INSTRUCTIONSFORUSE
BowforSurgicalandVascularProceduresWHA200
MODS.OPESCOPEACTIVOANDOPESCOPEPLENO
SHIMADZUCORPORATION
ImporterandDistributorinBrazil:
ShimadzudoBrasilComrcioLtda.
Av.MarqusdeSoVicente,n1771
SaoPauloSP
Authorization/MS:1036901

Manufacturer:
ShimadzuCorporation
1,NishinokyoKuwabaracho
NakagyoKu
Kyoto6048511Japan

InternationalDistributors:
ShimadzuCorporation
1,NishinokyoKuwabaracho
NakagyoKuKyoto6048511
Japan

ShimadzuCorporation
TokyoOffice,3
KandaNishikicho1chome
Chiyodaku,Tokyo1018448
Japan

ShimadzuMedicalSystems,
USACompany
20101SouthVermontAve.
Torrance,California90502
USA

ShimadzuEuropaGmbH
AlbertHahnStrasse610,
47269
Duisburg,FR
Germany

ShimadzuMedical(INDIA)
PVT.LTD.
1stfloor,1063,Munusamy
Salai
KKNagar,Chennai600078
India

ShimadzuMedicalSystems
(OCEANIA)PTY.LTD
UnitsE,1016SouthStreet
Rydalmere,NSW2116
Australia

Shimadzu(HongKong)LTD.
Suite1028OceanCenter
HarborCity,TsimShaTsui
Kowloon,HongKong

ShimadzuInternational
Trading(Shanghai)CO.LTD
BeijingOffice14F
ChinaLifeTower,No.16
ChaoyangmenwaiStreet
ChaoyangDistrict
Beijing100020PR
China

BeijingShimadzuMedical
EquipmentCO.LTD.
No.207LongShengIndustry
ZoneNo.7RongChangDong
street
BeijingEconomicTechnology,
Developmentarea
Beijing100176PR
China

Producttechnicalname:ARCOSURGICAL
TechnicalDirector:ConstantineLuizDiPIPICREA/SP0682526854
ANVISARegistrationNo:10369010062

https://translate.googleusercontent.com/translate_f

2/135

11/28/2016

OPERATIONMANUAL

Page3

ActiveOpescopeSetting:
XRayTube
XrayControllerUnit
Collimator
XrayGenerator
CarHolderwithCBow
ImageIntensifier
TVsystemwithCCDcamera
LCD/TFTmonitor(s)
Monitorstand(s)
Cassetteholder

FullOpescopeSetup
XRayTube
XrayControllerUnit
Collimator
XrayGenerator
CarHolderwithCBow
ImageIntensifier
TVsystemwithCCDcamera
LCD/TFTmonitors
MonitorStand
Cassetteholder9"

Optional
ClearprotectivecoverforCbow(disposablewith
50units/carton)
Transparentcoverformonotank(disposablewith
100units/carton)
TransparentprotectivecoverforImageEnhancer(disposable
With100units/box)

https://translate.googleusercontent.com/translate_f

3/135

11/28/2016

OPERATIONMANUAL

Page3

Sterilizablecoverforarcc,monotankand
images
Sterilizableprotectioncoverformonotankand
7"pictures
Sterilizableprotectioncoverformonotankand
9"pictures
Sterilizableprotectioncoverformonotankand
6pictures
IndicatorLaser(laserpointer)toII7"
IndicatorLaser(laserpointer)for9II"
IndicatorLaser(laserpointer)for6II"
Dualstagetrigger
FibergrillforII7"
FibergrillforII6"
FibergrillforII9"
ControlknobforIl
Cineloopmemory(ACTIVEonly)
DigitalImagingUnitwith
Digitalsubtraction(DSA)(ACTIVEonly).
MODrecorder(ACTIVEonly)
MODmedia(ACTIVEonly)
Videoprinter(thermalprinter)
DVDR/RWrecorder
LCD/TFTwithtouchscreen(touchscreen)
DICOMStorageKit(includedwithCDROMwithsoftware)
DICOMPrintKit(includingsoftwareCDROM)
DICOMWorklistKit(includedwithCDROMwithsoftwareandPC,
ConsistingofMonitor,CPU,MouseandKeyboard)
Fluoroimagedisplayfunctiononthearcscreen(Only
FULL)
XrayDoseMeterDAP(PLENARYOnly)
ExtensionofEmittingSourceSkinDistance.

