SERVICE MANUAL
Oct 22, 2013

Direction: CHSM i3-001


Rev. 2.2

CHISON MEDICAL IMAGING CO., LTD.

We reserve the right to make changes to this manual without prior notice.
CHISON i3 MANUAL V2.1

TABLE OF CONTENT
CHAPTER1 GENERAL DESCRIPTION ...................................................................................................... - 1 -
1.1 INTRODUCTION .......................................................................................................................................... - 1 -
1.2 SERVICE POLICY......................................................................................................................................... - 1 -
1.3 SERVICE PROCEDURE ................................................................................................................................. - 1 -
CHAPTER2 SYSTEM OVERVIEW .............................................................................................................. - 2 -
2.1 SYSTEM STRUCTURE .................................................................................................................................. - 2 -
2.2 ELECTRICAL SCHEMATIC DIAGRAM .......................................................................................................... - 4 -
CHAPTER 3 GETTING STARTED............................................................................................................... - 6 -
3.1 INSTALLATION FOR MAIN UNIT .................................................................................................................. - 6 -
3.2 INSTALLATION FOR VIDEO PRINTER ........................................................................................................ - 10 -
3.2.1 Connect the printer from the rear panel ................................................................................... - 10 -
3.2.2 Connect the printer from the side video box............................................................................ - 12 -
3.3 INSTALLATION FOR PC PRINTER .............................................................................................................. - 15 -
3.4 SETTING FOR NETWORK........................................................................................................................... - 20 -
3.5 SETTING FOR EXTERNAL MONITOR .......................................................................................................... - 22 -
3.6 SETTING FOR DICOM .............................................................................................................................. - 24 -
CHAPTER 4 DISASSEMBLY AND ASSEMBLY ...................................................................................... - 27 -
4.1 DISASSEMBLY AND ASSEMBLY OF MONITOR & TUMBLER ...................................................................... - 27 -
4.1.1 Disassembly and Assembly of Monitor ................................................................................... - 27 -
4.1.2 Disassembly and Assembly of Tumbler .................................................................................. - 29 -
4.2 DISASSEMBLY AND ASSEMBLY OF MAIN UNIT ........................................................................................ - 31 -
4.2.1 Disassembly and Assembly of Keyboard ................................................................................ - 31 -
4.2.2 Disassembly and Assembly of Ultrasound section .................................................................. - 32 -
4.2.3 Replacement and Cleaning of Filter Board.............................................................................. - 33 -
4.2.4 Disassembly and Assembly of Back Panel .............................................................................. - 35 -
4.2.5 Disassembly and Assembly of Rear cover .............................................................................. - 36 -
4.2.6 Disassembly and Assembly of PC section ............................................................................... - 37 -
4.2.7 Disassembly and Assembly of Power Section ......................................................................... - 38 -
4.2.8 Disassembly and Assembly of Printer ..................................................................................... - 39 -
4.2.9 Disassembly and Assembly of DVD-R/W Driver ................................................................... - 40 -
CHAPTER 5 SOFTWARE UPGRADE ....................................................................................................... - 42 -
5.1 ABOUT THE SYSTEM SOFTWARE .............................................................................................................. - 42 -
5.2 INSTRUCTION FOR UPGRADE .................................................................................................................... - 43 -
CHAPTER 6 TROUBLESHOOTING AND MAINTENANCE ................................................................. - 48 -
6.1 CONTACT INFORMATION .......................................................................................................................... - 48 -
6.2 TROUBLESHOOTING ................................................................................................................................. - 49 -
6.3 MAINTENANCE ........................................................................................................................................ - 50 -
6.3.1 Host Clean ............................................................................................................................... - 50 -
6.3.2 Probe maintenance ................................................................................................................... - 51 -
6.3.3 Disinfection ............................................................................................................................. - 52 -
CHISON i3 MANUAL V2.1

Chapter1 General Description

1.1 Introduction

This guide covers the preparation, use, and maintenance of the ultrasound system, transducers, and
accessories. Refer to the manufacturers instructions for specific information about peripherals.
Please read the information in this guide before providing service to i3 color ultrasound system. It
applies to the ultrasound system, and transducers.

1.2 Service policy

CHISON provides life-time service for all produced ultrasound machines. Meanwhile, according to
the warranty card enclosed in the packing box, the corresponding free warranty period is provided
for the main unit and probe (transducer) from the date of Ex-factory.

CHISON Service Dept provides efficient service solutions to customers timely, including technical
support, spare parts support, field service, upgrade, warranty extension, and on demand service.

The engineer must have ultrasound training before provide service, CHISON can provide varieties of
training, including operation, application , maintenance, and troubleshooting.

1.3 Service Procedure

If you encounter difficulty with proving service to CHISON products, please refer to the user manual
and service manual firstly to correct the problem.

