EFFECTIVE MARCH 18, 2005 (AISC 303-05). 11. SEE ARCHITECT'S DRAWINGS FOR THE EXTENT AND EXACT LOCATION OF MASONRY WALLS. OF 2,000 POUNDS FOR SEATS UP TO 3 1/2 INCHES DEEP AND 1,200 POUNDS FOR SEAT B. MECHANICAL ANCHORS INTO MASONRY LINTELS OR GROUT FILLED CELLS:
4. WATER REDUCING AND AIR ENTRAINING AGENTS SHALL BE INCLUDED IN DESIGN MIXES. OVER 3 1/2 INCHES DEEP. 1. USE THE FOLLOWING (UNO):
5. OPEN WEB STEEL JOISTS: SUPERPLASTICIZERS MAY BE USED AT THE CONTRACTOR'S OPTION. 12. WALL CONTROL JOINTS (WCJ): A. HILTI KWIK BOLT 3 MASONRY ANCHORS (ICC-ES ESR1385)
A. STANDARD SPECIFICATION FOR OPEN WEB STEEL JOISTS, K-SERIES (K-1.1-05). A. WALL CONTROL JOINTS SHALL BE PROVIDED IN ALL CONCRETE MASONRY STEEL ROOF DECK: B. SIMPSON STRONG-TIE WEDGE-ALL ANCHOR(WA)
B. CODE OF STANDARD PRACTICE FOR STEEL JOISTS AND JOIST GIRDERS, EFFECTIVE 5. A CONCRETE MIX DESIGN FOR EACH UNIQUE COMBINATION OF STRENGTH, COARSE AGGREGATE CONSTRUCTION AT LOCATIONS INDICATED ON THE STRUCTURAL OR ARCHITECTURAL 1. ROOF DECK SHALL BE 1-1/2" DEEP. SEE ROOF PLAN FOR GAGE AND PROFILE (ICBO-ES ER-3631) (FL5415)
MAY 03, 2005. GRADATION AND WATER CEMENT RATIO SPECIFIED SHALL BE PREPARED BY THE SUPPLIER OR DRAWINGS BUT UNLESS NOTED OTHERWISE AT A SPACING NOT GREATER THAN 24' DESIGNATION. 2. USE THE FOLLOWING ONLY WHERE SPECIFICALLY CALLED OUT ON THE
C. STANDARD SPECIFICATIONS, LOAD TABLES AND WEIGHT TABLES FOR STEEL JOISTS AN INDEPENDENT TESTING LABORATORY AND BE SUBMITTED FOR REVIEW PRIOR TO CASTING O.C. DOCUMENTS:
AND JOIST GIRDERS (SJI-02). ANY CONCRETE. MIXES THAT WILL BE TRANSPORTED AT THE PROJECT SITE BY PUMPING B. HORIZONTAL JOINT REINFORCING SHALL BE INTERRUPTED EACH SIDE OF WALL 2. ROOF DECK SHALL BE PLACED SO AS TO COVER AT LEAST TWO SPANS. NO SINGLE SPAN A. HILTI HUS-H SCREW ANCHOR (ICC-ES ESR2369)
D. STANDARD SPECIFICATIONS FOR LONGSPAN STEEL JOISTS, LH SERIES AND DEEP
LONGSPAN JOISTS, DLH SERIES (LH/DLH-1.1-05).
SHALL BE SPECIFICALLY DESIGNED FOR PUMPING.
C.
CONTROL JOINTS.
WALL CONTROL JOINTS SHALL NOT BE PLACED OVER OPENINGS OR WITHIN AN
CONDITIONS SHALL BE USED.
C.
B. SIMPSON STRONG-TIE TITEN HD (THD) (ICC-ES ESR1056)(FL2304)
ADHESIVE ANCHORS INTO CONCRETE:
Proton Therapy
E. STANDARD SPECIFICATION FOR JOIST GIRDERS (JG-1.1-05). 6. SLABS ON GRADE: UNLESS NOTED OTHERWISE, CONCRETE SLABS ON GRADE SHALL BE A
MINIMUM OF 4" THICK, REINFORCED WITH 6x6 W1.4Xw1.4 WWF PLACED 1-1/2" CLEAR
OPENING JAMB WIDTH. SEE PLANS AND/OR JAMB REINFORCING SCHEDULE FOR
MINIMUM JAMB WIDTHS.
3. DECK SHALL BE FABRICATED SO THAT DECK RUNS CONTINUOUSLY OVER OPENINGS. THE
OPENINGS IN THE DECK SHALL NOT BE CUT UNTIL THE OPENING IS NEEDED (PER OSHA).
1. USE THE FOLLOWING (UNO):
A. HILTI HIT-RE 500-SD ADHESIVE (ICC-ES ESR2322) LINAC Addition
6. MASONRY CONSTRUCTION: FROM THE TOP OF THE SLAB. SLABS SHALL BE PLACED OVER PROPERLY COMPACTED EARTH. D. SEE ARCHITECTURAL DRAWINGS FOR SEALANT REQUIREMENTS AT WALL CONTROL B. RED HEAD EPCON G5 ADHESIVE (ICC-ES ESR1137)(FL6582)
A. BUILDING CODE REQUIREMENTS FOR MASONRY STRUCTURES JOINTS. 4. STEEL DECK SHALL CONFORM TO ASTM A653 SQ GRADE 33 (Fy = 33,000 PSI). C. SIMPSON STRONG-TIE SET-XP EPOXY-TIE ADHESIVE (SETXP) Shands Jacksonville Campus
(ACI 530-05/ASCE 5-05/TMS 402-05). 7. CONCRETE TIE BEAMS: UNLESS NOTED OTHERWISE, CONCRETE TIE BEAMS SHALL BE A E. SEE THESE DRAWINGS FOR ADDITIONAL REQUIREMENTS. (ICC-ES ESR2508)
B. SPECIFICATIONS FOR MASONRY STRUCTURES
(ACI 530.1-05/ASCE 6-05/TMS 602-05).
MINIMUM OF 16" DEEP BY THE SUPPORTING WALL WIDTH, REINFORCED WITH 2 #5
CONTINUOUS TOP AND BOTTOM AND #3 TIES AT 24" O.C. 13. MASONRY WALLS SHALL BE BRACED EITHER BY OTHER INTERSECTING WALLS OR BY
5. STEEL DECK SHALL BE GALVANIZED WITH A PROTECTIVE ZINC COATING CONFORMING TO
ASTM A924, WITH COATING DESIGNATION G90.
