Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Caledon (Georgetown, ON). Reference standards of linear NH4OH in water. The cartridge was then purged with air and
PFOS [13C4-MPFOS (MPFOS) and LPFOS] and a mixture of the analytes were eluted with 6 mL of methanol (1% NH4OH)
linear and branched isomers of PFOS (brPFOSK) were into 15-mL centrifuge tubes. The extracts were evaporated
obtained from Wellington Laboratories (Guelph, ON). A under N2 to 0.5 mL and transferred to methanol rinsed
sample of technical PFOS was purchased from TCI America polypropylene microvials (Fisher), with polyethylene caps
(Portland, OR). (Supelco). Analysis by HPLC-MS/MS involved 10-L injec-
Freeze-dried human serum samples A and B repre- tions of extract onto a FluoroSep RP Octyl column (3 100
senting 10.0 and 10.7 mL, respectively, of reconstituted , 15 cm 2.1 mm, ES Industries, West Berlin, NJ). Gradient
human serum were acquired through participation in the elution conditions were 200 L/min, and starting conditions
2006 Worldwide PFC intercalibration study (22). Human were 40% A (water adjusted to pH 4.0 with ammonium
Serum Sample A (NIST SRM 1589a) was a freeze-dried formate) and 60% B (100% methanol). Initial conditions were
standard reference material (SRM) collected by NIST held for 0.3 min, ramped to 64% B by 1.9 min, increased to
consisting of 50 pooled blood samples of 5 mL from donors 66% B by 5.9 min, 70% B by 7.9 min, 78% B by 40 min, 88%
who consumed fish caught around the Great Lakes. Human B by 42 min, 100% B by 55 min, returning to initial conditions
Serum Sample B (NIST SRM 1957) was a freeze-dried SRM by 60 min, and allowing 30 min for equilibration. Mass
also collected by NIST. This sample was prepared from a spectral data were collected using a hybrid triple-quadrupole
serum pool of 200 L across the United States. Detailed linear ion trap mass spectrometer (4000QTRAP, MDS Sciex,
information on both SRMs and certificates of analysis can Concord, ON) equipped with an electrospray interface
be found at www.nist.gov. operating in the negative ion mode. The MS/MS tune
LC/ESI-MS and LC/ESI-MS/MS. Separation, identifica- parameters utilized for quantification were optimized for
tion, and determination of the response factors of the PFOS isomer specific transitions using the brPFOSK standard (Table
isomers were completed using HPLC/ESI-MS and HPLC/ S1). Chromatograms were recorded by selected reaction
ESI-MS/MS experiments on a Waters Acquity Ultra Perfor- monitoring in MS/MS mode.
mance LC (Waters, Milford, MA) interfaced to a Micromass
Quattro micro atmospheric pressure ionization (API) mass Results and Discussion
spectrometer (Micromass, Manchester, UK). Chromatogra- Comparison of PFOS Quantification by Total and Isomer
phy was performed on an Acquity UPLC BEH Shield RP18 Specific Methods. Quantification biases were investigated
column (1.7 m, 2.1 100 mm) with gradient elution. The in serum samples by first generating chromatograms using
gradient started at 46% 80:20 MeOH/ACN and 54% water previously developed isomer specific methods (19). These
(both containing 10 mM ammonium acetate) at a flow rate were subsequently analyzed using three distinct standard
of 350 L/min. The program was ramped to 49% 80:20 MeOH/ curve quantification techniques: (i) PFOS quantification
ACN by 6 min, held for 17 min, then ramped again to 90% using m/z 499f99 and m/z 499f80 transitions and summing
80:20 MeOH/ACN over 0.50 min, held for 1 min, and returned the response of all isomers relative to the technical brPFOSK
to initial conditions. The mass spectrometer was set up in standard, (ii) quantification of linear and total branched
negative-ion electrospray mode with the following conditions: isomers separately using m/z 499f99 and m/z 499f80
capillary voltage 2.00 kV, source temperature 110 C, cone transitions relative to a LPFOS and brPFOSK standard,
gas 60 L/h, desolvation gas flow 750 L/Hr, desolvation gas respectively, and (iii) the linear and each of the major
temperature 350 C, cone voltage 60 V, collision energy 40 perfluoromonomethyl PFOS isomers were quantified indi-
eV, and gas cell pressure 3.5e-3 mbar. The MS tune vidually using separate calibration curves generated from
parameters for the individual branched isomers were op- the known isomer composition of brPFOSK (determined by
19
timized, however essentially no difference in their response F NMR) and isomer specific transitions (Table S1 and Figure
factors was observed. Reduction in the cone voltage did not S1). For each quantification technique, six-point calibration
reduce the in-source fragmentation observed for isomers 2, curves (each curve the average of standards run before and
9, 10, and 11. Optimization of MS/MS tune parameters for after serum samples) were employed using concentrations
isomer specific transitions using the individual isomers also ranging from 0.5 to 50 ng/mL total PFOS. All of these methods
did not result in a significant increase in response for any of were applied to two human serum samples that many
the isomers (see Supporting Information (SI) for further laboratories have in their possession following the second
details and Table S1). Worldwide Interlaboratory Study on PFCs.