If the problem is not still solved there, please contact with CHISON Service Dept at the following
numbers or addresses:

CHISON website: www.chison.com.cn


Service Support: Tel: 0086-510-85311707
Fax: 0086-510-85310726
E-mail: service@chison.com.cn
CHISON i3 MANUAL V2.0

Chapter2 System Overview

CHISON i Series is fully digital and high resolution ratio Color Doppler diagnostic system. It
can create high qualified B/W& chromatic pictures to satisfy the demands of customers, and enhance
the veracity and acceptance. The system adopting the digital wave technology, which is based on PC,
brief interfaces, its your ideal choice.
Each product has 4 probe holders, they are compatible with convex, and linear phase-array
probe etc. CHISON i series has multiple measure function and well-integrated software.
By the advanced technology, this series has some distinctive functions which are image
post-processing, THI(Tissue Harmonic Imaging), Real-time, Triplex, Trapezoidal Imaging,
Auto-Doppler and Auto-Measurement etc. PC based Technology allows the additions and upgrading.
This system is appropriate for connecting with the other equipments, such as video printer, USB
port and LAN port etc.
DICOM(Digital Imaging and Communications in Medicine) is one selectable choice.

CAUTION: Chinese law prohibits to use medical appliance to identify the gender of fetus, apart
from in demand.

2.1 System Structure

(Fig 1)

-2-
CHISON i3 MANUAL V2.0

No. Name Function & Description

<1> Monitor Display ultrasound scanning and parameters


<2> Front frame of monitor Monitor Component, fixed part of monitor
<3> Tumbler Monitor Component, fixed part of monitor

<4> Keyboard upper panel Control Panel Component, man-machine interface

<5> Machine handle Handle for moving the whole unit

<6> Keyboard rear cover Control Panel Component, fixed part of monitor

<7> Video Printer Box Video printer storage room


<8> Left cover Safety block for main unit

<9> Front cover Safety block for main unit

<10> Foot wheel Move & fix part of the whole unit
<11> Back cover of monitor Safety block for Monitor

<12> Rear cover of monitor Safety block for Monitor

<13> First order cover of tumbler Monitor Component, fixed part of monitor

<14> Upper cover Safety block for main unit

<15> Back cover Safety block for main unit


<16> Back I/O panel Input/output of system signal

<17> Filter plate Safety block for main power section

<18> Wheel Move & fix part of the whole unit

<19> Front-upper cover Safety block for main unit

<20> Right cover Safety block for main unit

-3-
CHISON i3 MANUAL V2.0

2.2 Electrical Schematic Diagram

The whole system is mainly composed of Probe Module, Control Module, Ultrasound Module,
Power supply Module, Monitor Module, PC Module etc. Below diagram helps to learn basic
electrical schematic diagram when there is fault.

Control Panel <14>

<9> <13>
Hard disk

<12>
<8> CD Driver

<11>
Monitor <7> Video
Module <6> Distributor <10>

<23>
Ultrasound
Module <3> <2> PC Module

<15> <16> <17>

<4> <24> PCLINK BOARD


<5> Power Supply Module <18> <19> <20>

BACKLINK BOARD
<1>

<21>
Back Panel <22>

A/C
power
input

-4-
CHISON i3 MANUAL V2.0

No. Name Function & Description

<1> Power Input Cable General Power input to the Power section
<2> PC Power Input Cable Power Input Cable to PC Power Module
<3> Ultrasound Power Input Cable Power Input Cable to PC Power Module

<4> Monitor Power Input Cable Power Input Cable to Monitor

<5> Control panel Power Input Cable Power Input Cable to control panel

<6> Video distributor Power Input Cable Power Input Cable to Video distributor

<7> VGA signal output VGA signal to Monitor


<8> P CD driver ower Input Cable Power Input Cable to CD driver

<9> HDD Power Input Cable Power Input Cable to Hard disk

<10> VGA Signal input Cable Video input cable to Video Distributor
<11> VGA signal output VGA signal to BACKLINK board

<12> CD Driver Power input cable Power input to CD Driver

<13> HDD Power input cable Power input to Hard disk

<14> Control panel Power input cable Power input to Control panel

<15> Audio cable Audio input cable


<16> USB extension cable USB signal output

<17> Video & footswitch signal cable Print signal output

<18> Audio cable Audio input cable to PCLINK board

<19> USB extension cable USB signal output to PCLINK board


<20> Video & footswitch signal cable Print signal output to PCLINK board

<21> General signal output General signal output to back panel

<22> LAN cable LAN signal output


<23> Ultrasound Signal Cable Signal I/O Cable between Ultrasound Section and PC

<24> Power input cable Power input to BACKLINK board

-5-
CHISON i3 MANUAL V2.0

Chapter 3 Getting Started

3.1 Installation for Main unit

1) Acclimation Time
After being transported, the unit requires one hour for each 2.5 increment if its temperature
is below 10 C or above 40 C.

