VII-210 SCOPE (a) Casting Areas Prior to Finish Machining.

TheIQI
Metallic castings, due to their inherent complex shall be based on a thickness that does not
configurations, present examination conditions exceed the finished thickness by more than 20%
that are unique or
to this product form. 1
Radiographic examination may be performed on /4in. (6 mm),
castings when the modified provisions to Article whichever is greater. In no case shall an IQI size
2,asindicated herein, are satisfied. be based
VII-220 GENERAL REQUIREMENTS on a thickness greater than the thickness being
VII-224 SYSTEM OF IDENTIFICATION radiographed.
A system shall be used to produce permanent (b) Casting Areas That Will Remain in the As-Cast
identification on the radiograph traceable to the Condition. The IQI shall be based on the thickness
contract, component, or part numbers, as being
appropriate. In addition, radiographed.
each film of a casting being radiographed shall be VII-280 EVALUATION
plainly VII-282 RADIOGRAPHIC DENSITY
and permanently identified with the name or VII-282.1 Density Limitations.The transmitted film
symbol of density through the radiographic image of the
the Material Manufacturer, Certificate Holder, or body of
Subcontractor, job or heat number, date, and, if the appropriate hole IQI or adjacent to the
applicable, repairs (R1, R2, etc.). This designated
identification system does not wire of a wire IQI and the area of interest shall be
necessarily require that the information appear 1.5
as radiographic images. In any case, this minimum for single film viewing. For composite
information shall not obscure the area of interest. viewing
VII-270 EXAMINATION of multiple film exposures, each film of the
VII-271 RADIOGRAPHIC TECHNIQUE composite
VII-271.2 Double-Wall Viewing Technique.A set shall have a minimum density of 1.0. The
double-wall viewing technique may be used for maximum
cylindrical castings 3 density shall be 4.0 for either single or composite
1 viewing.
/2in. (88 mm) or less in O.D. or when the A tolerance of 0.05 in density is allowed for
shape of a casting precludes single-wall viewing. variations between densitometer readings.
VII-276 IQI SELECTION VII-290 DOCUMENTATION
VII-276.3 Additional IQI Selection Requirements. VII-293 LAYOUT DETAILS
The thickness on which the IQI is based is the 8
single-wall To assure that all castings are radiographed
thickness. consistently in the same manner, layout details
shall be provided. As a minimum, the layout
details shall include:
(a)sketches of the casting, in as many views as
necessary, to show the approximate position of
each location
marker; and
(b)source angles if not perpendicular to the film
15 VIII-230 EQUIPMENT AND MATERIALS
15 VIII-231 PHOSPHOR IMAGING PLATE
MANDATORY APPENDIX VIII VIII-231.1 Selection.Radiography shall be
RADIOGRAPHY USING PHOSPHOR IMAGING performed
PLATE using an industrial phosphor imaging plate
VIII-210 SCOPE capable of demonstrating IQI image
This Appendix provides requirements for using requirements.
phosphor imaging plate (photostimulable VIII-231.2 Processing.The system used for
luminescent phosphor) as an alternative to film processing a phosphor imaging plate shall be
radiography. capable of acquiring, storing, and displaying the
Radiography using phosphor imaging plate may digital image.
be performed on materials including castings and VIII-234 FACILITIES FOR VIEWING OF
weldments RADIOGRAPHS
when the modified provisions toArticle 2as Viewing facilities shall provide subdued
indicated background
herein and all other requirements ofArticle 2are lighting of an intensity that will not cause
satisfied. The term film,asusedwithinArticle 2, reflections, shadows, or glare on the monitor that
applicable interfere with the interpretation process.
to performing radiography in accordance with VIII-260 CALIBRATION
this VIII-262 DENSITOMETER AND STEP WEDGE
Appendix, refers to phosphor imaging plate. COMPARISON FILM
ASTM The requirements ofT-262are not applicable to
E2007,Standard Guide for Computed phosphor imaging plate radiography.
Radiography,may VIII-270 EXAMINATION
be used as a guide for general tutorial VIII-277 USE OF IQIS TO MONITOR
information regarding the fundamental and RADIOGRAPHIC EXAMINATION
physical principles of computed VIII-277.1 Placement of IQIs.
radiography (CR), including some of the (a) Source-Side IQI(s). When using separate
limitations of the blocks for
process. IQI placement as described inT-
VIII-220 GENERAL REQUIREMENTS 277.1(a),thethickness
VIII-221 PROCEDURE REQUIREMENTS of the blocks shall be such that the image
VIII-221.1 Written Procedure.A written procedure brightness at
is the body of the IQI is judged to be equal to or
required. In lieu of the requirements ofT-221.1, greater than
each procedure shall include at least the the image brightness at the area of interest for a
following information, as negative
applicable: image format. If verified by measurement, pixel
(a)material type and thickness range intensity
(b)isotope or maximum X-ray voltage used variations up to 2% are permitted in the
(c)minimum source-to-object distance (Din T- determination of
274.1) equal to.This image brightness requirement is
(d)distance from source side of object to the reversed
phosphor for a positive image format.
imaging plate (din T-274.1) (b)All other requirements ofT-277.1shall apply.
(e)source size (Fin T-274.1) VIII-277.2 Number of IQIs.
(f) phosphor imaging plate manufacturer and (a) Multiple IQIs. An IQI shall be used for each
designation applicable thickness range inTable T-276spanned
(g)screens used by the
(h)image scanning and processing equipment minimum-to-maximum thickness of the area of
manufacturer and model interest
VIII-221.2 Procedure to be radiographed
Demonstration.Demonstration of image quality
indicator (IQI) image requirements
of the written procedure on production or
technique
radiographs using phosphor imaging plate shall
be
consideredsatisfactoryevidenceofcompliancewith
that
procedure.
VIII-225 MONITORING DENSITY LIMITATIONS
OF RADIOGRAPHS
The requirements ofT-225are not applicable to
phosphor imaging plate radiography

