Sei sulla pagina 1di 23

Nothing is a waste of time if you use the experience wisely --- Rodin

BIOFLUID DYNAMICS OF THE HUMAN CIRCULATORY SYSTEM1


<Human Arjomandi> harjo001@fiu.edu <Stephanie S. Barcelona> sbarc002@fiu.edu
<Siobhain L. Gallocher> siobhain.gallocher@fiu.edu and <Martha Vallejo> mvalle02@fiu.edu 2

AbstractThe human circulatory system, consisting of INTRODUCTION


the cardiovascular and lymphatic systems, is a complex
network of branching vessels. The non-Newtonian The human circulatory system consists of both the
behavior of blood makes it a unique fluid. An analysis cardiovascular and lymphatic systems. Together, they help
of the fluid dynamics of the arteries is not an easy task to maintain the normal functioning of all the cells within
due to the many variables: fluid properties, vessel the body. The cardiovascular system functions in
properties, and Poiseuille flow was adapted to take into transportation, regulation, and protection, while the
account the pulsatility of flow in the large vessels. lymphatic system transports excess, accumulated fluid from
Finally, wave propagation speeds were derived which the interstitial spaces to the blood and plays an
took into account the compliant nature of the vessels. immunological role [1; 2; 3 and 7-8]. The purpose of this
Analog models of blood flow, such as the Windkessel study was to present a conceptual model, explain the fluid
model and Kirchoffs electrical model are presented. In mechanics, investigate the computational fluid dynamics,
the venous system, veins have less well-developed intima and discuss research advances of the circulatory system.
and media than comparable sized arteries. They
represent a large distensible sink or reservoir for blood DESCRIPTION OF THE SYSTEM
volume such that 50% of the circulating blood volume is
in the systemic venous system at rest. In addition to The cardiovascular system is comprised of the heart, blood,
viscosity of blood, the main determinants in finding the and blood vessels. It can be seen from Figure 1 that the
resistance and flow are the vessel diameter and length. heart, the muscular organ responsible for driving the blood
The capillaries are the most important part of the through the vessels, acts as two pumps in series. The right
cardiovascular system because they are the only part of side of the heart receives blood from the venous system and
the system in which exchange between blood and is responsible for pumping it through the pulmonary circuit.
interstitial fluid occurs. The rate of fluid exchange in The oxygenated blood returns to the left side of the heart
the capillaries is determined by hydrostatic pressure where it is pumped into the systemic circuit. Each different
and by oncotic pressure. The Starling hypothesis shows blood vessel type is responsible for performing a specific
their relationship to each other. The coronary function. Under normal resting conditions, 70% of the
circulation supplies the heart with blood. blood is located in the venous system, and only 17% is
Hemodynamics and the interaction of blood cells with located in the arterial system [1; 12; and 23].
vessel wall helps to understand blood flow behavior
within the heart. As is shown in figure 2, the cardiovascular and lymphatic
systems are intertwined, and they function in unison. The
Key wordsBlood, non-Newtonian, Hagen-Poiseuilles lymphatic system transports capillary ultrafiltrate,
Law, Reynolds number, Navier-Stokes, Bessel combined with proteins, bacteria, fat micelles, and immune
functions, Womersly parameter, Windkessel model, cells, from the interstitial spaces back to the cardiovascular
electrical analog model, CFD, Starling hypothesis, circulation. The lymphatic system consists of lymph, a
Ficks First Law of diffusion. network of lymphatic vessels, specialized cells called
lymphocytes, and lymphoid tissues and organs [1; 12; 19;
and 23].

The numbers in parenthesis refer to bibliographical


FUNCTIONS OF THE BIOFLUIDS
references with cited pages.
1 The body has three types of fluids. These include the
This review article was prepared for the Congress on
extracellular, intracellular and transcellular fluids. Table 1
Biofluid Dynamics of Human Body Systems. Biomedical
reveals that these fluids form 60% of the body weight for
Engineering Institute, Florida International University,
an average 70-kg male, which results in a total fluid
10555 West Flagler Streeet, Miami FL 33174,
volume of about 40 liters. The intracellular fluid makes up
organized on April 17, 2003. For details contact <Dr.
36% of the body weight, this fluid consists of the fluid
Megh R. Goyal> m_goyal@ece.uprm.edu or <Dr. Richard
found within the cells. Extracellular fluid consists of
T. Schoephoerster> schoepho@fiu.edu
2
Contributors are in the alphabetical order.

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-1
Table 1. Body fluids [4].
Fluid volume
Fluid % of Body Weight
(liters)
Intracellular 25 36
Extracellular 15 21
Interstitial 12 17
Plasma 4
Transcellular ------- ------
Total 40 57

At the alveoli capillaries, the blood picks up oxygen and


releases carbon dioxide. When the oxygenated blood
reaches the capillaries within the tissues, the oxygen is
released, and carbon dioxide, the by-product of cellular
respiration, is picked up for transporting it back to the
lungs.

The blood collects and releases nutrients and hormones for


distribution to target cells throughout the body, and waste
products are transferred to the kidneys for excretion in the
same manner. The blood regulates pH and ion
compositions of interstitial fluids. The capillary walls are
permeable to ions and water, so a diffusion gradient
between the interstitial fluid and the blood results in the
diffusion of water and ions down their concentration
Figure 1. Schematic diagram of the circulatory system gradient until normal physiologic conditions persist.
[4]. Excess ions become excreted by the kidneys and in
perspiration. Within the blood, a buffering system exists,
so if the pH drops, the blood accepts hydrogen ions and
vice versa. In this manner, the blood is capable of
regulating the pH. The blood prevents excessive fluid loss
at sites of injury by inducing a cascade of events that leads
to the formation of a clot to temporarily plug the injury site.
The blood contains leukocytes and antibodies that defend
against any toxins and pathogens that enter the system.
Finally, the blood regulates overall body temperature by
absorbing excess heat that is generated and redistributing it.
b If the body temperature rises, the blood is directed to the
surface of the skin where the heat can be lost to the
environment. If the body temperature is low, the blood is
redirected to the brain and other temperature-sensitive
organs [1; 12; and 23].

The lymphatic system plays an essential part in maintaining


a healthy body. The lymph consists of fluid and solutes
from the interstitial spaces. There is an imbalance in the
a c amount of fluid the capillaries add to and remove from the
tissues, so the lymphatic system removes this fluid
Figure 2. Overview of the lymphatic system [12, 753]. containing solutes as lymph for return to the venous
system. Many substances from the tissues that are not
capable of entering the bloodstream directly through the
interstitial fluid that accounts for 17% of the body weight, capillaries enter it through the lymphatic system. It is in
and the blood plasma that comprises around 4% of the body this fashion that the lymph transports hormones, nutrients,
weight [4]. waste products, and dietary lipids. The lymphatic system
produces, maintains, and distributes lymphocytes within the
Here we shall consider lymphatic fluid and natural blood. lymph. The lymphocytes are part of the immune response
The blood, the fluid portion of the cardiovascular system, and react to the presence of invading pathogens (viruses
has numerous functions, without which, the body would and bacteria), abnormal body cells, and foreign proteins by
cease to function. Table 2 shows the constituents and eliminating them or adapting them [1; 12; and 23].
functions of the blood.

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-2
Table 2. Constituents and functions of the blood cells [3]. Table 3. Blood flow to different organs and tissues under
basal conditions [10, 176].
Organ Percent ml/min ml/min/100g
Constituents Main functions
Brain 14 700 50
Formed elements (37-54%) Heart 4 200 70
Erythrocytes (RBC) Bronchial 2 100 25
CO2 and O2 transport
(99.9 %) Kidneys 22 1100 360
Leukocytes Neutrophil
Phagocytosis of bacteria Liver 27 1350 95
(WBC) (50-70%)
(<0.1%) Eosinophil Defense against parasitic Muscle (inactive
15 750 4
(2-4%) helminthes; modulation of state)
inflammatory processes Bone 5 250 3
Basophil Release of histamine and Skin (cool weather) 6 300 3
(<1%) other inflammation
mediators Thyroid gland 1 50 160
Monocyte Generation of monuclear- Adrenal glands 0.5 25 300
(2-8%) phagocyte system cells in Other tissues 3.5 175 1.3
tissues.
Lymphocytes T cells kill virus-infected Total 100.0 5000 ---
(20-30%) cells; B cells generate
antibody-producing terminal Table 3 summarizes the blood distribution to the different
cells; Natural killer cells kill organs and tissues in the human body. Table 4 summarizes
tumor and virus-infected the blood distribution by volume and flow. From these
cells
tables it can be seen that the greater the metabolism in an
Thrombocytes (Platelets)
Clotting of blood organ, the greater the blood flow.
(<0.1%)
Plasma (46-63%) PHYSICAL PROPERTIES OF THE BIOFLUIDS
Plasma Albumins Contributes to the osmotic
proteins (60%) pressure of plasma; The physical characteristics of blood remain the same for
(7%) transports fatty acids, all the different sources of blood (venous blood, arterial
thyroid hormones, and some blood, etc). Blood temperature is roughly 38 C (100.4 F),
steroid hormones. pH is between 7.35 and 7.45, and viscosity is five times
Globulins Immunoglobulins attack
greater than that of water. Table 5 lists the physical
(35%) foreign proteins and
pathogens; transport properties of an adult human blood. s the blood flows
globulins bind ions, through the body, materials are constantly being added and
hormones, and other removed, so the variations occur continuously. There are
compounds that might be differences among normal persons in every measurable
lost at kidneys or have very constituent. The composition of the blood is affected by
low solubility in water. age, sex, genetic factors, environmental conditions,
Fibrinogen Converted to fibrin, which physiological state of the individual, and disease [4].
(4%) provide the basic framework Lymph is derived from interstitial fluid that flows into the
for a blood clot.
lymphatics. It is a clear, watery fluid found in the
Regulatory Catalyze specific
proteins biochemical reactions and lymphatic vessels. There are several factors that affect
(<1%) regulate activities of other lymph flow, but essentially the primary factors that
cells. determine lymph flow are: (1) The interstitial fluid pressure
Other Electrolytes Maintain normal and (2) The activity of the lymphatic pump. Thus it can be
solutes extracellular fluid ion said that the rate of lymph flow is determined by the
(1%) composition essential for product of interstitial fluid pressure times the activity of the
cellular activities; contribute lymphatic pump [10].
to osmotic pressure of body
fluids.
Organic Needed for ATP production,
CHEMICAL AND BIOLOGICAL
nutrients growth, and cell COMPOSITION OF THE BIOFLUIDS
maintenance.
Organic Carried to sites of Blood is a fluid connective tissue that is composed of a
wastes breakdown or excretion. fluid matrix, plasma, with suspended formed elements.
Water Transports organic and inorganic molecules, The formed elements, whose percentage volume or
(92%) formed elements, and heat. hematocrit in blood is 37-54%, consist of three basic cell
types, namely, erythrocytes (red blood cells), leukocytes
(white blood cells), and thrombocytes (platelets).

