General GC1000
Specifications Analyzer Bus System

GS 11B06A01-01E

n GENERAL Other functions include data storage, parameter


Process analyzers are typically maintained by uploading/downloading, and network monitoring,
maintenance personnel from the front of the analyzer. with regard to the analyzers connected.
However, with the implementation of an analyzer Reduced Wiring Costs
bus system, maintenance can now be done entirely The analyzer bus system eliminates the need for
by monitoring and operating the analyzers from wiring between the individual field analyzers and the
a remote location. A personal computer can be associated devices, such as a DCS or other host
used as a terminal for controlling the analyzers, computers, and a PC for analyzer maintenance.
enabling the maintenance person to operate the Ethernet allows hubs and other general networking
gas chromatograph and field analyzers from a equipment to be used.
comprehensive computer screen interface instead of Network Scalability
the conventional gas chromatograph processor panel Ethernet facilitates connection to your network,
and chromatographic recorder. thereby enabling remote monitoring. (To ensure the
The analyzer bus system offers the following benefits: safety of your network, firewalls or other network
Comprehensive and centrally located remote security measures should be taken as appropriate.)
maintenance of analyzers. High-Speed Network
Reduction in the cost of wiring (among field analyzer Fast Ethernet supports data transfer rates of 100
to control room). Mbps, providing real-time monitoring of not only
Ease of expansion that is not limited by the physical measured values and alarm information but also
number of available DCS process inputs. chromatograms.
Optical Communication
n FEATURES The analyzer is available with either shielded twisted
Easy Maintenance of Analyzers pair wire or optical fiber. The optical fiber can be
Status of multiple analyzers can be monitored on a advantageously used in long distance networks and
PC connected to the network. Data are continuously electrically noisy environments.
stored in a server, allowing traceability of information Redundant System for Increased Reliability
necessary for maintenance. A redundant system is available for connection of
Stored data include chromatograms of detector analyzers with host computers such as a server and
signals as well as measured values and alarms. DCS.
Even after observation of an alarm or a change in
measured value, chromatograms can be checked
retrospectively.

Operation and display of Network monitoring,


Gas Chromatogram Automatic data storage Gateway [ ]:Software
PC PC STARDOM FCJ
[ASET] [PCAS] [ASGW]
RS-232C
DCS
DI/O

LAN (Ethernet)