Serum samples A and B were extracted in triplicate and An interesting result was the striking difference in the
analyzed using an isomer specific analytical method (19). branched isomer profiles of human serum samples A and B
Briefly, 1 mL of serum was diluted with 0.1 M formic acid (Figure 2). The HPLC-MS/MS chromatograms of these
containing MPFOS as the internal standard and sonicated samples qualitatively showed that each had a substantially
for 20 min. A 200 mg Oasis HLB cartridge (Waters) was different ratio of linear:branched isomers, and the relative
conditioned with HPLC grade methanol (6 mL) and 0.1 M profile of the branched isomers also varied. PFOS exposure
formic acid (6 mL). The samples were loaded onto the sources are not well characterized, thus the current state of
cartridge and slowly passed through under vacuum. The knowledge provides no explanation for these different isomer
cartridges were rinsed with 15 mL of 0.1 M formic acid, 6 mL profiles. Although limited to only two samples, these findings
of 50% 0.1 M formic acid/50% methanol, and 1 mL of 1% illustrate that PFOS exposure sources can vary and, irrespec-
VOL. 43, NO. 20, 2009 / ENVIRONMENTAL SCIENCE & TECHNOLOGY 9 7903
with brPFOSK. For simultaneous quantification of all of the
branched isomers (hereafter referred to as total branched
isomers) using brPFOSK, a calibration curve was constructed
by adjusting the concentration of the branched PFOS isomers
in each calibration standard to 21.2% of the total concentra-
tion based on 19F NMR spectroscopy data (supplied with the
standard). Quantification was achieved using both m/z
499f80 and 499f99 transitions (Table 1). Good agreement
was observed for the data obtained using the m/z 80 and
m/z 99 product ions for quantification of total branched
isomers (determined to not be significantly different using
a two-tailed students t test, p ) 0.32).
Finally, individual perfluoromonomethyl isomers were
quantified by building calibration curves for the individual
isomers using isomer specific transitions (Figure S1 and Table
1) after adjusting for their composition in the brPFOSK
standard. This can only be done for the PFOS isomers that
have been characterized by 19F-NMR. Utilizing the brPFOSK
standard, we were able to detect and quantify five perfluo-
romonomethyl isomers. The rank order of individual isomers
differed between the two samples, but the general rank order
was isomer 7 > 5 > 64 > 2 (Table 1). The sum concentration
of individual branched PFOS isomers is not a true total,
because some minor isomers are not accounted for, which
explains why the sum of all individual isomers was somewhat
less than that estimated by the total branched PFOS analysis
(Table 1). However, since the levels of perfluorodimethyl
isomers observed in the serum samples were lower than those
in the brPFOSK standard, any effect on the accuracy of the
FIGURE 2. HPLC-MS/MS chromatograms produced using the data resulting from the exclusion of the dimethyl branched
Benskin et al. method (19) showing the PFOS isomer profiles in
isomers during isomer specific quantification is expected to
(a) serum A, (b) serum B, and (c) the brPFOSK standard on the
be minimal; albeit this cannot necessarily be assumed for
FluoroSep RP Octyl column tracking m/z 499f80 (red), m/z
499f99 (green), and m/z 499f130 (blue). other human matrices or other organisms. It is also possible
that quantification of total branched isomers by summing
tive of improvements to accuracy, isomer specific analytical all branched peaks in the 499f80 transition may cause over-
methods could be a useful tool for exposure tracking in reporting because all but 3 isomers have relative response
biomonitoring studies. Nevertheless, the chromatograms of factors greater than 100% in this transition (Table 2).