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-3
formed elements, are involved in the bodys clotting
Table 4. Blood distribution by volume and flow [3, 513]. process [3 and 15].

Physiologic Typical Typical Plasma makes up 46-63% of the volume of whole blood. It
structure volume flow rate, is an aqueous, saline suspension medium that is 92% water
% mL mL/min by volume. The composition of the plasma closely
Aorta 4.5 156 5345 resembles that of the interstitial fluid, with the major
Arteries, larger 5.4 187 695 plasma ions being those similar to the interstitial fluid. The
distributing two major differences between the plasma and interstitial
Brain, cerebral 2.3 78 - fluids exist in the concentrations of dissolved proteins and
circulation the presence of respiratory gases. There is a continuous
Heart chambers 9.8 338 - exchange of water and ions across the capillary membrane,
so an equilibrium results in similar concentrations of ions.
Coronary circulation 2.3 78 267
The capillary membrane is impermeable to the plasma
Renal circulation 1.5 52 1149
proteins, resulting in the marked difference in
Lungs 15 520 5185
concentrations across the membrane. Plasma proteins
Lymphatic 0.4 15 5 make up approximately 6.5-8% by weight of the plasma.
circulation The three major plasma proteins are albumins, globulins,
Muscle, inactive 21 728 802 and fibrinogens. Albumins contribute to the osmotic
skeletal pressure of plasma and transport lipids and steroid
Skin 7.5 260 428 hormones. Globulins transport ions, hormones, and lipids,
Veins, large 26.7 925 - and participate in the immune response. Fibrinogen is a
collecting vital component of the clotting system. The remaining 2%
Vena cava 3.6 125 5185 of the plasma consists of mineral, trace elements, and
Total 100 3462 5345 electrolytes, some required by the cells for normal
functioning while the others contribute to the acid-base
Table 5. Physical properties of adult human blood. balance of the blood [3 and 15]. Table 6 shows a
laboratory analysis of blood for a healthy 20-year-old
Value female along with the normal ranges [2].
Property Ref
Range Normal
Lymph is considered to be a part of the extracellular fluid
Density (g/cm ) 3
1.050-1.064 1.057 [5, 475] system, and it is in many ways similar to plasma. Lymph
differs from plasma in that it contains a significantly lower
Viscosity (cP) 2.18-3.59 3.0 [5, 475] percentage of suspended proteins. In terms of the common
ions found in both the intracellular and extracellular
pH 7.35-7.45 7.4 [5, 475] compartments of the body, the difference in concentrations
of these ions between the lymph and plasma is not
Specific gravity 1.0501-1.0619 1.057 [5, 475] significant. Lymphocytes are the most abundant cells
found in lymph, accounting for 83% of the cells.
Refractive index 16.2-18.5 17.4 [2, 12]
Neutrophils, eosinophils, and erythrocytes are also found in
the lymph contributing to 0.9, 0.2, and 15% of the total
Specific heat (g/cal) - 0.92 [2, 12] number of cells respectively [15 and 19]. In Table 7, the
lymph components of a healthy male are given.
Surface tension
55.5-61.2 - [2, 12]
(dynes/cm) RHEOLOGY OF THE BLOOD
Colloid osmotic
Many factors govern pressure and flow in the circulation,
pressure: plasma 280-480 330 [2, 12]
and their interaction is complex. Blood is a non-
(mmHg)
homogenous, ionic, anisotropic, composite fluid, which is
formed of a mixture of blood cells and plasma. carried in a
The erythrocytes contribute to approximately 95% of the liquid that contains high molecular weight, long- chain
formed elements, and these cells are specialized for the polymers that behave in a complex way under shear stress
transportation of oxygen in the blood. The leukocytes are [3]. Thus blood exhibits a non-Newtonian, viscoelastic
far less numerous than the erythrocytes, accounting for less flow. In contrast, for a Newtonian fluid the shear-stress rate
than 0.15% of the formed elements. The leukocytes are an relationship is linear with the slope equal to the viscosity of
integral part of the bodys defense mechanism as they are the fluid [4]. For the non - Newtonian fluids, the slope of
capable of identifying and removing foreign substances the line at a given value of the shear rate is the apparent
from the body. The agranulocytes, the monocytes and viscosity. Blood can behave both as Newtonian and non-
lymphocytes, identify foreign matter, and the granulocytes, Newtonian fluid depending on the shear rate.
the neutrophils, eosinophils, and basophils, dispose of it.
The thrombocytes, accounting for approximately 5% of the
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-4
Table 6. Laboratory analysis of blood for a healthy 20- Table 7. Lymph components of a healthy male [2].
year-old female [2].
Component Units Average Range Thoracic duct
Constituent Units
lymph
Glucose mg/dl 91 65 - 115
Electrolytes
Sodium mEq/L 142 134 - 147
Chloride mEq/L 87 103
Potassium mEq/L 4.1 3.5 - 5.5
Phosphorous mEq/L 2.0 3.6
Chloride mEq/L 105 95 - 110
Potassium mEq/L 3.9 5.6
Carbon dioxide mEq/L 23.3 23.0 - 32.0
Sodium mEq/L 118 - 132
Urea nitrogen mg/dl 10 6.0 - 22.0
Nitrogenous substances
Creatinine mg/d 0.9 0.5 - 1.4
Protein, Total g/100mL 2.91 7.33
Bun/Creatinine ratio ------ 11.1 7 - 23 Albumin g/100mL 1.50 2.67
Uric Acid mg/dl 4.2 1.7 - 7.4 Globulin g/100mL 1.50 - 4.80
Phosphate mg/dl 3.7 2.4 - 4.9 Bilirubin g/100mL 0.8
Calcium mg/dl 9.5 8.5 - 10.5 Creatinine g/100mL 0.8 - 8.9
Protein, total g/dl 8.2 6.2 - 8.4 Uric Acid g/100mL 1.7 - 10.8
Albumin g/dl 4.6 3.4 - 5.1 Nitrogen, non-
g/100mL 13.4 - 139.0
Globulin, protein
g/dl 3.6 2.2 - 3.7
(Calculated) Lipids, carbohydrates
A/G ratio ------ 1.3 1.0 - 1.9 Cholesterol, Total mg/100mL 34 - 106
Bilirubin: Total mg/dl 0.5 0.2 - 1.3 Free mg/100mL 15 - 51
Bilirubin: Direct mg/dl 0.14 0.00 - 0.40 Fat, Total mg/100mL 1.1
Alkaline Glucose mg/100mL 136
U/L 76 20 - 150
phosphatase Enzymes
GGT U/L 11 1 - 70 Aldolase units/mL 10
AST U/L 14 1 - 45 Phosphatase units/mL 0.7
ALT U/L 8 1 - 45 Alkaline units/mL 1.9
LD U/L 126 80 - 235 Transaminase units/mL 12
Tryglycerides mg/dl 94 40 - 200
Cholesterol mg/dl 141 120 - 200 At low shear rates, the apparent viscosity of the blood is
Iron mg/dl 83 40 - 180 quite high due to the presence of rouleaux and aggregates,
TIBC mg/dl 357 250 - 390 thus blood behaves as non-Newtonian fluid. However, at
Trans. Sat. percent % 23 15.0 - 50.0 shear rates above 100/sec, only individual cells subsist, and
PBG mg/gm 0.73 0.00 - 1.20 blood behaves as if were a Newtonian fluid [4].
HDL Cholesterol mg/dl 53 35 - 95
Cholesterol/HDL
ratio
------------ 2.7 2.9 - 4.3 Blood flow in arteries, veins, and the smaller vessels is
LDL Cholesterol Newtonian near the vessel wall, and non-Newtonian as the
mg/dl 69 50 - 130 flow approaches the centerline of the vessel [4]. The
(Calculated)
CBC anomalies in the viscous behavior of blood result partly
WBC 103/mm3 6.3 3.9 - 11.4 from the presence and orientation of the cells suspended in
RBC 106/mm3 4.11 3.80 - 5.10 plasma and partly from the accumulation of RBCs in the
Hemoglobin, Hb g/dl 12.5 11.6 - 15.1 axial portion of the bloodstream. The result of this is a
Hematocrit, Hct % 37.2 34.0 - 45.0 decrease in apparent viscosity with increasing velocity of
MCV fl 90.5 80.0 - 96.0 flow and with decreasing vessel radius [17]. As blood
MCH PG 30.4 26.8 - 33.2 leaves the aorta and flows into capillaries, the velocity of
MCHC % 33.6 32.0 - 36.0 the blood decreases. The average velocity of blood in the
RDW % 11.9 11.0-15.0 aorta is approximately 0.3 m/s while the average velocity of
Platelet count 103/mm3 277 150 - 400
blood in a capillary is about 10-3 m/s. Blood velocity is
MPV fl 7.3 7.5 - 11.5
inversely related to the total cross-sectional area of the
Segmented
103/mm3 4.17 1.70 - 8.50 vessels carrying the blood as shown in figure 3 [21].
Neutrophils, ABS
Lymphocytes, ABS 103/mm3 1.61 1.10 - 3.50
Monocytes, ABS 103/mm3 0.40 0.04 - 0.90 The velocity of blood at the center of the vessel is greater
Eosinophil, ABS 103/mm3 0.06 0.03 0.55 than the velocity near the walls (Figure 4). This affects the
Basophils, ABS 103/mm3 0.05 0.00 - 0.13 distribution of RBCs in the circulatory system. It implies
Segmented that the distribution of RBCs is not uniform so that there
% 66.2 40.0 - 75.0
Neutrophils are more in the center than at the edges. The effect of this
Total lymphocytes % 25.6 15.0 50.0 is two-fold. First, as blood enters a small vessel branching
Monocytes % 6.3 0.0 10.0 from the side of a main vessel, the hematocrit (percentage
Eosinophils % 1.0 0.0 6.0 of red blood cells in blood) will be less than that of the
Basophils % 0.8 0.0 2.0 main vessel. Second, extremities have a greater percentage
ESR, westergreen mm/hr 17 0 - 20 of RBCs than blood from the heart [21].

