Sei sulla pagina 1di 40

Operating Instruction OI/S26-EN

Model S26 seals


for remote and direct mount

2600T Series Pressure Transmitters


Engineered solutions for all
applications
Aboutthisbook
ThisbookprovidesinstructionsforinstallingtheS26diaphragmsealsequippedtransmitters.Instructionscoveringallaspectsof
the transmitters, which are not related to installation of the remote seals, are included in the transmitter instruction books.
Pleasereadthismanualbeforeworkingwiththeproduct.
IncaseofadditionalinformationneedspleasecontactABBattheaddressyoufindatlastpagesofthismanualorattheweb
site:http://www.abb.com/pressure.
TheproductsdescribedinthepresentmanualarenotintendedforuseinNUCLEARQUALIFIEDenvironments/applications.

TheCompany
Weareanestablishedworldforceinthedesignandmanufactureofinstrumentation
for industrial process control, flow measurement, gas and liquid analysis and
environmentalapplications.
As a part of ABB, a world leader in process automation technology, we offer
customersapplicationexpertise,serviceandsupportworldwide.
We are committed to teamwork, high quality manufacturing, advanced technology
andunrivalledserviceandsupport.
Thequality,accuracyandperformanceoftheCompanysproductsresultfromover
100yearsexperience,combinedwithacontinuousprogramofinnovativedesignand
developmenttoincorporatethelatesttechnology.
TheNAMASCalibrationLaboratoryNo.0255(B)isjustoneofthetenflowcalibration
plantsoperatedbytheCompany,andisindicativeofourdedicationtoqualityand
accuracy.

ManualVersion
ThisInstructionmanualversionis01anditwasreleasedonFebruary26th2010.

2
WorlwideManufacturingLocations

ABBSpA ABBInc.
ViaStatale113 3450HarvesterRoad
22016Lenno(Co)Italy Burlington,OntarioL7N3W5Canada
Tel:+39034458111 Tel:+19056810565
Fax:+39034456278 Fax:+19056812810

ABBAutomationProductGmbH ABBLtd.
Shillerstrasse72 14MathuraRoad
D32425MindenGermany 121003Faridabad,HaryanaIndia
Tel:+49551905534 Tel:+911292275592
Fax:+49551905555 Fax:+911292279692

ABBInc. ABBEngineering(Shanghai)Ltd.
125E.CountyLineRoad No.5,Lane369,ChuangyeRoad
Warminster,PA18974USA KangqiaoTown,NanhuiDistrict
Tel:+12156746000 Shanghai,201319,P.R.China
Fax:+12156747183 Phone:+86(0)2161056666
Fax:+86(0)2161056677

Copyright2009byABBSpA
Subjecttochangeswithoutnotice
Thisdocumentisprotectedbycopyright.Itassiststheuserinsafeandefficientoperationofthedevice.
Thecontentsofthisdocument,whetherwholeorinpart,maynotbecopiedorreproducedwithoutpriorapprovalbythe
copyrightholder.

3
TableofContents

INTRODUCTION ..............................................................................................................................................................5
Productidentification ......................................................................................................................................................5
INSTALLATION................................................................................................................................................................7
General ............................................................................................................................................................................7
FLOWMEASUREMENT ...................................................................................................................................................7
HorizontalInstallation......................................................................................................................................................7
Verticalinstallation ..........................................................................................................................................................7
LIQUIDLEVELMEASUREMENT ..................................................................................................................................... 8
MeasurementReference ................................................................................................................................................. 8
OpenTankInstallation .................................................................................................................................................... 8
ClosedTankInstallationPressureService ..................................................................................................................... 9
DoubleSeal.............................................................................................................................................................. 9
SingleSeal ............................................................................................................................................................... 9
SingleSeal ..............................................................................................................................................................10
ClosedTankInstallationVacuumService.....................................................................................................................10
ClosedTankInstallationVacuumService..................................................................................................................... 11
CalibrationRequirements............................................................................................................................................... 11
CalibrationRequirements...............................................................................................................................................12
INTERFACELEVELMEASUREMENT .............................................................................................................................12
DENSITYMEASUREMENT.............................................................................................................................................12
GAUGEPRESSUREMEASUREMENT............................................................................................................................. 13
PressureService............................................................................................................................................................. 13
VacuumService ............................................................................................................................................................. 13
ABSOLUTEPRESSUREMEASUREMENT....................................................................................................................... 13
MOUNTINGTHETRANSMITTER ...................................................................................................................................14
CONNECTINGTHEREMOTESEALS..............................................................................................................................14
WaferRemoteSeal(S26WAandS26WE).......................................................................................................................14
ChemicalTeeRemoteSeal(S26CN)...............................................................................................................................16
Rotatingflangediaphragmsealsflushandextended(S26RA,S26REandS26RJ) ........................................................ 17
FlangedRemoteSealRingJoint(S26RR).....................................................................................................................19
Fixedflangediaphragmsealsflush(S26FA,S26FE) .....................................................................................................20
OfflineThreadedConnectionSeal(S26TT) ...................................................................................................................22
OffLineFlangedConnectionRemoteSeal(S26MA,S26ME) .........................................................................................23
RemoteSealforUreaService(S26PN) ...........................................................................................................................24
ButtonTypeRemoteSeal(S26BN) ................................................................................................................................25
UnionConnectionRemoteSeals(S26UN)...................................................................................................................... 27
SealwithoutWeldBushing ........................................................................................................................................27
SealwithWeldBushing .............................................................................................................................................27
SealwithChemicalTeeFlange ..................................................................................................................................28
FoodandSanitaryRemoteSeal(S26SS) ....................................................................................................................... 29
CherryBurrellAseptic............................................................................................................................................... 29
SanitarySealwithFlushDiaphragm ......................................................................................................................... 29
SanitarySealwithFlushDiaphragm ..........................................................................................................................30
SanitarySealwithExtendedDiaphragm....................................................................................................................30
Saddleandsocketdiaphragmseal(S26VN) ................................................................................................................... 33
Inlinediaphragmseal(S26JN).......................................................................................................................................35
PulpandPaperdiaphragmseals(S26KN).......................................................................................................................36
1in.and11/2in.sizewith1in.and11/2in.NPTthreadedconnection......................................................................36
1in.and11/2in.sizewithG1in.BandG11/2in.Bthreadedconnection .................................................................36
1in.and11/2in.sizetospudwith1or2screw(s)fixing.............................................................................................36
1in.and11/2in.sizetospudwith1or2screw(s)fixing............................................................................................. 37
1in.Gwithballvalveconnection ............................................................................................................................... 37
11/2in.sealingconnectiontothreadedspud ............................................................................................................ 37
WeldingProcedureforSanitaryandPulp&Paperspuds .............................................................................................38
SanitarySpuds(3Arecommendations) ......................................................................................................................38
CONNECTINGLOWSIDECOMPENSATIONLEGONADPWITHONESEAL ................................................................39
OpenTankInstillation ....................................................................................................................................................39
ClosedTankInstallation .................................................................................................................................................39

4
INTRODUCTION
TheS26SeriesSealSystemisaprotectivedeviceusedtoisolatetransmittersfromtheprocessfluid.
Whenusedwithoutdiaphragmseals,thesetransmittersaregenerallyconnectedtoprocesspipingorprocessvesselbyimpulse
lines. The process fluid leaves its enclosure, fills the impulse lines, and enters the body of the transmitter. For certain
applications, it is necessary to prevent the process fluid from leaving its enclosure. The seal system provides a flexible
diaphragm seal between the process fluid and a liquid filled capillary tube connected to the body of the transmitter. The
diaphragmisolatestheprocessfluidwhilethefilledcapillarytubehydraulicallytransmitstheprocesspressuretothetransmitter
body.
Somereasonswhyanisolatingsealisrequiredare:
The process fluid is highly corrosive. Remote seals offer a much wider choice of corrosion resistant materials
comparedtoconventionaltransmitters.
Theprocessfluidhassolidsinsuspensionorishighlyviscousandcanclogimpulselines.
Theprocessfluidcansolidifyinimpulselinesorinsidethetransmitterbody.
Theprocessfluidistoohazardoustoenterthecontrolareawherethetransmitterislocated.
Theprocesstemperatureexceedstherecommendedmaximumorminimumlimitsforthetransmitter.
Theapplicationisinterfacelevelordensitymeasurement.Remotesealsoffertherequiredconstantandequalspecific
gravityofthepressuretransferfluidonthehighandlowsidesofthetransmitter.
Thetransmitterbodymustbelocatedawayfromtheprocessforeasiermaintenance.
TheS26seriesisavailableinavarietyofsealelementtypes.TheyprovideprocessconnectiontoANSI,JISorENpipeflanges,
wedgeflowelements,chemicaltees,andthreadedpipefittings.Extendeddiaphragmremotesealsaresuitableforconnection
to2inch,3inchor4inchflangedtanknozzlesorflangedteespermitthesealdiaphragmtobelocatedflushwiththeinsideofa
tankorpipe.Sanitaryandsanitaryaseptictypesealsmeetthestringentrequirementsofsanitaryfood,dairy,pharmaceutical,
and Biotech applications. All sanitary seals comply with 3A Sanitary Standard No. 7402. These remote seal systems feature
unique all welded construction (see in Figures 1 and 2). This construction assures reliable operation over a wide range of
operatingtemperaturesandpressures.Itisespeciallyimportantinpreventingairleakageswhenthesystemisoperatingunder
vacuumconditions.