https://translate.googleusercontent.com/translate_f

4/135

11/28/2016

OPERATIONMANUAL

Page5

StorethismanualcarefullyalongwiththeOpescope.

Thesymbolsshowninthismanualfollowbelow.

Alertthatindicatesdirectdangerandmaycausedeathorseriousinjuryif
Notbeavoided.

Alertindicatingapotentialhazardwhichcouldcausedeathorserious
Damageifnotavoided.

Hazardwarningthatcancauseslightdamagetothepatientordamagetothe
Equipmentorfireifnotavoided

Alertswithinformationthatcanhelpyouusethesystemcorrectly.

IndicationofReferences
ProvidesinformationtomakeiteasiertouseOpescope
ProvidesinformationforreferencewhenOpescopeisused.
AbbreviationsandSpecialTerms

ReferringtoFig.
https://translate.googleusercontent.com/translate_f

5/135

11/28/2016

OPERATIONMANUAL

Page6

Introduction

WethankyouforchoosingtheArchforSurgicalandVascularProcedures
OPESCOPEACTIVO/OPESCOPEPLENO.
Beforeusingtheproduct,readtheoperationmanualandmakesurethe
Understandingofitscontent.
Afterreadingthismanual,storeitinasafeplace
Accessedeasily.

Copyright
ShimadzuCorporationownsthecopyrightsinthismanualandpartof
Contentorallofitcannotbecopiedorreproducedwithoutdue
AuthorizationfromShimadzuCorporation.
Thecontentsofthismanualcanbemodifiedforimprovementswithoutany
Notice.

ShimadzuCorporationAllrightsreserved.

https://translate.googleusercontent.com/translate_f

6/135

11/28/2016

OPERATIONMANUAL

Page7

OperatingPrecautions

1.Onlyqualifiedpersonnelshouldoperatetheequipment.
2.Wheninstallingtheequipment,payattentiontothefollowingitems:
(2.1)Donotinstalltheequipmentnearwatertapsorsimilar.
(2.2)Installawayfrompotentialsourcesofpressure,temperatureor
Moisture,directsunlight,hydrochloricorsulfurdustorgases.
(2.3)Duringtransportandoperationoftheequipmentavoidvibrations,
Inclinations,shocksandimpacts.
(2.4)Keepequipmentawayfromareaswheregasesorchemicals
Arestored.
(2.5)Useonlytherecommendedpowersource(frequency,currentand
voltage).
(2.6)Checktheconditionsofthebattery'spowersource(powerand
Polarity)beforeoperatingtheequipment.
(2.7)Properlygroundtheequipment.
3.Beforeoperatingtheequipment,payattentiontothefollowingitems:
(3.1)Checktheconditionsofthesockets,polarity,etc.
(3.2)Checkthatgroundingisconnectedcorrectly.
(3.3)Checktheelectricalinstallation.
(3.4)Payattentionwhenusingmorethanoneelectricalequipment,
Maycauseanincorrectdiagnosisorcausedamage.
(3.5)Checktheconditionoftheexternalelectricalcircuit,whichwillbe
Connecteddirectlytothepatient.
(3.6)Checktheconditionofthebatterypowersource
4.Whileoperatingtheequipment,payattentiontothefollowingitems:
(4.1)Donotexceedexposuretimefordiagnosisortherapy.
(4.2)Observethepatientandtheequipmentconstantlyfordetection
Problems.
(4.3)Whenaproblemisdetectedwiththeequipmentorpatient,