2. USE THE FOLLOWING ONLY WHERE SPECIFICALLY CALLED OUT ON THE
DOCUMENTS:
UF LM-4985
ANCHORAGE OR BRACING TO THE STRUCTURE ABOVE, OR TO ADJACENT WALLS, AS DETAILED A. HILTI HIT HY 150 MAX ADHESIVE (ICC-ES ESR2262)
7. STEEL DECK: REINFORCING STEEL: ON THE STRUCTURAL DRAWINGS. 6. SEE ROOF PLAN AND DETAILS FOR ROOF DECK ATTACHMENT AND FORCES IMPOSED DUE TO D. ADHESIVE ANCHORS INTO MASONRY LINTELS OR GROUT FILLED CELLS: Project Phase
A. NORTH AMERICAN SPECIFICATION FOR THE DESIGN OF COLD FORMED STEEL 1. REINFORCING STEEL: ASTM A 615, GRADE 60. UPLIFT AND DIAPHRAGM SHEAR UNDER WIND LOADING. SEE SPECIFICATIONS FOR 1. USE THE FOLLOWING (UNO): 100% Construction Documents
STRUCTURAL MEMBERS, INCLUDING 2004 SUPPLEMENT (NAS-01). 14. BLOCK LINTELS SHALL BE SPECIALLY FORMED U-SHAPED LINTEL OR LOW WEB LINTEL UNITS INSPECTION AND REPORTING REQUIRED ON ROOF DECK ATTACHMENT. A. HILTI HIT HY-150 MAX ADHESIVE (ICC-ES ESR1967)
B. SDI DESIGN MANUAL FOR COMPOSITE DECKS, FORM DECKS AND ROOF DECKS. 2. WELDED WIRE FABRIC: ASTM A 185 (FLAT SHEETS), MINIMUM YIELD STRENGTH OF 70,000 WITH REINFORCING BARS, OR PRECAST UNITS DESIGNED FOR THE WEIGHT OF MASONRY B. SIMPSON STRONG-TIE SET EPOXY-TIE ADHESIVE (SET)
Sheet Title
PSI. ABOVE AND OTHER APPLIED LOADS. 7. PROVIDE A MINIMUM END BEARING OF 2" OVER SUPPORTS. END LAPS OF SHEETS SHALL BE (ICC-ES ESR1772)(FL5550)
8. COLD FORMED METAL FRAMING: A MINIMUM OF TWO INCHES AND SHALL OCCUR OVER SUPPORTS.
A. NORTH AMERICAN SPECIFICATION FOR THE DESIGN OF COLD FORMED STEEL
STRUCTURAL MEMBERS, INCLUDING 2004 SUPPLEMENT (NAS-01).
3. MINIMUM REINFORCING STEEL CLEAR COVER (U.N.O.):
A.
B.
CONCRETE CAST DIRECTLY AGAINST EARTH .. 3"
INTERIOR SLABS ........................ 1"
15. ALL MASONRY WALLS SHOWN ON THE STRUCTURAL DRAWINGS HAVE BEEN DESIGNED TO RESIST
THE REQUIRED CODE VERTICAL AND LATERAL FORCES IN THE FINAL CONSTRUCTED
CONFIGURATION ONLY. IT IS THE RESPONSIBILITY OF THE CONTRACTOR TO ADEQUATELY
8. THE CONTRACTOR SHALL COORDINATE ALL TRADE REQUIREMENTS AND CONFIRM THE SIZE AND
LOCATION OF ALL OPENINGS. OPENINGS LARGER THAN 12", AND AS DETAILED, SHALL HAVE
STRUCTURAL DRAWING INDEX GENERAL NOTES
DESIGN LOADS: C. INTERIOR BEAMS ........................ 1-1/2" TO TIES BRACE THE WALLS FOR VERTICAL AND LATERAL LOADS THAT COULD POSSIBLY BE APPLIED STEEL FRAMING SUPPORTING ALL EDGES. SEE TYPICAL ANGLE FRAMING DETAILS. AND DRAWING
1. ROOF LIVE LOADS: D. SLABS ON GRADE ........................ 1-1/2" FROM TOP PRIOR TO COMPLETION OF LATERAL SUPPORT BY CONNECTIONS AT FLOORS OR ROOF FRAMING
A. UNIFORMLY DISTRIBUTED LIVE LOAD ON HORIZONTAL
4. WHERE REINFORCING BARS ARE NOTED AS CONTINUOUS, THE FOLLOWING SHALL BE COMPLIED
LEVELS. 9. STEEL MEMBERS SUPPORTING STEEL DECK AT THE PERIMETER OF THE BUILDING SHALL BE
CONTINUOUS, BUTT WELD PIECES WHERE SPLICES OCCUR.
Sheet No. Sheet Title/Drawing Description
INDEX
PROJECTION....................................... 20 PSF
B. SEE ROOF FRAMING PLAN FOR OTHER CONCENTRATED EQUIPMENT LOADS. WITH: 16. QUALITY ASSURANCE: ALL REINFORCED MASONRY SHALL BE TESTED/INSPECTED IN S001 GENERAL NOTES AND DRAWING INDEX
A. THE TERMINATION OF ALL CONTINUOUS REINFORCING BAR RUNS SHALL BE A CONFORMANCE WITH THE REFERENCED ACI 530/ASCE 5/TMS 402 CODES AND THE PROJECT
STANDARD HOOK UNLESS NOTED OTHERWISE. SPECIFICATIONS. QUALITY ASSURANCES SHALL MEET THE REQUIREMENTS OF SECTION 1.6, COLD FORMED METAL FRAMING:
S002 ABBREVIATIONS, SYMBOLS & WIND LOADS Scale Drawn By
2. FLOOR LIVE LOADS:
A. UNIFORMLY DISTRIBUTED LIVE LOADS: B. SPLICES IN CONTINUOUS TOP BARS, IF REQUIRED, SHALL OCCUR OVER PARALLEL TABLE 4 LEVEL B, UNLESS MORE RESTRICTIVE REQUIREMENTS ARE SPECIFIED ELSEWHERE 1. SEE NOTES ON PRIMARY CODES AND SPECIFICATIONS. S101 MACHINE PIT PLAN
TREATMENT ROOM FLOOR............................ 150 PSF CMU WALLS OR AT THE CENTER OF THE OPENING SPAN. IN THESE DOCUMENTS S102 LEVEL 1 SLAB PLAN CSH
B. CONCENTRATED FLOOR LIVE LOADS: C. SPLICES IN CONTINUOUS BOTTOM BARS, IF REQUIRED, SHALL OCCUR OVER CMU 2. ALL EXTERIOR WALLS, LINTELS, BEAMS, TRUSSES, ETC. AS WELL AS ALL INTERIOR
S103 PARTIAL STRUCTURAL CEILING PLAN
GANTRY VERTICAL LOAD........................ 24,500 LBS. WALLS OR CENTERED OVER COLUMNS. 17. TYPICAL SCHEDULED VERTICAL WALL REINFORCING SIZE AND SPACING SHALL ALSO BE BEARING WALLS, LINTELS, BEAMS, ETC. SHALL BE DESIGNED, SIGNED AND SEALED BY THE
GANTRY HORIZONTAL LOAD AT 710 CONTINUED ABOVE AND BELOW ALL OPENINGS. SUPPLIER'S DELEGATED ENGINEER REGISTERED IN THE STATE OF FLORIDA. S104 PARTIAL ROOF FRAMING PLAN
ABOVE TOP OF FOUNDATION...................... 7,300 LBS. 5. WHERE SPLICE LENGTHS ARE NOT SPECIFIED, USE 48 BAR DIAMETERS IN MASONRY AND S201 PARTIAL BUILDING SECTIONS Date Checked By
40 BAR DIAMETERS IN CAST CONCRETE.
S301 MASONRY SECTIONS AND DETAILS
3. SUPERIMPOSED DEAD LOADS: 09 DEC 2011 RWG
6. REINFORCING STEEL SHALL NOT BE TACK WELDED FOR ANY REASON. WELDED REINFORCING S302 FOUNDATION SECTIONS AND DETAILS
A. TYPICAL FLOORS................................... 25 PSF
STEEL SPLICES ARE NOT PERMITTED S303 CONCRETE SECTIONS AND DETAILS
S401 FRAMING SECTIONS AND DETAILS
11069
11662-00
12/8/2011 4:08:06 PM
Sheet No.
