INSTRUCTION MANUAL Read through the following safety instructions contained in this

manual before using the instrument.


Specification Instrument layout
Accuracy [ guaranteed at temperature & humidity: 23545 85% RH(*1) ] Instrument layout
#DANGER Function Range Accuracy Max. input voltage
Thank you for purchasing our instrument KEW1030. Never make measurement on a circuit in which electrical potential to ground over 4V 1.3%rdg 5dgt (50/60Hz)
Before using the instrument,read this manual thoroughly to obtain the maximum 600V exists. ACV 40V 1.7%rdg 5dgt ( 400Hz)
performance of this instrument and ensure the correct measurement. Do not attempt to make measurement in the presence of flammable gasses. Auto-ranging(*2) 400V 1.6%rdg 5dgt (50/60Hz)
Otherwise, the use of the instrument may cause sparking, which can lead to an 600V 2.0%rdg 5dgt ( 400Hz)
explosion. 400mV
Never attempt to use the instrument if its surface or your hand is wet. Otherwise, DCV
4V
0.8%rdg 5dgt
you may get electrical shock. Auto-ranging(*2)
40V
Never open the Bottom case and Battery cover during a measurement. 400V

Do not exceed the maximum allowable input of any measuring range. 600V 1.0%rdg 5dgt
Never try to make measurement if any abnomal conditions, such as broken 400
case is noted. 4k
The instrument should be used only in its intended applications or conditions 40k 1.0%rdg 5dgt Test pin (input terminal (+); red)
Auto-ranging 400k Test lead (input terminal (); black)
Otherwise, safety functions equipped with the instrument do not work, and
4M Connected to the negative () side or the earth of the circuit.
instrument damage or serious personal injury may be caused. 40M 2.5%rdg 5dgt
Keep your fingers and hands behind the barrier during measurement. DC 600V Protective cover
Diode check Test voltage:approx. 0.3V 1.5V Covering the Test pin for safety purpose.
Diode check/ AC 600Vrms(sin)
Continuity Buzzer sounds when Penlight
Continuity Check
#WARNING Check resistance is 120 or less. Barrier
Never attempt to make any measurement if any abnormal conditions, such as 50nF 3.5%rdg 10dgt It is a part providing protection against electrical shock and ensuring the
broken case and exposed metal parts are present on the instrument or test lead. Capacitance
500nF
minimum required air and creepage distances.
Do not install substitute parts or make any modification to the instrument. Return 5uF 3.5%rdg 5dgt
Auto-ranging Function switch
the instrument to your local Kyoritsu distributor for repair or re-calibration. 50uF

Do not turn the function switch with plugged in test leads connected to the circuit 100uF 4.5%rdg 5dgt
5Hz
OFF Power off (Battery will not be wasted.)
under test. AC voltage (ACV) Frequency (Hz) DUTY(%)
Do not try to replace the batteries if the surface of the instrument is wet. 50Hz
0.1%rdg 5dgt
Frequency 500Hz
Always switch off the instrument before opening the battery compartment cover Measurable input: Switches by pressing the "SELECT" key.
Auto-ranging 5kHz
for battery replacement. 1.5Vrms or more
50kHz DC voltage (DCV) REL (relative value display)
200kHz
#CAUTION
PEN TYPE DIGITAL MULTIMETER Always set the Function switch to the appropriate position before making
DUTY(pulsewidth/ pulse
cycle)
0.1 99.9%
2.5%rdg 5dgt(Accuracy is
guaranteed up to 10kHz.)
Switches by pressing the "SELECT" key.
Resistance Diode check Continuity check
measurement. Note:
1030 Do not expose the instrument to the direct sun, high temperatures and humidity Following abbreviations are used in above table.
rdg is an abbreviation of "reading", and it means the indicated value at a Switches by pressing the "SELECT" key.
or dew. Capacitance REL (relative value)
measurement.
This instrument is designed for in-door use. It can be used under the temperature dgt is an abbreviation of "digit", and it means the figure to be displayed at the
between 0 and 40 without impairing its safety characteristics. rightmost digit.
KYORITSU ELECTRICAL INSTRUMENTS This instrument doesn't have dust/water-proof construction. Do not use the (*1): Except for 40M range at Ohm function.
(*2): At Voltage function, the Auto-ranging function is released by pressing the
Switches by pressing the "SELECT" key.