neither human serum sample looked exactly like the brPFOSK Therefore, if laboratories undertake an isomer-specific
standard, thus some systematic analytical bias should be analysis, there may still be a benefit to reporting results from
anticipated from traditional PFOS methods. an accompanying total branched analysis.
Quantification of PFOS in the serum samples relative to Relative Response Factors of Branched PFOS Isomers.
brPFOSK, using either the m/z 499f80 or m/z 499f99 In an effort to understand the above analytical biases, the
transition, produced lower concentrations than the reference relative response factors of all PFOS isomers under electro-
values generated in the associated interlaboratory study (22) spray and MS/MS conditions were investigated using a more
(Table S2). It is notable that a pure linear standard (LPFOS) rapid UPLC-MS(/MS) method. The preparation and puri-
was provided for this interlaboratory study, however, its use fication of 10 different branched PFOS isomers (Figure 1)
was not made mandatory. Differences were also evident permitted the relative response factor of each to be studied,
between PFOS quantification with m/z 80 and m/z 99, and the determination of their elution order on a UPLC BEH
this depended on the sample. Specifically, the m/z 99 product Shield RP18 column (Figure 3), and hence the assignment of
ion produced a higher PFOS concentration than m/z 80 for the signals in various other technical PFOS mixtures. The
serum A (two-tailed students t test, p ) 0.006), whereas in elution order differed substantially from that previously
serum B the m/z 99 product ion produced a lower concen- observed using a perfluorophenyl (25) or the perfluorooctyl
tration relative to m/z 80 (p ) 0.003, Table 1). It is important (19) stationary phase used above. For example, on the UPLC
to note that the m/z 99 product ion is the most commonly Shield RP18 column the elution order was (in order of
used PFOS quantification ion due to endogenous interfer- increasing retention time) 9 < 10 < 8 < 11 < 6 < 5 < 4 < 7
ences associated with the m/z 80 product ion (19), however < 3 < 2 < 1, whereas on the perfluorophenyl column the
these interferences are chromatographically separated by order was 8-11 < 4 < 5 < 2 < 3 < 6 < 7 < 1, and on perfluorooctyl
the perfluorooctyl stationary phase used here and thus do it was 8-11 < 2 < 4 < 3 < 5 < 6 < 7 < 1. This highlights the
not contribute to the reported variation. importance of carefully identifying each isomer peak using
The linear PFOS isomer was then quantified using both isolated standards or CID patterns when undertaking an
m/z 80 and m/z 99 product ions, and LPFOS and brPFOSK isomer-specific analysis, rather than relying on elution orders
standards. The calibration curve for linear PFOS in the reported in the literature. This UPLC method is more rapid
brPFOSK standard was produced by simply adjusting the than existing isomer specific methods, however in our
calibrated mass of linear PFOS in the mixture to 78.8% of experience it had limited applicability for biological samples
the total mass based on 19F NMR spectroscopy data (supplied since it did not adequately separate endogenous steroid
with the standard). Regardless of the product ion (m/z 80 or interferences (11, 19). However, this UPLC method may be
m/z 99) chosen, or the standard used, the resulting con- very useful for abiotic samples such as water, air, or dust. It
centrations of linear PFOS were indistinguishable in both is also important to point out that, even with this highly
serum A and B (Table S3). This result was anticipated, and resolved method, not all branched isomers were baseline