S001
Affiliated
ABBREVIATIONS Engineers
Revisions
A
WINDOWS, -- 5 PRESSURE. S FOOTING STEP
>500 46
DOORS AND
CURTAIN WALLS
10 27 -36 A CONCRETE
MASONRY UNIT Proton Therapy
20 26 -34 A MOMENT CONNECTION
5 50 24 -31
A A SHEAR CONNECTION
LINAC Addition
100 23 -28 A = 12'-0"
>500 22 -23 Shands Jacksonville Campus
WALLS ROOF
10 98 -- UF LM-4985
20 83 --
5P 50 64 --
WIND PRESSURE DIAGRAMS
Project Phase
REFER TO 'COMPONENTS & CLADDING WIND
100 48 -- LOAD DESIGN PRESSURE SCHEDULE' ON THIS SHEET 100% Construction Documents
>500 46 --
Sheet Title
NOTES:
1. DESIGN WIND PRESSURES SHALL BE USED IN THE DESIGN OF ALL
COMPONENTS AND CLADDING ELEMENTS COMPRISING THE
BUILDING ENVELOPE. ABBREVIATIONS,
2. REFER TO THE WIND PRESSURE DIAGRAM FOR ZONE LOCATIONS
AND EXTENTS. SYMBOLS & WIND
3. POSITIVE PRESSURES ACT TOWARD COMPONENT SURFACES AND
NEGATIVE PRESSURES ACT AWAY FROM EACH COMPONENT LOADS
SURFACE.
4. LINEAR INTERPOLATION BETWEEN EFFECTIVE WIND AREAS MAY
BE USED TO OBTAIN THE REQUIRED COMPONENT AND CLADDING Scale Drawn By
DESIGN WIND PRESSURE.
5. VALUES FOR OVERHANGS INCLUDE PRESSURE CONTRIBUTIONS 12" = 1'-0" CSH
FROM BOTH UPPER AND LOWER SURFACES.
6. DIMENSION A = 12' - 0"
Date Checked By
11662-00
12/8/2011 4:08:07 PM
Sheet No.
S002
Affiliated
PLAN NOTES
PLAN NOTES ARE TYPICAL FOR THIS DRAWING UNLESS SPECIFICALLY
DESIGNATED OTHERWISE.
REFER TO DRAWING S001 FOR STRUCTURAL DRAWING INDEX.
1
ENGINEER OF RECORD
ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
1'-1 1/2"
5
TYP
S303
1'-0"
4
13'-7 1/2"
COUCH BASE PL Revisions
AND ANCHOR BOLTS
11'-6"
DRIVING UNIT
BASE PL AND
OUTLINE OF WALLS
ANCHOR BOLTS
ABOVE (TYP)
1
S201
21'-6"
ISOCENTER
8'-0"
5
11'-5"
TOSL -4'-3 1/4"
Project
TOSL -6'-3 1/4"
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
3'-0"
8
S302
1'-0"
Proton Therapy
2'-0" 10" 1'-0" 10'-6" 8'-0" 1'-0" 3'-4" 2'-0"
LINAC Addition
2'-5"
S201
MACHINE PIT
PLAN
Scale Drawn By
As indicated CSH
Date Checked By
11662-00
MACHINE PIT PLAN
12/8/2011 4:08:07 PM
N
Sheet No.
SCALE: 3/8" = 1'-0"
S101
Affiliated
11'-8 1/2"
REFERENCE ELEVATION TOP OF FOOTINGS (TOF) -2-0" UNO. AA26002009
1 7
S302 7 NOTE:
Structural
TEMPORARY SHORING OF CONCRETE SLAB ALL EXTERIOR 8" CMU WALLS SHALL BE WT1, UNO.
S302 8 Structural Engineers Group, Inc.
REQUIRED WHENEVER VERO MACHINE IS 4114 Sunbeam Rd, Bldg 200
TRANSPORTED ACROSS THE FLOOR. Jacksonville, Florida 32257
2
EJ
INFILL OPENING CUT IN EXISTING WALL W/ SOLID HEAVY WEIGHT CMU BLOCK. Tel 904.262.4000 Fax 904.262.4100
9 www.segonline.com
Cert of Auth - 6523
S EG
8" CONCRETE SLAB; REINFORCE W/ #6 @ 6" OC EACH WAY TOP AND
10
BOTTOM. ELEVATION TOP OF SLAB= 0'-0" (DATUM) MATCH EXISTING
2 CONCRETE. ENGINEER OF RECORD
4'-0" 3'-4" ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
S302 CONTRACTOR SHALL COORDINATE EDGE OF STRUCTURE WITH ARCHITECTURAL
11
DRAWINGS TO SUIT EACH TYPE OF ARCHITECTURAL FINISH.
1'-1 1/2"
12
13
9 6" DIAMETER PVC CONDUIT PIPE CAST IN MID SLAB DEPTH SPACED AT 18" OC.
1
4"
14
S302 13
3 14
1'-0"
8'-0"
SIM
11
3'-0"
S302
WF2.0
TOSL 0'-0"
TOSL 0'-0" 1'-2" 1'-2"
2
S302 3
3'-0" Revisions
S303
13'-7 1/2"
10
SJ
12'-2"
1'-9 3/8" 7 EQUAL SPACES @ 18" ON CENTER 1'-9"
14
8'-4 1/2"
TOSL -1'-0"
EJ
2
SJ
1
S201
21'-6"
ISOCENTER
5
S302 "
1/2
'- 3
8 R =5
6 3
S301 S302 11
1 10
S302 S302
3'-0"
8'-8"
Project
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
11'-5"
3'-0"
10
22'-5"
8
15'-5 1/2"
S302
Proton Therapy
1'-0"
4"
LINAC Addition
Shands Jacksonville Campus
2'-5"
UF LM-4985
19'-10"
Project Phase
SIM 100% Construction Documents
8
S302 Sheet Title
LEVEL 1 SLAB
PLAN
2
S201
Date Checked By
S102
Affiliated
11
S301
SIM
PLAN NOTES
PLAN NOTES ARE TYPICAL FOR THIS DRAWING UNLESS SPECIFICALLY
DESIGNATED OTHERWISE.
