Units R/h R/h Dose Rate Measurement 5R/h to 10R/h 1R/h to 10R/h Range Personal dose Measures area dose only? 0 to 1000R personal dose, Hp(10) The PED display features the dose man measurement graphical display for ease of understanding. Energy Range 45keV to 3MeV 33keV to 1.3MeV tested The PED type test was limited to 1.3MeV 33keV to 3MeV projected based on GM as this covers the vast majority of typical manufacturer data nuclides Alarms Visual (single LED) Visual (3 LEDs) The extra LEDs on the Tracerco PED Vibrating Vibrating means the user can easily distinguish Audible Audible dose, dose rate and error alarms Stay time prediction Yes, but user need to press Yes. Permanently displayed on readings button to achieve a temporary screen view Task dose No? Yes This is a useful function for recording dose and peak dose rate over a specific, user defined period. Gloved hand operation Poor Good Ease of use LCD screen is small and menu Modern mobile phone type screen allows system is difficult to navigate very easy navigation (no user manual through required) Intrinsic safety Not on the standard G Yes, approved to the highest level to US, EU Fire crews can use the PED without and global standards stopping to check for explosive FM approved for Class 1 Div 1 groups ABCD atmospheres and Class 1 Zone 0 IIC T4 EMC EN 50081-1 Disturbance EN 61526:2007 Radiation Protection emission Instrumentation EN 61000-6-2 Immunity EN 61000-4-2:1995 Electrostatic Discharge Immunity Test EN 61000-4-3:2006 Radiated Electromagnetic Field Immunity Test EN 61000-4-4:2004 Electrical Fast Transient/Burst EN 61000-4-6:1996 Conducted RF Immunity EN 61000-4-8:1993 Magnetic Field Immunity EN 61000-4-11:1994 Voltage Dips and Interruptions EN 55022:2006 Radiated Emissions EN 55022:2006 Conducted Emissions
Note that additional immunity tests
performed at 40 V/m Case material Hard plastic with separate soft Ruggedized housing with complete soft rubber overmold polymer overmold with antistatic properties Ingress protection IP65 IP67 The higher IP rating of the PED means Dust tight and water jet tested Dust tight and submersible in 1m of water that it is totally weatherproof and can be submerged in a sink for decontamination. Battery Life Two AAA batteries provide Rechargeable Li ion batteries provide 300 Note that the PED can be stored in a dock 600 hours of operation hours of operation to ensure full capacity when deployed. NiMH provides 300 hours The dock can be powered from mains, USB or 12v cabin supply. Dimensions 1.25 x 2.4 x 3.78 in. (3.1 x 6.1 x 3.9 x 2.4 x 0.8 in. 9.6cm) Weight 5.6 oz. (160g) 5.6 oz. (160g) Dose rate and dose data Additional cost Dock and DoseVision software included as reporting part of the standard supply Internal memory 1600 data point storage. 125,000 data point storage. The advanced data storage and also the Data stored every 120 secs Data can be logged every 0.01mR change in ease of viewing on the DoseVision dose or every minute. software means that meaningful post incident investigations can be performed Customised screen layout Presume not possible We can tailor the PED screen layout to suit Some of the fire services we have spoken the client. to in the UK are interested in just displaying the dose man and the stay time. Some of our NDT customers have requested just a large font dose rate display. We could even insert the NY fire service logo as a splash screen during start up.