https://translate.googleusercontent.com/translate_f

7/135

11/28/2016

OPERATIONMANUAL

Page8

Takethenecessarystepstostoptheequipmentwithout
Hurtthepatient.
(4.4)Donotlettheequipmenttouchthepatient.
5.Afteroperatingtheequipment,payattentiontothefollowingitems:
(5.1)Turnoffthekeysandreturnthedialtoitsoriginalposition.Then,turnoff
Themainkey.
(5.2)Donotpullthepowercordviolentlyfromtheoutlet.
(5.3)Whenstoringtheequipment,payattentiontothefollowing
items:
(I)Keepawayfromwater.
(Ii)Storeawayfrompotentialcausesofproblemssuchas
Abnormalpressure,temperatureandhumidity,airflow,sunlight
Direct,dust,chlorineorsulfurgas.
(Iii)Duringtransportationandstorageofequipment,avoid
Inclinations,vibrationsandimpacts.
(Iv)Storeequipmentawayfromstorageareas.
Chemicalsandgases.
(5.4)Cleanallaccessories,cables,andcontacts,andstoretheminalocation
Appropriate.
(5.5)Keepequipmentcleantoavoidproblemsduringoperation.
Nextuse.
6.Whenequipmentisbroken,donotattempttorepairit.
Properlyidentifyandcontacttheauthorizedservicecenter.
7.Donotmodifyanypartoftheequipment.
8.PreventiveMaintenance
(8.1)Theequipmentanditspartsshallbeperiodicallytested.
(8.2)Iftheequipmentisnotinoperationforanextendedperiod,
Testbeforeusetoensure
Evenifitisworkingproperlyandsafely.
9.Forotheritems,usethemaccordingtotheoperatinginstructions.

ReferringtoFig.

Page9
https://translate.googleusercontent.com/translate_f

8/135

11/28/2016

OPERATIONMANUAL

Usingthesystemsafely...

Onlyqualifiedpersonnelshouldusethesystem.
Theresponsibilitytocontroltheuseandmaintenanceofmedicalequipment
Fallsontheuser(s)ofthesystem.
Onlyqualifiedpersonnelortheirsubordinatesmayusethesystem.
Itisdangeroustocarryoutinternalrepairsorinspections,
Ifrepairs/inspectionsarenecessary.

Donotmodifythesystem.
Ingeneral,allmodificationsarestrictlyprohibitedbytherequirements
Regulatoryframeworks.Pleasecontactyourauthorizedservice
Anymodification.

Performperiodicinspections.
Tomaintainsystemsecurityandperformanceforalongtime,
Performpreventivemaintenance.

Repairsandmaintenanceoftheequipmentcanonlybeperformed
Technicalassistanceorauthorizedtechnicians

ReferringtoFig.

Page10

https://translate.googleusercontent.com/translate_f

9/135

11/28/2016

OPERATIONMANUAL

MainPointstoWatch

Beforeusingtheproduct,readtheoperationmanualandmakesurethe
Understandingofitscontent.
Theuseofthesystemwithoutreadingthismanual,orwitha
Incompleteunderstandingoftheinformationofitscontents,cancauseserious
Damagetotheequipment.

Donotturnonthepowerortouchthebuttonsorothercontrolswithoutfirst
Havereadtheinstructionmanual.
IfyoudonotunderstandanypartoftheInstructionManual,please
Contactyourauthorizedservicecenter.Contactdetailswillbe
Theendofthemanual.
Storeitinasafeplacethatcaneasilybeaccessedwhen
Necessary.
Familiarizeyourselfwiththelocationsofrisksandthenatureoftheserisks.
Makesureeverythinghasbeendonetoavoidsuchrisksandactions.
Correctivemeasurestobetakenintheeventofsomeofthese
Happen.

