Sei sulla pagina 1di 9

Accessed from 10.6.1.

1 by regis22 on Wed Apr 12 00:30:07 EDT 2017


USP 39 General Information / 1120 Raman Spectroscopy 1367

in which Ai is the absorbance for each data point; A is the mean absorbance over the spectral segment; and N is the number of
points per segment.
SPECTRAL REFERENCE LIBRARY is a collection of spectra of known materials for comparison with unknown materials. The term is
commonly used in connection with qualitative methods of spectral analysis (e.g., identification of materials).
SPECULAR (SURFACE) REFLECTANCE is the reflectance of the front surface of the sample.
STANDARD ERROR OF CALIBRATION (SEC) is a measure of the capability of a model to fit reference data. SEC is the standard devia-
tion of the residuals obtained from comparing the known values for each of the calibration samples to the values that are cal-
culated from the calibration. SEC should not be used as an assessment tool for the expected method accuracy (trueness and
precision of prediction) of the predicted value of future samples. The method accuracy should generally be verified by calculat-
ing the standard error of prediction (SEP), using an independent validation set of samples. An accepted method is to mark a
part of the calibration set as the validation set. This set is not fully independent but can be used as an alternative for the deter-
mination of the accuracy.
STANDARD ERROR OF CROSS-VALIDATION (SECV) is the standard deviation calculated using the leave-one-out method. In this
method, one calibration sample is omitted from the calibration, and the difference is found between the value for this sample
calculated from its reference value and the value obtained from the calibration calculated from all the other samples in the set.
This process is repeated for all samples in the set, and the SECV is the standard deviation of the differences calculated for all
the calibration samples. This procedure can also be performed with a group of samples. Instead of leaving the sample out, a
group of samples is left out. The SECV is a measure of the model accuracy that one can expect when measuring future samples
if not enough samples are available for the SEP to be calculated from a completely independent validation set.
STANDARD ERROR OF THE LABORATORY (SEL) is a calculation based on repeated readings of one or more samples to estimate the

General Chapters
precision and/or accuracy of the reference laboratory method, depending on how the data were collected.
STANDARD ERROR OF PREDICTION (SEP) is a measure of model accuracy of an analytical method based on applying a given cali-
bration model to the spectral data from a set of samples different from but similar to those used to calculate the calibration
model. SEP is the standard deviation of the residuals obtained from comparing the values from the reference laboratory to
those from the method under test for the specified samples. SEP provides a measure of the model accuracy expected when
one measures future samples.
SURFACE REFLECTANCE, also known as specular reflection, is that portion of the radiation not interacting with the sample but
simply reflecting back from the sample surface layer (sampleair interface).
TRANSFLECTION is a transmittance measurement technique in which the radiation traverses the sample twice. The second time
occurs after the radiation is reflected from a surface behind the sample.
TRANSMITTANCE is represented by the equation:

T = I/I0 or T = 10A

in which I is the intensity of the radiation transmitted through the sample; I0 is the intensity of the radiant energy incident on
the sample and includes losses due to solvent absorption, refraction, and scattering; and A is the absorbance.

1120 RAMAN SPECTROSCOPY

Change to read:

INTRODUCTION

(CN 1-May-2016) Raman is a vibrational spectroscopic technique and is therefore related to infrared (IR) and near-infrared (NIR)
spectroscopy. The Raman effect itself arises as a result of a change in the polarizability of molecular bonds during a given vibra-
tional mode and is measured as inelastically scattered radiation.
A Raman spectrum is generated by exciting the sample of interest to a virtual state with a monochromatic source, typically a
laser. Light elastically scattered (no change in wavelength) is known as Rayleigh scatter and is not of interest in Raman spec-
trometry, except for marking the laser wavelength. However, if the sample relaxes to a vibrational energy level that differs
from the initial state, the scattered radiation is shifted in energy. This shift is commensurate with the energy difference be-
tween the initial and final vibrational states. This inelastically scattered light is referred to as Raman scatter. Only about one
in 106108 photons incident on the sample undergoes Raman scattering. Thus lasers are employed in Raman spectrometers. If
the Raman-scattered photon is of lower energy, it is referred to as Stokes scattering. If it is of higher energy, it is referred to as
anti-Stokes scattering. In practice, nearly all analytically useful Raman measurements make use of Stokes-shifted Raman scatter.
The appearance of a Raman spectrum is much like an infrared spectrum plotted linearly in absorbance. The intensities, or
the number of Raman photons counted, are plotted against the shifted energies. The x-axis is generally labeled Raman
Shift/cm1 or Wavenumber/cm1. The Raman shift is usually expressed in wavenumber and represents the difference in the
absolute wavenumber of the peak and the laser wavenumber. The spectrum is interpreted in the same manner as the corre-

