The Theory of HPLC

Hydrophilic Interaction Chromatography

Wherever you see this symbol, it is important to access the on-line course
as there is interactive material that cannot be fully shown in this reference
i manual.
Aims and Objectives

Aims and Objectives

Aims

Introduce Hydrophilic Interaction Chromatography (HILIC) as a novel mode of


chromatography
Present advantages and limitations of the HILIC separation mode
Introduce Electrostatic Repulsion Hydrophilic Interaction Chromatography (ERLIC)
as a new HPLC option for hydrophilic analytes.
Present concepts of ion-exchange chromatography and its separation potential for
certain ionic samples

Objectives

At the end of this Section you should be able to:

Understand the benefits and limitations of using HILIC for certain applications
Recognise some of the most important parameters that can be used to alter the
separation in HILIC separation mode
Demonstrate an awareness of the benefits and limitations of HILIC over normal
and reversed phase chromatography
Understand the benefits and limitations of using ERLIC for certain hydrophilic
samples
Recognise some of the most important parameters that can be used to alter the
separation in an ERLIC separation
Demonstrate an awareness of the limitations of the ERLIC separation mode
Content

Introduction 3
General Considerations 4
Normal Phase Chromatography 5
HILIC Separation Mechanisms 6
Applications 9
Stationary Phases 15
Columns 17
Eluent Systems 17
Buffers and Additives 19
Gradient 19
ERLIC 21
Ion Exchange Chromatography 23
Anion-Exchange Mixed Modes of Separation I 25
Anion-Exchange Mixed Modes of Separation II 26
Anion-Exchange Mixed Modes of Separation III 28
ERLIC Vs. HILIC 29
References 30

Crawford Scientific www.chromacademy.com 2


Introduction

Chromatography is the collective name for a set of techniques used for the separation of a
mixture of components. In any mode of chromatography, the separation of components is
promoted by their difference in affinity for, or solubility in, two different phases the so
called Stationary and Mobile phases. The chromatographic separation of hydrophilic
compounds (highly water soluble, usually more polar), has traditionally been regarded as
difficult:

In Reversed Phase HPLC, the highly aqueous eluents required to gain any retention of
polar compounds are known to cause problems such as non-reproducible retention times
and low separation efficiencies, even with so-called Aqueous stationary phase types.

Crawford Scientific www.chromacademy.com 3


Hydrophilic interaction chromatography (HILIC) is a type of chromatography based on a
mixed mode of retention mechanisms. It describes a type of normal-phase
chromatography that employs a polar stationary phase (like silica or a polar bonded
phase) and an aqueousorganic mobile phase, in which the aqueous content is the
strong solvent (typical initial eluent composition is 98% organic / 2.0% water).[1, 2]

Despite being regarded as a new chromatographic method, the origins of HILIC


separation modes date back to the 1970s; however, the term HILIC was coined by Alpert
only in 1990.[2,3] Because of the relative polarities of the mobile and stationary phase,
HILIC chromatography is sometimes known as Reverse Reversed Phase
Chromatography.

General Considerations

In HILIC mode, the stationary phase is hydrated with a slow moving layer of water, where
hydrophilic compounds are preferentially retained compared with hydrophobic ones. To a
first approximation, the retention order of a series of analytes in HILIC is the opposite of
that in reversed-phase chromatography. HILIC can be used in certain situations where
reversed phase chromatography fails or is not efficient:[1, 2]

Samples with limited solubility in water or highly aqueous mobile phases


Samples that contain very polar compounds that are not retained adequately in
reversed phase
Hydrophilic water-soluble analytes, which are intractable to reversed-phase and/or
ion-exchange chromatography

HILIC presents the added advantage of using acetonitrile, which has low UV absorbance
(for better detection sensitivity) and low viscosity (for high chromatographic efficiency).
Drawbacks of the technique are related to the increasingly higher prices of acetonitrile, its
variable supply and the increasing focus on green chromatography where the use and
disposal of organic solvents is being driven down. Some of the general approaches to
reduce consumption of acetonitrile in HILIC include:

Reduce content of acetonitrile in the mobile phase


Replace of acetonitrile with alcohols, THF or other solvents
Use columns with smaller internal diameter

The use of ethanol and higher alcohols as alternative solvents, as well as the use of
reduced dimension chromatography columns continues to be the foundation of efforts to
reduce acetonitrile consumption.