NOTE:
Please keep at least 20 to 30 cm spare space away from the back of the system to ensure well
ventilation. Otherwise, with the increasing of the temperature inside the unit, malfunction
may occur.
2) Connecting and Using the System
To connect the system to the electrical supply:
Check the power voltage input labeling at rear panel of the system.
Ensure that the wall outlet is of the appropriate type and well grounded.
Ensure that the system powers off.
Unwrap the power cable, and allow sufficient slack in the cable so that the plug will not
be pulled out of the wall outlet if the system is moved slightly.
Attach the power plug to the system and secure it in place by using the retaining clamp.
Push the power plug securely into the wall outlet.

CAUTION
Use caution to ensure that the power cable does not disconnect during system use.
If the system is accidently unplugged, data may be lost.

WARNING
To avoid risk of fire, the system power must be supplied from a separate, properly
rated outlet.
Under no circumstances should the AC power plug be altered, changed, or adapted to
a configuration rated less than specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a hospital grade or hospital only
grounded power outlet.

-6-
CHISON i3 MANUAL V2.0

Specifications of Power Plug and Power Supply Cord delivered with this system:

object/part manufacturer/trademark type/model technical standard mark(s) of


no. data conformity1)
Zhenjiang Huayin
Power Plug VDE
Instrument and Electrical 3VTJ2 16A, 250V VDE 40012265
(EU) 0620-1
Equipment Co., Ltd.
Power Zhenjiang Huayin
Supply Cord Instrument and Electrical H05VV-F 3G 0.75mm2 VDE 0620 VDE 40026359
(EU) Equipment Co., Ltd.
Power Plug TAIWAN LINE TEK 13A, 125V, UL 430
LP-20 UL E70782
(US) ELECTRONICS CO LTD hospital grade UL 817
Power 300V, 105C,
TAIWAN LINE TEK
Supply Cord SJT VW-1, UL 62 UL E138949
ELECTRONICS CO LTD
(US) 16AWG

3) Power On

Press the main power switch on the back panel firstly, then press to turn on the
system.

Power Up Sequence:
The system is initialized and start-up status is reflected on the monitor:
The control panel flashes and get dark, the system is checking BIOS data
Booting the system
Loading software
HINTS
The power up procedure takes about approx. 150 seconds at most. If a problem occurs, take
a picture and record the error information for service reference.

4) Power Off
To power off the system:
Press the power switch on the control panel, then choose turn off item.
The shutdown process takes a few seconds and is completed when the illumination of
LCD and control panel shuts down.

NOTE
If the system hangs or has not fully shut down, press and hold down the power switch longer
than 4 seconds until the system shuts down.

-7-
CHISON i3 MANUAL V2.0

5) Probes
CAUTION
Only use the probes approved by Manufacturer.

Selecting probes
Always start out with a probe that provides optimum focal depths and penetration for the
patient size and exam.
Begin the scanning session by choosing the correct application and preset for the
examination by selecting the exam icon.
Begin an exam using the default Power Output setting for the probe and exam.

Connecting the Probe


When you connect the probes, please ensure that the probe ports are not active. Place the
system in EXAM screen by pressing to deactivate the probe ports.
To connect a probe:
Place the probes carrying case on a stable surface and open the case.
Carefully remove the probe and unwrap the probe cord.
DO NOT allow the probe head to hang free. Impact to the probe head could result in
irreparable damage. Use the probe cable hanger to wrap the cord.

CAUTION
Inspect the probe before and after each use for damage or degradation to the housing,
strain relief, lens, seal and connector. DO NOT use a probe that appears damaged until
its functional and safe performance is verified. A thorough inspection should be
performed during the cleaning process.

Align the connector with the probe port and carefully push into place with the cable
facing the front of the system.
Turn the probe connector locking lever to lock status.
Carefully position the probe cord so it is free to move and is not resting on the floor.
When the probe is connected, it will be automatically initialized.

CAUTION

Fault conditions can result in electric shock hazard. DO NOT touch the surface of
probe connector that is exposed when the probe is removed. DO NOT touch the patient
when connecting or disconnecting a probe.
Take precautions with probe cables DO NOT bend the cable acutely.

-8-
CHISON i3 MANUAL V2.0

Probe connector Unlock status Probe connector Lock status

Deactivating the Probe


When deactivating the probe, the probe is automatically placed in a standby mode.
To deactivate a probe:
Ensure the system is in EXAM screen. If necessary, press the key to return to
Probe Selection screen.
Gently wipe the excess gel from the probe surface.
Carefully slide the probe toward the probe holder, and place the probe gently in the
probe holder.

Disconnecting the Probe


Probes can be disconnected when the system is in Probe Selection screen.
To disconnect a probe:
Turn the connector locking lever to a Unlock position.
Pull the probe and connector straight out of the probe port.
Carefully slide the probe and connector away from the probe port.
Ensure that the probe head is clean before placing the probe in its storage box.

Transporting the Probe


When transporting a probe a long distance, store it in its original carrying case.

