Sei sulla pagina 1di 7

M e d i c a l P hy s i c s a n d I n f o r m a t i c s O r i g i n a l R e s e a r c h

Filippi et al.
MRI of the Lumbar Spine

Medical Physics and Informatics


Original Research
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

Improvements in Lumbar
Spine MRI at 3 T Using
Parallel Transmission
Christopher G. Filippi1 OBJECTIVE. Parallel transmission MRI at 3 T improves image quality by reducing di-
Morgan Carlson2 electric effects with radiofrequency shimming. The purpose of this study was to determine
Jason M. Johnson 3 whether parallel transmission MRI improves signal-to-noise ratio (SNR) and contrast-to-
Heather N. Burbank1 noise ratio (CNR) in lumbar spine MRI at 3 T.
Gary F. Alsofrom1 MATERIALS AND METHODS. Ten healthy volunteers underwent T1-weighted MRI
and nine healthy volunteers underwent T2-weighted MRI of the lumbar spine. Sagittal and ax-
Trevor Andrews1,4
ial T1- and T2-weighted images were acquired using parallel transmission MRI and conven-
Filippi CG, Carlson M, Johnson JM, Burbank HN, tional MRI. The percentage improvements in SNR and CNR were calculated, and statistical
Alsofrom GF, Andrews T significance was determined using a two-tailed Student t test with p < 0.05 for significance.
RESULTS. The CNR and SNR showed statistically significant improvements at all levels
of the lumbar spine except SNR at T11 on axial T2-weighted imaging. For sagittal T1-weight-
ed imaging, the average improvement with parallel transmission MRI was 53% in CNR and
19% in SNR. For axial T1-weighted imaging, the average improvement was 48% in CNR and
23% in SNR. For sagittal T2-weighted imaging, the average CNR improvement was 38% and
the average SNR improvement, 20%. For axial T2-weighted scans, the average percentage
improvement in CNR was greater than 100% and the average SNR improvement was 18%
with parallel transmission MRI.
CONCLUSION. The parallel transmission sequence improves image quality of lumbar
spine MRI at 3 T, which is quantitatively supported by statistically significant improvements
Keywords: 3-T MRI, lumbar spine MRI, parallel in SNR and CNR.
transmission MRI

T
echnologic developments have Recent advances in the use of multiple ra-
DOI:10.2214/AJR.11.8139
led to the widespread use of MRI diofrequency transmit sources, also known as
Received October 26, 2011; accepted after revision at 3 T. The benefits of imaging at parallel transmission MRI or multitransmit
January 27, 2012. a higher magnetic field strength MRI, have made it possible to improve ra-
include improvements in the signal-to-noise diofrequency uniformity. This correction can
Some of the equipment used in this study is supported
ratio (SNR) and increased spatial resolution be accomplished by performing a short addi-
by Department of Energy grant SC 0001753.
[1, 2]. However, there are drawbacks includ- tional calibration to determine the optimal out-
1
Department of Radiology, Fletcher Allen Health Care ing dielectric shading and a signal intensity put for the independent radiofrequency trans-
University of Vermont School of Medicine, inhomogeneity artifact (i.e., B1 inhomogene- mitters to produce radiofrequency uniformity
111 Colchester Ave, Patrick One, Burlington, VT 05401. ity). Nonuniform radiofrequency distribu- throughout the FOV. This calibration process
Address correspondence to C. G. Filippi
tion in the body is caused by changes of the is called radiofrequency shimming or B1
(christopher.filippi@vtmednet.org).
radiofrequency wave due to the electrical shimming and is analogous to the B0 shim-
2
Department of Physical Rehabilitation and Medicine, properties of the tissue and is more pro- ming calibration that is routinely performed
University of Washington School of Medicine, Seattle, WA. nounced at 3 T because the radiofrequency on all scanners to make the static magnetic
3
wavelength at 3 T ( 25 cm) approaches the field homogeneous. This improvement in ra-
Department of Radiology, Massachusetts General
Hospital, Boston, MA.
size of the body, resulting in standing waves diofrequency uniformity reduces dielectric
and B1 inhomogeneity [35]. The attenua- shading in images [8, 9]. With parallel trans-
4
Philips Healthcare, Cleveland, OH. tion of radiofrequency amplitude due to tis- mission MRI, radiofrequency shimming au-
sue conductivity contributes to B1 inhomo- tomatically adjusts the frequency, amplitude,
AJR 2012; 199:861867 geneity [4, 5]. Dielectric shading typically phase, and waveform of all radiofrequen-
0361803X/12/1994861
produces regions of decreased brightness at cy pulses for optimal uniformity adapted to
3 T and can even produce regions of absent each patients unique anatomy [8, 9]. Paral-
American Roentgen Ray Society signal [6, 7]. lel transmission imaging is not the same as

