Portable Suction Unit

Minic S -
OPERATING INSTRUCTIONS

MS2-1400

MS2-1500

Thank you very much for choosing a portable suction


unit, Minic S-II.
Read this manual well before using this unit so that you
can use it correctly. After reading it, keep it near you to
read it any time whenever necessary.
Main Features
Lightweight and compact, it is easier to carry as a portable suction unit.
A wide range of suction from a low-pressured suction to a high-pres-
sured suction is provided.
Equipped with the piston pump enabling a high suction vacuum and a
high low rate.
Equipped with the suction stop float valve, suctioning will stop auto-
matically when secretion suctioned becomes full in the suction bottle.
Equipped with the filter inside the cushion for the bottle cover, which is
hard for secretion such as mist like moisture to infiltrate into the pump.
Different sizes of the inlet adapter are used in the patient connection
side and in the suction unit side, eliminating the accidental misconnec-
tion to the bottle cover.
Equipped with the catheter holder made from resign for putting a cath-
eter or a patient tube in it, and with a hook groove to fix the patient-
tube. (It is used when stopping suctioning temporarily.)
As an optional accessory, a filter is prepared whose material is PTFE
(polytetrafluoroethylene), which is hard for secretion such as mist like
moisture to pass through to prevent bacteria pollution. (Refer to 37nd
page.)
Equipped with the exhaust filter, which does not release the stained air
as is.
Equipped with the Patient Tube Connection Port complying with the
ISO standard, the inner diameter of which is 6 mm, enabling the
smooth aspiration with less clogging.
The power cord is wound up inside the unit when the storage facilitat -
ing transportation.
Six hours of continuous operation is possible.

MS2-1400
The suction bottle is made from polycarbonate resin strong against
shock and its mouth is large about 11 cm. In addition, since the detach-
able button and the handle can make you pour out suction things easily
and they are easy to handle.

Table of Contents
For Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions on Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Proper Handling
1. Preparations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2. Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3. Handling after Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4. Sterilization and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . 23
Reference Information (About Sterilization
and Disinfection of Suction Unit). . . . . . . . . . . . . . . . . . . . . . 26
5. Cleaning of Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6. Replacement of Exhaust Filter. . . . . . . . . . . . . . . . . . . . . . . . 29
Maintenance and Inspection
1. Daily Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2. Periodical Inspection (User Inspection) Check List. . . . . 31
3. About Scheduled Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . 33
Troubleshooting before calling a serviceman . . . . . . . . . . . 34
About Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
About the Marks Shown on This Manual . . . . . . . . . . . . . . . . 45
About Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

This Operating Instructions are described mainly on the


models MS2-1400 .
For the models MS2-1500, only the exterior view of the suc-
tion bottle is different but the basic operations are exactly the
same.


FOR SAFETY
This Operating Instructions contains the important safety
Warnings, Cautions and Notes for safe use of this medi-
cal equipment. Read these Operating Instructions carefully
before use and become thoroughly familiar with the opera-
tions and maintenance of this equipment.

Observe the following safety Warnings, Cautions and Notes described


in this manual.
In order to prevent personal injury or property damage to the users, this
manual symbolizes items that are particularly important for safety with
the following symbols.
Always follow the guidance given in these items.

WARNING
Failure to comply with the following items may result in seriously
dangerous condition to the patient(s) and damage(s) to the equipment
even if the equipment can be operated normally.

CONTRAINDICATION AND PROHIBITION

The items regarding the improper handling of the equipment, the


prohibited use of this equipment for other than the intended use, the
patient(s) and disease(s) for which the equipment cannot be used, and the
medical equipment and the connection equipment with which concomitant
use with this equipment is prohibited are described. Failure to comply
with these items may result in seriously dangerous condition to the
patient(s) and damage(s) to the equipment.

CAUTION
In addition to the Warnings, Contraindications and Prohibitions, the items
to be followed when using this equipment, when storing this equipment
and when performing maintenance of this equipment are described.
Failure to comply with these items may result in the troubles that normal
performance of this equipment cannot be obtained, medical practice is
threatened, or damage occurs to the equipment.

IN ORDER TO USE THE UNIT SAFELY

The discontinuation of the medical treatment by failure or generating


of fault of the medical equipment may be concerned with a patients
life directly or indirectly.
Be sure to perform the maintenance check (user inspection) every 3
months. And also the scheduled inspection every 2 years by the manu-
facturer or the repair contractor and overhaul every 5 years are re-
quired. (Scheduled inspection is charged).

REQUEST

Before you will send this equipment for inspection or repair back to the
factory, empty the bottle and disinfect or sterilize it.
We cannot receive it when fear of infection is in it.