WORKS, LTD. instrument in dusty area or where it easily gets wet. It may lead to failure of the
instrument.
SELECT key. To measure a voltage again, turn the Function switch to the "OFF"
position once. Then set it to the Voltage function again.
LIGHTTurning on the Penlight. Set the Function switch to this position first,
and then turn it to any desirable function position. Then the Penlight turned
Set the Function switch to "OFF" position after use. Remove the batteries if the General specification on and illuminates the test point. (Measurement cannot be performed in this
instrument is to be stored and will not be in use for a long period. Method of operation : method switch position.)
Use a damp cloth and detergent for cleaning the instrument. Do not use abrasives Display : Liquid crystal display (max. 3999 counts)/Units/ Marks HOLD key
Safety warnings # or solvents. Over-range indication : "OL" displayed when exceeding the measuring range.
(except for AC/DC 600V range) Freezing the indicated value.
This instrument has been designed, manufactured and tested according to IEC Range switching : Fully-automatic range Turning on the LCD backlight. (Press this key at least 2 sec.)
61010: Safety requirements for Electronic Measuring apparatus, and delivered in Measurement categories (Over-voltage categories) (Single range is available at Continuity, Diode check and SELECT key
the best condition after passed the inspection. To ensure safe operation of measuring instruments, IEC61010 establishes safety DUTY range.) Switching the measurement modes. / Hz/ DUTY and/ /
This instruction manual contains warnings and safety rules standards for various electrical environments, categorized as CAT to CAT IV, and Range shifts to upper range:4000 counts or more. Enable/ Disable the REL function. (Only at DCV/ Capacitance)
called measurement categories. These are defined as indicated below. Range shifts to lower range:less than 360 counts.
which must be observed by the user to ensure safe operation Sample rate : twice per second LCD indication
of the instrument and retain it in safe condition. Therefore, read Higher-numbered categories correspond to electrical environments with greater Functional construction : OFF/ ACV/ DCV// Capacitance Low battery warning Diode check
through these operating instructions before using the instrument. momentary energy, so a measuring instrument designed for CAT environments Key : HOLD/ Hz/ DUTY/ / /
Auto-ranging Continuity check
can endure greater momentary energy than one designed for CAT . REL (only at DCV and Capacitance ranges)
#WARNING Power source
Low battery warning
: Button type battery LR44(SR44)1.5V x 2
:" " mark is displayed at 2.4V0.2V or less.
DC
REL
Read through and understand the instructions contained in this manual before


CAT Primary electrical circuits of equipment connected to an AC electrical Dimension : 190(L) x 39(W) x 31(D)mm AC
using the instrument.
outlet by a power cord. Weight : Approx. 100g (including batteries) Polarity Unit
Save and keep the manual at hand to enable quick reference whenever necessary. Location for use : Altitude up to 2000m, in-door use
The instrument is to be used only in its intended applications. CAT Primary electrical circuits of the equipment connected directly to the Hold
Operating temperature : 040, relative humidity 85% or less
Understand and follow all the safety instructions contained in the manual. distribution panel, and feeders from the distribution panel to outlets. & humidity range (no condensation)
The RESPONSIBLE BODY shall be made aware that, if the equipment is used Storage temperature : -2060, relative humidity 85% or less Protective cover
in a manner not specified by the manufacturer, the protection provided by the Designed to meet CAT 600V when the cap is attached to the test lead and to meet & humidity range (no condensation)
equipment may be impaired. CAT 600V when the cap is not attached to the test lead. Accessories : Carrying case x 1
Button type battery LR44(1.5V) x 2 #CAUTION
Instruction manual x 1 Do not apply excessive force to the Test pin and the Protective cover.
The symbol#indicated on the instrument means that the user must refer to the Standards (Safety) : IEC/EN 61010-1 Be careful not get hurt by the tip of the Test pin when setting or releasing the
related parts in the manual for safe operation of the instrument. Features Measurement category (CAT) 600V Protective cover.
Be sure to carefully read the instructions following each #symbol in the manual. Pollution degree 2 Designed to meet CAT III 600V when the cap is attached to the test lead and to
This instrument is a pen-type digital multimeter and can measure: AC/DC voltage, IEC/EN 61010-031 meet CAT II 600V when the cap is not attached to the test lead.
#DANGER is reserved for conditions and actions that are likely to cause resistance, capacitance and frequency/duty ratio. It also provides continuity check (EMC) : EN 61326
serious or fatal injury. and diode check functions. Electrical characteristics Use the Protective cover to cover the Test pin when carrying or storing the
#WARNINGis reserved for conditions and actions that can cause serious Temperature &
humidity range
: 235, relative humidity 85% or less
(no condensation) instrument.
or fatal injury.
Designed to meet the following safety standards. (guaranteed accuracy)
#CAUTION is reserved for conditions and actions that can cause injury or IEC61010-1 measurement category (CAT) 600V Supply voltage range : 3.4V till the " " mark is displayed. 1Method of setting the Protective cover
instrument damage. (guaranteed accuracy)
IEC61010-031 (for hand-held Probe assemblies) Pinch the tip of the Protective cover, and pull it towards the tip direction.
Insulation resistance : 10M or more/ DC1000V Then turn it 90 degrees as shown in the figure below to match the marks on the
Please refer to following explanation of the symbols used on the instrument Double molded main body and Function switch provide comfortable single handed (between electrical circuit and case enclosure)
and in this manual. grip. Withstand voltage : AC5.55kVrms, sine wave (50/60Hz for 1 min.) cover and on the instrument body.
Penlight illuminates brightly the point to be measured. (between electrical circuit and case enclosure)
# User must refer to the explanations in the instruction manual. Backlight LCD is highly visible, even in darkness. Overload protection : 720V (AC/DC) for 10 sec. at voltage function
(Over-voltage protection) 600V (AC/DC) for 10 sec. at all functions
Instrument with double or reinforced insulation REL function to check the difference (DC.V/ CAP). other than voltage function
Auto-power-off function to save battery. Rated supply voltage : DC3.0V Set CAT Release
AC Data hold function Rated power
Max. rated power
: Approx. 4mVA (when battery voltage is 3.0V)
: Approx.30mVA (when lights are on) 2M
 ethod of releasing the Protective cover
DC All ranges including Ohm range are protected against overload voltage of 600V. Continuous operating : Approx.80 hours (DCV measurement) Pinch the tip of the Protective cover, and pull it towards the tip direction.
Test lead is wrapped in its rear side compartment without difficulty. time Approx.15 hours (A operation; turning the light on for Then turn it 90 degrees as shown in the above figure. Then the cover is stored
Test pin can be covered by a unique cover mechanism for safety. 10 sec. and off for 20 sec., is repeated.) automatically and the Test pin (positive terminal) appears.