ASIU

GC1000 MarkII GC1000 MarkII F01.ai

Figure 1 Typical System Configuration

Yokogawa Electric Corporation GS 11B06A01-01E


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Copyright Jun. 2006
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 2nd Edition Feb. 09, 2010
<<Contents>> <<Index>> 
(A) Ethernet 2. Devices
2.1 Process Gas Chromatograph
n SYSTEM CONFIGURATION GC1000 Mark II which connects to Ethernet network
A typical network configuration of analyzer bus is needs to be equipped to Ethernet communication
shown in Figure 1. ports.
The network consists of process gas chromatograph Instrument specifications:
(GC), server, hub, etc. Analyzer type and style:
GC1000 Mark II Style 6.01 or higher
(1) Process Gas Chromatograph (GC1000 Mark II) Connection type:
Analyzers that is installed in the field and connected Ethernet communication ports of
to Ethernet network. GC1000 Mark II should be specified.
(2) Analyzer Server Twisted pair: RJ-45 1 or 2 ports
Controls network and stores the data. It is composed Optical fiber: SC 1 or 2 ports
of PC and application software. PC is connected to Number of GC1000 Mark II connected to a network:
analyzers with Ethernet network. Max. 240 units
Number of analyzer servers one GC1000 Mark II
(3) Engineering Terminal
can access: 2 servers
A human machine interface for analyzer control and
Functions: Ethernet connection allows commands,
data view. Its software runs on a PC connected to
status detection and data acquisition from
network.
analyzer server.
Two types of software are available for engineering
terminal: Analyzer Server Engineering Terminal For other functions, refer to General Specifications
(ASET) that can handle multiple analyzers via of the GC1000 Mark II, GS 11B03A03-01E.
analyzer network and GC Engineering Terminal 2.2 Analyzer Server
(GCET) that provides one-to-one communication Realized by analyzer server software (PCAS)
with analyzer. installed on PC.
(4) Gateway Unit PC Specifications
Exchanges data with the host systems. It is Type: IBM PC compatible
necessary for redundant network or address OS Microsoft Windows Microsoft Windows XP
mapping. (32-bit Vista Business Edition Professional with SP3
(5) Interface Unit versions
Interfaces field analyzers other than GC with only)
network. CPU 32 bit-1 GHz or more Pentium II 350 MHz or more
Equipped with analog inputs and contact Memory 1 GB or more 128 MB or more
inputs/outputs. Hard disk 10 GB or more of free space
capacity For data strorage 3.5 GB per 1 unit GC1000 Mark II
1.9 GB per 1 unit of ASIU
n SPECIFICATIONS Display SVGA (1024 x 768)
1. Network (Ethernet) Etehrnet 1 port (100BaseTX or 100BaseFX)
Type: IEEE802.3u 100Base-TX, 100Base-FX port 2 port (100BaseTX or 100BaseFX) when dual
Protocol: TCP/IP redundant configuration
Communication rate: 100 Mbps Oyher CD-ROM drive
Cabling: Shielded twisted pair wire or optical fiber
Max. number of nodes: 254 PC analyzer Server software (PCAS) specifications
Max. distance: Analyzer bus connection:
100 m (100Base-TX) Number of connected analyzers and analyzer bus
2 km (100Base-FX) interface units (FCN): Max. 64 sets
Total distance can be extended using Number of PCAS in one network: Max. 14 sets
multiple levels of switching hubs* in Automatic data storage:
cascade connection. Data of GC1000 Mark II and analyzer
*Refer to the switching hub. bus interface unit are stored on the
Redundant Network: server PC.
Two communication paths are provided. Storage Data Description Remarks
Both are equal and which is used is Analysis results Detailed analysis results are Storage
determined by device engaged in the stored for one year. Based capacity
communication. on this, historical data of depends on
GC1000 Mark II should have dual channel concentration and retention hardware and
Ethernet outputs and server PC has two time are made. settings.
Ethernet ports. Analyzer server, gateway Chromatogram Chromatograms for every ditto
unit, GC1000 Mark II and interface unit history (1) cycle for the past 2 months.
can be connected to a redundant network, Chromatogram Every 10 times of ditto
but engineering terminals cannot. history (2) chromatograms for one year.
Alarm history 100 kbyte
Event history 100 kbyte

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Dec. 15, 2010-02
<<Contents>> <<Index>> 
Stored data are read with analyzer bus 2.4 GC Engineering Terminal
engineering terminal software (ASET). Realized by GC engineering terminal software
Network monitoring screen: (GCET) installed on PC.
Analyzer server communication status is PC Specifications
monitored for troubleshooting purpose. Models supported:
2.3 Analyzer Server Engineering Terminal IBM PC compatible machine that caan
Realized by engineering terminal software (ASET) run Microsoft Windows and for which
installed on PC. Ethernet is available.
PC Specifications OS Microsoft Windows Microsoft Windows XP
Models supported: (32-bit Vista Business Professional with SP3
IBM PC compatible machine that caan versions Edition
run Microsoft Windows and for which only)
Ethernet is available. CPU 32 bit-1 GHz or more Pentium II 350 MHz or more
OS Microsoft Windows Microsoft Windows XP Memory 1 GB or more 125 MB or more
(32-bit Vista Business Edition Professional with SP3 Hard disk * At least 20 GB (for the program section only)
versions * For the hard disk, a data strage capacity should be
only) secured separately according to your PC system format,
CPU 32 bit-1 GHz or more Pentium II 350 MHz or more in addition to the capacity for the program.
Memory 1 GB or more 125 MB or more Communications function:
Hard At least 20 GB (for the program section only) Analyzer and your PC must also be
disk * connected via Ethernet communication.
* For the hard disk, a data strage capacity should be GC engineering terminal software (GCET) specifications
secured separately according to your PC system format, Function: Display of the status of GC1000 Mark
in addition to the capacity for the program. II
Communications function: Display of the results and alarms
Analyzers and analyzer server(s) stored on the same PC.
must be connected via Ethernet Operation of GC1000 Mark II
communication. The analyzer server GCET communicates with one
and your PC must also be connected via GC1000 Mark II at once.
Ethernet communication. GCET is not activated with ASET on
Analyzer server engineering terminal software the same PC.
(ASET) specifications Screen display: a) Analyzer Operation
Function: Display of the status of GC1000 Mark b) Analysis Result
II, Analyzer bus interface unit (FCN) c) Chromatogram
and network. d) Alarm Status
Display of the results and alarms e) LCD Emulator (EtherLCD)
stored on the servers (PCAS). 2.5 Analyzer Bus Gateway Unit
Operation of GC1000 Mark II and Established by Yokogawas FCJ Autonomous
analyzer bus interface unit. Controller and analyzer server gateway unit software
Screen display: a) Overview (ASGW).
b) Analyzer Operation Interfaces with DCS or other host system as a
c) Analysis Result Modbus RTU slave.
d) Chromatogram
Instrument specifications
e) Alarm Status
Analyzer bus connection:
f) LCD Emulator (EtherLCD)
Max. number of connected GC1000 Mark II and
Max. number of EtherLCD is 4.
interface units: 31 sets
Analyzer server connection:
DCS connection
Max. number of ASET connected to one
Serial connection:
PCAS: 4 sets
Modbus RTU slave, 2 ports available
ASET is connectable to PCAS on the
Hardware specifications
same PC.
Serial communication:
One ASET is activated on one PC.
Function: transmit and receive data.
ASET is not activated with GCET on the
Type: RS-232C
same PC.
Number of port: 2 ports
Protocol: Modbus
Transmission rate: 300, 1200, 2400, 4800, 9600,
14400, 19200, 28800, 38400, 57600,
115200 bps
Data length: 8 bits
Parity bit: selectable
Stop bit: 1 bit
Compliant standards
EMC standards: CE Mark
Standard for hazardous location equipment:
Non-Incendive
Refer to the specifications of FCJ.

