Technical Information

Proline Promag 50H, 53H


Electromagnetic Flow Measuring System
Flow measurement in food, beverage, pharmaceutical, or
process applications

Applications Features and benefits


Electromagnetic flowmeter for bidirectional Promag measuring devices offer you cost-effective flow
measurement of liquids with a minimum conductivity measurement with a high degree of accuracy for a wide
of 5 S/cm: range of process conditions.
Beverages, e.g. fruit juice, beer, wine
Milk products, fruit mixtures The Proline transmitter concept comprises:
Salt solutions Modular device and operating concept resulting in a
Acids and causic solutions, etc. higher degree of efficiency
Software options for batching, electrode cleaning and
Flow measurement up to 1250 gpm (4700 dm3/min) pulsating flow
Fluid temperature up to 300F (+150C) Uniform operating concept
Process pressures up to 580 psi (40 bar)
CIP/SIP cleaning The tried-and-tested Promag sensors offer:
Stainless steel housing No pressure loss
Approvals in the food industry/hygiene sector: Insensitivity to vibrations
3-A authorization, EHEDG-tested, conformity with Simple installation and commissioning
FDA
Application-specific lining materials:
PFA
Approvals for hazardous area:
ATEX, FM, CSA, TIIS
Connection to process control system:
HART, PROFIBUS DP/PA, FOUNDATION
Fieldbus, MODBUS RS485

TI048D/24/ae/09.07
Proline Promag 50H, 53H

Function and system design


Measuring principle Faradays law of induction states that a voltage is induced in a conductor moving in a magnetic field.
In electromagnetic measuring, the flowing medium corresponds to the moving conductor. The induced voltage
is proportional to the flow velocity and is detected by two measuring electrodes and transmitted to the ampli-
fier. Flow volume is computed on the basis of the pipe's diameter. The constant magnetic field is generated by
a switched direct current of alternating polarity.

Ue

V
B
I
L

A0003191
Ue = B L v
Q=Av

Ue = induced voltage
B = magnetic induction (magnetic field)
L = electrode gap
v = flow velocity
Q = volume flow
A = pipe cross-section
I = current strength

Measuring system The measuring system consists of a transmitter and a sensor.


Two versions are available:
Compact version: transmitter and sensor form a single mechanical unit.
Remote version: transmitter and sensor are installed separately.

Transmitter:
The Promag 50/53 transmitters incorporate a high impedance amplifier of 1 x 1012 ohms or greater.
Promag 50 (user interface with push buttons for operation, two-line display).
Promag 53 (Touch Control without opening the housing, four-line display).

Sensor nominal sizes:


1/12" to 4" (DN 2 to 100)

2 Endress+Hauser
Proline Promag 50H, 53H

Input
Measured variable Flow rate (proportional to induced voltage)

Measuring range Typically v = 0.033 to 33 ft/s (0.01 to 10 m/s) with the specified measuring accuracy

Operable flow range Over 1000 : 1

Input signal Status input (auxiliary input):


U = 3 to 30 VDC, Ri = 5 k, galvanically isolated.
Configurable for: totalizer(s) reset, measured value suppression, error-message reset.

Status input (auyiliary input) with PROFIBUS DP and MODBUS RS485:


U = 3 to 30 VDC, Ri = 3 k, galvanically isolated
Switching level: 3 to 30 VDC, independent of polarity
Configurable for: totalizer(s) reset, measured value suppression, error-message reset,
batching start/stop (optional), batch totalizer reset (optional)

Current input (for Promag 53 only):


Active/passive selectable, galvanically isolated, full scale value selectable, resolution: 3 A,
temperature coefficient: typ. 0.003% o.r./F (0.005% o.r./C), o.r. = of reading
active: 4 to 20 mA, Ri 150 , Uout = 24 V DC, short-circuit-proof
passive: 0/4 to 20 mA, Ri 150 , Umax = 30 V DC

Output
Output signal Promag 50
Current output:
active/passive selectable, galvanically isolated, time constant selectable (0.01 to 100 s),
full scale value selectable, temperature coefficient: typ. 0.003% o.r./F (0.005% o.r./C), o.r. = of reading,
resolution: 0.5 A
active: 0/4 to 20 mA, RL < 700 (HART: RL 250 )
passive: 4 to 20 mA, operating voltage VS 18 to 30 V DC, Ri 150

Pulse/frequency output:
passive, open collector, 30 V DC, 250 mA, galvanically isolated.
Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1,
pulse width max. 10 s.
Pulse output: pulse value and pulse polarity selectable, max. pulse width configurable
(0.5 to 2000 ms)

PROFIBUS DP interface:
Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A: 1998,
galvanically isolated
Profile version 3.0
Data transmission rate: 9.6 kBaud to 12 MBaud
Automatic data transmission rate recognition
Function blocks: 1 x analog input, 3 x totalizer
Output data: volume flow, totalizer
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device

Endress+Hauser 3
Proline Promag 50H, 53H

PROFIBUS PA interface:
Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
Profile version 3.0
Current consumption: 11 mA
Permissible supply voltage: 9 to 32 V
Bus connection with integrated reverse polarity protection
Error current FDE (Fault Disconnection Electronic): 0 mA
Function blocks: 1 x analog input, 1 x totalizer
Output data: volume flow, totalizer
Input data: positive zero return (ON/OFF), control totalizer, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device

Promag 53

Current output:
active/passive selectable, galvanically isolated, time constant selectable (0.01...100 s),
full scale value selectable, temperature coefficient: typically 0.005% o.r./C (o.r. = of reading),
resolution: 0.5 A
active: 0/4 to 20 mA, RL < 700 (HART: RL 250 )
passive: 4 to 20 mA, operating voltage VS 18 to 30 V DC, Ri 150

Pulse/frequency output:
active/passive selectable, galvanically isolated (Ex i version: only passive)
active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100
passive: open collector, 30 V DC, 250 mA

Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), EEx-ia: 2 to 5000 Hz;
on/off ratio 1:1; pulse width max. 10 s.
Pulse output: pulse value and pulse polarity adjustable, pulse width configurable
(0.05 to 2000 ms)

PROFIBUS DP interface:
Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A:
1998, galvanically isolated
Profile version 3.0
Data transmission rate: 9.6 kBaud to 12 MBaud
Automatic data transmission rate recognition
Function blocks: 2 x analog input, 3 x totalizer
Output data: volume flow, calculated mass flow, totalizer 1 to 3
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device
Available output combination Page 7 ff.

PROFIBUS PA interface:
Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
Profile version 3.0
Current consumption: 11 mA
Permissible supply voltage: 9 to 32 V
Bus connection with integrated reverse polarity protection
Error current FDE (Fault Disconnection Electronic): 0 mA
Function blocks: 2 x analog input, 3 x totalizer
Output data: volume flow, calculated mass flow, totalizer 1 to 3
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device

4 Endress+Hauser
Proline Promag 50H, 53H

MODBUS interface:
Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A:
1998, galvanically isolated
MODBUS device type: Slave
Adress range: 1 to 247
Bus address adjustable via miniature switches or local display (optional) at the measuring device
Supported MODBUS function codes: 03, 04, 06, 08, 16, 23
Broadcast: supported with the function codes 06, 16, 23
Transmission mode: RTU oder ASCII
Supported baudrate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
Response time:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
Available output combination Page 7 ff.

FOUNDATION Fieldbus interface:


FOUNDATION Fieldbus H1
Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
ITK version 4.01
Current consumption: 12 mA
Error current FDE (Fault Disconnection Electronic): 0 mA
Bus connection with integrated reverse polarity protection
Function blocks: 5 x analog input, 1 x discrete output, 1 x PID
Output data: volume flow, calculated mass flow, temperature, totalizer 1 to 3
Input data: positive zero return (ON/OFF), reset totalizer
Link Master (LM) functionality is supported

Signal on alarm Current output failure response selectable (e.g. in accord. with NAMUR Recom. NE 43)
Pulse/frequency output failure response selectable
Status output (Promag 50) non-conductive by fault or power supply failure
Relay output (Promag 53) de-energized by fault or power supply failure

Load See Output signal

Switching output Status output (Promag 50, Promag 53):


Open collector, max. 30 V DC / 250 mA, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values.

Relay outputs (Promag 53):


Normally closed (NC or break) or normally open (NO or make) contacts available (default: relay 1 = NO,
relay 2 = NC),
max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values,
batching contacts.

Low flow cutoff Switch points for low flow cutoff are selectable

Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.

Endress+Hauser 5
Proline Promag 50H, 53H

Power supply
Electrical connection
Measuring unit C
A B

d d
g g
b b

a
a
a

b d/(g) (d)

HART* PROFIBUS PA*


FOUNDATION Fieldbus*

27 PA()/FF() 27
+ 26 PA(+)/FF(+) 26
25 d 25 d
+ 24 + 24
f 23 e f 23 e
+ 22 + 22
21 21
+ 20 + 20

N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b

PROFIBUS DP* PROFIBUS DP**


MODBUS RS485**

A (RxD/TxD-N) 27 A (RxD/TxD-N) 27
B (RxD/TxD-P) 26 B (RxD/TxD-P) 26
25 d 25 d
+ 24 + 24
f 23 e f 23 e
+ 22 + 22
21 21
+ 20 g + 20 g

N (L-) 2 c N (L-) 2 c
L1 (L+) 1 L1 (L+) 1
b b

a0002441
Connecting the transmitter, cable cross-section max. 16 AWG (2.5 mm2)
AView A (field housing)
BView B (stainless steel field housing)
CView C (wall-mount housing)
*) non-changeable communication board
**) changeable communication board
a Cover of the connection compartment
b Cable for power supply: 85 to 260 VAC, 20 to 55 VAC,16 to 62 VDC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
c Ground terminal for protective conductor
d Signal cable: see Terminal assignment Page 7 ff.
Fieldbus cable:
Terminal No. 26: DP (A) / PA (+) / FF (+) / MODBUS RS485 (A) / (PA, FF: with reverse polarity protection)
Terminal No. 27: DP (B) / PA (-) / FF (-) / MODBUS RS485 (B) / (PA, FF: with reverse polarity protection)
e Ground terminal for signal-cable shield / Fieldbus cable / RS485 line
f Service connector for connecting service interface FXA 193 (Fieldcheck, Tof Tool - Fieldtool Package)
g Signal cable: see Terminal assignment Page 7 ff.
Cable for external termination (only for PROFIBUS DP with fixed assignment communication board):
Terminal No. 24: +5 V
Terminal No. 25: DGND

6 Endress+Hauser
Proline Promag 50H, 53H

Terminal assignment, Promag 50

Terminal No. (inputs / outputs)


Order variant 20 (+) / 21 () 22 (+) / 23 () 24 (+) / 25 () 26 (+) / 27 ()
Current output
50***-***********W
HART
Current output
50***-***********A Frequency output
HART
Current output
50***-***********D Status input Status output Frequency output
HART
PROFIBUS PA
50***-***********H

+5 V (external PROFIBUS DP
50***-***********J
termination)
Frequency output Current output IS
50***-***********S
IS, passive active, HART
Frequency output Current output IS
50***-***********T
Ex i, passive passive, HART
Ground connection, power supply Page 6

Terminal assignment, Promag 53


The inputs and outputs on the communication board can be either permanently assigned or variable, depending
on the version ordered (see table). Replacements for modules which are defective or which have to be replaced
can be ordered as accessories.