Storing the Probe


It is recommended that all probes should be stored in the original carrying case.
Place the probe connector into the carrying case.
Carefully wind the cable into the carrying case.
Carefully place the probe head into the carrying case. DO NOT use excessive force or
impact on the probe head.

-9-
CHISON i3 MANUAL V2.0

3.2 Installation for Video Printer

Note: When you connect the printers, please make sure both i3 system and printer are
power off.

3.2.1 Connect the printer from the rear panel

1) Connection. It needs three cables: Remote cable, Video signal cable, Power cable. See
picture in Fig.2-1.

Video signal cable

Remote cable

Power cable

Fig.2-1

2) Connect the remote cable to remote port on the rear panel of ultrasound system. See picture
in Fig.2-2.

Fig. 2-2

- 10 -
CHISON i3 MANUAL V2.0

3) Connect the video signal cable to the TV port of the ultrasound system as below. See picture in
Fig. 2-3.

Fig. 2-3

4) Turn on the printer.

5) Turn on the i3 system, the printer will work normally.

- 11 -
CHISON i3 MANUAL V2.0

If you choose to connect the printer from the side video box, please follow below steps to set
the printer and system.

3.2.2 Connect the printer from the side video box

1) Connect the video printer to the TV cable, Power input cable and Remote cable in the video
box.

2) System setting for Video Printer.

Press the setup key; enter "system" interface, then select "Display Setting ". See picture
in Fig.2-4.

Fig.2-4

- 12 -
CHISON i3 MANUAL V2.0

Choose "Extended Desktop" for Operation Mode.

Choose Primary device for Monitor and Secondary Device for Television, then press
apply. See picture in Fig.2-5

Fig.2-5

The below dialog box will appear, and press OK. See picture in Fig.2-6.

The screen will be black for a while.

Fig.2-6

Set the video print option to choose different methods for video print. Choose the
General setting, and select Keyboard submenu. See picture in Fig. 2-7.

Fig.2-7
- 13 -
CHISON i3 MANUAL V2.0

Choose video print under print key menu or Foot SW menu, and set the video print
option.

Fig.2-8

Note:
Image only means only print the ultrasound image.
Standard means print the ultrasound image with patient information.

Note:
It need to turn on the printer firstly, or i3 system may not recognize the printer.
When you use the video printers, do NOT attach any device to SVIDEO port on
the i3 system.
If choose to connect the video printer from the rear panel, there will no need to
change display settings on the main system.

- 14 -
CHISON i3 MANUAL V2.0

3.3 Installation for PC printer

Note: At present, we install HP universal Printing driver in the Q5 system before delivery, so
some of supported HP model printers will work once they are connected.

1) Below is an example for installing supported HP printer model.

- Place the printer smoothly.


- Connect the printer to the system, by LAN port or USB port.
- Set the print manager, choose HP universal Printing PCL6 as default.
- Choose PC print in system setting interface, and in the PC print selection, choose
print the image with information, or only print image. Besides this, also you can
choose different size of the print paper.

2) Install driver for unsupported printer.

Note: With the installation HP universal Printing PCL6 as an example. The Q5 system
adopts Windows XP as platform, so please choose the right driver for the system.

- Insert U disk with installation driver to the system, click Install Driver

- Select the driver, click Install, the system will jump out installation guide.

- 15 -
CHISON i3 MANUAL V2.0

Complete the installation by following up the guide. The steps are as below

- Click Yes

- Select Dynamic Mode, click Install

- Being installed automatically

- 16 -
CHISON i3 MANUAL V2.0

- Click Finish to finish the printer installation.

- AddClick Add, complete adding printer according to operation guide.

Click NEXT

- Select Local printer attached to this computer, and Click Next.

- 17 -
CHISON i3 MANUAL V2.0

- Select Use the following port, and click Next.

- Select the printer, and click "Next".

- Select Keep existing driver (recommended), and Click Next.

- 18 -
CHISON i3 MANUAL V2.0

- Select Yes, and Click "Next".

- Select Yes to print a test page, and click Next to continue.

- Click Finish.

Note: Dont forget to set the IP if you are using network printer. You may refer to below content.

- 19 -
CHISON i3 MANUAL V2.0

3.4 Setting for Network

1) Turn on the printer or other PC Devices, plug the cable into the LAN port of the system.

2) Press key to enter into the system setting unit. Click the page of Network Settings to
enter into the network settings interface.

REMARKS:

- Network Adapte: Display the existing network cards on this system. The user can select the network
card for current use
- Obtain an IP address automatically: When this option is selected, the system will obtain an IP
address automatically
- Use the following IP address (recommended): Set the IP address of the system manually. After
setting, click Apply to confirm the application.
- IP address: Input the IP address of the system manually, and make sure the IP address of the
network printer is in the same network segment.
- Subnet Mask: Input the subnet mask manually.
- Default Gateway: Input the gateway manually.
- DNS server: Input DNS server manually.
- Network Connection Status: Check the network connection status.
- Connected: The system is connected to the network successfully.
- Unconnected: The system is not connected to the network.
- Network Information: Display the current network information
- Configuration information: Click this button to display the network configuration information
- For more information: Click this button to display network details
- Release of the connection: Click this button to disconnect the current connection
- Update Connection: Click this button to update the current connection status

- 20 -
CHISON i3 MANUAL V2.0

- Network test: In the case of network connection status is connected, enter the IP address of the test,
click the test, you can test the network the connection status between the IP of current system and the IP
that needs to be tested.