Productidentification
TheS26SeriesSealSystemsareavailablewithavarietyofremotesealconstructionsidentifiedasfollows:
S26FAandS26FEFixedflangesealsaccordingtoASMEandENstandards
S26RA,S26REandS26RJRotatingflangesealsaccordingtoASME,ENandJISstandards
S26MAandS26MEOfflineflangedsealsaccordingtoASMEandENstandards
S26WAandS26WEWaferremotesealsaccordingtoASMEandENstandards
S26CNChemicalteeremoteseal
S26JNInlineseal
S26KNPulp&Paperconnectionseal
S26RRFlangedremotesealRingJointconstruction
S26SSFoodandsanitaryseal
S26TTOffLinethreadedconnectionseal
S26UNUnionconnectionremoteseal
S26PNRemotesealforUreaservice
S26BNButtontyperemoteseal
S26VNSaddleandsocketseal

TheS26SeriesSealSystemsareusedonthefollowingtransmitters:
266xR/266xD(wherexmeansanyletter)

Differentialpressuretransmittershaveoneortworemoteseals.Singlesideddifferentialpressuretransmittershaveasingle
remotesealfortanklevelapplications.Gaugeandabsolutepressuretransmittershaveasingleremoteseal.Thetransmitterand
sealsystemareidentifiedbyproductcodenumberswhicharestampedonthetransmitternameplate.Anexampleofthe
productcodestampingisasfollows:

266DRHGSRRAH / S26WAHD5FSM2AASNNN
TransmitterProductCode SealSystemProductCode

Thecharacterswhichcomprisethesecodenumbersidentifyspecificproductfeatures.Adetailedexplanationofthecode
numberstogetherwithphysicalandperformancespecificationsforboththetransmitterandsealsystemisprovidedinthe
productspecificationsheets.

6
INSTALLATION
Locatingthetransmitterandremoteseal
This section provides guidelines for determining where to locate the transmitter and seals for specific applications. Proper
locationisanimportantfactorinobtainingaccuratemeasurements.

General
The transmitters with remote seals are suitable for use on the following pressure and differential pressure measurement
applications:
FLOWMEASUREMENT
LIQUIDLEVELMEASUREMENT
INTERFACELEVELMEASUREMENT
DENSITYMEASUREMENT
GAUGEPRESSUREMEASUREMENT
ABSOLUTEPRESSUREMEASUREMENT
Eachapplicationhascertainuniquerequirementswhichinfluencetheselectionofthetransmitterandseallocations.Ingeneral
theflowandgaugepressureapplicationsallowthegreatestflexibilityinselectionofalocation.Liquidlevelapplicationsimpose
restrictionsontherelativeelevationsofthetransmitterandsealswhenthetankoperatesundervacuumandabsolutepressure
applications impose similar restrictions when the lower range value is less than atmospheric pressure (nominally 14.7 psia).
Transmitterswith remote seals measure pressure from two sources; (1) the applied process pressure or differential pressure,
and(2)theheadpressuredevelopedbytheweightoftheliquidcolumninthesealsystem.Theheadpressuresareafunctionof
transmitterandseallocationsasfollows:
Two Seals On a DP transmitter with a seal on both the high and low sides, the transmitter measures the head pressure
resulting from any difference in elevation between the seals and the output change is directly proportional to the elevation
difference. The head pressure has no effect on transmitter output when both seals are at the same elevation. Changing the
elevationofthesealsrelativetothetransmitterhasnoeffectaslongastheelevationofonesealrelativetotheotherisnot
changed.
OneSealOnatransmitterwithoneseal,theheadpressurefromthesealsystemismeasureddirectly.Anychangeinthe
elevationofthesealcausesachangeinthemeasuredheadpressureandthetransmitteroutputchangesindirectproportionto
theelevationchange.
Note:whenhandlingand/orinstallingthetransmitterwithremoteseal(s),pleaseconsiderthattheminimumbendingradiusof
thecapillarysystemis10cm(3.9in).
For most flow, gauge pressure, and absolute pressure installations the effect of head pressure is relatively minor and can be
canceled out by adjusting zero with the transmitter and seals in their installation position while a known process pressure is
applied. For liquid level installations, head pressures have a more significant impacton calibrationrequirementsbecause the
natureoftheinstallationrequiresasubstantialdifferenceinelevationbetweenthehighandlowsideseals.SeetheLiquidLevel
MeasurementSectionformoreinformation.

FLOWMEASUREMENT
Differential pressure transmitters with two remote seals can be used for both
horizontalandverticalflowmeasurementinstallations.BymeanofaWedgeFlow
Element(seeFigure3).Inbothhorizontalandverticalinstallations,thetransmitter
canbeplacedinanyconvenientlocationeitheraboveorbelowtheelevationofthe
sealelements.Thehighsidesealelementmustbelocatedontheupstreamsideof
theflowrestriction.Toensureaccuratemeasurementsonliquidflowapplications,
theflowrestrictionmustbelocatedinasectionofthepipelinewhichremainsfull
underallflowconditions.

HorizontalInstallation
Inahorizontalinstallationbothsealelementsareatthesameelevationandhead
pressuresresultingfromthefillingliquidareequalonthehighandlowsideofthe
transmitter.Thus,thetransmittercalibrationcanbezerobasedforthisinstallation.
Iftheprocesscontainssuspendedsolids,thesealelementsshouldbelocatedatthe
topofthepipetoavoidcollectionofsolidmaterialonthesealdiaphragmsurfaces.

Verticalinstallation
In a vertical installation, the elevation difference between the two seals applies a
differential pressure when there is no flow. When the pipe is full at no flow the
initialdifferentialisafunctionofthedistancebetweenthesealsandthedifference
indensitybetweentheprocessliquidandthefillliquid.
7
LIQUIDLEVELMEASUREMENT
Adifferentialpressuretransmitterwitheitheroneortworemotesealscanbeusedforliquidlevelmeasurementonopentanks,
closedtanksoperatingatpressuresaboveatmosphericandclosedtanksoperatingundervacuum.

MeasurementReference
Liquid level measurements are based on the height of a process liquid column with respect to predefined measurement
referencepointsonthetransmitter.Thetransmitterhasthreedatumlinesasfollows:
PrimaryDatumLineLocatedatthecenterofthesealsystemconnectiontothetransmitterbody.
HighSealDatumLineLocatedatthecenterofthehighsidesealdiaphragm.
LowSealDatumlineLocatedatthecenterofthelowsidesealdiaphragm.

OpenTankInstallation
Atransmitterwithoneremotesealcanbeusedforopentanklevelmeasurement(seeFigure1).
Thesealisonthehighsideofthetransmitter,andthesealelementmustbelocatednearthebottomofthetanksothatthe
highsealdatumlineisatorbelowtheminimumlevel.Thetransmittercanbelocatedeitheraboveorbelowthesealelement.
Thelowsideofthetransmittermustbeventedtoatmosphere.

Fig.1Levelmeasurementofaliquidinanopentank












8
ClosedTankInstallationPressureService
DoubleSeal
When a transmitter with two remote seals is used to measure level in a pressurized tank (see Figure 2a and 2b), the high
pressuresidesealelementmustbelocatednearthebottomofthetank.
Thehighsealdatumlinemustbeatorbelowtheminimumlevel.Thelowsidesealmustbelocatednearthetopofthetank,and
thelowsealdatumlinemustbeatorabovethemaximumlevel.
Thetransmittercanbelocatedbetweentheseals,abovebothsealsorbelowbothsealsonapressureserviceinstallation.
The preferred location is mid way between the seal elements (see Figure 2b). This location minimizes the required capillary
length,andusuallyprovidesthemostuniformdistributionofambienttemperaturesacrossthecapillarylength.

Fig.2a
Level measurement of liquid
in a closed tank, two seals
installation and transmitter
belowtheminimumlevel

Fig.2b
Level measurement of
liquid in a closed tank.
Recommended
installation.

9
SingleSeal
When a transmitter with one seal is used to measure level in a pressurized tank, a compensating leg must be connected
betweenthevaporspaceatthetopofthetankandthelowsideofthetransmitter(seeFigures3a,3b,4a,4band5).
Theremotesealisonthehighsideofthetransmitter,andthesealelementmustbelocatednearthebottomofthetanksothat
thehighsealdatumlineisatorbelowtheminimumlevel.
Iftheprocessvaporisnotreadilycondensable,orifthecompensatinglegisatahighertemperaturethanthetankinterior,adry
leg can be used (see Figure 3a and 3b). A trap installed at the bottom of the leg minimizes the possibility of condensate
collectinginthetransmitterbody.
When the process vapor is condensable, a wet leg is recommended. The wet leg is filled with a suitable liquid to maintain a
constantpressureonthelowsideofthetransmitter(seeFigures4aand4b).Thisapproachavoidstheproblemofprocessvapor
condensingandcollectinginthecompensatinglegresultinginseriousmeasurementerror.Thelegcanbefilledwithprocess
liquidoranyliquidwithalowvaporpressurerelativetotheprocess.Afillingteeisrequiredatthetopoftheleg.Thewetlegcan
beavoidedifthetransmittercanbelocatednearthetopofthetank(seeFigure5),sothatcondensatedrainsbackintothetank. tank.

Fig.3a
Level measurement of
liquid in a closed tank.
Transmitter installed
above the minimum
level with one seal and
dryleg.

Fig.3b
Level measurement of
liquid in a closed tank.
Transmitter installed
below the minimum
level with one seal and
dryleg.

10
ClosedTankInstallationVacuumService
When the transmitter is used on a tank in which the operating pressure range includes pressures below atmospheric, it is
essential that the transmitter be located below the highpressure seal datum line (see Figures 2a and 2b for two seal
installations, Figures 3a, 3b, 4a and 4b for one seal installation). The recommended minimum distance between the primary
datumlineandthehighsealdatumlineis1ft(0.3m).Sealelementlocationrequirementsforvacuumservicearethesameasfor
pressureservice.

Fig.4a
Levelmeasurementofliquidinaclosed
tank. Transmitter installed above the
minimum level with one seal and wet
leg.

Fig.4b
Level measurement of liquid in a closed tank.
Transmitter installed below the minimum level
withonesealandwetleg.