AJR:199, October 2012 861


Filippi et al.

parallel receive imaging, which uses multiple sion distribution and calculate the output required time was 7 minutes 16 seconds because the SAR
receiver coil elements and was first introduced from each transmitter to minimize radiofrequen- limits required more slice packages. For simi-
in the 1980s [10, 11]. Parallel receive technol- cy transmission inhomogeneity. The console soft- lar reasons, the scanning duration for axial T1-
ogy can shorten scanning time [12] and in- ware allowed the technologist to acquire each weighted imaging was 2 minutes 59 seconds with
crease SNR [10]. sequence thereafter either with or without radio- radiofrequency shimming and 5 minutes 58 sec-
Parallel transmission MRI has recently frequency shimming during the same scanning onds without radiofrequency shimming.
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

been released for clinical use. The purpose session. Each scan that was acquired with radio- It is necessary to first estimate noise to ob-
of this study was to evaluate the benefits of frequency shimming was followed by an identical tain SNR and CNR estimates. Several meth-
this technology for improving image quality scan acquired without radiofrequency shimming. ods have been used in the past [14], but general-
in routine lumbar spine MRI at 3 T. In one All scans were obtained using a 15-channel sensi- ly these methods have involved either using pixel
previous study describing the use of paral- tivity-encoding spinal receive coil. values in the tissue of interest for a set of repeat-
lel radiofrequency transmission on imaging For all acquisitions, a 4-mm slice thickness ed scans or using a region of interest (ROI) in air
of the spine [9], investigators reported a 30% was used with a gap of 1 mm and without paral- for a single image. For noise estimates in scans
reduction in total scanning time and image lel receive acceleration. The parameters for sag- of patients, the repeated-scans approach is often
quality comparable to that of standard sin- ittal T1-weighted imaging were as follows: TR/ avoided because the scanning session can be in-
gle-transmission sequences; however, the TE, 425/8.8; number of signals acquired, 3; FOV, tolerably long; in addition, patient movement be-
image quality assessment for that study was 320 mm; acquired matrix, 360 288; and band- tween scans during these long sessions can domi-
subjective and did not include quantitative width, 257 Hz. For axial T1-weighted imaging, the nate the noise estimates and lead to a significant
measurements of SNR or CNR [9]. In our following parameters were used: TR/TE, 400/8; and erratic patient-dependent overestimation bias.
study, routine 3-T axial and sagittal T1- and number of signals acquired, 3; FOV, 150 mm; ac- Unfortunately with spinal imaging, it is also often
T2-weighted images were directly compared quired matrix, 192 144; and bandwidth, 291 Hz. impractical to use the spinal image to draw a sig-
quantitatively with 3-T axial and sagittal The scanning parameters for sagittal T2-weighted nal-free ROI for noise estimation because there is
T1- and T2-weighted images obtained with imaging were TR/TE, 3500/110; number of sig- typically little, if any, air within the FOV. In cases
parallel transmission MRI during the same nals acquired, 3; FOV, 320 mm; acquired matrix, in which a region of air can be found, the region is
scanning session by calculating the SNR and 360 288; and bandwidth, 303 Hz. For axial T2- often contaminated by artifactual signal due to re-
CNR. Image quality of single-transmission weighted images, the scanning parameters were spiratory motion (as we noted in our preliminary
and parallel transmission MR images was TR/TE, 2800/120; number of signals acquired, 3; attempts to obtain these estimates). Therefore, to
compared qualitatively by two neuroradi- FOV, 150 mm, acquired matrix, 192 144; and obtain an accurate estimate of noise for SNR and
ologists, both of whom have a certificate of bandwidth, 150 Hz. CNR calculations, each acquisition in the current
