Sei sulla pagina 1di 5

Laboratory ABS Instruments Pvt. Ltd.

, Unit 21, Block 1, SIDCO Electronics Complex,


Thiru-Vi-Ka Industrial Estate, Guindy, Chennai, Tamil Nadu

Accreditation Standard ISO/IEC 17025:2005

Discipline Mechanical Calibration Issue Date 24.02.2015

Certificate Number C-0332 Valid Until 23.02.2017

Last Amended on - Page 1 of 5

Quantity Measured / Range/ Frequency * Calibration Measurement Remarks


Instrument Capability ()

I. FORCE

1. UNIAXIAL MATERIAL TESTING MACHINE

a. VERIFICATION OF
THE FORCE
MEASURING SYSTEM
Tension 5 N to 1000 kN 0.23 % Using Force Proving
Compression 5 N to 3000 kN 0.23 % Dynamometers, Load cells
As per IS 1828 (Part 1) &
ISO 7500

b. VERIFICATION OF 0 to 500 mm 0.13 mm Using Liner Dimensional


DISPLACEMENT Instruments
SYSTEM USED IN THE as per ASTM E 2309-05
MATERIAL TESTING
MACHINE

c. VERIFICATION OF 0 to 1000 mm/min 0.098 mm/min Using Liner Length Measuring


TEST SPEED OF Instruments And Stop Watch
MATERIAL TESTING As Per ASTM E2658-11
MACHINES

II. HARDNESS
Vishal Shukla Avijit Das
Convenor Program Manager
Laboratory ABS Instruments Pvt. Ltd., Unit 21, Block 1, SIDCO Electronics Complex,
Thiru-Vi-Ka Industrial Estate, Guindy, Chennai, Tamil Nadu

Accreditation Standard ISO/IEC 17025:2005

Discipline Mechanical Calibration Issue Date 24.02.2015

Certificate Number C-0332 Valid Until 23.02.2017

Last Amended on - Page 2 of 5

Quantity Measured / Range/ Frequency * Calibration Measurement Remarks


Instrument Capability ()

1. INDIRECT HRA 0.43 HR Using Standard Test Blocks


VERIFICATION OF HRB 0.70 HR As per IS 1586 (Part 2)
ROCKWELL HRC 0.50 HR ISO 6508/ ASTM E18
HARDNESS TESTER
HR 15 N 0.61 HR Using Standard Test Blocks
HR 30 N 0.53 HR As per IS 1586 (Part 2)
HR 45 N 0.60 HR ISO 6508
HR 15 T 0.80 HR
HR 30 T 0.86 HR
HR 45T 0.86 HR

HR 15 Y 1.03 HR Using Standard Test Blocks


HR 30 Y 1.03 HR As per IS/ASTM E18

2. DIRECT Pre Load Using Force Proving Devices &


VERIFICATION OF 2.492 N & 4.930 N 0.5 % Load Cells as per IS 286 (Part 2) &
ROCKWELL Total Load ISO 6508
HARDNESS TESTER 147.1 N, 294.2 N,
TEST FORCE 441.3 N, 588.4 N,
980.7 N & 1471.0 N

3. INDIRECT Micro Vickers Using Standard Test Blocks


VERIFICATION OF HV0.3 & HV0.5 2.0 % As per IS 1754/ ISO 6507
VICKERS & MICRO ASTM E92
VICKERS HARDNESS
TESTER Vickers scales Using Standard Test Blocks
HV1, HV5, HV10, 2.0 % As per IS 1754/ ISO 6507
HV20, HV30 & HV50

4. DIRECT Test Forces Using Force Proving Devices &


VERIFICATION OF 0.49033 N to 490.33 N 0.5 % Load Cells

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory ABS Instruments Pvt. Ltd., Unit 21, Block 1, SIDCO Electronics Complex,
Thiru-Vi-Ka Industrial Estate, Guindy, Chennai, Tamil Nadu

Accreditation Standard ISO/IEC 17025:2005

Discipline Mechanical Calibration Issue Date 24.02.2015

Certificate Number C-0332 Valid Until 23.02.2017

Last Amended on - Page 3 of 5

Quantity Measured / Range/ Frequency * Calibration Measurement Remarks


Instrument Capability ()