REFER TO DRAWING S001 FOR STRUCTURAL DRAWING INDEX.
1
Architecture
3 REFER TO ARCHITECTURAL DRAWINGS FOR DIMENSIONS, DETAILS AND LOCATIONS OF Walker Architects, Inc.
INTERIOR PARTITIONS, DOORS AND WINDOWS, AND TO VERIFY DEPTH AND EXTENT OF 4055 NW 43rd St, #28
SLAB DEPRESSIONS. Gainesville, Florida 32606
Tel 352.672.6448 Fax 352.672.6468
www.walker-arch.com
4 LIFT CAPACITY = 11,000# (MINIMUM). AA26002009
Structural
Structural Engineers Group, Inc.
4114 Sunbeam Rd, Bldg 200
Jacksonville, Florida 32257
Tel 904.262.4000 Fax 904.262.4100
www.segonline.com
Cert of Auth - 6523
S EG
ENGINEER OF RECORD
ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
MAINTENANCE OP'G
BELOW
4
W8x31
W8x31
W8x31
13'-7 1/2"
Revisions
1
S201
ISOCENTER
11
S301
4
S401
11'-5"
Project
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
12'-4" 12'-4"
MAINTENANCE OP'G
BELOW
Proton Therapy
LINAC Addition
Shands Jacksonville Campus
UF LM-4985
Project Phase
100% Construction Documents
Sheet Title
PARTIAL
STRUCTURAL
CEILING PLAN
2
S201 Scale Drawn By
As indicated CSH
Date Checked By
11662-00
CEILING PLAN N
12/8/2011 4:08:08 PM
S103
Affiliated
6
S301
PLAN NOTES
PLAN NOTES ARE TYPICAL FOR THIS DRAWING UNLESS SPECIFICALLY
DESIGNATED OTHERWISE.
6
REFER TO DRAWING S001 FOR STRUCTURAL DRAWING INDEX.
S401 1
5'-10"
REFERENCE ELEVATION TO TOP OF STEEL JOISTS IS 16'-2 1/2". FOR TOP OF JOISTS OR
2
BEAMS AT OTHER ELEVATIONS, TOP OF STEEL IS NOTED THUS (-2 1/2") IN RELATION TO THE
REFERENCE ELEVATION.
Architecture
Walker Architects, Inc.
ROOF DECK SHALL BE GALVANIZED 1.5B20 STEEL ROOF DECK OR APPROVED EQUAL A
3 4055 NW 43rd St, #28
CONFORMING TO SDI SPECIFICATIONS. REFER TO ROOF DECK DETAILS FOR Gainesville, Florida 32606
12K3 ATTACHMENT PATTERN: 2 Tel 352.672.6448 Fax 352.672.6468
www.walker-arch.com
AA26002009
USE STANDARD SJI CONTINUOUS TOP AND BOTTOM CHORD BRIDGING AT BRIDGING LINES
4 INDICATED. FOR ROOF JOISTS, PROVIDE AN ADDITIONAL SINGLE LINE OF BOTTOM CHORD
BRIDGING NEAR THE FIRST BOTTOM CHORD PANEL POINT AT EACH END OF ALL JOISTS. Structural
Structural Engineers Group, Inc.
ADD CONTINUOUS BOLTED `X' BRIDGING WHERE INDICATED AND AS OTHERWISE REQUIRED
4 4114 Sunbeam Rd, Bldg 200
TO COMPLY WITH SJI. Jacksonville, Florida 32257
12 Tel 904.262.4000 Fax 904.262.4100
S401
5
CONTRACTOR SHALL VERIFY MECHANICAL OPENING DIMENSIONS WITH EQUIPMENT
SUPPLIED. NOTIFY ENGINEER AND ARCHITECT IF DIMENSIONS VARY MORE THAN 1'-0".
www.segonline.com
Cert of Auth - 6523
S EG
EXTERIOR MASONRY WALLS SHALL BE WT1, UNO. MASONRY WALL TYPES CONTINUE FROM
6
12K3 FLOOR PLAN BELOW UNLESS NOTED OTHERWISE ON THE PLANS. IF NOT MARKED OR ENGINEER OF RECORD
CONTINUED FROM THE LOWER FLOOR, WALLS SHALL BE WT1. ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
12K3
7
S401
8
S401
Revisions
TYP
12K3
5'-10"
1
S201
11
S301
6
S401
Project
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
Proton Therapy
LINAC Addition
Shands Jacksonville Campus
UF LM-4985
Project Phase
100% Construction Documents
Sheet Title
PARTIAL ROOF
FRAMING PLAN
Scale Drawn By
As indicated CSH
Date Checked By
S104
Affiliated
1 Engineers
TOW 18'-0"
2 3
S401 S201
TOW TOW
14'-10" 14'-10"
3'-4"
3'-4"
1
W8x31 W8x31 W8x31 2'-3" 2'-0" 5'-0" 2'-0"
S401 3
4'-3" S401
5'-8 7/8" 5'-8 7/8" Architecture
Walker Architects, Inc.
4055 NW 43rd St, #28
Gainesville, Florida 32606
11'-6"
Tel 352.672.6448 Fax 352.672.6468
11'-6"
www.walker-arch.com
5'-0" 7'-0" AA26002009
1
12'-4" 12'-4" 6'-5" 12'-0" 6'-7 1/2"
S302 Structural
7 Structural Engineers Group, Inc.
4'-0" 25'-0 1/2" VARIES
S302 4114 Sunbeam Rd, Bldg 200
(4'-6" MAX) Jacksonville, Florida 32257
Tel 904.262.4000 Fax 904.262.4100
= 10'-7"
.
LEVEL 1 -
www.segonline.com
Cert of Auth - 6523
S EG
LEVEL 1 -
TOSL
8"
TOSL
0'-0"
1'-6"
0'-0"
1'-6"
ENGINEER OF RECORD
ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
2'-9 1/4"
2'-9 1/4"
MACHINE MACHINE
PIT FLOOR PIT FLOOR
-4'-3 1/4"
1'-6"
-4'-3 1/4"
1'-6"
SIM
9
1'-0" 19'-6" 1'-0"
9 S302
S302 1'-0" 21'-6"
1 SECTION 2 SECTION
Revisions
PL 3/4x21x1'-9"
14'-10"
"
/16
1'-9"
PROVIDE HOLES TO
MATCH PROVIDED
EXACTRAC CEILING Project
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
12'-10" 1'-9"
Proton Therapy
LINAC Addition
3/4"
PARTIAL BUILDING
SECTIONS
Scale Drawn By
As indicated CSH
Date Checked By
11662-00
12/8/2011 4:08:10 PM
Sheet No.
S201
Affiliated
VERTICAL REINFORCING:
SEE CMU WALL REINFORCING SCHEDULE
USE OPEN BOTTOM LOW WEB Engineers
BOND BEAM BLOCK WHERE Affiliated Engineers, Inc.
