General GC1000
Specifications Analyzer Bus System
GS 11B06A01-01E
LAN (Ethernet)
ASIU
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Dec. 15, 2010-02
<<Contents>> <<Index>>
Stored data are read with analyzer bus 2.4 GC Engineering Terminal
engineering terminal software (ASET). Realized by GC engineering terminal software
Network monitoring screen: (GCET) installed on PC.
Analyzer server communication status is PC Specifications
monitored for troubleshooting purpose. Models supported:
2.3 Analyzer Server Engineering Terminal IBM PC compatible machine that caan
Realized by engineering terminal software (ASET) run Microsoft Windows and for which
installed on PC. Ethernet is available.
PC Specifications OS Microsoft Windows Microsoft Windows XP
Models supported: (32-bit Vista Business Professional with SP3
IBM PC compatible machine that caan versions Edition
run Microsoft Windows and for which only)
Ethernet is available. CPU 32 bit-1 GHz or more Pentium II 350 MHz or more
OS Microsoft Windows Microsoft Windows XP Memory 1 GB or more 125 MB or more
(32-bit Vista Business Edition Professional with SP3 Hard disk * At least 20 GB (for the program section only)
versions * For the hard disk, a data strage capacity should be
only) secured separately according to your PC system format,
CPU 32 bit-1 GHz or more Pentium II 350 MHz or more in addition to the capacity for the program.
Memory 1 GB or more 125 MB or more Communications function:
Hard At least 20 GB (for the program section only) Analyzer and your PC must also be
disk * connected via Ethernet communication.
* For the hard disk, a data strage capacity should be GC engineering terminal software (GCET) specifications
secured separately according to your PC system format, Function: Display of the status of GC1000 Mark
in addition to the capacity for the program. II
Communications function: Display of the results and alarms
Analyzers and analyzer server(s) stored on the same PC.
must be connected via Ethernet Operation of GC1000 Mark II
communication. The analyzer server GCET communicates with one
and your PC must also be connected via GC1000 Mark II at once.
Ethernet communication. GCET is not activated with ASET on
Analyzer server engineering terminal software the same PC.
(ASET) specifications Screen display: a) Analyzer Operation
Function: Display of the status of GC1000 Mark b) Analysis Result
II, Analyzer bus interface unit (FCN) c) Chromatogram
and network. d) Alarm Status
Display of the results and alarms e) LCD Emulator (EtherLCD)
stored on the servers (PCAS). 2.5 Analyzer Bus Gateway Unit
Operation of GC1000 Mark II and Established by Yokogawas FCJ Autonomous
analyzer bus interface unit. Controller and analyzer server gateway unit software
Screen display: a) Overview (ASGW).
b) Analyzer Operation Interfaces with DCS or other host system as a
c) Analysis Result Modbus RTU slave.
d) Chromatogram
Instrument specifications
e) Alarm Status
Analyzer bus connection:
f) LCD Emulator (EtherLCD)
Max. number of connected GC1000 Mark II and
Max. number of EtherLCD is 4.
interface units: 31 sets
Analyzer server connection:
DCS connection
Max. number of ASET connected to one
Serial connection:
PCAS: 4 sets
Modbus RTU slave, 2 ports available
ASET is connectable to PCAS on the
Hardware specifications
same PC.
Serial communication:
One ASET is activated on one PC.
Function: transmit and receive data.
ASET is not activated with GCET on the
Type: RS-232C
same PC.
Number of port: 2 ports
Protocol: Modbus
Transmission rate: 300, 1200, 2400, 4800, 9600,
14400, 19200, 28800, 38400, 57600,
115200 bps
Data length: 8 bits
Parity bit: selectable
Stop bit: 1 bit
Compliant standards
EMC standards: CE Mark
Standard for hazardous location equipment:
Non-Incendive
Refer to the specifications of FCJ.
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Dec. 15, 2010-02
<<Contents>> <<Index>>
Hardware IOP occurs when value exceeds 106.3 % of
Model Name Item Remarks input range or is below -6.3 %. If IOP occurs, the
NFJT100-S10x FCJ previous value will be held as AI data.
