Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
available at www.sciencedirect.com
a r t i c l e i n f o a b s t r a c t
Article history: Objectives. The aim of the study was to characterize the chemical composition, microstruc-
Received 27 May 2009 ture and roughness of two commercially available zirconia dental implants (WhiteSky and
Received in revised form Zit-Z).
27 September 2009 Methods. The chemical composition of the cervical collar and threaded root parts of the
Accepted 17 November 2009 implants (n = 2) were studied by XPS and HV-EDX. LV-SEM was used for morphological
assessment, Raman microanalysis for microstructural characterization and optical pro-
filometry for surface roughness measurements. XRD, HV-EDX and Raman microanalysis
Keywords: of bulk regions (longitudinal sections) were used as reference.
Zirconia dental implants Results. XPS showed the presence of C, O, Zr and Y (collar) plus Al (root) at implant surfaces.
Surface analysis More C (10–26 at%) and a lower Al/Zr ratio were found in WhiteSky (1.05 vs 1.26 in Zit-Z). Zr,
XPS Y and Al were detected in single, fully oxidized states. The same elements, plus Hf, were
Raman microanalysis identified by HV-EDX at bulk and surface regions, with a Al/Zr ratio higher in WhiteSky
SEM/EDX (0.17 vs 0.09 in Zit-Z). Na, K and Cl contaminants were traced at implant root parts by both
Optical profilometry methods. XRD analysis of cross-sectioned specimens revealed the presence of monoclinic
and tetragonal zirconia along with cubic yttria phases. Raman microanalysis showed that
the monoclinic zirconia volume fraction was higher at root surfaces than the collar. No
monoclinic phase was found at bulk regions. Significantly higher Sa and Sq values were
recorded in WhiteSky than Zit-Z, whereas Zit-Z showed higher Rt value.
Significance. The differences found between the implants in the extent of carbon contami-
nation, residual alumina content, tetragonal to monoclinic ZrO2 phase transformation and
3D-roughness parameters may contribute to a substantial differentiation in the cellular and
tissue response.
© 2009 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
∗
Corresponding author at: Department of Biomaterials, University of Athens, School of Dentistry, 2 Thivon Str-Goudi, 115 27 Athens,
Greece. Tel.: +30 210 7461101; fax: +30 210 7461306.
E-mail address: geliad@dent.uoa.gr (G. Eliades).
0109-5641/$ – see front matter © 2009 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.dental.2009.11.079
296 d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305
long-term human trials to establish their clinical success are 2.2. Low-vacuum scanning electron microscopy and
still missing [10]. high vacuum X-ray energy dispersive microanalysis
From the material standpoint, concerns have been (LV-SEM/HV-EDX)
expressed on the low temperature degradation of the tetrag-
onal to monoclinic ZrO2 phase, that has been associated with The same specimens analyzed by XPS were subjected to
in-service failures of orthopaedic implants [11,12]. Treatments LV-SEM imaging. Secondary electron images were acquired
to avoid or reverse such transformations have been advocated employing a large field detector (LFD) attached to a SEM
during manufacturing or sterilization [13]. Moreover, it has unit (Quanta 200, FEI, Hillsboro, OR, USA) operated at 30 kV
been shown that the osseointegration capacity of machined accelerating voltage, 90 A beam current, 1 Torr pressure
ZrO2 surfaces is substantially increased after modification (without electron conductive coating to avoid masking off
by Al2 O3 sandblasting [14]. Although such techniques have microstructural features) at 40×, 300×, 600× and 2400×
already been adopted in some products, there is lack of infor- magnifications.
mation on the surface chemistry, structure and morphology The elemental composition of cervical collar, threaded
of currently commercially available ZrO2 dental implants root and bulk regions (the latter prepared for XRD analy-
that would facilitate understanding of the implant–bone sis as described below) of each implant were determined
interactions and the bone augmentation mechanisms by HV-EDX analysis (this technique provides better accuracy
involved. in quantitative determinations than LV-EDX). All the spec-
The aim of the present study was to investigate the surface imens were coated with a thin layer of conductive carbon
chemistry, morphology and structure of two commercially in a sputter-coating unit (SCD 004 Sputter-Coater with OCD
available zirconia implants. The testing hypothesis was that 30 attachment, Bal-Tec, Vaduz, Liechtenstein). EDX analy-
significant differences exist in these properties between the sis was performed using a liquid N2 -cooled Si(Li) detector
two implants. with super ultra-thin Be window (Sapphire SUTW+ CDU,
EDAX Int, Mahwah, NJ, USA) attached to the SEM unit under
the following conditions: 30 kV energy range, 10−6 Torr pres-
2. Materials and methods sure, 110 A beam current, 128 eV resolution, 250 s acquisition
time, 210 m × 210 m sampling window and 28–34% detector
The products tested were WhiteSky (Ø: 3.5 mm, l: 12 mm, dead time. The depth of analysis was estimated as to 1 m.
