Sei sulla pagina 1di 7

MAGNETOM

Maintenance Protocol
System
Preventive Maintenance Protocol

Customer:

Address:

Fax / E-Mail:
Fax / E-Mail:
Department:

Room:

Material-No.:

Serial-No.:

Contract-No.:

Expire date:

Order-No.:

System - ID:

Aera
Skyra
The instructions M7-010.831.05.01.XX are required for this protocol

Answers for life.

Protocol

Date:

Serial-No.:

Evaluating the Condition of the System


The system has no deficiencies. *
The system has slight deficiencies that have no effect on contin
ued operation of the system. The deficiencies should be corrected
preventively. *
The system has serious deficiencies. For safety reasons, continued
operation of the system is permitted only after successfully cor
recting the deficiencies.
*) For imaging systems: The result of the image quality check shows no deviation from
the reference values.
The evaluation was performed after completing all work steps.
Signature:
Date:

Name:

If required by country-specific regulations:


The customer or a representative has taken note of the result of the evaluation of the
system condition.
Signature:
Date:

Name:

Remarks:

MAGNETOM

M7-010.832.05.01.02
12.15

Page 2 of 7
CS SD MR

Siemens, 2015

Protocol

Date:

Serial-No.:

Remarks Regarding the Protocol


The chapter numbers refer to the chapters in the instruction, which is referenced on the
cover page.
All pages have to have the serial number of the system and the date of maintenance in
the page header.
The assignment n.a. (not applicable) indicates that the checkpoint or measured value is
not used for this system.
On page 2 the completeness and the results of the maintenance work is confirmed.

Explanation of Abbreviations in the Protocol


Abbrev. Explanation

Abbrev. Explanation

SI

Safety Inspection

PMF

Preventive Maintenance, Operating


Value Check, Function Check

SIE

Electrical Safety Inspection

System Quality, Image Quality

SIM

Mechanical Safety Inspec


tion

QIQ

Image Quality

PM

Preventive Maintenance

QSQ

System Quality Check

PMP

Periodic Preventive Mainte


nance

SW

Software Maintenance

PMA

Preventive Maintenance Ad
justments

CSE

Customer Service Engineer

Siemens, 2015

M7-010.832.05.01.02
12.15

Page 3 of 7
CS SD MR

MAGNETOM

Protocol

Date:

Serial-No.:

Measuring Equipment and Measuring Instruments Used


Measuring instruments and measuring devices (phantoms, MR coils, etc.) may not be en
tered in the table if they have already been entered in the mobile device.
Measuring equipment / -instruments

MAGNETOM

M7-010.832.05.01.02
12.15

Page 4 of 7
CS SD MR

Serial No.

Date used

Siemens, 2015

Protocol

Date:

Serial-No.:

5
OK

not
OK

n.a.

2 Maintenance Instructions
2.2

Computer MRC (MRAWP) - MRSC (MRWP)


PM

2.3
2.3.2

Cleaning the computer

TFT monitor
Check
PMA Checking the TFT monitor

2.4

Vacuum pump (comfort kit option)


PMP

Replacing the vacuum pump filter (every 2 years)

Startup Date:

Date of Last Replacement:

2.5

2.8.1

Checking the gradient filter fans

Cooling system - general checks

SEP (option)
PMP

2.8

Checking the water circuit of the SEP

Magnet system
Cold Head
PM
PM

Cold head is monitored (cold head replacement on demand)


Replacing the cold head (w/o cold head monitoring replace
ment every 24 months)

Startup Date:

Date of Last Replacement:


2.8.2

DD

MMM

YYYY

Replacing the adsorber (F70; every three years)

Startup Date:

Date of Last Replacement:


2.9.1

Adsorber
PM

2.9

YYYY

Cooling
PMP

2.7

MMM

Gradient filters
PM

2.6

DD

DD

.
MMM

YYYY

Patient table
Hydraulic system of the patient table
PM

Siemens, 2015

Checking the hydraulic system of the patient table

M7-010.832.05.01.02
12.15

Page 5 of 7
CS SD MR

MAGNETOM

Protocol

Date:

Serial-No.:
OK

2.9.2

2.10.1

Door

2.12.1

Cleaning up directories

2.13.1

Deleting the MR save log files

Site-specific data
SW

2.13

Quality assurance measurement

Software
SW

2.12.2

Checking the RF room door

Quality Assurance (QA)


PM

2.12

Checking the docking station (dockable table only)

RF room door
PMP

2.11

n.a.

Option dockable table: docking station


PM

2.10

not
OK

Saving site-specific data

Option: Combi Dockable Table


Option Combi Dockable Table: Hydraulic System
PM

2.13.2

Option Combi Dockable Table: Checking the hydraulic sys


tem
Option Combi Dockable Table: Docking Station
PM

2.14

Option Combi Dockable Table: Checking the docking station

End of Maintenance Instructions


Maintenance of this section was performed by:
Signature:
Date:

MAGNETOM

Name:

M7-010.832.05.01.02
12.15

Page 6 of 7
CS SD MR

Siemens, 2015

All documents may only be used for


rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re
served.
healthcare.siemens.com/services

Siemens Healthcare Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens.com/healthcare

Print No.: M7-010.832.05.01.02 | Replaces: n.a.


Doc. Gen. Date: 12.15 | Language: English
Siemens Healthcare GmbH, 2015

siemens.com/healthcare