Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
SOP No.:
QUALITY CONTROL DEPARTMENT Page 1 of 3
QCG0029
Supersedes : Effective:
STANDARD OPERATING PROCEDURE
00 April 2010
1.0 OBJECTIVE:
2.0 SCOPE:
3.0 RESPONSIBILITY:
4.0 ACCOUNTABILITY:
HOD QC
5.0 PROCEDURE :-
5.1 Ensure that the instrument is clean and free from dust.
5.2 Ensure that the magnifying lens is fixed on its appropriate plate.
5.3 Ensure that the marker is connected to either right or left side of
the instrument.
5.9 When the counting is over, the screen can be set to 000 by
pressing the manual button.
5.11 Switch OFF the instrument and put under cover when not
required.
6.0 CALIBRATION:-
Operate the instrument as per the operating procedure and press the tip
of the marker 10 times and then check the number displayed on the
screen. Reset to 000 and again press the manual button 25 times and
check the number displayed on the screen. Reset to 000 and again
press the manual button 50 times and check the number displayed on the
screen. In each the cases the number displayed on the screen is also
10, 25 and 50.
10 10
25 25
50 50
6.0 RECORDS:
Calibration Record of Digital Colony Counter. (Annexure-1).
7.0 ANNEXURES:
Annexure-1 : Format for Calibration Record of Digital Colony Counter.
8.0 ABBREVIATIONS & DEFINITIONS:
SOP : Standard Operating Procedure.
Dept. : Department.
Annexure-1
CALIBRATION OF COLONY COUNTER
10 10
25 25
50 50