added qualification, in blinded fashion. We The TR was kept the same for correspond- study was run twice with identical settings (e.g.,
hypothesized that parallel transmission MRI ing shimmed and unshimmed scans so that im- with identical bandwidth and receiver scaling) ex-
improves SNR and CNR and that these im- provements in SNR and CNR could be directly cept that the gradients and radiofrequency trans-
provements result in improved image quality. attributable to the radiofrequency shimming. In mitter were switched off by the MR console dur-
a multislice sequence when the number of slices ing the second scan. This method has been used
Materials and Methods exceeds the maximum number of slices that can previously to estimate noise in coronary MR stud-
This HIPAA-compliant and institutional re- be acquired in a single TR interval, the slices are ies [15]. Although acquiring these noise scans
view boardapproved study was performed at the grouped in packages of slices that can fit in a effectively doubles the duration of the total scan-
University of Vermont MRI Center for Biomed- single TR interval and the total scanning time ning session, this approach made it possible to ob-
ical Imaging. Ten healthy volunteers (nine men, is proportional to the number of these packages. tain noise ROIs free of motion artifacts for all cor-
one woman; age range, 2547 years; average age, On sagittal sequences, the number of packages responding spinal images acquired in this study.
30.7 years) were recruited for T1-weighted imag- was kept identical. Thus, the scanning duration SNR and CNR measurements were calculated
ing and nine healthy volunteers (six male, three for each sagittal T1-weighted scan was 4 minutes from ROI measurements. To compare the SNR of
female; age range, 1646 years; average age, 30.2 8 seconds; for each sagittal T2-weighted scan, it parallel transmission MRI versus the SNR of non
years) were recruited for T2-weighted imaging. was 4 minutes 30 seconds. On axial sequences, parallel transmission MRI, T1-weighted scans in
T1- and T2-weighted scans of the same patient the number of packages was reduced from eight the axial plane of vertebral bodies T11, T12, and L1
were not obtained during the same session be- to five on both the axial T1- and T2-weighted ra- were chosen because these areas represent the re-
cause of the scanning time needed for all of the diofrequency-shimmed images because radiofre- gion in which there is the greatest degree of dielec-
required scans. quency shimming reduced the specific absorption tric shading. Equivalent ROI measurements were
All images were obtained on a 3-T Achieva MR rate (SAR). More specifically, because the SAR obtained with respect to size at all levels using a
system (Philips Healthcare) after a commercial- safety limits of the system require a dead time minimum ROI size of 15 15 mm for both scans
ly available dual-source parallel radiofrequency at the end of the TR interval (after the acquisi- with and scans without radiofrequency shimming.
transmission upgrade to an Achieva TX MR sys- tion readouts) during which there are no radiofre- CNR values were calculated in the axial plane
tem. The upgraded system uses two independent- quency transmissions, the reduced SAR also re- by placing an ROI on the spinal cord and vertebral
ly controlled transmitters to provide radiofrequen- duces the dead time and allows more slices to be marrow and comparing this ROI with the para-
cy power to the separate ports of a standard body acquired in each TR interval. Thus, for axial T2- spinal musculature on equivalent slices at the T11,
transmit coil [13]. At the beginning of a session, weighted imaging with radiofrequency shimming, T12, and L1 levels on parallel transmission and
a radiofrequency shimming calibration test is per- the scanning duration was 3 minutes 58 seconds. nonparallel transmission images. The minimum
formed to estimate the radiofrequency transmis- Without radiofrequency shimming, the scanning ROI size for the spinal cord was 5 5 mm, and the