Crawford Scientific www.chromacademy.com 4


i

The HILIC-MS process, oversimplified

A mobile phase composed of a low fraction of water and a large fraction of organic solvent
(like acetonitrile) provides an ideal composition for efficient desolvation in electrospray
ionisation mass spectrometry.

Normal Phase Chromatography

In order to fully understand the nature of HILIC chromatography lets first look at the
nature of Normal phase chromatography in order that we can distinguish between Normal
Phase and HILIC modes.

Normal phase chromatography was, chronologically, the first Liquid Chromatographic


technique. Tswett used this mode to separate plant pigments using a calcium carbonate
stationary phase with a petroleum ether mobile phase.[4]

By definition, normal-phase HPLC utilises a stationary phase that is more polar than the
mobile phase.

Typical stationary phases include bare silica as well as cyano, diol, and amino bonded
phases. Typical mobile phase constituents include organic solvents such as hexane and
ethyl acetate. One of which is non-localising (the weak solvent), whilst the other so
called localising solvent competes for surface retention sites with the analyte and
therefore acts as the strong solvent.

The retention mechanism in normal phase HPLC is based on polar adsorption of either
the solvent molecules or the analyte onto the polar stationary phase surface. If the solvent
molecules are localising they will be adsorbed onto the stationary phase surface,
displacing the analyte and hence effecting analyte elution.

The order of elution is least polar first, followed by increasingly polar (less hydrophobic)
analytes.

Crawford Scientific www.chromacademy.com 5


i

The normal phase chromatographic process (oversimplified)

A dynamic competition for sites on the stationary phase between the analyte (phenol) and
eluent (acetonitrile) molecules is established, and when the eluent concentration
increases, its presence on the stationary phase becomes dominant and analyte molecules
are displaced.

HILIC Separation Mechanisms

As was previously described, HILIC describes a type of normal-phase chromatography


that employs a polar stationary phase but uses aqueousorganic mobile phases, as
opposed to normal phase modes that we have seen use organic solvents only.

In contrast to reversed-phase chromatography, the aqueous component of the mobile


phase (for example, water or buffer) serves as the strong solvent, and the organic
component (for example, acetonitrile) is the weak solvent. Analytes are eluted in order of
increasing hydrophilicity.[1]

Due to the polar nature of the stationary phase, water molecules concentrate at the
surface, and a water enriched layer is thus created at the silica surface.

It has been proposed that in HILIC mode retention occurs as the analyte partitions
between the bulk mobile phase and the water-enriched layer which hydrates the
hydrophilic stationary phase. This is in contrast to retention in conventional normal-phase
chromatography, which occurs by adsorption at the polar stationary phase surface.[5,6]

Crawford Scientific www.chromacademy.com 6


Some authors have suggested that the HILIC mechanism is similar to that in normal
phase chromatography. Several attempts have been made to establish whether the HILIC
mechanism involves partitioning or adsorption; however, more research is required for
absolute clarity and the debate still continues.[7]

HILIC separation mechanism (partition in the aqueous mobile phase)

Crawford Scientific www.chromacademy.com 7


i

HILIC separation mechanism (interaction with the stationary phase)

Crawford Scientific www.chromacademy.com 8


Applications

HILIC can be used to address certain limitations of reverse phase chromatography:

Samples with limited solubility in water or highly aqueous mobile phases


Samples that contain very polar compounds that are not retained adequately in
reversed phase mode

To list the full range of HILIC application areas is prohibitive, its flexibility makes it suitable
to a multitude of application types. Examples of some interesting applications are shown
below:

Agrochemistry:[8] Chlormequat and mepiquat are quaternary ammonium compounds


used as plant growth regulators to reduce unwanted longitudinal shoot growth without
lowering plant productivity.

Full-scan HILIC LCMS chromatograms of chlormequat and mepiquat standard solutions.

Column: bare silica 150mm2.1mm, 3m.

Eluent system: The gradient elution started with a 0.5 min isocratic step at 60% of
solvent A (acetonitrile) and 40% solvent B (50mM formic acidammonium formate buffer
solution at pH 3.75), followed by a linear gradient of solvent A down to 40% in 0.5 min and
followed by an isocratic step of 2.5 min at 40% of solvent A.

Eluent flow rate: 400L/min

Crawford Scientific www.chromacademy.com 9


Bioanalysis:[9] Sildenafil citrate, is used to treat erectile dysfunction and pulmonary
arterial hypertension.