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
1368 1120 Raman Spectroscopy / General Information USP 39

sponding mid-infrared spectrum. The positions of the (Raman shifted) wavenumbers for a given vibrational mode are identical
to the wavenumbers of the corresponding bands in an IR absorption spectrum. However, the stronger peaks in a Raman spec-
trum are often weak in an IR spectrum, and vice versa. Thus the two spectroscopic techniques are often said to be comple-
mentary.
Raman spectroscopy is advantageous because quick and accurate measurements can often be made without destroying the
sample (solid, semisolid, liquid or, less frequently, gas) and with minimal or no sample preparation. The Raman spectrum con-
tains information on fundamental vibrational modes of the sample that can yield both sample and process understanding. The
signal is typically in the visible or NIR range, allowing efficient coupling to fiber optics. This also means that a signal can be
obtained from any medium transparent to the laser light; examples are glass, plastics, or samples in aqueous media. In addi-
tion, because Raman spectra are ordinarily excited with visible or NIR radiation, standard glass/quartz optics may be used.
From an instrumental point of view, modern systems are easy to use, provide fast analysis times (seconds to several minutes),
and are reliable. However, the danger of using high-powered lasers must be recognized, especially when their wavelengths are
in the NIR and, therefore, not visible to the eye. Fiber-optic probes should be used with caution and with reference to appro-
priate government regulations regarding lasers and laser classes.
In addition to normal Raman spectroscopy, there are several more specialized Raman techniques. These include resonance
Raman (RR), surface-enhanced Raman spectroscopy (SERS), Raman optical activity (ROA), coherent anti-Stokes Raman spectro-
scopy (CARS), Raman gain or loss spectroscopy, and hyper-Raman spectroscopy. These techniques are not widely employed in
pharmaceutical laboratories, and are not addressed in this general information chapter.

QUALITATIVE AND QUANTITATIVE RAMAN MEASUREMENTS


General Chapters

There are two general classes of measurements that are commonly performed by Raman spectrometry: qualitative and
quantitative.

Qualitative Raman Measurements

Qualitative Raman measurements yield spectral information about the functional groups that are present in a sample. Be-
cause the Raman spectrum is specific for a given compound, qualitative Raman measurements can be used as a compendial ID
test, as well as for structural elucidation.

Quantitative Raman Measurements

For instruments equipped with a detector that measures optical power (such as Fourier transform [FT]-Raman spectrome-
ters), quantitative Raman measurements utilize the following relationship between signal, Sn, at a given wavenumber, n, and
the concentration of an analyte, C:
Sn = Ksn(nL nb )4P0C

in which K is a constant that depends on laser beam diameter, collection optics, sample volume, and temperature; sn is the
Raman cross section of the particular vibrational mode; nL is the laser wavenumber; nb is the wavenumber of the vibrational
mode; and P0 is the laser power. The Raman cross section, sV, is characteristic of the nature of the particular vibrational mode.
The sample volume is defined by size of the focus of the laser beam at the sample, the optic being used for focusing, and the
optical properties of the sample itself. Spot sizes at the sample can range from less than 1 mm for a microprobe to 6 mm for a
large area sample system. For Raman spectrometers that measure the number of photons per second (such as change-coupled
device [CCD]-Raman spectrometers) the corresponding equation is:
Sn = KsnnL(nL nb )3P0C

From the above equations, it is apparent that peak signal is directly proportional to concentration. It is this relationship that is
the basis for the majority of quantitative Raman applications.

FACTORS AFFECTING QUANTIFICATION

Sample-Based Factors

The most important sample-based factors that deleteriously affect quantitative Raman spectrometry are fluorescence, sample
heating, absorption by the matrix or the sample itself, and the effect of polarization. If the sample matrix includes fluorescent
compounds, the measured signal will usually contain a contribution from fluorescence. Fluorescence will be observed only if
the laser excitation wavelength overlaps with an absorption band of a fluorescent compound. Fluorescence is typically ob-
served as a broad sloping background underlying the Raman spectrum. Fluorescence can cause both a baseline offset and re-
duced signal-to-noise ratio. The wavelength range and intensity of the fluorescence is dependent on the chemical composition
of the fluorescent material. Because fluorescence is generally a much more efficient process than Raman scattering, even very

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
USP 39 General Information / 1120 Raman Spectroscopy 1369