862 AJR:199, October 2012


MRI of the Lumbar Spine

minimum ROI size for the paraspinal musculature of nine cases on axial T2-weighted imaging,
was 10 10 mm. Average pixel measurements were and six of nine cases on sagittal T2-weighted
obtained along with SDs of the measurements. imaging. In all cases, radiofrequency shim-
To compare SNRs in the sagittal plane on T1- ming images were judged to be as good as or
weighted scans, vertebral bodies L1 through and better than the images without radiofrequen-
including L5 were identified, and ROI averag- cy shimming.
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

es were obtained from equivalent slices on both For all sequences studied, improvements
scans with and those without radiofrequency in image quality met statistical significance
shimming. CNR values were calculated for MR with the Friedman rank coefficient test (p <
images with and MR images without radiofre- 0.05). The Cohen kappa statistic exceeded
quency shimming by measuring ROIs of the T11 0.75, which is indicative of excellent agree-
12 disk through and including the L45 disk space ment between the two neuroradiologists. The
levels. These values were compared with the ROIs reasons noted for improvements in image
of the vertebral body marrow (T12 through and quality included less dielectric shading at the
including L5). For example, an average ROI mea- Fig. 1Sagittal T2-weighted image (TR/TE, thoracolumbar junction with better visual-
3500/100; 3 signals acquired; 4-mm thickness;
surement from the T1112 disk was compared 1-mm gap; 320-mm FOV) of healthy 27-year-old ization of conus medullaris; improved con-
with an ROI measurement from the T12 vertebral male volunteer shows regions of interest () trast between the spinal cord, vertebral mar-
body. The minimum ROI measurement for disk placed on vertebral body marrow and within CSF row, and CSF; and more uniform fat signal
for measurement of signal-to-noise ratio. Min =
spaces was 3 2 mm, and the minimum ROI mea- minimum, max = maximum, US = unsigned short across the entire FOV.
surement for vertebral bodies was 15 15 mm. (integer that cannot be less than zero or greater than With multitransmit MRI, there were marked
The SNR calculations on sagittal T2-weighted 216 1), avg = average, StDv = SD. improvements on axial T1-, axial T2-, sagittal
images were performed the same way as on the T1-, and sagittal T2-weighted imaging with re-
sagittal T1-weighted images. For the CNR cal- experience, respectively) were asked to judge im- spect to SNR and CNR values (Tables 1 and
culations on sagittal T2-weighted images, verte- age quality of axial T1-, axial T2-, sagittal T1-, 2). On sagittal T1-weighted imaging, the av-
bral body ROI averages were compared with CSF and sagittal T2-weighted images with and with- erage percentage improvement in CNR was
and ROI averages in the spinal cord. The mini- out radiofrequency shimming on the PACS in a 53% (p < 0.001 at all levels) and the average
mum ROI size for a vertebral body was 15 15 blinded fashion; neither identifiable imaging pa- percentage improvement in SNR was 19%
mm; for spinal cord, 3 3 mm; and for CSF, 5 2 rameters nor text was on the images. The grad- (p < 0.006 at all levels). On axial T1-weight-
mm. SNR and CNR calculations were performed ing system the neuroradiologists used was as ed imaging, the average percentage improve-