HILIC-MS determination of sildenafil citrate in human plasma.

Column: bare silica 50mm5.0mm, 3m.


Eluent system: acetonitrilewaterTFAacetic acid (92:8:0.025:1, v/v/v/v).
Eluent flow rate: 0.3mL/min

Cosmetics:[10] Due to its moisturizing and keratolytic properties, allantoin as an active


ingredient widely used in the preparation of cosmetics.

HILIC determination of allantoin from a commercial lotion

Column: triazol-bonded silica column 4.6mm ID 250mm, 5m.


Eluent system: acetonitrile and water in the ratio 90:10 ammonium acetate buffer (2.0
mM, pH 4.0)
Eluent flow rate: 1.0mL/min

Crawford Scientific www.chromacademy.com 10


Environmental analysis:[11] Estrogens are a group of strong endocrine-disrupting
chemicals that can cause the feminization of male fish at trace level concentrations,
estriol-3-sulfate is one of such compounds

MRM chromatogram of estriol-3-sulfate from a water river sample. Detection ESI negative
ion mode.

Column: Zwittterionic HILIC mixed mode 100mm 2.1mm ID, 5m.


Eluent system: Mobile phase A consisted of acetonitrile/aqueous ammonium acetate
(5mM, pH 6.80) (5/95, v/v) and mobile phase B consisted of acetonitrile/aqueous
ammonium acetate (5mM, pH 6.80) (95/5, v/v). The step gradient started at 20% B for 2.0
min; 30% B for 8.0 min; 80% B for 8.0 min; and 20% B for 7.0 min.
Eluent flow rate: 0.15mL/min

Food analysis:[12] Ascorbic acid and its derivatives are of overriding importance in human
diet.

Crawford Scientific www.chromacademy.com 11


HILIC chromatogram of a standard mixture of ascorbic acid derivatives.

Column: Diol 4.6 i.d.250 mm, 5m.


Eluent system: acetonitrile and water (85:15, v/v) 66.7mM ammonium acetate.
Eluent flow rate: 0.7mL/min

Forensic analysis:[13] Saxitoxin, one of the most potent natural toxins known. It is
naturally produced by certain species of marine species. Human toxicity and mortality can
occur after ingestion of these substances.

HILICMS analyses of an Alexandrium tamarense (dinoflagellate) extract

Column: Cyano 250mm2.0mm or 4.6mm i.d.


Eluent system: acetonitrile and water (95:5, v/v) with 10.0mM ammonium acetate buffer
(pH=2.5).
Eluent flow rate: 0.8mL/min

Crawford Scientific www.chromacademy.com 12


Metabolomics:[14] The metabolome represents the collection of all metabolites in a
biological organism, which are the end products of its gene expression.

HILIC-MS determination of citric acid.

Column: bare silica 4.6 i.d.100 mm, 5m.


Eluent system: The separation was done at 30 oC with a linear gradient elution of mobile
phases A and B at a flow rate of 0.3 mL/min. Solvents A, B were acetonitrile and a 10mM
ammonium formate buffer with pH adjusted to 2.5 by formic acid, respectively. B was kept
at 25% for 3 mins, then it was changed from 25 to 50% in 25 mins.
Eluent flow rate: 0.3mL/min

Pharmaceutical chemistry:[15] Analysis of highly polar pharmaceutical products.

Crawford Scientific www.chromacademy.com 13


HILIC separation of anthracyclines (antitumor antibiotics used for the treatment of cancer
chemiotherapy).

Column: C18 250mm4.6mm i.d. 5m.


Eluent system: acetonitrile and water (90:10, v/v) with 20.0mM sodium formate buffer
(pH=2.9).
Eluent flow rate: 0.7mL/min

Pharmacokinetics:[16] Zanamivir has been used in preventing, controlling, or rapidly


reducing certain types of influenza.

HILIC-MS determination of zanamivir from rat plasma.

Column: bare silica 50mm2.1mm, 3m.


Eluent system: 80% acetonitrile and 20% ammonium acetate (10mM) with 1% methanol
Eluent flow rate: 0.3mL/min

Crawford Scientific www.chromacademy.com 14


Proteomics:[17] Fetuins are blood proteins that are made in the liver and secreted into the
blood stream.

MALDI TOF mass spectra obtained from the fetuin tryptic digest after HILIC separation.