Method of storing the test lead AC voltage(ACV), Frequency and DUTY ratio measurement 3 Press the SELECT key to conduct the Diode check. Battery replacement
Test lead is stored in the rear side compartment of the instrument. Connect the Test pin and the test lead to the equipment under
Cord is winded around the cord holder. 1 Set the Function switch to" "position. test. When following indication is confirmed, the diode is good.
#WARNING
2 Connect the Test pin and test lead to AC circuit as shown in the figure below to Following measurements can be done To avoid getting electrical shock, be sure to remove the measuring terminals from
measure AC voltage (ACV). by pressing the SELECT key. the equipment under test; set the Function switch to OFF position before replacing
Diode check Continuity check batteries.
#DANGER
Do not mix new and old batteries. Never mix the different kinds of batteries.
1.Forward direction 2.Backward direction Make sure to install batteries in correct polarity as marked inside.
Cathode Anode Anode Cathode Be sure to fasten the Battery case-fixing screws after the battery replacement.

3 Press the SELECT key and select the Frequency range to measure a frequency. #CAUTION
Functions In this case, the unit "Hz" is displayed on the LCD. Dispose the used batteries according to the rules, which are defined by each
(Example) (Example) community.
Auto-ranging (AUTO) Following measurements can be done Forward voltage "OL"
A function to automatically select the appropriate measurement range based on is displayed. is displayed. Set the Function switch to OFF position.
the input signal. The "AUTO" mark is displayed on the LCD while this function is by pressing the SELECT key.
ACV Frequency DUTY ratio Loosen one Battery cover-fixing screw, and remove the Battery cover.
activated. This function is not available in Diode check, Continuity check and Duty Note Replace the batteries with new ones. Make sure to install batteries in correct polarity
ratio measurements. The "AUTO" mark is not displayed. as marked inside. Always replace all two batteries with new ones at the same time.
When the forward voltage of diode is out of the range of 0.3V1.5V,
Hold function ( ) measurement may not be done. (Zener diode, LED and etc.) Put the Battery case at the original position, and fasten the screws.
A function to freeze the measured value on the LCD. (Not available in Frequency  ress the SELECT key and select the DUTY ratio range to measure a DUTY ratio
4P
measurement) (Pulse width/ Pulse cycle). In this case, the unit "%" is displayed on the LCD. Battery cover-xing screw
The " " mark is displayed on the LCD when the HOLD key is pressed.  ress the SELECT key to conduct the Continuity check. Connect the
4P
Then the measured value is frozen. Press this key again or switch the Test pin and the test lead to the equipment under test.
measurement function to others to release the Hold function. Buzzer sounds when continuity is ok. (120 or less) Button type battery
LR44(SR44) 1.5V2pcs
REL function ( )
Resistance value of 400 or less is displayed on the LCD.
A function to display the difference between the measured values (relative value) Note
on the LCD at DCV and Capacitance functions. The "" mark is displayed on the At ACV function, a few dgts may remain displayed on the LCD after removing the
LCD when the HOLD key is pressed. Then the value being measured is stored.
After that, the difference between the stored value and the measured value is input.
displayed on the LCD. Press this key again or switch the measurement function to Connect the test lead (minus terminal) to the earth side of the circuit under test.
others to release the REL function. When the circuit under test does not have the earth, any connection is allowed.