on axial T2-weighted images the same way as on follows: The image quality of the parallel trans- ment in CNR was 48% (p < 0.02 at all levels)
T1-weighted images (Fig. 1). Noise measurements mission MR image was graded as equivalent to and the average percentage improvement in
were calculated from the background noise scans (2 points), better than (3 points), or worse than SNR was 23% (p < 0.008 at all levels). On
that were acquired for each and every imaging (1 point) the nonparallel transmission MR im- sagittal T2-weighted imaging, the average
scan. The following formula was used to calcu- age. These results were recorded for all 19 cases percentage improvement in SNR was 20%
late SNR [16]: for all four sequences. The Cohen kappa statistic (p< 0.05 at all levels) and the average per-
Meantissue was used to measure the degree of agreement be- centage improvement in CNR was 38% (p<
(SDnoise) 0.655
tween the two readers concerning diagnostic im- 0.005). On axial T2-weighted imaging, the
aging quality. Kappa values of 0.400.75 are good average percentage improvement in SNR
where Meantissue is the mean value of the ROI agreement and greater than 0.75, excellent agree- was 18%, and p was < 0.005 at T12 and L1,
drawn in the corresponding tissue and SDnoise is the ment. The Friedman rank coefficient test was used but statistical significance was not observed
SD of the ROI on the corresponding noise scan. to compare image quality of the parallel transmis- at T11. The average percentage improvement
In addition, CNR was calculated as follows [16]: sion MR images and routine MR images, with sta- on axial T2-weighted imaging in CNR was
Meantissue1 Meantissue2
tistical significance set to p < 0.05. greater than 100%, with p < 0.05 at all lev-
els (Figs. 25).
(SDnoise) 0.655
Results
where Meantissue1 is the mean value of the ROI On qualitative examination of the MR im- Discussion
drawn in tissue 1 and Meantissue1 is the mean value ages, one neuroradiologist judged the radio- Nonuniform radiofrequency distribution in
of the ROI drawn in tissue 2. frequency-shimmed images to be superior to the body often results in dielectric shading,
Percentage improvements in SNR and CNR the images without radiofrequency shimming which creates artifacts that can impede image
values were calculated at all levels measured for in nine of 10 cases on axial T1-weighted, eight interpretation [1, 2, 17]. We have often ob-
all the volunteers. The statistical significance of of nine cases on axial T2-weighted, nine of served these artifacts on routine lumbar spine
the average percentage improvements in SNR 10 cases on sagittal T1-weighted, and eight of MRI at 3 T, particularly near the lumbosacral
and CNR comparing images with radiofrequency nine cases on sagittal T2-weighted imaging. and thoracolumbar junctions on both sagittal
shimming to those without radiofrequency shim- The other neuroradiologist judged the images and transaxial imaging. Dielectric shading has
ming was obtained with a two-tailed Student t test with radiofrequency shimming to be superior been reported to occur predominantly in heavi-
with significance set for p values < 0.05. to the corresponding images without radio- er patients [18]. However, we have observed
Two board-certified neuroradiologists with frequency shimming in eight of 10 cases on significant dielectric effects across the spectrum
certificates of added qualification (4 and 25 years sagittal and axial T1-weighted imaging, eight of body types including children (Johnson A, et