minor amounts of fluorescent impurities can lead to significant degradation of the Raman signal. Fluorescence can be reduced
by using longer wavelength excitation sources such as 785 nm or 1064 nm. However, it should be remembered that the
strength of the Raman signal is proportional to (nL nb )4, so the advantage of using a long-wavelength excitation laser to
minimize fluorescence is at least partially offset by the reduced strength of the Raman signal. The greatest signal-to-noise ratio
will be obtained by balancing fluorescence rejection, signal strength, and detector response.
Fluorescence in solids can sometimes be mitigated by exposing the sample to the laser radiation for a period of time before
measurement. This process is called photobleaching, and operates by degrading the highly absorbing species. Photobleaching
is less effective in liquids, where the sample is mobile, or if the amount of fluorescent material is more than a trace.
Sample heating by the laser source can cause a variety of effects, such as physical form change (melting), polymorph con-
version, or sample burning. The chance for sample heating is greatest when the spot size at the sample is the smallest, i.e.,
when a microprobe is being used. This is usually an issue for colored, highly absorbing species, or very small particles that have
low heat transfer. The effects of sample heating are usually observable either as changes in the Raman spectrum over time or
by visual inspection of the sample. Besides decreasing the laser flux, a variety of methods can be employed to diminish laser-
induced heating, such as moving the sample or laser during the measurement or improving the heat transfer from the sample
with thermal contact or liquid immersion.
Absorption of the Raman signal by the matrix or the sample itself can also occur. This problem is more prevalent with long-
wavelength FT-Raman systems where the Raman signal can overlap with an NIR overtone absorption. This effect will be de-
pendent on the optics of the system as well as on the sample presentation. Associated with this effect is variability from scatter-
ing in solids as a result of packing and particle-size differences. The magnitude of all of these effects, however, is typically less
severe than in NIR because of the limited depth of penetration and the relatively narrower wavelength region sampled in Ram-

General Chapters
an spectroscopy.
Finally, it should be recognized that laser radiation is polarized and the Raman spectra of crystalline materials and other ori-
ented samples can differ significantly depending on the way that the sample is mounted. If the Raman spectrometer is capable
of producing linearly polarized radiation at the sample then a polarization scrambler is recommended for routine sample analy-
sis.

Sampling Factors

Raman spectroscopy is a zero-background technique, in that the signal at the detector is expected to be zero in the absence
of a sample. This situation can be contrasted with absorption spectrometry, where the signal at the detector is at a maximum
in the absence of a sample. Zero-background techniques are inherently sensitive because small changes in sample concentra-
tion lead to proportionate changes in the signal level. The instrument will also be sensitive to other sources of light that can
cause sample-to-sample variations in the measured signal level. In addition, a large background signal caused by fluorescence
will lead to an increased noise level (photon shot noise). Thus it may be very difficult to use the absolute Raman signal for
direct determination of an analyte. Other potential sources of variation are changes in the sample opacity and heterogeneity,
changes in the laser power at the sample, and changes in optical collection geometry or sample position. These effects can be
minimized by sampling in a reproducible, representative manner. Careful design of the instrumentation can reduce these ef-
fects but they cannot be eliminated entirely.
Use of an internal reference standard is the most common and robust method of eliminating variations caused by absolute
intensity fluctuations. There are several choices for this approach. An internal standard can be deliberately added, and isolated
peaks from this standard can be employed; or a band due to a moiety such as an aromatic ring, the Raman cross-section of
which does not change with the way the sample is prepared, can also be used. For solution spectra, an isolated solvent band
can be employed because the solvent will remain relatively unchanged from sample to sample. Also, in a formulation, an exci-
pient peak can be used if it is in substantial excess compared to the analyte. The entire spectrum can also be used as a refer-
ence, with the assumption that laser and sample-orientation changes will affect the entire spectrum equally.
A second important sampling-based factor to consider is spectral contamination. Raman scattering is a weak effect that can
be masked by a number of external sources. Common contamination sources include sample-holder artifacts (container or
substrate) and ambient light. Typically, these issues can be identified and resolved by careful experimentation.

APPARATUS

Components

All modern Raman measurements involve irradiating a sample with a laser, collecting the scattered radiation, rejecting the
Rayleigh-scattered light, differentiating the Raman photons by wavelength, and detecting the resulting Raman spectrum. All
commercial Raman instruments therefore share the following common features to perform these functions:
1. Excitation source (laser)
2. Sampling device
3. Device to filter/reject light scattered at the laser wavelength
4. Wavelength processing unit
5. Detector and electronics

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
1370 1120 Raman Spectroscopy / General Information USP 39

EXCITATION SOURCE (LASER)

Table 1 identifies several common lasers used for pharmaceutical applications or Raman spectrometry. UV lasers have also
been used for specialized applications but have various drawbacks that limit their utility for general analytical measurements.
As more applications for UV lasers are described, it is likely that they may become more common for Raman spectrometry.
Table 1. Lasers Used in Pharmaceutical Applications
Wavelength Range, nm
Laser l, nm (nearest Typical Power (Stokes Region, 100 cm1
whole number) Type at Laser to 3000 cm1 shift) Comments
NIR Lasers
1064 Solid state Up to 3 W 10751563 Commonly used in Fourier transform in-
(Nd:YAG) struments
830 Diode Up to 300 mW 827980 Typically limited to 2000 cm1; Raman shift
because of CCD spectral response; less
common than the other lasers
785 Diode Up to 500 mW 7911027 Most widely used dispersive
Raman laser
Visible Lasers
632.8 HeNe Up to 500 mW 637781 Relatively small fluorescence risk
532 Doubled Up to 1 W 535632.8 High fluorescence risk
General Chapters

