Sei sulla pagina 1di 6

CLINICAL SCIENCE

Comparison of Human Tear Film Osmolarity Measured by


Electrical Impedance and Freezing Point
Depression Techniques
Alan Tomlinson, PhD, DSc, Louise C. McCann, MC Optom, and Edward I. Pearce, PhD

Purpose: Tear hyperosmolarity is diagnostic of dry eye disease


(DED), yet difficulty in measurement has limited its utility; develop-
D ry eye disease (DED), as recently defined by the Dry Eye
Workshop,1 is a multifactorial disease of the tears and
ocular surface that results in symptoms of discomfort, visual
ment of new instruments could facilitate its clinical application. This disturbance and tear film instability, with potential damage to
study compares the new OcuSense TearLab osmometer (OcuSense, the ocular surface. It is accompanied by increased osmolarity
Inc, San Diego, CA), based on electrical impedance lab-on-a-chip of the tear film and inflammation in the ocular surface
nanoliter technology, with the freezing point depression Clifton functional unit.1 Tear osmolarity is a function of the rate of
Osmometer (Clifton Technical Physics, Hartford, NY). tear secretion and elimination from the eye.2 As tear secretion
Methods: Thirty-six subjects were recruited: 15 DED (9 women, decreases in dry eye, osmolarity increases because the volume
6 men age: 41 6 16 years) and 21 controls (12 women, 9 men age: of isotonic tears is inadequate to overcome the loss of fluid,
35 6 12 years); criteria for DED were noninvasive tear breakup time mainly by evaporation.2
,10 seconds, Schirmer I test ,5 mm, and positive symptoms. Attempts have been made to provide a reliable, highly
Samples were collected from the inferior tear meniscus for testing sensitive, specific, and clinically applicable test for the
with both osmometers. measurement of DED. However, because of the multifactorial
etiology of this disease, diagnosis is difficult. Clinically, the
Results: Osmolarity values measured with OcuSense TearLab were most common tests used in the assessment of DED are
308 6 6 and 321 6 16 mOsm/L for controls and dry eye, respectively, Schirmer,36 tear breakup time,1,711 and fluorescein staining.12
and those measured with Clifton were 310 6 7 and 323 6 14 mOsm/L Such measurements are quick and easy to perform, but their
for controls and dry eye, respectively; these values were significantly invasiveness and lack of repeatability (in the case of Schirmer)
different. Significant correlation was found between OcuSense and lead to the need for a more effective measure of dry eye.
Clifton measurements (r = 0.904; P = 0.006). BlandAltman analysis To address this requirement, investigators have studied
revealed agreement between techniques; the majority of points fell numerous tear physiology measures in the detection of DED;
within the 95% confidence limits, and actual values differed by less parameters such as evaporation,1316 tear production,1720 and
than 1%. A cutoff value of .316 mOsm/L, derived from the osmolarity2,21,22 have been employed. However, the need for
distribution of osmolarity values, was used to diagnose DED with an high levels of expertise, expense, and length of time required
effectiveness of 73% sensitivity, 90% specificity, and 85% positive render these tests impractical in a clinical setting.
predictive value for the OcuSense and 73% sensitivity, 71% specificity, Tear hyperosmolarity has been shown to be diagnostic of
and 65% positive predictive value for the Clifton in the study samples. DED1,2,11,2124 and the most effective single measure.22
Hyperosmolarity has been shown to be a primary cause of
Conclusions: Tear film osmolarity measured with the OcuSense
ocular surface damage and inflammation. Early studies of
TearLab system correlates well with the Clifton Osmometer. The
rabbits found osmolarity to be a function of evaporation and
new instrument has the potential to provide clinicians with a readily
tear production.2,25 However, the difficulty in measuring
available clinically applicable measure, which could become the gold
osmolarity has led to its limited utility in a clinical context.
standard in DED.
The osmolarity of a sample can be determined in
Key Words: tear film, dry eye, osmolarity a number of ways, both in situ and by sampling, using methods
that measure the colligative properties of the tears. These
(Cornea 2010;29:10361041)
properties, such as freezing point depression and vapor
pressure, depend on the number of dissolved particles in
a solution but are not dependent on the identity of the particles.
Received for publication July 31, 2009; revision received October 8, 2009; The freezing point depression nanoliter osmometer is at
accepted November 1, 2009. present the most commonly applied principle in osmolarity
From the Department of Vision Sciences, Glasgow Caledonian University, measurement. In this method, the temperature of the freezing
United Kingdom. point is directly proportional to the total number of dissolved
Funding was received from OcuSense, Inc.
Reprints: Alan Tomlinson, Glasgow Caledonian University, 30 Cowcaddens
particles in the solution. Therefore, the osmolarity can be
Road, Glasgow, G4 0BA, United Kingdom (e-mail: a.tomlinson@gcal.ac.uk). calculated from the depression in the freezing point. The most
Copyright 2010 by Lippincott Williams & Wilkins frequently applied freezing point depression techniques in tear