FOR REQUIRED SPLICE LENGTHS. INDICATED ON DRAWINGS OR 3007 SW Williston Rd.
HOLD IN POSITION AT INTERVALS OF Gainesville, Florida 32608
VERTICAL NOTES: AS REQUIRED. Tel 352.376.5500 Fax 352.375.3479
REINFORCING APPROXIMATELY 4'-8". www.aeieng.com
1. SEE PLAN & SCHEDULE FOR FILLED CELL LOCATIONS,
FL Cert of Auth - 0005140
REINFORCING SIZE, AND SPACING.
2. BED WEBS AS WELL AS FACE SHELLS ADJACENT TO PLACE METAL LATH UNDER BOND BEAM
CELLS TO BE FILLED TO PREVENT GROUT LEAKAGE TO TOP OF FOOTING, FLOOR VERTICAL REINFORCING TO CONFINE GROUT OVER OPEN CORES CORES CONTAINING REINFORCING
COVER UNITS WITH ADJACENT CELLS UNLESS ADJACENT CELLS ARE ALSO OR PREVIOUSLY OR OTHERWISE PLUG CORES IN SUCH A ARE FILLED SOLIDLY WITH GROUT;
ABSORPTIVE PAPER TO BE FILLED. GROUTED WALL MANNER THAT GROUT WILL KEY INTO VERTICAL CORES SHALL PROVIDE A
3. SEE SPECIFICATIONS AND GENERAL NOTES FOR GROUT CELL BELOW APPROXIMATELY 1". USE CONTINUOUS CAVITY FREE OF
REQUIREMENTS. OF BUILDING PAPER IS NOT ALLOWED. MORTAR DROPPINGS. TOP OF
SAW CUT 4" WIDE x 3" HIGH
4. ALL REINFORCING SHALL BE TIED OR SUPPORTED SLOT AT CELLS TO BE FILLED PREVIOUS LIFT SHALL ALSO BE
SECURELY IN POSITION. INSPECT REINFORCING AND OR USE PREFORMED BLOCK. REINFORCING IN BOND CLEAN. CLEAN-OUTS ARE NOT
GROUT TEST PRISM 3 CELL CONTINUITY PRIOR TO CLOSING CLEAN-OUTS. REQUIRED FOR LOW LIFT GROUTING.
BEAMS SHALL BE SET IN
1/2 x 3 1/2 x 7 5. GROUTING SHALL NOT COMMENCE UNTIL 48 HOURS HAS PLACE (ON CHAIRS) AS
ELAPSED SINCE THE SECTION OF THE WALL TO BE WALL IS LAID UP.
GROUTED HAS BEEN COMPLETED TO ITS FULL HEIGHT.
6. A GROUT PUMP SHALL BE USED TO PLACE GROUT FOR
THE HIGH LIFT METHOD.
7. PLACE THE FIRST LIFT OF GROUT FOR A SECTION OF
WALL TO MAXIMUM HEIGHT OF 5 FEET, VIBRATING THE FOOTING, FLOOR PLACE GROUT IN LIFTS NOT TO EXCEED 4'-8",
GROUT IN EACH FILLED CELL WITHIN 10 MINUTES OF OR PREVIOUSLY PUDDLE AND VIBRATE IN PLACE. HOLD DOWN
TAPE GROUTED WALL 1 1/2" BETWEEN LIFTS.
PLACEMENT.
8. PLACE THE NEXT SUCCESSIVE LIFT NOT LESS THAN 30
FRONT BLOCK NOT MINUTES NOR MORE THAN 60 MINUTES LATER AND NOTES:
Architecture
3 1/2 x 3 1/2 x 5/8 PLACE MORTAR ON CROSS WEB ADJACENT TO CORES Walker Architects, Inc.
SHOWN FOR CLARITY VIBRATE 12 INCHES INTO THE PREVIOUSLY PLACED LIFT. 1. CLEAN OUT MORTAR DROPPINGS AT MIDDAY AND AT THE END
WOOD PALLET THAT WILL BE GROUTED TO PREVENT LEAKAGE OF 4055 NW 43rd St, #28
(NON-ABSORPTIVE) USE SPACERS OR WIRE TIES 9. CONTINUE PLACING SUCCESSIVE LIFTS UNTIL OF EACH WORKING DAY. PLACE AND POSITION CELL GROUT INTO ADJACENT CELLS. Gainesville, Florida 32606
NOTES:
TO HOLD REINFORCING IN GROUTING OF SECTION OF WALL IS COMPLETED TO REINFORCING AFTER FINAL CLEANOUT. Tel 352.672.6448 Fax 352.672.6468
PROPER POSITION. SPACE AT THE TOP. 2. PLACE REINFORCING IN CELLS AFTER LAYING MASONRY. 1. CONSTRUCT WALL IN LIFTS, GROUTING CELLS AND HORIZONTAL BOND BEAMS
www.walker-arch.com
NOTES: 10. SEE CMU WALL REINFORCING SCHEDULE FOR AA26002009
INTERVALS NOT EXCEEDING 3. SEE GENERAL NOTES AND SPECIFICATIONS FOR INSPECTION PRIOR TO LAYING BLOCK ABOVE.
1. TEST PER ASTM C1019.
192 BAR DIAMETERS OR 10'-0". REQUIRED SPLICE LENGTHS. & ADDITIONAL REQUIREMENTS.
2. SEE SPECIFICATIONS FOR FREQUENCY REQUIREMENTS.
2. BOND BEAMS MAY NOT OCCUR IN ALL LIFTS. REFER TO PLANS AND Structural
METHOD OF CASTING GROUT TESTING PRISMS WALL SECTIONS FOR BOND BEAM LOCATIONS. Structural Engineers Group, Inc.