lot 59235, Bredent Medical, Senden, Germany) and Zit-Z (Ø: The quantitative analysis was performed in non-standard
3.5 mm, l: 13 mm, lot 49744, Ziterion, Uffenheim, Germany). mode using ZAF and coating corrections employing Gene-
Five specimens from each product were subjected to the fol- sis v. 5.2 software (EDAX, Int). Elemental mapping of regions
lowing testing procedures: of interest was based on compositional backscattered elec-
tron images, obtained with a solid-state detector (SSD) at
2.1. X-ray photoelectron spectroscopy (XPS) the same conditions as above, but at 15 kV accelerating
voltage.
The samples as received (n = 2) were placed in the ultra-high
vacuum chamber of an X-ray photoelectron spectrome- 2.3. Raman microanalysis
ter (Scienta 300 ESCA system, NCESS, Cheshire, UK). Two
regions, one at the cervical collar and the other at the The cervical collar, threaded root parts bulk regions (pre-
threaded root portion, were located on each implant and pared for XRD analysis as described below) of the implants
analyzed under the following conditions: Al Ka (1486.7 eV) (n = 2) were analyzed by Raman microscopy to identify and
monochromated rotating anode, 14 kV accelerating voltage, map the distribution of the tetragonal (145 and 262 cm−1
0.2 A current emission, 2.8 kW maximum power, 10−9 mbar Raman shift) and monoclinic (180 and 190 cm−1 Raman
pressure, ∼90◦ electron take-off angle, 0.30 eV energy reso- shift) ZrO2 phases at the surface region [15]. A Raman
lution and 6 mm × 0.5 mm sampling area. An electron flood microscope was used (LabRAM Aramis, Horiba Jobin-Yvon,
gun operated at 4 eV was used for charge compensation. Villeneuve d’Ascq, France) operated under the following
Energy calibration was performed based on Ag3d5/2 peak stan- conditions: Ar laser (532 nm), 10 mW power at sample,
dard. Survey scans (150 eV pass energy) were taken from 50× LWD objective, 1800 grit/mm grating, 1000 m confo-
each region to identify the elements present on the surface. cal hole (defocused mode), 100 m slit, 10 s acquisition.
Then, high resolution narrow scans (150 eV pass energy) were Three regions were randomly located at each implant sur-
recorded over the predominant peaks and the elemental bind- face location (collar/root) with the optical system of the
ing states were determined. All spectra were aligned on the microscope (10× optical objective) and analyzed. For map-
binding energy scale to the C 1s peak arising from adven- ping of the monoclinic ZrO2 phase, 20 m × 35 m areas
titious hydrocarbons (–CH2 –, 285 eV BE). The core level data were selected and scanned at 5 m steps and 5 s acquisition
were analyzed using the CASA XPS software package. A Shirley time.
background was subtracted from the data and 80:20 Gaus- From the net peak height intensities of the monoclinic and
sian:Lorenzian peaks fitted to give binding energy positions. tetragonal ZrO2 phases, the percentage volume of the mono-
The fitted peak areas in conjunction with relative inten- clinic phase (Vm %) was calculated according to the equation:
sity factors allowed the elemental ratios at the surface to Vm % = {I180 + I190 /0.97(I145 + I262 ) + I180 + I190 } × 100, where I the
be quantified. The depth of analysis was estimated as to net peak height intensities at the corresponding Raman shifts
3 nm. [15].
d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305 297
2.4. X-ray diffraction (XRD) ley height) was performed by Veeco-Vision software at 40×
magnification (160 m × 120 m).
In order to examine the presence of the tetragonal and mon-
oclinic ZrO2 phases in the entire implant, one specimen from 2.6. Statistical analysis
each implant was embedded in epoxy resin and sectioned at
a longitudinal direction using a microtome under continuous The results of elemental at% obtained from XPS and EDX were
water cooling. The specimens were ground to a smooth sur- subjected to a two-way ANOVA (implant type and region as
face using SiC paper up to 1200 grit size, polished with a 3 m independent variables, and at% per element as dependent
diamond paste and ultrasonically cleaned in a distilled water variable), whereas a two-way ANOVA was used for Vm % com-
bath for 3 min. The sections were studied in an X-ray diffrac- parisons (implant type and region as independent variables).
tometer (D8 Focus, Bruker AXS, Karlsruhe, Germany) under the A Tukey’s test was used for pairwise multiple comparisons.
following conditions: CuKa anode, 40 kV accelerating voltage, Finally, the differences in the roughness parameters between
40 mA beam current, 20–90◦ 2 range, 0.02◦ step, 2 s exposure the implants were evaluated by a t-test. In all comparisons
per step. The identification of phases was based on the ICSD a 95% confidence level was used (˛: 0.05). Statistical analysis
database [16]. was performed by SigmaStat software (Jandel, St. Raphael, CA,
USA).