Terminal No. (inputs / outputs)


Order variant 20 (+) / 21 () 22 (+) / 23 () 24 (+) / 25 () 26 (+) / 27 ()
Fixed communication boards (fixed assignment)
Current output
53***-***********A Frequency output
HART
Current output
53***-***********B Relay output 2 Relay output 1 Frequency output
HART
PROFIBUS PA
53***-***********F
IS
FOUNDATION
53***-***********G
Fieldbus, IS
PROFIBUS PA
53***-***********H

PROFIBUS DP
53***-***********J

FOUNDATION
53***-***********K
Fieldbus
MODBUS RS485
53***-***********Q Status input

Frequency output Current output, IS


53***-***********S
IS active, HART
Frequency output Current output, IS
53***-***********T
IS passive, HART
Flexible communication boards
Current output
53***-***********C Relay output 2 Relay output 1 Frequency output
HART
Current output
53***-***********D Status input Relay output Frequency output
HART
Current output
53***-***********L Status input Relay output 2 Relay output 1
HART

Endress+Hauser 7
Proline Promag 50H, 53H

Terminal No. (inputs / outputs)


Order variant 20 (+) / 21 () 22 (+) / 23 () 24 (+) / 25 () 26 (+) / 27 ()
Current output
53***-***********M Status input Frequency output Frequency output
HART
53***-***********N Current output Frequency output Status input MODBUS RS485
53***-***********P Current output Frequency output Status input PROFIBUS DP
53***-***********V Relay output 2 Relay output 1 Status input PROFIBUS DP
Current output
53***-***********2 Relay output Current output Frequency output
HART
Current output
53***-***********4 Current input Relay output Frequency output
HART
Current output
53***-***********5 Status input Current input Frequency output
HART
53***-***********7 Relay output 2 Relay output 1 Status input MODBUS RS485
Ground connection, power supply Page 6

Electrical connection remote


version
Electrode circuit Coil circuit

Meas. signal Pipe EPD

S1 E1 E2 S2 GND E S

6 5 7 8 4 37 36 42 41

yellow
brown
white

green

2
1
n.c. n.c. n.c.

5 7 4 37 42 41

E1 E2 GND E

1-1/2 to 4 1/12 to 1

F06-5xHxxxxx-04-xx-xx-en-001
n.c. = isolated cable shields, not connected

Cable entry Power-supply and signal cables (inputs/outputs):


Cable entry M20 x 1.5 (8 to 12 mm)
Threads for cable entries 1/2" NPT, G 1/2"

Connecting cable for remote version:


Cable entry M20 x 1.5 (8 to 12 mm)
Threads for cable entries 1/2" NPT, G 1/2"

8 Endress+Hauser
Proline Promag 50H, 53H

Cable specifications Coil cable:


remote version 2 x 18 AWG (0.75 mm2) PVC cable with common, braided copper shield ( approx. 0.28" / 7 mm)
Conductor resistance: 37 /km
Capacitance core/core, shield grounded: 120 pF/m
Permanent operating temperature: -5 to +180F (20 to +80C)
Cable cross-section: max. 16 AWG (2.5 mm2)

Signal cable:
3 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield ( approx. 0.28" / 7 mm) and
individually shielded cores.
With Empty Pipe Detection (EPD): 4 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield
( approx. 0.28" / 7 mm) and individually shielded cores.
Conductor resistance: 50 /km
Capacitance core/shield: 420 pF/m
Permanent operating temperature: -5 to +180F (20 to +80C)
Cable cross-section: max. 16 AWG (2.5 mm2)

1
2
3
4
5
6
7
a b

A0003194
a = signal cable, b = coil current cable (cross-section: max. 16 AWG / 2.5 mm2)
1 = core, 2 = core insulation, 3 = core shield, 4 = core jacket, 5 = core strengthening,
6 = cable shield, 7 = outer jacket

Optionally, Endress+Hauser also supplies flexible metal conduit. We recommend such cables for the following
cases:
Cables laid underground
Danger of rodent attack
Device used with ingress protection NEMA 6P (IP 68)

Operation in zones of severe electrical interference:


The measuring device complies with the general safety requirements in accordance with EN 61010,
the EMC requirements of EN 61326/A1, and NAMUR recommendation NE 21.

Caution!
Grounding is by means of the ground terminals provided for the purpose inside the connection housing.
Keep the stripped and twisted lengths of cable shield to the terminals as short as possible.

Supply voltage 85 to 260 VAC, 45 to 65 Hz


20 to 55 VAC, 45 to 65 Hz
16 to 62 VDC

PROFIBUS PA and FOUNDATION Fieldbus


Non-hazardous: 9 to 32V DC
Intrinsically safe: 9 to 24 VDC
Explosion proof: 9 to 32 VDC

Power consumption AC: <15 VA (including sensor)


DC: <15 W (including sensor)

Switch-on current:
max. 13.5 A (< 50 ms) at 24 VDC
max. 3 A (< 5 ms) at 260 VAC

Endress+Hauser 9
Proline Promag 50H, 53H

Power supply failure Lasting min. 1 power cycle:


EEPROM or T-DAT (Promag 53 only) retain the measuring system data in the event of a power supply failure
S-DAT: exchangeable data storage chip which stores the data of the sensor (nominal diameter, serial number,
calibration factor, zero point, etc.)

Potential equalization Perfect measurement is only ensured when the medium and the sensor have the same electrical potential.

Metal process connections


Potential matching usually takes place over the metallic process connection in contact with medium process
connections which are directly mounted on the measuring transmitter. This usually means that additional
potential matching measures are unnecessary.

Plastic process connections


For plastic process connections, potential matching must be ensured between sensor and medium using addi-
tional ground rings. If these ground rings are missing, this can influence accuracy or destroy the measuring
transmitter through the electrochemical decomposition of electrodes.

2 3 2 4

A0002651
1 Allen screw (process connection)
2 O-ring seals
3 Plastic washer (spacer) or ground ring
4 Sensor

When using ground rings, note the following points:


Depending on the option ordered, plastic washers may be installed at the process connections instead of
ground rings. These plastic washers serve only as spacers and have no potential equalization function. In
addition, they provide a sealing function at the interface between the sensor and process connection. For this
reason, with process connections without ground rings, these plastic washers/seals must not be removed,
or must always be installed.
Ground rings can be ordered separately from Endress+Hauser as an accessory.
Also make sure that the ground rings are compatible with the electrode material. Otherwise the danger
exists that the electrodes could be destroyed by electrochemical corrosion. You can find material data on
Page 38 ff.
Ground rings, incl. seals, are mounted inside the process connection.
This has no influence on the installation length. You can find the dimensions of ground rings on Page 33.

10 Endress+Hauser
Proline Promag 50H, 53H

Performance characteristics
Reference operating To DIN EN 29104 and VDI/VDE 2641:
conditions Medium temperature: +28C 2 K
Ambient temperature: +22C 2 K
Warm-up period: 30 minutes

Installation:
Inlet run >10 x Dia.
Outlet run > 5 x Dia.
Sensor and transmitter grounded.
Sensor centered relative to the pipe.

Maximum measured error Promag 50:


Pulse output: 0.5% o.r. 1 mm/s (o.r. = of reading)
Current output: plus typically 5 A

Promag 53:
Pulse output: 0.2% o.r. 2 mm/s (o.r. = of reading)
Current output: plus typically 5 A

Supply voltage fluctuations have no effect within the specified range.

[%]
2.5

2.0
0.5 %
1.5
0.2 %
1.0

0.5

0
0 3 6 13 20 26 33
(1) (2) (4) (6) (8) (10)
v in ft/s [m/s]

F06-5xxxxxxx-05-xx-xx-xx-000
Max. measured error in % of reading

Repeatability max. 0.1% o.r. 0.5 mm/s (o.r. = of reading)

Endress+Hauser 11
Proline Promag 50H, 53H

Operating conditions

Installation conditions
Installation instructions Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
Highest point of a pipeline. Risk of air accumulating.
Directly upstream from an open pipe outlet in a vertical pipe.

h  2 x Dia

F06-5xxxxxxx-11-00-00-xx-000

Installation of pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and the
consequent risk of damage to the lining of the measuring tube. Information on the lining's resistance to partial
vacuum can be found on Page 18.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic
pumps. Information on the measuring system's resistance to vibration and shock can be found on Page 17.

F06-5xxxxxxx-11-00-00-xx-001

12 Endress+Hauser
Proline Promag 50H, 53H

Partially filled pipes


Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection (EPD)
function offers additional protection by detecting empty or partially filled pipes.

Caution!
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a
cleaning valve.

 2 x Dia.

 5 x Dia.

F06-5xxxxxxx-11-00-00-xx-002

Vertical pipes
Install a siphon (b) or a vent valve (a) downstream of the sensor in vertical pipes longer than 15 ft (5 meters).
This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube.
These measures also prevent the system losing prime, which could cause air inclusions. Information on the
linings resistance to partial vacuum can be found on Page 18.

> 15 ft (5 m)

F06-5xxxxxxx-11-00-00-xx-003
a = vent valve, b = siphon

Endress+Hauser 13
Proline Promag 50H, 53H

Orientation
An optimum orientation helps avoid gas and air accumulations and deposits in the measuring tube. Promag,
nevertheless, supplies a range of options and accessories for correct measuring of problematic mediums:
Electrode Cleaning Circuitry (ECC) to remove electrically conductive deposits in the measuring tube, e.g. in
accretive mediums.
Empty Pipe Detection (EPD) for recognition of partially filled measuring tubes, or for degassing mediums or
for applications with fluctuating process pressure (only for 1/2" to 4" / DN 15 to 100).

Vertical orientation:
This orientation is ideal for self-emptying piping systems and for use in conjunction with Empty Pipe Detection.

F06-5xxxxxxx-11-00-00-xx-004

Horizontal orientation:
The measuring electrode-plane should be horizontal. This prevents brief insulation of the two electrodes by
entrained air bubbles.

Caution!
Empty Pipe Detection functions correctly only when the measuring device is installed horizontally and the
transmitter housing is facing upward. Otherwise there is no guarantee that Empty Pipe Detection will respond
if the measuring tube is only partially filled or empty.

2 2
A

F06-5xxxxxxx-11-00-xx-xx-001
1 = EPD electrode (Empty Pipe Detection) except for Promag H 1/12" and 5/32" (DN 2, 4) sensors
2 = Measuring electrodes (signal detection)

14 Endress+Hauser
Proline Promag 50H, 53H

Vibrations
Secure the piping and the sensor if vibration is severe.

Caution!
It is advisable to install sensor and transmitter separately if vibration is excessively severe. Information on
resistance to vibration and shock can be found on Page 17.
The sensor might require support or additional attachments, depending on the application and the length of
the piping run. When plastic process connections are used, the sensor must be additionally supported
mechanically. A wall-mounting kit can be ordered separately from Endress+Hauser as an accessory.

> 30 ft ( 10 m)

F06-5xxxxxxx-11-00-00-xx-006

Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the
following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy:
Inlet run 5 x Dia.
Outlet run 2 x Dia.

5 x Dia. 2 x Dia.