CAUTIONS ON USE
WARNING
Don't use this machine other than the skilled man.
Operate the unit under direction of medical practi-
cians or well-trained men
Maintenances and inspections should be performed
following this manual and the package insert.
Use the patient tube so that it should not be bent or
folded in any manner.
Strictly follow the specified type, dose and usage of
the chemicals.
Pour out secretion as early as possible before the
bottle becomes full.
Don't touch the power cord or the power supply
plug with wet hands.
The patient tube, suction bottle, accessories of suc-
tion bottle, and catheter holder should be sterilized
after washing.
(When this equipment is used at home, sterilize and
disinfect them as necessary.)
The consumables should be replaced periodically.
When replacing parts, use always the genuine parts
of our company.
Dont block the ventilation air outlet on the rear of
the housing.
When unplugging the power cord, hold it by the
plug. Pulling on the wire can cause the damage,
electrical shock and fire.
If secretion is suctioned and has entered in the suc-
tion pump side, it is very dangerous that the piping
is chocked inside the tube by the secretion.
The check and inspection by the repair contractor or
the manufacturer is definitely required.

WARNING
The PTFE suction filter (an optional accessory)
degenerates gradually as the mist of the secre-
tion adheres to the filter, causing clogging. The
effect of suction filter will be down in a month or
so. Replace the PTFE suction filter with a new
one when the suction power becomes weak and
it seems that the filter clogs.
If secretion is suctioned into the PTFE suction
filter (an optional accessory), the PTFE suction
filter will be clogged even at one time, and it
may not be able to suction any more.

CONTRAINDICATION AND PROHIBITION

Use this equipment only for its intended use as


described in this manual. Never use this equip-
ment for blood collecting and others.
This PTFE suction filter (an optional accessory)
cannot be reused even it is sterilized and disin-
fected. Replace it with a new one.
Do not run this equipment continuously longer
than the specified continuous run time. (The
OI
maximum continuous run time is six hours.)
L

Do not place this equipment in a direct sunshine,


high temperature or high humidity.
Don't lubricate any parts.
Don't touch the internal mechanisms. Don't alter
it in any manner.
Use only the genuine parts of our company.


CAUTIONS ON USE

CAUTION

Read the Operating Instructions and the package


insert carefully and master it before using this de-
vice so that you can use it correctly.
The power supply voltage, frequency and the
maximum allowable current of the power outlet.
Be sure to use this equipment within the ratings.
Be sure to set the suction pressure in accordance
with the doctors prescription, and make before
use that the suction pressure is set in accordance
with the doctors prescription.
Constantly monitor the state of the equipment and
patient that there is no abnormalities in the equip-
ment and patient.
If any abnormality should be found, turn off the
power switch and take proper measures including
request for repair.
Use this equipment in the environment of ambi-
ent temperature from 10 to 40C and relative hu-
midity from 30 to 75%.
Do not splash water on this equipment.
Place this equipment on the level and stable
place. Use this equipment without giving a vibra-
tion nor a shock.
Do not damage a power cord. Do not place any-
thing on the cord.
When winding or unplugging the power cord,
hold it by the plug.
Clean occasionally not to accumulate dust on
the power plug and wall outlet. Not doing so can
cause that the dust ignites by excess heat of the
outlet.

CAUTION

Clean and dry the patient tube, the patient tube connection port, the
suction bottle and the accessories of the suction bottle thoroughly.
If cleaning and/or drying are insufficient, it may result in mold
(black or pink mold). After use, clean and dry them thoroughly.
Insert the suction bottle holder until it comes to the top of the suc-
tion bottle. Then, place the bottle cover. If the suction bottle holder
is stopped in the middle of the suction bottle, the bottle cover may
not be closed.
Carry out periodical cleaning and check. In order to use this unit
safely long, and prevent sudden failure and the like, be sure to ac-
cept the maintenance inspection (user inspection) and scheduled
inspection. (Refer to 30th page.)
When the equipment will not be used for a long period of time, dis-
connect the power plug from the wall outlet.
When this equipment was not used for a long period of time, use it
after taking it out to check for safety, and use it.
Be careful that the metal parts and the peripherals can be so hot
during the continuous run (maximum six hours).
Do not set the Power switch to On when the suction pressure re-
mains (i.e., in the state that the vacuum gauge needle does not re-
turn to 0) in this equipment. Doing so may result in the trouble
that this equipment does not work, or fuse may blow.
When you will dispose of this equipment (inclusive of the acces-
sories and optional parts), please be directed by the administrator
of your residential area. It can cause environmental pollution.


USER INTERFACE
MS2-1400

Suction Unit Connecting Tubing


HIGH-LOW Pressure Control Dial
Suction Unit Tube Inlet (Adjusting the Suction Pressure)
Adapter (yellow)
Vacuum Gauge
Cushion
Power Switch
Patient Tube
Inlet Adapter
(with nozzle)

Suction Bottle
Cover
Power Cord
Suction Bottle
Holder Patient Tube
Cover
Detachable
L - Joint
Button
Patient Tube Receptacles
Suction Bottle (right and left)
(Plastic)
Catheter Holder (1 pc.)
Suction Bottle Holder

REAR of HOUSING

Ventilation Air Outlet

Exhaust Filter Cover

Exhaust Filter

10
MS2-1500

Suction Unit Connecting Tubing HIGH-LOW Pressure Control Dial


Suction Unit Tube Inlet (Adjusting the Suction Pressure)
Adapter (yellow)
Vacuum Gauge
Cushion
Power Switch
Suction Bottle
Head

Patient Tube
Connection Port
(Inner Diameter
6 mm)
Power Cord
Suction Bottle
Cap
Patient Tube
Cover
Suction Bottle
L - Joint
Holder
Patient Tube Receptacles
(right and left)
Suction Bottle (Glass)
Catheter Holder (1 pc.)

