Sei sulla pagina 1di 2

SPECIFICATIONS INFUSION PUMP TE-172

Product Name TERUFUSION INFUSION PUMP TE-172


Model TE 172
Pumping System MIDPRESS peristaltic finger system ( A newly system
to provide accurate infusion without pressing the
tubing completely )
Delivery Rate 0.1 1200 mL/h ( 0.1 99.9 mL/h: 0.1 mL/h step,
100 1200 mL/h: 1 mL/h step ). The delivery rate can
be set in 1 mL/h step throughout the range by the
internal mode select switch
Accuracy Within 5 % infusion set: Specified TERUFUSION
infusion sets for pump ( TSPA )
Solution: water/saline
Test time: second hour
Delivery rate: 1.0 mL/h or more
Temperature: 250 C
Volume Delivered range 0.0 to 9,999 mL
Delivery Limit range 1 9,999 mL ( 1 mL step ) or
- - - - ( Delivery Limit Free )
Purge flow function Higher than 500 mL/hours
Alarms Air, Occlusion, Flow Error, Door, Low Battery

Operation indicator flashes red and the alarm message


come on. Pump stops except during the LOW BAT
alarm
Occlusion Detection Pressure 13.3 133.3 kPa ( 100 1.000 mmHg / 0.1 1,4
kgf/cm2 )
The detection pressure can be adjusted in 3 level ( L,
M, H ) within the above range
Infusion Complete Function When the total volume delivered reaches the preset
( Completion ) delivery limit, light and buzzer sound notifies the
operator
Nurse Call Alarm Output an alarm relay contact ( Can be connected to
nurse call )
Contact rating: DC 12 Volt, 1.0 Ampere
Computer Interface RS-232C/RS-485
Special Functions KEEP VEIN OPEN ( KVO ) function
KVO function continues infusion at flow rate of
1 mL/h after delivery limit ( preset volume )
has been reached.
Tubing clamp function
When the door is opened, tubing is
automatically clamped
Volume delivered clear function
Holding down the [ mL CLEAR ] switch for
about 0.5 second clears the volume delivered
reading
Time remaining display
Battery capacity indicator ( in 3 levels )
Adjustable display brightness ( can be switched
to 2 level )
Adjustable buzzer volume ( in 3 level )
Operation history function ( Storage / display
of operation history )
Standby function
Repeat alarm function
Start reminder alarm
Operating conditions + 10 to +400 C, Humidity : 30 to 85 %
Storage Conditions Temperature : - 20 to +450 Celcius
Humidity : 10 95 % RH ( no condensation )
Power Supply AC : 100 240 Volt, 50/60 Hz
DC : 12 15 VDC
Internal Nickel Cadmium Battery can run the pump
continously for about 3 hours.
Condition : New battery, Delivery rate : 25 mL/h,
Temperature: 250C, Battery charging 15 hours for the
first use
Power Consumption 28VA ( Watt )/AC 100 240 Volt
11 VA ( Watt )/ DC power
Classification Class I and Internal Power Supply
Type CF, IPX1 Continuous Operation
Dimensions 100 mm (W) x 215 mm ( H) x 208 mm ( D )
Weight Approx 3.7 kg
Body Weight Mode Dose Rate : 0.01 99.9 g/kg/min
( Dose Rate Calculation ) ( 0.01g/kg/min step )
Body Weight : 0.1 300.0 kg ( 0.1 kg step )
Drug mass : 0.1 999.9 mg ( 0.1 mg step )
Solution Volume : 0.1 999.9 mL ( 0.1 mL
step )

Memory Function The latest five patterns of the above four parameter
are automatically saved and can be displayed