1036 | www.corneajrnl.com Cornea  Volume 29, Number 9, September 2010


Cornea  Volume 29, Number 9, September 2010 Comparison of Human Tear Film Osmolarity

research use nanoliter samples, most commonly with a Clifton Subject Recruitment
nanoliter osmometer (Clifton Technical Physics of Hartford, Thirty-six volunteers were recruited for this study,
NY).26 Although used in the diagnosis of DED, this method 15 with mild to moderate dry eye (9 women, 6 men; mean age:
requires significant expertise, takes considerable time, and 41.7 6 16.9 years) and 21 controls (12 women, 9 men; mean
is open to potential errors because of evaporation of test age: 35.0 6 12.8 years). Inclusion criteria for DED were
samples.27 Other techniques using the freezing point depres- noninvasive tear breakup time (NITBUT) ,10 seconds using
sion technique such as the Advanced Tear Osmometer the HirCal grid10 and 2 or more positive symptoms using
(Advanced Instruments, Inc, Norwood, MA) and the Otago McMonnies30 questionnaire. Schirmer I test was carried out on
Osmometer (Otago Osmometers, Ltd, Dunedin, New Zealand) all participants to determine contribution of aqueous tear
are also available. deficiency to the dry eye condition (10 of the dry eye patients
Vapor pressure techniques have also been employed in were aqueous deficient with a Schirmer I score of #8 mm in 5
the measurement of osmolarity.28 These work on the principle minutes). The controls were asymptomatic and had a Schirmer
that the vapor pressure of a solution is lower than that of score of $15 mm in 5 minutes and a tear breakup time of .10
the pure solvent at the same temperature and pressure; the seconds. Tear samples were collected from the inferior-
decrease in vapor pressure, like depression of freezing point, temporal conjunctival sac of the right eye of each subject for
is proportional to the number of dissolved particles in the both the TearLab (OcuSense, Inc, San Diego, CA) and Clifton
solution. Thus, the osmolarity of a solute can be calculated (Clifton Technical Physics, Hartford, NY) osmometers. The
from its vapor pressure. Original vapor pressure osmometers order of test measurements was as follows:
engaged a precision thermocouple hygrometer to measure dew  Symptom questionnaire
point depression and required large sample volumes.28 This  Osmolarity (samples by each instrument was randomized
necessitated the collection of reflex tears, which in turn could to avoid bias)
lower the osmolarity values obtained.29 More recently, vapor  NITBUT
pressure osmometers, such as the Wescor (Wescor, Inc, Logan,  Schirmer test
UT) have been used. However, although easier to operate and All measurements were on the right eyes only.
more streamlined than freezing point depression osmometers,
they are still not suitable for the quick easy application Symptoms
required in clinical practice. Although many diagnostic tests are unreliable, the
There is a need for a new instrument to facilitate clinical response of patients to questioning about dry eye symptoms
application and the adoption of osmolarity as a primary has been found to be repeatable.7,3133 The most common
diagnostic test in DED. Recently the OcuSense TearLab questionnaire used clinically is the McMonnies30 dry eye ques-
system (OcuSense, Inc, San Diego, CA) has been developed. tionnaire. If 2 or more subjective responses are positively
This new osmometer is based on electrical impedance and reported, this is indicative of dry eye.30
lab-on-a-chip technology, which allows the calculation of
osmolarity. This technique allows osmolarity testing with Schirmer I Test
a very small volume (less than 20 nL), is a quick and accurate The Schirmer I test is the most widely used test for
measurement of osmolarity of the tear film in a clinical setting, investigating tear production in patients with dry eye. A
and reduces the evaporation of the fluid. However, although Schirmer Test Strip (SNO Strip; Clement Clarke International,
the device measures charged particles, corrections or Ltd, Harlow, Essex, United Kingdom) was placed in the lower
assumptions are made with regards the contribution made outer fornix of the unanesthetized eye, and the patient was
by noncharged particles in the tear sample. The OcuSense advised to close his/her eyes. After 5 minutes, the strip was
TearLab system is in the final stages of approval as a medical removed from the eye and the length of wetting from the notch
device by Food and Drug Administration. was measured (in millimeters).34 A cutoff of 15 mm in 5
This study sets out to compare the new OcuSense minutes for the normal eye was initially suggested by
TearLab osmometer (OTO) with the Clifton Osmometer to Schirmer, with further work indicating values less than 5
determine the comparability of results between the instruments mm in 5 minutes are pathological. We used a cutoff of #8 mm
and the diagnostic efficacy of each test for dry eye. to differentiate dry eye from normal.