BOTTOM of HOUSING

Fuse Holder

Fuse

11
USER INTERFACE

STANDARAD ACCESSORIES
Check before use whether any accessories are missing.

Patient Tube (with Adapter)

Adapter

Catheter Connector Suction Bottle Filter, 2 pcs (*2)


S-6 Packing (*1) (Spare)
(Spare, for Glass
Suction Bottle)

Spare Fuse Dust Cover Operating Instructions

*1: It will only come with MS2-1500.


*2: This filter can be used either as a filter for cushion and as an exhaust
filter.

12
PROPER HANDLING
1. Preparations
CAUTION When using this equipment for the first time in a
day, perform the daily inspection without fail.
(Refer to 30th page.)

1 Place this equipment on the level and stable place.
2 Check that the bottle cover (suction bottle cap) and the connecting
tubes are set so securely that there may be no leak.

3 Pull the power cord from Power Switch [On]


the equipment and plug
it into the wall outlet and set
the power switch to On.
Note: When the power cord
is pulled until the red
tape on the power
cord became visible,
do not pull it any
more.
4 Set the patient tube to the patient tube connection port or to the patient
tube inlet adapter (with nozzle) securely. Fold the patient tube with fin-
gers, and turn the HIGH-LOW pressure control dial to regulate the suc-
tion pressure to the specified one (the pressure prescribed by doctor).
After this, set the power switch to Off once.

HIGH-LOW Pressure Control Dial


Patient Tube Inlet Adapter Power Switch
(with nozzle)

13
PROPER HANDLING

5 Connect a catheter and the catheter connector S6 with the patient tube.

Patient Tube

Catheter Connector S-6 Catheter

14
2. Operations
1 When the power switch is set to On, the suction pump starts suction-
ing.

2 Put a catheter in the patient's affected part such as mouth cavity, throat
or nasal cavity.

3 When the suction pressure is low or high, turn the HIGH-LOW pres-
sure control dial gently while watching the vacuum gauge dial carefully
to regulate the suction pressure to the specified value.

Cushion

Suction Bottle
Cover
LOW
HIGH

HIGH-LOW Pressure
Control Dial
(Adjusting the Suction
Pressure)
Vacuum Gauge

4 Although suctioning will stop automatically when secretion suctioned


becomes full in the suction bottle, pour out secretion as early as pos-
sible before the bottle becomes full.

5 Even if secretion is suctioned from the suction bottle to the pump


side, this equipment is still designed to store the secretion suctioned
in the cushion of the bottle cover. However, replace the cushion if
secretion is suctioned in the cushion. Be careful that secretion should
not remain in the cushion. (Use the cushion always in an empty state.)
15
PROPER HANDLING

About the float in the section bottle

The more secretion in the suction bottle, the float will rise along
with the volume of the secretion until the float activates the float
stopper (the suction stop float valve). When the suction stop float
valve is activated, it reduces the infiltration of secretion to the
suction pump.
If the float stopper does not work, the float remains rose in the top
of the suction bottle. In such a case, open the bottle cover and make
sure that the float can move up and down.

Float Stopper

Float

CAUTION Since the suction unit is suctioning continuously


the misted secretion in its suction piping and pump,
suction pressure and suction flow rate are affected
with its operating frequency. Be sure to have the unit
inspected by qualified repairers periodically. (It is
charged).

16
When stop suctioning temporarily
When stop suctioning temporarily, hold to fix the patient tube in the
receptacle for it, and put the catheter (Example 1) or patient tube
(Example 2) in the catheter holder as illustrated below.

Patient Tube

Cathete

Patient Tube
Receptacle
Patient Tube
Receptacle Catheter
Holder
Catheter
Holder

Example 1 Example 2

17
PROPER HANDLING

3. Handling after Use


The appropriate handling after use is important in order to use this equip-
ment properly for a long period of time. Handle this equipment following
the steps below.
1 After use, suction some refined Catheter
water or distilled water, wash away
the solid stuck in the tube.
* Stains can be removed more eas-
ily by suctioning tepid water Refined Water or
(about 40C) Distilled Water

2 Switch off the power, and pull out


the power cord. (In unplugging the Power Switch [Off ]
power cord, be sure to hold it by
the plug.)

3 Remove the suction bottle, inlet


adapters and bottle cover from
the housing, and pour out suction
things.
MS2-1400

Push the detachable
button of the suction
bottle holder, and Remove the Inlet Adapter.
remove the bottle
Suction Bottle Cover
cover.

Detachable Button

Suction Bottle
Holder

18
Removing the Cover for Cushion

When removed, make


While holding the sure that the filter for
bottle cover, remove cushion is in it, and not
the cover for cushion. stained.

