AURO PEPTIDES LIMITED.

Annexure Title: Certificate of Calibration


Department: Engineering Page No.: 01 of 01

Calibration certificate No Date of issue


Calibration Done Date 07/01/2017 Calibration Due Date 05/04/2017

Details of UUC:
Instrument Name: Pressure Gauge Make: SMC
Instrument ID: PRP005/PG/01 Least Count: 0.05MPA
Equipment Name: Preparative HPLC System Acceptance criteria : 1 % FS
Equipment Code: PRP005 Physical condition: Active
Range: 0-1 Mpa

Details of standard instruments Used:


Valid up Report no: /
S.No Name of the instrument Sr.No Range
to Certificate no:
01 High pressure calibrator MPC1511P0415 0-700 Bar 02/11/2017 1611020400109
02 Mini data logger 36915499 -20 to 70 C 23/08/2017 1608200700809

Applied Procedure : EN/067-00 ,EN/038-03,EN/060-00 & EN/065-00.

Environmental Details Temperature : 24.2 C Humidity : 56.4 %RH

Calibration Test Result


Standard Value in Mpa
Mean value UUC measured
S.No Ascending Descending Error ()
In Mpa value in Mpa
order order
01 0.000 0.000 0.000 0.0 0.000
02 0.200 0.202 0.201 0.2 0.001
03 0.401 0.403 0.402 0.4 0.002
04 0.603 0.604 0.603 0.6 0.003
05 0.804 0.806 0.805 0.8 0.005
06 1.007 1.008 1.007 1.0 0.007

Remark: Put the tick mark which is applicable

Instrument calibrated and found accuracy within acceptance criteria.

Instrument calibrated and found accuracy out of acceptance criteria.

Calibrated By Checked By Approved By

Annexure-1 of SOP EN/054-00


AURO PEPTIDES LIMITED.
Annexure Title: Certificate of Calibration
Department: Engineering Page No.: 01 of 01

(QA)

Annexure-1 of SOP EN/054-00