(b)As an alternative to(a)above, a minimum of VIII-284 EXCESSIVE BACKSCATTER


two For a negative image format, the requirements
IQIs representing the minimum and maximum ofT-284
thicknesses of the area of interest may be used, shall apply. For a positive image format, if a dark
provided the image of
requirements ofVIII-288are met. theB,as described inT-223, appears on a lighter
(c)All other requirements ofT-277.2shall apply. background of the image, protection from
(d)Comparators such as digitized film strips, gray backscatter is insufficient and the radiographic
scale image shall be considered
cards, etc., may be used to aid in judging unacceptable. A light image of theBon a darker
displayed image background is not cause for rejection.
brightness. When comparators are used to judge VIII-287 DIMENSIONAL MEASURING
areas VIII-287.1 Measuring Scale Comparator.The
within the image, they need not be calibrated. measuring scale used for interpretation shall be
Pixel intensity values may also be used to capable of
quantify image brightness providing dimensions of the projected image. The
comparisons. measurement scale tool shall be based on one of
VIII-277.3 Shims Under Hole IQIs.For welds with the following:
reinforcement or backing material, a shim of (a)a known dimensional comparator that is
material radiographically similar to the weld placed on
metal and/or backing the cassette prior to exposure
material shall be placed between the part and (b)a known dimensional comparator that is
the IQIs, inscribed
such that the image brightness at the body of the on the imaging plate prior to processing
IQI is (c)a known comparator scale placed on the
judged to be equal to or greater than the image imaging
brightness plate prior to processing
at the area of interest for a negative image VIII-287.2 Alternative Comparator.As an
format. If verified by measurement, pixel alternative
intensity variations up to 2% to a measuring scale comparator, a dimensional
are permitted in the determination ofequal calibration of the measuring function based upon
to.This image brightness requirement is reversed a verifiable
for a positive image format. scanned pixel size may be used.
The shim dimensions shall exceed the IQI VIII-288 INTERPRETATION
dimensions Final radiographic interpretation of the area of
such that the outline of at least three sides of the interest
IQI shall shall be performed within the identified IQI image
be visible in the radiograph. contrast and brightness values or, if multiple IQIs
VIII-280 EVALUATION are used,
VIII-281 SYSTEM-INDUCED ARTIFACTS the overlapping portions of the identified
The digital image shall be free of system-induced contrast and
artifacts in the area of interest that could mask or brightness values for the intervening thickness
be confused with the image of any discontinuity. ranges.
VIII-282 IMAGE BRIGHTNESS The IQI and the area of interest shall be of the
The image brightness through the body of the same image
hole-type format (positive or negative). Additionally, where
IQI or adjacent to the designated wire of the applicable
wire-type IQI, (a)the contrast and brightness range that
shall be judged to be equal to or greater than the demonstrates the required IQI sensitivity shall be
image used for interpretation. When more than one IQI
brightness in the area of interest for a negative is used to qualify
image format. If verified by measurement, pixel multiple thicknesses, the overlapping portions of
intensity variations each
up to 2% are permitted in the determination IQIs determined contrast and brightness range
ofequal to. shall be
This image brightness requirement is reversed for considered valid for interpretation of intervening
a positive image format. Additionally, the thicknesses. Pixel intensity values may also be
requirements of used to quantify
T-282are not applicable to phosphor imaging image brightness comparisons in accordance with
plate VIII-277.1, VIII-277.3, and VIII-282.
radiography. (b)the digital image may be viewed and evaluated
VIII-283 IQI SENSITIVITY in a
VIII-283.1 Required Sensitivity.Radiography shall negative or positive image format.
be performed with a technique of sufficient (c)independent areas of interest of the same
sensitivity image
to display the designated hole-type IQI image and may be displayed and evaluated in differing image
the essential hole, or the essential wire of a wire- formats, provided the IQI and the area of interest
type IQI. The are viewed
radiographs shall also display the IQI identifying and evaluated in the same image format
numbers
and letters. Multiple film technique is not
applicable to
phosphor imaging plate radiography.

VIII-290 DOCUMENTATION (i) distance from source side of object to storage


VIII-291 DIGITAL IMAGING TECHNIQUE phosphor media (din T-274.1)
DOCUMENTATION DETAILS (j) storage phosphor manufacturer and
The organization shall prepare and document the designation
radiographic technique details. As a minimum, (k)image acquisition (digitizing) equipment
the following information shall be provided: manufacturer, model, and serial number
(a)the requirements ofArticle 1, T-190(a) (l) single- or double-wall exposure
(b)identification as required by T-224 (m) single- or double-wall viewing
(c)the dimensional map (if used) of marker (n)procedure identification and revision level
placement (o)imaging software version and revision
in accordance with T-275.3 (p)numerical values of the final image processing
(d)number of exposures parameters, i.e., filters, window (contrast), and
(e)X-ray voltage or isotope used level
(f) source size (Fin T-274.1) (brightness) for each view
(g)base material type and thickness, weld The technique details may be embedded in the
reinforcement thickness, as applicable data file.
(h)source-to-object distance (Din T-274.1) When this is performed, ASTM E1475, Standard
Guide for
Data Fields for Computerized Transfer of Digital
Radiological Test Data, may be used as a guide for
establishing
data fields and information content