(Nd:YAG)
514.5 Ar+ Up to 1 W 517608 High fluorescence risk
488632.8 Ar+ Up to 1 W 490572 High fluorescence risk

SAMPLING DEVICE

Several sampling arrangements are possible, including direct optical interfaces, microscopes, fiber optic-based probes (either
noncontact or immersion optics), and sample chambers (including specialty sample holders and automated sample changers).
The sampling optics can also be designed to obtain the polarization-dependent Raman spectrum, which often contains addi-
tional information. Selection of the sampling device will often be dictated by the analyte and sample. However, considerations
such as sampling volume, speed of the measurement, laser safety, and reproducibility of sample presentation should be evalu-
ated to optimize the sampling device for any given application.

FILTERING DEVICE

The intensity of scattered light at the laser wavelength (Rayleigh) is many orders of magnitude greater than the Raman sig-
nal and must be rejected prior to the detector. Notch filters are almost universally used for this purpose and provide excellent
rejection and stability combined with small size. The traditional use of multistage monochromators for this purpose, although
still viable, is now rare. In addition, various filters or physical barriers to shield the sample from external radiation sources (e.g.,
room lights, laser plasma lines) may be required depending on the collection geometry of the instrument.

WAVELENGTH PROCESSING UNIT

The wavelength scale may be encoded by either a scanning monochromator, a grating polychromator (in CCD-Raman
spectrometers) or a two-beam interferometer (in FT-Raman spectrometers). A discussion of the specific benefits and drawbacks
of each of the dispersive designs compared to the FT instrument is beyond the scope of this chapter. Any properly qualified
instruments should be suitable for qualitative measurements. However, care must be taken when selecting an instrument for
quantitative measurements, as dispersion and response linearity might not be uniform across the full spectral range.

DETECTOR

The silicon-based CCD array is the most common detector for dispersive instruments. The cooled array detector allows
measurements over the spectral range from 4500 to 100 cm1 Raman shift with low noise when most visible lasers, such as
frequency-doubled neodymium-doped yttriumaluminumgarnet (Nd:YAG) (532 nm) or heliumneon (632.8 nm) lasers, are
used. When a 785-nm diode laser is used, the wavelength range is reduced to about 3100 to 100 cm1. The most commonly
used CCD has its peak wavelength responsivity when matched to the commonly used 632.8-nm HeNe gas laser or 785-nm
diode laser. FT instruments typically use single-channel germanium or indiumgalliumarsenide (InGaAs) detectors responsive
in the NIR to match the 1064-nm excitation of a Nd:YAG laser.

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
USP 39 General Information / 1120 Raman Spectroscopy 1371

Calibration

Raman instrument calibration involves three components: primary wavelength (x-axis), laser wavelength, and intensity (y-
axis).

PRIMARY WAVELENGTH (X-AXIS)

In the case of FT-Raman instruments, primary wavelength-axis calibration is maintained, at least to a first approximation,
with an internal HeNe laser. Most dispersive instruments utilize atomic emission lamps for primary wavelength-axis calibra-
tion. In all instruments suitable for analytical Raman measurements, the vendor will offer a procedure of x-axis calibration that
can be performed by the user. For dispersive Raman instruments, a calibration based on multiple atomic emission lines is pre-
ferred. The validity of this calibration approach can be verified subsequent to laser wavelength calibration by using a suitable
Raman shift standard. For scanning dispersive instruments, calibration might need to be performed more frequently, and pre-
cision in both a scanning and static operation mode may need to be verified.1

LASER WAVELENGTH

Laser wavelength variation can impact both the wavelength precision and the photometric (signal) precision of a given in-
strument. Even the most stable current lasers can vary slightly in their measured wavelength output. The laser wavelength
must therefore be confirmed to ensure that the Raman shift positions are accurate for both FT-Raman or dispersive Raman in-

General Chapters
struments. A reference Raman shift standard material such as those outlined in ASTM E1840-96 (2002)1 or other suitably veri-
fied materials can be utilized for this purpose. [NOTEReliable Raman shift standard values for frequently used liquid and solid
reagents, required for wavenumber calibration of Raman spectrometers, are provided in the ASTM Standard Guide cited.
These values can be used in addition to the highly accurate and precise low-pressure arc lamp emission lines that are also avail-
able for use in Raman instrument calibration.] Spectrometric grade material can be purchased from appropriate suppliers for
this use. Certain instruments may use an internal Raman standard separate from the primary optical path. External calibration
devices exactly reproduce the optical path taken by the scattered radiation. [NOTEWhen chemical standards are used, care
must be taken to avoid contamination and to confirm standard stability.]
Unless the instrument is of a continuous calibration type, the primary wavelength axis calibration should be performed, as
per vendor procedures, just prior to measuring the laser wavelength. For external calibration, the Raman shift standard should
be placed at the sample location and measured using appropriate acquisition parameters. The peak center of a strong, well-
resolved band in the spectral region of interest should be evaluated. The position can be assessed manually or with a suitable,
valid peak-picking algorithm. The software provided by the vendor might measure the laser wavelength and adjust the laser
wavelength appropriately so that this peak is at the proper position. If the vendor does not provide this functionality, the laser
wavelength should be adjusted manually. Depending on the type of laser, the laser wavelength can vary with temperature,
current, and voltage. Wavelength tolerances can vary depending on the specific application.