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Dec. 15, 2010-02
<<Contents>> <<Index>> 
Hardware IOP occurs when value exceeds 106.3 % of
Model Name Item Remarks input range or is below -6.3 %. If IOP occurs, the
NFJT100-S10x FCJ previous value will be held as AI data.
(c) Data of
Software license GC1000 Mark II
License Code Item Remarks - Stream number
NT711AJ-LM05E FCN/FCJ basic Required for each - First peak number
software for single ASGW - Number of peak
CPU including JAVA - Sampling time
NT8035J-LW11A Modbus Required for each - Analysis result
Portfolio ASGW - Retention time
- Calibration coefficient
Software media Interface unit
CR-ROM Code Item Remarks - AI read data
NT203AJ-PC11E Software media Loading a license for (Actual number in 2 words: Range 0.0-1.0)
installation When multiple requests are received in update
Modification of IP period, the last request is executed.
address
2.6 Analyzer Bus Interface Unit
Analyzer server gateway software (ASGW): Realized by Yokogawa's FCN Autonomous
Data mapping of GC1000 Mark II, Controller and analyzer server interface unit
interface unit and host system such as software (ASIU).
DCS for interface with host system. Function: Read and Write I/O interface data every
Capable of commanding, detecting 200 ms.
status, and reading data as follows. Compliant standards
(a) Commands for EMC standards: CE Mark
GC1000 Mark II: Standard for hazardous location equipment:
Non-Incendive
- Request of collective analyzer clock setting (all
Refer to the specifications of FCN.
connected analyzers)
- Request of individual analyzer clock setting Hardware specifications
(specified analyzers) Max. number of contact inputs: 16
GC1000 Mark II through mapping: Max. number of contact outputs: 16
- Run command Max. number of analog inputs: 16
- Stop command Specification Item Remarks
- Stream sequence assign NFBU200-Sxx Back board
- Calibration (validation) command NFCP100-S0x CPU
- Stream (continuous) assign
NFPW441-1x Power card 100-120 V AC
Interface unit through mapping
NFPW442-1x 220-240 V AC
- DO on command
- DO off command NFPW444-1x
(b) Status detection of NFDV151-P10/B5S00 DI card 32 points, 24 V DC
GC1000 Mark II and interface unit in common NFDV551-P10/D5S00 DO card 32 points, 24 V DC
- In operation NFDR541-P00/C4S70 Relay output 16 points, 24-110 V
- Communication error card DC/100-240 V AC
- Write error NFAI135 AI card 4-20 mA, 8 points
GC1000 Mark II NFAI143 AI card 4-20 mA, 16 points
- Analyzer normality/failure NFFCV01 Dummy cover for I/O card
- Change of analyzer alarm status NFFCV00 Dummy cover for power card
- Measurement, stop, or maintenance status
- Progress of stream sequence Software license
- Rejection of request of stream (continuous) License Code Item Remarks
- Rejection of request of calibration/validation NT711AJ-LS05E FCN/FCJ basic Required for each
- Data update software for single ASIU
- Calibration coefficient update CPU excluding JAVA
- Data validity NT8035J-LW11A Modbus Required for each
- Each alarm condition Portfolio ASIU
Concentration alarm of each peak, retention
time alarm, variation coefficient alarm, tailing Software media
coefficient alarm CD-ROM Code Item Remarks
Interface unit
NT203AJ-PC11E Software Loading a license for
- Unit normality/failure media installation
- Main power supply failure, 24 V DC failure, IO Modification of IP
module failure address
- DI data
- DO data
- AI IOP