METHOD OF HIGH LIFT GROUTING CLEAN-OUT OPENINGS AT FILLED CELLS 4114 Sunbeam Rd, Bldg 200
Jacksonville, Florida 32257
REQUIRED WHERE HIGH LIFT GROUTING IS USED METHOD OF LOW LIFT GROUTING Tel 904.262.4000 Fax 904.262.4100
(4'-8" MAXIMUM LIFT) www.segonline.com
Cert of Auth - 6523
S EG
ENGINEER OF RECORD
1 2 3 4
ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
CLR
4"
JAMB CONCRETE OR
REINFORCING CMU ROOF BEAM
STANDARD SASH BLOCK,
SCHEDULED
EACH SIDE OF CONTROL JOINT PREFABRICATED
SPLICE
PREMOLDED DUR-O-WAL RAPID PREFABRICATED TEE
PLAN DETAIL
SCHEDULED
CONTROL JOINT WITH WIDE CORNER JOINT JOINT REINFORCING
SPLICE
FLANGE, OR APPROVED EQUAL REINFORCING
SEE PLAN FOR CMU LINTEL Revisions
WALLS & REINFORCING REINFORCING DOWEL W/ STANDARD
HOOK OR TERMINATE
VERTICAL WITH A
STANDARD HOOK
REINFORCED CELL EACH
SIDE OF CONTROL JOINT, WALL REINFORCING AT CORNERS / INTERSECTIONS
CLR
FILL WITH GROUT
2"
1'-4" CMU WALL CMU WALL
ADD CORNER BARS TO ALL HORIZONTAL WALL REINFORCING
REINFORCING, LAP 48 BAR DIAMETERS (HOOKS NOT SHOWN), 5 TYP
SEE PLAN S401
7 SCHEDULED
S301 REINFORCING
48 BAR DIAMETER
CONTROL JOINT -
24'-0" OC, MAX ON EXTERIOR WALLS
LAP, TYP
CLR
28'-0" OC, MAX ON INTERIOR WALLS
2"
OR SCHEDULES FOR BEAM SLAB ON GRADE
DO NOT PLACE OVER OPENINGS OR
SCHEDULED
CONTINUE THROUGH CAST IN PLACE REINFORCING
SPLICE
BEAMS, UNO
SCHEDULED
APPLY SEALANT ON BOTH SIDES CONCRETE OR FOOTING
MASONRY BEAMS
SPLICE
NOTES:
1. SEE PLAN FOR LOCATIONS. SILL
2. INTERRUPT HORIZONTAL JOINT REINFORCING & CMU BOND REINFORCING
BEAM REINFORCING AT CONTROL JOINTS.
DOWEL WITH
METHOD OF WALL CONTROL JOINT (WCJ) CONSTRUCTION BEAM REINFORCING AT CORNERS / INTERSECTIONS
CLR
REINFORCING SIZE AND STANDARD HOOK
3"
SPACING SAME AS WALL
5 DETAIL 6 DETAIL
ADJACENT TO OPENING
(UNO)
NOTES:
1. REFER TO PLANS AND SCHEDULES FOR SIZE, NUMBER AND LOCATION OF
VERTICAL REINFORCING.
2. CLEAN-OUTS AND HJR NOT SHOWN BUT ARE REQUIRED, SEE SPECIFICATIONS. Project
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
EQ
1 PER CELL,
DIAGRAMMATIC ELEVATION
SEE SCHEDULE
WINDOW OPENING IN CMU (DOOR SIMILAR)
Proton Therapy
LINAC Addition
t
t
CLR
WALL THICKNESS VERTICAL MAXIMUM SPLICE NO SCALE NO SCALE 100% Construction Documents
S301 S301
WALL TYPE "t" REINF. SPACING LENGTH HJR SPACING REMARKS
Sheet Title
WT1 7 5/8" 1-#6 16" 42" 16"
2 PER CELL,
SEE SCHEDULE
NEW EXIST
MASONRY
MASONRY WALL JAMB REINFORCING SCHEDULE CONSTRUCTION SECTIONS AND
EXPANSION JOINT
WALL MASONRY NUMBER OF REINFORCED CELLS PER JAMB METAL WALL ANCHORS FASTENED MATERIAL DETAILS
TYPE OPENING OPENING IN OPENING IN SPLICE LENGTH TO EXIST COLUMN W/ 1/4" TAPCON
WIDTH EXTERIOR WALL INTERIOR WALL ANCHORS. SPACE TO MATCH WALL
*SEE SEPARATE DETAIL/SCHEDULE FOR JAMB REINFORCING. HORIZ JOINT REINFORCING, TYP EXIST WALL CONCRETE WALL, Scale Drawn By
UP TO 3'-4" 1-#6 PER CELL 1-#6 PER CELL 42"
IN 2 CELLS IN 1 CELL NOTES: SEE PLAN WALL,
WT1 SEE PLAN As indicated CSH
1. VERTICAL REINFORCING SHALL RUN FROM FOOTING TO 4" CLEAR FROM TOP OF UPPERMOST
UP TO 6'-8" 2-#6 PER CELL 1-#6 PER CELL 42" SUPPORTED BEAM (ROOF BEAM OR OPENING LINTEL). VERTICALS MAY BE LAP SPLICED AS
IN 2 CELLS IN 2 CELLS
REQUIRED FOR EASE OF BLOCK INSTALLATION. PROVIDE HOOKED DOWEL FROM FOOTING OR
SUPPORT BEAM AT EACH FILLED BLOCK CELL. PROVIDE HOOK AT TOP OF VERTICAL. EACH POUR OF
Date Checked By
GROUT SHALL BE STOPPED AT LEAST 1 1/2" BELOW THE TOP OF THE LAST COURSE OF BLOCK LAID
(EXCEPT AT PRECAST LINTELS).
2. SEE TYPICAL DETAILS AND CODE REQUIREMENTS FOR CLEAN-OUTS. 09 DEC 2011 RWG
WALL,
SEE PLAN
TYPICAL FILLED CELLS WITH 1 BAR OR 2 BARS PER CELL
11069
2" CLR
3/8" EXPANSION JOINT FOOTING & TOF
MATERIAL & SEALANT REINFORCING,
SCHEDULED
SEE PLAN SEE PLAN
DEPTH
EXTERIOR SLAB OCCURS AT SLAB, SEE PLAN EMBEDED PVC 3" CLR
SOME LOCATIONS, CONDUIT, SEE PLAN TYP, UNO
SEE ARCHITECTURAL/CIVIL
TOSL SJ, SEE PLAN
2" SLAB, SEE PLAN CONTINUOUS FOOTING CORNER BARS
SEE PLAN SLAB, CLR SCHEDULED
CLR
REINFORCING
2"
SEE PLAN MINIMUM LAP SPECIFIED SCHEDULED
TOSL
IF CONSTRUCTION JOINT OCCURS AT WIDTH
0'-0" SPLICE IN CONT BARS
VAPOR RETARDER
CLR
NOTES:
1'-0"
1'-0"
2"
6"
FOOTING OR 40 BAR DIAMETERS (2'-0" MIN) INTO COLUMN www.walker-arch.com
h
FOOTING. AA26002009
FILL ALL CELLS W/ GROUT 4. LAP ALL CONTINUOUS REINFORCING 40 BAR DIAMETERS (2'-0" MIN).
CLR
5. REFER TO CONCRETE SUPPORT BLOCK DETAIL FOR ADDITIONAL
3"
3" TO FINISHED FLOOR OR CONTINUOUS KEY TIES Structural
TOP OF GRADE, INFORMATION.
CLR
CLR EXTEND WWF 8" INTO VAPOR RETARDER NOTE: Structural Engineers Group, Inc.
