Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
The performance of the GEM Premier 3000 with iQM was compared to the
following competitive systems: Bayer Rapidlab 855, Bayer Rapidpoint
405, Nova pHOx plus L, and Radiometer ABL77. The systems were com-
pared for accuracy using available reference materials and methods, and for
the response of the systems to blood samples containing interfering sub-
stances and clots.
This study shows the GEM Premier 3000 to have clinically acceptable accu-
racy when evaluated using available reference materials and methods for
PO2, PCO2, Na+, K+, Ca++ glucose and hematocrit. Overall, the GEM
Premier 3000 demonstrated superior accuracy to the competitive systems for
measurement of these analytes. iQM was shown effective in flagging the
presence of interfering substances in samples and the presence of blood clots
on sensors. In all cases, interfering substances were flagged by iQM, while
in some cases the competitive systems reported unflagged, erroneous
results. The clot-detection and clearing mechanisms of iQM make the GEM
Premier 3000 superior to the competitive systems for handling problems cre-
ated by the presence of blood clots in samples.
The systems were compared for the following: 2. Interference produced by two substances known
1. Accuracy using available to interfere with electrode-based critical care
reference materials and methods: systems: benzalkonium and thiopental,
Na+, K+, Ca++: SRM 956a from the representing a macromolecular cation and
National Institute for Standards and anion, respectively.
Technology (NIST) (certified reference
material) 3. Problems produced by introduction of clots to
glucose: SRM 965 from NIST the systems and evaluation of clot-handling
(certified reference material) mechanisms of the systems.
____________________________
*Instrumention Laboratory, Lexington, MA
Materials and Methods Interfering Substances
Benzalkonium chloride was spiked into whole blood at
All instruments were set up and maintained according to concentrations of 5 and 10 g/mL. Sodium thiopental
the manufacturers instructions. Table 1 is a was spiked into whole blood at concentrations of 30 and
summary of the analytes available and the quality 50 g/mL. Samples were run on all the systems and
control system for each instrument. Each instrument was results compared to unspiked blood to determine the
required to pass its own quality control prior to beginning amount of interference, which was further compared
the evaluation. against clinically acceptable error, based on CLIA 88
limits.
Table 1
Analytes and Quality Control
Table 1. Systems for the
Analytes and
GEM Premier Quality
3000 andControl Systems Instruments
Competitive for the Clots
GEM Premier 3000 and Competitive Instruments
System Available Analytes Quality Control System Clots were formed by addition of thrombogenic
GEM Premier pH, PCO2, PO2, Na+, K+, Ca++, glucose, lactate, iQM i ii
CLIA 88 limits.
(Test Instrument -
PO2 Offset
Calculated
-20 0 50 100 150 200 250 300 350 400 450 500 550 600 650 700 750
ABL 77
Rapidpoint 405
-70
Rapidlab 855
Table 2 GEM Premier 3000
Concentrations ofofElectrolytes
Concentrations and
Electrolytes and Glucose
Glucose in in pHOx plus L
-120
NIST SRM
NIST SRM956a andSRM
956a and SRM 965a
965a
SRM 956a SRM 965a PO2 (mm/Hg)
Na+ (mmol/L) K+ (mmol/L) Ca++ (mmol/L) Glucose (Calculated Tonometered Value)
(mg/dL)
Level 1 121.4 6.008 1.741 34.56
Figure 1
Level 2 141.0 3.985 1.411 78.50
Level 3 160.9 2.025 1.091 122.1 PO2 Accuracy in Competitive and GEM Premier 3000 Instruments
Level 4 292.6 Test Instrument - Expected Tonometered Value
ABL 77 Rapidlab 855 ABL 77 Rapidlab 855
Rapidpoint 405 pHOx plus L Rapidpoint 405 pHOx plus L
GEM Premier 3000 p GEM Premier 3000
25 0.3 ABL 77 Rapidlab 855
(Test Instrument -Calculated
Rapidpoint 405
(Instrument - Value
)
of NIST Standard)
15 Rapidlab 855
PCO2 Offset
Ca++ Offset
10 pHOx plus L
Value)
0.0
5 0.5 1.0 1.5 2.0 2.5
-0.1
0
5 15 25 35 45 55 65 75 85 95 105 115 125 -0.2
-5
-10 -0.3
Figure 2 Figure 5