F06-5xxxxxxx-11-00-00-xx-005

Endress+Hauser 15
Proline Promag 50H, 53H

Adapters Suitable adapters to DIN EN 545 (double flange junction sections) can be used to install the sensor in larger-
diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving
fluids.
The nomogram shown here can be used to calculate the pressure loss caused by reducers and expanders. The
nomogram applies only to fluids of viscosity similar to water.

1. Calculate the ratio of the diameters d/D.


2. From the nomogram read off the pressure loss as a function of flow velocity (downstream from the reduc-
tion) and the d/D ratio.

psi [mbar]
1.45 (100)
26 ft/s
8 m/s
23 ft/s
7 m/s
20 ft/s
6 m/s
0.14 (10) 16 ft/s
5 m/s
13 ft/s
4 m/s
max. 8
10 ft/s
3 m/s
d D 6 ft/s
2 m/s
0.014 (1)

3 ft/s
1 m/s

0.5 0.6 0.7 0.8 0.9


d/D

F06-5xxxxxxx-05-05-xx-xx-000

Length of connecting cable Permissible cable length Lmax depends on the conductivity of the medium. A minimum conductivity of
20 S/cm is required for measuring demineralized water.

(S/cm]

200

L max 100

33 330 650 ft
(10) (100) (200) [m]

L max

F06-5xxxxxxx-05-xx-xx-xx-006
Gray shaded area = permissible range for medium conductivity
Lmax = length of connecting cable in feet [m]
Medium conductivity in [S/cm]
In order to ensure measuring accuracy, moreover, comply with the following instructions when installing the
remote version:
Secure the cable run or route the cable in a conduit. Movement of the cable can falsify the measuring signal,
particularly if the conductivity of the medium is low.
Route the cable well clear of electrical machines and switching elements.
Ensure potential equalization between sensor and transmitter, if necessary.

16 Endress+Hauser
Proline Promag 50H, 53H

Environment
Ambient temperature Transmitter:
Standard: -5 to +140F (20 to +60C)
Optional: -40 to +140F (40 to +60C)

Note!
At ambient temperatures below -5F (20C) the readability of the display may be impaired.

Sensor:
Flange material carbon steel: 15 to 140F (10 to +60C)
Flange material stainless steel: -40 to +140F (40 to +60C)

Caution!
It is not allowed to use the device beyond the min. and max. lining specified temperature values
( Medium temperature range).

Note the following points:


Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
If both fluid and ambient temperatures are high, install the transmitter at a remote location from the sensor
( Medium temperature range).

Note!
HE or Harsh Environment Protection for sensor:
Provides enhanced moisture ingress protection for humid ambient environments. Apply in applications with
severe temperature cycling and humid environments. Refer to ordering information, page 43.

Storage temperature The storage temperature corresponds to the ambient temperature range of the transmitter and sensor (see
Ambient temperature).
The measuring device must be protected against direct sunlight during storage in order to avoid unacceptably
high surface temperatures.
Choose a storage location where moisture does not collect in the measuring device. This will help prevent
fungus and bacteria infestation which can damage the liner.
Do not remove the protective plates or caps on the process connections until the device is ready to install.

Degree of protection NEMA 4X (IP 67) for transmitter and sensor

Shock and vibration resistance Acceleration up to 2 g by analogy with IEC 60068-2-6

CIP cleaning Possible

SIP cleaning Possible

Electromagnetic To EN 61326/A1 and NAMUR recommendation NE 21


compatibility (EMC)

Endress+Hauser 17
Proline Promag 50H, 53H

Process conditions
Medium temperature range The permissible medium temperature depends on the sensor and the sealing material:

Sensor:
1/12" to 4" (DN 2 to 100): -5 to +300F (20 to +150C)

Seal:
EPDM: -5 to +265F (20 to +130C)
Silicon: -5 to +300F (20 to +150C)
Viton: -5 to +300F (20 to +150C)
Kalrez: -5 to +300F (20 to +150C)

Conductivity Minimum conductivity:


5 S/cm for fluids generally
20 S/cm for demineralised water

Note that in the case of the remote version, the minimum conductivity is also influenced by the length of the
connecting cable see Length of connecting cable

Medium pressure range The permissible nominal pressure depends on the process connection and seal:
(nominal pressure) 580 psi (40 bar): flange, weld nipple (with O-ring seal)
230 psi (16 bar): all other process connections

Pressure tightness Nominal diameter Measuring tube Resistance to partial vacuum of measuring tube lining
(liner) lining Limit values for abs. pressure [mbar] at various
fluid temperatures

[mm] [inch] 75F 175F 212F 265C 300F 356F

2 to 100 1/12 to 4" PFA 0 0 0 0 0 0

Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity
of flow is 6 to 10 ft/s (2 to 3 m/s). The velocity of flow (v), moreover, has to be matched to the physical prop-
erties of the medium:
v > 6 ft/s (2 m/s:) for media forming coatings, e.g. in full-fat milk, etc.

Flow characteristics (SI units)


Nominal Recommended Factory settings
diameter flow rate

Min./max. full scale value Full scale value Pulse weighting Creepage
[mm] [inch] (v ~ 0.3 or 10 m/s) (v ~ 2.5 m/s) (~ 2 pulse/s) (v ~ 0.04 m/s)

2 1/12" 0.06 to 1.8 dm3/min 0.5 dm3/min 0.005 dm3 0.01 dm3/min

4 5/32" 0.25 to 7 dm3/min 2 dm3/min 0.025 dm3 0.05 dm3/min

8 5/16" 1 to 30 dm3/min 8 dm3/min 0.10 dm3 0.1 dm3/min

15 1/2" 4 to 100 dm3/min 25 dm3/min 0.20 dm3 0.5 dm3/min

25 1" 9 to 300 dm3/min 75 dm3/min 0.50 dm3 1 dm3/min

40 1 1/2" 25 to 700 dm3/min 200 dm3/min 1.50 dm3 3 dm3/min

50 2" 35 to 1100 dm3/min 300 dm3/min 2.50 dm3 5 dm3/min

65 2 1/2" 60 to 2000 dm3/min 500 dm3/min 5.00 dm3 8 dm3/min

80 3" 90 to 3000 dm3/min 750 dm3/min 5.00 dm3 12 dm3/min

100 4" 145 to 4700 dm3/min 1200 dm3/min 10.00 dm3 20 dm3/min

18 Endress+Hauser
Proline Promag 50H, 53H

Flow characteristics (US units)


Nominal Recommended Factory settings
diameter flow rate

Min./max. full scale value Full scale value Pulse weighting Creepage
[inch] [mm] (v ~ 1.0 or 33 ft/s) (v ~ 8 ft/s) (~ 2 pulse/s) (v ~ 0.1 ft/s)

1/12" 2 0.0150.5 gal/min 0.1 gal/min 0.001 gal 0.002 gal/min

5/32" 4 0.072 gal/min 0.5 gal/min 0.005 gal 0.008 gal/min

5/16" 8 0.258 gal/min 2 gal/min 0.02 gal 0.025 gal/min

1/2" 15 1.027 gal/min 6 gal/min 0.05 gal 0.10 gal/min

1" 22 2.565 gal/min 18 gal/min 0.20 gal 0.25 gal/min

1 1/2" 40 7190 gal/min 50 gal/min 0.50 gal 0.75 gal/min

2" 50 10300 gal/min 75 gal/min 0.50 gal 1.25 gal/min

2 1/2" 65 16500 gal/min 130 gal/min 1 gal 2.0 gal/min

3" 80 24800 gal/min 200 gal/min 2 gal 2.5 gal/min

4" 100 401250 gal/min 300 gal/min 2 gal 4.0 gal/min

Pressure loss With 5/16" to 4" (DN 8 to 100), no pressure loss if the sensor is installed in a pipe of the same nominal
diameter.
Pressure losses for configurations incorporating adapters to DIN EN 545 Page 16.

Measuring-tube specifications
Nominal Diameter Pressure rating * Inside diameter of
measuring tube **

[mm] [inch] EN (DIN) [bar] PFA, inches (mm)

2 1/12" 230 / 580 psi (PN 16 / PN 40) 0.09 (2.25)

4 5/32" 230 / 580 psi (PN 16 / PN 40) 0.12 (4.5)

8 5/16" 230 / 580 psi (PN 16 / PN 40) 0.35 (9.0)

15 1/2" 230 / 580 psi (PN 16 / PN 40) 0.63 (16.0)

1" 230 / 580 psi (PN 16 / PN 40) 0.89 (22.6)

25 230 / 580 psi (PN 16 / PN 40) 1.02 (26.0)

40 1 1/2" 230 psi (PN 16) 1.39 (35.3)

50 2" 230 psi (PN 16) 1.89 (48.1)

65 2 1/2" 230 psi (PN 16) 2.36 (59.9)

80 3" 230 psi (PN 16) 2.86 (72.6)

100 4" 230 psi (PN 16) 3.84 (97.5)

* The pressure rating depends on the process connections and the seals (see Page 18).
** Inside diameters of process connections Page 26 ff.

Endress+Hauser 19
Proline Promag 50H, 53H

Mechanical construction
Design / dimensions Dimensions: Wall-mount housing (non hazardous area and II3G / zone 2)

6.27"
Esc
(159.5)
- + E

9.84"
(250)

3.56"
(90.5)

8.46" 3.54" 1.77"


(215) (90) (45)
5.31"
(135)

3.18" 2.08" 3.18"


> 1.97" (81) (53) (81)
(50)

4.01"
(102)

2.08"
(53)

3.74"
3.20" (95)
(81.5)

0.45" 7.55" 0.45"


(11.5) (192) (11.5)

A0001150

Dimensions: Remote field housing (II2GD / zone 1)

10.4"
(265)
9.53" *
(242)
9.45 ~8.11
(240) (206)
8.54 * 7.32
(217) (186)

6.57"
(167)

13.9"
(355)

7.40"
3.94" 5.24" (188)
(100) (133)

* Blind version
0.34"
3.94"
(8.6 M8)
(100)
4.84"
(123)

A0002128

20 Endress+Hauser
Proline Promag 50H, 53H

There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an
accessory. The following installation variants are possible:
Panel-mounted installation
Pipe mounting

Panel-mounted installation

+0.02

8.27 -0.02
+0.5
-0.5
(210 )

+0.02
9.65 -0.02
+0.5
(245 -0.5
)

4.33
(110)

A0001131

Pipe mounting

0.79 to 2.75
(20 to 70)

~6.10
(155)

A0001132

Endress+Hauser 21
Proline Promag 50H, 53H

Compact version, aluminum field housing 1/12" to 1" (DN 2 to 25)

8.94" 7.68"
(227) (195)
8.15" * 6.61"
(207) (168)

6.30"
(160)
9.92" 12.5"
(252) (317)

DI 2.56"
L (65)
K

* Blind option dimension


F06-53Hxxxxx-06-00-xx-xx-001

Compact version, stainless-steel field housing 1/12" to 1" (DN 2 to 25)

8.66" 6.02"
(220) (153)

6.7"
(171) 10.3"
(263) 12.9"
(328)

DI 2.56"
(65)
K L
F06-53Hxxxxx-06-00-xx-xx-002

Nominal size Pressure ** DI L K M

[mm] [inch] psi in (mm) in (mm) in (mm) [mm]

2 1/12 230/580 0.09 (2.25) 3.39 (86) 1.69 (43) M 6x4

4 5/32 230/580 0.18 (4.5) 3.39 (86) 1.69 (43) M 6x4

8 5/16 230/580 0.35 (9.0) 3.39 (86) 1.69 (43) M 6x4

15 1/2 230/580 0.63 (16.0) 3.39 (86) 1.69 (43) M 6x4

1 230/580 0.89 (22.6) 3.39 (86) 2.09 (53) M 6x4

25 1 230/580 1.02 (26.0) 3.39 (86) 2.09 (53) M 6x4

Fitting length depends on process connections Page 27 ff.

