Sei sulla pagina 1di 4

Detection of Sulfur Compounds According

to ASTM D5623 in Gasoline with Agilent's


Dual Plasma Sulfur Chemiluminescence
Detector (G6603A) and an Agilent 7890A
Gas Chromatograph
Application

Hydrocarbon Processing

Authors Frequently, petroleum feeds and products contain


varying amounts and types of sulfur compounds.
Wenmin Liu Many sulfur compounds can be corrosive to equip-
Agilent Technologies Co. Ltd. ment, can inhibit or destroy catalysts employed in
412 Ying Lun Road downstream processing, and can impart undesir-
Waigaoqiao Free Trade Zone able odors to products. The ability to speciate
Shanghai, 200131 sulfur compounds in various petroleum liquids is
China useful in controlling sulfur compounds in finished
products and is frequently more important than
Mario Morales
the ability to simply measure total sulfur content
Agilent Technologies, Inc. alone.
2850 Centerville Road
Wilmington, DE 19808 The SCD burner easily mounts on the 6890 and
USA 7890A GCs, and incorporates features for easier
and less frequent maintenance. The DP technology
harnesses the power of dual flame plasma combus-
Abstract tion, optimizing combustion of the sample matrix
and formation of sulfur monoxide (SO).
Sulfur components in gasoline samples are detected uti-
lizing ASTM D 5623 with an Agilent 7890A GC configured The 355 DP SCD response is inherently linear,
with Agilent 355 dual plasma sulfur chemiluminescence equimolar, and far less susceptible to hydrocarbon
detector (DP SCD). The 355 SCD response is linear, interference. These advantages eliminate the need
equimolar, and provides a linear range from 0.1 to 10 ppm. for linearizing data or determining separate
The coelution of hydrocarbon and sulfur peaks does not response factors for individual sulfur compounds.
present a problem for the detector. The detection limits of Furthermore, hydrocarbons are virtually invisible
sulfur compounds in gasoline are down to 20 ppb, while to the DP SCD. The coelution of hydrocarbon and
no quenching was found in the gasoline sample analysis. sulfur peaks does not show quenching. Frequent
column changes are required for analysis of vari-
Introduction ous hydrocarbon products by flame photometric
detectors (FPD) to avoid serious quenching and
Gas chromatography with sulfur chemilumines- inaccurate results. ASTM Method D5623 utilizes
cence detection (SCD) provides a rapid and highly the sulfur chemiluminescence detector (SCD) for
specific means to identify and quantify various the detection of sulfur compounds in gasoline.
sulfur compounds that may be present in petro-
leum feed stocks and products, such as gasoline.
Experimental Experimental Conditions

An Agilent 7890A GC configured with split/split- 7890A GC Conditions


less inlet (Sulfinert-treated), 7683B autosampler, Front inlet Split/splitless (Sulfinert-treated
capillary inlet system)
and Agilent 355 DP SCD were used. The sulfur
standards in toluene and iso-octane (10:90) were Heater 275 C
purchased from Supelco (Bellefonte, PA). The com- Pressure 10.951 psi
ponent information is in Table 1.
Septum purge flow 3 mL/min
Mode Split
Table 1. Sulfur Standards Components
Gas saver 20 mL/min after 2 min
Concentration Split ratio 10 :1
Components Formula (ppm) (w/w)
Split flow 15 mL/min
1 Ethyl mercaptan CH3CH3SH 11.62
Oven 30 C (1 min) 10 C/min 250 C (1 min)
2 Dimethyl sulfide (CH3)2S 11.92
Column HP-1 30 m 0.32 mm 4 m
3 Carbon disulfide C2S 17.84
(P/N 19091Z-613)
4 Isopropyl mercaptan (CH3)2CHSH 34.32
SCD Conditions Agilent G6603A
5 T-butyl mercaptan (CH3)3CSH 11.28
Burner temperature 800 C
6 n-propyl mercaptan CH3CH2CH2SH 5.93
Vacuum of burner 372 torr
7 Methylethyl sulfide CH3CH2SCH3 11.87
Vacuum of reaction cell 5 torr
8 Thiophene C4H4S 14.81
H2 40 mL/min
9 Sec-butyl mercaptan CH3CH2CH(SH)CH3 23.26
10 n-butyl mercaptan CH3(CH2)3SH 5.89 Air 53 mL/min

11 Dimethyl disulfide CH3SSCH3 14.75


12 2-methyl thiophene C5H6S 14.29
Results and Discussion
13 3-metyl thiophene C5H6S 21.35 Several commercially available sulfur detectors are
14 Diethyl disulfide (C2H5)2S2 27.99 available for the determination of sulfur com-
15 Benzo(b)thiophene C8H6S 40.49 pounds in various matrices. When compared to
flame photometric detection (FPD), pulsed flame
NIST Standard Reference Material 2299, Sulfur in photometric detection (PFPD), atomic emission
Gasoline, was used. The total sulfur in gasoline is detection (AED), and inductively coupled plasma-
13.6 1.5 g/g based on analyses by isotope dilu- mass spectrometry (ICP-MS), the SCD shows the
tion thermal ionization mass spectrometry (ID- best performance based on stability, cost, and
TIMS). Homogeneity testing was performed using quantification. [2,3]
X-ray fluorescence spectrometry. With DP technology, the performance of Agilent
355 SCD has been further enhanced, for unsur-
passed stability, selectivity, and lack of quenching.
Table 2 lists the DP SCD stability at different sulfur
levels.