NITBUT
The HirCal grid consists of a white on black grid pattern
MATERIALS AND METHODS fitted to a Bausch and Lomb keratometer.10 The subject, seated
at the instrument, was instructed to blink normally and then
Study Design refrain from blinking. The grid image is reflected from the
This study adopted a single visit for comparative surface of the tears, and the first point of disruption of the grid
measurements with the 2 instruments. It was conducted lines after the complete blink was noted; 3 measurements were
according to the principles contained in the Declaration of taken, and the average calculated.10
Helsinki. Ethical approval was obtained from the Glasgow
Caledonian University Ethics Committee. Written informed Clifton Osmometer
consent was obtained from all subjects after explanation of the Osmolarity was measured with the freezing point depres-
procedures and study requirements. sion technique using a Clifton nanoliter osmometer (Clifton

q 2010 Lippincott Williams & Wilkins www.corneajrnl.com | 1037


Tomlinson et al Cornea  Volume 29, Number 9, September 2010

Technical Physics, Hartford, NY). A micro capillary glass tube RESULTS


(Bilbate, Ltd, Northants, United Kingdom) was drawn over
a flame to produce a fine tip. By capillary action, a sample of Statistical Analyses
tears was taken from the marginal tear strip while the patient All of the data were tested for normality using a
was gazing supranasally. This was to avoid reflex tearing.35 The ShapiroWilk test, and found to be normally distributed (SPSS
sample was then injected into a vessel containing heavy oil to Version 11 (SPSS, Inc); a 2-sampled t test was then applied to
avoid evaporation. Samples of nanopure water and 290 standard the data.
solution (Clinitrol 290, Advanced Instruments, Inc) were also
injected (used for calibration during each analysis to control the Osmolarity
variation in environmental conditions). A small sample of each Significant differences were found in this study between
fluid was in turn drawn from the vessels into a capillary tube the mean test values of osmolarity for controls [308 6
attached to a syringe by plastic tubing, supported by a clamp. 6.2 mOsm/L (TearLab); 310 6 7.2 mOsm/L (Clifton)] (P ,
Up to 8 samples were loaded and measured per run. Samples 0.001) and subjects with dry eye [321 6 16.5 mOsm/L
were then loaded into the sample plate and frozen and then (TearLab); 323 6 14.7 mOsm/L (Clifton)] (P , 0.001)
thawed, with the process being viewed through a stereomicro- (Fig. 1). Subjects with dry eye had higher values indicating
scope. The point at which the last crystal of each sample melted increased tear osmolarity. The osmolarity with the Clifton
was noted as the freezing point of the sample. This was repeated instrument was generally higher than with the TearLab system,
3 times and an average taken. The osmolarities of the tear although this difference is not significant, either statistically
samples were then calculated. or clinically (r = 0.904; P = 0.006) with the actual values
differing by less than 1% (approximately 2 mOsm/L).
OcuSense TearLab Nanoliter Osmometer A significant correlation between osmolarity measure-
The use of electrical impedance to measure the electrical ments with the TearLab and Clifton systems was found (r =
conductivity of a fluid has been well documented.3640 Fluids 0.904; P = 0.006). A scatter graph was plotted, and a Pearson
can be characterized by their ionic content.41 Any change in correlation was determined. This is shown in Figure 2.
composition and concentration of the ions within a fluid can BlandAltman analysis assessed the level of agreement
change the electrical conductivity, an indirect function of between the results with the OTO and Clifton osmometers
osmolarity.39,42 Previously Fouke et al40 had used thin film (Fig. 3). This analysis showed a high level of agreement with
microelectronic technology to produce a flexible sensor, which only a minimal number of points falling out with the 95%
was placed directly on to the ocular surface to calculate the confidence limits.
electrical conductivity of the fluids present. A similar technique Cutoff values (316 mOsm/L for TearLab; 317.2 mOsm/L
was adopted by Ogasawara et al42 who placed a sensor directly for Clifton) for diagnosis were derived from the intercept of
onto the temporal conjunctival cu-de-sac. Tear fluid conduc- the distribution curves of osmolarity values for dry eyes and
tivity was monitored on a display, and the sodium chloride controls. The effectiveness of the individual measurement
concentration of the tears calculated from a calibration curve techniques in the differential diagnosis of the dry eye from
and converted to the equivalent electrolyte concentration.42 One the normal was evaluated. These cutoff values yielded
obvious benefit of this technique is the in situ measurement of
osmolarity, which reduces the chance of evaporation of the tear
fluid on removal from the eye. However, as the sensor is placed
on the ocular surface, reflex tear lacrimation may ensue thus
altering the osmolarity of the tear sample.
Recent work has been done on a novel osmolarity mea-
suring system, the OTO.26 This osmometer employs a proprie-
tary impedance spectroscopy technique, a similar technique as
used in blood glucose monitors where it is used to characterize
tissues based on the knowledge of their electrical properties
in the frequency spectrum. Studies have shown that such
measures of dielectric properties can indicate the pathological
status of tissues.43
The OTO employs a single-use disposable test card
mounted to a collection pen, to obtain a small sample of
tear fluid by passive capillary action from the inferior-temporal
tear meniscus. The pen monitors the collection process and
provides an audible and visual signal when the sample of
tear has been collected. The measurement of the electrical
impedance is carried out within the pen. The pen is then
docked into the reader, which calculates and displays the
osmolarity result. This technology has recently received 510k
approval by the Food and Drug Administration for marketing
in the United States as an aid in the diagnosis of DED. FIGURE 1. Mean osmolarity values for Clifton and OTOs.

1038 | www.corneajrnl.com q 2010 Lippincott Williams & Wilkins


Cornea  Volume 29, Number 9, September 2010 Comparison of Human Tear Film Osmolarity