it furthermore validates the 19F NMR spectral data provided separated from each other or the linear isomer. Specifically,
7904 9 ENVIRONMENTAL SCIENCE & TECHNOLOGY / VOL. 43, NO. 20, 2009
TABLE 1. Quantification of PFOS (ng/mL) in Human Serum A and B by the Traditional PFOS Method, a Total Branched Method,
and an Isomer Specific Method (All Methods Used the brPFOSK Technical Standard; All Data Were Collected in Selected
Reaction Monitoring Mode Using m/z 499 as the Precursor Ion (m/z Values in the Table Are Product Ions))
PFOS and total branched PFOS isomer specific analysis
PFOS a
PFOS b
total branched PFOS linear 7 6 5 4 2
sum of branched sum of all
m/z 99 80 99 80 80 80 130 330 130 419
isomers isomers
Serum A
mean 3.72 2.85 1.15 0.92 1.8 0.29 0.07 0.19 0.12 0.06 0.7 2.5
95% CI (0.53 (0.82 (0.85 (0.22 (0.7 (0.02 (0.04 (0.09 (0.03 (0.05 (0.13
Serum B
mean 18.1 21.1 9.6 9.5 10.6 3.58 1.06 1.44 0.82 0.39 7.3 17.9
95% CI (4.20 (4.89 (2.31 (2.54 (2.86 (0.80 (0.18 (0.31 (0.12 (0.14 (1.41
a b
Quantification by summing the areas of the m/z 80 product ions for all of the peaks. Quantification by summing the
areas of the m/z 99 product ions for all of the peaks.
TABLE 2. MS and MS/MS Relative Response Factors (%) for the Various PFOS Isomers Relative to Linear PFOS (Isomer 1)
(Uncertainty of All Measurements is 16%)
isomer 1 2 3 4 5 6 7 8 9 10 11
MS relative response factor for m/z 499 using electrospray ionization
100 40 106 80 101 109 91 84 69 73 74
In-source fragmentation (% ) 100 A/B)a
3.0 77 6 4.6 1.7 1.7 1.8 7.3 44 38 22
MS/MS relative response factors
m/z 499 f m/z 99 100 117 97 49 39 43 78 10 0 0 19
m/z 499 f m/z 80 100 0 78 135 241 142 123 113 118 220 90
a
A is the sum of the signals from the different channels set at m/z 80, 99, 130, 180, 230, 280, 330, 169, 219, 269, 419,
and B is equal to A + signal at m/z 499.
isomer 2 coeluted with linear PFOS (Figure 3), however tions using the m/z 499f99 or m/z 499f80 transitions (23),
quantification by using their unique product ions in MS/MS thus certain branched PFOS isomers can be missed entirely,
(m/z 419 and 80, respectively) allowed for their resolution and furthermore the two transitions will never yield the same
and accurate quantification. quantitative result in a PFOS analysis. The MS/MS differ-
The relative response factors for the branched PFOS ences appear to be due to some of the branched PFOS isomers
isomers under electrospray ionization were determined in preferring alternate CID pathways as discussed below. It
separate injections by UPLC-MS using MPFOS [13C4-PFOS] should also be noted that the isomers produced vastly
as an internal standard under selected ion monitoring different m/z 80:99 product ion ratios (Table 2), and this
conditions optimized for linear PFOS (1). In the selected ion factor will therefore contribute to systematic analytical bias
monitoring mode (m/z 499), isomers 2, 9, 10, and 11 showed when PFOS or total branched PFOS concentrations are
a substantially lower response (<75%) compared to the linear reported without an isomer specific analysis.
isomer (Table 2). The full scan spectra of these particular Preferential CID Pathways for the Branched PFOS
isomers confirmed considerable in-source fragmentation Isomers. Careful examination of the product ion mass spectra
which largely explained the reduction in signal strength of the branched PFOS isomers revealed interesting details
observed for m/z 499 (Figure S2). As an example, the full concerning preferential CID pathways. As expected, isomer
scan spectrum of isomer 2 gave fragments at m/z 99, 169, 1 fragmented mainly to m/z 80 ([SO3]-) and m/z 99 ([SO3F]-).