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-5
Figure 6. The viscosity-shear rate relations in whole
blood (), defibrinated blood (x----x), and washed
cells in Ringer solution (-----), at 45% and 90% red cell
colume concentrations.
Figure 3. Velocity of blood (solid line) decreases as total
cross-sectional area increases (dashed line) [21]. The rheology of blood can be significantly affected by the
hematocrit and the plasma protein concentration. Increasing
the hematocrit increases the viscosity of blood (Figure 5).
The same effect is also seen when the fibrinogen
concentration increases. Fibrinogen is the primarily
responsible for the non-Newtonian fluid behavior of the
blood. Other plasma proteins like albumin, do not
contribute to the non-Newtonian behavior of blood [7].

Blood is a nonhomogeneous, anisotropic, ionized fluid


containing a suspension of viscoelastic particles, and it
displays non-Newtonian, viscoelastic deformation
characteristics that can be modeled by the Power Law:
n
du /1/
= k + y
dy
Figure 4. (a) Velocity profile of blood flow. (b) Non-
uniform distribution of blood cells [21]. Where: is the time dependent stress, du/dy is the time
dependent strain, K is a material dependent constitutive
parameter which reduces to the real dynamic coefficient of
viscosity for a Newtonian fluid, n is the material-dependent
non-linearity index that is > 0, and y is the yield stress. y,
for blood, in dynes per square centimeter, can be calculated
using the equation:

y = ( H 0.1)( C F + 0.5 )
/2/

Where: H is the hematocrit expressed as a fraction, and CF


is the plasma concentration of fibrinogen, expressed in
grams per 100 mL. Typical values for K and n are
0.1352x10-1 Pa-s and 0.7844 respectively [3, 482].

It has been determined that the deformation properties of


RBC suspensions is influenced by shear rate, hematocrit,
Figure 5. Viscosity versus Hematocrit percentage [10]. degree of cellular aggregation, and the mechanical
properties of each of the cells themselves [3, 477].

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-6
Figure 6 reveals that an increase in shear rate leads to a
decrease in viscosity, and an increase in hematocrit leads to
an increase in viscosity. In Figure 7, the effects of shear
rate and temperature are observed. Again, viscosity is
shown to decrease with increasing shear rate, and a
decrease in temperature led to an increase in viscosity.

Deformability is a term used to describe the structural


response of a body or cell to applied forces. The effect of
RBC deformability in influencing general fluidity of whole
blood is clearly revealed in Figure 8. This figure shows the
relative viscosity of blood at a shear rate >100 s-1 (at which
particle aggregation is negligible, isolating RBC
deformability) compared with that of suspensions with
rigid spheres. At 50% concentration, the viscosity of a
suspension of rigid spheres reaches almost infinity so that
the suspension is not able to flow. On the contrary, normal
blood remains fluid even at a hematocrit of 98%, on
Figure 7. The variation of the viscosity of human blood with shear
account of the deformability of its RBCs [32]. rate and temperature for a male donor with a hematocrit of 44.8 %
[6, 68].
Cassons equation is another method used to describe the
shear stress of whole blood:
du
= y +
dy /3/

Where: the shear stress, yield stress, and shear rate are
defined by equation /1/ and /2/, and is a constant.
Experimental data of whole blood at 37C and a hematocrit
of 51.7% are shown in Figure 9. It is observed that
Cassons equation can be fitted to the experimental data.

In Figure 10, the impact of hematocrit on the yield stress of


whole blood is represented. As is expected, the yield stress
increases with increasing hematocrit.
Figure 8. Variation of relative viscosity of blood and
Figure 11 shows the impact of shear rate on shear stress. It suspension with rigid spheres at a shear rate < 100 s-1 [32].
can be seen that at low shear rates, the blood behaves like a
non-Newtonian, Casson fluid, but as the shear rate
increases, the blood begins to act more and more like a
Newtonian fluid.. In the figure, the data to the right of the
dashed curve represents the Newtonian nature of the fluid,
whereas the non-Newtonian region is located on the left
hand side of the dashed curve.

The most contributing factor in making blood non-


Newtonian is RBC aggregation. Human RBCs have the
ability to form aggregates known as rouleaux. Rouleaux
formation is highly dependent on the concentration of
fibrinogen and globulin in plasma. As RBCs form
rouleaux, they will tumble while flowing in large vessels.
The tumbling disturbs the flow and requires the
consumption of energy, thus increasing blood viscosity at
low shear. As shear rate increases, blood aggregates tend
to be broken up resulting in drop in blood viscosity. In
short, rouleaux formation increases blood viscosity,
whereas breaking up rouleaux decreases blood viscosity Figure 9. Casson plot for whole blood at a hematocrit of
[32]. 51.7% and at a temperature of 37C [6, 69].

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-7
as the Fahreus-Linqvist effect is shown clearly in Figure
12. While the blood is flowing through the vessels, the
RBCs drift towards the centerline and equilibrate and align
themselves at a particular radial location in order to
decrease the amount of profile drag in the direction of flow.
With this arrangement of RBCs, a cell free layer persists at
the walls of the vessel, so fluid flowing along the walls of
the vessels consists only of plasma. The cell free layer is
what leads to a decrease in the hematocrit with decreasing
vessel diameter, as the size of the cell free layer remains
constant in all sized vessels. What this means is that with
decreasing vessel size, the cell free layer occupies a greater
volume of the vessel leading to decreased apparent
viscosity. A phenomenon called plasma skimming occurs
in the body, as small branching vessels draw the majority
of their fluid from the cell free layer, thereby decreasing the
hematocrit in the smaller vessels [3 and 6]. Figure 13
shows the effect of the size of the blood vessel on the
relative viscosity of the blood.
Figure 10. Variation of the yield stress of blood with
hematocrit [8, 70].

Figure 13. The change of relative viscosity of blood with the size
Figure 11. Casson plot for a higher shear rate rheological data
of blood vessel [6, 172].
for blood [6, 71].
Lymph is thought to behave like a Newtonian fluid and it
obeys Newtons law of viscosity:
du
= /4/
dy
Where: is the dynamic coefficient of viscosity [3 and 6].

CONCEPTUAL MODEL OF THE


CIRCULATORY SYSTEM

The Figure 14 shows an analog model of the human


cardiovascular system, but it provides a basic idea of how it
functions. In this model, the Oxygen Plant represents the
Figure 12. Elevated blood viscosity at low shear rates lungs where oxygen is absorbed and carbon dioxide is
indicates RBC aggregation (rouleaux formation). Blood removed from the blood. The pipeline leading from the
viscosity decreases with increasing shear rates as RBC plant to pump 1 represents the pulmonary veins, pump 1
aggregations breaks up to individual RBCs [32]. represents the left hand side of the heart, and the pipeline
leading from pump 1 to the factory represents the systemic
For blood flowing through a cylindrical vessel, such as is arterial circulation. The factory represents the capillaries
found in the human circulatory system, the apparent bed where oxygen and nutrients diffuse into the tissues.
viscosity decreases with decreasing vessel diameter when The pipeline leading from the factory to the inlet valve
all other parameters are kept constant. This effect, known represent the systemic venous system, pump 2 represents

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-8
the right hand side of the heart, and the pipeline leading There are three primary factors that determine the
from pump 2 to the plant represents the pulmonary arteries resistance to blood flow within a single vessel: diameter (or
[36, 1-2]. radius), length, and viscosity of the blood. Other factors
are the organization of the vascular network (series and
parallel arrangements), and extravascular mechanical
forces acting upon the vasculature [36].

If only the primary factors are considered, Poiseuilles law


can be used to find the resistance and flow. In Poiseuilles
law, vessel resistance R is directly proportional to the
length (l) of the vessel and the viscosity () of the blood,
and inversely proportional to the radius to the fourth power
a4, [36].