Column: Zwittterionic HILIC mixed mode C18, 50mm 2.1mm ID, 3.5m.
Eluent system: 80% acetonitrile and 20% of an aqueous solution of formic acid at 0.5%.
Eluent flow rate: 0.3mL/min

Stationary Phases

In essence HILIC requires a hydrophilic stationary phase to adsorb a water layer for the
partitioning process to take place, however there are several ways in which to achieve this
and also to fine tune the HILIC separation process.

The separation mechanism in HILIC is not well understood, but it is known that
parameters such as eluent pH and analyte-stationary phase interactions will affect the
separation to some extent.

Modern HILIC stationary phases are either neat silica or use ionic (or ionisable) ligands
bonded to the silica surface. The use of ligands capable of undergoing electrostatic
interactions can add an extra dimension to the separation when analysing ionisable
compounds.

Accordingly, HILIC stationary phases can be divided into three main categories:[18,19]

Neutral: polar surface with no electrostatic interactions

Charged: strong electrostatic interaction -the stationary phase contains anionic or cationic
functional groups

Zwiterionic: weak electrostatic interaction -the stationary phase contains both positive
anionic and cationic functional groups

Crawford Scientific www.chromacademy.com 15


i

In HILIC, different mixed retention modes can occur according to the analyte and column
functionality.

Crawford Scientific www.chromacademy.com 16


Columns

A huge variety of different types of polar stationary phase have been used in HILIC,
including bare silica, aminopropyl, diol and zwitterionic phases bonded to silica or polymer
matrices; nonetheless, bare silica is by far the most widely used stationary phase. HILIC
columns from selected manufacturers are presented below.

Eluent Systems

Mobile phase strength is one of the important parameters in HILIC. Unlike reverse phase
chromatography, water in HILIC is the stronger eluting solvent.

In a typical HILIC mobile phase, acetonitrile is used as a weak eluent component and
water or aqueous buffer as a strong component. Due to the recent shortage of acetonitrile,
alcohol had been considered as an alternative to replace acetonitrile; however, the efforts
to use alcohol as a weak eluent in HILIC had often fail due to insufficient retention and
quality of the separation.[8]

Mobile phase pH plays an important role in the HILIC separation. Acidic or neutral mobile
phase pH is commonly used due to the instability of silica-based columns at high pH.[20]

Solvent strength (from weakest to strongest) for HILIC is in generally as follows: THF <
acetone < acetonitrile < isopropanol < ethanol < methanol < water. Unfortunately, this is
not always true, and sometimes is not possible to predict solvent strength. The interactive
experience presented opposite, illustrates a separation where the solvent strength is as
follows: acetonitrile < THF < isopropanol < methanol, slightly different to what is expected.

Very strong solvents, such as dimethylformamide or dimethylsulfoxide, will usually result


in poor peak shapes and are not recommended. These solvents will generally have to be
Crawford Scientific www.chromacademy.com 17
diluted with a weaker solvent, such as acetonitrile, before satisfactory peak shape can be
obtained.

HILIC separation of anthracyclines (antitumor antibiotics used for the treatment of cancer
chemiotherapy) on a porous silica column. Mobile phase: sodium formate buffer (20
mM, pH 2.9)

Crawford Scientific www.chromacademy.com 18


Buffers and Additives

HILIC uses buffers and additives to achieve high column efficiency and reproducibility. In
general terms, the buffer concentration should be in the range of 10 20 mM and
additives should be used in concentrations usually not exceeding 1.0%.

Due to their poor solubility in high organic mobile phases, phosphate buffers are not
recommended for HILIC applications. Phosphate buffers are also incompatible with MS
detection.

Buffers or additives above pH 6 usually are not recommended for HILIC applications
because they may enhance the slow dissolution of the silica support; however, with the
advent of new HILIC stationary phases, modern columns are capable of dealing with
severe conditions of pH (up to about pH of 10).[21]

Buffer/Additive pKa Used for HILIC? Further information


TFA 0.3 YES Ion pair additive, can suppress MS signal.
Used in the 0.01 0.1% range
Formic Acid 3.75 YES Used in the 0.1 1.0% range
Acetic Acid 4.76 YES Used in the 0.1 1.0% range
Formate 3.75 YES Used in the 1.0 10.0 mM range
Acetate 4.76 YES Used in the 1.0 10.0 mM range. Sodium
or potassium salts are not volatile.
Phosphate 2.15, NO Will reduce column lifetime.
7.2