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Dec. 15, 2010-02
<<Contents>> <<Index>> 
Analyzer server interface unit software (ASIU) 2.8 Analyzer Bus Network
Capable of commanding, detecting status, and (a) Redundancy
reading data as follows. In a redundant network, if one of the paths fails,
(a) Command communication line of analyzed data information is
- DO on/off command switched to the other automatically, minimizing the
(b) Status detection influence.
- Unit normality/failure Redundant network requires analyzer bus gateway
- Main power supply failure, 24 V AC failure, IO unit (FCJ) and its software (ASGW).
module failure The following licenses and media are required to
- DI data duplicate server PC.
- DO data License Code
- AI IOP
Licence Code Item Remarks
IOP occurs when value exceeds 106.3 % of
input range or is below -6.3 %. If IOP occurs, the NT783AJ-LW11A Redundant network Required for each
previous value will be held as AI data. function license PC
for FCN/FCJ OPC
(c) Data Server
- AI read data (media:NT203AJ)
(Actual number in 2 words: Range 0.0-1.0)
2.7 Network Components Software media
CD-ROM Code Item Remarks
(a) Hub
Switching type 100Base hub is recommended. NT203AJ-PC11E media Available for ASGW
or ASIU.
example; MOXA made P/N:EDS-308-MM-SC
(multi mode, wave length 1310 nm Communication lines of server PC, gateway unit
100Base-FX (SC connector) 2 ports, (FCJ), GC1000 MarkII and interface unit (FCN) are
TP (RJ45) 6 ports) duplicated.
(Area classification: class 1 Div 2/Zone 2)
Communication line of engineering terminal PC
(b) Cable (ASET or GCET) is not duplicated.
Optical Fiber Cable
It is recommended to use optical fiber cable when Engineering Analyzer Server
extending distance, wiring between separate Terminal Gateway
buildings or using the device in electrically noisy Unit
environments.
Optical fiber cable does not require an Electric
energy blocking-off circuit for explosion proof.
The type of optical fiber is the multi mode which core
diameter is 50/125 mmor 62.5/125 mm.
GC1000 Mark II has port(s) with 100Base-FX SC
connector.
Twisted Pair Cable
GC1000 Mark II has port(s) with RJ-45 connector for
twisted pair output. Interface
GC1000 MarkII GC1000 MarkII Unit
Use Category 5 or higher of ScTP (Screened twisted F02.ai

pair) or STP cable. CE Mark is declared on the Figure 2


condition with ScTP or STP cable.
Twisted pair cable requires an Electric energy (b) Restrictions on Total Distances of Network
blocking-off circuit for explosion proof. Refer to According to the 100Base-TX standards, cables of
section 2.11, Explosion-proof. up to 100 meters can be used. Total distance can be
extended using multiple levels of switching hubs* in
There are two types of pair cables: straight-through cascade connections.
or cross-over. Generally, a straight-through cable
is used to connect between a node and a hub, a * Refer to the switching hub.
cross-over cable is used to directly connect between 2.9 Modbus Communication
nodes. Realized by the following methods.
(c) Media Converter (a) Connection with analyzer bus gateway unit
Since gateway interface unit (FCJ) and general PC (FCJ)
do not have optical interface. For connection to an Analyzer server gateway unit can interface as a
optical fiber network, media converter interface is Modbus RTU slave. Refer to section 2.5, Analyzer
required. Bus Gateway Unit.
Select the appropriate devices according to the (b) Direct connection to GC1000 Mark II
conditions of infrastructures. Ethernet communication of GC1000 Mark II and
example; MOXA made P/N:IMC-101-M-SC analyzer server interface unit support Modbus/TCP
(multi mode, wave length 1310 nm client.
100Base-FX (SC connector) 1 ports,
TP (RJ45) 1 ports)
(Area classification: class 1 Div 2/Zone 2)