About the Filter for Cushion


The filter for cushion is to reduce the infiltration of the misted
secretion and disinfectant to the suction pump. Replace the filter
for cushion with a new spare filter when the suction power be-
comes weak and it seems that the filter clogs.

If a spare filter is not available, and if urgent, this equipment


can be used without filter. Obtain the filter for cushion with
spares from the seller.

4 Remove all the parts shown on the next page and wash them in clean
water. Be careful of the parts missing when washing them.

19
PROPER HANDLING

Suction Unit Tube Inlet Adapter (yellow)

Patient Tube Inlet Tubing between Bottle Cover and Suction Unit
Adapter (white)
(with nozzle)

Patient Tube
Section

"O"-ring
Cover for (2 pcs) Toward the Bottle
Cushion Cover

Filter for
Cushion

Suction Bottle
Cover How to Fix the Bottle Packing
In fixing the bottle packing, match the
Bottle Packing hook on the bottle packing and the
recess on the bottle cover, and set it.
(Foolproof mechanism)
Float Stopper

Float Top Up
The float stopper has an
Float Guide orientation. Pay attention
to the orientation when
fixing it.
Float

Suction
Bottle

Suction Bottle Holder When removing the bottle holder,


holding the suction bottle and pull
down the bottle holder with hand.

20
MS2-1500

Suction Unit Tube Inlet Adapter Suction Unit Connecting Tubing


(yellow)

"O"-ring (2 pcs) Note: When fixing or removing


the suction bottle cap,
Cover for Cushion
remove the cover for
cushion first.
Filter for Cushion

Suction Bottle
Cap

Suction Bottle
Head

Removing and Fixing


Suction Bottle While tilting the suction bottle
Packing cap, put it on the suction bottle
head.
Float Stopper

Up
Float Top
The float stopper has an
orientation. Pay attention
Float Guide to the orientation when
fixing it.
Float

Suction Bottle

21
PROPER HANDLING

How to Remove the Patient Tube Cover

Turn the patient tube cover


Patient Tube Cover
counter-clockwise to
remove it. Remove it for
cleaning when it is stained.
(When fixing it, align the
housing and the hook of the
patient tube cover, and turn
it clockwise.)
Hook

5 Check to see that the stains are removed well. Then, disinfect or
sterilize the parts. (Refer to 23th page.)
* When using at home, sterilize and disinfect the parts as needed.

When Using at Home


1. When Using at Home without Sterilization and Disinfection
Wash all the parts thoroughly, and then dry them thoroughly.
After that, set them securely as they were set before.
2. When there is not enough time to dry
After setting all the parts and connected to the housing, perform
the idle run until the moisture in the patient tube and the suction
bottle are fully dried. (Occurrence of mold should be prevent.)

CAUTION Ifcleaning and/or drying of the following parts are


insufficient, it may result in mold (black or pink
mold). After use, clean and dry them thoroughly.
(Patient tube, patient tube
connection port, suction
bottle and accessories of
the suction bottle)
Dry the patient tube in
the manner of letting the
water droplets fall.

22
4. Sterilization and Disinfection
CAUTION

IMPORTANT POINTS FOR CHEMICAL DISINFECTION


1. Sterilize after remove stains well.
(Effective sterilization will not be expected if the stains are not
removed well.)
2. Do not mix one disinfectant with others.
(Chemical effect may lose or a dangerous element may be gen-
erated.)
3. Observe the density of the disinfectant prescribed in the manu-
al and/or package insert of the chemicals.
(Failure to observe the prescription may make the appliances
inferior.)
4. Be careful not to touch disinfectant or not to inhale the vapor-
ized chemicals. (It may chap your hands poisonous gas may be
inhaled.
5. After disinfection, wash out, dry well and store.

1. Method of Sterilization and Disinfection at the Medi -


cal Facilities such as Hospital

As to sterilization, high pressure steam sterilization (autoclaved steril-
ization) cannot be applied. Suggest you use Ethylene Oxide Gas
(EOG) or immersion disinfection using disinfectant.

Ethylene Oxide Gas (EOG) Sterilization (Refer to 26nd page.)


In the event of gas sterilization, perform it at below at below 50C
and remove the remaining gas by providing enough aeration after
sterilization.
Immersion Disinfection with Disinfectant (Refer to 26nd page.)
Remove the suction bottle and the accessories of the suction bottle,
and use the disinfectant which suites each material. (Observe the
concentration of a disinfectant, that is described in the package in -
sert or the attached document of the disinfectant.)

23
PROPER HANDLING

CAUTION Dont use sodium hypochlorite or povidone-


iodine. They may erode metallic parts.
Dont use phenolic, ampholytic surface active
agent or alcohol. They may crack resin.
(When using the PTFE suction filter, do not use
alcohol and the like solution. They do not suit
PTFE filter.)
Do not use strong acid water as it erodes metal-
lic parts

Suggest you use the disinfectants such as chlorhexidine gluconate (such as


Hibitane solution) or invert soaps (such as Osvan solution) that is gentle
with resin.