219, and 419. The cumulative signal count for these four However, the preferred CID pathways for some of the
fragments largely accounted for the 60% reduction in signal branched isomers can be very different from that of 1 (19, 25).
response for m/z 499 (Table 1). These results indicate that For example, isomer 2 fragmented through the perfluoroalkyl
in-source fragmentation can be a major contributor to route to give the 9-series ions (25) and m/z 99, however,
analytical bias in PFOS methods where it is an inherent m/z 80 was totally absent from the mass spectrum. It appears,
assumption that all isomers behave similarly. Branched in this case, that the primary CID pathway involved the loss
isomers exhibiting lower response factors (isomers 2, 4, 9, of SO3 with the charge being retained by the perfluoroalkyl
10, and 11) have been observed to vary in amounts present fragment m/z 419 [CF3(CF2)7]- instead of the m/z 80 [SO3]-
(range of between 3 and 10%) in technical PFOS samples fragment. The perfluoroalkyl fragment (m/z 419) then
(27). The overall bias will depend on the tuning parameters underwent secondary fragmentation to give the expected
of the instrument. fragments at m/z 119, 169, 219, and 269 (26). This observation
The relative response factors for the branched PFOS was explained by the formation of more stable secondary
isomers (to the linear isomer) were then determined by UPLC- perfluoroalkyl anions after primary CID. This pathway was
MS/MS using MPFOS [13C4-PFOS] as an internal standard not seen for the linear isomer 1 since loss of SO3 would have
and MS/MS conditions optimized for linear PFOS (1): only led to a much less stable linear primary perfluoroalkyl anion,
the m/z 499f99 and 499f80 transitions were monitored for therefore, the charge preferred to stay on the SO3 fragment.
native PFOS (Table 2). Isomers 9 and 10 did not fragment Some of the isomers also produced perfluoroalkyl fragments,
to m/z 99, while isomer 2 did not fragment to m/z 80. PFOS for example, isomer 9 fragmented to produce both the 9-
quantification is typically performed under MS/MS condi- series ([CF3(CF2)n]-) and 0-series ([CF2(CF2)nSO3]-) frag-
VOL. 43, NO. 20, 2009 / ENVIRONMENTAL SCIENCE & TECHNOLOGY 9 7905
FIGURE 3. Chromatograms of the individual PFOS isomers and brPFOSK obtained in SIM mode (m/z 499) on a Waters Acquity UPLC
using a Shield RP18 column at a concentration of 500 ng/mL. Compounds in brackets are impurities in the sample.
ment ions (25) while isomers 10 and 11 only produced the distribution between the various sources of commercial
0-series ions. technical PFOS (27) and the aforementioned differences in
The consequence of these preferred CID pathways on the response factors of the branched isomers. Under our
quantitative data is dependent on the transitions chosen experimental conditions it was possible to calculate the
during selected reaction monitoring analyses. Transitioning overall relative response factors (RRFs) to the linear isomer
to only the 80 or 99 product ions during MS/MS analysis for commercial PFOS standards, which is relevant in a PFOS
could result in the exclusion of some PFOS isomers from analysis. This involves knowing the MS/MS RRFs for the most
reported findings. This is of importance since these isomers abundant PFOS isomers (Table 2) and the isomer content of
may possess unique biological properties. For example, the several standards as measured by 19F NMR spectroscopy
alpha-PFOS branched isomer (2) is more bioaccumulative (Table S4) (27). Table S4 summarizes examples in which the
than linear PFOS in rats (28, 29) but will not be quantified theoretical and measured RRFs for both the m/z 499f99
when using the m/z 80 product ion as it does not produce and m/z 499f80 transitions were calculated and measured
this ion. Acquiring and analyzing accurate isomer patterns for brPFOSK and a commercial PFOS sample (TCI America).
could also give clues as to the exposure sources and In both cases the measured values were higher than the
environmental fate of perfluorinated substances. calculated values, but they are generally in good agreement.