To maintain maximum HILIC separation performance, always use high quality eluent,
buffers and additives, and use good eluent preparation practice:

Filter all aqueous buffers prior to use


Particulate solvents will generally clog the column
Bacterial growth can be prevented by adding small amounts of organic modifier to
the buffer system
Degas all solvents before use
Use freshly prepared eluent systems wherever possible

Gradient

In reverse phase chromatography, highly hydrophilic compounds will elute close to hold-
up time, which may adversely affect their resolution and reliable quantification. The
increased retention of highly polar, ionisable or ionic compounds in the HILIC mode
effectively overcomes this problem.[22]

Retention in HILIC decreases with mobile phase polarity (which in practical terms means
that retention decreases with the water content).

In HILIC gradient applications, the water composition is increased during the analysis. The
gradient usually begins with only 210 percent water and can reach values in the order of
60-70% or even above.

As with reversed-phase separations, the solvent used to inject the sample should, as
closely as possible, resemble the strength and type used in the eluent starting
composition, i.e. solute injections should be carried out in the mobile phase or in a weaker
Crawford Scientific www.chromacademy.com 19
solvent (that is, containing less water) to prevent loss of efficiency. An alternative for
samples not very soluble in high concentrations of organic is to use small sample volumes
(<10L).

The use of gradients in HILIC should take into consideration the extended equilibration
times sometimes required to establish the initial gradient composition and equilibrate the
column.[23]

The time required for a non-retained particle to pass through the column is known as the
hold-up time (to).

HILIC separation of peptides on a HILIC reverse phase silica column.

Column: C18, 200 m ID40 cm.


Mobile phase: gradient; A: aqueous solution of 0.2% formic acid, B: Acetonitrile+0.2%
formic acid. Gradient A 10-90% linear gradient in 20 min
Solutes: 1. Leu-Enkefalin; 2. VGSE; 3. EH, 4. DSDPR

Crawford Scientific www.chromacademy.com 20


ERLIC

Electrostatic Repulsion Hydrophilic Interaction Chromatography (ERLIC) is a new mode of


chromatography in which an ion-exchange column is eluted with a predominantly organic
mobile phase. Here a water enriched layer (which hydrates the polar stationary phase)
permits hydrophilic analytes to partition and gain retention (even if they have the same
charge as the stationary phase.).[5,24]

Under ERLIC conditions, analytes and stationary phase functional groups present the
same charge

Under ERLIC conditions ionic analytes and functional groups from the stationary phase
present the same charge, and as a consequence analyte retention is increased by
hydrophilic interactions (partition of ionic molecules in the aqueous phase) and
selectively antagonized (decreased) by electrostatic repulsion with functional groups.

Crawford Scientific www.chromacademy.com 21


Very basic and very hydrophilic species exhibit strong retention in HILIC, and, when
present in a multi-component mixture of less polar analytes, require gradient elution for
separation within a reasonable time frame. ERLIC provides an alternative for dealing with
complex mixtures with components at the extreme of retention (poorly vs. highly retained).

Under HILIC conditions ionic analytes and functional groups from the stationary phase
present affinity by each other, and as a consequence analyte retention is increased by
hydrophilic interactions (partition of molecules in the aqueous phase) and probably by a
normal phase like retention mechanism on the stationary phase.

Crawford Scientific www.chromacademy.com 22


As previously explained, in HILIC mode, the stationary phase is hydrated with a slow
moving layer of water, where hydrophilic ions are:

Preferentially partitioned in the aqueous phase, i.e. their retention is increased by


hydrophilic interactions
Retained on functional groups from stationary phase (retention is increased by a
normal phase like mechanism)

A similar situation takes place under ERLIC conditions, where ionic analytes are:

Preferentially partitioned in the aqueous phase, i.e. their retention is increased by


hydrophilic interactions
Repelled by electrostatic interactions with functional groups (of the same charge)
present on the stationary phase i.e. retention is decreased

Ion Exchange Chromatography

In order to fully explain the HILIC and ERLIC modes of chromatography we first need to
study the simple mechanisms of Ion Exchange HPLC.

This mode of chromatography is based on the different affinities of the analyte ions for
oppositely charged ionic centres in the stationary phase or adsorbed counter-ions in the
hydrophobic stationary phase.[25]

Ion-exchange chromatography uses a stationary phase that possesses electrical charges


on its surface. Ionic groups such as SO3 , CO2 , NH 3 , etc, are incorporated into a resin
or gel.
i

Anion exchange Cation exchange

Crawford Scientific www.chromacademy.com 23


Ionic sample molecules compete with counter ions present in the mobile phase for binding
at such stationary phase charged sites. This process is illustrated opposite.