8. .l
R= /6/
4
.a

Flow, then, can be related to the resistance as follows:

P
Figure 14. Circulatory system analogy using industrial Q = /7/
system [36, 1]. R

Where: P= Pressure difference between two points and Q


FLUID MECHANICS is the volume flow rate.
The description of any flow problem requires the
determination of the three major component equations of In the body, however, flow does not conform quantitatively
motion and pressure as functions of position and time. to this relationship because this relationship assumes long,
However, the problem of treating the pulsatile flow of straight tubes (blood vessels), a Newtonian fluid (e.g.,
blood through the cardiovascular system in precise water, not blood which is non-Newtonian), and steady,
mathematical terms is overwhelming. This can be reasoned laminar flow conditions. Nevertheless, the relationship
as blood is non-Newtonian; flow is not steady but pulsatile; clearly shows the dominant influence of vessel radius on
vessels are elastic, multi branched conduits of constantly resistance and flow and therefore serves as an important
changing diameter and shape. Therefore, a precise concept to understand how physiological and pathological
calculation of blood is required in each segment of the changes in vessel radius affect flow [22, 2].
cardiovascular system [33, 35].
In Figure 15, the relationship between flow and radius for a
Blood flow through an organ or any vascular network is single vessel is shown. Laminar flow conditions are
driven by a pressure gradient or perfusion pressure that is assumed at pressure, viscosity, and vessel length. As vessel
normally represented by the difference between the arterial radius decreases, there is a dramatic fall in flow because
and venous pressures across the organ. The actual blood flow is directly related to radius to the fourth
flow at any given pressure gradient is determined by the power. Therefore very small changes in vessel radius can
resistance to blood flow [14, 80-86]. have dramatic effects on flow [35].

Blood flow can be either laminar or turbulent. Flow is Within a given organ, the vascular arrangement is a
laminar when the fluid travels smoothly or in regular paths. combination of series and parallel elements. For example,
The velocity, pressure, and other flow properties at each in the Figure 16, the vascular segments labeled A (large
point in the fluid remain constant. Turbulent flow occurs artery), a (arterioles), c (capillaries), v (venules), and V
when the fluid undergoes irregular fluctuations or mixing. (large vein) are in-series with each other. All of the blood
The speed of the fluid at a point is continuously undergoing that flows through the large artery (A), also flows through
changes in magnitude and direction, which results in each of the other segments. Within each of the series
swirling and eddying as the bulk of the fluid moves in a segments, there may be many parallel components (e.g.,
specific direction. Reynolds number can be used to find if several parallel capillaries may arise from a single
the flow is laminar or turbulent [8, 84-99]: arteriole). Each vascular segment will have a resistance
um d value (Rx) that is determined by the length and radius of
Re = /5/ each of the individual parallel vessels [35].

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-9
results from partial or complete blockage of the coronary
arteries due to the formation of a fatty deposit, or plaque, in
the wall of a coronary vessel. Blood flow is reduced as a
result of the narrowing of the passageway [28].

During maximum exertion, oxygen demand rises and blood


flow to the heart increases nine times that of resting levels.
Coronary artery disease (CAD) is the partial or complete
blockage of coronary circulation. CAD is one of the
leading causes of mortality and morbidity in western
society. CAD results from atherosclerotic lesions
(narrowing) in the main branches or from diffuse disease in
the microvascular bed. Both lead to reduced blood flow
where the oxygen demand of the heart muscles is not met
resulting in myocardial ischemia and angina pectoris [28;
Figure 15. Change in radius with flow [35]. 13, 415].

Complex hemodynamics plays a critical role in the


development of atherosclerosis. It is important to
understand not only the forces and movement of blood cells
and whole blood but also the interaction between blood
cells and the vessel wall. Geometry and elasticity of vessel
walls help determine the flow behavior. High velocity
fluctuations that disturb flow should be avoided. The
Figure 16. Veins in series [35] following dynamic factors are responsible for the deposit of
blood cells and lipids, a leading cause of atherosclerosis:
For the in-series resistance network, the total resistance
(RT) equals the sum of the individual resistances [35]: 1. Flow rate ratio.
2. Pressure and velocity gradients.
R = R A + R a + R c + R v + RV /8/ 3. Flow behavior, velocity distribution, and shear stress
on the wall and on blood cells.

For the parallel resistance network, with 3 parallel vessels, Interaction between blood cells and vessel wall leads to the
total resistance is used [35]: formation of plaques and agglomeration. These deposits
are found predominantly at arterial bends or wherever
1 1 1 1 blood flow is disturbed such as bifurcations, wherever a
= + + /9/ secondary flow is created, and wherever flow separation
R R1 R2 R3 regions can be found [13, 415].

The total resistance of a network of parallel vessels is less CORONARY ARTERY FLOW
than the resistance of the vessel having the lowest
resistance. Therefore, a parallel arrangement of vessels The following factors are important to wall adaptation and
greatly reduces resistance to blood flow. That is why to the development of atherosclerosis in coronary arteries:
capillaries, which have the highest resistance of individual local wall shear stress induced by coronary artery flow and
vessels because of their small diameter, constitute only a local wall stress (or strain) induced by wall deformation.
small portion of the total vascular resistance of an organ or The wall shear stress induced by coronary artery flow is
microvascular network [35]. dependent on the following: (1) Vessel geometry which is
three-dimensional, non-planar, and time-dependent as a
result of wall motion induced by pressure pulse acting on a
CORONARY CIRCULATION distensible wall and movements of the heart (Figure 17);
(2) Flow, a time-dependent pulsatile (Figure 18); and (3)
Since the heart works continuously, cardiac muscle cells Rheological properties of blood that is exhibiting shear
need reliable supplies of oxygen and nutrients. The thinning behavior as a result of RBC deformation (Figure
coronary circulation supplies the heart with blood while the 19).
heart pumps the blood into the systemic circulation.
During maximum exertion, oxygen demand rises and blood The wall stress induced by wall deformation is dependent
flow to the heart increases nine times that of resting levels. on the following: (1) Vessel geometry which is three-
Coronary artery disease (CAD) is the partial or complete dimensional and time-dependent exhibiting a non-uniform
blockage of coronary circulation. CAD is one of the wall thickness (Figure 14); (2) Motion of the heart (Figure
leading causes of mortality and morbidity in western 17); (3) Pressure wave in coronary arteries (Figure 18); and
society. Reduced circulatory supply, or coronary ischemia, (4) Material properties of the arterial wall [24, 17-15].
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-10
Figure 20. Laminar flow in a round tube.

Blood flow in the arteries is pulsatile, and the pulsatility


Figure 17. Systolic and diastolic geometry of the right decreases with distance from the heart. In order to
coronary artery [24, 17]. investigate the flow of blood through the arteries, a
simplified model is needed, and its complexity must then
be increased in order to give an accurate representation of
blood flow [5, 108-112; 18, 1-8].

Laminar Flow in a Tube

The vessels of the arterial system are for the most part
circular, therefore the laws governing the flow of liquids in
in a long straight tube were considered for steady state,
laminar flow and Newtonian fluid. In Figure 20, the radius
of a tube is a. The velocities in the x, y, and z directions
are: u = u(y, z); v = 0; w = 0.

In order to perform the analysis, flow of an incompressible


The fluid entering the system does not accumulate, so we
can use following the continuity equation:

u v w
+ + =0 /10/
x y z

The Navier-Stokes equation for flow through a cylindrical


Figure 18. Relevant pressure and flow waves [24, 15]. tube is:

2 2
u u 1 dp
+ = /11/
y
2
z
2 dx

dp/dx is a pressure gradient. The Navier-Stokes equation in


cylindrical coordinates (x, r, ) is as follows:

2
1 u 1 u 1 dp
r + 2 2
= /12/
r r r r dx

If the flow is assumed symmetrical, the equation reduces


to:

1 d du 1 dp
r = /13/
r dr dr dx

No-slip condition at the tube wall at r = a and the symmetry


Figure 19. Shear thinning behavior of blood [24, 15]. at the center line yield following boundary conditions:
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-11
2
u = 0 at r = a u 1 u 1 p 1 u
+ + = /18/
2 r r x t
r
du
= 0 at r = 0 For the case of pulsatile flow, the pressure gradient is
dr
assumed to be a simple harmonic motion:
On integrating equation /13/, we get:
p * i t
= A e /19/
u=
1
4
(a 2
r
2 dp
) dx /14/
x

Where: A* is the complex conjugate of A, and =2f, the


angular frequency in radians per second. Substituting
The equation /14/ shows is a parabolic velocity profile for equation /18/ into equation /19/, we get:
the Hagen-Poiseuille flow. Volume flow rate is defined as:
2 *
a u 1 u 1 u A i t
Q = 2 0 urdr /15/ + = e /20/
2 r r t
r
Equations /14/ and /15/ result in the following Hagen- Equation /20/ can be simplified by substituting
Poiseuille formula: U = ue it , and by canceling out the exponential term
throughout the equation. We get:
4 4
a dp a p p
Q = = = /16/ 2 *
8 dx 8 L R d U 1 dU i A
+ U = /21/
2 r dr
dr
Where: p is the pressure drop in a segment of vessel of
length L, and R is the resistance. The shear stress at the Equation /21/ is a form of the Bessels equation. The
vessel wall, (u/r) at r = a is as follows: solution with appropriate boundary conditions is as
follows:
a dp a p 4

w
=
2 dx
=
2 L
= Q /17/
a3 A 0 r
* [ ( / )i ]
3/ 2

As was shown in Figure 15, the most important factor


U =
i
1
0 R [ ( / )i ]
3/ 2
/22/

controlling blood flow through and artery is the radius of


the vessel. If the size of the vessel is halved, the flow rate
decreases by a factor of 16, and if the vessel size is
The equation /22/ contains the Bessel function. The
doubled, the flow rate increases by a factor of 16. A
decrease in vessel size by a factor of 2 would require an quantity, R ( / ) , is a non-dimensional parameter
increase in pressure by a factor of 16 in order to maintain referred as Womersly number (). It is a ratio of unsteady
the constant flow rate, and vice versa. Physiologically
to viscous effects. When becomes large, the velocity
speaking, a successful way to adjust blood pressures is by
profile becomes blunt, and this can be seen in Figure 21.
changing vessel diameters.
The volume flow can be calculated in the same manner as
Hagen-Poiseuilles equation is not applicable to large
Poiseuille flow rate. The Bessel function provides a
arteries with turbulent, pulsatile flow. Another requirement
solution to the Navier-Stokes equation for pulsatile,
for equation /16/ is that the tube should be rigid and long
unsteady flow, and this more closely approximates what is
enough to avoid entrance effects. Thus many of the
happening in the arteries [5, 130-134; 18, 35-39; 25, 137-
assumptions are not applicable under actual physiological
161].
conditions. Therefore, Hagen-Poiseuilles formula can
only be used for blood flow under very specific conditions
Wave Propagation in Arteries
[5, 114-120; 18, 12-17; 31].
The arterial walls are highly elastic. Therefore, when a
Pulsatile Flow in a Tube pressure waveform traverses the artery, a certain amount of
stretch is observed in the arterial wall. This stretch is
The Navier-Stokes equation for unsteady, laminar flow and
proportional to the pressure at the specific location.
an incompressible liquid is given by:

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-12
u u 1 p
+u + = 0, /24/
t z z

Where: A is the tube cross sectional area, u is the uniform


axial velocity of blood, p is the pressure in the tube, is the
fluid density, z is the axial coordinate, and t is time. It is
assumed that the elasticity of the wall is given by a
relationship between pressure and cross-sectional area. It is
assumed that the wavelength is large compared with the
tube radius, so the conservation of mass and momentum
equations reduce to:

A u
= A /25/
Figure 21. Theoretical velocity flow profiles of an t z
oscillating flow [12, 3.9].
u 1 p
= /26/
t z

Equation /25/ is differentiated with respect to time, and


equation /26/ is differentiated with respect to z. We obtain:

2 2 2
p A p 2 p
= =c /27/
2 A p t 2 2
z t

We can solve for the wave speed (c) to obtain:

A p
c= /28/
A
Figure 22. Cross-section of artery showing change of
volume and volume flow rate [12, 3.7]. The solution to this equation states that waves are
propagated at speed, c, from the heart to the distal
circulation. The high-pressure portion of the pressure wave
will tend to travel at a higher speed than the low-pressure
portion of the wave, resulting in a steepening of the
pressure wave as it travels from the heart to the distal
circulation. This is represented in figure 23. It can be
Figure 23. Steepening of a pressure pulse with distance
observed that wave speed varies with age due to the
along the artery [12, 3.8].
ultimate stiffening or loss of elasticity in the vessel walls.
To explain wave propagation, we assumed an ideal case An assumption was made that the vessels were infinitely
with an infinitely long, straight, isolated, circular, long, but in reality, this is not true, so it is possible for the
cylindrical, elastic tube containing a homogeneous, wave to reflect backwards up the artery and add to new
incompressible, and nonviscous fluid (Figure 22). wave forms being generated [5, 140-150; 12, 3.6-3.8].

If this tube is disturbed, a wave will propagate along the VENOUS SYSTEM
tube. Our aim is to determine the speed of wave
propagation. For one-dimensional flow, the conservation The main characteristic of the venous system is having
of mass is defined by: walls that are thinner than arteries. This gives the venous
system a lower pressure than in arteries, and a possibility of
collapsing. Furthermore, presence of valves and anatomical
A proximity of arterial pulses gives the venous system
+ ( Au ) = 0 , /23/
t z pressure fluctuations [35]. In the venous system, the
Reynolds number does not exceed 3,300. Table 8
The conservation of momentum equation is as follows: illustrates Reynolds numbers in various part of the venous
system.

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-13
Table 8. Some properties of the venous system [34].

Vessel Velocity Diameter


Re #
(cm/sec) (cm)
Venule
0.2 0.002 0.01
Vein
10 0.5 140
Vena
Cava 38 3.0 3300

MICROCIRCULATION

Physically blood in the capillaries can be viewed as a two-


phase fluid consisting a liquid plasma phase and an
elastically deformable solid phase filled with blood cells.
Before going into the actual mechanics in the capillaries it
is necessary to explain the different forces that act upon the
capillaries. The first force and probably the most relevant is
pressure. In the cardiovascular system the pressure gradient Figure 24. Vessel diameter, total cross sectional area,
is the circulatory pressure that averages100 mm Hg. This average blood pressure and velocity of blood flow in the
is value is so high in part because this pressure is needed to blood vessels [15, 704].
force blood through the arterioles and the peripheral
capillaries. The circulatory pressure is often separated in
three classes: (1) Blood pressure: capillary blood flow is
directly proportional to blood pressure, (2) Capillary
hydrostatic pressure (CHP): it normally ranges from 35 mm
Hg to about 18 mm Hg, and (3) Venous pressure: this
pressure is very low.

The resistance should also be considered. The flow is


directly proportional to the pressure gradient, and inversely
proportional to the resistance. The resistance of the
cardiovascular system opposes the movement of blood.
The greater the peripheral resistance, the slower the blood
flow. Peripheral resistance comes from different sources
such as vascular resistance, viscosity and turbulence.
Vascular resistance is the resistance of the blood vessels Figure 25. Forces acting across capillary walls [15, 709].
and depends both on the vessel length and the vessel
diameter. Viscosity is the resistance to flow caused by the If on the other hand both numbers are larger than 1, fluid
interactions of molecules and suspended materials in a viscosity is ignored, and the flow is said to be
liquid. And finally turbulence is causes the flow to slow macrocirculation. These numbers are dimensionless [15,
and the resistance to increase. The variation of pressure 196]:
through the different blood vessels and the variation in
cross sectional area, diameter, and velocity are shown in
D
figure 24.. It can be seen that pressure and the speed of =N = /30/
blood flow are proportional to the cross-sectional area of w 2
the vessels involved, also blood flow velocity decreases in
velocity as the total cross sectional area of the vessel
Where: D is the blood vessel diameter, is the kinematic
increases [6, 702-703].
viscosity of the blood, and is the circular frequency of
oscillation of the blood velocity fluctuations.
Capillary blood flow is often identified with the
microcirculation. The microcirculation also includes blood
flow to the arterioles, and venules. In order to distinguish VD
Re = /31/
between macrocirculation and microcirculation, two basic

parameters are taken into account: Womersley number (),
and the Reynolds number (Re). If the Womersley number
Where: V is the mean velocity of flow in the vessel, D is
and the Reynolds number are both smaller than 1, then
the diameter of the vessel and is the kinematic viscosity
inertial force is not taken into account, and flow is said to
of the blood. It should be mentioned that Equation /5/ and
be microcirculation.
/31/ are the same. Some of the general features of the
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-14
microcirculation are: hematocrit is lower in the blood that The second term is the osmotic pressure difference between
in the heart, apparent viscosity of blood decreases due to the capillary and the interstitial fluid. If this term is
the declining concentration of red blood cells, and activated positive osmotic flow of water into the capillary will occur,
white blood cells have a greater chance to stick to the wall and if it is negative, the opposite will occur. If the first
vessel because of closer contact with the endothelium [6, term in brackets is positive, this net osmotic pressure flows
179 and 196]. opposite to the flow induced by the hydrodynamic pressure
difference. From equation /32/, if the value in brackets is
The velocity and hematocrit distribution in any capillary positive, then there is a net flow from the capillary to the
bed fluctuates constantly. One of the most important interstitial fluid, on the other hand if this term is negative,
features of the microcirculation is the appearance of the fluid will go into the opposite way (interstitial to
RBCs. In small blood vessels (capillaries) RBCs are so capillary) [4, 8-9].
big that they fill the vessel from wall to wall. The non-
uniform RBC distribution results in the Fahraeus effect that Also from the Starling equation, the filtration coefficient
emerges when the microvessel hematocrit is smaller than can be defined as the ratio of J and P. This coefficient is a
the feed or discharge hematocrit; due to the fact that RBCs constant and represents the properties of the pore. Lp for
move with a higher velocity than blood plasma, and the blood, in m2 sec/ kg can be calculated using equation:
Fahraeus-Lindqvist effect in which the apparent viscosity
of blood is less than the bulk viscosity. The apparent Ap r 2
viscosity of a fluid through a cylindrical vessel under a lp = /33/
S 8tm
pressure diffference is known as the Poiseuilles law,
equation /11/.
The stokes equation for the plasma, which states the
This law states that the rate of blood flow is directly pressure-flow relationship in a capillary blood vessel:
proportional to the fourth power of the radius of the vessel,
which demonstrates that that the diameter of a blood vessel 1 p 2
is the most important factor in determining the rate of blood 0= + u i /34/
flow to a vessel. The units of this equation are kg /m sec [3, xi
31-33].
Since this is flow in the capillaries, it can be assumed that
The exchange of materials between blood and interstitial the Reynolds number and the Womersley number are less
fluid across the capillary membrane is regulated by three than 1, thus the acceleration term on the left side of the
specific mechanisms: (1) Hydrostatic pressure on each side equation can be set to zero. Here ui is the velocity vector. In
of the capillary wall, (2) Osmotic pressure in the plasma order to compute the flow in the vessel the pressure
and in the tissue fluid, and (3) The characteristics of the distribution and other stresses at the entry and exit sections
capillary wall (Figure 25). The pressure outside the need to be specified [6, 203].
capillary wall plays an important role in mass transport; the
flow of fluid across the capillary wall is driven by a Flow through a vessel is based on two factors: (1) Pressure
difference in pressure across this capillary wall. This difference and (2) Vascular resistance. Equation /11/
difference takes place due to the hydrodynamic effects, and shows that blood flow is directly proportional to the
the difference in osmotic pressure between the fluids pressure difference and inversely proportional to the
separated by the membrane [4, 8-9]. resistance [10, 146].