3) Below is an example for setting, finally click apply.

- Device setting

IP: 192.168.21.202
Subnet mask: 255.255.255.0
Gateway: 192.168.21.254

- System setting

IP Address: 192.168.21.112 (The 4th letter should be different.)


Subnet mask: 255.255.255.0
Gateway: 192.168.21.254
DNS Server: Define a name as you like.

- 21 -
CHISON i3 MANUAL V2.0

3.5 Setting for External monitor

1) Choose Display setting

2) Display Settings: The settings include display device, display Settings, color correction, shortcut, etc.

3) Choose IntelRDual Display Clone, then choose monitor item if there is an monitor connected.

Remarks:

- Single DisplaySingle display


- IntelRDual Display Clone: The display content of two screens are the same. you need to
choose this item if there has an external monitor connected.
- Extended DesktopExpanded display

- Display SettingsSettings of display.


- Color Quality: 32Bit, 16Bit can be chosen.

- 22 -
CHISON i3 MANUAL V2.0

- Screen ResolutionResolution of the Screen.


- Refresh RateRefresh Rate of screen
- RotationRotation of the screen

- Color CorrectionIn this interface, users can correct the color. Adjust the color of monitor.

- Hot Keys: The setting of hot keys. In this interface, users can set the shortcut key

- 23 -
CHISON i3 MANUAL V2.0

3.6 Setting for DICOM

Note: If the DICOM function is not open, the system setting interface will not display DICOM
interface. Please make sure the DICOM function is opened before you use DICOM function.

1) Connect the machine into LAN before setting DICOM

2) Press tool key to enter into system setting interface. Press DICOM item to enter into DICOM
interface.

Remarks:
- Add: Press this icon to add a DICOM service link
- Modify: Select the DICOM link which need to be modified. Press this icon to modify the
parameters inside.
- Delete: Select the DICOM link which need to be deleted. Press this icon to delete the
DICOM link.
- Verify: Select the DICOM link which need to be verified. Press this icon to verify the
current DICOM link status.
- Default: If there are several service links of same type, you can select one of them as the
default one.

- 24 -
CHISON i3 MANUAL V2.0

3) Below is parameters setting for DICOM storage , after all the settings are completed, please
click OK to save current setting. Click clear to clear all the setting. Clicking cancel will not save
the current setting.

Remarks:
- Service: Click the drop list to select storage function.
- AE title: Name the server at will
- Host: Set the IP of the service host
- Port: Make the port of the work station the same as the port of the server.
- Alias: The name of current link
- Time out: time delay
- Mode: batch, follow, manual selectable
Batch: all the saved images in current exam will be sent to the server and printed
automatically in one time
Follow: When one image is saved, the system will send this image to the server.
Manual: The user has to select the image by manual.

4) Operation for DICOM storage

- Batch mode: In batch mode, when the patient finishes his exam, press End key to end the
current exam. All the saved images will be saved as DCM format and sent to the server
automatically.

- Follow mode: In follow mode, during the exam, press key to save the image. The
current image will be saved as DCM format and sent to the server.

- Manual mode: In manual mode, the user has to select the image by manual and do DICOM

transfer. Press key to enter into review interface. Select the images need to be sent,

- 25 -
CHISON i3 MANUAL V2.0

press key to send.


5) Below is parameters setting for DICOM print , after all the settings are completed, please click OK to
save current setting. Click clear to clear all the setting. Clicking cancel will not save the current setting.

Service: Click the drop list to select print function.


AE title: Name the server at will
Host: Set the IP of the server
Port: Make the port of the work station the same as the port of the server.
Alias: The name of current link
Time out: time delay
Mode: batch, follow, manual selectable
Batch: all the saved images in current exam will be sent to the server and printed
automatically in one time
Follow: When one image is saved, the system will send this image to the server and
print it.
Manual: The user has to select the image by manual and print it.

6) Operation for DICOM print

- Batch mode: In batch mode, when the patient finishes his exam, press End key to end the
current exam. All the saved images will be saved as DCM format and sent to the server
automatically to be printed.

- Follow mode : In follow mode, during the exam, press save key to save the image. The current
image will be saved as DCM format and sent to the server to be printed.

- Manual mode: In manual mode, the user has to select the image by manual and do DICOM
transfer. Press browse key to enter into review interface. Select the images need to be sent,

click icon to print.