Auto-power-off function At Frequency and DUTY ratio measurement, the measurable min. input is approx.
A function to turn off the instrument when 30 min. have elapsed after the Function 1.5Vrms. Note Maintenance
switch is switched from OFF to the other measurement function. Indicated value may not be "0" after shorting the tip of the test lead.
Press the HOLD key again or switch the measurement function to others to DC voltage(DCV) measurement However, this is because of the resistance of the test lead and not a failure. Cleaning
restore from the Auto-power-off state. Use a cloth dipped in water or neutral detergent for cleaning the instrument.
Over-range indication  et the Function switch to" "position. Do not use abrasives or solvents. Otherwise, instrument get damaged,deformed
1S Capacitance measurementnF , F or discolored.
When the measured value exceeds the max. indication range, "OL" is displayed 2 Connect the Test pin to the positive (+) side of the equipment under test and the test
on the LCD. (This indication is not displayed at AC/DC 600V range.) This 1 Set the Function switch to " " position.
indication is not displayed while the Hold function is activated. lead to the negative (-) side as shown in the figure below. When test lead is
Low battery warning ( ) connected to the positive (+) side, the "-" mark is displayed on the LCD.
When the battery voltage drops to 2.4V0.2V or less, the " " mark is displayed
on the LCD.
Penlight
Set the Function switch to "LIGHT" position DISTRIBUTOR
to turn on the Penlight. Turn the switch to 2 Press the SELECT key to make the indicated value to "0" before
Penlight ON
any desirable function position. Select any connecting the test lead to the equipment under test.
(Measurement cannot be performed when Penlight function to make
measurement
3 Press the SELECT key to display a REL value (relative value).
the switch is in "LIGHT" position.) Turn the OFF Measurement function Press this key and store the initial measured value. After that, the difference between Press the SELECT key to make the indication
switch to "OFF" position to turn off the light. the stored value and the measured value is displayed on the LCD. Auto-ranging to 0.
LCD backlight function doesn't activate when this function is enabled. The first selected range will
The LCD backlight lights up by pressing down the HOLD key at any measurement be held. The relative measurement is allowed in the following range.
function other than OFF at least 2 sec.. 3 Connect
 the Test pin and the test lead to the equipment under test as shown in the
Press down this key again at least 2 sec. or turn the Function switch to OFF once to * Measuring range = Full scale value at a range initial value
Press this key again or switch the measurement function to others to release the REL figure below.
turn off the light.
function.
Note
Following measurements can be done
Penlight and LCD backlight are not turned off automatically. Be sure to turn them off
when they are not in use. by pressing the SELECT key.
When turning on/ off the LCD backlight, the " " mark is displayed on the LCD and DCV REL (relative value)
"" mark is displayed Note
the Hold function is activated. Press the HOLD button for a while to release the
function and perform the next measurement. on the LCD. Measuring time varies depending on the capacitance to be measured.
Capacitance
4 40 100
to be measured
Measurement
Resistance ( ) measurement, Diode/ Continuity check Measuring
2 sec. 7 sec. 15 sec.
time
#DANGER 1 Set the Function switch to" "position.
To prevent electrical shock to person and damage to the instrument, following 2 Connect the Test pin and test lead to the equipment under test as shown in the Kyoritsu reserves the rights to change specifications or designs desscribed in this manual
instructions must be observed. figure below. without notice and without obligationss.
The max. rated voltage to ground is AC/DC600V. Never attempt to make
measurement on a circuit in which electrical potential to the ground exceeding this
voltage exists.
The max. input voltage is DC600V/AC600Vrms (sin). Never attempt to make any
measurement on a circuit in which electrical potential exceeding this voltage exists.
Do not operate the Function switch during a measurement.
Never make a measurement with the Bottom case is removed.
Keep your fingers and hand behind the barrier (see 4-1) of the instrument and test
lead.
Be careful not to short-circuit the line under test with the metal part of the instrument
or the test lead during a measurement.
Never make measurement on an energized circuit at Resistance, Diode check,
Continuity check and Capacitance function of this instrument.
Always attach the cap firmly when using the test lead under CAT environment.
When the test lead is connected to the instrument, the lower category either of them
belongs to is applied. 2-13 92-1634E