Onlyqualifiedpersonnelshouldusethesystem.
Theresponsibilitytocontroltheuseandmaintenanceofmedicalequipment
Fallsontheuser(s)ofthesystem.
Onlyqualifiedpersonnelortheirsubordinatesmayusethesystem.
Itisdangeroustocarryoutinternalrepairsorinspections,
Ifrepairs/inspectionsarenecessary.

Donotmodifythesystem.
Ingeneral,allmodificationsarestrictlyprohibitedbytherequirements
Regulatoryframeworks.Pleasecontactyourauthorizedservice
Anymodification.

10

Page11

https://translate.googleusercontent.com/translate_f

10/135

11/28/2016

OPERATIONMANUAL

LocationofRisks

HighVoltageSection

Nevertouchtheinternalpartsoftheequipment,asitcontains
Highvoltagesectors.
Thiscouldresultinseriousaccidentsincludingdeathby
electrocution.

Figure1HighVoltageSection

Inspectthecableusedinthesystem.
Checktheinsulationofthecableoranydamagebeforeusingit.
inthesystem.
Insulationfaultscancauseseriousaccidentsincludingdeath
byelectrocution

11

Page12

https://translate.googleusercontent.com/translate_f

11/135

11/28/2016

OPERATIONMANUAL

MovingSection

Beawareofhazardswhentouchingequipment
Youcaninjureyourarmsorlegsbypositioningthembetween
Carmandtherestofthestructure.
WhenmovingtheCarm,grasponlyattheappropriatepointsand
Operatecarefully.

Donotplaceanyperipheraldevicesoranyother
Intherangeofmovementoftheequipment
Thelaserbeamgeneratedhasawavelengthof630nm
(Nanometers)andtheoutputislessthanmWandconsequentlyis
CharacterizedasaClassIIlaserbytheFDAandasalaser
Class3AbytheIEC.

Figure2RangeofMotion

12

Page13

https://translate.googleusercontent.com/translate_f

12/135

11/28/2016

OPERATIONMANUAL

PreventiveMaintenanceNotes

PerformPeriodicInspections.
Preventivemaintenancewillberequiredtomaintainsafety
Andsystemperformance.
Maintenanceshouldonlybecarriedoutbytechniciansor
Authorizedtechnicalassistance.

Repairsandmaintenanceshouldonlybeperformedby
Techniciansand/orauthorizedtechnicalassistance.

RadiationCare

Beawareofxrayirradiation.
Iftheequipmentisnotusedproperlybytheuser,the
Patientandothersmayreceiveahigherdoseof
Radiationthannecessary.Nooneelse,exceptthepatientandthe
Personsstrictlynecessaryfortheprocedure(staff
Doctor)shouldremainintheexaminationroomduring
Exposureofxrays.Ifforsomereasonanotherperson
Needtostayinsidetheexaminationroom,itshouldtakethe
Measurestoprotectagainstradiation.

Tominimizetheabsorbeddose,focusthe
Asfaraspossible.

13

Page14

Careofthelaserbeam
https://translate.googleusercontent.com/translate_f

13/135

11/28/2016

OPERATIONMANUAL

Thelaserbeamgeneratedhasawavelengthof630nm
(Nanometers)andtheoutputislessthanmWandconsequentlyis
CharacterizedasaClassIIlaserbytheFDAandasalaser
Class3AbytheIEC.

Neverlookdirectlyatthelaserbeam.
Whenlookingatthelaserbeam,theeyescanbeinjured.
Turnontheequipmentonlywhennecessarytoavoid
Unnecessaryexposure.
Donotattemptanyoperation,adjustmentorprocedurethatdoesnot
Describedinthismanual.

Explosioncare

Equipmentnotsuitableforuseinthepresenceofa
Inflammableanestheticmixturewithair,oxygenor
nitrous.