2.5. Optical profilometry
Fig. 1 – XPS spectra of implant surfaces.(a) Survey spectrum of WhiteSky; (b) Survey spectrum of Zit-Z; High resolution
spectra of Zr 3d (c) and Y 3d peaks (d).
298 d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305
Fig. 2 – Low-vacuum secondary electron images of the implant surfaces. WhiteSky: (a) Low magnification view (40×, bar:
1 mm); (b) cervical collar-threaded root transitional zone (300×, bar: 100 m); (c) threaded root (600×, bar: 50 m); (d)
threaded root at high magnification (2400×, bar: 20 m). Zit-Z: (e) Low magnification view (40×, bar: 1 mm); (f) cervical
collar-threaded root transitional zone (300×, bar: 100 m); (g) threaded root (600×, bar: 50 m); (h) threaded root at high
magnification (2400×, bar: 20 m).
d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305 299
Table 1 – XPS elemental composition of the Zirconia implant surfaces (mean and sd in parentheses)* .
Element (%) WhiteSky Zit-Z
The implant surfaces have similar qualitative composition. listed in Table 1, ranged from 0.05 to 0.07 (p > 0.05), with the
Al was identified only at threaded root regions. WhiteSky exception of Zit-Z collar, where the highest value of 0.15 was
demonstrated more C contamination and less O content at recorded (p < 0.05). The Al/Zr atomic ratios for WhiteSky and
the cervical collar. The highest Zr content was found at Zit-Z Zit-Z were 1.05 and 1.27 respectively (p < 0.05). The Zr 3d and
collar. The XPS Y/Zr atomic ratios, as derived from the data Y 3d peaks showed only one doublet arising from 3d5/2 and
Fig. 3 – High-vacuum EDX spectra of the implants tested along with the elemental distributions of the dispersed phases at
threaded root regions. (a) Spectra of cervical collar for WhiteSky; (b) Spectra of threaded root for Zit-Z; Al X-ray maps (green)
imaged onto the corresponding compositional backscattered electron images for WhiteSky (c) and Zit-Z (d); X-ray maps of
Al–O–C for WhiteSky (e) and of Al–Na–K–C for Zit-Z (f) imaged onto the corresponding backscattered electron images.
300 d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305
Table 2 – EDX elemental composition of the Zirconia implants (mean and sd in parentheses)* .
Element (%) WhiteSky Zit-Z
3d3/2 spin orbit splitting, indicating that Zr and Y exist in one The results of the percentage volume of the monoclinic phase
oxidation state in the implants (Fig. 1c and d). The binding (Vm %) are presented in Table 3. No interaction was found
energy of the Zr 3d5/2 peak at 182.4 eV corresponds to ZrO2 , between the independent variables. The threaded root parts
whereas that of theY 3d5/2 peak at 156.8 eV to Y2 O3 . For Al, the demonstrated the highest Vm % (p = 0.05). No monoclinic phase
binding energy of the Al 2p peak at 74.68 eV corresponds to was detected at bulk regions. Raman mapping showed that at
Al2 O3 . Traces of Na were also identified (Na 1s at 1023 eV and cervical collars the monoclinic phase distribution was related
Na Auger at 499 eV) in both implants. to deep machining grooves, whereas a more uniform distribu-
tion was found at the treaded parts (Fig. 4e and f).
3.2. LV-SEM/HV-EDX
3.4. X-ray diffraction (XRD)
Fig. 2(a and e) illustrates low-vacuum secondary electron
images of the threaded roots of the implant surfaces at low Representative XRD graphs are given in Fig. 5(a and b). The
magnification, which revealed distinct morphological differ- monoclinic ZrO2 phase was clearly identified (2: 28.2◦ ) along
ences. WhiteSky demonstrated a double thread self-cutting with the tetragonal ZrO2 (2: 30.2◦ ) and cubic Y2 O3 (2: 29.1◦ )
design with a longitudinal retentive groove, whereas Zit-Z a phases. The presence of the latter phase was verified by the
cylindrical screw design. Fig. 2(b and f) exhibits the transitional second peak as well (2: 48.5◦ ). No cubic zirconia phase was
zone between the cervical collar and the threaded root part of identified.
the implants. The cervical collars showed horizontal machin-
ing serrations, more pronounced in Zit-Z. The transition of
3.5. Optical profilometry
the collar to the rough threaded root part was more clearly
defined in WhiteSky. In Zit-Z the machining tracks were vis-
3D-topometric images of the threaded root regions are given
ible even in the roughened threaded root region (Fig. 2c and
in Fig. 6(a–c).
g). Both implants showed a grainy texture at the root regions.