Ion-exchange chromatography is useful for the separation of ionic or ionisable species


and resolves solutes based on the strength of their ionic interactions with ionic functional
groups on the stationary phase.

Ion-exchange chromatography uses an aqueous mobile phase (usually buffered to a


certain pH) to separate analytes of opposite charge to the stationary phase. The mobile
phase contains counter-ions with the same type of charge as the analyte that actively
compete with analyte ions for ion pair interactions with the stationary phase.

Oversimplified separation mechanism of cationic analytes (K+ in this example) on a cation-


exchange resin column

Crawford Scientific www.chromacademy.com 24


Resin SO3 H K Resin SO3 K H

Continued elution of the column with an acidic eluent (H+) introduces competition of H+,
and K+ for the exchange sites SO3 causing the K+ to move down the column.

Anion-Exchange Mixed Modes of Separation I

Hydrophilic interaction can be superimposed as a mixed mode on an ion-exchange


column in the presence of high concentrations of an organic solvent. The important
feature of this combination is the independence of hydrophilic interaction and electrostatic
effects;[1,3,5] or in other words, with sufficient organic solvent in the mobile phase,
hydrophilic interaction dominates the separation. In this way, the behaviour of charged
solutes can be controlled to address extremes of retention. For anion-exchange
chromatography, three important situations can be considered:

Samples with limited solubility in water or highly aqueous mobile phases


Samples that contain very polar compounds that are not retained adequately in
reversed phase
Hydrophilic water-soluble analytes, which are intractable to reversed-phase and/or
ion-exchange chromatography

The first situation is the late elution of acidic analytes on an anion-exchange column; in
this case, the pH of the eluent system can be used to control analyte retention.

Lets consider a reduction of the mobile phase pH below the pKa of certain analyte of
interest; in this case, the analyte will remain as a neutral compound and its interaction with
the stationary phase is minimised and as a consequence, its retention is reduced. As
expected, if the pH is raised above the pKa, then the analyte will remain in the anionic
state, the electrostatic interaction with the stationary phase will be maximised and its
retention will increase.

Crawford Scientific www.chromacademy.com 25


i

Anion-Exchange Mixed Modes of Separation II

The second important case of mixed mode of separation in anion-exchange


chromatography is related to the strong retention of basic analytes.

As previously described,[1,3] with sufficient organic solvent in the mobile phase, hydrophilic
interaction dominates the chromatographic separation even under anion-exchange
conditions.

The positively charged basic analytes partition into the aqueous phase and despite the
electrostatic repulsion with functional groups (also positively charged) present on the
stationary phase, analytes are strongly retained.

Crawford Scientific www.chromacademy.com 26


This is the same mechanism that accounts for analyte retention under HILIC conditions;
however, note that hydrophilic stationary phases for HILIC are neutral and not positively
charged that is the case in anion-exchange chromatography.

Note how analytes ( R NH3 in the present case) presenting the same polarity than the
stationary phase and are precluded from accessing certain regions within the stationary
phase, from which their eluting time is reduced compared with other species. In some
cases this reduction is so dramatic than their eluting time become shorter than the void
volume (to).

Crawford Scientific www.chromacademy.com 27


Anion-Exchange Mixed Modes of Separation III

The third important case of mixed mode of separation in anion-exchange chromatography


is related to the poor retention (prior to the void volume) of basic analytes.

As previously described,[1,3] with sufficient organic solvent in the mobile phase, hydrophilic
interaction dominates the chromatographic separation even under anion-exchange
conditions; however, when the aqueous content exceeds a certain value (application
dependent), the electrostatic repulsion between the analytes and positively charged
functional groups (from the stationary phase) become important. HILIC retention is
counteracted by electrostatic repulsion.

In this case, the two superimposed modes antagonize each other's extremes of retention
and can permit isocratic elution of complex mixtures of analytes (as mixtures of peptides).

This last case forms the basis for Electrostatic Repulsion Hydrophilic Interaction
Chromatography (ERLIC) of analytes.

In the example opposite, vasopressin has a net positive charge, is repelled


electrostatically, and is eluted before the void volume since it is excluded from the pore
system (note that neutral solvent molecules are not affected by the positively charged
stationary phase, they are not precluded from reaching any site within the stationary
phase).