VICKERS & MICRO as per IS 1754/ISO 6507


VICKERS HARDNESS
TESTER TEST FORCE

5. INDIRECT Brinell Scales Using Standard Test Blocks


VERIFICATION OF HBW 2.5/187.5 2.0 % as per IS 1500 (Part 2) &
BRINELL HARDNESS HBW 5/750 ISO 6506/ ASTM E10
TESTER HBW 10/3000

6. DIRECT Test force Using Force Proving Devices &


VERIFICATION OF 1.839 kN 0.5 % Load Cells
BRINELL HARDNESS 2.452 kN As per IS 1500 (Part 2) &
TESTER TEST 7.355 kN ISO 6506
FORCE 29.42 kN

7. DIRECT 0 to 7 mm 0.008 mm Using Stage micrometer


VERIFICATION OF As per IS 1500 (Part 2)
BRINELL HARDNESS
TESTER MEASURING
DEVICE

8. VERIFICATION AND Using Weighing Balance and


CALIBRATION OF Load Cells
IRHD HARDNESS

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory ABS Instruments Pvt. Ltd., Unit 21, Block 1, SIDCO Electronics Complex,
Thiru-Vi-Ka Industrial Estate, Guindy, Chennai, Tamil Nadu

Accreditation Standard ISO/IEC 17025:2005

Discipline Mechanical Calibration Issue Date 24.02.2015

Certificate Number C-0332 Valid Until 23.02.2017

Last Amended on - Page 4 of 5

Quantity Measured / Range/ Frequency * Calibration Measurement Remarks


Instrument Capability ()

TESTERS

a. IRHD Micro Minor Load 0.2 mN Using Weighing Balance and


Total Load 0.5 mN Load Cells
As per ASTM 1415-06,
b. IRHD Normal/Soft/Hard Minor Load 0.5 mN ISO 48-2003 &
Total Load 6.0 mN ISO 18898-06

c. Verification And 0 to 100 Shore 0.35 Shore Using Shore Calibrator, Weigh
Calibration Of Shore Balance and Load Cell
A/B/O/D/C/DO As per ASTM D2240-05
(Spring Force) ISO 18898-06

9. VERIFICATION OF 0 to 50 mm 0.008 mm Using Calibration Jig and Linear


EXTENSOMETER Displacement Device as per ASTM
0 to 500 mm 0.06 mm E83-10a, ISO 9513 &
IS 12872: 1990

10. VERIFICATION OF 0 to 450 J 0.73 % Using Clinometer, Load Cell,


CHARPY /IZOD Measuring Tape and other
IMPACT TESTING Measuring Instruments
MACHINE as per ISO 14:2008
ASTM E23-07, BS131-Part 4
1972 (Amendment 199)
ASTM D 256-05, D 256-05,
D 6110-4, ISO 13802;1999

11. VERIFICATION AND 0 to 100 Shore 0.27 Shore Using Shore Calibrator, Weigh
CALIBRATION OF Balance and Load Cell as per
SHORE A AND D$ ASTM D2240-05
ISO 18898-06

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory ABS Instruments Pvt. Ltd., Unit 21, Block 1, SIDCO Electronics Complex,
Thiru-Vi-Ka Industrial Estate, Guindy, Chennai, Tamil Nadu

Accreditation Standard ISO/IEC 17025:2005

Discipline Mechanical Calibration Issue Date 24.02.2015

Certificate Number C-0332 Valid Until 23.02.2017

Last Amended on - Page 5 of 5

Quantity Measured / Range/ Frequency * Calibration Measurement Remarks


Instrument Capability ()

12. VERIFICATION & 0 to 50 mm 0.008 mm Using Calibration Jig


CALIBRATION OF as perASTM E83-10a &
EXTENSOMETER$ IS 12872: 1990, ISO 9513

13. VERIFICATION AND 0 to 100 Shore 0.32 Shore Using Digital Displacement
CALIBRATION OF Measurement Device
SHORE A/B/O/D/C/DO Liner Instruments
(INDENTER TRAVEL
DISTANCE) $

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$
Only in Permanent Laboratory

Only for Site Calibration


#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual
environmental conditions and master equipment used.

Vishal Shukla Avijit Das


Convenor Program Manager

Potrebbero piacerti anche