** The permissible nominal pressure depends on the process connection and seal:
580 psi (40 bar): flange EN 1092-1 (DIN 2501), welded nipples for DIN EN ISO 1127 pipes and ODT
(with O-ring seal)
230 psi (16 bar): all other process connections

22 Endress+Hauser
Proline Promag 50H, 53H

Remote version 1/12" to 1" (DN 2 to 25)

4.92" 2.76"
(125) (70)

2.95"
(75)
B
8.39"
(213) M

DI
L K
F06-xxHxxxxx-06-05-xx-xx-001

Nominal size Pressure * DI L B C K M

[mm] [inch] psi in (mm) in (mm) in (mm) in (mm) in (mm) [mm]

2 1/12 230/580 0.09 (2.25) 3.39 (86) 5.83" (148) 2.56 (65) 1.69 (43) M 6x4

4 5/32 230/580 0.18 (4.5) 3.39 (86) 5.83" (148) 2.56 (65) 1.69 (43) M 6x4

8 5/16 230/580 0.35 (9.0) 3.39 (86) 5.83" (148) 2.56 (65) 1.69 (43) M 6x4

15 1/2 230/580 0.63 (16.0) 3.39 (86) 5.83" (148) 2.56 (65) 1.69 (43) M 6x4

1 230/580 0.89 (22.6) 3.39 (86) 5.83" (148) 2.56 (65) 2.09 (53) M 6x4

25 1 230/580 1.02 (26.0) 3.39 (86) 5.83" (148) 2.56 (65) 2.09 (53) M 6x4

Fitting length depends on process connections Page 27 ff.


Dimensions wall-mounted housing Page 20

* The permissible nominal pressure depends on the process connection and seal:
580 psi (40 bar): flange EN 1092-1 (DIN 2501), welded nipples for DIN EN ISO 1127 pipes and ODT
(with O-ring seal)
230 psi (16 bar): all other process connections

Wall-mounting kit

 0.28"
4.33" (7)
(110) 3.46"
(88)

5.51 4.92"
(140) 4.72"
(125)
(120)

F06-xxHxxxxx-06-07-06-xx-000

Endress+Hauser 23
Proline Promag 50H, 53H

Compact version, aluminum field housing 1-1/2" to 4" (DN 40 to 100)

8.94" 7.68"
(227) (195)
8.15" * 6.61"
(207) (168)

6.30"
(160)
B
A
M

DI C
K L

* Blind option dimension


F06-53Hxxxxx-06-00-xx-xx-000

Compact version, stainless-steel field housing 1-1/2" to 4" (DN 40 to 100)

8.66" 6.02"
(220) (153)

6.7"
(171)
B
A
M

DI C
K L
F06-53Hxxxxx-06-00-xx-xx-003

Nominal Press. DI L A* B* C K M

[mm] [inch] psi in (mm) in (mm) in (mm) in (mm) in (mm) in (mm) (mm)

40 1 1/2" 230 1.39 (35.3) 5.51 (140) 12.6 (319) 10.0 (255) 2.52 (64) 5.03 (128) M 8x4

50 2" 230 1.89 (48.1) 5.51 (140) 13.5 (344) 10.5 (267) 3.03 (77) 6.02 (153) M 8x4

65 2 1/2" 230 2.36 (59.9) 5.51 (140) 13.5 (344) 10.5 (267) 3.03 (77) 6.02 (153) M 8x4

80 3" 230 2.86 (72.6) 7.87 (200) 15.5 (394) 11.5 (292) 4.02 (102) 7.99 (203) M 12x4

100 4" 230 3.84 (97.5) 7.87 (200) 15.5 (394) 11.5 (292) 4.02 (102) 7.99 (203) M 12x4

Fitting length depends on process connections Page 27 ff.


* Dimensions for aluminum housing, see table below for stainless steel housing.

A, SS housing B, SS housing

in (mm) in (mm)

13.0 (330) 10.5 (266)

14.0 (355) 10.9 (278)

14.0 (355) 10.9 (278)

16.0 (405) 11.9 (303)

16.0 (405) 11.9 (303)

24 Endress+Hauser
Proline Promag 50H, 53H

Remote version 1-1/2" to 4" (DN 40 to 100)

4.92" 2.76"
(125) (70)

2.95"
(75)
B
A M

DI
L K

F06-xxHxxxxx-06-05-xx-xx-000

Nominal size Press. DI L A B C K M

[mm] [inch] psi in (mm) in (mm) in (mm) in (mm) in (mm) in (mm) [mm]

40 1 1/2" 230 1.39 (35.3) 5.51 (140) 8.50 (216) 5.96 (151.5) 2.54 (64.5) 5.03 (128) M 8x4

50 2" 230 1.89 (48.1) 5.51 (140) 9.49 (241) 6.46 (164.0) 3.03 (77) 6.02 (153) M 8x4

65 2 1/2" 230 2.36 (59.9) 5.51 (140) 9.49 (241) 6.46 (164.0) 3.03 (77) 6.02 (153) M 8x4

80 3" 230 2.86 (72.6) 7.87 (200) 11.4 (290) 7.42 (188.5) 3.99 (101.5) 7.99 (203) M 12x4

100 4" 230 3.84 (97.5) 7.87 (200) 11.4 (290) 7.42 (188.5) 3.99 (101.5) 7.99 (203) M 12x4

Fitting length depends on process connections Page 27 ff.


Dimensions wall-mounted housing Page 20

Endress+Hauser 25
Proline Promag 50H, 53H

Front view 1/12" to 1" (DN 2 to 25), without process connection

1/12 to 1/2 1
(DN 2 to 15) (DN 255)
A A-A B B-B
0.15"
0.15" (4)
(4)
2.36" 0.004"
0.004"
(0.1)
0
(0.1) 0
(60) 1.34" -0.004" 2.76" 1.91" 1.73" -0.004"

1.64" C 0 D 0

0.004" (34 )
-0.1" (70) (48.5) (44 )
-0.1"

(41.6) (0.1)
0.004"
(0.1)

M6 M6
0.004" 0.004" 0.004" 0.004"
A (0.1) (0.1) B (0.1) (0.1)
0.23"
0.23"
(6) (6)
0.33" 1.10" 0.33"
0.94"
(8.5) (28) (8.5)
(24)
2.05"
1.65"
(52)
(42)

F06-5xHxxxxx-06-05-08-xx-000

Nominal size C D (DIN) D (ANSI)


(DIN) in (mm) in ([mm) inches

1/12" to 5/16" (2 to 8) 0.35 (9)

1/2" (15) 0.63 (16)

25 (DIN) 1.02 (26)

1" ANSI (25) 0.89"

26 Endress+Hauser
Proline Promag 50H, 53H

Process connections with O-ring seals 1/12" to 1" (DN 2 to 25)

Weld nipples for DIN Sensor Fits to di G L HxB


1.4404 / 316L SS
5*H**-B*********** DN [mm] Piping [mm] [mm] [mm] [mm]
DIN EN ISO 1127

2 to 8 13.5 x 1.6 10.3 13.5 20.3 60 x 42

15 21.3 x 1.6 18.1 21.3 20.3 60 x 42

25 (DIN) 33.7 x 2 29.7 33.7 20.3 70 x 52


G HxB
di Fitting length = (2 x L) + 86 mm

L
F06-xxHxxxxx-06-

Weld nipples for IPS/SMS pipe Sensor Fits to di G L HxB


1.4404 / 316L SS
DN [mm] Piping OD/SMS [mm] [mm] [mm] [mm]
5*H**-C***********

2 to 8 13.5 x 2.3 9.0 13.5 20.3 60 x 42

15 21.3 x 2.65 16.0 21.3 20.3 60 x 42

25 (1" ANSI) 33.7 x 3.25 27.2 33.7 22.3 70 x 52


G HxB
di Fitting length = (2 x L) + 86 mm

L
F06-xxHxxxxx-06-09-

Flange PN 40 EN 1092-1 Sensor Fits to di G L LK M HxB


(DIN 2501), Form B
1.4404 / 316L SS DN [mm] Flange EN1092-1 [mm] [mm] [mm] [mm] [mm] [mm]
(DIN 2501)
5*H**-D***********

2 to 8 DN 15 17.3 95 56.2 65 14 60 x 42
M 15 DN 15 17.3 95 56.2 65 14 60 x 42

LK 25 (DIN) DN 25 28.5 115 56.2 85 14 70 x 52


G HxB
di Fitting length = (2 x L) + 86 mm
Fitting length to DVGW (200 mm)

L
F06-xxHxxxxx-06-09-07-xx-014

Flange CI 150 / ANSI B16.5 Sensor Fits to di G L LK M HxB


1.4404 / 316L SS
Flange ANSI B16.5 inches inches inches inches inches inches
5*H**-E***********

1/12" to 5/16" 1/2" 0.62 3.50 2.60 2.38 0.62 2.36 x 1.65

M 1/2" 1/2" 0.63 3.50 2.60 2.38 0.62 2.36 x 1.65

1" 1" 1.05 4.25 2.83 3.12 0.62 2.76 x 2.05


LK HxB
Fitting length = (2 x L) + 3.39" (86 mm)
G di

L
F06-xxHxxxxx-06-09-07-xx-015

Endress+Hauser 27
Proline Promag 50H, 53H

Flange 20 K / JIS B2238 Sensor Fits to di G L LK M HxB


1.4404 / 316L SS
DN [mm] Flange B2238 [mm] [mm] [mm] [mm] [mm] [mm]
5*H**-F***********

2 to 8 ND 15 15 95 67 70 15 60 x 42

M 15 ND 15 16 95 67 70 15 60 x 42

25 (DIN) ND 25 26 125 67 95 19 70 x 52
LK HxB
G Fitting length = (2 x L) + 86 mm
di

L
F06-xxHxxxxx-06-09-07-xx-016

Flange PN 16 / EN 1092-1 Sensor Fits to di G L M LK HxB


(DIN 2501)
DN [mm] Flange EN1092-1 [mm] [mm] [mm] [mm] [mm] [mm]
PVDF
(DIN 2501)
5*H**-G***********
2 to 8 DN 15 15.7 95 57 14 65 60 x 42
M 15 DN 15 15.7 95 57 14 65 60 x 42
LK 25 (DIN) DN 25 27.3 115 57 14 85 70 x 52
G HxB Fitting length = (2 x L) + 86 mm
di Fitting length to DVGW (200 mm)
The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)

L
F06-xxHxxxxx-06-09-07-xx-029

Flange CI 150 / ANSI B16.5 Sensor Fits to Flange di G L M LK HxB


PVDF
5*H**-H*********** ANSI B16.5 inches inches inches inches inches inches

1/12" to 5/16" 1/2" 0.62 3.74 2.24 0.63 2.36 2.36 x 1.65
M 1/2" 1/2" 0.62 3.74 2.24 0.63 2.36 2.36 x 1.65
LK 1" 1" 1.07 4.53 2.24 0.63 3.11 2.76 x 2.05
G HxB Fitting length = (2 x L) + 3.39" (86 mm)
di The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)