unlikely to require any stimulation of reflex tearing and is


therefore suitable for basal measurements of tear film. How-
ever, its expense, need for significant expertise, and consider-
able time required to perform this technique have hindered its
adoption as a suitable clinical technique.
Electrical conductivity of tear samples has been shown
to be beneficial in determining tear osmolarity; previous
techniques measured osmolarity directly on the eye surface.
This decreased the potential for evaporative loss from the tear
samples, allowed measures of tear film osmolarity in real
time, and was relatively unobtrusive.42 However, electrical
conductivity is not a colligative property and measures only
charged particles and is dependent on the number of such
particles. This is unlike the freezing point depression and
vapor pressure techniques. Also with previous electrical con-
ductivity techniques, measurement directly on the eye may
precipitate reflex tearing and thus alter the osmolarity of the
sample. In addition, the required sample size, although small,
was actually dependent on the area of the electrode sensor.42
Some of these problems with electrical conductivity as
a measurement of osmolarity have been addressed by recent
FIGURE 2. Scatter plot showing a significant correlation developments in this area. These include the design of the
between osmolarity measurements with the OcuSense TearLab TearLab system using electrical impedance and lab-on-
and Clifton systems (r = 0.904; P = 0.006). Black line shows 45- a-chip technology.26,44 This technique measures osmolarity
degree line. The dashed line shows the regression line.
immediately after sample acquisition and in real time,
unlike freezing point and other techniques such as vapor
a sensitivity of 73%, a specificity of 90%, and a positive pressure. This is obviously beneficial because it reduces
predictive value of 85% for the TearLab system and for the the evaporation of the tear samplean inherent problem with
Clifton values of 73% sensitivity, 71% specificity, and 65% the other lab-based techniques. Furthermore, this technique
positive predictive value, respectively. requires a very small tear sample (;0.2 mL), unlike the
previous electrical conductivity technique, and eliminates the
DISCUSSION problem of reflex tearing because the micro scale electrode
Osmolarity is measured with an instrument that employs does not come into contact with the ocular surface.26,44
colligative properties such as freezing point depression and is The osmotic pressure of a solution is a colligative
a powerful tool in the differential diagnosis of dry eye.22 property depending solely on the number of dissolved particles
Previous work with the Clifton nanoliter osmometer (Clifton in a solution and not on their size or weight. The electrical
Technical Physics of Hartford, NY) shows that it can be used conductivity is dependent on the number of particles and their
to differentiate dry eye from the normal with 90% accuracy.22 charge; it could be argued that the conductivity technique for
It requires only microliter samples of tear fluid and thus is measuring osmolarity is limited by its inability to recognize all
particles. The freezing point depression method is the only one
that measures concentration based solely on the number
of dissolved particles in a solution, and it can be considered as
the superior technique. Although both techniques seem to be
measuring different properties of the tear sample, they have
been found to be highly correlated in this study, indicating
that electrical conductivity of the TearLab system may be
considered as a valid alternative to freezing point depression
with the Clifton.
The association between Clifton and TearLab is illu-
strated in a scatter plot fitted with a regression line (Fig. 2).
It can be seen that there is a linear relationship between the
data obtained with these 2 tests, with the Clifton generally
reading slightly higher than the TearLab system. However, the
difference was small (less than 2 mOsm/L or 1%) and not
significant (P . 0.001).
FIGURE 3. BlandAltman analysis showing mean value plotted In the literature, the cutoff for osmolarity in dry eye has
against the difference between osmolarity values for Clifton been set between 293 and 320 mOsm/L.21,22,35,45,46 Our
and TearLab osmometers. Dashed lines show 95% confidence findings of a cutoff of 316 mOsm/L for TearLab, 317.2
limits. mOsm/L for Clifton from the intercept of the distribution