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
<<Contents>> <<Index>> 
2.10 OPC (B) ARCNET
OPC requires analyzer bus gateway unit (FCJ) and The New ARCNET-based bus system enables
OPC server software. connection between an ARCNET system and an
Refer to FCN/FCJ OPC Server for Windows (GS Ethernet system. This realizes addition of Ethernet
34P02Q61-01E). analyzers to existing ARCNET systems. YOKOGAWA
In addition, redundant network function license will propose the optimum systems based on your
for FCN/FCJ OPC Server software is required for existing analyzer bus system configurations.
connecting to the redundant network.
2.11 Explosion-proof n SYSTEM CONFIGURATION
The following conditions should be satisfied. A typical network configuration of analyzer bus is
(a) Process Gas Chromatograph (GC1000 Mark II) shown in Figure 4.
Refer to the General Specifications of GC1000 Mark The network consists of analyzers (Gas
II (GS11B03A03-01E) chromatograph etc.) and "Analyzer bus systems for
ARCNET". This "Analyzer bus systems for ARCNET"
(b) Twisted Pair Wiring consist of Gateway unit,analyzer server and
Twisted pair line should be disconnected when the ARCNET / Ethernet Converter.
explosion-proof status is not kept at the GC1000
The communication to the analyzer connected to
Mark II side.
Ethernet is enabled.
A 12 V DC output of the analyzer should be used to
Each software (PCAS, ASET) for Analyzer server
turn on the power line of the hub. (Figure 3)
and Analyzer Server Engineering Terminal should
Box be used ARCNET edition /ARC
The communication to the analyzer connected to
Hub Ethernet is enabled.
Power
Source - "Analyzer bus systems for ARCNET" should be
used so that DCS communicates with analyzer
To Network
Relay connected to ARCNET.
- Redundancy for ARCNET is enabled.
Redundancy for Ethernet can be done by using
Explosion-proof enclosure double "Analyzer bus systems for ARCNET".
- One "Analyzer bus systems for ARCNET" can
communicate with up to 30 sets of analyzers
GC1000 MarkII connected to ARCNET. In case of communicating
with more than 31 sets, plural number of
"Analyzer bus systems for ARCNET" should be
used.

F03.ai

Figure 3

(c) Optical Fiber


Optical fiber cable does not require an Electric
energy blocking-off circuit for explosion proof.
(d) Hub
Non-incendive certified hubs can be installed in
Class 1 Division 2 hazardous areas. For compliance
with the standard, the hubs must be installed in a
cabinet approved by local explosion-proof testing
organization.
Select the appropriate devices according to the
conditions of location and wiring.
(e) Analyzer Bus Interface Unit
FCN used as analyzer bus interface unit is
approved that the product meets the non-incendive
requirements of the FM Standards.
Refer to the General Specifications of FCN/FCJ (GS
34P02Q01E).
2.12 Security
PC should be protected against computer viruses.
Accesses from external network should be restricted
by fire-wall.

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
<<Contents>> <<Index>> 

Operation and display of Network monitoring,


Gas Chromatograph Automatic data storage Gateway [ ]:Software
PC PC STARDOM FCJ
[ASET/ARC] [PCAS/ARC] [ASGW]
RS-232C
DCS
DI/O

LAN (Ethernet)

Analyzer bus system


for ARCNET

ASIU ARCNET/Ethernet
(added) Converter

GC1000 MarkII
(added) ARCNET

GCIU GCIU

GC1000 MarkII GC1000 GC1000 Analyzer


pH analyzer,
Oxygen analyzer, etc. F06.ai

Figure 4 Typical System configuration


The former type of a typical network configuration is shown in Figure 5.
However, the order to this former type has been stopped since 2007.