L
F06-xxHxxxxx-06-09-07-xx-029

Flange 10K / JIS B2238 Sensor Fits to di G L M LK HxB


PVDF
DN [mm] Flange B2238 [mm] [mm] [mm] [mm] [mm] [mm]
5*H**-J***********

2 to 8 ND 15 15.7 95 57 15 70 60 x 42
M 15 ND 15 15.7 95 57 15 70 60 x 42

LK 25 (DIN) ND 25 27.3 125 57 19 90 70 x 52


G HxB Fitting length = (2 x L) + 86 mm
di The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)

L
F06-xxHxxxxx-06-09-07-xx-029

28 Endress+Hauser
Proline Promag 50H, 53H

External pipe thread ISO 228 / Sensor Fits to di G L S HxB


DIN 2999
Internal thread
1.4404 / 316L SS
inches in (mm) inches in (mm) in (mm) inches
5*H**-K***********

1/12" to 5/16" R 3/8" 0.39 (10) 3/8" 1.57 (40) 0.40 (10.1) 2.36 x 1.65
S 1/2" R 1/2" 0.63 (16) 1/2" 1.57 (40) 0.52 (13.2) 2.36 x 1.65

1" R 1" 0.98 (25) 1" 1.65 (42) 0.65 (16.5) 2.76 x 2.05
G HxB
di Fitting length = (2 x L) + 3.39" (86 mm)

L
F06 H 06 09 07 025

Internal pipe thread ISO 228 / Sensor Fits to di G D L S HxB


DIN 2999, 1.4404 / 316L SS
5*H**-L*********** External thread
inches in (mm) inches in (mm) in (mm) in (mm) inches

1/12" to 5/16" Rp 3/8" 0.35 (9.0) 3/8" 0.87 (22) 1.77 (45) 0.51 (13) 2.36 x 1.65
S
1/2" Rp 1/2" 0.63 (16.0) 1/2" 1.06 (27) 1.77 (45) 0.55 (14) 2.36 x 1.65

1" Rp 1" 1.07 (27.2) 1" 1.57 (40) 2.01 (51) 0.67 (17) 2.76 x 2.05
D HxB
G Fitting length = (2 x L) + 3.39" (86 mm)
di

L
F06-xxHxxxxx-06-09-07-xx-027

Hose connection Sensor Fits to di LW L HxB


1.4404 / 316L SS
Inside diameter
5*H**-M/N/P***********
in (mm) in (mm) in (mm) in (mm) inches

1/12" to 5/16" 0.51 (13) 0.39 (10.0) 0.51 (13) 1.93 (49) 2.36 x 1.65

1/2" 0.63 (16) 0.50 (12.6) 0.63 (16) 1.93 (49) 2.36 x 1.65

1/2" 0.75 (19) 0.63 (16.0) 0.75 (19) 1.93 (49) 2.36 x 1.65
HxB
Fitting length = (2 x L) + 3.39" (86 mm)
di

L
F06-xxHxxxxx-06-09-07-xx-024

Adhesive fitting Sensor Fits to di G L HxB


PVC
pipe (mm) (mm) (mm) (mm)
5*H**-R/S***********

1/12" to 5/16" 1/2" 0.85" (21.5) 1.07" (27.3) 1.52" (38.5) 2.36" x 1.65"

DN 2 to 8 20 x 2 [mm] (20.2) (27.0) (38.5) (60 x 42)


(DIN 8062)
G HxB DN 15 20 x 2 [mm] (20.2) (27.0) (28.0) (60 x 42)
di
(DIN 8062)

Fitting length = (2 x L) + 3.39" (86 mm)


The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)
L
F06-xxHxxxxx-06-09-07-xx-028

Endress+Hauser 29
Proline Promag 50H, 53H

Process connections with aseptic gasket seal 1/12" to 1" (DN 2 to 25 )

Weld nipple for DIN Sensor Fits to di G L HxB


1.4404 / 316L SS
5*H**-U*********** DN (mm) Piping DIN 11850 (mm) (mm) (mm) (mm)

2 to 8 14 x 2 9 14 23.3 60 x 42

15 20 x 2 16 20 23.3 60 x 42

25 (DIN) 30 x 2 26 30 23.3 70 x 52
G HxB Fitting length = (2 x L) + 86 mm
di
If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.

L
A0003870

Weld nipples for ODT/SMS Sensor Fits to di G L HxB


1.4404 / 316L SS
Piping OD/SMS in (mm) in (mm) in (mm) inches
5*H**-V***********

1/12" to 5/16" 1/2" 0.35 (9.0) 0.50 (12.7) 0.63 (16.1) 2.36 x 1.65

1/2" 3/4" 0.63 (16.0) 0.75 (19.1) 0.63 (16.1) 2.36 x 1.65

1" 1" 0.89 (22.6) 1.0 (25.4) 0.63 (16.1) 2.76 x 2.05
G HxB Fitting length = (2 x L) + 3.39" (86 mm)
di
If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.

L
A0003871

Clamp ISO 2852, Fig. 2 Sensor Fits to Clamp ISO 2852 di G L HxB
1.4404 / 316L SS
Piping ISO 2037 / Diameter in (mm) in (mm) in (mm) inches
5*H**-W***********
BS 4825-1 in (mm)

1" Tube 1" x 0.64" 0.98 (25) 0.89 (22.6) 1.99 (50.5) 1.74 (44.3) 2.76 x 2.05

Fitting length = (2 x L) + 3.39" (86 mm)


If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.
G HxB
di

L
F06 xxHxxxxx 06 09 07 xx 023

Clamp DIN 32676 Sensor Fits to di G L HxB


1.4404 / 316L SS
5*H**-0*********** DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm]

2 to 8 Tube 14 x 2 (DN 10) 10 34.0 41.0 60 x 42

15 Tube 20 x 2 (DN 15) 16 34.0 41.0 60 x 42

25 (DIN) Tube 30 x 2 (DN 25) 26 50.5 44.5 70 x 52


G HxB
Fitting length = (2 x L) + 86 mm
di
If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.

L
F06-xxHxxxxx-06-09-07-xx-019

30 Endress+Hauser
Proline Promag 50H, 53H

Tri-clamp L14 AM7 Sensor Fits to di G L HxB


1.4404 / 316L SS
Piping OD in (mm) in (mm) in (mm) inches
5*H**-1***********

1/12" to 5/16" Tube 12.7 x 1.65 0.37 (9.4) 0.98 (25).0 1.12 (28.5) 2.36 x 1.65
(ODT 1/2")

1/2" Tube 19.1 x 1.65 0.62 (15.8) 0.98 (25).0 1.12 (28.5) 2.36 x 1.65
(ODT 3/4")
G HxB
di 1" Tube 25.5 x 1.65 0.87 (22.1) 1.98 (50.4) 1.12 (28.5) 2.76 x 2.05
(ODT 1")

Fitting length = (2 x L) + 3.39" (86 mm)


If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
L connection (di) into account.
A0003872

Coupling SC DIN 11851 Sensor Fits to di G L HxB


Threaded adapter
DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm]
1.4404 / 316L SS
5*H**-2***********

2 to 8 Tube 12 x 1 10 Rd 28 x 1/8" 44 60 x 42
(DN 10)

15 Tube 18 x 1 or 1.5 16 Rd 34 x 1/8" 44 60 x 42


G HxB (DN 15)
di 25 (DIN) Tube 28 x 1 or 1.5 26 Rd 52 x 1/6" 52 70 x 52
(DN 25)

Fitting length = (2 x L) + 86 mm
L If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-017

Coupling DIN 11864-1 Sensor Fits to di G L HxB


Aseptic threaded adapter, Form A
DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm]
1.4404 / 316L SS
5*H**-3***********

2 to 8 Tube 13 x 1.5 (DN 10) 10 Rd 28 x 1/8" 42 60 x 42

15 Tune 19 x 1.5 (DN 15) 16 Rd 34 x 1/8" 42 60 x 42

25 (DIN) Tube 29 x 1.5 (DN 25) 26 Rd 52 x 1/6" 49 70 x 52


G HxB
di Fitting length = (2 x L) + 86 mm
If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.

L
F06-xxHxxxxx-06-09-07-xx-021

Flange DIN 11864-2 Sensor Fits to di G L LK M HxB


Aseptic grooved flange, Form A
DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm] [mm] [mm]
1.4404 / 316L SS
5*H**-4***********
M 2 to 8 Tube 13 x 1.5 (DN 10) 10 54 48.5 37 9 60 x 42

15 Tube 19 x 1.5 (DN 15) 16 59 48.5 42 9 60 x 42


LK 25 (DIN) Tube 29 x 1.5 (DN 25) 26 70 48.5 53 9 70 x 52
G HxB
di Fitting length = (2 x L) + 86 mm
If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
connection (di) into account.

L
F06-xxHxxxxx-06-09-07-xx-022

Endress+Hauser 31
Proline Promag 50H, 53H

Coupling SMS 1145 Sensor Fits to SMS 1145 di G L HxB


Threaded adapter
DN [mm] Piping OD Diameter [mm] [mm] [mm] [mm]
1.4404 / 316L SS
5*H**-5*********** [mm]

25 (1" ANSI) 1" 25 22.6 Rd 40 x 1/6" 30.8 70 x 52

Fitting length = (2 x L) + 86 mm
If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
G HxB connection (di) into account.
di

L
F06-xxHxxxxx-06-

Process connections ordered only as accessories (with O-ring seal, 1/12" to 1" / DN 2 to 25)

External pipe thread Sensor Fits to di G L S HxB


1.4404 / 316L SS
DKH**-GD** DN [mm] NP internal thread [mm] [inch] [mm] [mm] [mm]

2 to 8 NPT 3/8" 10 3/8" 50 15.5 60 x 42


S 15 NPT 1/2" 16 1/2" 50 20.0 60 x 42

25 (1" ANSI) NPT 1" 25 1" 57 25.0 70 x 52


G HxB
di Fitting length = (2 x L) + 86 mm

L
F06-xxHxxxxx-06-09-07-xx-025

Internal pipe thread Sensor Fits to NPT di G D L S HxB


1.4404 / 316L SS
DKH**-GC** external thread in (mm) inches in (mm) in (mm) in (mm) inches

1/12" to 5/16" NPT 3/8" 0.35 (8.9) 3/8" 0.87 (22) 1.77 (45) 0.51 (13) 2.36 x 1.65
S
1/2" NPT 1/2" 0.63 (16.0) 1/2" 1.06 (27) 1.77 (45) 0.55 (14) 2.36 x 1.65

1" ANSI NPT 1" 1.07 (27.2) 1" 1.57 (40) 2.01 (51) 0.67 (17) 2.76 x 2.05
D HxB
G Fitting length = (2 x L) + 3.39" (86 mm)
di

L
F06-xxHxxxxx-06-09-07-xx-027

32 Endress+Hauser
Proline Promag 50H, 53H

Process connections orderable only as accessories (with aseptic gasket seal)

