Scope of work for supply of Turbine Flow Meters
1. The vendor shall supply the turbine flow meters as per the bill of materials and specifications
mentioned herewith,
2, The vendor shall ensure that the specifications mentioned herewith are fully satisfied by the
product being offered,
3. All the tests mentioned in the acceptance criteria shall be carried out by the vendor. Intimation
to DRDL shall be sent to DRDL prior to conducting the acceptance tests. Hydro test shall be
done in the presence of rep DRDL. Calibration of flow meters shall be done in the presence of
rep DRDL. Calibration shall be done at 10 operating points evenly distributed over the whole
operating range and covering the extreme ends of flow range.
4. After successfully completing the acceptance tests at firms premises in the presence of rep
DRDL, the items shall be dispatched to DRDL.
6. Packaging and transportation of the flow meters is the responsibility of the firm.
7. The firm shall supply all the flow meters along with relevant inspection/ test reports and
operating manual / instructions / installation procedure etc. All calibration certificates shall be
supplied along with flow meters.
heAnnexure ~ 3
Specifications of Turbine Flow Meter 8" for Air
Parameter Value
Operating pressure | 150 bar (max)
30 bar (min)
Line size 8"
Flow ran; 320 t0 3200 AM3/hr
Operating, 20 deg Cto + 45 deg C
temperature
Working media Diy Air
‘Connections Flanged end (class 900 or higher)
[Make Rockwin / Flotech or equivalent
Calibration 10 point calibration
Output Voltage output, reading to be corrected for pressure
and temperature using flow computer
Installation Horizontal
Linearity +1 % of reading
Power suppl 220 V AC
Excitation voltage _| 24 VDC
Proofin; Alli weather proof, to be operated in open atmosphere
Attachments Necessary standard attachments like pressure sonsor,
temperature sensor and flow computer in order to
provide output in Voltage with corrections for
pressure and temperature
Leakage Bubble Tight
Acceptance criteria:
1. Hydro tests at 1.5 times of maximum working pressure bar for 10 minutes in the présence of
rep DRDL.
2. Chemical, mechanical and ultrasonic test certificates for materials of construction,
te
3.10 point calibration covering extreme ends of flow range. Calibration to eb. done in the
presence of rep DRDL.
4. Certificate of conformity for specifications.
W teoory® tnt)
Oe‘Specifications of Turbine Flow Meter 3" for Air
Parameter | Value
‘Operating pressure | 150 bar (max)
30 bar (min)
Line size 3
Flow range 4010 400 AM3/ar
Operating, “20 deg C to + 45 degC
temperature
Working media Dry Air
Connections Flanged end (class $00 or higher)
Make Rockwin / Flotech or equivalent
Calibration {10 point calibration
Output Voltage output, reading to be corrected for pressure
and temperature using flow computer
Installation —
~__ | Horizontal
[Linearity +1 % of reading
Power suppl 220 VAC
Excitation voltage. 24 VDC
Proofing lll weather proof, to be operated in open atmosphere
‘Attachments Necessary standard attachments like pressure sensor,
temperature sensor and flow computer in order to
provide output in Voltage with corrections for
pressure and temperature.
Leakage Bubble Tight
Acceptance criteri
1, Hydro tests at 1.5 times of maximum working pressure bar for 10 minutes in the presence of
rep DRDL.
2. Chemical, mechanical and ultrasonic test certificates for materials of construction,
we
3.10 point calibration covering extreme ends of flow range. Calibration to eb done in the
presence of rep DRDL.
4. Certificate of conformity for specifications.
i
§ EL
eo
Go ete gc
Beer”~
List of Deliverables
DrINEXUREC!
SNo
Item
Turbine Flow Meter 8° for air as per specifications
given in annexure A
02 (Two only)
Turbine Flow Meter 3” for air as per specifications
mentioned in annexure B
02 (Two only)
ih.
Caled:
SUNITA DEVE JENA
Scientist ‘G
Head, LRPD
DRODL, Hyderabad,