The following Starling equation represents the fluid flow Another phenomenon that takes place in the capillaries and
across the capillary membrane, and it can also be used to that is important to mention is diffusion. Diffusion is the
describe the flow across any permeable membrane. The most important mechanism for transferring nutrients and
units of this equation are mm Hg. waste products between blood and tissues. Lipid-soluble
substances such as oxygen and carbon dioxide diffuse
J
LpS
[( )(
= Pc Pif c if )] = P /32/
rapidly down their concentration gradients across the
lipophilic membranes. Water and water-soluble substances
(figure 26) such as ions and glucose pass through the
capillary wall at a slower rate, mainly through intercellular
In this equation: Lp stands for hydraulic conductance, J clefts and fenestrae [10, 164].
represents the volumetric fluid transfer rate, S is the total
circumferential surface area of the capillary wall, and P is MECHANISMS OF CAPILLARY EXCHANGE
the effective pressure. The first term in brackets is the
difference between hydrodynamic pressure between the Fluid, electrolytes, gases, and other substances transverse
capillaries and the interstitial fluid. If this term is positive the capillaries by different mechanisms: (1) Diffusion; (2)
fluid will leave the capillary, if it is negative it fluid will bulk flow; (3) Vesicular transport; and (4) Active
flow from the interstitial fluid into the capillary. transport. Diffusion is important for O2, CO2, and lipid-
soluble substances (a in Figure 27).

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-15
Figure 26. Diffusion of fluid molecules and dissolved Figure 29. Hydrostatic forces within the capillaries [29].
substances between the capillary and interstitial fluid
spaces [10, 164].
Capillary Fluid Exchange

Water, electrolytes, and small molecules are feely


exchanged between the intravascular and extravascular
compartments of the body. The primary site for this
exchange is the capillary. Several mechanisms are
involved in this exchange with bulk flow and diffusion
being the most important ones. The rate of exchange
(Figure 28), however, is determined by physical factors
Figure 27. Mechanisms of capillary exchange [27]. such as hydrostatic pressure and oncotic pressure [38].

HYDROSTATIC PRESSURE

The hydrostatic pressure is the principal force in filtration


across capillary walls. It consists of two opposing forces:
Capillary Hydrostatic Pressure (PC) and Tissue (Interstitial)
Hydrostatic Pressure (PT). Since PC >>PT, the net
hydrostatic pressure gradient across the capillary is positive
Figure 28. Forces that affect the rate of fluid exchange and fluid flows from the capillary into the interstitium [38].
across the capillary [38].
Capillary Hydrostatic Pressure, Pc
Ficks First Law of diffusion describes movement (or flux)
across capillaries: Capillary hydrostatic pressure is dependent on the arterial
pressure, the venous pressure, precapillary (artery and
dn = DA C arteriolar) resistance, and postcapillary (venules and small
dt X /35/ veins) resistance (Figure 29).

dn An increase in arterial or venous pressure results in an


Where: = flux (moles/sec) increase in capillary hydrostatic pressure while a decrease
dt
D = diffusion constant in each has the opposite effect. Likewise, an increase in
A = surface area arteriolar resistance results in a decrease in capillary
C = concentration difference pressure while an increase in venous resistance results in an
increase in capillary pressure. Thus, since the capillary
X = thickness of barrier to diffusion.
hydrostatic pressure is highest at the arteriolar end of the
capillary and lowest at the venular end, filtration is favored
Bulk flow of fluid and electrolytes occurs through the pores
at the arteriolar end and reabsorption is favored at the
and intercellular clefts (d,e,f in Figure 27). Vesicles can
venular end [29; 38; and 39]. Capillary hydrostatic
also fuse together creating pores across endothelial cells to
pressure [29] is determined by:
allow bulk flow of fluid across the capillary (c in Figure
25). Unlike bulk flow, vesicular transport involves the
Rv
translocation of macromolecules across capillary Pa + Pv
endothelium (b in Figure 27). Finally, in active transport, Ra
Pc = /36/
vascular endothelial cells take up molecules such as ions, Rv
1+
glucose, and amino acids by transport mechanisms [27]. Ra
Where: Pc = Capillary hydrostatic pressure.
Pa = Arterial pressure.

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-16
Pv = Venous pressure. Since C >> T, then the oncotic pressure difference, = C
Ra = Precapillary resistance (on the arterial side). - T, if unopposed by hydrostatic forces, would reabsorb
fluid from the interstitium into the capillary [29].
Rv = Postcapillary resistance (on the venous
side). BALANCE OF HYDROSTATIC AND ONCOTIC
PRESSURESTARLING HYPOTHESIS
Tissue (Interstitial) Hydrostatic Pressure, PT
The Starling hypothesis expounds on the relationship
This pressure is dependent on the interstitial fluid volume between hydrostatic pressure and oncotic pressure and their
and by tissue compliance. Normal value for PT is nearly role in regulating fluid passage across the capillaries. This
zero. However, during enhance filtration or lymphatic relationship can be expressed by the following equations:
blockage, PT dramatically increases as a result of increased
interstitial fluid volume. It results in a decrease in the J = k (P ) /38/
hydrostatic gradient across the capillary thereby limiting
filtration [38; 39].
[( ) (
J = k Pc Pi p i )] /39/
J = k (Pc Po ) /40/
ONCOTIC PRESSURE
Where: J = Rate of fluid movement.
There are also two opposing oncotic pressures influencing
k = Filtration constant for the capillary
fluid exchange: capillary plasma oncotic pressure (C) and
endothelium.
tissue (interstitial) oncotic pressure (T).
Pc = Capillary hydrostatic pressure.
Capillary Plasma Oncotic Pressure Pi = Interstitial fluid hydrostatic pressure.
= Reflection coefficient.
Since the capillary barrier is permeable to ions, the osmotic p = Plasma protein oncotic pressure.
pressure within the capillary is determined by plasma
proteins that are impermeable. This pressure, which is i = Intersititial fluid oncotic pressure.
generated by colloids, is referred to as capillary plasma ( )
Po = p i + Pi
oncotic pressure. Albumin is responsible for about 70% of
the oncotic pressure, which is typically at 25-30 mm Hg.
Along the length of the capillary especially along Filtration occurs when the algebraic sum of the hydrostatic
capillaries with high net filtration, C increases along its and oncotic pressures is positive while absorption occurs
length because the filtering fluid leaves behind proteins when it is negative. In an idealized capillary, filtration
thereby increasing concentration of these proteins. occurs at the arterial end while absorption occurs at the
venous end. However, this is not the case in most
Tissue (Interstitial) Oncotic Pressure capillaries, such as in the renal glomerulus, where filtration
has been observed over their entire lengths (Figure 30). In
This pressure depends on the interstitial protein other capillaries, such as in the intestinal mucosa, Starling
concentration and the reflection coefficient of the capillary forces change when intestinal epithelium actively pumps
wall such that increased permeability of the capillary water from the intestinal lumen into the interstitium,
thereby promoting absorption of fluid (Figure 31).
barrier to proteins has a positive effect on T. This pressure
is also dependent on the amount of fluid filtered into the
Rate of fluid movement across the capillary (J) also
interstitium such that an increase in capillary filtration
depends on the filtration constant of the capillary
decreases interstitial protein concentration and reduces T
endothelium (k):
[29; 38; and 39].
k = Lp A /41/
Oncotic pressure difference, = C - T, is determined by
the following equation:
Where: k = Capillary filtration constant.
= ( )RT (Ci Co ) /37/ Lp = Hydraulic conductivity of the endothelium.
A = Area of the capillary wall available for
Filtration.
Where: = Oncotic pressure difference.
= Reflection coefficient. Hydraulic conductivity of the endothelium is not affected
R = Gas constant. by arteriolar dilation nor is it affected by capillary
T = Absolute temperature. distension. However, capillary injury such as those due to
Ci = Solute concentration inside the capillary. toxins and severe burns increases capillary permeability
such that large amounts of fluid and protein leaks out of the
Co = Solute concentration outside the capillary.
capillaries into the interstitial space. Total capillary surface
available for filtration is dependent on the number of open
capillaries in a capillary bed [29].
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-17
Windkessel Model

The heart beats rhythmically and creates pressure and flow


wave forms within the arteries. When the heart contracts,
pressure within the left ventricle increases until it is greater
than the pressure in the aorta, leading to the opening of the
aortic valve. Blood flows out of the ventricle into the aorta,
and through the contraction cycle, both flow and pressure
decrease until pressure within the aorta becomes higher
than in the ventricle, leading to the closing of the aortic
valve. The pressure in the ventricle decreases rapidly as
the muscle relaxes; however, the decrease in pressure in the
aorta occurs at a much slower rate as the elastic vessel acts
as a reservoir. The Windkessel model (Figure 32) is a
linear mechanical model of circulatory system. The aorta
and large vessels are modeled as a compliant chamber,
while the peripheral vessels are modeled as rigid tubes.
According to the model:

P
V =CP and Qout = , /42/
R

Where: V, P, and C are the volume, pressure, and


Figure 30. When PC > PO, capillaries exhibit filtration over compliance in the compliant chamber; Qout is the outflow
their entire lengths [29]. of the chamber; R is the linear resistance of the rigid tube;
and Pvenous is assumed to be zero. Law of Conservation of
Mass requires:
dP P
Qin = C + /43/
dt R

Figure 32. (a) The Windkessel model of the aorta and


peropheral circulation. (b) Electrical analog of the
Windkessel model [12, 3.21].