Fig.5
Level measurement of liquid in a
closed tank. Transmitter installed on
the topof the tank, oneseal and dry
leginstallation.

11
CalibrationRequirements
The location of the high side seal near thebottom of the tank and the low seal nearthe top of the tank provides increasing
transmitteroutputforincreasingliquidlevel.Theactualdifferentialpressureappliedtothetransmitterisuptomaximumvalue
onlywhentheprocessisatitsminimumlevelduetotheelevationdifferencebetweentheseals.Asthetankfills,theprocess
liquidtendstocanceltheinitialsealsystemheadpressureandthemeasureddifferentialpressureisthereforeattheminimum
valuewhenthelevelreachesitsmaximumvalue.
Notealsothatthemaximumdifferentialisappliedfromthelowsidebecausethelowsealhasthehigherelevation.
Toaccommodatetheheadpressureconditionsimposedbyaliquidlevelinstallation,thetransmitterzeromustbeadjustedto
elevatetheoutput.Thiscancelstheeffectofthelowsideheadpressureandprovideszeropercentoutputfortheminimumlevel
condition.Forexample,therangevaluesforalevelinstallationmightbe120to5inchesofwater.Thisindicatesthatthelevel
change(span)is115inchesofwater;thedifferentialatminimumlevelis120andatmaximumlevelis5inchesofwater.
Theminussignindicatespositivedifferentialpressuresappliedtothelowsideofthetransmitter.
Determination of the span and range values for a specific installation must take into account the level change, elevation
difference between the seals, height of the process liquid above the high seal datum line at minimum level, and the specific
gravityoftheprocessliquidandthefillliquidinthesealsystem.

INTERFACELEVELMEASUREMENT
Adifferentialpressuretransmitterwithtworemotesealscanbeusedtomeasureinterfacelevel(seeFigure6).Thehighside
sealelementmustbelocatednearthebottomofthetank.Thehighsealdatumlinemustbeatorbelowthelowestinterface
elevation. The low side seal must be located near the top of the tank, and the low seal datum line must be at or above the
highest interface elevation. The minimum level of liquid in the tank must be above the low seal element under all interface
conditions.Thetransmittercanbelocatedbetweentheseals,abovebothsealsorbelowbothsealswhenthetankisopenor
operating under positive pressure. The preferred location for 266DLH the transmitter is directly mounted on the bottom
connection (see Figure 7). This location enables the user to eliminate the capillary and consequently allows a more accurate
measurement.Iftheoperatingpressurerangeincludespressuresbelowatmospheric,thetransmittermustbelocatedbelowthe
highsideseal(refertothevacuumserviceinformationinLiquidLevelMeasurementSection).

Fig.6Interfacelevelmeasurement

DENSITYMEASUREMENT
Adifferentialpressuretransmitterwithtworemotesealscanbeusedtomeasureliquiddensityorspecificgravity(seeFigure7).
Thehighsidesealelementmustbelocatednearthebottomofthetank,andthelowsidesealmustbelocatednearthetopof
the tank. The minimum level of liquid in the tank must be above the low seal element under all density conditions. The
transmittercanbelocatedbetweentheseals,abovebothsealsorbelowbothsealswhenthetankisopenoroperatingunder
positivepressure.Thepreferredlocationismidwaybetweenthesealelements.Thislocationminimizestherequiredcapillary
length,andusuallyprovidesthemostuniformdistributionofambienttemperaturesacrossthecapillarylength.Iftheoperating
pressure range includes pressures below atmospheric, the transmitter must be located below the high side seal (refer to the
vacuumserviceinformationinLiquidLevelMeasurementSection).

12

Fig.7Densityorspecificgravitymeasurement

GAUGEPRESSUREMEASUREMENT
Agaugepressuretransmitterwitharemotesealcanbeusedfor
measurement of pressure in either a process pipe or tank (see
Figure8).Locatetheremotesealinthesideortopofthepipeto
avoidcollectionofsedimentonthesealdiaphragm.

PressureService
Whentheoperatingpressurerangehasaminimumvalueabove
atmospheric pressure (0 psig), the transmitter can be placed in
any convenient location, either at the same elevation as the
remoteseal,aboveorbelowtheseal.

VacuumService
When the operating process pressure range includes pressures
belowatmospheric,itisessentialthatthetransmitterbelocated
below the elevation of the seal element to ensure that the
pressure in the transmitter body is always greater than 0 psia.
The recommended minimum distance between the transmitter
datumlineandthesealdatumlineis0.3m(1ft).

ABSOLUTEPRESSUREMEASUREMENT
Anabsolutepressuretransmitterwitharemotesealcanbeused
formeasurementofabsolutepressureineitheraprocesspipeor
tank(seeFigure8).Locatethesealelementinthesideortopof
the pipe to avoid collectionof sediment on the seal diaphragm.
When the operating absolute pressure range has a minimum
value above atmospheric pressure (nominally 14.7 psia), the
transmitter can be placed in any convenient location, either at
thesameelevationastheremoteseal,abovetheseal,orbelow
the seal. When the pressure range includes pressures below
atmospheric, the transmitter datum line must be below the
elevation of the seal element to ensure that the pressure in the
transmitter body is always greater than 0 psia. The
recommended minimum distance between the transmitter
Fig.8Possibleinstallationsforgaugepressure
datumlineandthesealdatumlineis0.3m(1ft).
measurement
13
MOUNTINGTHETRANSMITTER
The transmitter with remote seals can be mounted on a 60 mm (2 in.) horizontal or vertical pipe using the optional pipe
mounting bracket available with all 2600T series transmitters. Refer to all the documents listed on page 3 "Supplementary
documentation"formountinginformation.

CONNECTINGTHEREMOTESEALS
WaferRemoteSeal(S26WAandS26WE)
The wafer remote seal is designed to be clamped between two ASME (S26WA) or EN raised face flanges (S26WE). The
diaphragmsideofthesealfacestheprocessflangeandabackupflangeisusedontheothersideoftheseal.TheS26WAwafer
sealscanbeusedwith:ANSICL150,CL300,CL600,CL900orCL1500flangeswhereastheS26WEwithENDN1040,DN64160.
FormountingdimensionsforthewaferremotesealpleaseseeFigure12.Therequiredbackupflangecanbesuppliedbythe
user,oritcanbeobtainedasanoptionwiththesealsystem.Bolts,nuts,andaflangegasketmustbesuppliedbytheuser.

Note.IncaseofDPstyletransmitterwithoneremoteseal,pleasereferto"LiquidLevelMeasurement"Section.

Model S26WA (manufactured according to ASME B16.5) is characterized by a maximum working pressure up to 41.37 MPa
(413.7baror6000psi).
ModelS26WE(manufacturedaccordingtoEN10921)canhavedifferentpressurelimitsbecauseoftherelevantforms:
FormB1upto40MPa(400baror5800psi)
FormDupto16MPa(160baror2320psi)
FormEupto10MPa(100baror1450psi)

Connectthesealelementasfollows:
1.Removeprotectivecoveringfromseal.
2.Installagasketbetweentheprocessflangeandthesealelementgasketsurface.
3.Clampthesealbetweentwoflanges(seeFigure12).Besurethatthegasketisproperlypositionedbetweentheprocessflange
andthesealelementgasketsurface.
4.Tightenboltsfirmlyanduniformlyinaccordancewithstandardindustrialflangeboltingpractices.

Fig.9Wafer/Pancakestyleremoteseal

14

ASMEandEN10921FormB1 FormD FormE

Fig.10Wafer/Pancakestyleremotesealdifferentconfigurations

Dimensionsmm(in)forS26W
Size/Rating Diaphragm(dia) Aflushingring B(dia) C(dia) D(dia)
Std.thickness Lowthickness internaldia
1in.ASMEB16.5 47(1.85) 47(1.85) 52(2.05) NA NA 73(2.87)
2in.ASMEB16.5 60(2.36) 58(2.28) 62(2.44) NA NA 92(3.62)
3in.ASMEB16.5 89(3.5) 75(2.95) 92(3.62) NA NA 127(5)
DN40EN10921FormB1 47(1.85) 47(1.85) 52(2.05) NA NA 88(3.46)
DN50EN10921FormB1 60(2.36) 58(2.28) 62(2.44) NA NA 102(4.02)
DN80EN10921FormB1 89(3.5) 75(2.95) 92(3.62) NA NA 138(5.43)
DN40EN10921FormD 47(1.85) 47(1.85) NA 60(2.36) 76(2.99) 88(3.46)
DN50EN10921FormD 60(2.36) 58(2.28) NA 72(2.83) 88(3.46) 102(4.02)
DN80EN10921FormD 89(3.5) 75(2.95) NA 105(4.13) 121(4.76) 138(5.43)
DN40EN10921FormE 47(1.85) 47(1.85) NA 75(2.95) NA 88(3.46)
DN50EN10921FormE 60(2.36) 58(2.28) NA 87(3.42) NA 102(4.02)
DN80EN10921FormE 89(3.5) 75(2.95) NA 120(4.72) NA 138(5.43)

ForanyotherdetailedinformationregardingS26Wxsealssuchasvacuumservicerecommendations,processtemperature
limits,gasketseatfinish,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

15
ChemicalTeeRemoteSeal(S26CN)
Thechemicalteeremotesealisdesignedtoconnecttoa1630LWedgeFlowElementortoanyprocessfittingwithappropriate
matingsurfaces.ChemicalteeelementscannotbeconnectedtoastandardASMEorENpipeflange.
Pleasenotethatthepressurelimitforatransmitterwiththisremotesealisdeterminedbythesealflange.Maximumworking
pressurefortheflangeis2MPa(20baror290psi).
ThegasketsrequiredtoconnectthesealareavailablewiththesealelementsystemandcanbemanufacturedeitherinPTFEor
ingraphite.Theirrelevantprocesstemperaturelimitsarelistedintheproductdatasheet.
Theeightcapscrewsrequiredtoconnecteachsealaresuppliedwiththewedgeflowelement.Thesegasketsandboltshave
been specifically selected to meet the sealing and pressure rating requirements of the chemical tee element. Substitution of
thesesuppliedelementswithotherselectedbytheuserisnotrecommended.