3"
WHICHEVER IS HIGHER. 6. ALL FOOTING MARKS MAY NOT BE USED ON THIS PROJECT. 4114 Sunbeam Rd, Bldg 200
THICK SLAB a. USE ONE 2x4 KEY FOR 'h' 18"
Jacksonville, Florida 32257
COMPACTED EARTH b. USE ONE 2x6 KEY FOR 18" < 'h' 30" Tel 904.262.4000 Fax 904.262.4100
FOOTING, SEE PLAN c. USE TWO 2x6 KEYS FOR 'h' > 30" www.segonline.com S EG
SLAB STEP WALL FOOTING SCHEDULE Cert of Auth - 6523
WALL FOOTING CONSTRUCTION JOINT TOP BOTT. TIES
MARK WIDTH DEPTH BARS BARS TYPE SIZE SPACING REMARKS
WF2.0 2'-0" 1'-0" 3-#4 3-#4 B #3 12" ENGINEER OF RECORD
1 2 3
ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
EXISTING
WALL & SLAB EXISTING
3
4-#5 CONT W/ #3 3'-0" S303
11 SAWCUT SLAB AS SOON AS CUT CAN BE 3/8" EXPANSION JOINT STIRRUPS @ 12" OC SLAB,
MADE WITHOUT DAMAGING EDGES OF MATERIAL + SEALANT SEE PLAN
(1" MIN)
OTHER WIRE AT JOINT
AGAINST WALLS, COLUMNS, ETC. AS TOSL
REQUIRED TO COMPLETE SLAB JOINT. 0'-0"
FILL W/ SEALANT.
Revisions
TOSL
LEVEL 1 CONCRETE SLAB,
SEE PLAN SEE PLAN
0'-0"
#6 @ 12"
1'-6"
TOSL 1'-6"
TOSL
3'-1"
-1'-0"
-1'-0" 4-#6 @ 12" OC
3/8" EXPANSION
COMPACTED EARTH VAPOR RETARDER
1'-0"
JOINT MATERIAL
+ SEALANT 1'-6"
2'-0"
5 SECTION 6 DETAIL 7 SECTION
ADDITIONAL
INFORMATION
EW T&B
CLR
2"
S302 NO SCALE S302 NO SCALE S302 NO SCALE SUMP PIT
-6'-3 1/4"
CONT 2x4
KEY W/
3/8" EJ MATERIAL + SEALANT WATERSTOP
1'-6"
Project
C:\Users\phi.segonline\Documents\Revit Structure 2012 Local\UF_PROTON_11069 - STRT_phi.rvt
FLOOR SLAB
CLR
CLR
3" CLR
1"
3"
(NTS)
2x4 CONT Level 01 -
TOSL
1"
KEYWAY
0'-0" COMPACTED EARTH Proton Therapy
8"
NEW EXIST
8 Shands Jacksonville Campus
CLR
DETAIL
1"
2'-6" LAP
MACHINE EXIST
-1'-0"
PIT FLOOR
6"
10"
1'-0"
-4'-3 1/4"
2-#4 x (OPNG - 4") Scale Drawn By
2" P
3007 SW Williston Rd.
TY
14 GAGE WIRE TIE
2" P
Gainesville, Florida 32608
TY
Tel 352.376.5500 Fax 352.375.3479
SEE MECHANICAL FOR SIZE AND LOCATION OF PAD www.aeieng.com
3000 PSI CONCRETE SUPPORT BLOCK - FL Cert of Auth - 0005140
USE 2 @ 4'-0" OC IN CONTINUOUS WALL
4" EQUIPMENT SIZE
FOOTINGS, 1 PER 16 SQUARE FEET (4 MIN)
2-#5 x 4'-0" DIAGONAL BARS,
IN COLUMN FOOTINGS, AND @ 3'-0" OC,
1 TOP & BOTTOM INNER 2-#5 x 4'-0" DIAGONAL
EACH WAY IN SLABS, TYPICAL 1'-0" ROUGHEN TOP OF STRUCTURAL SLAB REINFORCING LAYERS, BARS, 1 EACH FACE,
MIN (FLOAT FINISH OR HEAVY BROOM) TYPICAL EACH CORNER TYP EACH CORNER WALL &
REINFORCING,
SEE MECHANICAL SEE PLAN
FOR EQUIPMENT
APPLY APPROVED
ANCHORAGE
3/4" CHAMFER BONDING AGENT
ALL SIDES
1-#4 CONTINUOUS 6x6 W2.1xW2.1
TYP
ALL SIDES
3"
CENTERED
TYP
3"
3 1/2"
UNO
4" TOSL
SEE PLAN PROVIDE 180 HOOKS ON
ALL INTERRUPTED BARS Architecture
3 1/2" 3'-0" OPENING Walker Architects, Inc.
NOTE: OPENING
NOTES: MAX 4055 NW 43rd St, #28
AS AN ALTERNATE, 3" CASTLE CHAIR 3'-0" MAX Gainesville, Florida 32606
BY AZTEC CONCRETE ACCESSORIES NOTES:
Tel 352.672.6448 Fax 352.672.6468
MAY BE USED 1. ADD HALF OF THE PARALLEL INTERRUPTED BARS ON EACH SIDE OF THE 1. ADD HALF OF THE PARALLEL INTERRUPTED BARS ON EACH SIDE OF THE www.walker-arch.com
SLAB, SEE PLAN OPENING & ENSURE THAT MINIMUM REINFORCING REQUIREMENTS ARE OPENING & ENSURE THAT MINIMUM REINFORCING REQUIREMENTS ARE MET. AA26002009
MET. ADDED REINFORCING SHALL BE NOT LESS THAN THE FOLLOWING: ADDED REINFORCING SHALL BE NOT LESS THAN THE FOLLOWING.
a. 2-#5 MIN @ SLABS 10" THICK, (1 TOP & BOTTOM, EACH WAY) a. 2-#5 MIN @ WALLS 12" THICK, (1 EACH FACE, EACH WAY)
CONTRACTOR FABRICATED CONCRETE SUPPORT BLOCK
TYPICAL EQUIPMENT PAD b. 2-#7 MIN @ SLABS > 10" THICK, (1 TOP & BOTTOM, EACH WAY) b. 2-#7 MIN @ WALLS > 12" THICK, (1 EACH FACE, EACH WAY) Structural
Structural Engineers Group, Inc.