SIGNAL LEVEL (Y-AXIS)

Calibration of the photometric axis can be critical for successful quantification by using certain analytical methods (chemo-
metrics) and method transfer between instruments. Both FT-Raman and dispersive Raman spectrometers should undergo simi-
lar calibration procedures. The tolerance of photometric precision acceptable for a given measurement should be assessed dur-
ing the method development stage.
To calibrate the photometric response of a Raman instrument, a broad-band emission source should be used. There are two
accepted methods. Method A utilizes a tungsten white light source.2 The output power of such sources is traceable to the Na-
tional Metrology Institute (NMI). In the United Kingdom, the National Physical Laboratory also provides calibrated light bulbs.
Several other vendors also provide NIST-traceable irradiance calibration standards. This method is applicable to all common
laser excitation wavelengths listed in Table 1. In Method B, NIST standard reference materials (SRMs) are utilized.3 Several do-
ped-glass fluorescence standards are currently available.
Method AThe source should be placed at the sample location with the laser off and the response of the detector meas-
ured (using parameters appropriate for the instrument). The output for the source used for calibration should be known. The
ratio of the measured response to the true response should be determined and a correction file generated. This correction
should be applied to all spectra acquired with the instrument. Most manufacturers will provide both appropriate calibration
sources and software for this approach. If the manufacturer does not provide a procedure or method, the user can accomplish

1 ASTM E1840-96 (2002) Standard Guide for Raman Shift Standards for Spectrometer Calibration, ASTM International, 100 Barr Harbor Drive, PO Box C700, West

Conshohocken, PA, USA 19428-2959.


2 NIST-traceable tungsten white light source statement: While the calibration of the Raman frequency (or Raman shift, cm1) axis using pure materials and an

existing ASTM standard is well accepted, techniques for calibration of the Raman intensity axis are not. Intensity calibrations of Raman spectra can be accomplish-
ed with certified white light sources.
3 NIST SRM 2241: Ray KG, McCreery RL. Raman intensity correction standard for systems operating with 785-nm excitation. Appl. Spectrosc. 1997, 51, 108116.

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
1372 1120 Raman Spectroscopy / General Information USP 39

the task using a source obtained from NIST and appropriate software. If a manufacturer's method is used, attention must be
paid to the calibration procedure and source validity. The user should obtain appropriate documentation from the manufac-
turer to ensure a qualified approach.
Method BThe fluorescence standard should be placed at the sample location. With the laser on, a spectrum of the SRM
should be obtained (using parameters appropriate for the instrument). The output of the source used for calibration should be
known. The ratio of the measured response to the true response should be determined and a correction file generated. This
correction should be applied to all spectra acquired with the instrument. Most manufacturers will provide both appropriate
calibration sources and software for this approach. If the manufacturer does not provide a procedure or method, the user can
accomplish the task using a source obtained from NIST and appropriate software. If a manufacturer's method is used, attention
must be paid to the calibration procedure and source validity. The user should obtain appropriate documentation from the
manufacturer to ensure a qualified approach. [NOTEMethod B is currently appropriate for systems with 785-nm (SRM 2241),
532-nm (SRM 2242), and both 514.5-nm and 488-nm (SRM 2243) laser excitation. NIST is currently developing other SRMs
that will be wavelength-specific for 1064-nm (SRM 2244) and 632.8-nm excitation (expected to be available in 2006).]

EXTERNAL CALIBRATION

Detailed functional validation employing external reference standards is recommended to demonstrate instrumental suitabil-
ity for laboratory instruments, even for instruments that possess an internal calibration approach. The use of external reference
standards does not obviate the need for internal quality control procedures; rather, it provides independent documentation of
the fitness of the instrument to perform the specific analysis or purpose. For instruments installed in a process location or in a
General Chapters

reactor where positioning of an external standard routinely is not possible, including those instruments that employ an internal
calibration approach, the relative performance of an internal versus an external calibration approach should be periodically
checked. The purpose of this test is to check for changes in components that might not be included in the internal calibration
method (process lens, fiber-optic probe, etc.), e.g., photometric calibration of the optical system.