Influence of Pure Linear or Technical Calibration The variation between the calculated and measured values
Standards. Even if a technical PFOS sample is chemically was larger for brPFOSK compared to the TCI sample, but
pure (i.e., just branched and linear PFOS), the use of such this may be due to the presence of other isomers that were
material as a reference standard will introduce interlaboratory not accounted for in the theoretical calculations. The results
data variability because of the significant differences in isomer highlight that laboratories using a PFOS method and
7906 9 ENVIRONMENTAL SCIENCE & TECHNOLOGY / VOL. 43, NO. 20, 2009
quantifying with a linear standard using the m/z 499f99 (7) Fromme, H.; Schlummer, M.; Moller, A.; Gruber, L.; Wolz, G.;
transition may underestimate the true PFOS concentration Ungewiss, J.; Bohmer, S.; Dekant, W.; Mayer, R.; Liebel, B.;
because the RRFs for the branched isomers are lower than Twardella, D. Exposure of an adult population to perfluorinated
substances using duplicate diet portions and biomonitoring
for linear PFOS in this transition. Conversely, only using the data. Environ. Sci. Technol. 2007, 41, 79287933.
m/z 499f80 transition and a linear PFOS standard would (8) Apelberg, B. J.; Goldman, L. R.; Calafat, A. M.; Herbstman, J. B.;
overestimate PFOS content in a real sample because the RRFs Kuklenyik, Z.; Heidler, J.; Needham, L. L.; Halden, R. U.; Witter,
for the branched isomers are higher than that for linear PFOS. F. R. Determinants of fetal exposure to perfluorinated com-
This is based on our instrument conditions and may be pounds in Baltimore, Maryland. Environ. Sci. Technol. 2007,
41, 38913897.
instrument dependent. The actual difference between labo-
(9) Calafat, A. M.; Kuklenyik, Z.; Reidy, J. A.; Caudill, S. P.; Tully,
ratories analyzing the same sample will vary, thus making it J. S.; Needham, L. L. Serum concentrations of 11 polyfluoroalkyl
difficult to compare interlaboratory data. To limit ambiguity compounds in the U.S. population: Data from the national
when comparing data generated by different laboratories, health and nutrition examination survey (NHANES) 1999-2000.
we suggest that a common standard should be used in all Environ. Sci. Technol. 2007, 41, 22372242.
laboratories so that interlaboratory data can be more easily (10) Karrman, A.; Ericson, I.; van Bavel, B.; Darnerud, P. O.; Aune,
M.; Glynn, A.; Lignell, S.; Lindstrom, G. Exposure of perfluori-
compared.
nated compounds through lactation: levels of matched human
The use of linear PFOS (1) as the reference standard for milk and serum and a temporal trend, 1996-2004, in Sweden.
PFOS analyses would remove the unknown variable of Environ. Health Perspect. 2007, 115, 226230.
branched isomer content. Consistency between analytical (11) Ehresman, D. J.; Froehlich, J. W.; Olsen, G. W.; Chang, S.-C.;
laboratories would probably improve, however accuracy Butenhoff, J. L. Comparison of human blood, plasma, and serum
would remain an issue since no environmental or human matrices for the determination of perfluorooctanesulfonate
(PFOS), and perfluorooctanoate (PFOA), and other fluoro-
sample, to our knowledge, will contain only linear PFOS.
chemicals. Environ. Res. 2007, 103, 176184.
For any PFOS analysis, unless a standard of exactly the (12) Tittlemier, S. A.; Pepper, K.; Seymour, C.; Moisey, J.; Bronson,
same isomer composition as the unknown sample is used, R.; Cao, X.-L.; Dabeka, R. W. Dietary exposure of Canadians to
the isomer specific RRFs will always lead to a quantification perfluorinated carboxylates and perfluorooctane sulfonate via
bias of unknown proportions. However, any technical PFOS consumption of meat, fish, fast foods, and food items prepared
standard will always more closely resemble a sample, thus in their packaging. J. Agric. Food Chem. 2007, 55, 32033210.
