Monitoring load

current with a DMM


Application Note

A digital multimeter (DMM) has The current clamp is designed


become a necessity for electrical to clamp around a single conduc-
system measurements, and most tor and uses the fluctuating mag-
electricians and electronic tech- netic field around the conductor
nicians would feel lost without to induce a current. Because it
one. The features available in uses a transformer action, the
todays meters make them versa- current transformer is only
tile enough to find most prob- usable on ac circuits. However,
lems in electrical and electronic there are other clamps like the
circuits. Fluke i410 designed to use the
There are, however, some Hall Effect sensor which allows
measurement tasks that make a them to be used on both ac and
technician think twice about his dc circuits.
DMM. As long as the DMM is safety
Take for instance load cur- rated for the environment in
rents. Many DMMs are designed which you will be making the
to safely measure current up to measurement, say CAT III for
2, 10, or 20 amps depending on electrical panels and switch
the DMM. Yet load currents found gear, you can safely attach an
in most industrial environments appropriately rated current clamp
are much higher and backed by to your DMM and make the
high energy. On the surface it is measurement. Just remember
not an attractive match. that when you are using a
Thats where a DMM acces- DMM/current transformer combi-
sory called a current transformer nation you need to combine the
comes in handy. A current trans- measurement specifications of
former, like the Fluke 80i-400, the current transformer and the
can be employed to overcome DMM to get overall measurement
the DMMs limitations for meas- accuracy.
uring ac current. Like a power Using the current clamp to
transformer, the current clamp overcome the current limitations
uses windings around a lami- of the DMM allows you to use a period, the meter will constantly
nated core to step down the number of DMM features to ana- update built-in storage registers
circuit current within the meas- lyze load currents. with the latest minimum and
urement range of the DMM. The maximum measured value. As an
example, lets say you want to
step-down ratio is commonly Recording minimum and determine the maximum load
1,000 to 1. So measuring 100 maximum current
amps in a load circuit through a current drawn on a branch cir-
current clamp will apply 100 mA Some DMMs are capable of cuit for a 24-hour period of time.
to the DMMs current input. recording the maximum and Connect the current clamp leads
Thats well within the DMMs minimum measurement over to the current inputs of a DMM.
capability and, conveniently, a time. The Fluke 170 Series, for In the case of the Fluke 179, one
direct reading from the DMM dis- instance, has this capability in its lead goes to the mA input and
play. Just switch the DMM to mA min/max mode. After setting the the other lead to the COM input.
ac mode and think amps when DMM up for a measurement, Turn on the meter and place it in
reading the display. attach the clamp accessory to the the mA ac function. Clamp the
DMM and place the clamp current clamp around the single
around the conductor to be conductor carrying the current
measured. Then simply press the you want to measure. Activate
min/max button and the record- the min/max function by press-
ing begins. During that time

From the Fluke Digital Library @ www.fluke.com/library


ing the button and let the DMM DMM. Place the clamp around Interval logging
sit clamped to the circuit. the lead of a motor, turn the A specific measurement period,
Twenty- four hours later, you DMM function switch to the ac such as one second or 15 min-
can return to the meter and read mA/A function and activate the utes, is the characteristic that
both the maximum and the mini- fast min/max by pressing the sets interval logging apart from
mum current measured by the corresponding buttons. Turn on event logging in the Fluke 189.
meter by simply pressing the the motor and let it come to full Through the DMMs setup func-
min/max button. As an extra speed. Shut down the motor and tion, you can configure the
bonus, the Fluke 179 will also recall the maximum value measurement interval from one
display the average of all read- recorded in the DMM. second to 99 minutes 59 sec-
ings during the measurement If you are using a 189, these onds in one-second increments.
period. readings can be stored in the Setting the interval to 00 min-
This type of measurement is DMMs memory and recalled utes and 00 seconds turns off
fine for monitoring changing load later or downloaded to a PC for interval logging.
currents lasting 200 ms or further analysis or record stor- If, for example, you wanted to
longer. A special form of age. monitor the current drawn on a
min/max is needed for capturing branch circuit for a 24-hour
faster changes in current such as Logging period, you could set up the
inrush current from fractional Fluke 189 to perform logging in
horsepower motors. The Fluke 189 Logging Meter can
five minute intervals. After 24-
be used to capture a series of
hours, the DMM would collect
readings over time using either
Capturing fast changes interval or event triggers. In
288 measurement periods of
in current high, low and average measure-
either case, the logging function
ment data. This data would be
Some DMMs have a function that stores the highest, lowest, and
extracted from the DMM through
can measure the value of signals average values measured over
FlukeView Forms and displayed
as short as 250 microseconds. In time. In addition, the logging
on a PC.
the Fluke 180 Series for example, process time stamps each start
the fast min/max function will and stop time of the measure- Event logging
capture both the positive and ment period in real time. It is the Triggering a new measurement
negative peak of an ac signal. triggering of the measurement period when the measured value
This feature comes in handy period that differentiates interval appears unstable or becomes
when you need to capture the logging from event logging. To stable is the differentiating char-
value of circuit transients. To view all the data from these acteristic of event logging in the
measure the start current of a measurement periods however, Fluke 189. Stable is defined as
motor, attach a current clamp to you must have access to a PC and a measured value that stays
the COM and mA inputs of the use FlukeView Forms software. within 4 percent for one sec-
ond or more. If the measurement
increases or decreases more than
four percent, an unstable period
is triggered. The unstable period
ends when the signal meets the
criteria for a stable value.
So, in our 24-hour branch
current measurement example,
each time the load current
changes more than four percent,
the DMM will start a new meas-
urement period. If the jump is
quick enough, there will not be a
recorded unstable period, just a
new stable period with new val-
ues. For current measurements,
the unstable period must last 50
milliseconds or more to be
recorded by the DMM.

Monitoring load current with a DMM Fluke Corporation 2


FlukeView Forms As can be seen in Figure 1, In conclusion
FlukeView Forms is a very pow- the monitoring of the branch cir-
erful software package that pro- cuit resulted in both stable and Combining a DMM with a current
vides access to stored unstable periods. This can be clamp accessory enables accu-
information in a 180 Series DMM, expected with equipment coming rate measurement and analysis
as well as other Fluke products. on and off during the day. If this of load currents.
With FlukeView Forms loaded on were a circuit of a motor running Fluke has a wide range of
a PC, you can transfer logging constantly, then we might be current measurement accessories
and measurement data from detecting a problem with the to fit just about any application.
DMMs to a PC and display that unstable readings. Such conclu- Current clamps with a maximum
data in a graphical as well as sions of course, will depend on current rating ranging from 200
tabular form. the application. But the branch to 2,000 amps are available at
circuit activity can be clearly your nearest Fluke distributor.
seen in the graphical display These same distributors stock a
and further analyzed with the collection of Fluke DMMs with
measurement periods shown in varying feature sets so you can
tabular form. purchase the features you specif-
ically need in your work.

Figure 1.

Fluke. Keeping your world


up and running.

Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
2002 Fluke Corporation. All rights reserved.
Printed in U.S.A. 5/2002 1989128 A-ENG-N Rev A

Potrebbero piacerti anche