- 26 -
CHISON i3 MANUAL V2.0

Chapter 4 Disassembly and Assembly

4.1 Disassembly and Assembly of Monitor & Tumbler

4.1.1 Disassembly and Assembly of Monitor

Maintenance of monitor needs to take off the monitor from the tumbler first, then dismount the
monitor and go on with maintenance.

1) Dismounting the back cover of monitor: Remove the 2 screws on the back cover of the
monitor, see Fig 4-1

Back cover of
monitor

2 screws

Fig 4-1

2) After dismounting the back cover of monitor, appears 6 screws(M4)take the power supply
line and data linethen remove the screws, see Fig.4-2:

- 27 -
CHISON i3 MANUAL V2.0

Power supply cable


6 screws on monitor back panel

VGA data line


8 pcs screws on the 4 edges of back
cover

Fig 4-2

3) Remove the back cover of monitor:8 pcs screws on the 4 edges of back cover, when
removed, the back cover is movable.

4) Disassembly and assembly of front frame of monitor: Remove 16 pcs around the front
frame of monitor, then take out the front frame ,see Fig 4-3

Carrier of monitor

Front frame of
monitor

Fig 4-3
5) Remove screws on the shield of case, then take out monitor;

- 28 -
CHISON i3 MANUAL V2.0

4.1.2 Disassembly and Assembly of Tumbler

1) Take out second order cover of tumbler;

2) Remove 3 pcs screws with socket head wrench;

3) Move up the tumbler, take out the tumbler from main unit, see Fig 4-4

2nd order cover of tumbler

Fig 4-4

4) Remove the cover of tumbler: The screws of first order cover and second order cover of
tumbler are located at the bottom of the tumbler, see Fig 6,the cover of second order can be
removed by finger.

- 29 -
CHISON i3 MANUAL V2.0

5) Separate first order tumbler and second order tumbler :Remove socket head cap screw,
pressing piece of tumbler, damping ring of tumbler on the first order of tumbler, see Fig 4-5
and Fig 4-6 and Fig 4-7

Upper cover of first


order of tumbler

Rear cover of 1st


order tumbler

Upper cover of 2nd


order of tumbler

Rear cover of 2nd Fig 4-5 Fig 4-6


order tumbler

Fig 4-7

- 30 -
CHISON i3 MANUAL V2.0

4.2 Disassembly and Assembly of Main Unit

4.2.1 Disassembly and Assembly of Keyboard

Remove screws, then separate the keyboard. Fixing screws which keep the keyboard fixed
are located at the bottom side of keyboard,16 pcs self-tappig screws totally see Fig 4-8

(Fig 4-8)

- 31 -
CHISON i3 MANUAL V2.0

4.2.2 Disassembly and Assembly of Ultrasound section

Ultrasound power section located at the back of the Front-rear cover, take the front-rear cover to take
power supply out .

1) Disassemble front cover: remove 2 pcs scews, see Fig.4-9,then press cover dowm, then take
out the front cover;

Front cover

Screws,2 pcs

Fig 4-9

2) Take out the ultrasound section: Remove 4 pcs screws in front of the ultrasound section,
take out the ultrasound section, see Fig 4-10

CAUTION: REMOVE POWER SUPPLY LINE, AVOID TAKING OUT THE CABLE!

- 32 -
CHISON i3 MANUAL V2.0

Handle

Ultrasound unit

Screws 6 pcs

Fig 4-10

3) Remove the cover of ultrasound section: Remove 8 pcs screws, then take out the cover of
ultrasound section and replace boards, see Fig 4-11

Fig 4-11

4.2.3 Replacement and Cleaning of Filter Board

Pull the left-side of the filter board, take out the filter, see Fig 4-12

- 33 -
CHISON i3 MANUAL V2.0

Rubber cap 7 pcs

Filter board

Fig 4-12
Wash filter board with water, dry it up, then put it back.

- 34 -
CHISON i3 MANUAL V2.0

4.2.4 Disassembly and Assembly of Back Panel

1) Remove filter board, see 1-1-5,remove 7 pcs of rubber caps, see Fig 4-12
2) Remove 10 pcs screws ,see Fig 4-13

Rubber cap
7pcs

Bottom of the rear


cover,3 pcs of screws

Fig 4-13

- 35 -
CHISON i3 MANUAL V2.0

4.2.5 Disassembly and Assembly of Rear cover

Remove 3 pcs of screws located at right side of the rear cover, see Fig 4-14

Screws 3 pcs

Fig 4-14

- 36 -
CHISON i3 MANUAL V2.0

4.2.6 Disassembly and Assembly of PC section

1) Remove 3 pcs screws on the back of the Pc section, see Fig 4-15

CAUTION: REMOVE THE POWER CABLE AND LINE FIRST!

PC section

Screws 3pcs

Fig 4-15
2) Remove screws on the PC section ,then remove the cover of PC section, see Fig 4-16

Screws 8 pcs

Fig 4-16

- 37 -
CHISON i3 MANUAL V2.0

4.2.7 Disassembly and Assembly of Power Section

1) Remove screws 3pcs, then take out the power section, see Fig 4-17:

CAUTION: REMOVE THE POWER CABLE AND LINE FIRST.