Takecarewhenusingdisinfectantspray.
Donotusespraydisinfectantsthatareflammableastheymay
Causeexplosions.
Vaporfromadisinfectantspraymaytriggeranexplosion
Causingdeaththreatstothesurroundingpeopleanddamageto
system.

14

Page15

FlameCare
https://translate.googleusercontent.com/translate_f

14/135

11/28/2016

OPERATIONMANUAL

Implementmeasurestoeliminateflames.
Usingthesysteminanunsuitableenvironmentmaycause
Fireorexplosion.
Installappropriatefireextinguishersandensurethatthe
Procedurefordealingwithfiresiswidelyknown.
Usefireextinguishersuitableforfire
Chemicalandelectrical.
Usingwaterorotherliquidsinfirefighting
Cancauseseriousdamage.

Cellphonerelatedcare

Keepcellphonesorthelikeoutoftheroom.
exam.
Donotplacedevicesthatemitelectromagneticwaves(eg
Cellphones)withintheexaminationroom,evenif
Off.Suchdevicesmayexceedthewave
Electromagneticfields,andundercertainconditionsmay
Functioningofthesystem.Intheworstcase,itcancauseserious
Damagesorclinicalerrors.

15

Page16

Careaboutfieldcompatibility
electromagnetic
https://translate.googleusercontent.com/translate_f

15/135

11/28/2016

OPERATIONMANUAL

TheWHA200requiresspecialprecautionsregardingEMC
(ElectromagneticCompatibility)andneedstobeinstalledand
UsedinaccordancewiththeEMCinformationprovidedinthis
Operatingmanual.

Checkthattheelectromagneticfieldsarecompatible.
Allperipheraldevicesmustconformtothe
EMCstandardsfortheemissionofelectromagneticwavesand
Sensitivitytoemitelectromagneticwaves.Devicesthat
Donotmeetthesestandardsmaycausethe
Thesystem.Intheworstcase,itcouldcauseseriousdamageor
Errors.

TheWHA200shouldnotbeusedtooclosetoorstackedwith
Otherequipment,butifuseinthiswayisnecessary,the
WHA200shouldbeconstantlymonitoredtoverifythe
Normaloperationintheconfigurationwhichwillbeused.

16

Page17

Cleaninganddisinfectioncare

https://translate.googleusercontent.com/translate_f

16/135

11/28/2016

OPERATIONMANUAL

Cleaninganddisinfectionofthesystem.
Cleaninganddisinfectionareveryimportanttoensurethat
Thesystemwillbecleanandsafetouse.

Disconnectthemainpowercablebeforestartingthe
cleaning.
Itisstrictlynecessarytoswitchtheunitoffbeforestartingthe
Cleaninganddisinfecting.
Otherwise,thereisariskofelectricshock.

Donotusewaterinthesystem.
Cleanthesystembywipingwithaclothmoistenedwithdisinfectant.
Usingwatertocleanthesystemmaycause
Electricalinstallations.

Donotapplydisinfectantorspraydirectlyonthe
equipment.
Ifdisinfectantentersthesystem,itmaycauseproblemsor
Accidents.

17

Page18

Cautiononuseandstorage

https://translate.googleusercontent.com/translate_f

17/135

11/28/2016

OPERATIONMANUAL

Installasatisfactoryground.
Thesystemrequiresgrounding.
Unsatisfactorygroundingmaycauseelectricshock.

Usetheequipmentinanenvironmentthatmeetsthe
Guidelinesaddressedhere.
Usetheequipmentonlyinanareaorroomthatis
Accordancewithgovernmentalandelectricallysafelawsforthis
kindofequipment.

Conditionsforproperuseandstorage.
Thelocationwheretheequipmentwillbeusedandstoredshall
Followingconditions:
Atmosphere:Cannotbeexplosiveorcontaincorrosivegases
Temperature:10Cto40C
Relativehumidity:30%to80%(noncondensing)
Atmosphericpressure:700hPaat1060hPa
Note:theenvironmentshouldnotundergosudden
Humidityortemperature,asthismaycausesystemproblems.