WhiteSky demonstrated a rougher surface texture com-
However, the size of the grainy domains was much smaller in
pared with Zit-Z. The machining tracks were apparent in the
Zit-Z (Fig. 2d and h).
root part of Zit-Z. Fig. 6(c) illustrates a top-view image of
The HV-EDX analysis showed that the implants were com-
the surface between two successive threads after automatic
posed of O, Hf, Y, and Zr. On the threaded root regions, Al was
plane subtraction; the latter was used for calculation of the
identified as well (Fig. 3a and b). The quantitative results are
surface roughness parameters. The surface roughness param-
summarized in Table 2. A significant interaction was found
eters calculated at 40× are summarized in Table 4. WhiteSky
among the independent variables. Comparison between the
demonstrated significantly higher roughness parameters than
implants per same region showed that Zit-Z exhibited the
Zit-Z (p < 0.05).
lowest O and the highest Zr content at the threaded root part.
The EDX Y/Zr atomic ratios as derived from the results pre-
sented in Table 2 ranged from 0.10 to 0.12 (p > 0.05), whereas 4. Discussion
the Al/Zr atomic ratios were 0.17 (WhiteSky) and 0.09 (Zit-Z,
p < 0.05). Elemental mapping showed that the Al distribution The results of the present study confirmed the testing hypoth-
followed the dark regions (low atomic number regions) of the esis, that significant differences exist between the implants in
backscattered images (Fig. 3c and d). At some of these regions the properties tested.
C, Na, K and Cl contaminants were mapped (Fig. 3e and f). HV-EDX analysis of bulk sectioned specimens confirmed
that in both implants yttria-stabilized zirconia was used as
3.3. Raman microanalysis raw material with a composition complying with the lim-
its described for biomedical applications (4.5–5.5% Y2 O3 , <5%
Raman spectra taken from cervical collar and the treaded root HfO2 , ZrO2 + Y2 O3 + HfO2 ∼99%) [17,18]. According to the man-
parts are illustrated in Fig. 4(a–d). The low intensity peaks at ufacturers, sintering of the implants was performed by hot
180 and 190 cm−1 Raman shift correspond to the monoclinic isostatic pressure (HIP). XRD revealed that a fraction of the
ZrO2 phase and the high intensity peaks at 145 and 262 cm−1 monoclinic ZrO2 phase existed in the implants along with the
Raman shift to the tetragonal ZrO2 phase. The threaded parts predominating tetragonal ZrO2 phase. However, since XRD as
exhibited more intense monoclinic peaks in both implants. applied, lacks of lateral resolution, it is unclear whether the
d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305 301
Fig. 4 – Raman microanalysis of the implant surfaces. Spectra taken at 3 different locations at cervical collar (a: WhiteSky, c:
Zit-Z) and threaded root locations (b: WhiteSky, d: Zit-Z). Mapping of the monoclinic ZrO2 phase distribution at cervical
collar (e) and threaded root regions (f) for WhiteSky implant.
302 d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305
Fig. 6 – 3D-optical profilometric images of threaded root regions of WhiteSky (a, 100X) and Zit-Z (b, 100×). For calculation of
the surface roughness parameters, the top-view images were subjected to automatic plane subtraction (c: WhiteSky).
destabilized tetragonal phase more aggressively than machin- formation and the distribution of the stresses developed
ing. The tetragonal to monoclinic ZrO2 phase transformation [23].