(13) Karrman, A.; Langlois, I.; van Bavel, B.; Lindstrom, G.; Oehme,
minimizing bias, and if the linear and branched isomers
M. Identification and pattern of perfluorooctane sulfonate
have been quantified by 19F NMR spectroscopy, it can also (PFOS) isomers in human serum and plasma. Environ. Int. 2007,
be used for an isomer specific quantitative analysis. 33, 782788.
Therefore, until quantitative standards of the branched (14) Butenhoff, J. L.; Olsen, G. W.; Pfahles-Hutchens, A. The
PFOS isomers are available, this approach provides an applicability of biomonitoring data for perfluorooctanesulfonate
improvement in the methodology relating to the quan- to the environmental public health continuum. Environ. Health
Perspect. 2006, 114, 17761782.
tification of total PFOS.
(15) Reagen, W. K.; Ellefson, M. E.; Kannan, K.; Giesy, J. P. Comparison
of extraction and quantification methods of perfluorinated
Supporting Information Available compounds in human plasma, serum, and whole blood. Anal.
Tables summarizing the optimization of MS/MS tune Chim. Acta 2008, 628, 214221.
parameters for isomer specific transitions, a comparison (16) 3M Company. Fluorochemical Isomer Distribution by 19F NMR
Spectroscopy; US EPA Public Docket AR-226-0564, 1997.
of our data to that obtained in the 2nd Worldwide (17) Arsenault, G.; Chittim, B.; McAlees, A.; McCrindle, R. Separation
Interlaboratory Study, the linear PFOS concentrations and fluorine nuclear magnetic resonance spectroscopy (19F-
obtained using the brPFOSK versus the LPFOS standard, NMR) analysis of the individual branched isomers present in
and the measured versus calculated response factors of technical perfluorooctanesulfonic acid (PFOS). Organohalogen
the linear and branched PFOS isomers. Figures illustrating Compd. 2005, 67, 818821.
the chromatograms obtained during isomer specific (18) Vyas, S. M.; Kania-Korwell, I.; Lehmler, H.-J. Differences in the
isomer composition of perfluorooctanesulfonyl (PFOS) deriva-
quantification and expressing the inverse association of tives. J. Environ. Sci. Health Part A 2007, 42, 249255.
isomer response factors with in-source fragmentation. This (19) Benskin, J. P.; Bataineh, M.; Martin, J. W. Simultaneous
material is available free of charge via the Internet at http:// characterization of perfluoroalkyl carboxylate, sulfonate, and
pubs.acs.org. sulfonamide isomers by liquid chromatography-tandem mass
spectrometry. Anal. Chem. 2007, 79, 64556564.
(20) Martin, J. W.; Kannan, K.; Berger, U.; De Voogt, P.; Field, J.;
Literature Cited Franklin, J.; Giesy, J. P.; Harner, T.; Muir, D. C. G.; Scott, B.;
(1) Ishibashi, H.; Iwata, H.; Kim, E.-Y.; Tao, L.; Kannan, K.; Amano, Kaiser, M.; Jarnberg, U.; Jones, K. C.; Mabury, S. A.; Schroeder,
M.; Miyazaki, N.; Tanabe, S.; Batoev, V. B.; Petrov, E. A. H.; Simcik, M.; Sottani, C.; Van Bavel, B.; Karrman, A.; Lindstrom,
Contamination and effects of perfluorochemicals in Baikal Seal G.; Van Leeuwen, S. Analytical challenges hamper perfluoroalky
(Pusa sibirica). 1. Residue level, tissue distribution and temporal research. Environ. Sci. Technol. 2004, 38, 248A255A.