Figure 31. When PC < PO, absorption of fluid occurs [29].

ANALOG MODELS OF BLOOD FLOW

The circulatory system consists of a complex network of


branching tubes, and it is capable of adjusting itself in
response to certain stimuli from different systems within
the body. Analog models provide a means to analyze Figure 33. Electrical analog for blood flow in a
single vessels, or even the entire cardiovascular system. compliant vessel [12, 3.22].

2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-18
It is possible to solve equation /46/ analytically with use of systems of algebraic equations that replace the
specific boundary conditions. Therefore, this equation can governing partial differential equations used normally to
be used to determine flow, pressure, compliance, or solve fluid dynamics problems. CFD creates a solution
resistance in the cardiovascular system [5, 170; 12, 3.21- based on these algebraic equations by using iterations, due
3.22; and 18, 125]. to the non-linearity of the governing equations [9]. By
using a process of discretization the algebraic equations are
Electrical Analog Models linked to the values of the dependent variables at the
adjacent point. One obstacle that is present in this type of
Electrical analogs are used to estimate blood flow trends, computational models is the generation of computational
and they result in the derivation of linear differential meshes that accurately resolve both the complex geometry
equations. In the electrical models, the vessels resistance and the physiologically flow features. Various studies have
to blood flow is represented by a resistor, the tube been conducted in the field of circulatory system by using
compliance by a capacitor, and the blood inertia by CFD this is due to the fact that the equations that are used
inductance. An electrical current can represent the blood to model the blood flow are non linear, complex and not
flow and the voltage can represent the blood pressure. The well constrained. Some of the studies done include: the
system can be modeled as shown in Figure 33. Using mechanism of aggregation of red blood cells in capillary
Kirchoffs law of currents, the governing differential vessels. A discrete-particle model in 3D was used to model
equation for the system becomes: the flow of plasma and red blood cells inside a capillary
tube. This particular work did not use classical CFD, but a
2 fluid particle model that can accurately measure scales
d P dP dQ from 0.001 to 100 m. Different models were done such as
+ RC +P=L + RQ /44/
2 dt dt sickle cell and normal [26]. Examples are shown in figures
dt
35 to 38.

The next study is based on arterial stenosis, a new method


of calculating coronary flow reserve (CFR) and fractional
flow reserve (FFR) based on pressure measurements
around the stenosed region was developed. In vitro studies
using CFD method was tested. The blood vessel was
modeled as a rigid tube, and the vascular bed as a porous
media. The blood was assumed to be a Newtonian fluid,
Figure 34. Lumped-parameter electrical analog and Navier-Stokes equations were solved by using the CFD
model of a multicompartment arterial segment. package FLUENT [20].
When vessels are attached in series as shown in Figure 34,
a more complex model is required to explain the system Now we shall present frequency dependence of dynamic
that is composed of numerous compartments, i = 1, 2, curvature effects on flow through coronary arteries. The
3etc. Kirchoffs law can be applied to each loop, entry to the tube was fixed but the radius of curvature was
resulting in: varied sinusoidally in time at two different frequencies (1
or 5 Hz). Computational studies were done on the flow
through a curved tube model of a coronary artery (Figure
dQi 35). From this model different studies on the time varying
Pi Pi 1 = Ri Qi + Li /45/
dt curvature on flow patterns that have been related to
arteriosclerosis were done. CFD software was used to
model this particular curved tube [16].
Vi
Pi = /46/ FLOW IN THE ASCENDING AORTA
Ci
Flow in the ascending aorta is complex, pulsatile, and
dVi three-dimensional. Kilner et al. (1993) and Bogren and
Qi Qi 1 = /47/ Bounocore (1999) visualized aortic flow patterns with a
dt Magnetic Resonance Velocity Mapping technique. They
found that the right-handed helical flow through the
Again, the system can be solved with the appropriate ascending aorta to the aortic arch was unstable.
boundary conditions. The RCL characteristics can be
derived from the specific properties of the blood and They attributed the skewed pattern to flow in a curved tube.
vessels [12, 3.22-3.23]. Both groups also noticed a retrograde flow in the inner
curvature of the ascending aorta during late systole and
COMPUTATIONAL FLUID DYNAMICS (CFD) diastole but could not speculate on the cause [11, 13]. A
combination of magnetic resonance imaging (MRI) and
Computational techniques are widely used for studying CFD was used by Jin et al (2001) to simulate flow patterns
hemodynamics in the circulatory system CFD involve the in the ascending aorta.
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-19
Figure 35. (a) Fluid particle model, (b) RBC model made
of flexible particles, (c) RBC model after volume Figure 37. Color-encoded wall shear rate in a curved
rendering, and (d) real image of deformed RBC [26]. bifurcation model of a coronary artery whose curvature is
being varied in time [16].

Figure 36. Computational flow model for a stenosed


vessel [20].

Their results revealed the following: (1) in contrast to the


previously reported retrograde flow in the inner curvature
of the ascending aorta by Kilner et al. (1993) and Bogren
and Bounocore (1999), a relatively steady, large vortex
during late systole and most of diastole was discovered and
(2) the skewing flow pattern seemed to be formed as a
Figure 38. Effects of the unsteady movement of the lumen
result of the acceleration movement of the ascending aorta
on flow patterns [11, 14].
rather than the geometric influence of curvature as
previously thought [11, 13-14].
RESEARCH ADVANCES
Unsteady movement of the lumen had a strong effect on the
Biologists at the California Institute of Technology have
flow patterns. Figure 38a shows velocity contours of MRI
uncovered an unsuspected but fundamental genetic rule
data at middle systole; 38b shows computed velocity
governing the formation of the cardiovascular system. The
contours of the moving wall model at same time and same
cells destined to form arteries and veins are already
location as 38a; 38c shows computed contours with the
genetically distinct at the earliest stages of blood vessel
wall radial expansion-contraction model; and 36d shows
formation in the embryo. It is likely that arteries and veins
contours of a rigid wall model. While Figures 36a and 36b
will differ in their expression of many other genes that have
show similar peak flow velocity positions during systole,
yet to be discovered. Such genes may lead to the
Figures 38c and 38d provide evidence of different pattern
development of new artery- or vein-specific drugs, or may
behavior [11, 14].
help to target known drugs specifically to either arteries or
veins. Such advances could potentially enhance the efficacy
or specificity of blood vessel-directed anticancer drugs
[30].
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-20
CONLUSIONS Fluid, electrolytes, gases, and other substances transverse
the capillaries by different mechanisms: (1) Diffusion; (2)
The cardiovascular and lymphatic systems work in unison Bulk flow; (3) Vesicular transport; and (4) Active
to form the human bodys circulatory system. Blood, the transport. The rate of exchange, however, is determined by
biofluid of the cardiovascular system, can be broken down physical factors such as hydrostatic pressure and oncotic
into two constituents, namely formed elements and plasma. pressure. The hydrostatic pressure is the principal force in
Fluid, electrolytes, respiratory gases, and leukocytes leave filtration across capillary walls. It consists of two opposing
and enter the blood at the level of the capillaries, and some forces: capillary hydrostatic pressure and tissue
of these excess components accumulate in the interstitial (interstitial) hydrostatic pressure. There are also two
fluid. It is the job of the lymphatic system to constantly opposing oncotic pressures influencing fluid exchange:
drain the excess interstitial fluid and return it to the venous capillary plasma oncotic pressure and tissue (interstitial)
blood, thereby maintaining the fluid balance within the oncotic pressure. The Starling hypothesis expounds on the
human body. It is because of this that the composition of relationship between hydrostatic pressure and oncotic
lymph closely resembles that of plasma, except for the pressure and their role in regulating fluid passage across the
abundance of proteins and lipids. Lymph has a lower capillaries. Filtration occurs when the algebraic sum of the
concentration of proteins and a higher concentration of hydrostatic and oncotic pressures is positive while
lipids than the blood plasma. absorption occurs when it is negative.