NOTE:IfthetransmitterisaDPwithoneremotesealforliquidlevelmeasurement,refertoSectionConnectingtheLowSide
CompensatingLegonDPwithOneSeal beforeconnectingtheseal.

Connectthesealasfollows:
1.Removeprotectivecoveringfromseal.
2.Placethegasketonsealingsurfaceandinsertsealintoconnectionopening.
3.Insert8capscrewsandfingertighteneachscrew.
4.Followingadiagonalsequence,tighteneachcapscrewtoatorqueof12.4Nm.

IfyouexperiencealargeZeroshiftduetotemperaturechangeorerraticoutput:
1.Takethechemicalteesealsandmountthemfaceupseveralfeetabovethetransmitter.
2.Waitanhour.
3.Withyourthumbsatthe9oclockand3oclockpositionsonasinglediaphragm,lightlyapplythumbpressurealternating
rightthumb/leftthumb.Asyoupresswiththeright,theleftsideofthediaphragmshouldriseevenly.
Ifitdoesnotrisethenthereisalackoffill.Replacetheseal.
Ifitrisesandthensnaps(likeanoilcan)thenthediaphragmhasbeenstressed.Replacetheseal.
4.Lightlytapthediaphragmswiththeflatpartofyourfingernail.Adullthudisheardwhenfillisbehindthediaphragm.
Ifthereisairpresent,thesoundwillbeTinny.Replacetheseal.

Fig.11ChemicalTeeremoteseal

ThiskindofdiaphragmsealhasbeendesignedtobeconnectedtoABBWedgeprimaryelementmodelWRC.

ForanyotherdetailedinformationregardingS26CNsealsuchasvacuumservicerecommendations,processtemperaturelimits,
temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

16
Rotatingflangediaphragmsealsflushandextended(S26RA,S26REandS26RJ)
The extended and flush diaphragm remote seal is designed to connect to flanged pipe fitting, according to ASME B16.5
(S26WA),EN10921(S26WE)orJISB2220standards(S26RJ).
Forliquidlevelmeasurementinstallationsthesealconnectstoaflangedtanknozzle,complianttotherelevantstandard.The
sealingisprovidedbyaselectablesmoothorserratedgasketseatsurfacefinish.
ThemountingflangeisofrotatingtypeandcanbesuppliedoncarbonsteelorinstainlesssteelAISI316.

Fig.12Flushrotatingflangeseals(flushingringasanoption)

Fig.13Extendedrotatingflangeseals

Dimensionsmm(in)forS26RJ
Size/Rating A(dia)flushdiaphragm B(dia) C(dia) D(dia) E(dia) F G Nofholes

A50Class10K 60(2.36) 96(3.78) 120(4.72) 155(6.1) 19(0.75) 16(0.63) 9.5(0.37) 4


A50Class20K 60(2.36) 96(3.78) 120(4.72) 155(6.1) 19(0.75) 18(0.71) 9.5(0.37) 8
A50Class40K 60(2.36) 104.3(4.11) 130(5.12) 165(6.5) 19(0.75) 26(1.02) 9.5(0.37) 8
A80Class10K 89(3.5) 126(4.96) 150(5.91) 185(7.28) 19(0.75) 18(0.71) 9.5(0.37) 8
A80Class20K 89(3.5) 132(5.2) 160(6.3) 200(7.87) 23(0.91) 22(0.87) 9.5(0.37) 8
A80Class40K 89(3.5) 139.4(5.49) 170(6.69) 210(8.27) 23(0.91) 32(1.26) 9.5(0.37) 8
A100Class10K 89(3.5) 151(5.94) 175(6.89) 210(8.27) 19(0.75) 18(0.71) 9.5(0.37) 8
A100Class20K 89(3.5) 160(6.3) 185(7.28) 225(8.86) 23(0.91) 24(0.94) 9.5(0.37) 8

17

Dimensionsmm(in)forS26RA
A(dia)
Size/Rating
extended flushdiaphragm flushing B(dia) C(dia) D(dia) E(dia) F G Holes
diaphragm std. lowthick ringint.dia
2in.ASMECL150 48(1.9) 60(2.36) 58(2.28) 62(2.44) 92(3.62) 120.65(4.75) 152.4(6) 19.1(0.79) 17.5(0.6) 9.5(0.37) 4
2in.ASMECL300 48(1.9) 60(2.36) 58(2.28) 62(2.44) 92(3.62) 127(5) 165.1(6.5) 19.1(0.79) 20.8(0.8) 9.5(0.37) 8
2in.ASMECL600 NA 60(2.36) 58(2.28) 62(2.44) 92(3.62) 127(5) 165.1(6.5) 19.1(0.79) 25.4(1) 9.5(0.37) 8
2in.ASMECL900 NA 60(2.36) 58(2.28) 62(2.44) 92(3.62) 165(6.5) 215.9(8.5) 26(1.02) 38.1(1.5) 9.5(0.37) 8
2in.ASMECL1500 NA 60(2.36) 58(2.28) 62(2.44) 92(3.62) 165(6.5) 215.9(8.5) 26(1.02) 38.1(1.5) 9.5(0.37) 8
3in.ASMECL150 72(2.83) 89(3.5) 75(2.95) 92(3.62) 127(5) 152.4(6) 190.5(7.5) 19.1(0.79) 22.4(0.88) 9.5(0.37) 4
3in.ASMECL300 72(2.83) 89(3.5) 75(2.95) 92(3.62) 127(5) 168.15(6.62) 209.6(8.25) 22.4(0.88) 26.9(1.1) 9.5(0.37) 8
3in.ASMECL600 NA 89(3.5) 75(2.95) 92(3.62) 127(5) 168.15(6.62) 209.6(8.25) 22.4(0.88) 31.8(1.3) 9.5(0.37) 8
3in.ASMECL900 NA 89(3.5) 75(2.95) 92(3.62) 127(5) 190.5(7.5) 241(9.48) 26(1.02) 38.1(1.5) 9.5(0.37) 8
3in.ASMECL1500 NA 89(3.5) 75(2.95) 92(3.62) 127(5) 203.2(8) 266.7(10.5) 31.75(1.25) 47.7(1.88) 9.5(0.37) 8
4in.ASMECL150 94(3.7) 89(3.5) 75(2.95) 92(3.62) 157.2(6.2) 190.5(7.5) 228.6(9) 19.1(0.79) 22.4(0.88) 9.5(0.37) 8
4in.ASMECL300 94(3.7) 89(3.5) 75(2.95) 92(3.62) 157.2(6.2) 200.2(7.88) 254(10) 22(0.86) 30.2(1.19) 9.5(0.37) 8


Dimensionsmm(in)forS26RA
A(dia)
Size/Rating
extendeddiaphragm flushdiaphragm flushingring B(dia) C(dia) D(dia) E(dia) F G Holes
std. lowthick internaldia
DN5oENPN16 48(1.9) 60(2.36) 58(2.28) 62(2.44) 102(4.02) 125(4.92) 165(6.5) 18(0.71) 15(0.58) 9.5(0.37) 4
DN5oENPN40 48(1.9) 60(2.36) 58(2.28) 62(2.44) 102(4.02) 125(4.92) 165(6.5) 18(0.71) 18(0.67) 9.5(0.37) 4
DN5oENPN63 NA 60(2.36) 58(2.28) 62(2.44) 102(4.02) 135(5.31) 180(7.08) 22(0.86) 23(0.9) 9.5(0.37) 4
DN5oENPN100 NA 60(2.36) 58(2.28) 62(2.44) 102(4.02) 145(5.71) 195(7.67) 26(1.02) 27(1.06) 9.5(0.37) 4
DN8oENPN16 NA 60(2.36) 58(2.28) 62(2.44) 138(5.43) 160(6.3) 200(7.87) 18(0.71) 17(0.67) 9.5(0.37) 8
DN8oENPN40 72(2.83) 89(3.5) 75(2.95) 92(3.62) 138(5.43) 160(6.3) 200(7.87) 18(0.71) 21(0.83) 9.5(0.37) 8
DN8oENPN63 72(2.83) 89(3.5) 75(2.95) 92(3.62) 138(5.43) 170(6.7) 215(8.46) 22(0.86) 25(0.98) 9.5(0.37) 8
DN8oENPN100 NA 89(3.5) 75(2.95) 92(3.62) 138(5.43) 180(7.08) 230(9.05) 26(1.02) 33(1.3) 9.5(0.37) 8
DN100ENPN16 NA 89(3.5) 75(2.95) 92(3.62) 158(6.22) 180(7.08) 220(8.66) 18(0.71) 17(0.67) 9.5(0.37) 8
DN100ENPN40 NA 89(3.5) 75(2.95) 92(3.62) 162(6.38) 190(7.48) 235(9.25) 22(0.86) 21(0.83) 9.5(0.37) 8


ForanyotherdetailedinformationregardingS26Rxsealssuchaspressurelimits,vacuumservicerecommendations,gaskets
characteristics,processtemperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.