ERLIC separation of selected peptides.

Mobile phase: 10.0 mM sodium methylphosphonate (pH=2) plus 10%Acetonitrile;


Column: anion-exchange.

In the present example, the determination of vasopressin was performed under ERLIC
conditions and vasopressin is the only compound that elutes before the hold-up time.

As far as all peaks within the chromatogram (even the ones eluting before the hold-up
time) are well resolved and well shaped then the separation is OK.

Crawford Scientific www.chromacademy.com 28


ERLIC Vs. HILIC

Under anion exchange conditions certain analytes can elute before the void volume.

In the example shown opposite, vasopressin (a polypeptide) holding a net positive charge,
is repelled electrostatically, and is eluted before the void volume since and is excluded
from the pore system.

The comparative performance of HILIC and ERLIC to resolve a mixture of peptide


standards is also shown. In HILIC mode, a concentration of acetonitrile that promotes
retention of the basic peptides 15 and 17 affords inadequate retention of acidic and
neutral peptides. In the ERLIC mode, electrostatic repulsion of peptides 15 and 17 allows
the acetonitrile concentration to be increased to a level where all peptides can be retained
and well resolved.[1]

Very basic and very hydrophilic species exhibit strong retention in HILIC, and, when
present in a multi-component mixture of less polar analytes, require gradient elution for
separation within a reasonable time frame. The technique of ERLIC provides an
alternative strategy for dealing with complex mixtures with components at the extreme of
retention (poorly vs. highly retained).

HILIC vs ERLIC separation of selected peptides

Crawford Scientific www.chromacademy.com 29


Where:

Tryptic Peptides Acidic Peptides Basic Peptides


(1) Thr-Tyr-Ser-Lys (7) Asp-Val (14) ACTH
(2) Asp-Leu-Trp-Gln-Lys (8) Val-Asp (15) Arg-Lys-Arg-Ser-Arg-
(3) Tyr-Gly-Gly-Phe-Leu-Arg (9) Glu-Ala-Glu Lys-Glu
(4) Leu-Val-Val-Tyr-Pro-Trp- (10) Asp-Ala-Asp-Glu-(pTyr)-Leu (16) Lys-Arg-Gln-His-Pro-
Thr-Gln-Arg (11) Trp-Ala-Gly-Gly-Asp-Ala- Gly-Lys-Arg
(5) Phe-Ser-Trp-Gly-Ala-Glu- Ser-Gly-Glu (17) Lys-Pro-Val-Gly-Lys-
5
Gly-Gln-Arg (12) [isoAsp ]-DSIP Lys-Arg-Arg-Pro-Val-Lys-Val-
5
(6) Val-Gln-Gly-Glu-Glu-Ser- (13) [phosphoSer ]-DSIP. Tyr-Pro
Asn-Asp-Lys

References

1. Tim Wehr. Electrostatic Repulsion Hydrophilic Interaction Chromatography LCGC


North America. Mar 1, 2009
2. David V. McCalley. Hydrophilic Interaction Chromatography. April 1, 2008.
Chromatography Online.com The Global Website of LC/GC
3. Andrew J. Alpert. Hydrophilic-interaction chromatography for the separation of
peptides, nucleic acids and other polar compounds Journal of Chromatography A,
Volume 499, 19 January 1990, Pages 177-196
4. Normal Phase (absorption) Chromatography. Theory of HPLC, from the HPLC
Channel.
5. Eric S. Grumbach, Diane M. Wagrowski-Diehl, Jeffrey R. Mazzeo, Bonnie Alden, and
Pamela C. Iraneta. Hydrophilic Interaction Chromatography Using Silica Columns for the
Retention of Polar Analytes and Enhanced ESI-MS Sensitivity LCGC NORTH AMERICA
VOLUME 22 NUMBER 10 OCTOBER 2004
6. Eugene P. Kadar , Chad E. Wujcik, David P. Wolford, Olga Kavetskaia. Rapid
determination of the applicability of hydrophilic interaction chromatography utilizing ACD
Labs Log D Suite: A bioanalytical application Journal of Chromatography B, 863 (2008)
18
7. Tatsunari Yoshida. Peptide separation by Hydrophilic-Interaction Chromatography: a
review Journal of Biochemical and Biophysical Methods. 60 (2004) 265280
8. X. Esparza, E. Moyano, M.T. Galceran. Analysis of chlormequat and mepiquat by
hydrophilic interaction chromatography coupled to tandem mass spectrometry in food
samples Journal of Chromatography A, 1216 (2009) 44024406
9. Wilson Z. Shou, Weng Naidong. Simple means to alleviate sensitivity loss by
trifluoroacetic acid (TFA) mobile phases in the hydrophilic interaction chromatography
electrospray tandem mass spectrometric (HILICESI/MS/MS) bioanalysis of basic
compounds Journal of Chromatography B, 825 (2005) 186192
10. Takahiro Doi, Keiji Kajimura, Satoshi Takatori, Naoki Fukui, Shuzo Taguchi, Shozo
Iwagami. Simultaneous measurement of diazolidinyl urea, urea, and allantoin in
cosmetic samples by hydrophilic interaction chromatography Journal of Chromatography
B, 877 (2009) 10051010
11. Feng Qin, Yuan Yuan Zhao, Michael B. Sawyer, Xing-Fang Li. Column-switching
reversed phasehydrophilic interaction liquid chromatography/tandem mass spectrometry
method for determination of free estrogens and their conjugates in river water analytica
chimica acta 627 (2008) 9198