2. Method of Disinfection at Home


Suggest you use the immersion disinfection using disinfectant.
Immersion Disinfection with
Disinfectant (Refer to 26nd page.)
Remove the suction bottle and the
accessories of the suction bottle,
and use the disinfectant which
suites each material.
(Observe the concentration of
a disinfectant, that is described in
the package insert or the attached
document of the disinfectant.)

3. After Disinfection
After disinfection, wash out and
dry well. After disinfection, set the
suction bottle assembly as it was
before, and put it on the unit. Be
sure and put the packing in properly. Hook
(Refer to 20th page.)
In fixing the bottle cover on the
suction bottle, hook it on the bottle
holder and set it.
24
CAUTION Insert the suction
bottle holder until it
comes to the top of the
suction bottle. Then,
place the bottle cover.
If the suction bottle
holder is stopped
in the middle of the
suction bottle, the
bottle cover may not
be closed.

When not in use, the patient tube can be wrapped around the patient
tube cover. Hold the tip of the patient tube as shown in the figure.
To remove the patient tube cover, turn it counter-clockwise. (Refer to
22th page.)

Patient Tube Cover

Patient Tube

For storage, use the dust cove


supplied.

25
PROPER HANDLING
Reference Information About Sterilization and Disinfection of
Usable Usable (There are changes in appearance or of feeling.) Prohibited

Suction Bottle Suction Suction Bottle Bottle


Bottle Holde Packing
(1400 mL) Cover
Name of Parts

(Polycarbonate) (PC /ABS) (PC /ABS) (Silicon)


(Material)

(Type of disinfectant)
Name of typical product
(Chlorhexidine gluconate)
Hibitane solution, Hexizac solution

(Invert soaps)

Osvan solution, Hyamine solution
(Phtharal)
Disopa

(Alcohol)
Ethanol disinfectant, Isopropanol

(Sodium hypochlorite)
Milton, Purelox
* Metal corrosion

(Ampholytic surface active agent)


Usunon 51, Kinsaaru
Crack
(Phenol)
Creosol soap solution

(Povidone-iodine)

Isodine
Boiling sterilization

(Inclusive of washer-disinfectors)
High pressure steam sterilization

(autoclaved sterilization)
Ethylene Oxide Gas (EOG)

disinfection
* As the bottle holder contains the stainless steel parts, metal corrosion of these parts
will occur.
Phenol, and povidone-iodine may cause crack, dissolution or coloration.
26
Suction Unit

Patient Tube Adapter


Float Top Float C o n n e c t i o n Patient Tube Catheter Catheter Cover for
Float Guide Stopper Holder Cushion
Port Connector S-6
(Polycarbonate) (Polycarbo-
(Polycarbonate) (Silicon) O-ring (Vinyl chloride) (EVA resin) (PMP resin)
nate)
(Silicon)



Stain


O-ring
Tube hardening
hardening

O-ring Crystallization
deterioration inside the tube


Crack


O-ring
Pungent odor Tube hardening
hardening

Boiling sterilization and high pressure steam sterilization (autoclaved sterilization) may
cause deformation.
Note: This filter cannot be reused even it is sterilized. Replace it with a new one.
27
PROPER HANDLING
Reference Information About Sterilization and Disinfection of Suction Unit
Usable Usable (There are changes in appearance or of feeling.) Prohibited

Name of Parts Suction Bottle Suction Suction Suction


Bottle Bottle
(1500 mL) Head Bottle Cap Packing
(Material) (Glass) (PC /ABS) (ABS resin) (NBR)

(Type of disinfectant)
Name of typical product

(Chlorhexidine gluconate)
Hibitane solution, Hexizac solution

(Invert soaps)

Osvan solution, Hyamine solution
(Phtharal)

Disopa
(Alcohol)
Ethanol disinfectant, Isopropanol

(Sodium hypochlorite)
Milton, Purelox
Deterioration

(Ampholytic surface active agent)



Usunon 51, Kinsaaru
(Phenol)

Creosol soap solution Swelling
(Povidone-iodine)

Isodine
Boiling sterilization
(Inclusive of washer-disinfectors)
Damage

High pressure steam sterilization



(autoclaved sterilization)
Ethylene Oxide Gas (EOG)

disinfection
Phenol, and povidone-iodine may cause crack, dissolution or coloration.
Boiling sterilization and high pressure steam sterilization (autoclaved sterilization) may
cause deformation.
Note: This filter cannot be reused even it is sterilized. Replace it with a new one.

28
5. Cleaning of Housing
Wash the dirt off the outside of the housing with a wet cloth adding a little
bit neutral cleanser, and wipe it off with a dry cloth.
CAUTION

Never wash in water.


Do not wash with benzine, thinner,
abrasive cleaner (such as cleanser).
Doing so will cause the crack and
break of resin. Use only neutral deter-
gent.
Dont use phenol, ampholytic surface
active agent or alcohol. They may
thinner benzin
crack resin. (When using the PTFE
filter, do not use alcohol and the like
solution.)

