test will rarely give a false
positive results

114
Annexure VII.....contd.
(B) The 99% Specificity of an HIV test means that:
(i) The test will correctly identify (ii) The test will be negative in
99 to 100 infected individuals 99% of uninfected persons
(iii) The technician must be careful in (iv) The accuracy of the test
running the test divided by its cost
(v) The test will rarely give a false
positive results

115
Annexure VIII
Name of the NRL / SRL Conducting the Training Programme
(Sample Questionnaire)
Workshop on External Quality Assessment Programme for HIV Testing

Participant Group: Laboratory supportive personnel

Pre-Test / Post-Test Evaluation


Time :30 min.

Name of the Participant:

Name of the Institute:

TICK ( ) THE CORRECT ANSWER

1. 3rd generation ELISA uses the following as antigen of HIV


(i) Infected cell lysate (ii) Recombinant glycopeptides
(iii) Synthetic peptides

2. Internal kit controls (serum) refers to following:


(i) Controls supplied with the kit (ii) Controls prepared out of own
Laboratory samples
(iii) Controls procured from commercial (iv) Controls procured from the
agency reference laboratory

3. External control (serum) refers to


(i) Controls supplied with the kit (ii) Controls prepared out of
own laboratory samples
(iii) Control procured from commercial (iv) Controls procured from the
agents reference laboratory

4. ELISA ratio or E ratio refers to


(i) Ratio of sample OD/ cut off OD (ii) Ratio of sample OD upon
negative control OD
(iii) Ratio of sample OD upon positive
control OD

5. Calculation of standard deviation of external control values over 20 observations requires.

(i) Mean value of the 20 observations (ii) Square of the individual


values
(iii) Square of the mean values of the
20 observations.

116
Annexure VIII.....contd.
6. The recommended storage temperature of most of the HIV diagnostic kits is

(i) 0C (ii) 2-8C


(iii) <-30C (iv) <-70C

7. For general purpose use the recommended concentration of sodium hypochlorite is


(i) 0.01% (ii) 0.1%
(iii) 1% (iv) 10%

8. To mop up accidental spillage of blood, the recommended concentration of


Sodium hypochlorite is
(i) 0.01% (ii) 0.1%
(iii) 1% (iv) 10%

9. In the multiple test strategy for diagnosis of HIV infection in an individual

(i) The specimen is tested thrice by a particular rapid test with high sensitivity.
(ii) The specimen is tested by indirect EIA from three different companies that employ identical antigens for
coating the EIA plate.
(iii) The specimen is tested by three tests (EIA/rapid) based on different principles / different antigen
preparations

10. Ideal choice of a kit for use in a blood bank is the one with
(i) Maximum sensitivity(100%) (ii) Maximum specificity
Third generation/fourth generation

11. Discrimination between HIV-1 and HIV-2 infections is possible by


(i) All EIA kits (ii) All rapid / visual kits
(iii) Some rapid / visual kits

12. On a particular day while running routine EIA for detection of HIV reactive units among donated blood
samples it was observed that all the wells in the plate including the control well did not show any colour at the
end of the run. The possibilities considered are: (Mark ) the one you think valid.

(i) The plate is not coated with (ii) Washing has not been proper or
antigen omitted at the stage following
addition of conjugate
(iii) Substrate has been broken (iv) Plate has been incubated more than
down before addition the recommended time
(v) Stop solution has been added (vi) The plate has been exposed to light
late in the final stage for long time after addition of substrate

13. On a day of EIA run, the whole plate, including the positive control wells show colour. The possibilities
considered are as follows: (Mark ) the one you consider valid:
(i) Washing has not been done (ii) Substrate has lost its potency
properly
(iii) Conjugate has been deteriorated (iv) Incubation has been carried
out for less than recommended period
117
Annexure VIII.....contd.
14. Most commonly employed laboratory tests for diagnosis of HIV infection.
(i) Serological diagnosis by (ii) HIV isolation
antibody detection
(iii) Serological diagnosis by antigen (iv) Haematological studies
detection
(iv) Skin test

15. Collection of blood for diagnosis of HIV infection by ELISA


(i) Clotted blood (ii) Heparinised blood
(iii) Blood in tissue culture media

16. Mention the appropriate storage temperature for 3 days for serum for serological test of HIV infection:
(i) 2-8C (ii) 0C
(iii) Less than 20C

17. Mention the storage temperature of ELISA kits for HIV


(i) 2-8C (ii) 0C
(iii) Less than 20C

18. The most commonly used disinfectant in a blood bank / HIV testing laboratory is
(i) Absolute alcohol (ii) Formaldehyde
(iii) Dettol (iv) Sodium hypochlorite

19. The recommended working concentration of sodium hypochlorite is (put tick against the appropriate
concentration)
(i) ELISA plate / tips used - .001% .01% 0.1% 1% 5-10% 1
in routine serology 100%
(ii) Blood spillage - .001% .01% 0.1% 1% 5-10%
100%