18
FlangedRemoteSealRingJoint(S26RR)
TheflangedflushdiaphragmremotesealringjointisdesignedtoconnecttoASMEflangedpipefitting.Onthe1in.sealthe
processflangeandnutsaresuppliedbytheuser.Onthe11/2in.and2in.sealstheprocessflange,boltsandnutsaresupplied
bytheuser.Ausersuppliedgasketmustbeinstalledbetweentheprocesspipeflangeandthelowerhousingofeachseal.
Thesesealsareavailablewithaflushingconnectioninthelowerhousing.Theconnectionprovidesa1/4in.NPTportintothe
cavityaroundthesealdiaphragmtopermitflushingtoremovetrappedsolids,whennecessary.Theflushingconnectionalso
providesameansofcheckingcalibrationwithoutdisconnectingtheelement.Agateorballvalveinstalledintheprocessline
near the seal connection allows the process to be isolated and calibration pressure can then be applied via the flushing
connection.Useofagateorballtypevalveavoidsrestrictingtheprocesslineduringnormaloperation.

Connectthesealasfollows:
1.Installagasketbetweentheprocessflangeandthesealelementgasketsurface.
2.Boltthesealtotheprocessflange.
3.Tightenbolts(nutson1in.seal)firmlyanduniformlyinaccordancewithstandardindustrialflangeboltingpractices.
4.Ifthesealhasaflushingconnection,connectaflushinglinewithappropriatevalvingtothe1/4in.connection.

Fig.14RotatingflangeremotesealRingJointconstruction

Dimensionsmm(in)forS26RR
Size/Rating A(dia) B(dia) C(dia) D(dia) E(dia) F G H(dia) R holes

11/2in.ASMECL150 48(1.89) 83(3.27) 98.6(3.88) 127(5) 15.75(0.62) 17.5(0.69) 17.3(0.68) 65.1(2.56) R19 4
11/2in.ASMECL300 48(1.89) 90(3.54) 114.4(4.5) 155(6.1) 22.35(0.88) 20.6(0.81) 17.3(0.68) 68.3(2.69) R20 4
11/2in.ASMECL600 48(1.89) 90(3.54) 114.3(4.5) 155(6.1) 22.35(0.88) 22.4(0.88) 17.3(0.68) 68.3(2.69) R20 4
11/2in.ASMECL900/1500 48(1.89) 92(3.62) 124(4.88) 177.8(7) 28.45(1.12) 31.8(1.25) 20.8(0.82) 68.3(2.69) R20 4
11/2in.ASMECL2500 48(1.89) 114(4.49) 146.1(5.75) 203.2(8) 31.75(1.25) 44.5(1.75) 20.8(0.82) 82.6(3.25) R23 4
2in.ASMECL150 60(2.36) 102(4.02) 120.65(4.75) 152.4(6) 19.05(0.75) 19.05(0.75) 17.3(0.68) 82.6(3.25) R22 4
2in.ASMECL300 60(2.36) 108(4.25) 127(5) 165.1(6.5) 19.05(0.75) 22.35(0.88) 17.3(0.68) 82.6(3.25) R23 8
2in.ASMECL600 60(2.36) 108(4.25) 127(5) 165.1(6.5) 19.05(0.75) 25.4(1) 17.3(0.68) 82.6(3.25) R23 8
2in.ASMECL900/1500 60(2.36) 124(4.88) 165(6.5) 215.9(8.5) 25.4(1) 38.1(1.5) 20.8(0.82) 95.3(3.75) R24 8
2in.ASMECL2500 60(2.36) 133(5.24) 171.5(6.75) 235(9.25) 28.45(1.12) 50.8(2) 20.8(0.82) 101.6(4) R26 8
3in.ASMECL150 89(3.5) 133(5.24) 152.4(6) 190.5(7.5) 19.05(0.75) 23.87(0.94) 17.3(0.68) 114.3(4.5) R29 4
3in.ASMECL300 89(3.5) 146(5.75) 168.15(6.62) 209.55(8.25) 22.35(0.88) 28.44(1.12) 17.3(0.68) 123.8(4.87) R31 8
3in.ASMECL600 89(3.5) 146(5.75) 168.15(6.62) 209.55(8.25) 22.35(0.88) 31.75(1.25) 17.3(0.68) 123.8(4.87) R31 8
3in.ASMECL900 89(3.5) 155(6.10) 190.5(7.5) 241.3(9.5) 25.4(1) 38.1(1.50) 20.8(0.82) 123.8(4.87) R31 8
3in.ASMECL1500 89(3.5) 168(6.61) 203.2(8) 266.7(10.5) 31.75(1.25) 47.8(1.88) 20.8(0.82) 136.5(5.37) R35 8
3in.ASMECL2500 89(3.5) 168(6.61) 228.6(9) 304.8(12) 35.05(1.38) 66.5(2.62) 20.8(0.82) 127(5) R32 8

ForanyotherdetailedinformationregardingS26RRsealsuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

19
Fixedflangediaphragmsealsflush(S26FA,S26FE)
Theseflushdiaphragmsealaredesignedtoconnecttoflangedpipefitting,accordingtoASMEorENstandards.
Forliquidlevelmeasurementinstallations,thesealconnectstoaflangedtanknozzle,complianttorelevantstandard.
Thesealingisprovidedbyaselectablegasketseatsurfacefinish.Thefixedmountingflangeisintegralwiththeseal.

ASMEandEN10921smoothandFormB1
(flushingringasoption)

EN10921FormE EN10921FormD

Fig.15 Fixedflangediaphragmsealsdifferentversions

ForanyotherdetailedinformationregardingS26Fxsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

20
Dimensionsmm(in)forS26FA
A(dia)
Size/Rating
flushdiaphragm flushingring B(dia) C(dia) D(dia) E(dia) F G Holes
std. lowthick internaldia
2in.ASMECL150 60(2.36) 58(2.28) 62(2.44) 92(3.62) 120.65(4.75) 152.4(6) 19.1(0.79) 17.5(0.6) 2(0.08) 4
2in.ASMECL300 60(2.36) 58(2.28) 62(2.44) 92(3.62) 127(5) 165(6.5) 19.1(0.79) 20.8(0.8) 2(0.08) 8
2in.ASMECL600 60(2.36) 58(2.28) 62(2.44) 92(3.62) 127(5) 165(6.5) 19.1(0.79) 25.4(1) 7(0.27) 8
3in.ASMECL150 89(3.5) 75(2.95) 92(3.62) 127(5) 152.4(6) 190.5(7.5) 19.1(0.79) 22.4(0.86) 2(0.08) 4
3in.ASMECL300 89(3.5) 75(2.95) 92(3.62) 127(5) 168.15(6.62) 209.6(8.25) 22.4(0.86) 26.9(1.1) 2(0.08) 8
3in.ASMECL600 89(3.5) 75(2.95) 92(3.62) 127(5) 168.15(6.62) 209.6(8.25) 22.4(0.86) 31.8(1.3) 7(0.27) 8
4in.ASMECL150 89(3.5) 75(2.95) 92(3.62) 157.2(6.2) 170(6.7) 228.6(9) 19.1(0.79) 22.4(0.86) 2(0.08) 8

Dimensionsmm(in)forS26FEsmoothandFormB1
A(dia)
Size/Rating
flushdiaphragm flushingring B(dia) C(dia) D(dia) E(dia) F G Holes
std. lowthick internaldia
DN50ENPN16 60(2.36) 58(2.28) 62(2.44) 102(4.02) 125(4.92) 165(6.5) 18(0.71) 15(0.58) 3(0.12) 4
DN50ENPN40 60(2.36) 58(2.28) 62(2.44) 102(4.02) 125(4.92) 165(6.5) 18(0.71) 18(0.67) 3(0.12) 4
DN50ENPN63 60(2.36) 58(2.28) 62(2.44) 102(4.02) 135(5.31) 180(7.08) 22(0.86) 23(0.9) 3(0.12) 4
DN50ENPN100 60(2.36) 58(2.28) 62(2.44) 102(4.02) 145(5.71) 190.5(7.5) 26(1.02) 27(1.06) 3(0.12) 4
DN80ENPN16 89(3.5) 75(2.95) 92(3.62) 138(5.43) 160(6.3) 200(7.87) 18(0.71) 17(0.67) 3(0.12) 8
DN80ENPN40 89(3.5) 75(2.95) 92(3.62) 138(5.43) 160(6.3) 200(7.87) 18(0.71) 21(0.83) 3(0.12) 8
DN80ENPN63 89(3.5) 75(2.95) 92(3.62) 138(5.43) 170(6.7) 215(8.46) 22(0.86) 25(0.98) 3(0.12) 8
DN80ENPN100 89(3.5) 75(2.95) 92(3.62) 138(5.43) 180(7.08) 230(9.05) 26(1.02) 33(1.3) 3(0.12) 8
DN100ENPN16 89(3.5) 75(2.95) 92(3.62) 138(5.43) 180(7.08) 220(8.66) 18(0.71) 17(0.67) 3(0.12) 8

Dimensionsmm(in)forS26FEFormE
A(dia)
Size/Rating
flushdiaphragm B(dia) C(dia) D(dia) E(dia) F G Holes
std. lowthick
DN50ENPN16 60(2.36) 58(2.28) 87.(3.42) 125(4.92) 165(6.5) 18(0.71) 13.5(0.53) 4.5(0.18) 4
DN50ENPN40 60(2.36) 58(2.28) 87.(3.42) 125(4.92) 165(6.5) 18(0.71) 15.5(0.61) 4.5(0.18) 4
DN50ENPN63 60(2.36) 58(2.28) 87.(3.42) 135(5.31) 180(7.08) 22(0.86) 21.5(0.85) 4.5(0.18) 4
DN50ENPN100 60(2.36) 58(2.28) 87.(3.42) 145(5.71) 190.5(7.5) 26(1.02) 25.5(1) 4.5(0.18) 4
DN80ENPN16 89(3.5) 75(2.95) 120(4.72) 160(6.3) 200(7.87) 18(0.71) 15.5(0.61) 4.5(0.18) 8
DN80ENPN40 89(3.5) 75(2.95) 120(4.72) 160(6.3) 200(7.87) 18(0.71) 19.5(0.77) 4.5(0.18) 8
DN80ENPN63 89(3.5) 75(2.95) 120(4.72) 170(6.7) 215(8.46) 22(0.86) 23.5(0.92) 4.5(0.18) 8
DN80ENPN100 89(3.5) 75(2.95) 120(4.72) 180(7.08) 230(9.05) 26(1.02) 31.5(1.24) 4.5(0.18) 8
DN100ENPN16 89(3.5) 75(2.95) 149(5.87) 180(7.08) 220(8.66) 18(0.71) 15(0.59) 5(0.20) 8