is followed by a 4–6% increase in volume that generates Dental implants are not similar to orthopaedic implants,
stress fields, which are associated with the development of where many problems have been encountered from the low
defects, like surface and subsurface cracking, grain pull-out temperature degradation and transformation of the tetrago-
and damage, that may affect load bearing interfaces like nal to monoclinic ZrO2 phase, that resulted in grain pull-out
the bone–implant interface [23]. The Vm % found on implant and catastrophic 3-body abrasive wear of the zirconia articu-
surfaces was quite low (0–7.2%). However, studies on trans- lating surfaces (hip and acetabular cap) [12,26]. Nevertheless,
formation kinetic parameters have shown that the growth of osseointegration with a superficial zirconia zone rich in mono-
initial fraction of monoclinic phase further affects the surface clinic phase could create subsurface cracking under functional
degradation mechanism, despite that formation of new mon- loading due to the great interfacial elastic modulus mismatch
oclinic nuclei is considered as the predominant effect [24]. It with bone. Additionally, in-service development of the mono-
is quite interesting that no requirement on the pre-existing clinic zone, starting from the surface and progressing into the
monoclinic phase content is listed in the relevant ISO specifi- material bulk, might further contribute to bulk and interfacial
cation [17]. problems. The exact mechanism of tetragonal to monoclinic
The Raman analysis as performed in the present study, ZrO2 transformation in aqueous medium is unknown. Three
probed only the surface layer, extending to a depth of a few zir- mechanisms have been proposed so far [11]: First is the cor-
conia grain diameters. However, the importance of this zone rosion mechanism, where the transformation proceeds from
is high, since a low percentage of monoclinic phase extending the hydrolytic degradation of the Zr–O–Zr bonds at the crack
to a depth of some microns has been found to severely reduce tips leading to formation of Zr–OH, that relieve the strains
the strength of Y-TZP [25]. stabilizing the tetragonal phase, while at the same time OH-
Further studies employing confocal Raman microanalysis groups diffuse into the lattice to occupy oxygen vacancies. It
could provide important information regarding the in-depth should be mentioned that the addition of Y3+ in Zr4+ sites of
extent of the tetragonal to monoclinic ZrO2 phase trans- Y-TZP creates oxygen vacancies due to establishment of Zr–O
304 d e n t a l m a t e r i a l s 2 6 ( 2 0 1 0 ) 295–305
containing tetragonal zirconia. J Am Ceram Soc [23] Pezzotti G, Yamada K, Sakakura S, Pitto RP. Raman
1982;65:284–8. spectroscopic analysis of advanced ceramic composite for
[16] http://icsdweb.fiz-karlsruhe.de/. hip prosthesis. J Am Ceram Soc 2008;91:1199–206.
[17] ISO 13356. Implants for surgery-ceramic materials based on [24] Chevalier J, Cales B, Drouin JM. Low-temperature aging of
yttria stabilized tetragonal zirconia (Y-TZP). Geneva: Y-TZP ceramics. J Am Ceram Soc 1999;82:2150–4.
International Standardization Organization; 1997. [25] Thompson I, Rawlings RD. Mechanical behaviour of zirconia
[18] ASTM F1873-98: High-purity dense yttria-tetragonal and zirconia-toughened alumina in a simulated body
zirconium oxide polycrystal (Y-TZP) for surgical implant environment. Biomaterials 1990;11:505–8.
applications. American Society for Testing and Materials [26] Fernandez-Fairen M, Sala P, Gil FJ. Failures of yttria stabilized
1998:13.01;908 [withdrawn]. tetragonal zirconia: 52 retrieved ceramic femoral heads of
[19] Deville S, Chevalier J, Gremillard L. Influence of surface total hip prostheses. Biomed Mater Eng 2006;16:
finish and residual stresses on the aging sensitivity of 415–22.
biomedical grade zirconia. Biomaterials 2006;27: [27] Mǔnoz MC, Gallego S, Beltrán JI, Cerdá J. Adhesion at
2186–92. metal–ZrO2 interfaces. Surf Sci Rep 2006;61:303–44.
[20] Piattelli A, Degidi M, Paolantonio M, Mangano C, Scarano A. [28] Macdonald W, Campbell P, Fisher J, Wennerberg A. Variation
Residual aluminum oxide on the surface of titanium in surface texture measurements. J Biomed Mater Res Part B:
implants has no effect on osseointegration. Biomaterials Appl Biomater 2004;70B:262–9.
2003;24:4081–9. [29] Thomas KA, Cook S. An evaluation of variables influencing
[21] Aita H, Hori N, Takeuchi M, Suzuki T, Yamada M, Anpo M, et implant fixation by direct implant apposition. J Biomed
al. The effect of ultraviolet functionalization of titanium Mater Res 1985;19:875–901.
on integration with bone. Biomaterials 2009;30:1015–25. [30] Shalabi MM, GortemakerA, Van’t Hof MA, Jansen JA,
[22] Massaro C, Rotolo P, De Riccardis F, Milella E, Napoli A, Creugers NH. Implant surface roughness and bone healing:
Wieland M, et al. Comparative investigation of the surface a systematic review. J Dent Res 2006;85:
properties of commercial titanium dental implants. Part I: 496–500.
chemical composition. J Mater Sci Mater Med 2002;13:538–48.