20. ELISA positivity in an HIV infected means that person has AIDS:
(i) True (ii) False

118
Annexure IX

STATES ALLOTED TO THE NATIONAL REFERENCE CENTERS

Sr.No. National Reference Centers State

1. Christian Medical College, Vellore Kerala, Lakshadweep

2. Madras Medical College, Chennai Tamil Nadu, Pondicherry

3. National AIDS Research Institute, Pune Maharashtra, Goa, Gujarat

4. Institute of Immunohaematology, Mumbai Mumbai, Madhya Pradesh

5. All India Institute of Medical Sciences, Himachal Pradesh,


New Delhi Chandigarh and Punjab

6. National Institute of Communicable Rajasthan, Haryana, Delhi,


Diseases, Delhi Jammu & Kashmir

7. National Institute of Cholera & Enteric Assam, Orissa, Andaman &


Diseases, Calcutta Nicobar Islands

8. School of Tropical Medicine, Calcutta West Bengal, Bihar, Sikkim

9. Regional Institute of Medical Sciences, Manipur, Tripura, Arunachal


Imphal Pradesh, Mizoram, Nagaland

10. National Institute of Biologicals, Noida Uttar Pradesh

11. MGR Medical University, Chennai Andhra Pradesh

12. National Institute of Mental Health & Karnataka


Neurosciences (NIMHANS), Bangalore

119
ANNEXURE X
PIPETTING TECHNIQUE
The following information is consolidated from operational instruction manuals from several pipette manufacturers.
Complete information and more detailed instructions may be seen in the specific pipette instruction manuals. Read
the manufacturer's manual carefully before beginning the pipetting procedure.

Choose the appropriate disposable pipette tips for your application.


Hold the pipette in one hand and use a slight twisting movement to seat the tip firmly on the shaft of
the pipette to ensure an air-tight seal
Select the desired volume (with manual pipettes, higher volumes should be set first; if adjusting from
a lower to a higher volume, first surpass the desired volume and then slowly decrease the volume
until the required setting is reached).
If applicable, select the desired mode (e.g., reverse pipette). This is recommended for optimal
precision and reproducibility.
Reverse pipetting can be done with a manual pipette by pressing the control button slightly past the
first stop when aspirating, taking up more liquid than will be dispensed, then pressing the control
button only to the first stop when dispensing. A small volume will remain in the tip after dispensing.
Select an appropriate tip.

Pre-rinsing
The following procedures will help ensure optimal precision and accuracy.

Volumes >10 L: Prerinse pipette 23 times for each new tip (this involves aspirating and dispensing liquid
several times). Reasons for prerinsing include the following:
to compensate for system pressure, for slight differences in temperature between pipette and liquid,
and for properties of the liquid;
to clear the thin film formed by the liquid on the inside of the pipette. Without prerinsing, retention of a
thin film on the inside wall of the tip would cause the first volume to be too small. The thickness and
nature of this film, and therefore the potential source of error, will vary depending on the nature of the
liquid being pipetted.

Volumes <10 L: Do not prerinse pipette, but rinse tip after dispensing to ensure that the whole volume was
dispensed. For smaller volumes, prerinsing is not recommended because the dispensed volume may be
higher than the requisite volume.

Filling
Make sure tip is securely attached.
Hold pipette upright.
When aspirating, try to keep the tip at a constant depth below the surface of the liquid.
Glide control button slowly and smoothly (electronic pipettes perform this step automatically).
When pipetting viscous liquids (e.g., whole blood), leave the tip in the liquid for 12 seconds after
aspirating before withdrawing it.
After liquid is in the tip, never lay the pipette on its side.

120
Dispensing
Hold the tip at an angle, against the inside wall of the vessel/ tube if possible.
Glide control button slowly and smoothly (electronic pipettes perform this step automatically).

Testing pipetting precision


The precision of pipetting should be evaluated periodically (e.g., monthly or the lab can establish its own frequency) to
ensure the accuracy of results. All records of this evaluation procedure must be retained for quality assurance
purposes.
Using the reverse pipetting technique, pipette 10 replicates of blood and record the weights. Select a
volume normally used in the performance of the assay.
Using the reverse pipetting technique, pipette 10 replicates of bead suspension and record the
weights (this applies to methods in which the beads must be pipetted into the tubes).
Calculate the mean, standard deviation, and coefficient of variation (CV). The CV for replicates
should be <2%.

Testing pipetting accuracy


The following procedure can be used to test the pipette and how accurately it measures volume. Water is used
because the weight of 1 L of water is 1 g.
Using the reverse pipetting technique, pipette 10 replicates of distilled water and record the weight.
(100 L of water should weigh 0.1000 grams.) (Table 1)
Calculate the mean, standard deviation, and CV. The CV must be <2% (range: 0.0980.102).

Table 1. EVALUATION OF PRECISION AND ACCURACY

TO EVALUATE : ACCURACY PRECISION PRECISION


Replicate # 100 l of water (grams) 100 l of blood (grams) 100l of microbeads
(grams)
1 0.1036 0.1072 0.1056
2 0.1018 0.1071 0.1056
3 0.1020 0.1067 0.1055
4 0.1026 0.1069 0.1056
5 0.1008 0.1067 0.1052
6 0.1002 0.1060 0.1055
7 0.0989 0.1072 0.1056
8 0.1019 0.1090 0.1047
9 0.1009 0.1070 0.1050
10 0.1027 0.1066 0.1050
Mean 0.1015 0.1070 0.1053
S.D. 0.0014 0.0008 0.0003
% CV 1.35 0.73 0.31

121
Other Recommendations :

To ensure optimal performance, the temperature of the pipetted solution and the pipette and tips
should be the same (volume errors may occur because of changes in air displacement and viscosity
of the liquid). Do not pipette liquids with temperatures >70C.
Volume errors may also occur with liquids that have a high vapor pressure or a density/viscosity that
differs greatly from water. Water is most commonly used to calibrate pipettes and to check inaccuracy
and imprecision. A pipette could possibly be recalibrated for liquids with densities that vary greatly
from that of water.
Pipettes should be checked regularly for precision and accuracy.
Regular maintenance (e.g., cleaning) should be performed either by the user or a service technician
according to manufacturer's instructions.
All calibration and maintenance records must be maintained.

122

Potrebbero piacerti anche