q 2010 Lippincott Williams & Wilkins www.corneajrnl.com | 1039


Tomlinson et al Cornea  Volume 29, Number 9, September 2010

curves for dry eye and normals agree with recent studies22 6. Hamano H, Hori M, Hamano T, et al. A new method for measuring tears.
where values of 316 mOsm/L gave a sensitivity of 59% and CLAO J. 1983;9:281289.
7. Nichols KK, Mitchell GL, Zadnik K. The repeatability of clinical
a specificity of 94%.47 measurements of dry eye. Cornea. 2004;23:272285.
In the present study, 2 groups of patients, those with 8. Lemp MA, Hamill JR Jr. Factors affecting tear film breakup in normal
and without DED, were evaluated for tear osmolarity with eyes. Arch Ophthalmol. 1973;89:103105.
2 osmometers. The results for osmolarity in DED and normals 9. Lemp MA. Evaluation and differential diagnosis of keratoconjunctivitis
sicca. J Rheumatol Suppl. 2000;61:1114.
are similar to those reported recently in a meta-analysis22 10. Hirji N, Patel S, Callander M. Human tear film pre-rupture phase time
(averages of 302 6 9.7 mOsm/L for normal and 326 6 22.1 (TP-RPT)a non-invasive technique for evaluating the pre-corneal tear
mOsm/L for dry eye). The means in the present study for film using a novel keratometer mire. Ophthalmic Physiol Opt. 1989;9:
normals were higher at 308 6 6.2 mOsm/L with the TearLab 139142.
and 310 6 7.2 mOsm/L with the Clifton and slightly lower for 11. Lemp MA. Report of the National Eye Institute/Industry workshop on
clinical trials in dry eyes. CLAO J. 1995;21:221232.
subjects with dry eye with means of 321 6 16.5 mOsm/L with 12. Mengher LS, Bron AJ, Tonge SR, et al. Effect of fluorescein instillation on
the TearLab and 323 6 14.7 mOsm/L with the Clifton. The the pre-corneal tear film stability. Curr Eye Res. 1985;4:912.
difference in values is most likely because of smaller and 13. Mathers W. Evaporation from the ocular surface. Exp Eye Res. 2003;78:
different sample sizes in the present study. 389394.
14. Miano F, Calcara M, Giuliano F, et al. Effect of meibomian lipid layer on
On the basis of the present study, the osmolarity values evaporation of tears. J Phys. 2004;16:S2461S2467.
found with the TearLab system seem to correlate well with the 15. Mathers WD. Ocular evaporation in meibomian gland dysfunction and
Clifton (P = 0.006) (Fig. 3). This suggests TearLab is a suitable dry eye. Ophthalmology. 1993;100:347351.
alternative technique for the diagnosis of DED, giving 16. Tsubota K, Yamada M. Tear evaporation from the ocular surface. Invest
a sensitivity of 73%, a specificity of 90%, and a positive Ophthalmol Vis Sci. 1992;33:29422950.
17. Pearce EI, Keenan BP, McRory C. An improved fluorophotometric
predictive value of 85%. This does not reach the level of method for tear turnover assessment. Optom Vis Sci. 2001;78:3036.
effectiveness in diagnosis of some other reports26 in which 18. van Best JA, Benitez del Castillo JM, Coulangeon LM. Measurement of
TearLab had a sensitivity of 94% and a specificity of 84%. This basal tear turnover using a standardized protocol. European concerted
is probably because of differences in the participants in the action on ocular fluorometry. Graefes Arch Clin Exp Ophthalmol. 1995;
233:17.
studies with a different (and greater) spectrum of severity in 19. Macri A, Rolando M, Pflugfelder S. A standardized visual scale for evalu-
the previous study.26 In that study, the mean value of ation of tear fluorescein clearance. Ophthalmology. 2000;107:13381343.
osmolarity for normals was lower (306 mOsm/L compared 20. Macri A, Pflugfelder S. Correlation of the Schirmer 1 and fluorescein
with 308 mOsm/L in the present study) and the subjects with clearance tests with the severity of corneal epithelial and eyelid disease.