4114 Sunbeam Rd, Bldg 200
2. EXTEND REINFORCING 48 BAR DIAMETERS BEYOND THE OPENING OR 2. EXTEND REINFORCING 48 BAR DIAMETERS BEYOND THE OPENING OR Jacksonville, Florida 32257
PROVIDE A STANDARD HOOK AT THE SLAB EDGES AROUND THE OPENING. PROVIDE A STANDARD HOOK AT THE WALL EDGES AROUND THE OPENING. Tel 904.262.4000 Fax 904.262.4100
www.segonline.com
Cert of Auth - 6523
S EG
TYPICAL SMALL SLAB OPENING TYPICAL SMALL WALL OPENING ELEVATION
ENGINEER OF RECORD
1 2 3 4
ROBERT W. GIVENS, P.E. FL. P.E. NO. 31961
1'-4"
0" TYP
2" CLR
HALF OF TYPICAL OF REQUIRED STEPS Revisions
VERTICAL SPACING SEE PLAN FOR WALL ALUMINUM GRATING,
& REINFORCING TOF SEE ARCHITECTURAL
48 BAR DIAMETER
TYP
SEE PLAN
0",
UNO
SLAB, SEE PLAN
8"
LAP, TYP
ADDITIONAL BARS
UNO
8"
AS REQUIRED TOSL
TOF
SEE PLAN
3 1/2"
MIN
SEE PLAN
SEE PLAN FOR WALL 48 BAR DIAMETER
& REINFORCING CONTINUOUS KEY STEP BARS TO
3"
h
(2x4, MIN) LAP, TYP MATCH NUMBER &
CL
SIZE OF FOOTING CONTINUOUS STAINLESS
R
REINFORCING STEEL L4x3x3/8 (LLV),
TWO LAYERS OF REINFORCING
TWO LAYERS OF REINFORCING MITER AT CORNERS AND
h TIES TO MATCH FOOTING TIES WELD JOINTS. CONNECT
TYPICAL WALL CORNER TYPICAL WALL CONSTRUCTION JOINTS W/ 3/4" SS HILTI KWIK
LAP 40 BAR , MIN BOLT 3 @ 8" FROM
CORNERS W/ REMAINDER
@ 24" OC
NOTE: TYPICAL WALL FOOTING STEP
SEE SECTIONS, DETAILS, ETC FOR REINFORCING LAYERING REQUIREMENTS, TYPICAL
NOTED THUS "$" ON PLAN
Proton Therapy
LINAC Addition
Shands Jacksonville Campus
UF LM-4985
Project Phase
100% Construction Documents
Sheet Title
CONCRETE
SECTIONS AND
DETAILS
Scale Drawn By
As indicated CSH
Date Checked By
11662-00
12/8/2011 4:08:13 PM
Sheet No.
S303
Affiliated
2"
CONTRACTOR TO HAVE
10"
TESTING AGENCY
ACCURATELY LOCATE
8"
REINFORCING IN RELATION TO 10"
2 1/2"
BEAM ANCHOR PLATE AND
EMBED
BOC BOC
7 1/2"
HDA ANCHORS. REINFORCING TOS
1'-0"
1'-4"
11'-6" IS NOT TO BE CUT. 11'-6"
(VERIFY) (VERIFY) 11'-5"
SEE 1/S401 BOC 14'-10"
1/4
3"
1/4
Architecture
1/4 1/4 EXISTING REINFORCING - Walker Architects, Inc.
2"
CONTRACTOR TO HAVE PL 3/4x6x1'-0" W/ 4055 NW 43rd St, #28
TESTING AGENCY ACCURATELY 2 HILTI HDA-PF 30M16x190/40 Gainesville, Florida 32606
1 1/2" 4 1/2" 4 1/2" 1 1/2"
2 1/2"
3" Tel 352.672.6448 Fax 352.672.6468
LOCATE REINFORCING IN CENTER PLATE OVER www.walker-arch.com
2" 1'-0" 2" RELATION TO BEAM ANCHOR MAINTENANCE OPENINGS
W8x31, SEE PLAN AA26002009
W8x31 6" PLATE AND HDA ANCHORS. 1'-0" ON LINAC FLOOR
1'-4" REINFORCING IS NOT TO BE
W8x31, SEE PLAN CUT. Structural
MAINTENANCE HOIST Structural Engineers Group, Inc.
CONNECTOR AS REQUIRED. 4114 Sunbeam Rd, Bldg 200
Jacksonville, Florida 32257
PL 1x16x1'-4" W/ 4-HILTI HDA-PF 37 Tel 904.262.4000 Fax 904.262.4100
M20x250/100 ANCHORS
SPACED AT 12" OC EACH WAY
www.segonline.com
Cert of Auth - 6523
S EG
ENGINEER OF RECORD
BOND BEAM W/ CONNECT OVERLAPPED SIDE LAPS WITH #10 HEX HEAD
SEE PLAN
#5 CONTINUOUS SELF-DRILLING SCREWS. SEE ROOF PLAN FOR NUMBER
TOW CONTINUOUS L4x4x1/4 W/ REQUIRED IN EACH SPAN BETWEEN SUPPORTS.
3/16 2
TYP 18'-0" 3/4" INSERTS @ 32" OC, INTERLOCKED STANDING SIDE LAPS NOT ACCEPTABLE.
EXISTING CONCRETE
3/16 2 EMBED 4"
TOW WALL, SEE PLAN
1'-4"
18'-0"
16'-8"
ROOF DECK, SEE PLAN TOJ PROVIDE @ 6" OC AT
SECTION STRUCTURAL
DISCONTINUED EDGES AND SUPPORT AT
1'-4"
SEE PLAN
2 1/2"
MIN
4"
REINF
1'-4"
2-#5 CONTINUOUS
TOP & BOTTOM,
#3 TIES @ 16" OC
END LAP
JOIST, SEE PLAN JOIST, SEE PLAN
MIN
2"
EXTEND WALL
L5x3x3/8x0'-10" (LLH) W/ 2- 1/2" x 6" REINFORCING
HCA, PROVIDE 1/4" VENT HOLE @ TO PARAPET L6x4x1/2x1'-0" (LLV) W/ 2-3/4"
7 5/8"x16" CONCRETE BEAM CENTER OF ANGLE INSERTS @ 8" OC, EMBED 6" MIN
BRIDGING, SEE PLAN JOIST TOP CHORD
W/ 2-#5 TOP & BOTTOM,
PLAN SHOWN,
#3 TIES @ 16" OC
BEAM SIMILAR
WALL, SEE PLAN TYPICAL BRIDGING ANGLE L3x3x1/4 x JOIST
7 5/8"
DEPTH W/ 2-3/4" EXPANSION ANCHORS NOTES: SECTION
W/ 5" EMBEDMENT, MIN CONNECT DECK TO ALL SUPPORTS BY WELDING AT SIDE LAPS AND AT
INTERMEDIATE RIBS AT THE FOLLOWING SPACINGS:
WALL, SEE PLAN
PATTERN 'A' AT 6" ON CENTER (EVERY RIB)
PATTERN 'B' AT 12" ON CENTER (EVERY OTHER RIB)
1'-4".
A
SEE ROOF PLAN FOR LOCATIONS OF WELDING PATTERN - SHOWN THUS
PRECAST LINTEL BLOCK & THE NUMBER OF SIDE LAP SCREWS BETWEEN SUPPORTS - SHOWN THUS
1'-4".
SEE ARCHITECTURAL OR
MECHANICAL DRAWINGS
FOR SIZE AND LOCATION OF
SEE PLAN FOR ROOF DECK.
WELD WITH 5/8" PUDDLE
Proton Therapy
LINAC Addition
DIRECTION OF
TYPICAL REINFORCEMENT USE THIS DETAIL WHEN SUPPORTING MISCELLANEOUS MINOR AEI Project No.
CONCENTRATED LOADS, OR FOR LOADS SHOWN ON FRAMING PLANS,
AT SMALL OPENINGS IN ROOF DECK FROM JOIST BOTTOM CHORDS.
11662-00
12/8/2011 4:08:14 PM