AJR:199, October 2012 863


Filippi et al.

TABLE 1: Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) Values on Sagittal and Axial T1-Weighted
Imaging and Percentage Improvement for Parallel Transmission Sequence Compared With NonParallel
Transmission Sequence
Mean SNR Mean CNR
T1-Weighted
Imaging Plane Multitransmit Nonmultitransit % Improvement Multitransmit Nonmultitransit % Improvement
Sagittal
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

L1 50.08 41.76 19.9a 14.26 9.46 50.7a


L2 48.19 41.83 15.2a 15.60 11.02 41.6a
L3 49.00 42.54 15.2a 17.08 11.72 45.7a
L4 47.06 40.02 17.6a 15.89 10.49 51.5a
L5 48.97 39.24 24.8a 18.62 10.71 73.9a
Axial
T11 26.16 20.41 28.2a 26.12 18.97 37.7b
T12 26.55 21.93 21.1a 20.78 13.85 50.0a
L1 24.91 20.90 19.2a 22.76 14.47 57.3a
ap < 0.01.
bp < 0.05.

TABLE 2: Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) Values on Sagittal and Axial T2-Weighted
Imaging and Percentage Improvement for Parallel Transmission Sequence Compared With NonParallel
Transmission Sequence
Mean SNR Mean CNR
T2-Weighted
Imaging Plane Multitransmit Nonmultitransit % Improvement Multitransmit Nonmultitransit % Improvement
Sagittal
L1 29.36 24.72 18.8a 108.08 79.55 35.9a
L2 29.50 20.69 42.6b 107.94 77.96 38.5a
L3 30.18 27.37 10.3b 107.27 76.91 39.5a
L4 29.63 26.15 13.3b 107.82 78.12 38.0a
L5 30.48 26.11 16.7a 106.97 78.16 36.9a
Axial
T11 15.96 14.03 13.8 11.67 1.38 > 100b
T12 15.17 12.89 17.7a 12.58 1.30 > 100b
L1 15.20 12.47 21.8a 40.94 3.53 > 100b
ap < 0.01.
bp < 0.05.

al., 2010 annual meeting of the International In our study, there were statistically sig- To isolate the direct effect of parallel trans-
Society for Magnetic Resonance in Medicine nificant improvements in both CNR and SNR mission on SNR, CNR, and subjective image
[ISMRM]), athletic and healthy adults [15], values on axial and sagittal T1-weighted scans quality, almost identical protocols were per-
average-weight and overweight adults [18], and on axial and sagittal T2-weighted scans formed using both conventional and parallel
patients with ascites [13, 19], and pregnant compared with conventional MR images at 3 T. transmission MRI at 3 T. However, because
women (Filippi CG, et al., 2010 ISMRM In our study, we kept both the TR and TE iden- these protocols were optimized for conven-
meeting). We have previously attributed di- tical on the nonmultitransmit and multitrans- tional MRI, they were not necessarily opti-
electric effects to problems with the position mit MR images to assess how radiofrequency mal for parallel transmission. Specifically, the
of the receiver coil, but this loss of signal uni- shimming improved both CNR and SNR. On parallel transmission MR scans did not ben-
formity seems more likely to be related to B1 axial T1- and axial T2-weighted images, the efit from reduced SAR that would enable de-
inhomogeneity or difficulties with radiofre- number of packages was reduced, but all other creases in TR and scanning time. In a more
quency transmission. This nonuniformity of imaging parameters were kept constant includ- practical imaging scenario, parallel transmis-
signal from dielectric shading degrades im- ing the TR, TE, number of signals acquired, sion would allow either a substantial decrease
age quality on routine lumbar spine imaging FOV, and slice thickness; this reduction in the in scanning time or, if the scanning time is
at 3 T and could potentially obscure impor- number of packages resulted in 50% reductions maintained, an additional increase in SNR and
tant pathologic findings. in scanning time. CNR through more signal averaging.

864 AJR:199, October 2012


MRI of the Lumbar Spine

Fig. 2Healthy 36-year-old male volunteer.


A and B, Routine axial T1-weighted image (A)
(scanning time, 5 minutes 58 seconds; TR/TE, 400/8;
3 signals acquired; 4-mm thickness; 1-mm gap; 320-
mm FOV) and axial parallel transmission T1-weighted
image (B) (scanning time, 2 minutes 59 seconds;
same parameters as A; number of stacks reduced
from 8 to 5). There is dielectric shading on routine
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

image (A), particularly across spinous process region


posteriorly and anterior to aorta and inferior vena
cava. These areas show more uniformity of signal on
parallel transmission image (B).

A B

Fig. 3Healthy 38-year-old female volunteer.