Power section

3 Screws

Fig 4-17
2) Remove screws 9 pcs, then remove the cover of the power section, see Fig 4-18:

Screws 9 pcs

Fig 4-18

- 38 -
CHISON i3 MANUAL V2.0

4.2.8 Disassembly and Assembly of Printer

Take out the cover of printer, pull the printer out, see Fig 4-19 and Fig 4-20

CAUTION: REMOVE THE POWER CABLE AND LINE FIRST!

Cover of printer Printer

Fig 4-19 Fig 4-20

- 39 -
CHISON i3 MANUAL V2.0

4.2.9 Disassembly and Assembly of DVD-R/W Driver

1 Remove the rear cover.


2 Remove the cover :remove the back cover, and remove 4 pcs screw, take out the cover, see
Fig 4-21:

4 Screws

Fig 4-21
3 Remove the front cover, see Fig 4-22
4 Remove the front-upper cover: remove 2 pcs screws, take out the cover, see Fig 24:

Screws 2 pcs

(Fig 4-22

- 40 -
CHISON i3 MANUAL V2.0

5 Remove fixing 4 screws, see, Fig 4-23,Fig 4-24:

2 screws

2 Screws
Fig 4-23 Fig 4-24

- 41 -
CHISON i3 MANUAL V2.0

Chapter 5 Software Upgrade

5.1 About the System Software

The ultrasound system contains software that controls its operation. Typically new software provides
new capabilities, CHISON will provide the software upgrade service for lifetime.

A single Flash disk can be used to upgrade one or more systems, and software upgrade use the USB
port on the left hand side of the system.

Upgrades provided may be required or optional. Normally, we dont suggest the owner upgrade the
system if not necessary, because there has risk cause the system corrupt by incorrect operation or
power failure.

- 42 -
CHISON i3 MANUAL V2.0

5.2 Instruction for upgrade

1) Please contact with Chison Service Dept get the newest upgrade package, including two files:
fpga.ZIP and upgrade.ZIP

2) Unzip them, you will get two folders as below; upgrade folder and fpga folder, copy them to
the root of U disk.

- 43 -
CHISON i3 MANUAL V2.0

3) Please confirm when you open the folders, there should be subfile in the folder;

4) Insert the U disk to the USB port on side panel.

- 44 -
CHISON i3 MANUAL V2.0

5) Press key on the keyboard to the system settings interface. Select System item. You
can see the color of items SW upgrade and FW upgrade change to white from gray.

6) Click the SW upgrade item firstly and confirm YES , the system will upgrade the software
automatically.

Note: The whole procedure is about 5minutes.

- 45 -
CHISON i3 MANUAL V2.0

7) The system will restart when you confirm OK after upgrade successed.

Note: The upgrade procedure is not finshed after you upgrade the software, you have to
upgrade the FPGA to work with the software.

8) Please go to System interface again and click the item FW upgrade to upgrade, then the
system will upgrade automatically after you confirm YES.

Note: The whole procedure is about 16 minutes.

- 46 -
CHISON i3 MANUAL V2.0

9) The unit will turn off when you confirm OK after upgrading is successful.

10) Please check the system information after the whole upgrade procedure.

Possible causing for upgrade failure:


1) .Pls confirm the U disk is a good one, or you may try with another U disk.
2).The volume for U disk (2GB is the best), too small or too big volume is not easy for machine to
recognize.
3) .Never unplug the U disk during upgrade process.
4) .Please confirm there will be subfiles once you open the upgrade folder or fpga folder.
5). The upgrade package is damaged during the download procedure, you can try to download
again.
6). Please check the system information, whether you forget upgrade software or fpga

- 47 -
CHISON i3 MANUAL V2.0

Chapter 6 Troubleshooting and Maintenance

6.1 Contact information

This chapter contains information to help you correct problems with system operation and provides
instructions on the proper care of the system, transducer, and accessories.

Troubleshooting:
If you encounter difficulty with the system, use the information in this chapter to help correct the
problem. If the problem is not covered here, contact Chison Service Dept at the following numbers
or addresses:

CHISON website: www.chison.com.cn


Service Support: CHISON Medical Imaging Co., Ltd.
Tel: 0086-510-85311707
Fax: 0086-510-85310726
E-mail: service@chison.com.cn
Service Skype ID: chison-frank chison-cathy chison-nick