18

Page19

Caringforthesystem

https://translate.googleusercontent.com/translate_f

18/135

11/28/2016

OPERATIONMANUAL

Becarefulofimpactsorsystemcrashes.
Thesystemweighsmorethan200kg.
Impactsandfallswilldamagethesystemandcauseseriousdamageto
Peoplewhoaremanipulatingit.
Whenmovingthesystem,payattentiontothefollowingpoints:
CheckthattheCarmisinthetransportposition.
Whenyouarecrossingrampsorentrances,and
Movethesystemassmoothlyaspossible.Becareful
Particularlyonsteepramps.
Alwaysusethebrakewhenstoppingthesystem.
Onrampswithaslopegreaterthan5,thesystemmayfall.
Donotmoveoroperatetheequipmentinlocationswherethe
Exceed5.
Whenmovingthesysteminplacesthatgeneratestrongvibrations,
Outdoors,avoidrapidprogress,stop
Abrupt,andsteepramps.Donotstoptheequipmentonroutes
Sloping.
Whilemovingthesystem,becareful
obstacles.

19

Page20

WarningLabels

Donotremoveordamagethewarninglabels.
Warninglabelsindicateimportantcarewhenusing
thesystem.
https://translate.googleusercontent.com/translate_f

19/135

11/28/2016

OPERATIONMANUAL

Ifthelabelisunreadablecontacttheauthorizedservice.

Figure3WarningLabels

20

Page21

CAUTION

Fordriving,turntheCArmtothe
Position.

https://translate.googleusercontent.com/translate_f

20/135

11/28/2016

OPERATIONMANUAL

CAUTION
CRUSHING
Donotputyourhandhere.
WARNING
ThisXrayunitcanbedangerous
Forthepatientandoperatorunlessthe
Safeexposurefactors,exposure
Operationandmaintenancescheduling
Areobserved.
WARNING
RISKOFEXPLOSION
Donotuseinthepresenceofanesthetics
Flammable.
CAUTION
RISKOFFALL
Donotusetheequipmentonafloor
Withaslopegreaterthan5degrees.

CAUTION
RISKOFELECTRICSHOCK
Keeptheequipmentdry.
CAUTION
LASERRADIATION
DONOTLOOKFORLASER
1.0mWMax.CW645nm
ClassIILaserProductIEC608251

21

Page22

Warrantyterm
ShimadzudoBrasilComrcioLtdaguaranteesthenormaloperationfor
Allequipmentsuppliedbyit,againstdefectsorfailuresthatunderuse,
Maintenanceandoperation,occurdueto
Ormaterial,foraperiodoftwelve(12)months,countedfromthedateof
Terminationoftheinstallationandconsequentacceptanceofthe
Maynotexceedfifteen(15)months
OftheBillofLadingorBillofSale,
https://translate.googleusercontent.com/translate_f

21/135

11/28/2016

OPERATIONMANUAL

Whicheveroccursfirst.Thebuyershallcommunicateinwritingtothe
Shimadzu,anydefectsthatmayoccur,withtheevidence
Satisfactory,assoonaspossible.
Defectiveparts,whichareundertheWarrantyperiod,willbe
ReplacedorrepairedatShimadzu'sdiscretionandwithoutchargetothepurchaser.
ThecostsrelatedtothestayandtraveloftheEngineersand/or
Shimadzuanditsrepresentatives,whentheequipmentisinstalled
OutsidethecityofSoPaulo,oroutsidethehostcityofoneoftheRepresentatives
Aswellasotherexpensesnecessaryforsuch
Becoveredbythebuyer.Thereplacementofpartswillbedoneinthesame
Purchaseofequipment.Therepairorreplacementofpart
Equipmentdoesnotextendorinterruptthewarrantyperiod.AShimadzu
Reservestherighttorequestthereturnofreplacedpartsandcomponents
UnderWarranty.
ExceptfortheGuarantee,anyandallobjectssubjecttothe
Deterioration,naturalwearorconsumption,suchas:accumulators,batteries
Droughts,rubberorplasticobjects,includingprotective,filmandphoto
Chemicalproducts,incandescentbulbs,cablesingeneral,materials
Processing,etc.
Defectsordamagesresultingfromaccidentsarealsoexcluded.
Suchas:fires,floods,electricpowerlines,shocks
Maintenance,cleaning,improperuseof
Equipmentornegligence.Alsonotincludedinthiswarranty
Anyindemnityforlossofprofit,personalinjuryandotherproperty.