Dimensionsmm(in)forS26FEFormD
Size/Rating A(dia) B(dia) C(dia) D(dia) E(dia) F H(dia) I(dia) L Holes
std. lowthick
DN50ENPN16 60(2.36) 58(2.28) 125(4.92) 125(4.92) 165(6.5) 18(0.71) 15(0.59) 72(2.83) 88(3.46) 4(0.16) 4
DN50ENPN40 60(2.36) 58(2.28) 125(4.92) 125(4.92) 165(6.5) 18(0.71) 18(0.71) 72(2.83) 88(3.46) 4(0.16) 4
DN50ENPN63 60(2.36) 58(2.28) 135(5.31) 135(5.31) 180(7.08) 22(0.86) 23(0.91) 72(2.83) 88(3.46) 4(0.16) 4
DN50ENPN100 60(2.36) 58(2.28) 145(5.71) 145(5.71) 190.5(7.5) 26(1.02) 27(1.06) 72(2.83) 88(3.46) 4(0.16) 4
DN80ENPN16 89(3.5) 75(2.95) 160(6.3) 160(6.3) 200(7.87) 18(0.71) 17(0.67) 105(4.13) 121(4.76) 4(0.16) 8
DN80ENPN40 89(3.5) 75(2.95) 160(6.3) 160(6.3) 200(7.87) 18(0.71) 21(0.83) 105(4.13) 121(4.76) 4(0.16) 8
DN80ENPN63 89(3.5) 75(2.95) 170(6.7) 170(6.7) 215(8.46) 22(0.86) 25(0.92) 105(4.13) 121(4.76) 4(0.16) 8
DN80ENPN100 89(3.5) 75(2.95) 180(7.08) 180(7.08) 230(9.05) 26(1.02) 33(1.3) 105(4.13) 121(4.76) 4(0.16) 8
DN100ENPN16 89(3.5) 75(2.95) 180(7.08) 180(7.08) 220(8.66) 18(0.71) 17(0.67) 128(5.04) 149(5.91) 4.5(0.18) 8

21
OfflineThreadedConnectionSeal(S26TT)
TheofflinethreadedremotesealconnectsdirectlytoaprocesspipeviatheNPTconnectioninthelowerhousing.Connections
for1/4,1/2,3/4,1,and11/2in.processpipingareavailable.Inadditiontotheprocessconnection,mechanicalsupportforthe
weightofthesealmayberequired,especiallyforthe1/4and1/2inchpipesizes.
Theseelementsareavailablewithaflushingconnectioninthelowerhousing.Theconnectionprovidesa1/4in.NPTportinto
thecavityaroundthesealdiaphragmtopermitflushingtoremovetrappedsolids,whennecessary.Theflushingconnectionalso
providesameansofcheckingcalibrationwithoutdisconnectingtheelement.Agateorballvalveinstalledintheprocessline
nearthesealelementconnectionallowstheprocesstobeisolatedandcalibrationpressurecanthenbeappliedviatheflushing
connection.Useofagateorballtypevalveavoidsrestrictingtheprocesslineduringnormaloperation,notagasket.
ConnectthesealelementtotheprocesspipeattheNPTconnectionintheendofthelowerhousing.
Iftheelementhasaflushingconnection,connectaflushinglinewithappropriatevalvingtothe1/4in.NPTconnectioninthe
sideofthelowerhousing.

Fig.16Offlinethreadeddiaphragmseal

Notethatthisdiaphragmsealfeaturestwoflushingconnections.

Dimensionsmm(in)forS26TT
Size(thread) D(dia) Q
1/4in.NPT 109.2(4.3) 53.3(2.1)
1/2in.NPT 109.2(4.3) 53.3(2.1)
3/4in.NPT 109.2(4.3) 63.5(2.5)
1in.NPT 109.2(4.3) 63.5(2.5)
11/2in.NPT 109.2(4.3) 63.5(2.5)

S26TTcanbesuppliedwithdifferentkindsofbolts:stainlesssteelAISI316,carbonsteeloralloysteel(NACEcompliant)

ForanyotherdetailedinformationregardingS26TTsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

22
OffLineFlangedConnectionRemoteSeal(S26MA,S26ME)
Available with ASME (S26MA) and EN (S26ME) flanged process connection, the Offline model matches small process
connections.Recommendedforcleanprocessfluidmeasurements.
Theseelementsareavailablewithaflushingconnectioninthelowerhousing,selectableonrequest.

Fig.17Offlineflangeddiaphragmseal

Notethatthisdiaphragmsealfeaturestwoflushingconnections.

Dimensionsmm(in)forS26MAandS26ME
Size/Rating C(4studs)
A(dia) B(dia) Length Thread D(dia) E(dia) F
1/2in.ASMECL150 110(4.33) 60.5(2.38) 39(1.53) 1/2in.13UNC 35.1(1.38) 15.8(0.62) 1.6(0.06)
1/2in.ASMECL300 110(4.33) 66.5(2.62) 39(1.53) 1/2in.13UNC 35.1(1.38) 15.8(0.62) 1.6(0.06)

1in.ASMECL150 110(4.33) 79.4(3.12) 39(1.53) 1/2in.13UNC 50.8(2) 26.7(1.05) 1.6(0.06)


1in.ASMECL300 124(4.88) 88.9(3.5) 51(2) 5/8in.11UNC 50.8(2) 26.7(1.05) 1.6(0.06)
11/2in.ASMECL150 127(5) 98.4(3.87) 39(1.53) 1/2in.13UNC 73(2.87) 41(1.61) 1.6(0.06)
11/2in.ASMECL300 155(6.1) 114.3(4.5) 57(2.24) 3/4in.10UNC 73(2.87) 41(1.61) 1.6(0.06)
DN25ENPN1640 115(4.52) 85(3.34) 42(1.65) M12 68(2.67) 28.5(1.12) 2(0.08)

DN40ENPN1640 150(5.9) 110(4.33) 48(1.89) M16 88(3.46) 43.1(1.69) 3(0.12)

ForanyotherdetailedinformationregardingS26Mxsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,bolts,gasketseatfinishing,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

23
RemoteSealforUreaService(S26PN)
ThisspecializedremotesealismanufacturedfrommaterialswhichmatchtheaggressivelycorrosiveconditionswhenUreais
presentintheprocess.
Theureagradewettedmaterialsensurestableperformanceseveninhightemperature/highvacuumconditions.

Fig.18UreaserviceextendedsealswithASMEflange.

For any other detailed information regarding S26PN seal such as pressure limits, vacuum service recommendations, process
temperaturelimits,bolts,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

24
ButtonTypeRemoteSeal(S26BN)
ThisremotesealisdesignedtoconnecttoaprocessviatheNPTthreadedconnectionortomatchpipefittingwithaninterface
suitablefortheprovidedmatingflange.Thebuttonseals,thankstotheiraccuratedesign,areprojectedtocarryoutaccurate
measurementswithmedium/highcalibratedspan(2MPa20bar290psiapproxorgreater).
Thebuttontyperemotesealistobeusedonlyforgaugepressure.
Thistypeofconnectionisespeciallysuitableforplasticandresinsproductionprocesses.
ForanyotherdetailedinformationregardingS26BNsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

Fig.1931/4in.flangeextendedtype91

Fig.2031/4in.flangeextendedtype91modified

25

Fig.2111/2in.threadeduniontype92or92modified

Fig.22Brackettype89

Fig.23Universaltype90

26
UnionConnectionRemoteSeals(S26UN)
Theunionconnectionremotesealsareusedexclusivelyforpressuremeasurement.Thesealisavailablewithoutaweldbushing,
withanoptionalweldbushing,orwithanoptionalchemicalteeflange.AnOringisusedtosealtheprocessconnection.TheO
ringmaterialcanbesiliconerubberfortemperaturesupto350F(177C)orTeflonTFEfortemperaturesupto400F(204C).
Note that the operating temperature limit for the seal can be lower than the Oring temperature limit because of fill fluid
limitations (refer to the fill fluid parameter table in the Specification Sheet). Connect the seal as described in the applicable
followingsection.

SealwithoutWeldBushing
Thesealwithoutaweldbushing,seeFigure23,mustbeconnectedtoausersuppliedfittinghavingmatingsurfacedimensions
asshowninthefigure.Connectthesealasfollows:
1.LubricatethesealingOringwithSiliconegasketgrease.
2.MounttheOringontheendoftheseal.
3.InsertsealintomatingfittingandtightentheunionconnectionnuttopresstheOringagainstthefittingsurface.

Fig.24Unionconnectionremotesealbasicversion

SealwithWeldBushing
Theremotesealwithaweldbushingincludesabushingwhichprovidesthematingsurfaceforthesealelement.Thebushing
mustbeweldedtotheprocesspipingorvesselbeforeinstallingtheseal.Usethefollowingprocedure:
1.Weldbushingtopipeorvesselinaccordancewithstandardindustrialpractices.
2.Removeprotectivecoveringfromseal.
3.LubricatethesealingOringwithsiliconegasketgrease.
4.MounttheOringontheendoftheelement.
5.InsertelementintomatingfittingandtightentheunionconnectionnuttopresstheOringagainstthefittingsurface.