Tri-Clamp L14 AM7 Sensor Fits to di G L HxB


1.4404 / 316L SS
DKH**-HF*** Piping OD in (mm) in (mm) in (mm) in (mm)

1/2 (DN 15) Tube 25.4 x 1.5 0.87 (22.1) 1.98 (50.4) 1.12 (28.5) 2.36 x 1.65
(ODT; 1") (60 x 42)

Fitting length = (2 x L) + 3.39" (86 mm)


G If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 22) and process
HxB
connection (di) into account.
di

L
F06-xxHxxxxx-06-

Ground rings (accessories for PVDF flanges / PVC adhesive fitting)

Ground ring Sensor di D B C


1.4435/316L SS, Alloy C-22,
in (mm) in (mm) in (mm) in (mm)
tantalum
DK5HR ****

1/12" to 5/16" 0.35 (9.0) 1.33 (33.9) 0.87 (22.0) 0.69 (17.6)

1/2" 0.63 (16.0) 1.33 (33.9) 1.14 (29.0) 0.97 (24.6)

1" 0.89 (22.6) 1.73 (43.9) 1.44 (36.5) 1.23 (31.2)

25 (DIN) 1.02 (26.0) 1.73 (43.9) 1.53 (39.0) 1.36 (34.6)


0.14"
(3.5)
B C di D
0.02"
(0.5)

0.07"
(1.9)
0.13"
(3.4)
0.17"
(4.5)

Endress+Hauser 33
Proline Promag 50H, 53H

Front view 1-1/2" to 4" (DN 40 to 100), without process connection

K L
Z-Z
Z

90 0.5 E D C B A

60 0.5
F
Z
G
H

F06-5xHxxxxx-06-05-08-xx-001

Nomi- A B C D E F G H L K
nal size

in (mm) in (mm) in (mm) in (mm) in (mm) in (mm) [mm] in (mm) in (mm) Thread holes

1-1/2 4.80 3.39 2.79 2.01 1.39 M8 0.59 0.71 4


(40) (122.0) (86) (71.0) (51.0) (35.3) (15) (18)

2 5.79 3.90 3.29 2.50 1.89 M8 0.59 0.71 4


(50) (147.0) (99) (83.5) (63.5) (48.1) (15) (18)

2-1/2 5.79 4.53 3.94 2.99 2.36 M8 0.59 0.71 6


(65) (147.0) (115) (100.0) (76.1) (59.9) (15) (18)

3 7.76 5.55 4.76 3.50 2.86 M 12 0.59 0.79 4


(80) (197.0) (141) (121.0) (88.9) (72.6) (15) (20)

4 7.76 6.38 5.57 4.50 3.84 M 12 0.59 0.79 6


(100) (197.0) (162) (141.5) (114.3) (97.5) (15) (20)

34 Endress+Hauser
Proline Promag 50H, 53H

Process connections with gasket seal, 1-1/2" to 4"(DN 40 to 100)

Weld nipples for DIN Sensor Fits to di G D L L1 LK


1.4404 / 316L SS DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm] [mm] [mm]
5*H**-U***********

40 42 x 2 38.0 43 92 42 19 71.0

50 54 x 2 50.0 55 105 42 19 83.5


LK 65 70 x 2 66.0 72 121 42 21 100.0
G D
di
80 85 x 2 81.0 87 147 42 24 121.0

100 104 x 2 100.0 106 168 42 24 141.5

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


L1 If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
L connection (di) into account.
F06-xxHxxxxx-06-

Weld nipples for ODT/SMS Sensor Fits to di G D L L1 LK


1.4404 / 316L SS DN (mm) Piping OD/SMS (mm) (mm) (mm) (mm) (mm) (mm)
5*H**-V*********** (mm)

40 38.1 x 1.65 35.3 40 92 42 19 71.0

50 50.8 x 1.65 48.1 55 105 42 19 83.5


LK 65 63.5 x 1.65 59.9 66 121 42 21 100.0
G D
di 80 76.2 x 1.65 72.6 79 147 42 24 121.0

100 101.6 x 1.65 97.5 104 168 42 24 141.5

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


L1 If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
L connection (di) into account.
F06-xxHxxxxx-06-

Clamp ISO 2852 Sensor Fits to Clamp ISO di G D L LK


1.4404 / 316L SS size Piping ISO 2037 / 2852 in (mm) in (mm) in (mm) in (mm) inches
5*H**-W*********** BS 4825-1 Diameter

1-1/2" 1.50" 1.5" 1.40 (35.6) 1.99 (50.5) 3.62 (92) 2.70 (68.5) 2.79

2" 2.00" 2" 1.91 (48.6) 2.52 (64.0) 4.13 (105) 2.70 (68.5) 3.29
LK 2-1/2" 2.50" 2.5" 2.37 (60.3) 3.05 (77.5) 4.76 (121) 2.70 (68.5) 3.94
G D
di
3" 3.00" 3" 2.87 (72.9) 3.58 (91.0) 5.79 (147) 2.70 (68.5) 4.76

4" 4.00" 4" 3.84 (97.6) 4.69 (119.0) 6.61 (168) 2.70 (68.5) 5.57

Fitting length = (2 x L) + 5.35" (1-1/2" to 2-/12"), + 7.72" (3" to 4")


L If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
F06-xxHxxxxx-06-09-07- connection (di) into account.

Clamp DIN 32676 Sensor Fits to di G D L LK


1.4404 / 316L SS DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm] [mm]
5*H**-0***********

40 42 x 2 38 50.5 92 61.5 71.0

50 54 x 2 50 64.0 105 61.5 83.5

LK 65 70 x 2 66 91.0 121 68.0 100.0


G
di D
80 85 x 2 81 106.0 147 68.0 121.0

100 104 x 2 100 119.0 168 68.0 141.5

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


L If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
F06-xxHxxxxx-06- connection (di) into account.

Endress+Hauser 35
Proline Promag 50H, 53H

Tri-Clamp L14 AM7 Sensor Fits to di G D L LK


1.4404 / 316L SS
DN [mm] Nominal Piping OD in (mm) in (mm) in (mm) in (mm) in (mm)
5*H**-1***********

40 1-1/2" 1.5" 1.37 (34.8) 1.98 (50.4) 3.62 (92) 2.7 (68.6) 2.80 (71.0)

50 2" 2" 1.87 (47.5) 2.52 (63.9) 4.13 (105) 2.7 (68.6) 3.29 (83.5)
LK
65 2-1/2" 2.5" 2.37 (60.2) 3.05 (77.4) 4.76 (121) 2.7 (68.6) 3.94 (100.0)
G di D
80 3" 3" 2.87 (72.9) 3.58 (90.9) 5.79 (147) 2.7 (68.6) 4.76 (121.0)

100 4" 4" 3.83 (97.4) 4.68 (118.9) 6.61 (168) 2.7 (68.6) 5.57 (141.5)

Fitting length = (2 x L) + 5.35" (1-1/2" to 2-1/2"), + 7.72" (3" to 4")


L If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
F06-xxHxxxxx-06- connection (di) into account.

Coupling SC DIN 11851 Sensor Fits to di G D L LK


Threaded adapter DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm] [mm]
1.4404 / 316L SS
5*H**-2***********

40 42 x 2 38 Rd 65 x 1/6" 92 72 71.0

50 54 x 2 50 Rd 78 x 1/6" 105 74 83.5

65 70 x 2 66 Rd 95 x 1/6" 121 78 100.0


G
di LK D 80 85 x 2 81 Rd 110 x 1/4" 147 83 121.0

100 104 x 2 100 Rd 130 x 1/4" 168 92 141.5

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
L
F06-xxHxxxxx- connection (di) into account.

Coupling DIN 11864-1 Sensor Fits to di G D L LK


Aseptic threaded adapter, Form A DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm] [mm]
1.4404 / 316L SS
5*H**-3***********

40 42 x 2 38 Rd 65 x 1/6" 92 71 71.0

50 54 x 2 50 Rd 78 x 1/6" 105 71 83.5

LK 65 70 x 2 66 Rd 95 x 1/6" 121 76 100.0


G di D
80 85 x 2 81 Rd 110 x 1/4" 147 82 121.0

100 104 x 2 100 Rd 130 x 1/4" 168 90 141.5

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


L If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
F06-xxHxxxxx- connection (di) into account.

Flange DIN 11864-2 Sensor Fits to di G D L LK1 LK2


Aseptic flat flange, Form A DN [mm] Piping DIN 11850 [mm] [mm] [mm] [mm] [mm] [mm]
1.4404/316L SS
5*H**-4***********

40 42 x 2 38 82 92 64 71.0 65

50 54 x 2 50 94 105 64 83.5 77

LK 1 65 70 x 2 66 113 121 64 100.0 95


G LK 2 di
D 80 85 x 2 81 133 147 98 121.0 112

100 104 x 2 100 159 168 98 141.5 137

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
L
connection (di) into account.
F06 xxHxxxxx 06 09 07 xx 007

36 Endress+Hauser
Proline Promag 50H, 53H

Coupling SMS 1145 Sensor Fits to SMS 1145 di G D L LK


Threaded adapter DN [mm] Piping OD Diameter [mm] [mm] [mm] [mm] [mm]
1.4404 / 316L SS [mm]
5*H**-5***********

40 38.1 x 1.65 38.0 35.5 Rd 60 x 1/6" 92 63 71.0

50 50.8 x 1.65 51.0 48.5 Rd 70 x 1/6" 105 65 83.5

LK 65 63.5 x 1.65 63.5 60.5 Rd 85 x 1/6" 121 70 100.0


G di D
80 76.2 x 1.65 76.0 72.0 Rd 98 x 1/6" 147 75 121.0

100 101.6 x 1.65 101.6 97.6 Rd 132 x 1/6" 168 70 141.5

Fitting length = (2 x L) + 136 mm (DN 4065) / + 196 mm (DN 80100)


If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
L
connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-

Coupling ISO 2853 Sensor Fits to ISO 2853 di G D L LK


Threaded adapter Piping ISO 2037 Diameter in (mm) in (mm) in (mm) in (mm) in (mm)
1.4404 / 316L SS / BS 4825-1 in (mm)
5*H**-6***********

1-1/2" 1-1/2" 1.50 (38.0) 1.40 (35.6) 1.99 (50.6) 3.62 (92) 2.42 (61.5) 2.80 (71.0)

2" 2" 2.00 (51.0) 1.91 (48.6) 2.52 (64.1) 4.13 (105) 2.42 (61.5) 3.29 (83.5)

G LK 2-1/2" 2-1/2" 2.50 (63.5) 2.37 (60.3) 3.06 (77.6) 4.76 (121) 2.42 (61.5) 3.94 (100.0)
D
di 3" 3" 3.00 (76.1) 2.87 (72.9) 3.59 (91.1) 5.79 (147) 2.42 (61.5) 4.76 (121.0)

4" 4" 4.00 (101.6) 3.84 (97.6) 4.65 (118.1) 6.61 (168) 2.42 (61.5) 5.57 (141.5)

Fitting length = (2 x L) + 5.35" (1-1/2" to 2-/12"), + 7.72" (3" to 4")


If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 24) and process
L
connection (di) into account.
F06-xxHxxxxx-06-

Endress+Hauser 37
Proline Promag 50H, 53H

Weight
Weight data in lbs [kg]