Blood is nonhomogeneous and anisotropic, and it contains The human circulatory system is a complex network of
a suspension of particles that display viscoelastic branching arteries, each with a different compliant nature
properties. All of this contributes to why blood does not and flow characteristics. The pressure and flow in the
obey Newtons linearized law of viscosity. Instead, bloods circulatory system is under the autonomic control, so
viscoelastic deformation can be modeled by the power law certain vessels can dilate and constrict, and the properties
or by its manipulation, the Casson equation. Many trends of the fluid can change over time. It is almost impossible
in the deformation properties of blood have been witnessed. to accurately model the circulatory system as it is not
It has been found that increasing shear rate, decreasing possible to include all known variables in a single model.
hematocrit, decreasing vessel size, and increasing If all the known variables were included, the model would
temperature all lead to a decrease in the apparent viscosity become complex and difficult to solve. Due to this, the
of blood. The Casson equation could be fitted to the approach adopted included a simplified model whose
experimental data between shear rate and shear stress; complexity was increased in order to provide a better
however, at high shear rates, blood tended to display more understanding of blood flow within the arteries. Initially,
linear, Newtonian qualities. In addition, increasing laminar flow in a rigid, cylindrical tube was investigated,
hematocrit percent led to increases in yield stress. and this gave rise to Poiseuilles formula for flow rate. The
results showed an inverse relationship between pressure
The flow in the venous system is dependent on the diameter and vessel radius and the direct relationship between flow
(or radius), length, and viscosity of the blood. Other rate and radius.
factors are the organization of the vascular network (series
and parallel arrangements), and extravascular mechanical Pulsatile flow, as is observed in the aorta and other large
forces acting upon the vasculature. In the venous system, arteries, was also considered. It was seen that as the
veins have less well-developed intima and media than Womersly parameter was increased, the velocity profile
comparable sized arteries. They represent a large became more blunt. Wave propagation in elastic arteries
distensible sink or reservoir for blood volume such that was calculated, and it was found that the speed of
50% of the circulating blood volume is in the systemic propagation was dependent on the pressure change with
venous system at rest. Most veins are 1-9 mm in diameter area.
with a well-developed adventitia. Small and medium sized
veins have valves to prevent backflow within their lumens. Analog models were used to understand the blood flow. In
a mechanical model, the Windkessel model, the flow was
found to be dependent of the pressure change with time, the
The valves consist of two folds of the tunica intima compliance of the elastic vessel, and the resistance of the
composed of elastic tissue and covered with endothelium peripheral vessels. In an electrical model using Kirchoffs
that project into the lumen. law of currents: pressure, flow, compliance, resistance, and
fluid inertial effects were each given electrical analogs.
To understand capillary fluid exchange requires knowledge Although it was possible to describe blood flow in the
of the capillary. There are three different types of arteries using mathematical, mechanical, and electrical
capillaries: (1) Continuous with the lowest permeability models, none of the models, however, could encompass all
consists of a basement membrane that is continuous and properties of the arteries and blood.
intercellular clefts that are tight; (2) Fenestrated with
relatively high permeability consists of perforations in the
endothelium; and (3) Discontinuous with extremely high
permeability consists of large intercellular gaps and gaps in
the basement membrane.
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-21
ACKNOWLEDGEMENTS 20. Shalman, E. 2001. Pressure-based simultaneous CFR
and FFR measurements: understanding the physiology
This study was aided in part by the teachings of Dr. James of a stenosed vessel. Computers in biology and
E. Moore Jr, Florida International University, Fluid medicine.
Mechanics Applications in Physiologic Systems. Our 21. Skofronich, J.G., Cameron J.R., and Grant, R.M.
thanks to Megh R. Goyal, University of Peurto Rico for 1999. Physics of the Body. 2nd edition.. Wisconsin :
reviewing this article. Medical Physics Publishing.
22. Stanely, A. B., Stroud, J.S. and Rayz, V.
REFERENCES Hemodynamic Simulations of Flow in Veins. 2001.
ASME International Mechanical Engineering
1. Aaronson, P.I. 1999. The Cardiovascular System at a Congress and Exposition Journal.
Glance. Massachusetts: Blackwell Science. 23. Tortora, G.J. and Grabowski, R.R. 1996. Principles of
2. Altman, P.L. 1961. Blood and Other Body Fluids. Anatomy and Physiology. New York: Harper Collins.
Washington, D.C.: Federation of American Societies 24. Van de Vosse, F. N., Gijsen, F. J. H., and Wolters, B.
of Experimental Biology, 1961. J. B. M. 2001. Numerical analysis of coronary artery
3. Enderle, J.D., Blanchard, S.M. and Bronzino, J.D. flow. 2001 Bioengineering Conference.
2000. Introduction to Biomedical Engineering. San 25. Wood, N.B. 1999. Aspects of Fluid Dynamics
Diego: Academic Press. Applied to Larger Arteries Journal of Theoretical
4. Fournier, R.L. 1998. Basic Transport Phenomena in Biology, 199:137-161.
Biomedical Engineering. Pennsylvania: Taylor & 26. Yuen, D.A. 2003. A discrete-particle model of blood
Francis. dynamics in capillary vessels. Journal of colloid and
5. Fung, Y.C. 1997. Biomechanics: Circulation. New interface science.
York: Springer-Verlag 27. http://cvphysiology.com/Microcirculation/M016.htm
6. Fung, Y.C. 1993. Biomechanics: Mechanical 28. http://cwx.prenhall.com/bookbind/pubbooks/martini7/
Properties of Living Tissues. New York: Springer- chapter20/medialib/CH20/html/ch2036.html.
Verlag. The Heart.
7. Fung, Y.C. 1990. Biomechanics- Motion, Flow, Stress, 29. http://human.physiol.arizona.edu/SCHED/CV/Baldwi
and Growth. New York: Springer-Verlag. n/Bald19/Baldwin.L19.html. Transcapillary Fluid
8. Geankoplis, C. J. 1993 Transport Processes and Unit Exchange: Starling Forces.
Operations 3rd Edition, New Jersey: Prentice Hall 30. http://pr.caltech.edu/media/lead/052998DA.html.
9. Gutierrez, Ronald. 2003. Personal communication. Tinidol, R. May 1998.
Miami-Fl: Tissue and Bioengineering Lab, Florida 31. http://scienceworld.wolfram.com/physics/Navier-
International University. StokesEquations.html. Navier-Stokes Equations.
10. Guyton, A.C. and Hall, J.E. 2000. Medical Wolfram Research: Scienceworld
Physiology. 10th edition. Philadelphia, PA: W.B. 32. http://thesis.library.drexel.edu/archive/00000084
Saunders Company. 33. http://thesis.library.drexel.edu/view/subjects/Non-
11. Jin, S., Oshinki, J. and Giddens, D. 2001. Effects of Newtonian%20fluids.html. Drexel University
inflow conditions and wall motion on flow in the Sangho, K. 2002. Study of Non-Newtonian viscosity
ascending aorta. 2001 Bioengineering Conference. and yield stress of blood in a scanning capillary-tube
12. Kutz, M. 2003. Standard Handbook of Biomedical rheometer.
Engineering Design. New York: McGraw-Hill. 34. http://www.coheadquarters.com/PennLibr/MyPhysiolo
13. Liepsh, D. 2002. An introduction to biofluid g/lect5/Tables5.01.htm. Hemodynamics
mechanicsbasic models and applications. Journal 35. http://www.cvphysiology.com/Hemodynamics/H001.h
of Biomechanics. tm. Hemodynamics.
14. Lightfoot, E. N. 1974 Transport Phenomena and 36. http://.www.cvrti.utah.edu/~macleod/bioen/be6010/no
Living Systems. New York: John Wiley & Sons Inc. tes/hemo-bw.pdf
15. Martini, F.H. 2001. Fundamentals of Anatomy and 37. http://www.jdaross.cwc.net/heart2. The Circulatory
Physiology. New Jersey: Prentice Hall. System. John Ross
16. Moore, J.E. 2001. Frequency dependence of dynamic 38. http://www.oucom.ohiou.edu/cvphysiology/
curvature effects on flow through coronary arteries. M011.htm. Cardiovascular Physiology Concepts:
Journal of biomechanical engineering. Capillary Fluid Exchange.
17. MountCastle, V.B. 1974. Medical Physiology. 13th 39. http://www.oucom.ohiou.edu/cvphysiology/
edition, Saint Louis,MO: C. V. Mosby Company. M012.htm. Cardiovascular Physiology Concepts:
18. Nichols, W.W. and ORourke, M.F. 1990. Hydrostatic and Oncotic Pressures.
McDonalds Blood Flow in Arteries: Theoretical,
Experimental, and Clinical Principles. London:
Hodder and Stoughton.
19. Nieweg, O.E., Essner, R., Reintgen, D.S. and
Thompaon, J.F. 2000. Lymphatic Mapping and Probe
Applications in Oncology. New York: Marcel Dekker
Inc.
2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-22
GLOSSARY OF TERMS Vein Any one of a series of blood vessels of the vascular
system that carries blood from various parts of the body
Abdominal aorta - The portion of the aorta in the back to the heart; returns oxygen-depleted blood to the
abdomen. heart.
Alveoli - Air sacs in the lungs where oxygen and carbon Ventricle (right and left) One of the two lower
dioxide are exchanged. chambers of the heart, either on the left or right.
Aorta - The largest artery in the body and the initial blood- Ventricular fibrillation A condition, in which the
supply vessel from the heart ventricles contract in a rapid, unsynchronized fashion.
Aortic valve - The valve that regulates blood flow from the When fibrillation occurs, the ventricles cannot pump blood
heart into the aorta. throughout the body.
Arterioles - Small, muscular branches of arteries. When
they contract, they increase resistance to blood flow, and
blood pressure in the arteries increases.
Artery - A vessel that carries oxygen-rich blood to the
body.

Blood Pressure The force or pressure exerted by the


heart in pumping blood; the pressure of blood in the arteries

Cardiac Pertaining to the heart


Cardiac output - The amount of blood the heart pumps
through the circulatory system in one minute.
Cardiology The study of the heart and its function in
health and disease
Cardiovascular (CV) Pertaining to the heart and blood
vessels. The circulatory system of the heart and blood
vessels is the cardiovascular system.
Circulatory System Pertaining to the heart, blood
vessels and the circulation of blood.
Coronary arteries - Two arteries arising from the aorta
that arch down over the top of the heart and divide into
branches. They provide blood to the heart muscle.

Inferior vena cava - The large vein returning blood from


the legs and abdomen to the heart.

Lipid - A fatty substance insoluble in blood.

Platelets One of the three types of cells found in blood;


they aid in the clotting of the blood

Pulmonary Referring to the lungs and respiratory system


Pulmonary valve The heart valve between the right
ventricle and the pulmonary artery. It controls blood flow
from the heart into the lungs.
Pulmonary vein The blood vessels that carry newly
oxygenated blood from the lungs back to the left atrium of
the heart.

Sodium - A mineral essential to life found in nearly all


plant and animal tissue. Table salt (sodium chloride) is
nearly half sodium.
Stenosis - The narrowing or constriction of an opening,
such as a blood vessel or heart valve.
Superior vena cava - The large vein that returns blood
from the head and arms to the heart.

Thrombosis - A blood clot that forms inside the blood


vessel or cavity of the heart.

Vascular - Pertaining to the blood vessels.


2003 Congress on Biofluid Dynamics of Human Body Systems at Biomedical Engineering - FIU, Miami - FL A-23

Potrebbero piacerti anche