6. Replacement of Exhaust Filter


Remove the exhaust filter cover to pick the exhaust filter out of the suction
bottle stand.
Replace the exhaust filter periodically (about once a month).

Exhaust Filter

Exhaust Filter Cover

29
MAINTENANCE AND INSPECTION
1. Daily Inspection
CAUTION Before operation be sure to inspect the following
items.
Inspect that the AC power cord plug is not deformed.
Inspect that the appearance of the suction bottle and its accessories
shows no crack and any other abnormalities..
Inspect that the connecting parts, dial and the like are not loose, and
work smoothly.
Check that on and off of the power switch moves smooth. Check that
the vacuum gauge indicator changes smoothly when the patient tube is
folded with fingers and turning the HIGH-LOW pressure control dial.
Check whether the exhaust filter stains or not. (Refer to 29th page.)

Vacuum Gauge

HIGH-LOW Pressure
Control Dial

Patient Tube

30
2. Periodical Inspection (User Inspection)
Check List
This is a user's inspection. Make use of the following list, and inspect the
unit periodically every 3 months. (Copy P31 to 32 and use them)

Function Check
Serial No.
Inspection Date : Year, month, day Inspector :

CONTENTS RESULTS SITUATION REMARKS

When turning the HIGH-


LOW pressure control
dial to HIGH, the suction Good Wrong
pressure increases. (Refer
to P30 to P35.)
When turning the HIGH-
LOW pressure control
Good Wrong
dial to LOW, the suction
pressure decreases.
No noise is generated. The
Dial is not loose. Good Wrong

The vacuum gauge


Good Wrong
indicator moves smooth.

The vacuum gauge


indicator indicates 0 Good Wrong
when the unit is stopped.

Call the seller when function check results in wrong and request the seller
or repair contractor for repair.

31
MAINTENANCE AND INSPECTION

The component parts are equipped or not.


Serial No.
Inspection Date : Year, month, day Inspector :

Crack or no Crack or
Description Equipped or not crack on deterioration, Re-
deformity, marks
appearance stain
Suction Unit Connecting
equip not equip equip not equip
Tubing
Patient Tube (with Adapter) equip not equip equip not equip
Catheter Connector
equip not equip equip not equip
S-6
Catheter Holder equip not equip equip not equip
HIGH-LOW Pressure
equip not equip
Control Dial
Power Switch equip not equip
Fuse Holder equip not equip
Vacuum Gauge equip not equip
Exhaust Filter equip not equip equip not equip
Power Cord equip not equip
Outside of Housing equip not equip
Patient Tube Cover equip not equip equip not equip
Cover for Cushion equip not equip equip not equip
Filter for Cushion equip not equip equip not equip
Others
Suction Bottle and Related Parts
MS2-1400
Suction Bottle (Plastic) equip not equip
Suction Bottle Cover equip not equip
Suction Bottle Holder equip not equip
Bottle Packing equip not equip equip not equip
MS2-1500
Suction Bottle (Glass) equip not equip
Suction Bottle Head equip not equip
Suction Bottle Packing equip not equip equip not equip
When the components are missing or degrading for spare, place orders with
the seller.
32
3. About Scheduled Inspection
In order to prevent sudden accidents and use this product in safety, the
scheduled inspection every 2 years by the manufacturer or the repair
contractor and overhaul every 5 years are required. In scheduled inspec-
tion, perform function and performance test in addition to replacement of
consumables such as rubber products.
On the structure of the unit, the longer the hours worked become, the
more the degeneration of parts progress. And also accumulations of dust,
mist suctioned and the like have progressed. It becomes difficult for the
unit to keep the ability by only performing the periodical inspection over
a certain period of time. The overhaul with the parts replacements for
new ones and a very thorough examination is essential to returning the
unit to the ability as it was. (The scheduled inspection is charged sepa-
rately.)

33
TROUBLE SHOOTING BEFORE CALLING A SERVICEMAN
Before contacting the store of purchase, serve the following remedies.
It may operate normally. If this does not remedy the phenomenon, and
when you cannot operate normally, do not disassemble the unit, but con-
tact the store of purchase.
PHENOMENON CAUSE REMADY

The pump The power cord separat- Insert the power cord in
does not ed from the wall outlet the outlet correctly. Call
run. (The or it disconnected. the store you bought after
electric moving a code and check-
power is not ing condition, when it
switched disconnected or may have
on.). disconnected.
The pump
stops occa-
sionally.

The fuse blew. See and check the fuse. If


(There is a fuse holder it blew, replace it with a
under the base of the new one. After replace it
housing.) with the new one, and if it
If the Power switch blew again, call the seller.
is set to On while
the suction pressure Fuse Holder
remains (i.e., in the
Fuse
state that the vacuum
gauge needle does not
return to 0) in this
equipment, it may
result in the trouble
that this equipment
does not work, or fuse
may blow.

It is thinkable of th Call the seller.


trouble of the inside
mechanism (motor or
others).