- 48 -
CHISON i3 MANUAL V2.0

6.2 Troubleshooting

Symptom Solution

1. The system cant power on. 1)Check power connections, e.g. power cord connection
on the rear panel;
2)Check the fuse: if it is burnt due to mains fluctuation,
use spare fuse for replacement.
2. When starting the system, the 1) Switch off the system, and check probe connection.
monitor has signal but no 2) Check whether the probe knock is broken, so you
ultrasound image. cant connect it tightly.
3) Check whether you use the right version probe, even
the probe model is right.
3.The whole system image 1) Adjust the LCD monitor position for a better viewing
quality is not good. angle;
2) Change a place, socket for a stable voltage;
3) Adjust the image parameters, e.g. Gain, Dynamic
range.
4.No OB calculation package 1)Select the OB application before scanning.
menu.
5. PRINT-key doesnt work. 1) Check if the approved printer is connected;
2) Check if the printer power is on;
3) Check printer connection;
4) Check printer setting in system setup. (Refer to
Chapter 3)
6.The External monitor doesnt 1) Check the monitor connections;
work. 2) Check if the monitor power is on and is set up
correctly.
7.CFM or PW Doppler image 1) Adjust CFM or PW gain value properly;
has noise. 2) Check if there is appliance or equipment resulting in
strong electromagnetic interference;
8. Image has interference. 1) Change a room to avoid interference source;
2) Use separate power outlet;
3) Perform good ground protection
9.The gray scale is S- twisted in 1) Adjust the power supply to normal voltage or use a
the image area. voltage stabilizer
10.The date and time on the 2) Press Setup-key to display General Setting screen,
screen is not correct. and correct time and date.

- 49 -
CHISON i3 MANUAL V2.0

6.3 Maintenance

It is necessary to maintain the system regularly, as it can ensure the system being operated under safe
state by eliminating possible trouble, and it can shorten the checking and repair period, lower the
service costs and reduce the operation danger.

6.3.1 Host Clean

Caution

Before clean any part of the system, please make sure that the system is turned off and the
power cable is disconnected from the power supply socket. Otherwise there will be danger of
electricity shock.
Cleaning method:
Please use a piece of soft and dry cloth to clean the system. If theres some dirty difficult to be
cleaned, please use wet cloth to clean system, and then use dry cloth to wipe off the water on the
system.

Caution:

1. Please dont use organic solvent to clean; otherwise it will damage the exterior paint.
2. Please never allow any liquid get inside the system, otherwise it will damage the system and
cause electronic short.
If the probe connector and any peripheral device are required to be cleaned, please contact our
authorized agent in your country in advance. Any cleaning by unauthorized person may result in
system malfunction or affect its features.

- 50 -
CHISON i3 MANUAL V2.0

6.3.2 Probe maintenance

According to the purposes, the probes of this system can be divided into 2 categories: those used
on the surface of patient body, or those used inside the body of patient.

Caution

No matter which type of examination is performed, please always try to reduce the unnecessary
radiation of ultrasound wave to the patient during the ultrasound examination

Caution

1. Probe can only be used by professional doctor who has received professional training of
ultrasound.
2. It is forbidden to sterilize and disinfect probe by high pressure. If it needs to be used in
sterilized occasion, please use a sterilized disposable probe cover on the probe.
3. Please avoid drop off or hitting the probe by anything.
4. Dont scratch the probe surface while using it.
5. Please use the authorized ultrasound gel during scanning. Using un-authorized gel may
cause scratch or damage to probe surface.
6. Dont bend or pull the probe cable with force.
7. Please dont put the probe connector and its adjacent probe cable part into any liquid.
8. Please keep the probe clean and dry.
9. Please connect or disconnect the probe after the system is turned off or freeze the image.
10. Please dont use or preserve the probe where it is over 50.
11. Please carefully check the probe surface, probe cable and probe connector before using. If
there is any abnormal phenomenon (e.g. theres a leakage on the probe surface), please stop
using the probe immediately and contact our authorized agent in your country as soon as
possible. If you dont know the contact number of your authorized agent, please contact us
by detail contact information at the end of this chapter.

Note:

For surface probe, it shouldnt be immersed in the liquid deeper than the acoustic lens. For
transvaginal probe, it shouldnt be deeper than the insertion region. It is strictly prohibited to
let probe connected plug into any liquid

- 51 -
CHISON i3 MANUAL V2.0

6.3.3 Disinfection

Disinfection is applicable to transvaginal probe.


If required in the operation of using the probe, please follow the guidance of personnel to
disinfect the probe.
Disinfection procedures
1. Clean the probe thoroughly according to the procedure prescribed.
2. Prepare the glutaraldehyde solution (concentration is 2%) as the disinfection solution
according to the manufacturers instructions.
3. Put the probe into the liquid, the depth shouldnt be deeper than the acoustic lens. It is
strictly prohibited to let probe connected plug into any liquid
4. Immersion probe for 3 hours
5. After removing the probe, clean it with sterile water or sterile saline immediately. In order
to ensure that there is no solution residue must perform manufacturer rinsing procedures
strictly according to the provisions, including adequate rinsing with water and rinse times.
6. If it needs to be used in sterilized occasion, please use a sterilized disposable probe cover
on the probe.

Note:

1. It is strictly prohibited to let probe connected plug into any liquid.


2. For transvaginal probe, it shouldnt be immersed in the liquid deeper than the insertion
region.

- 52 -