Operation and display of


Gas Chromatograph [ ]:Software
PC
[ASMT]
Network monitoring, RS-232C
DCS
Automatic data storage
Gateway

LAN (Ethernet)
GCAS DI/O

ARCNET

GCIU GCIU

GC1000 MarkII GC1000 GC1000 Analyzer


pH analyzer,
Oxygen analyzer, etc. F07.ai

Figure 5 the former type of typical System configuration

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 10, 2010-01
<<Contents>> <<Index>> 
n SPECIFICATIONS 2. PC Analyzer Server Software (PCAS)
Optional
Devices Model Suffix Code
Code
Description

1. Process Gas Chromatograph PCAS Software package


GC1000 Mark II which connects to ARCNET network Function -A01 Standard
needs to be equipped to ARCNET communication Language E English
ports. J Japanese
2. Analyzer Server -N Always -N
This is realized by the analyzer server software N Always N
PCAS installed on PC. Additional /ARC ARCNET
A one unit of the analyzer server can communicate supported version
with up to 32 units ARCNET Analyzer.
3. Analyzer Server Engineering Terminal
3. Analyzer Server Engineering Terminal
This is realized by the software ASET installed on
PC. Software (ASET)
4. Analyzer Bus Gateway Unit Model Suffix Code
Optional
Description
This is established by Yokogawas FCJ Autonomous Code
controller and analyzer server gateway unit software ASET Software package
(ASGW). Function -A01 Standard
Interfaces with DCS or other host system as a Language E English
Modbus RTU slave. J Japanese
Maximum number of connections from one unit of -N Always -N
ASGW to GC and PCAS must be less than 31 N Always N
units.
Additional /ARC ARCNET
5. ARCNET/Ethernet Converter supported version
Direct purchased by customer.
Model: AECV01 made by VECTOR INFORTECH
PTE LTD 4. GC Engineering Terminal Software (GCET)
Specifications: Optional
482.6(W) 88.9(H) 220(D), 4.7(kg) Model Suffix Code Description
Code
The number of Port: ARCNET(2), Ethernet(1)
GCET Software package
Power: 100 to 264V AC
Routing: ARCNET/Ethernet converting Function -A01 Standard
Redundant for ARCNET is enabled. Language E English
J Japanese
n MODEL AND SUFFIX CODES -N Always -N
N Always N
1. Process Gas Chromatograph (GC1000)
Any of the following Suffix Codes should be
specified for Ethernet analyzer bus. 5. Analyzer Server Gateway software (ASGW)
Suffix Optional Optional
Model Description Model Suffix Code Description
Code Code Code
GC1000 Gas chromatograph ASGW Software package
(Note) Function -A01 Standard
Analyzer bus A Ethernet twisted pair Language E English
dual-channel J Japanese
B Ethernet optical fiber -N Always -N
dual-channel
C Ethernet twisted pair N Always N
single-channel
D Ethernet optical fiber
single-channel 6. Analyzer Server Interface Unit Software
(ASIU)
Note: For specifications other than GC1000 Mark
II analyzer buss, consult our salesperson or Optional
representative. Model Suffix Code Description
Code
ASIU Software package
Function -A01 Standard
Language E English
J Japanese
-N Always -N
N Always N

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
<<Contents>> <<Index>> 
n TERMINAL ARRANGEMENT
1. Process Gas Chromatograph GC1000 Mark II
1.1 Ethernet (Twisted Pair Wiring)

Ethernet (Twisted pair) Terminal B (M)

Hub terminal (L)

Ethernet (TP) Terminal A (M)

Terminal Number
- Power off signal (A) 3 7 Power off signal (B) -
+ Power off signal (A) 4 8 Power off signal (B) +
F04.ai

1.2 Ethernet (Optical Fiber Wiring)

Ethernet (Optical fiber) Terminal (N)


A (Lower)
B (Upper)

T T Transmit TX

R R Recieve RX

A B
(Lower) (Upper)
F05.ai

2. Analyzer Bus Gateway Unit


Please refer to the General Specifications, FCN/FCJ Autonomous Controller Functions GS 34P02Q01-01E.

3. Analyzer Bus Interface Unit


Please refer to the General Specifications, FCN/FCJ Autonomous Controller Functions GS 34P02Q01-01E.

All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
Subject to change without notice.