QUALIFICATION AND VERIFICATION OF RAMAN SPECTROMETERS

The suitability of a specific instrument for a given method is ensured by a thorough technology-suitability evaluation for the
application; a routine, periodic instrument operational qualification; and the more frequent performance verification (see Defi-
nition of Terms and Symbols). The purpose of the technology-suitability evaluation is to ensure that the technology proposed is
suitable for the intended application. The purpose of the instrument qualification is to ensure that the instrument to be used is
suitable for its intended application and, when requalified periodically, continues to function properly over extended time peri-
ods. When the device is used for a specific qualitative or quantitative analysis, regular performance verifications are made. Be-
cause there are many different approaches to measuring Raman spectra, instrument operational qualification and performance
verification often employ external standards that can be used on any instrument. As with any spectrometric device, a Raman
instrument needs to be qualified for both wavenumber (x-axis and shift from the excitation source) and photometric (signal
axis) precision.
In performance verification, a quality-of-fit to an initial scan or group of scans (often referred to in nonscanning instruments
as an accumulation) included in the instrumental qualification can be employed. In such an analysis, it is assumed that refer-
ence standard spectra collected on a new or a newly repaired, properly operating instrument represent the best available spec-
tra. Comparison of spectra taken over time on identical reference standards (either the original standard or identical new
standards, if stability of the reference standards is a concern) forms the basis for evaluating the long-term stability of a Raman
measurement system.

Frequency of Testing

Instrumental qualification is performed at designated intervals or following a repair or significant optical reconfiguration,
such as the replacement of the laser, the detector or the notch or edge filters. Full instrument requalification might not be
necessary when changing between sampling accessories such as a microprobe, a sample compartment, or a fixed fiber-optic
probe. Performance verification tests may be sufficient in these cases; instrument-specific guidance from the vendor on qualifi-
cation requirements should be followed. Tests include wavelength (x-axis and shift from the excitation source) and photomet-
ric (signal axis) precision. Instrument qualification tests require that specific application-dependent tolerances be met.
Performance verification is carried out on the instrument configured for the analytical measurements and is performed more
frequently than instrument qualification. Performance verification includes measurement of the wavelength uncertainty and in-
tensity-scale precision. Wavelength precision and intensity-scale precision tests may be needed prior to any data collection on
a given day. Performance is verified by matching the current spectra to those collected during the previous instrument qualifi-
cation.

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
USP 39 General Information / 1120 Raman Spectroscopy 1373

Instrument Operational Qualification

It is important to note that the acceptance specifications given in both the Instrument Operational Qualification and Perform-
ance Qualification sections are applicable for general use; specifications for particular instruments and applications can vary de-
pending on the analysis method used and the desired accuracy of the final result. ASTM standard reference materials are also
specified, with the understanding that under some circumstances (specifically remote on-line applications) calibration using
one of these materials may be impractical, and other suitably verified materials can be employed. At this juncture it is impor-
tant to note that specific parameters such as spectrometer noise, limits of detection (LOD), limits of quantification (LOQ), and
acceptable spectral bandwidth for any given application should be included as part of the analytical method development.
Specific values for tests such as spectrometer noise and bandwidth will be dependent on the instrument chosen and the pur-
pose required. As a result, specific instrument tests for these parameters are not dictated in this information chapter.

WAVELENGTH (X-AXIS) ACCURACY

It is important to ensure the accuracy of the wavelength axis via calibration to maintain the integrity of Raman peak posi-
tions. Wavelength calibration of a Raman spectrometer consists of two parts: primary wavelength axis and laser wavelength
calibration. After both the primary wavelength axis and the laser wavelength are calibrated, instrument wavelength uncertain-
ty can be determined. This can be accomplished using a Raman shift standard such as the ASTM shift standards or other suita-
bly verified material. Selection of a standard with bands present across the full Raman spectral range is recommended so that
instrument wavelength uncertainty can be evaluated at multiple locations within the spectrum. The tolerance of wavelength

General Chapters
precision that is required for a given measurement should be assessed during the method-development stage. [NOTEFor
scanning dispersive instruments, calibration might need to be performed more frequently, and precision in both a scanning
and static operation mode may need to be verified.]

PHOTOMETRIC PRECISION

Laser variation in terms of the total emitted photons occurring between two measurements can give rise to changes in the
photometric precision of the instrument. Unfortunately, it is very difficult to separate changes in the photometric response as-
sociated with variations in the total emitted laser photons from the sample- and sampling-induced perturbations. This is one of
the reasons why absolute Raman measurements are strongly discouraged and why the photometric precision specification is
set relatively loosely. The tolerance of photometric precision required for a given measurement should be assessed during the
method-development stage.

PERFORMANCE QUALIFICATION

The objective of performance qualification is to ensure that the instrument is performing within specified limits with respect
to wavelength precision, photometric axis precision, and sensitivity. In certain cases when the instrument has been set up for a
specific measurement (for example, installed in a process reactor), it might no longer be possible or desirable to measure the
wavelength and photometric (signal) qualification reference standards identified above. Provided instrument operational quali-
fication has shown that the equipment is fit for use, a single external performance verification standard can be used to reverify
function on a continuing basis (for example, a routinely used process solvent signal, for both wavelength and photometric
precision, following reactor cleaning). The performance verification standard should match the format of the samples in the
current analysis as closely as possible and use similar spectral acquisition parameters. Quantitative measurements of an external
performance verification standard spectrum check both the wavelength (x-axis and laser wavelength) and the photometric
(signal) precision. Favorable comparison of a series of performance verification spectra demonstrates proper continued opera-
tion of the instrument.