A and B, MR images of lumbar spine obtained
using conventional (A) and parallel transmission
(B) sequences with identical parameters (TR/
TE, 425/8.8; 3 signals acquired; 4-mm thickness;
1-mm gap; 150-mm FOV). Only difference between
images is application of B1 shimming for parallel
transmission sequence. Parallel transmission image
(B) better shows less dielectric effect or shading
along superior aspect of image acquisition than non
parallel transmission image (A). Scanning time for
both sequences was 4 minutes 30 seconds.
A B

In our study, because of the time need- design could be seen as a potential limitation cy shimming translate into improvements in
ed for radiofrequency shimming, obtaining to this study. Furthermore, we recruited only thoracic and cervical spine MRI at 3 T.
noise scans, and directly comparing parallel healthy volunteers so we cannot address is- One of the reasons for obtaining estimates of
transmission with nonparallel transmission sues of improved lesion conspicuity or diag- SNR and CNR for this study was to provide a
MR sequences, which surpassed 2 hours, we nostic accuracy for lesion detection because basis for quantitative comparison of scans with
opted to scan volunteers just for the sagittal we were interested just in the determination potentially different scaling factors for the pur-
and axial T1-weighted scans and not for the of improvements in SNR and CNR with the pose of examining the effects of radiofrequency
T2-weighted component. If the T2-weight- addition of parallel transmission MRI. How- shimming. The gradient coils were disabled
ed component had been included, each ex- ever, image quality was judged to be supe- during the noise scan to avoid any potentially
amination would have required more than 4 rior by both neuroradiologists. Further stud- erratic noise bias due to the gradients them-
hours of continuous imaging and none of the ies will be helpful in determining whether selves (e.g., radiofrequency spikes) because
volunteers had agreed to undergo imaging observed gains in SNR and CNR on rou- even small contributions can, in principle, affect
for that long. This characteristic of our study tine lumbar spine MRI with radiofrequen- the noise estimate without affecting the scaling.

AJR:199, October 2012 865


Filippi et al.

Fig. 4Healthy 46-year-old male volunteer.


A and B, Axial T2-weighted images of lumbar
spine obtained using conventional MRI (A) and
parallel transmission (B) sequences. Routine image
(A; scanning time, 7 minutes 16 seconds) shows
dielectric shading or B1 inhomogeneity at level of
spinal canal that is obscuring visualization of nerve
roots. Signal contrast between nerve roots and CSF
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

is markedly improved on multitransmit image (B;


scanning time, 3 minutes 58 seconds). Images were
obtained using identical scanning parameters (TR/
TE, 2800/120; 3 signals acquired; 4-mm thickness;
1-mm gap; 150-mm FOV) but there were fewer stacks
for B. Scan time reduction of 50% was observed
for parallel transmission sequence compared with
conventional sequence.

A B

Fig. 5Healthy 33-year-old male volunteer.


A and B, Sagittal T2-weighted images of lumbar
spine obtained using conventional (A) and
parallel transmission (B) sequences with identical
parameters (TR/TE, 3500/110; 3 signals acquired;
4-mm slice thickness; 1-mm gap; 150-mm FOV).
With use of B1 shimming for parallel transmission
sequence, there is less dielectric shading both
anterior to lumbar spine vertebral bodies and at level
of conus medullaris and contrast-to-noise ratio is
markedly improved, resulting in better conspicuity
between CSF and conus medullaris as well as
vertebral body marrow. Hemangioma at T11 level is
clearly better visualized on multitransmit image (B).
A B
One could argue, however, that obtaining such In conclusion, parallel transmission MRI fect of B1 inhomogeneity on breast MR imaging
noise scans with the gradients turned on is pref- shows statistically significant improvements at 3.0 T. Radiology 2007; 244:929930
erable because having the gradient coils turned in both SNR and CNR on axial and sagit- 3. Zhang Z, Yip C, Grissom W, Noll DC, Boada FE,
on is more similar to the imaging scan. Such an tal T1-weighted and on axial and sagittal T2- Stenger VA. Reduction of B1 inhomogeneity with
approach can provide additional insight into the weighted images of the lumbar spine when transmit SENSE slice-select pulses. Magn Reson
noise and the general performance of the MR compared with routine MRI at 3 T. The im- Med 2007; 57:842847
scanner; because the contribution of gradients age quality of parallel transmission MR im- 4. Yang QX, Wang J, Zhang X, et al. Analysis of
to the noise can vary from one scan to another, ages was judged to be superior to that of wave behavior in lossy dielectric samples at high
however, accurately characterizing such effects nonparallel transmission MR images be- field. Magn Reson Med 2002; 47:982989
might require repeated noise scanning with a cause there was less dielectric shading and 5. Collins CM, Liu W, Schreiber W, Yang QX, Smith
correspondingly longer scanning session. Our the signal was more uniform. MB. Central brightening due to constructive inter-
weekly quality control phantom tests showed ference with, without, and despite dielectric reso-
no significant artifacts or change in SNR during References nance. J Magn Reson Imaging 2005; 21:192196
the course of this experiment even for more gra- 1. Merkle EM, Dale BM. Abdominal MRI at 3.0 T: 6. Jin J, Chen J. On the SAR and field inhomogeneity
dient-intensive functional MRI and diffusion- the basics revisited. AJR 2006; 186:15241532 of birdcage coils loaded with the human head.
tensor imaging sequences. 2. Kuhl CK, Kooijman H, Gieseke J, Schild HH. Ef- Magn Reson Med 1997; 38:953963