Nominal Compact version Remote version (without cable)


diameter

[mm] [inch] Sensor Wall housing

2 1/12" 11.5 (5.2) 5.5 (2.5) 13 (6.0)

4 5/32" 11.5 (5.2) 5.5 (2.5) 13 (6.0)

8 5/16" 11.7 (5.3) 5.5 (2.5) 13 (6.0)

15 1/2" 11.9 (5.4) 5.7 (2.6) 13 (6.0)

25 1" 12.1 (5.5) 6.2 (2.8) 13 (6.0)

40 1 1/2" 14.3 (6.5) 9.9 (4.5) 13 (6.0)

50 2" 19.8 (9.0) 15.4 (7.0) 13 (6.0)

65 2 1/2" 20.9 (9.5) 16.5 (7.5) 13 (6.0)

80 3" 41.8 (19.0) 37.5 (17.0) 13 (6.0)

100 4" 40.8 (18.5) 36.4 (16.5) 13 (6.0)

Transmitter Promag (compact version): 3.4 kg


(Weight data valid for standard pressure ratings and without packaging material)

Materials Transmitter housing:


Compact housing: powder coated die-cast aluminium or stainless-steel field housing
1.4301/316L SS
Wall-mounted housing: powder coated die-cast aluminium

Sensor housing: 1.4301/304 SS


Wall mounting (holder panel): 1.4301/304 SS

Measuring tube: stainless steel 1.4301 or 1.4306/304LSS

Flange:
All connections 1.4404/316L SS
Flanges (EN (DIN), ANSI, JIS) made of PVDF
Adhesive fitting made of PVC

Ground disks (accessory): 1.4435/316L SS, Tantalum, Alloy C-22

Electrodes:
Standard: 1.4435/316L SS
Option: Alloy C-22, tantalum, platinum/rhodium 80/20 (up to 1" / DN 25 only)

Seals:
1/12" to 1" (DN 2 to 25): O-ring (EPDM, Viton, Kalrez) or gasket seal (EPDM, silicone, Viton)
1-1/2" to 4" (DN 40 to 100): gasket seal (EPDM, silicone)

38 Endress+Hauser
Proline Promag 50H, 53H

Material load diagrams Caution!


The following diagrams contain material load curves (reference curves) for various process connections relating
to the fluid temperature. But the maximal permissible fluid temperature always depends on the lining material
of the sensor and/or the sealing material (see Page 18).

Weld nipple to DIN EN ISO 1127, ODT / SMS, coupling to ISO 228 / DIN 2999
Weld nipple material: 1.4404 / 316L SS (with O-ring)

psi (bar)
725 (50)

580 (40)

507 (35)

435 (30)

363 (25)

0
-76 -40 -4 32 68 104 140 176 212 248 284 320 356 [F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [C]

F06-xxHxxxxx-05-xx-xx-xx-000

Weld nipple to DIN 11850, ODT / SMS, Clamp (ODT, ISO 2852, DIN 32676, L14 AM7),
coupling (DIN 11851, DIN 11864-1, ISO 2853, SMS 1145), flange DIN 11864-2
Weld nipple material: 1.4404 / 316L SS (with gasket seal)

psi (bar)
290 (20)

217 (15)

145 (10)

73 (5)

0
-76 -40 -4 32 68 104 140 176 212 248 284 320 356 [F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [C]

F06-xxHxxxxx-05-xx-xx-xx-001

Flange connection to EN 1092-1 (DIN 2501), adhesive fitting


Flange material: 1.4404 / 316L SS, PVDF, PVC-U

psi (bar)
580 (40)

507 (35) 316L SS

435 (30)

363 (25)

290 (20)

217 (15)

145 (10) PVDF


PVC-U
73 (5)

0
-76 -40 -4 32 68 104 140 176 212 248 284 320 356 [F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [C]

F06-xxHxxxxx-05-xx-xx-xx-002

Endress+Hauser 39
Proline Promag 50H, 53H

Flange connection to ANSI B16.5


Flange material: 1.4404 / 316L SS, PVDF

psi (bar)
435 (30)

363 (25)

290 (20)

217 (15) Cl 150 (316L)

145 (10)

73 (5) Cl 150 (PVDF)

0
-76 -40 -4 32 68 104 140 176 212 248 284 320 356 [F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [C]

F06-xxHxxxxx-05-xx-xx-xx-003

Flange connection to JIS B2238


Flange material: 1.4404 / 316L SS, PVDF

F06-xxHxxxxx-05-xx-xx-xx-004

Fitted electrodes Measuring electrodes and EPD electrodes


Standard with: 1.4435/316L SS, Alloy C-22, tantalum, platinum/rhodium 80/20
1/12" and 5/32" (DN 2 to 4): without EPD electrode

Process connection With O-ring: welding nipples (DIN EN ISO 1127, ODT / SMS), flanges (EN (DIN), ANSI, JIS), PVDF flanges
(EN (DIN), ANSI, JIS), external pipe thread, internal pipe thread, hose connection, PVC adhesive fittings
With gasket seal: weld nipples (DIN 11850, ODT / SMS), clamps (ISO 2852, DIN 32676,
L14 AM7), threaded fasteners (DIN 11851, DIN 11864-1, ISO 2853, SMS 1145), flanges DIN 11864-2

Surface roughness PFA liner: 16 in (0.4 m)


Electrodes:
1.4435/316L SS, Alloy C-22: 12 to 20 in (0.3 to 0.5 m)
Tantalum, platinum/rhodium: 12 to 20 in (0.3 to 0.5 m)
Process connection: 32 in (0.8 m)

(all data relate to parts in contact with the medium)

Human interface
Display elements Liquid-crystal display: backlit, two lines (Promag 50) or four lines (Promag 53) with 16 characters per line
Custom configurations for presenting different measured-value and status variables
Totalizer:
Promag 50: 2 totalizers (7-digit plus 7-digit overflow with sign and units)
Promag 53: 3 totalizers (7-digit plus 7-digit overflow with sign and units)

40 Endress+Hauser
Proline Promag 50H, 53H

Operating elements Unified control concept for both types of transmitter:


Promag 50:
Local operation with three push buttons (, +, E)
Quick Setup menus for straightforward commissioning

Promag 53:
Local operation with Touch Control (, +, E)
Application-specific Quick Setup menus for straightforward commissioning

Language group Language groups available for operation in different countries:

Promag 50, Promag 53:


Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
Eastern Europe and Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
South and east Asia (SEA):
English, Japanese, Indonesian
Promag 53:
China (CIN):
English, Chinese
You can change the language group via the operating program ToF Tool - Fieldtool Package.

Remote operation Promag 50: Remote control via HART, PROFIBUS DP/PA
Promag 53: Remote control via HART, PROFIBUS DP/PA, MODBUS RS485, FOUNDATION Fieldbus

Certificates and approvals


Ex approvals Information about currently available Ex versions (ATEX, FM, CSA) can be supplied by your Endress+Hauser
Sales Center on request. All explosion protection data are given in separate documentation which is available
upon request.

Sanitary compatibility Meets 3-A standards and EHEDG-tested


Seals in conformity with FDA (except Kalrez seals)

CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser
confirms successful testing of the device by affixing to it the CE mark.

C-Tick mark The measuring system is in conformity with the EMC requirements of the Australian Communications
Authority (ACA).

Pressure Equipment Directive Flow meters with a nominal diameter smaller or equal DN 25 are covered by Art. 3(3) of the
European directive 97/23/EG (Pressure Equipment Directive) and are designed according to sound engineer
practice. For larger nominal diameter, optional approvals according to Cat. III are available when required
(depends on fluid and process pressure).

PROFIBUS DP/PA The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS User Organization). The device thus meets all the requirements of the following specifica-
tions:

Certified to PROFIBUS PA, profile version 3.0 (device certification number: on request)
The device can also be operated with certified devices of other manufacturers (interoperability)

FOUNDATION Fieldbus The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus Foundation. The device thus meets all the requirements of the following specifications:

Certified to FOUNDATION Fieldbus Specification


The device meets all the specifications of the FOUNDATION Fieldbus H1.
Interoperability Test Kit (ITK), revision status 4.0 (device certification number: on request)
The device can also be operated with certified devices of other manufacturers
Physical Layer Conformance Test of the Fieldbus Foundation

Endress+Hauser 41
Proline Promag 50H, 53H

MODBUS certification The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the MOD-
BUS/TCP Conformance Test Policy, Version 2.0. The measuring device has successfully passed all the test
procedures carried out and is certified by the MODBUS/TCP Conformance Test Laboratory of the University
of Michigan.

Other standards, guidelines EN 60529:


Degrees of protection by housing (IP code)

EN 61010:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and
Laboratory Procedures.

EN 61326/A1 (IEC 6326):


Electromagnetic compatibility (EMC requirements)

NAMUR NE 21:
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.

NAMUR NE 43:
Standardization of the signal level for the breakdown information of digital transmitters with analog
output signal.

NAMUR NE 53:
Software of field devices and signal-processing devices with digital electronics.

Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor. The Endress+Hauser service organization can provide detailed information on request.

Supplementary documentation
Flow Measurement (FA005D/06/en)
Operating Instructions Promag 50 (BA046D/06/en, BA049D/06/en)
Operating Instructions Promag 50 PROFIBUS PA (BA055D/06/en, BA056D/06/en)
Operating Instructions Promag 53 (BA047D/06/en, BA048D/06/en)
Operating Instructions Promag 53 PROFIBUS DP/PA (BA053D/06/en, BA054D/06/en)
Operating Instructions Promag 53 FOUNDATION Fieldbus (BA051D/06/en, BA052D/06/en)
Operating Instructions Promag 53 MODBUS (BA117D/06/en und BA118D/06/en)
Supplementary documentation on Ex-ratings: ATEX, FM, CSA, etc.

Registered trademarks
HART
Registered trademark of HART Communication Foundation, Austin, USA

PROFIBUS
Registered trademark of the PROFIBUS User Organisation, Karlsruhe, Germany

FOUNDATION Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA

MODBUS
Registered trademark of the MODBUS Organisation

HistoROM, S-DAT, T-DAT, F-CHIP, ToF Tool - Fieldtool Package, Fieldcheck, Applicator
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

42 Endress+Hauser
Proline Promag 50H, 53H

Ordering Information

NOTE: Endress+Hauser reserves the right to change or modify product, specifications, and ordering information at any time without notice.
Please consult Endress+Hauser or your local representative for the most recent information.