PCO2 Accuracy in Competitive and GEM Premier 3000 Instruments Ca++ Accuracy Using NIST Standards in Competitive
Test Instrument - Expected Tonometered Value and GEM Premier 3000 Instruments
Instrument Values - NIST Reference Values
pHOx plus L
(Instrument - Value of 30
NIST Standard)
NIST Standard)
Glucose Offset
20
10
115 120 125 130 135 140 145 150 155 160 165
0
0 25 50 75 100 125 150 175 200 225 250 275 300 325
-10
NIST Value for each Standard Level NIST Value for each Standard Level
(mmol/L of Na+) (mg/dL of Glucose)
Figure 3 Figure 6
Na+ Accuracy Using NIST Standards in Competitive Glucose Accuracy Using NIST Standards in Competitive
and GEM Premier 3000 Instruments and GEM Premier 3000 Instruments
Instrument Values - NIST-Certified Values Instrument Values - NIST-Certified Values
Hematocrit Offset
0.2
NIST Standard)
3
Hematocrit)
K+ Offset
0.0 1
1 2 3 4 5 6 7
-0.2 -1
10 20 30 40 50 60
-0.4 -3
-0.6 ABL 77
-5
Rapidlab 855
GEM Premier 3000
-7 Rapidpoint 405
NIST Value for each Standard Level pHOx plus L % Hct in Spun Hematocrit
(mmol/L of K+) % Hct in Spun Hematocrit
Figure 4 Figure 7
K+ Accuracy Using NIST Standards in Competitive Hematocrit Accuracy of Blood in Competitive
and GEM Premier 3000 Instruments and GEM Premier 3000 Instruments
Instrument Values - NIST-Certified Values Instrument Values - Spun Hematocrit
Interfering Substances QC samples were run while a clot was trapped
Table 3 below shows the amount of interference by in the sensor cassette. In one case, the QC
benzalkonium and thiopental on the GEM Premier 3000 result was so errant for pH that the sample was
and competitive systems for the following analytes: identified as the wrong level of QC, and could
pH, PCO2, Na+, Ca++ and PO2. Other analytes were not have been accepted as such.
affected by the presence of these substances in any of When a clot was trapped in the sensor cassette,
the systems. Shaded areas show where the interference, if the operator observed the sample path, it was
expressed as the delta in the reported value due to the obvious that sample was not flowing normally.
interfering substance, exceeded the clinically acceptable However, it was not until the next two-point
level or other problems were detected. calibration was initiated that the instrument
flagged a problem (sensitivity below minimum
range) and no further samples would be
Table 3
Effect of Benzalkonium and Thiopental on the accepted.
GEM Premier 3000 and Competitive Instruments Clots were cleared out of the sensor cassette
Interfering Substance
Spiked into Whole Blood GEM ABL 77 pHOx RP 405 Bayer 855
with great difficulty (forcing air and liquid in
Interfering Substance Rapidpoint Rapidlab 855
Spiked into Whole Blood Premier plus L 405 and out of the sensor cassette with a syringe).
3000*
Benzalkonium (5ug/mL) Normal system washes would not remove clots
- pH -0.03, -0.02 -0.02, 0 -0.033 -0.005 -0.038
- PCO2 (mmHg) -1, 2 0, -1.5 5.5 -4.3 2.6
from the sensor cassette.
- Na+ (mmol/L) 2,3 2.5, 2.5 -3.4 off scale -9.3
- Ca++ (mmol/L) 0.30, 0.20 0.03 -0.02 0.05 0.09
- PO2 (mmHg) 0, -4.5 0, 3 -0.1 0.7 2.6, 4.9 Rapidlab 855
Benzalkonium (10 ug/mL)
- pH -0.04 0 -0.037 -0.053 -0.019 A total of 10 samples containing clots were introduced
- PCO2 1 -0.5 3.9 did not read -0.2
- Na+ 5 5 -7.8 off scale -0.5
into the Rapidlab 855 with the following results.
- Ca++ 0.36 -0.02 0.01 0.04 0.03 A blood sample containing a clot was rejected
- PO2 -5.5 4 -0.7 -4.8 did not read
Thiopental (30 ug/mL) by the analyzer due to bubbles detected in
- pH -0.01 0 -0.007 -0.002 0.01
- PO2 34 0 0.5 6.4 -1.1 sample, and no sample results were reported.
- Na+ -2 0 -0.7 -1.1 -2.1
- Ca++ -0.21 -0.02 -0.01 -0.03 -0.04
A clot became trapped in the sample
- PO2 -1 0 -2.2 2.4 0.9 introduction area and an obstruction message
Thiopental (50 ug/mL)
- pH 0.01 0.01 -0.005 0.012 -0.001 appeared. The sample was not aspirated and no
- PCO2 5 -3 -0.1 1.6 0
- Na+ 5 1 -0.1 1.6 0.2 results reported. A user-initiated wash was
- Ca++ -0.41 -0.01 -0.03 -0.03 -0.01
- PO2 -1 0 -2.1 3.1 1.7
needed to clear the obstruction.