dry eye had a higher mean compared with the present study Arch Ophthalmol. 2000;118:16321638.
21. Gilbard JP, Farris RL, Santamaria J II. Osmolarity of tear microvolumes in
(334 mOsm/L compared with 321 mOsm/L). However, it keratoconjunctivitis sicca. Arch Ophthalmol. 1978;96:677681.
could be argued that the present study is still a rigorous 22. Tomlinson A, Khanal S, Ramaesh K, et al. Tear film osmolarity:
assessment of the diagnostic techniques because of the determination of a referent for dry eye diagnosis. Invest Ophthalmol Vis
composition of the participants. The patients with dry eye Sci. 2006;47:43094315.
23. Farris RL, Gilbard JP, Stuchell RN, et al. Diagnostic tests in
in the present study were generally in the mild to moderate keratoconjunctivitis sicca. CLAO J. 1983;9:2328.
region of the diagnostic spectrum. If a technique is sensitive to 24. Farris RL. Tear osmolaritya new gold standard? Adv Exp Med Biol.
such patients, its effectiveness is validated because it is this 1994;350:495503.
group who are often difficult to diagnose. 25. Gilbard JP, Dartt DA. Changes in rabbit lacrimal gland fluid osmolarity
with flow rate. Invest Ophthalmol Vis Sci. 1982;23:804806.
26. Sullivan B. Clinical resorts of a first generation lab-on-chip nanolitre tear
film osmometer. Ocul Surf. 2005;3.
CONCLUSIONS 27. Nelson JD, Wright JC. Tear film osmolality determination: an evaluation
The present study demonstrates that tear film osmolarity of potential errors in measurement. Curr Eye Res. 1986;5:677681.
28. Terry JE, Hill RM. Human tear osmotic pressure: diurnal variations and
measured with the OcuSense TearLab system correlates well the closed eye. Arch Ophthalmol. 1978;96:120122.
with the Clifton osmometer. The new instrument has the 29. White KM, Benjamin WJ, Hill RM. Human basic tear fluid osmolality. I.
potential to provide clinicians with a readily available Importance of sample collection strategy. Acta Ophthalmol (Copenh).
clinically applicable measure, which could become the gold 1993;71:524529.
standard in DED. It is also effective as a single test for the 30. McMonnies CW. Key questions in a dry eye history. J Am Optom Assoc.
1986;57:512517.
discrimination of those with dry eye from those without the 31. Nichols JJ, Ziegler C, Mitchell GL, et al. Self-reported dry eye disease
condition. across refractive modalities. Invest Ophthalmol Vis Sci. 2005;46:
19111914.
32. Nichols KK, Nichols JJ, Mitchell GL. The reliability and validity of
REFERENCES McMonnies dry eye index. Cornea. 2004;23:365371.
1. Foulks GN. DEWS report. Ocul Surf. 2007;5. 33. Nichols KK. Patient reported symptoms in dry eye disease. Ocul Surf.
2. Gilbard JP, Farris RL. Tear osmolarity and ocular surface disease in 2006;4:137145.
keratoconjunctivitis sicca. Arch Ophthalmol. 1979;97:16421646. 34. Tomlinson A, Blades KJ, Pearce EI. What does the phenol red thread test
3. Clinch TE, Benedetto DA, Felberg NT, et al. Schirmers test. A closer actually measure? Optom Vis Sci. 2001;78:142146.
look. Arch Ophthalmol. 1983;101:13831386. 35. Craig JP, Tomlinson A. Effect of age on tear osmolality. Optom Vis Sci.
4. Pinschmidt NW. Evaluation of the Schirmer tear test. South Med J. 1970; 1995;72:713717.
63:1256 passim. 36. Ackmann JJ, Seitz MA. Methods of complex impedance measurements in
5. Danjo Y. Diagnostic usefulness and cutoff value of Schirmers I test in biologic tissue. Crit Rev Biomed Eng. 1984;11:281311.
the Japanese diagnostic criteria of dry eye. Graefes Arch Clin Exp 37. Sheppard NF. Electrical conductivity measurements using microfabricated
Ophthalmol. 1997;235:761766. interdigitated electrodes. Anal Chem. 1993;65:1199.