(c) Data of
Software license GC1000 Mark II
License Code Item Remarks - Stream number
NT711AJ-LM05E FCN/FCJ basic Required for each - First peak number
software for single ASGW - Number of peak
CPU including JAVA - Sampling time
NT8035J-LW11A Modbus Required for each - Analysis result
Portfolio ASGW - Retention time
- Calibration coefficient
Software media Interface unit
CR-ROM Code Item Remarks - AI read data
NT203AJ-PC11E Software media Loading a license for (Actual number in 2 words: Range 0.0-1.0)
installation When multiple requests are received in update
Modification of IP period, the last request is executed.
address
2.6 Analyzer Bus Interface Unit
Analyzer server gateway software (ASGW): Realized by Yokogawa's FCN Autonomous
Data mapping of GC1000 Mark II, Controller and analyzer server interface unit
interface unit and host system such as software (ASIU).
DCS for interface with host system. Function: Read and Write I/O interface data every
Capable of commanding, detecting 200 ms.
status, and reading data as follows. Compliant standards
(a) Commands for EMC standards: CE Mark
GC1000 Mark II: Standard for hazardous location equipment:
Non-Incendive
- Request of collective analyzer clock setting (all
Refer to the specifications of FCN.
connected analyzers)
- Request of individual analyzer clock setting Hardware specifications
(specified analyzers) Max. number of contact inputs: 16
GC1000 Mark II through mapping: Max. number of contact outputs: 16
- Run command Max. number of analog inputs: 16
- Stop command Specification Item Remarks
- Stream sequence assign NFBU200-Sxx Back board
- Calibration (validation) command NFCP100-S0x CPU
- Stream (continuous) assign
NFPW441-1x Power card 100-120 V AC
Interface unit through mapping
NFPW442-1x 220-240 V AC
- DO on command
- DO off command NFPW444-1x
(b) Status detection of NFDV151-P10/B5S00 DI card 32 points, 24 V DC
GC1000 Mark II and interface unit in common NFDV551-P10/D5S00 DO card 32 points, 24 V DC
- In operation NFDR541-P00/C4S70 Relay output 16 points, 24-110 V
- Communication error card DC/100-240 V AC
- Write error NFAI135 AI card 4-20 mA, 8 points
GC1000 Mark II NFAI143 AI card 4-20 mA, 16 points
- Analyzer normality/failure NFFCV01 Dummy cover for I/O card
- Change of analyzer alarm status NFFCV00 Dummy cover for power card
- Measurement, stop, or maintenance status
- Progress of stream sequence Software license
- Rejection of request of stream (continuous) License Code Item Remarks
- Rejection of request of calibration/validation NT711AJ-LS05E FCN/FCJ basic Required for each
- Data update software for single ASIU
- Calibration coefficient update CPU excluding JAVA
- Data validity NT8035J-LW11A Modbus Required for each
- Each alarm condition Portfolio ASIU
Concentration alarm of each peak, retention
time alarm, variation coefficient alarm, tailing Software media
coefficient alarm CD-ROM Code Item Remarks
Interface unit
NT203AJ-PC11E Software Loading a license for
- Unit normality/failure media installation
- Main power supply failure, 24 V DC failure, IO Modification of IP
module failure address
- DI data
- DO data
- AI IOP
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Dec. 15, 2010-02
<<Contents>> <<Index>>
Analyzer server interface unit software (ASIU) 2.8 Analyzer Bus Network
Capable of commanding, detecting status, and (a) Redundancy
reading data as follows. In a redundant network, if one of the paths fails,
(a) Command communication line of analyzed data information is
- DO on/off command switched to the other automatically, minimizing the
(b) Status detection influence.
- Unit normality/failure Redundant network requires analyzer bus gateway
- Main power supply failure, 24 V AC failure, IO unit (FCJ) and its software (ASGW).
module failure The following licenses and media are required to
- DI data duplicate server PC.
- DO data License Code
- AI IOP
Licence Code Item Remarks
IOP occurs when value exceeds 106.3 % of
input range or is below -6.3 %. If IOP occurs, the NT783AJ-LW11A Redundant network Required for each
previous value will be held as AI data. function license PC
for FCN/FCJ OPC
(c) Data Server
- AI read data (media:NT203AJ)
(Actual number in 2 words: Range 0.0-1.0)
2.7 Network Components Software media
CD-ROM Code Item Remarks
(a) Hub
Switching type 100Base hub is recommended. NT203AJ-PC11E media Available for ASGW
or ASIU.
example; MOXA made P/N:EDS-308-MM-SC
(multi mode, wave length 1310 nm Communication lines of server PC, gateway unit
100Base-FX (SC connector) 2 ports, (FCJ), GC1000 MarkII and interface unit (FCN) are
TP (RJ45) 6 ports) duplicated.