34
PHENOMENON CAUSE REMADY

The pump CHECK POINT


runs, but Fold the tube of between
does not the bottle and housing with
suction. fingers and shut the air.
When turning the HIGH-
Or the suc-
LOW pressure control dial
tion is weak. to HIGH and operating
it, it is normal that the
vacuum gauge indicates
around -75kPa (see note
*), indicating the normal
running of the pump. (The
vacuum gauge indicator
moves.)
* The default setting is
-75kPa. However,
since the performance
changes depending
on the period and
frequency of use, use
-75kPa as a guideline.

The suction bottle cover Replace the suction bot-


(suction bottle head or cap) tle cover (suction bottle
cracks head or cap) with a new
one.

The packing of the suction Set the packing properly.


bottle cover is not set or If it is old, replace it
set bent. with a new one. (Refer
to 20 or 21th page)

35
TROUBLE SHOOTING BEFORE CALLING A SERVICEMAN

PHENOMENON CAUSE REMADY

The pump The suction stop float Remove the bottle cover,
runs, but valve is activated. and separate the float from
does not (The float adheres to the float stopper. When any
suction. the float stopper.) secretion or the like is rec-
Or the suc- ognized attached to the float
tion is weak. guide, disassemble them.
After washing, cleaning
and drying, set them as they
were set before. (Refer to 20
or 21th page.)

PTFE filter (an Replace the PTFE filter (an


optional accessory) optional accessory) with a
clogs. new one.

The cover for cushion Set it properly.


is not set properly.
(Loose)

Crack of cover for Replace the cover for cush-


cushion ion with a new one.

The O-ring is not Set the O-ring properly. If


set in the cover for it is old, replace with a new
cushion. one.
Or, it is set bent. (Refer to 20 or 21th page.)

Others Call the seller.

36
ABOUT OPTIONAL ACCESSORIES
The following products are prepared as the optional accessories of Minic
S-II. Use them in order to use Minic S-II effectively.
- Optional Accessories -
Description Usage
PTFE Suction Filter Prevent the mist of the secre-
(Water breakthrough point: 14 psi) tion and the liquid moisture
(*) from flowing into the pump.
Recommend you to use this
filter for preventing infection.

With "O"-ring

Catalog number: 200191750


Packing unit: 1 piece

Suction Catheter MT This is a catheter for


Packing unit: Each 20 pieces suctioning.
There are four types of outer
diameter and color variations.

8FR (Outer diameter2.6 mm) / light blue


Catalog number: 340790040
10FR (Outer diameter: 3.3 mm) / gray
Catalog number: 340790001
12FR (Outer diameter: 4.0 mm) / white
Catalog number: 340790002
14FR (Outer diameter: 4.6 mm) / green
Catalog number: 340790003

* Water breakthrough point means the capacity not to pass solution if


it is within a regulated value in the canal natured filter. In addition,
if solvents and antibacterial, such as alcohol, are poured, care should
be taken since the solution may also pass by the pressure within a regu-
lated value.
37
ABOUT OPTIONAL ACCESSORIES

Description Usage
Olive Tube Kit It is used to suction the
Component parts: nasal discharge. Since it is
Olive (ultra-small, small, large, ball) available in four different
Joint sized, it can be used from
Patient Tube (with Adapter) children to adults. It is made
Catalog number: 140030155 of silicon rubber. Suctioning
Packing unit: 1 set without damaging the nasal
cavity is possible.

Suctioning Toothbrush By connecting it to the patient


Catalog number: 200190210
tube or catheter connector
Packing unit: 3 pieces
(S-6), user can brush teeth
while suctioning rinse water or
saliva. Due to the small brush
part, the brush can reach a
narrow part or inner part of the
oral cavity. (Toothbrush and
suction catheter are integrated
into a single piece.)

Suctioning Sponge By connecting it to the patient


Catalog number: 200190211
tube or catheter connector
(S-6), user can perform suction
Packing unit: 5 pieces
while cleaning the inside of
the oral cavity. Suction can
be switched on or off with the
heart-shaped hole attached to
the holder.

38
Description Usage
Special Carry Bag With the shoulder belt
Catalog number: 200190224
supplied, it is convenient
Packing unit: 1 piece
to carry. Since the top
cover can be opened
immediately, user can
respond quickly in an
emergency.

When placing orders with the seller, specify the name of the product,
catalog number and quantity.

39
ABOUT OPTIONAL ACCESSORIES

1. Installation of PTFE Suction Filter


Caution
1. The PTFE suction filter degenerates gradually as the mist of the
secretion adheres to the filter, causing clogging. The effect of suc-
tion filter will be down in a month or so. Replace the PTFE suction
filter with a new one when the suction power becomes weak and it
seems that the filter clogs.
2. If secretion is suctioned into the PTFE suction filter, the PTFE suc-
tion filter will be clogged even at one time, and it may not be able
to suction any more.
3. When the suction power becomes weak due to leakage from the
inlet adapter, check the O-ring, and replace it when necessary.
4. This filter cannot be reused even it is sterilized. Replace it with a
new one.
5. The life of the filter depends on the frequency of use, properties of
secretion and chemicals for disinfection used.
6. It is recommended to prepare spare filters all the time to enable re-
placement of the filter with a new spare filter when the suction
power becomes weak and it seems that the filter clogs.