2
Table 2. Repeatability of Sulfur Compounds at Different Concentrations (Refer to Table 1 for peak identification.)
1 2 3 4 5 6 7 8 10 11 12 13 14 15
Con. ppm 1.16 1.79 1.78 3.42 1.13 0.59 1.19 3.80 0.59 1.47 1.43 2.14 2.80 4.0
RSD (%) 2.8 3.6 3.1 1.9 3.0 2.7 3.9 3.9 2.9 2.1 2.2 2.9 0.4 3.7
Con. ppm 0.12 0.18 0.18 0.34 0.11 0.06 0.12 0.38 0.06 0.15 0.14 0.21 0.28 0.4
RSD (%) 5.7 7.4 3.4 3.7 6.6 4.8 5.7 4.8 8.0 4.0 3.3 4.7 7.3 3.1

Correlation coefficients of the tested sulfurs over compound. Figure 1 shows the chromatogram of
three orders of magnitude were better than 0.99 sulfurs in a hydrocarbon matrix without interfer-
(R2). Table 3 shows the linearity of each sulfur ence (1 to approximately 2 g/kg).

Table 3. Linear Ranges of Tested Sulfurs (Refer to Table 1 for 34


8/9 11
14 15
13
peak identification.) 3500

Linearity (R2)
3000
Analytes Concentration Range 12
2500 7 12

1 0.1ppm to 10ppm 0.9975 2000


5
6
2 0.1ppm to 10ppm 0.9982 1500

10
1000

3 0.1ppm to 10ppm 0.9991 500

4 0.1ppm to 10ppm 0.9992 0


2.5 5 7.5 10 12.5 15 17.5 20 m in

5 0.1ppm to 10ppm 0.9995


Figure 1. Chromatogram of sulfur standard in hydrocarbon
6 0.1ppm to 10ppm 0.9996 matrix. (Refer to Table 1 for peak identification.)
7 0.1ppm to 10ppm 0.9998
8/9 0.1ppm to 10ppm 0.9998 The data in Table 4 illustrate the sensitivity (S/N)
10 0.1ppm to 10ppm 0.9994 of 355 SCD for trace-level analysis (approximately
20 ng/kg) of sulfurs in a hydrocarbon matrix.
11 0.1ppm to 10ppm 0.9999 Figure 2 shows the chromatogram of trace sulfurs,
12 0.1ppm to 10ppm 0.9997 which also indicates no interferences in the analy-
13 0.1ppm to 10ppm 0.9995 sis.
14 0.1ppm to 10ppm 0.9997
15 0.1ppm to 10ppm 0.9999

Table 4. Sulfur Sensitivity (Refer to Table 1 for peak identification.)


Peak No. 1 2 3 4 5 6 7 8/9 10 11 12 13 14 15
S/N 2.0 2.5 5.0 4.6 1.8 1.6 2.4 5.0 1.5 3.6 2.0 4.6 3.2 5.2

15V

1000

800

600 15
400
3 4 8/9 11 1213 14
1 2 5 67 10
200

Figure 2. Chromatogram of sulfurs at trace levels (20 ng/g).


(Refer to Table 1 for peak identification.)

3
www.agilent.com/chem
Gasoline Samples Analysis Conclusions
NIST Standard Reference Material 2299 was The Agilent DP SCD is used for the detection of
detected on an Agilent SCD and the mass concen- sulfur compounds in a complex hydrocarbon
tration of total sulfur in sample was calculated by matrix. The results show that the DP SCD has
summing the sulfur content of all sulfur compo- linear response to sulfur compounds without
nents (known and unknown) in the sample to quenching, yielding MDLs down to 20 ng/g. This
arrive at its total sulfur value as recommended by solution is available as an Agilent preconfigured
ASTM 5623. system; please refer to Agilent SP1 7890-0365 for
ordering information.
Figure 3A shows the chromatogram of the sulfur
standard and Figure 3B shows the chromatogram
of the standard gasoline sample. The total amount
References
of sulfur is calculated by summing all the peak 1. ASTM 5623: Standard test method for sulfur
areas in Figure 3B and quantitated as thiophene compounds in light petroleum liquids by gas
11.8 g/g of total sulfur with an RSD 2.7% (n = 5) chromatography and sulfur selective detection
was calculated in the gasoline sample. This was
2. Roger L. Firor and Bruce D. Quimby, Analysis
within the specified range of 13.6 1.5 g/g.
of Trace Sulfur Compounds in Beverage-Grade
Carbon Dioxide, Agilent Technologies publica-
15V 14
tion 5988-2464EN
1000 4
800
3 8/9
11
15
3. Roger L. Firor and Bruce D. Quimby, A Com-
13
600

400
12 7 12
parison of Sulfur Selective Detectors for Low-
56
200
10
Level Analysis in Gaseous Streams, Agilent
0
0 2 4 6 8 10 12 14 16 18 min Technologies publication 5988-2426EN

Figure 3A. Chromatogram of sulfur standard.


For More Information
For more information on our products and services,
8/9

15V
visit our Web site at www.agilent.com/chem.
1000

800
2 11

600
3 7 12 14
400
6 13
200 4 5
0
0 2 4 6 8 10 12 14 16 18 min

Figure 3B. Chromatogram of gasoline standard sample. (Refer


to Table 1 for peak identification.)

Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this
material.

Information, descriptions, and specifications in this publication are subject to change


without notice.

Agilent Technologies, Inc. 2008

Printed in the USA


August 12, 2008
5989-9233EN

Potrebbero piacerti anche