Allowable error A clot entered the CO-Ox sample path and was
- pH 0.04
- PCO2 5 very difficult to remove. The CO-Ox stopped
- Na+ 4
- Ca++ 0.1
functioning after that event requiring a service
- PO2 10% call to correct the problem.
* Interference for the GEM Premier 3000 detected by iQM in all cases.
Rapidpoint 405
Clots A total of ten samples containing clots were introduced
ABL 77 into the Rapidpoint 405 with the following results.
A total of nine samples containing clots were intro- The clot-catcher stopped one clot. No sample
duced into the ABL 77 with the following results. was aspirated into the system and the sampling
Clots produced ??? on the results screen with was aborted.
air in sample message. Another clot was stopped by the clot-catcher,
Following a clot being trapped in the sensor but evidently a small piece of the clot passed
channel, incorrect values were reported for through the probe and became trapped in the
whole blood samples which followed, but the internal fluidics of the cartridge. The
system did not flag them as such. instrument requested that the cartridge be
Example: removed and discarded.
Another small clot was aspirated into the
pH PCO2 PO2 Na+ K+ Ca++ Cl- Hct system. The analyzer determined that there was
Expected values 7.19 64 44 131 7.7 1.28 107 40
Values reported with a problem with the sample because certain
clots in sensor channel 7.46 28 187 122 1.1 0.5 82 --- analytes were not reported (appeared with X
Bias vs. expected value: 0.27 -56% 325% -7% -86% -61% -24%
over the analyte symbol). It took 17 minutes
for all sensors to start reporting again.
When a clot was stopped by the clot-catcher, The competitive systems showed a variety of problems
the clot-catcher needed replacement. This was with accuracy, based on the accuracy limits applied in
a tedious procedure, requiring approximately ten this study.
minutes.
pHOx plus L: High bias for PCO2 at the high
pHOx plus L end of the range (>100 mmHg)
A total of ten samples containing clots were introduced High bias for Ca++ at the normal
into the pHOx plus L with the following results. and mid concentrations.
Clots produced fluidic problems flagged as High bias for glucose at all four
insufficient sample, sample flow and std A levels
flow errors, with instructions to check the flow High bias at the low end of the
path and prime the standards. After repeated range and low bias at the high end
attempts to introduce clots with resulting fluidic of the range for hematocrit
errors, the system would not accept any more ABL 77: Low bias for PO2 at the high end
samples. Tubing had to be removed and flushed. of the range (700 mmHg)
After the clots were removed and the system High bias for PCO2 at the high
was recalibrated, controls were run. For an end of the range (>100 mmHg)
unknown reason, the pH, Na+, K+, Ca++ and High bias for Ca++ at the mid-level
Hct channels did not report any results for three Rapidpoint 405: High bias for PCO2 at the high
samples after the clot was cleared. end of the range (>100 mmHg)
Rapidlab 855: Low bias for K+ at the low end of
GEM Premier 3000 with iQM the range
A total of 21 samples containing clots were introduced Low bias for Ca++ at the high end
into the GEM Premier 3000 with the following results. of the range
For six out of 21 samples, the narrow inlet
probe did not allow clots to enter the system. Interfering Substances
Error messages 1.06 (no sample solution All the systems evaluated showed some degree of
detected) and 1.07 (no air before sample interference from benzalkonium and thiopental. The
solution detected) were the most frequent advantage of the GEM Premier 3000 was that in all
messages. The clots were trapped in the tip of cases, the presence of an interfering substance was
the inlet probe and automatically wiped off by flagged by iQM. With the competitive systems, in
the probe wiper. In every instance, the analyzer some cases, erroneous result were reported without
was ready to accept another sample after the flagging of an error. See Table 3.
automatically wiping off the clot.
For three samples, a clot did enter the system Clots
and was trapped on a sensor triggering an iQM Clots which were introduced into the competitive
clot pattern detected. All three clots were systems produced mainly various fluidic errors which
immediately cleared by the iQM corrective were more of an inconvenience, resulting in down-time
action. No sensors were disabled due to clots and maintenance, than a risk of reporting incorrect
during this study. results. The exception was the ABL 77, where incorrect
The 12 remaining samples were analyzed blood results were reported, with no flags present, when
without problem, indicating that clots were not clots were present in the sensor cassette. See Results
trapped on any sensors. Section.