PTFE Suction Filter

Set the seal printed


side facing to the
front.

40
2. Installation of Olive Tube Kit

Patient Tube
Joint
Olive

3. Installation of Suctioning Toothbrush

4. Installation of Suctioning Sponge

41
ABOUT OPTIONAL ACCESSORIES

- Replacement Parts -

1. Catheter Connector 2. 7x11 Suction Tube 3. A Set of Plastic


(S-6) with Adapter 2 m Suction Bottle Head
Catalog number: 200190225 Catalog number: 200190082 Catalog number: 200191250
Packing unit: 5 pieces Packing unit: 1 piece Packing unit: 1 piece

4. Suction Bottle 5. 1400 mL Plastic 6. 110 x 130 Silicon


Holder Suction Bottle Packing
Catalog number: 200191251 Catalog number: 200191264 Catalog number: 200191254
Packing unit: 1 piece Packing unit: 1 piece Packing unit: 1 piece


7. Cover for Cushion 8. Patient Tube 9. Suction Unit


(with filter) Connection Port Connecting Tube
with Nozzle (V-1) (YEL YEL) V-2
Catalog number: 200191253 (Bottle head side)
Catalog number: 200191263 V-3
Packing unit: 1 piece

Packing unit: 5 pieces (Suction unit side)
Catalog number: 200191255

Packing unit: 1 piece

42
10. Filter 11. Catheter Holder 12. Suction Bottle
* Only filter of above item Tube (blue)
Catalog number: 200190037
7. Cover for Cushion (with
Packing unit: 1 piece Catalog number: 200191262
filter)
Packing unit: 5 pieces
Catalog number: 200191256
Packing unit: 5 pieces

13. Float COMP 14. Silicon Float 15. A Set of Glass


* The item 14. Silicon Float Stopper Suction Bottle
Stopper is excluded. Head
Catalog number: 200191260
Catalog number: 200191261 Packing unit: 1 piece Catalog number: 200191270
Packing unit: 1 piece Packing unit: 1 piece

16. 1500 mL Glass 17. 55 x 65 Bottle


Suction Bottle Packing
Catalog number: 200191342 Catalog number: 200190692
Packing unit: 1 piece Packing unit: 5 pieces

When placing orders with the seller, specify the name of the product,
catalog number and quantity.

43
SPECIFICATIONS
Model Minic S-II
Approval No. 22300BZX00311000
Type MS2-1400 MS2-1500

Classification by Japanese Law Medical Equipment (32), Aspirators


for medical use
Japanese Medical Device Nomen- Portable electric aspirator
clature (JMDN code) (36616030)
Controlled medical devices
Medical device classification
(Class II equipment)
Suction pressure (Maximum) - 85kPa ( - 63.75cmHg )
Suction volume (Maximum) 20 L/minute
Suction bottle capacity 1400 1500
Power supply AC mains, 50, 60 Hz
Power consumption 90 VA50Hz100VA60Hz

Motor Single-phase induction motor, AC


mains, 30 W, 50 or 60Hz
Pump Piston type single pump head
Type of protection against elec- Class II equipment
tric shock
Classification according to the
degree of protection against Type B applied part
electric shock
Continuous run time Maximum six hours
Room temperature 10 to 40C
Applicable temperature range
Relative humidity 30 to 75%
Storage/transportation Environ temperature -10 to 60C
temperature range Relative humidity 10 to 75%
Dimensions mm 395 157 274 398 157 288
(Exclusive of Hook) )

Mass Approx. 5.0kg


Specification expresses the performance at the time of the delivery in which tolerance by the
makers regulation is included. Each above-mentioned performance is an initial value. Since
the above-mentioned performance falls with the frequency to be used, in order to use mainte-
nance of a performance and to use this unit safely, perform the maintenance check in every
three months (user inspection) and periodical inspection by a specialist once 2 years.
44
ABOUT THE MARK SHOWN ON THIS MANUAL
The meanings of the marks shown on the label of the rear of this unit are
as follows:

Mark Meanings Details

It shows that Type B is the


classification of the applied part
(The part that the appliance touches
Type B Applied to the patient) according to the type
Part of protection against electric shock.
It conforms to the claims postulated
about Type B applied part specified
to JIS T 0601-1.

It is shown that the degree of


protection against electric shock is
class II. Class II equipment is the
Class II equipment which does not depend
equipment for the protection to an electric
shock only on basic insulation, but
was equipped with the additional
safety measures.

Caution. Read the OPERATING


Warning INSTRUCTIONS carefully and
(Refer to the
master it before using this device so
OPERATING that you can use it correctly.
INSTRUCTIONS)

Electric power source of this


Alternating
Current
equipment is alternating current.

45
ABOUT DISPOSAL
When you will dispose of this unit (inclusive of the accessories
and optional parts), please be directed by the administrator of your
residential area. (It can cause environmental pollution.)