Fig.25Unionconnectionremotesealwithweldbushing

27
SealwithChemicalTeeFlange
Theunionconnectionsealwithachemicalteeflangeisdesignedtoconnecttoanyprocessfittingwhichacceptsachemicaltee
sealelement(refertoChemicalTeeSealformoreinformation).Theunionsealconnectstothechemicalteeflangewhichserves
asanadaptertopermitconnectionoftheunionsealtoachemicalteetypefitting.
In addition to the oring required on the remote seal, this option requires a gasket to seal the chemical tee flange at its
connectiontotheprocessfitting.Thisgasketisincludedwhenthechemicalteeflangeoptionisspecified.Thegasketmaterialis
PTFEwithsilicafillerandthemaximumoperatingtemperatureis400F(204C).Notethatthesealoperatingtemperaturelimit
canbelowerthanthisgaskettemperaturelimitiftheremotesealoringissiliconerubberwitha350F(177C)rating,orifthefill
fluid has a lower temperature limit (refer to the fill fluid parameter table in the Specification Sheet). The eight cap screws
requiredtoconnecttheflangearesuppliedwiththematingprocessfitting.Thegasketandboltshavebeenspecificallyselected
tomeetthesealingandpressureratingrequirementsofthechemicalteeflange.
Substitutionofusersuppliedboltsandgasketsisnotrecommended.

Connectthesealelementusingthefollowingprocedure:
1.Connectthechemicalteeflangeasfollows:
Placethegasketontheflangesealingsurfaceandinsertflangeintotheprocessfitting.
Insert8capscrewsandfingertighteneachscrew.
Followingadiagonalsequence,tighteneachcapscrewtoatorqueof12.4Nm(110inchlbs).

2.Connecttheunionsealtothechemicalteeflangeasfollows:
Removeprotectivecoveringfromseal.
LubricatethesealingOringwithSiliconegasketgrease.
MounttheOringontheendoftheseal.
InsertthesealintothechemicalteeflangeandtightentheunionconnectionnuttosealtheOringagainstitsmating
surface.

Fig.26UnionconnectionremotesealwithChemicalTeeflange

NOTE:Thepressurelimitforatransmitterwiththissealisdeterminedbythechemicalteeflange.Maximumworkingpressure
fortheflangeis2000KPa(300psi).

ForanyotherdetailedinformationregardingS26UNsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

28
FoodandSanitaryRemoteSeal(S26SS)
Sanitary diaphragm seals have been specifically developed for food, sanitary, chemical and pharmaceutical applications,
complyingwiththestringent3Arequirements.
Available with different process fittings (Triclamp, Cherry Burrell, Union Nut and Sanitary), this model highlights ABBs
commitmenttosatisfyusersneedsapproachingeventhemostdemandingprocessessuccessfully.

CherryBurrellAseptic
The sanitary aseptic remote seal is designed to connect to a 4in sanitary fitting: either an aseptic tank spud or a 4in Cherry
Burrellasepticferrule.Thetankspud,gasketsandVbandclampareavailableoptionwiththesealelement.
NOTE: The tank spud or ferrule required for connection of this seal element must be welded to the process vessel prior to
connectingtheelement,followingrecommendedweldingandpressuretestingprocedure.WeldtheCherryBurrellferruletothe
processvesselinaccordancewithmanufacturersrecommendations.
Connectthesanitaryasepticsealtothetankspudorferruleasfollows:
1.Removeprotectivecoveringfromseal.
2.InstalltwoOringsontheperipheryofthesealandanOringundereachsteamconnectionfitting.
3.Insertthesealintothetankspudorferruleandplacethesteamconnectioncapagainstthebackoftheelement.
4.PositiontheclampringasshownandtightentosealtheOringsagainstthespud.
5.Connecttheinletandoutletsteamlinestothe1/8.NPTsteamconnectionsinthecap.Steamtemperaturemustnotexceed
149C(300F);pressuremustnotexceed358.5KPa(52psi).
(52psi).

4in.CherryBurrellaseptic

4in.Asepticflangedconnection

Fig.274in.sanitaryseal
29
SanitarySealwithFlushDiaphragm
Thesanitaryremotesealwithflushdiaphragmisdesignedtoconnecttoa4inchsanitarytankspud.
ThetankspudandprocessOring,madeofBunaorViton,areavailablewiththeseal.
Connectthesealasfollows:
1.RemoveprotectivecoveringfromsealandinstalltheprocessOring.
2.Insertthesealintothetankspud.
3.PositiontheclampringasshownandtightentosealtheOringagainstthespud.

Fig.28Sanitarysealwithflushdiaphragm

SanitarySealwithExtendedDiaphragm
Thesanitaryremotesealwithextendeddiaphragmisdesignedtoconnecttoa4inchsanitarytankspudwithanextendedneck
(2,4or6inches).Thetankspud,anEthylenePropyleneprocessOringandVBandclampringareavailablewiththeseal.
Connectthesealasfollows:
1.Weldthetankspudintotheprocessvessel.Besuretoorientthespudsothatthedrainholeisatthelowestpoint.
2.RemoveprotectivecoveringfromsealandInstalltheOring.
3.Insertthesealintothetankspud.
4.Positiontheclampringoverthejointbetweenthesealandspudandtightentheclamp.
NOTE:Thetankspudrequiredforconnectionofthissealelementmustbeweldedtotheprocessvesselpriortoconnectingthe
seal,followingarecommendedweldingandpressuretestingprocedure.

Fig.29Sanitarysealwithextendeddiaphragm
30
UnionNutandTriclampremoteseal
TheUnionNutandTriclampsealsaredesignedforconnectionbyUnionNutaccordingtoDIN11851F50orF80and2in,3in,4in
Triclampsanitaryfittings.Avarietyofgasketsandclampringsforthesealsareavailable.

Note: this kind of diaphragm seal


doesnotcomplywiththestringent3A
standard requirements. Therefore, it
is not covered by the relevant
certification.

Dimensionsmm(in)forS26SS
UnionNuttoDIN11851
Size A(dia) B(dia) C(dia) D
F50 42(1.65) 78(3.07) 92(3.62) 22(0.87)
Fig.30UnionNutsanitaryseal F80 72(2.83) 110(4.33) 127(5) 29(1.14)

Dimensionsmm(in)forS26SS
Triclamp
Size A(dia) B(dia)
2in. 56.3(2.2) 64(2.5)
3in. 83(3.26) 91(3.58)
4in. 110.3(4.34) 119(4.68) Fig.31Triclampsanitaryseal

31
CherryBurrellseal
TheCherryBurrellsealsaredesignedforconnectionto2in,3inor4inCherryBurrellILinesanitaryfittings.
A4in.Vbandclampisoptionallyavailableforthe4in.variant.

Fig.32CherryBurrellsanitaryseal

Size Dimensions mm (in) for S26SS Cherry Burrell


A (dia) B (dia) C (dia) D (dia) E F G H
2 in. 67 (2.64) 56 (2.2) 42 (1.65) 57 (2.24) 3.2 (0.13) 6.5 (0.26) 12.5 (0.49) 3 (0.12)
3 in. 98.4 (3.87) 81 (3.19) 72.42 (2.85) 83.8 (3.3) 2.4 (0.09) 7.9 (0.31) 15 (0.59) 3 (0.12)
4 in. 124 (4.88) 111.25 (4.38) 72.42 (2.85) 109.3 (4.3) 2.4 (0.09) 7.9 (0.31) 15 (0.59) 3 (0.12)

Beverageboltedseal
Thissealhasbeendesignedtobesuitedforspecificapplicationsinbeverageprocesses.Thesanitaryboltedsealisavailableonly
forgaugeandabsolutetransmitterswithdirectmountsealconstruction.Spudprocessfittingisavailableonrequest.

Note: this kind of diaphragm seal


doesnotcomplywiththestringent3A
standard requirements. Therefore, it
is not covered by the relevant
certification.

Fig.33Beverageboltedsealfor266HDHand266NDHtransmitters

Foranyother detailedinformationregardingS26SSsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

32
NOTE(3Arequirements)
AllsanitaryS26SSremotediaphragmsealsandotherassociateddirectmountsanitarydiaphragmsealsmustbemountedin
suchawayastoallowfordrainageofallprocessfluids.Tankspudfittingsmustbemountedwiththeleakdetectionholelocated
atthebottomforproperdrainagebygravity.Sanitaryapplicationsmustbeperformedsothatallweldingsurfacearesmoothin
ordertopreventdirtfromlodginginsidesurfaceirregularities.
3Asealsareequippedwithaspecificlabel(seebelow)indicatingthetemperaturerangeandgasketcharacteristics.

Saddleandsocketdiaphragmseal(S26VN)
Thesaddleandsocketsealarethebestsolutionwhenthediaphragmneedstobeasclosestaspossibletotheprocessmedia.
Thesearetypicallyinstalledbyweldingtotheprocesspipeswithfluidathighviscosity.Saddleandsocketprocessconnection
fittingsareavailableasoptionselection,availableonlyinAISI316Lss.

Fig.34S26VNSaddletype

Dimensionsmm(in)forS26VNsaddletype
Fittingconnection/Size
A(dia) B C(dia) R
Saddle2in. 55(2.17) 48(1.89) 40(1.57) 30
Saddle21/2in. 76(3.0) 45(1.77) 52(2.05) 45
Saddle3in. 76(3.0) 45(1.77) 50(1.97) 45
Saddle4in. 76(3.0) 41(1.61) 50(1.97) 57
Saddle5in. 76(3.0) 40(1.57) 50(1.97) 70
Saddle6in. 76(3.0) 36(1.42) 50(1.97) 85

33

Fig.35S26VNSockettype

Dimensionsmm(in)forS26VNsockettype
Fittingconnection/Size
A(dia) B C
Socket1/2in. 21.8(0.86) 15.9(0.63) 86(3.39)
Socket3/4in. 27(1.06) 21.2(0.83) 96(3.78)
Socket1in. 33.6(1.32) 26.8(1.06) 101(3.98)
Socket11/2in. 48.5(1.91) 41(1.61) 121(4.76)
Socket2in. 60.5(2.38) 52.5(2.07) 121(4.76)

ForanyotherdetailedinformationregardingS26VNsealssuchaspressurelimits,vacuumservicerecommendations,process
temperaturelimits,temperatureeffectsandconfiguration,pleaserefertoproductdatasheet.