Promag 50H Electromagnetic flowmeter


0 1 2 3 4 5 6 7 8 9 10 11 12
Promag 50 H -
Nominal Diameter - Liner, PFA
7 Housing
0 02 1/12
A NEMA 4X (IP 67) compact aluminum housing
04 5/32
B NEMA 4X (IP 67) compact SS housing
08 5/16
C NEMA 4X (IP 67) remote wall-mounted (only for approvals A or R)
15 1/2
G NEMA 4X (IP 67) remote aluminum field housing for non-hazardous areas
22 1
P NEMA 4X (IP 67) compact aluminum housing, HE (harsh environment) *
40 1-1/2
R NEMA 4X (IP 67) compact SS housing, HE (harsh environment) *
50 2
S NEMA 4X (IP 67) remote wall-mounted housing, HE (harsh environment) *
65 2-1/2
1 -40F (-40C) ambient temperature, NEMA 4X (IP 67), compact,
80 3
aluminum field housing (only for approvals A and R)
1H 4
2 -40F (-40C) ambient temperature, NEMA 4X (IP 67), compact SS
1 Process connection housing for transmitter (only for approvals A and R)
For o-ring seal 3 -40F (-40C) ambient temperature, NEMA 4X (IP 67), compact,
A Without process connection (only for approvals A and R)
C Weld nipple for IPS pipe, 316L SS
8 Cable for remote
E Class 150 ANSI flange, 316L SS (1/12 to 1 only)
0 Without cable
H Class 150 ANSI flange, PVDF (1/12 to 1 only)
1 15 foot coil and signal cable
K External pipe thread, 316L SS (1/12 to 1 only)
2 30 foot coil and signal cable
L Internal pipe thread, 316L SS (1/12 to 1 only)
5 Coil and signal cable, specify length (maximum 650 ft depending on
M Hose connection for 13 mm ID hose, 316L SS (1/12 to 5/16 only)
conductivity of process material)
N Hose connection for 16 mm ID hose, 316L SS (1/2 only)
P Hose connection for 19 mm ID hose, 316L SS (1/2 only) 9 Cable entries
R Adhesive coupling for 1/2 pipe, PVC (1/12 to 5/16 only) B 1/2 NPT
For gasket seal L 1/2 NPT fieldbus connector (only for approval A and R)
V Weld nipple for ODT, 316L SS X Sensor only, without cable, compact version only
W Clamp, ISO 2852, 316L SS (1 to 4 only) 10 Power supply / display
1 Tri-clamp for ODT, 316L SS 7 85 to 260 VAC, without display, remote configuration only (not for
6 Coupling, ISO 2853 threaded adapter, 316L SS (1-1/2 to 4 only) wall mount or SS housing)
2 Seal material 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
A EPDM o-ring (up to 1 only) only (not for wall mount or SS housing)
B Viton o-ring (up to 1 only) A 85 to 260 VAC, with display, push button operation (language: EN,
C Kalrez o-ring (up to 1 only) ES, FR, IT, NL, PT, DE)
F EPDM gasket seal, all sizes B 20 to 55 VAC / 16 to 62 VDC, with display, push button operation
H Silicon gasket seal, standard, all sizes (language: EN, ES, FR, IT, NL, PT, DE)
G Viton gasket seal (up to 1 only) X Sensor only
O Without seal (without process connections only) 11 Software
3 Electrode material A Standard software
A 316L SS with grounding rings, 316L SS (1/12 and 5/32 no EPD C ECC electrode cleaning circuit (only for approvals A and R)
electrodes), only up to 1 sensor J Batching
B Alloy C-22 with grounding rings (1/12 and 5/32 no EPD electrodes), L Batching and ECC electrode cleaning circuit (only for approvals A and R)
only up to 1 sensor 12 Outputs / Inputs
D Tantalum with grounding rings (1/12 and 5/32 no EPD electrodes), A Current HART, frequency
only up to 1 sensor D Current HART, frequency, status output, status input
0 Measuring, reference and EPD electrodes, 316L SS (1/12 and 5/32, H PROFIBUS-PA (approvals A and R only)
no EPD electrodes) J PROFIBUS-DP (approvals A and R only)
1 Measuring, reference and EPD electrodes, Alloy C22 (1/12 and 5/32, S Current HART, frequency; active (not for approvals A and R)
no EPD electrodes) T Current HART, frequency; passive (not for approvals A and R)
2 Measuring, reference and EPD electrodes, tantalum (1/12 and 5/32, W Current HART
no EPD electrodes) X Sensor only
5 Measuring, reference and EPD electrodes, Platinum/Rhodium 80/20%
up to 1 only (1/12 and 5/32, no EPD electrodes) * Harsh environment (HE) option is available for process conditions where
4 Calibration cool process temperatures in tropical (high humidity) environments or process
B 3-point calibration, 0.2% fluids which undergo large cyclical temperature variations which can cause
E SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate high amounts of moisture that could condense onto the measurement tube.
traceable according to ISO 9000 (specify range)
5 Certificates
1 Standard, no certificate
2 3.1B material certificate for pipe and flanges
3 2.3 pressure test certificate (1.5 x PN, 3 minutes)
4 3.1B material and 2.3 pressure test certificate
6 Approvals
A For use in non-hazardous areas
N FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for aluminum
field housing, compact version)
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2

Endress+Hauser 43
Promag 53H Electromagnetic flowmeter
0 1 2 3 4 5 6 7 8 9 10 11 12
Promag 53 H -
Nominal Diameter - Liner, PFA
7 Housing
0 02 1/12
A NEMA 4X (IP 67) compact aluminum housing
04 5/32
B NEMA 4X (IP 67) compact SS housing
08 5/16
C NEMA 4X (IP 67) remote wall-mounted (only for approvals A or R)
15 1/2
G NEMA 4X (IP 67) remote aluminum field housing for non-hazardous areas
22 1
P NEMA 4X (IP 67) compact aluminum housing, HE (harsh environment) *
40 1-1/2
R NEMA 4X (IP 67) compact SS housing, HE (harsh environment) *
50 2
S NEMA 4X (IP 67) remote wall-mounted housing, HE (harsh environment), *
65 2-1/2
1 -40F (-40C) ambient temperature, NEMA 4X (IP 67), compact,
80 3
aluminum field housing (only for approvals A and R)
1H 4
2 -40F (-40C) ambient temperature, NEMA 4X (IP 67), compact SS
1 Process connection housing for transmitter (only for approvals A and R)
For o-ring seal 3 -40F (-40C) ambient temperature, NEMA 4X (IP 67), compact,
A Without process connection (only for approvals A and R)
C Weld nipple for IPS pipe, 316L SS
8 Cable for remote
E Class 150 ANSI flange, 316L SS (1/12 to 1 only)
0 Without cable
H Class 150 ANSI flange, PVDF (1/12 to 1 only)
1 15 foot coil and signal cable
K External pipe thread, 316L SS (1/12 to 1 only)
2 30 foot coil and signal cable
L Internal pipe thread, 316L SS (1/12 to 1 only)
5 Coil and signal cable, specify length (maximum 650 ft depending on
M Hose connection for 13 mm ID hose, 316L SS (1/12 to 5/16 only)
conductivity of process material)
N Hose connection for 16 mm ID hose, 316L SS (1/2 only)
P Hose connection for 19 mm ID hose, 316L SS (1/2 only) 9 Cable entries
R Adhesive coupling for 1/2 pipe, PVC (1/12 to 5/16 only) B 1/2 NPT
For gasket seal L 1/2 NPT fieldbus connector (only for approval A and R)
V Weld nipple for ODT, 316L SS X Sensor only, without cable, compact version only
W Clamp, ISO 2852, 316L SS (1 to 4 only) 10 Power supply / display
1 Tri-clamp for ODT, 316L SS 7 85 to 260 VAC, without display, remote configuration only (not for
6 Coupling, ISO 2853 threaded adapter, 316L SS (1-1/2 to 4 only) wall mount or SS housing)
2 Seal material 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
A EPDM o-ring (up to 1 only) only (not for wall mount or SS housing)
B Viton o-ring (up to 1 only) A 85 to 260 VAC, with display, push button operation (language: EN,
C Kalrez o-ring (up to 1 only) ES, FR, IT, NL, PT, DE)
F EPDM gasket seal, all sizes B 20 to 55 VAC / 16 to 62 VDC, with display, push button operation
H Silicon gasket seal, standard, all sizes (language: EN, ES, FR, IT, NL, PT, DE)
G Viton gasket seal (up to 1 only) X Sensor only
O Without seal (without process connections only) 11 Software
3 Electrode material A Standard software
A 316L SS with grounding rings, 316L SS (1/12 and 5/32 no EPD C ECC electrode cleaning circuit (only for approvals A and R)
electrodes), only up to 1 sensor 12 Outputs / Inputs
B Alloy C-22 with grounding rings (1/12 and 5/32 no EPD electrodes), Fixed communication boards
only up to 1 sensor A Current HART, frequency
D Tantalum with grounding rings (1/12 and 5/32 no EPD electrodes), B Current HART, frequency, 2 relays
only up to 1 sensor F PROFIBUS-PA, IS
0 Measuring, reference and EPD electrodes, 316L SS (1/12 and 5/32, G Foundation Fieldbus, IS
no EPD electrodes) H PROFIBUS-PA
1 Measuring, reference and EPD electrodes, Alloy C22 (1/12 and 5/32, J PROFIBUS-DP
no EPD electrodes) K Foundation Fieldbus
2 Measuring, reference and EPD electrodes, tantalum (1/12 and 5/32, Q Modbus RS485, status input
no EPD electrodes) S Current HART, frequency; active I.S
5 Measuring, reference and EPD electrodes, Platinum/Rhodium 80/20% T Current HART, frequency; passive I.S
up to 1 only (1/12 and 5/32, no EPD electrodes) Flexible communication boards
4 Calibration C Current HART, frequency, 2 relays, flexible module
B 3-point calibration, 0.2% D Current HART, frequency, relay, and status input, flexible module
E SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate L Current HART, 2 relays and status input/output
traceable according to ISO 9000 (specify range) M Current HART, 2 frequency output, status input
5 Certificates N Modbus RS 485, current and frequency output, status input Subject to modification
1 Standard, no certificate P PROFIBUS-DP, current and frequency output, status input
2 3.1B material certificate for pipe and flanges V PROFIBUS-DP, two relay outputs, status input
3 2.3 pressure test certificate (1.5 x PN, 3 minutes) 2 Current HART, relay, current, frequency ouputs
4 3.1B material and 2.3 pressure test certificate 4 Current HART, relay, frequency outputs, current input
International
6 Approvals Head Quarter
5 Current HART, frequency output, status, relay inputs
7 Modbus RS485, two relay outputs, status input
A For use in non-hazardous areas
X Sensor only
N FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for aluminum
Endress+Hauser housing, compact version)
field
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2 * Harsh environment (HE) option is available for process conditions where
GmbH+Co. KG cool process temperatures in tropical (high humidity) environments or process
Instruments International fluids which undergo large cyclical temperature variations which can cause
Colmarer Str. 6 high amounts of moisture that could condense onto the measurement tube.
79576 Weil am Rhein
United States
Deutschland Canada Mexico

Tel. +49 76 21Inc.


Endress+Hauser, 9 75 02 Endress+Hauser Canada Endress+Hauser, Mxico, S.A. de C.V.
Fax +49
2350 76 Place
Endress 21 9 75 34 5 1075 Sutton Drive Fernando Montes de Oca 21 Edificio A Piso 3
Greenwood, IN 46143
www.endress.com Burlington, ON L7L 5Z8 Fracc. Industrial San Nicols
Tel. 317-535-7138 Tel. 905-681-9292 54030. Tlalnepantla de Baz
info@ii.endress.com
Sales 888-ENDRESS 800-668-3199 Estado de Mxico
Service 800-642-8737 Fax 905-681-9444 Mxico
fax 317-535-8498 info@ca.endress.com Tel: +52 55 5321 2080
inquiry@us.endress.com www.ca.endress.com Fax +52 55 5321 2099
www.us.endress.com eh.mexico@mx.endress.com
www.mx.endress.com

TI048D/24/ae/09.07
2007 Endress+Hauser, Inc.