Crawford Scientific www.chromacademy.com 30


12. Akihiro Tai, Eiichi Gohda. Determination of ascorbic acid and its related compounds
in foods and beverages by hydrophilic interaction liquid chromatography Journal of
Chromatography B, 853 (2007) 214220
13. Carmela DellAversano, Philipp Hess, Michael A. Quilliam. Hydrophilic interaction
liquid chromatographymass spectrometry for the analysis of paralytic shellfish poisoning
(PSP) toxins Journal of Chromatography A, 1081 (2005) 190201
14. Yanping Lin, Duanyun Si, Zongpeng Zhang, Changxiao Liu. An integrated
metabonomic method for profiling of metabolic changes in carbon tetrachloride induced
rat urine Toxicology 256 (2009) 191200
15. Ruiping Li, Junxiong Huang. Chromatographic behavior of epirubicin and its
analogues on high-purity silica in hydrophilic interaction chromatography Journal of
Chromatography A, 1041 (2004) 163169
16. Todd M. Baughman, Wayne L. Wright, Kathryn A. Hutton. Determination of zanamivir
in rat and monkey plasma by positive ion hydrophilic interaction chromatography
(HILIC)/tandem mass spectrometry Journal of Chromatography B, 852 (2007) 505511
17. Cosima D. Calvano, Carlo G. Zambonin, Ole N. Jensen. Assessment of lectin and
HILIC based enrichment protocols for characterization of serum glycoproteins by mass
spectrometry Journal of Proteomics. 71 (2008) 304 317
18. Petrus Hemstrm. Hydrophilic Separation Materials for Liquid Chromatography
ISBN 978-91-7264-406-9. Print och Media : 2003591 Ume University, UME. 2007
19. A PRACTICAL GUIDE TO HILIC. A Tutorial and Application Book ISBN 978-91-
631-8370-6. http://www.sequant.com/hilicguide
20. Min Liu, Emily X. Chen, Ruthie Ji, David Semin. Stability-indicating hydrophilic
interaction liquid chromatography method for highly polar and basic compounds Journal
of Chromatography A, 1188 (2008) 255263
21. ZIC-pHILIC HYDROPHILIC POLYMER PHASE FOR LIQUID
CHROMATOGRAPHY http://www.sequant.com/
22. LLOYD R. SNYDER and JOHN W. DOLAN. HIGH-PERFORMANCE GRADIENT
ELUTION. The Practical Application of the Linear-Solvent-Strength Model Pp 361-364.
Copyright 2007 by John Wiley & Sons, Inc. All rights reserved
23. J.C.Valette, C. Demesmay, J. L. Rocca, E. Verdon. Separation of Tetracycline
Antibiotics by Hydrophilic Interaction Chromatography Using an Amino-Propyl Stationary
Phase Chromatographia 59, January (No. 1/2), Pp 55-60. 2004.
24. Andrew J. Alpert. Electrostatic Repulsion Hydrophilic Interaction Chromatography for
Isocratic Separation of Charged Solutes and Selective Isolation of Phosphopeptides
Analytical Chemistry 2008, 80, 62-76
25. Ion Exchange Chromatography. Principles and Methods Amersham Biosciences.
ISBN 91 970490-3-4