WAVELENGTH PRECISION

The wavelength precision should be measured by collecting data for a single spectrum of the selected Raman shift standard
for a period equal to that used in the photometric consistency test. When appropriate, powdered samples should be repacked
between each set of measurements. Peak positions across the spectral range of interest are used to calculate precision. Per-
formance is verified by matching the current peak positions to those collected during the previous instrument qualification and
should not vary with a standard deviation of more than 0.3 cm1, although this specification can be adjusted according to the
required accuracy of the measurement.

PHOTOMETRIC PRECISION

The photometric precision should be measured by collecting data for a single spectrum of a suitably verified reference stand-
ard material for a specified time. After suitable baseline correction, the areas of a number of bands across the spectral range of

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
1374 1120 Raman Spectroscopy / General Information USP 39

interest should be calculated by means of an appropriate algorithm. The area of the strongest band is set to 1, and all other
envelopes are normalized to this band. Performance is verified by matching the current band areas to the respective areas col-
lected during the previous instrument qualification. The areas should vary by no more than 10%, although this specification
can be adjusted according to the required accuracy of the measurement.

LASER POWER OUTPUT PRECISION AND ACCURACY

This test is applicable only to Raman instruments with automatic, internal laser power meters. Instruments without laser
power measurement should utilize a calibrated laser power meter from a reputable supplier. The laser output should be set to
a representative output, dictated by the requirements of the analytical measurement and the laser power measured. The out-
put should be measured and checked against the output measured at instrument qualification. The power (in milliwatts or
watts) should vary by no more than 25% compared to the qualified level. If the power varies by more than this amount, the
instrument should be serviced (as this variation might indicate, among other things, a gross misalignment of the system or the
onset of failure of the laser).
For instruments with an automatic, internal laser power meter, the accuracy of the values generated from the internal power
meter should be compared to a calibrated external laser power meter at an interval of not more than 12 months. The internal-
ly calculated value should be compared to that generated by the external power meter. Performance is verified by matching
the current value to that generated during the previous instrument qualification. The manufacturer might provide software to
facilitate this analysis. If the instrument design prevents the use of an external power meter, then the supplier should produce
documentation to ensure the quality of the instrument and provide a recommended procedure for the above analysis to be
General Chapters

accomplished during a scheduled service visit.

METHOD VALIDATION

Validation of Raman methods will follow the same protocols described in Validation of Compendial Procedures 1225 in terms
of accuracy, precision, etc. However, several of these criteria are affected by variables specific to Raman spectrometry. Fluores-
cence is the primary variable that can affect the suitability of a method. The presence of fluorescent impurities in samples can
be quite variable and have little effect on the acceptability of a material. The method must be flexible enough to accommo-
date different sampling regimes that may be necessary to minimize the effects of these impurities.
Detector linearity must be confirmed over the range of possible signal levels. Fluorescence might drive both the signal base-
line and the noise higher than that used in the validation, in which case the fluorescence must be decreased, or the method
modified to accommodate the higher fluorescence levels. This is also true for the precision, limit of detection, and limit of
quantification of the method, as increased baseline noise will negatively impact all of these values. Because fluorescence can
also affect quantification caused by baseline shifts, acceptable quantification at different levels of photobleaching, when used,
should also be confirmed.
The impact of the laser on the sample must be determined. Visual inspection of the sample and qualitative inspection of the
Raman spectrum for measurements with differing laser powers and exposure times will confirm that the sample is not being
altered (other than by photobleaching). Specific variables to confirm in the spectrum are shifts in peak position, changes in
peak height and band width, and unexpected changes in background intensity.
Method precision must also encompass sample position. The sample presentation is a critical factor for both solids and liq-
uids, and must be either tightly controlled or accounted for in the calibration model. Sample-position sensitivity can often be
minimized by appropriate sample preparation or sample holder geometry, but will vary from instrument to instrument based
on excitation and collection optical configuration.

DEFINITION OF TERMS AND SYMBOLS

CALIBRATION MODEL is a mathematical expression that relates the response from an analytical instrument to the properties of
samples.
INSTRUMENT BANDPASS (OR RESOLUTION) is a measure of the capability of a spectrometer to separate radiation of similar wave-
lengths.
OPERATIONAL QUALIFICATION is the process by which it is demonstrated and documented that the instrument performs accord-
ing to specifications, and that it can perform the intended task. This process is required following any significant change such
as instrument installation, relocation, major repair, etc.
PERFORMANCE QUALIFICATION is the process of using one or more well-characterized and stable reference materials to verify
consistent instrument performance. Qualification may employ the same or different standards for different performance char-
acteristics.
RAMAN SPECTRA4 are plots of the radiant energy, or number of photons, scattered by the sample through the indirect interac-
tion between the molecular vibrations in the sample and monochromatic radiation of frequency much higher than that of the
vibrations. The abscissa is usually the difference in wavenumber between the incident and scattered radiation.