34
Inlinediaphragmseal(S26JN)
Inlinesealsaresuitableformeasuringthepressureoffluidsinpipes.Thepressuremeasuringdiaphragmformsthepipewall
makingthistypeofsealsuitableformeasuringthepressureofflowingfluidsparticularlythosethatarehighlyviscousorcontain
solids.Thisparticulardiaphragmsealisavailableonlyondirectmountgaugeandabsolutepressuretransmitters.

Fig.36Inlinediaphragmseal

Dimensionsmm(in)forS26JN
Size/Rating D(dia) Mb(dia)
1in./DN25 63(2.48) 28.5(1.12)
11/2in./DN40 85(3.35) 43(1.69)
2in./DN50 95(3.74) 54.5(2.15)

3in./DN80 130(5.12) 82.5(3.25)

35
PulpandPaperdiaphragmseals(S26KN)
Models266HDH,266NDH,266GDHand266ADHintegrateadirectmountsealonthepositiveside,havingthenegativeside
referenceatatmosphericorvacuumpressure,respectivelyforgaugeorabsolutemeasurements.
The integral seal is available in 1in and 11/2in diaphragm sizes specifically designed for pulp & paper industry applications,
providingaflushdiaphragmtotheprocess;thissolutionensuresthemeasurereliabilityavoidingallproblemsofcrystallization/
polymerization,typicalofpulpshavingviscosityasinthepapermill.
Useofproperproceduresandfixturesavoidtherisksofpersonalinjuryorplantdamage.
Therecommendedminimumradiusofcurvatureofthetankis0.91m(3feet).
TM TM.
AllowedfillfluidsareSiliconeOilDC200 andmineraloilMarcol82

1in.and11/2in.sizewith1in.and11/2in.NPTthreadedconnection
This particular process connection can withstand a maximum working pressure of 345bar (5000psi). The wetted parts are
availableinAISI316L,HastelloyC276TMandDiaflex(antiabrasiondiaphragmtreatment).

Fig.37 1in.and11/2in.NPTthreadedconnection

1in.and11/2in.sizewithG1in.BandG11/2in.Bthreadedconnection
This particular process connection can withstand a maximum working pressure of 600bar (8700psi). The wetted parts are
availableinAISI316L,HastelloyC276TMandDiaflex(antiabrasiondiaphragmtreatment).

Fig.38 1in.BandG11/2in.Bthreadedconnection

36
1in.and11/2in.sizetospudwith1or2screw(s)fixing
Considerthatthiskindofpulpandpapersealingwithgasketsealcanwithstandamaximumworkingpressureof:
30bar(435psi)incaseof1screwfixing
50bar(725psi)incaseof2screwsfixing
ThewettedpartsareavailableinAISI316L,HastelloyC276TMandDiaflex(antiabrasiondiaphragmtreatment).
Whileinstalling,pleaseusetheappropriateVitongasket.

Fig.40 1in.and11/2in.sizetospud

1in.Gwithballvalveconnection 11/2in.sealingconnectiontothreadedspud
This special connection allows easy instrument mounting (M44x1.5)
on ball valves which are extensively used in industry since Consider that this particular process connection can
they are easy to install and repair. They offer user the withstandamaximumworkingpressureof50bar(725psi).
opportunity of avoiding complete plant shutdown. Whileinstalling,pleaseusetheappropriatePTFEgasket.
Consider that this particular process connection can ThewettedpartsareavailableinAISI316L,HastelloyC276
TM
withstandamaximumworkingpressureof40bar(580psi). andDiaflex(antiabrasiondiaphragmtreatment).
ThewettedpartsareavailableonlyinHastelloyC276TM.

Fig.41 BallValveconnection Fig.42 Threadedconnectiontospud


37
WeldingProcedureforSanitaryandPulp&Paperspuds
1.Usingappropriatesizeholesaw,cutaholeintheprocessvesseltoacceptthespud.Theholeshouldcouplewiththespud
perfectly.
2.RemovetheweldspudfromthetransmitterandsecondlythePTFEgasketfromtheweldspud(ifany).
3.Positiontheweldspudinthevesselholeandstartwelding.Forthisimportantphase,pleaselookatandfollowthewelding
sequence(Figure43).
4.Cooleachsectionproperlybeforeproceedingtothenextsection.
5.Usebetween100and150amps.andadjusttheamperage,ifneeded,forspudpenetration.
6.Aftertheweldspudhascooled,installtheTeflongasketintotheweldspud.Ensurethegasketisproperlypositionedwithin
theweldspud;improperplacementcancauseaprocessleak.
7.Positionthetransmitterintothespudandengagethethreads.Rotatethetransmitterandthreaditcompletely.
8.Tightenthetransmitteror,ifincaseofpulpandpapercylindricalspud,installtheneededbolt(s).

NOTE:usersshouldinstallspudscarefully.

Pleasefollowthenotesbelow:
Position inside gasket (if any) perfectly in order to avoid
processleaks.
Improperinstallationmayresultinspuddistortion.
Excessiveheatwilldistortthespud.
Allow adequate cooling between welding and the following
operations.
Weld the spud properly on tank or vessel walls in order to
avoidpersonalinjuryorplantdamage.

Fig.43 Spudweldingsequence

SanitarySpuds(3Arecommendations)
Useofproperproceduresandfixturespermitstheflanges
tobeweldedtothemaximumwallthickness.

The recommended minimum radius of curvature of the


tankis0.91m(3feet).

The spud has to be welded so that the welding surface


remainssmoothandwithoutirregularitieswheredirtcan
lodge.Thespudleakdetectionholehastobepositioned
atthebottomofthespud.

For sanitary spud welding procedure, please refer to


figure43.

Fig.44 Sanitaryspudandweldingpoint

38
CONNECTINGLOWSIDECOMPENSATIONLEGONADPWITHONESEAL
Differentialpressuretransmitterswithasingleremotesealelementareusedforliquidlevelinstallations.Thesealelementis
connectedusingtheprocedureapplicabletothetypeofsealelement.Theconnectionrequirementsforthelowsidedependson
whetherthetankisopentoatmosphereorclosed(pressurizedorevacuated).Makethelowsideconnectionasdescribedbelow.

OpenTankInstillation
Thelowsideofthetransmittermustbeventedtoatmosphereforopentankinstallations.Besuretherearenoplasticshipping
plugsorotherrestrictionsinthelowsideflangeports.Connectthesealelementusingtheprocedureapplicabletotheelement
type.RefertotheConnectingtheRemoteSealssection.

ClosedTankInstallation
Beforeconnectingthesealelementtothetankfitting,determinewhetherawetordrycompensatinglegisrequiredonthelow
sideofthetransmitter.RefertotheLocatingtheTransmitterandRemoteSealssection.
Thetransmitterisshippedwiththevent/drainpluglocatednearthetopofthelowsideflange.Inthispositionthevalvepermits
ventingofentrappedgas(normallyair)whenawet(liquidfilled)compensatinglegisrequired.Ifadrylegisrequired,remove
theboltsandrotatethelowsideflange180sothatthevent/drainscrewispositionedatthebottomoftheflangetopermit
drainingofcondensatefromtheprimarycavity.Retightentheboltstoatorqueof20ftlb.
Afterproperlylocatingthevent/drainscrew,connectthelowsidetothecompensatingleg,andconnectthesealelementusing
theprocedureapplicabletotheelementtype.RefertotheConnectingtheRemoteSealssection.
Contact us

ABB Automation Products GmbH ABB Inc. Note

OI/S26-EN 03.2010 | 3KXP600003R4201


We reserve the right to make technical changes or
Borsigstr. 2 Automation Technology Products
modify the contents of this document without prior
63755 Alzenau, Germany 125 E. County Line Rd notice. With regard to purchase orders, the agreed
Phone: +49 551 905 534 Warminster PA 18974-4995, USA particulars shall prevail. ABB does not accept any
Fax: +49 551 905 555 Phone: +1 215 674 6000 responsibility whatsoever for potential errors or
possible lack of information in this document.
Fax: +1 215 674 7183
ABB Limited We reserve all rights in this document and in the
Oldends Lane ABB Engineering (Shanghai) Ltd. subject matter and illustrations contained therein.
Stonehouse No.5, Lane 369, Chuangye Road Any reproduction, disclosure to third parties or
utilization of its contents - in whole or in parts is
Gloucestershire GL 10 3TA, UK Kangqiao Town, Nanhui District
forbidden without prior written consent of ABB.
Phone: +44 1453 826 661 Shanghai, 201319, P.R. China
Fax: +44 1453 829 671 Phone: +86 (0) 21 61056666 Copyright 2010 ABB
Fax: +86 (0) 21 61056677 All rights reserved
ABB S.p.A.
Hastelloy C-276 is a Cabot Corporation
Via Statale 113 trademark
22016 Lenno (CO), Italy Hastelloy C-2000 is an Haynes International
Phone: +39 0344 58111 trademark
Monel is an International Nickel Co. trademark
Fax: +39 0344 56278
Viton is a Dupont de Nemour trademark
DC200 is a Dow Corning Corporation trademark
www.abb.com/pressure DC704 is a Dow Corning Corporation trademark
Galden is a Montefluos trademark
Halocarbon is a Halocarbon Products Co.
trademark
Baysilone is a GE Bayer trademark
Neobee M-20 is a Stepan Company trademark
Esso Marcol 122 is a Esso Italiana trademark
Syltherm is a Dow Chemical Company trademark

Potrebbero piacerti anche