trend. Environ. Sci. Technol. 2008, 42, 22952301. (21) Yoo, H.; Jones, P. D.; Bradley, P. W.; Guzvic, M.; Upham, B. L.;
(2) Kim, S.-K.; Kannan, K. Perfluorinated acids, in air, rain, snow, Trosko, J. E.; Newsted, J. L.; Giesy, J. P. Separation and
surface runoff and lakes: Relative importance of pathways to characterization of structural isomers of perfluorinated com-
contamination of urban lakes. Environ. Sci. Technol. 2007, 41, pounds. In Proceedings of the 2005 International Fluoros
83288334. Symposium: An international Symposium on Fluorinated Alkyl
(3) Van de Vijver, K. I.; Holsbeek, L.; Das, K.; Blust, R.; Joiris, C.; De Organics in the Environment; Toronto, Ontario, August 18-20,
Coen, W. Occurrence of perfluorooctane sulfonate and other 2005; Mabury, S., Ed.; University of Toronto: Ontario, Canada;
perfluorinated alkylated substances in harbor porpoises from Abstract TOX009.
the black sea. Environ. Sci. Technol. 2007, 41, 315320. (22) Lindstrom, G.; Karrman, A.; van Bavel, B. Accuracy and precision
(4) Butt, C. M.; Muir, D. C. G.; Stirling, I.; Kwan, M.; Mabury, S. A. in the determination of perfluorinated chemicals in human
Rapid response of Arctic ringed seals to changes in perfluoroalkyl blood verified by interlaboratory comparisons. J. Chromatogr.
production. Environ. Sci. Technol. 2007, 41, 4248. A 2009, 1216, 394400.
(5) Hauks, M.; Berger, U.; Hop, H.; Gulliksen, B.; Gabrielsen, G. W. (23) de Voogt, P.; Saez, M. Analytical chemistry of perfluoroalkylated
Bioaccumulation of per- and polyfluorinated alkyl substances substances. Trends Anal. Chem. 2006, 25, 326342.
in selected species from the Barents sea food web. Environ. (24) Villagrasa, M.; de Alda, M. L.; Barcelo, D. Environmental analysis
Pollut. 2007, 147, 360371. of fluorinated alkyl substances by liquid chromotagraphy-
(6) Powley, C. R.; George, S. W.; Russell, M. H.; Hoke, R. A.; Buck, (tandem) mass spectrometry: a review. Anal. Bioanal. Chem.
R. C. Polyfluorinated chemicals in a spatially and temporally 2006, 386, 953972.
integrated food web in the western arctic. Chemosphere 2007, (25) Langlois, I.; Oehme, M. Structural identification of isomers
70, 664672. present in technical perfluorooctane sulfonate by tandem mass
VOL. 43, NO. 20, 2009 / ENVIRONMENTAL SCIENCE & TECHNOLOGY 9 7907
spectrometry. Rapid Commun. Mass Spectrom. 2006, 20, 844 (28) Benskin, J. P.; De Silva, A. O.; Martin, L. J.; Arsenault, G.;
850. McCrindle, R.; Riddell, N.; Mabury, S. A.; Martin, J. W.
(26) Arsenault, G.; McAlees, A.; McCrindle, R.; Riddell, N. Analysis Disposition of Perfluorinated Acid Isomers in Sprague-Dawley
of perfluoroalkyl anion fragmentation pathways for perfluo- Rats; Part 1: Single Dose. Environ. Toxicol. Chem. 2009, 28 (3),
roalkyl carboxylates and sulfonates during liquid chromatog- 542554.
raphy/tandem mass spectrometry: evidence for fluorine mi- (29) De Silva, A. O.; Benskin, J. P.; Martin, L. J.; Arsenault, G.; McCrindle,
gration prior to secondary and tertiary fragmentation. Rapid R.; Riddell, N.; Martin, J. W.; Mabury, S. A. Disposition of
Commun. Mass Spectrom. 2007, 21, 38033814. Perfluorinated Acid Isomers in Sprague-Dawley Rats; Part 2:
(27) Arsenault, G.; Chittim, B.; McAlees, A.; McCrindle, R.; Riddell, Subchronic Dose. Environ. Toxicol. Chem. 2009, 28 (3), 555
N.; Yeo, B. Some issues relating to the use of perfluorooctane 567.
sulfonate (PFOS) samples as reference standards. Chemosphere
2008, 70, 616625. ES901261V
7908 9 ENVIRONMENTAL SCIENCE & TECHNOLOGY / VOL. 43, NO. 20, 2009