866 AJR:199, October 2012


MRI of the Lumbar Spine

7. Ibrahim TS, Lee R, Abduljalil AM, Baertlein BA, Mueller OM. The NMR phased array. Magn Re- encoding: application to coronary MR angiogra-
Robitaille PM. Dielectric resonances and B(1) field son Med 1990; 16:192225 phy. J Magn Reson Imaging 2011; 33:13301340
inhomogeneity in UHFMRI: computational analy- 12. Hutchinson M, Raff U. Fast MRI data acquisition 16. Henkelman RM. Measurement of signal intensi-
sis and experimental findings. Magn Reson Imag- using multiple detectors. Magn Reson Med 1988; ties in the presence of noise in MR images. Med
ing 2001; 19:219226 6:8791 Phys 1985; 12:232233
8. Katscher U, Brnert P. Parallel RF transmission 13. Willinek WA, Gieseke J, Kukuk GM, et al. Dual- 17. Erturk SM, Alberich-Bayarri A, Hermann KA,
Downloaded from www.ajronline.org by 36.82.102.157 on 03/04/17 from IP address 36.82.102.157. Copyright ARRS. For personal use only; all rights reserved

in MRI. NMR Biomed 2006; 19:393400 source parallel radiofrequency excitation body Martin-Bonnati L, Ros PR. Use of 3.0-T MR im-
9. Nelles M, Knig R, Gieseke J, et al. Dual-source MR imaging compared with standard MR imag- aging for evaluation of the abdomen. Radio-
parallel RF transmission for clinical MR imaging ing at 3.0 T: initial clinical experience. Radiology Graphics 2009; 29:15471563
of the spine at 3.0 T: intraindividual comparison 2010; 256:966975 18. Vernickel P, Roschmann P, Findeklee C, et al.
with conventional single-source transmission. Ra- 14. Dietrich O, Raya JG, Reeder SB, Reiser MF, Eight-channel transmit/receive body MRI coil at
diology 2010; 257:743753 Schoenberg SO. Measurement of signal-to-noise 3T. Magn Reson Med 2007; 58:381389
10. Hyde JS, Jesmanowicz A, Froncisz W, Kneeland ratios in MR images: influence of multichannel 19. Kukuk GM, Gieseke J, Weber S, et al. Focal liver
JB, Grist TM. Parallel image acquisition from coils, parallel imaging, and reconstruction filters. lesions at 3.0 T: lesion detectability and image
noninteracting local coils. J Magn Reson 1986; J Magn Reson Imaging 2007; 26:375385 quality with T2-weighted imaging by using con-
70:512517 15. Yu J, Agarwal H, Stuber M, Schar M. Practical ventional and dual-source parallel radiofrequency
11. Roemer PB, Edelstein WA, Hayes CE, Souza SP, signal-to-noise ratio quantification for sensitivity transmission. Radiology 2011; 259:421428

AJR:199, October 2012 867

Potrebbero piacerti anche