4 Chalmers, J., Griffiths, P., Eds. Handbook of Vibrational Spectroscopy; John Wiley & Sons, Ltd: New York, 2002.

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by regis22 on Wed Apr 12 00:30:07 EDT 2017
USP 39 General Information / 1121 Nomenclature 1375

(NORMAL) RAMAN SCATTERING4 is the inelastic scattering of radiation that occurs because of changes in the polarizability, of the
relevant bonds during a molecular vibration. Normal Raman spectra are excited by radiation that is not in resonance with elec-
tronic transitions in the sample.
RAMAN WAVENUMBER SHIFT4,

is the wavenumber of the exciting line minus the wavenumber of the scattered radiation. SI unit: m1. Common unit: cm1 =
100 m1.

where b is the differential Raman cross section, is positive for Stokes scattering and negative for anti-Stokes scattering.

1121 NOMENCLATURE

INTRODUCTION

General Chapters
The USP (or NF) titles for monograph articles are legally recognized under the Federal Food, Drug, and Cosmetic Act as the
designations for use in labeling the articles to which they apply.
The value of designating each drug substance by one and only one nonproprietary1 name is important in terms of achieving
simplicity and uniformity in drug nomenclature. In support of the U.S. Adopted Names program (see Mission and Preface in
USPNF), of which the U.S. Pharmacopeial Convention is a co-sponsor, the USP Council of Experts gives consideration to the
adoption of the U.S. Adopted Name, if any, as the official title for any compound that attains compendial recognition.
A compilation of the U.S. Adopted Names (USAN) published from the start of the USAN program in 1961, as well as other
names for drugs, both current and retrospective, is provided in the USP Dictionary of USAN and International Drug Names. This
publication serves as a book of names useful for identifying and distinguishing all kinds of names for drug substances, whether
public, proprietary, chemical, or code-designated names.2
A nonproprietary name of a drug serves numerous and varied purposes. Its principal function is to identify the substance to
which it applies by means of a designation that may be used by the professional and lay public free from the restrictions asso-
ciated with registered trademarks. Teaching in the health sciences requires a common designation, especially for a drug that is
available from several sources or is incorporated into a combination drug product; nonproprietary names facilitate communi-
cation among healthcare providers; nonproprietary names must be used as the titles of the articles recognized by official drug
compendia; a nonproprietary name is essential to the pharmaceutical manufacturer as a means of protecting trademark rights
in the brand name for the article concerned; and, finally, the manufacturer is obligated by federal law to include the establish-
ed nonproprietary name in advertising and labeling.
Under the terms of the Drug Amendments of 1962 to the Federal Food, Drug, and Cosmetic Act, which became law Octo-
ber 10, 1962, the Secretary of Health and Human Services is authorized to designate an official name for any drug wherever
deemed necessary or desirable in the interest of usefulness and simplicity.3
The Commissioner of Food and Drugs and the Secretary of Health and Human Services published in the Federal Register
regulations effective November 26, 1984, which state, in part:
Sec. 299.4 Established names of drugs.
(e) The Food and Drug Administration will not routinely designate official names under section 508 of the act. As a re-
sult, the established name under section 502(e) of the act will ordinarily be either the compendial name of the drug or, if
there is no compendial name, the common and usual name of the drug. Interested persons, in the absence of the desig-
nation by the Food and Drug Administration of an official name, may rely on as the established name for any drug the
current compendial name or the USAN adopted name listed in USAN and the USP Dictionary of Drug Names.4
It will be noted that the monographs on the biologics, which are produced under licenses issued by the Secretary of the
U.S. Department of Health and Human Services, represent a special case. Although efforts continue toward achieving uni-
formity, there may be a difference between the respective title required by federal law and the USP title. Such differences
are fewer than in past revisions of the Pharmacopeia. The USP title, where different from the FDA Center for Biologics
Evaluation and Research title, does not necessarily constitute a synonym for labeling purposes; the conditions of licensing
the biologic concerned require that each such article be designated by the name appearing in the product license issued

1 The term generic has been widely used in place of the more accurate and descriptive term nonproprietary with reference to drug nomenclature.
2 USP Dictionary of USAN and International Drug Names is obtainable on order from U.S. Pharmacopeia, Customer Service Department, 12601 Twinbrook Parkway,
Rockville, MD 20852.
3 F.D.&C. Act, Sec. 508 [358].
4 53 Fed. Reg. 5369 (1988) amending 21 CFR 299.4.

Official from December 1, 2016


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.

Potrebbero piacerti anche