(Area classification: class 1 Div 2/Zone 2)
Communication line of engineering terminal PC
(b) Cable (ASET or GCET) is not duplicated.
Optical Fiber Cable
It is recommended to use optical fiber cable when Engineering Analyzer Server
extending distance, wiring between separate Terminal Gateway
buildings or using the device in electrically noisy Unit
environments.
Optical fiber cable does not require an Electric
energy blocking-off circuit for explosion proof.
The type of optical fiber is the multi mode which core
diameter is 50/125 mmor 62.5/125 mm.
GC1000 Mark II has port(s) with 100Base-FX SC
connector.
Twisted Pair Cable
GC1000 Mark II has port(s) with RJ-45 connector for
twisted pair output. Interface
GC1000 MarkII GC1000 MarkII Unit
Use Category 5 or higher of ScTP (Screened twisted F02.ai
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
<<Contents>> <<Index>>
2.10 OPC (B) ARCNET
OPC requires analyzer bus gateway unit (FCJ) and The New ARCNET-based bus system enables
OPC server software. connection between an ARCNET system and an
Refer to FCN/FCJ OPC Server for Windows (GS Ethernet system. This realizes addition of Ethernet
34P02Q61-01E). analyzers to existing ARCNET systems. YOKOGAWA
In addition, redundant network function license will propose the optimum systems based on your
for FCN/FCJ OPC Server software is required for existing analyzer bus system configurations.
connecting to the redundant network.
2.11 Explosion-proof n SYSTEM CONFIGURATION
The following conditions should be satisfied. A typical network configuration of analyzer bus is
(a) Process Gas Chromatograph (GC1000 Mark II) shown in Figure 4.
Refer to the General Specifications of GC1000 Mark The network consists of analyzers (Gas
II (GS11B03A03-01E) chromatograph etc.) and "Analyzer bus systems for
ARCNET". This "Analyzer bus systems for ARCNET"
(b) Twisted Pair Wiring consist of Gateway unit,analyzer server and
Twisted pair line should be disconnected when the ARCNET / Ethernet Converter.
explosion-proof status is not kept at the GC1000
The communication to the analyzer connected to
Mark II side.
Ethernet is enabled.
A 12 V DC output of the analyzer should be used to
Each software (PCAS, ASET) for Analyzer server
turn on the power line of the hub. (Figure 3)
and Analyzer Server Engineering Terminal should
Box be used ARCNET edition /ARC
The communication to the analyzer connected to
Hub Ethernet is enabled.
Power
Source - "Analyzer bus systems for ARCNET" should be
used so that DCS communicates with analyzer
To Network
Relay connected to ARCNET.
- Redundancy for ARCNET is enabled.
Redundancy for Ethernet can be done by using
Explosion-proof enclosure double "Analyzer bus systems for ARCNET".
- One "Analyzer bus systems for ARCNET" can
communicate with up to 30 sets of analyzers
GC1000 MarkII connected to ARCNET. In case of communicating
with more than 31 sets, plural number of
"Analyzer bus systems for ARCNET" should be
used.
F03.ai
Figure 3
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
<<Contents>> <<Index>>
LAN (Ethernet)
ASIU ARCNET/Ethernet
(added) Converter
GC1000 MarkII
(added) ARCNET
GCIU GCIU
LAN (Ethernet)
GCAS DI/O
ARCNET
GCIU GCIU
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 10, 2010-01
<<Contents>> <<Index>>
n SPECIFICATIONS 2. PC Analyzer Server Software (PCAS)
Optional
Devices Model Suffix Code
Code
Description
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
<<Contents>> <<Index>>
n TERMINAL ARRANGEMENT
1. Process Gas Chromatograph GC1000 Mark II
1.1 Ethernet (Twisted Pair Wiring)
Terminal Number
- Power off signal (A) 3 7 Power off signal (B) -
+ Power off signal (A) 4 8 Power off signal (B) +
F04.ai
T T Transmit TX
R R Recieve RX
A B
(Lower) (Upper)
F05.ai
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E Feb. 09, 2010-00
Subject to change without notice.