1040 | www.corneajrnl.com q 2010 Lippincott Williams & Wilkins


Cornea  Volume 29, Number 9, September 2010 Comparison of Human Tear Film Osmolarity

38. Mitsubayashi K. Flexible conductimatric sensor. Anal Chem. 1993; measured by a flexible conductimetric sensor. Graefes Arch Clin Exp
65:3586. Ophthalmol. 1996;234:542546.
39. Mitsubayashi K, Ogasawara K, Yokoyama K, et al. Measurement of tear 43. Han A, Frazier AB. Ion channel characterization using single cell
electrolyte concentration and turnover rate using a flexible conductimetric impedance spectroscopy. Lab Chip. 2006;6:14121414.
sensor. Technol Health Care. 1995;3:117121. 44. Sullivan BD. Tear film osmometry. In: OcuSense. vol B2. The Regents of
40. Fouke JM, Wolin AD, Saunders KG, et al. Sensor for measuring the University of California; 2006.
surface fluid conductivity in vivo. IEEE Trans Biomed Eng. 1988;35: 45. Mathers WD, Lane JA, Zimmerman MB. Tear film changes associated
877881. with normal aging. Cornea. 1996;15:229234.
41. van Haeringen NJ. Clinical biochemistry of tears. Surv Ophthalmol. 1981; 46. Mathers WD, Daley TE. Tear flow and evaporation in patients with and
26:84. without dry eye. Ophthalmology. 1996;103:664669.
42. Ogasawara K, Mitsubayashi K, Tsuru T, et al. Electrical conductivity of 47. Khanal S, Tomlinson A, McFadyen A, et al. Dry eye diagnosis. Invest
tear fluid in healthy persons and keratoconjunctivitis sicca patients Ophthalmol Vis Sci. 2008;49:14071414.

q 2010 Lippincott Williams & Wilkins www.corneajrnl.com | 1041

Potrebbero piacerti anche