22

Page23

Chapter1

https://translate.googleusercontent.com/translate_f

22/135

11/28/2016

OPERATIONMANUAL

WHA200Overview

OPESCOPEACTIVE/FULL

ThischapterprovidesanoverviewoftheWHA200OPESCOPEACTIVEand
Describestheenvironmentalcharacteristicsandconditionswhichshouldbeused.

23

Page24

1.1Outline
Thisequipmentisamobileimagingsystemforradiographyand
Fluoroscopyforuseduringsurgery,
AswellasinXrayexaminationrooms.

Pleaseattachtheoptional"ExtensionKitforSourceSkinDistance"
Whentheequipmentisusedforanotherapplicationthan
Forsurgerypurposes.

https://translate.googleusercontent.com/translate_f

23/135

11/28/2016

OPERATIONMANUAL

1.1.1AbouttheWHA200Active/FullOpescope
Innormalconfiguration,themonitorstandhasamonitor
(Forlivefluoroscopyonly).Theexamplebelowreferstothe
Configurationwith2monitors,whichoneisoptional.

Figure1.1SystemOverview

24

Page25

1.2Features
Thisunitincorporatesthefollowingsupportfeaturestoenable
Radiographyandsmoothedfluoroscopywhileensuringfunctionalityand
Requiredintheoperatingroomsand
Emergencies.
Highimagequality
ThesystemincorporatesaCCDcameraandaveryhighxraytube
Capacityandhighdissipationforhigh
Withoutexposuretimelimits.
Lowirradiation
https://translate.googleusercontent.com/translate_f

24/135

11/28/2016

OPERATIONMANUAL

Inthefluoroscopypulse,xraysareemittedasapulse.ITIS
Possibletochoosethelowdosemodethatminimizesirradiationand
Highqualitypicturemodeaccordingtoyourneeds.
The"LastImageHold"functionis
Forcaseswherefluoroscopyisnotrequiredforconfirmation.
Eightimagescanbestoredinmemory.
Speed
Thefreedomofmovementfacilitatesthepositioningofthesystemto
Workonanylimborspine,whichhelpsreducetime
Positionandthetimerequiredforsurgery.
Lightweightandcompact
Thesystemwasmadewithfreeandcompactangle.
Cleaning
ThecablesareembeddednexttotheCarm,sotheydonotgointo
Contactwiththepatientorsurgicalinstruments,allowingthesafeuseof
ofthesystem.
LastImageMaintenance
(LIH)
ThelastXrayimageisdisplayedonthemonitor.

25

Page26

1.3EnvironmentalConditions
1.3.1ConditionsofUse
Inorderfortheequipmenttoachievetheintendedperformance,
Itmustbeusedundertheconditionslaiddown.

Environmentalconditions
Usetheequipmentinanenvironmentwiththe
Specifiedbelow.

https://translate.googleusercontent.com/translate_f

25/135

11/28/2016

OPERATIONMANUAL

Evenwhenthefollowingconditionsaremet,
Makesurethattherearenosuddenchangesin
Temperatureorhumidity.
Thisleadstocondensation,whichcancauseproblems.

Condition

Stipulation

Temperature

10Cto40C

Relativehumidity

30%to85%(noncondensing)

Atmosphericpressure

700hPaat1060hPa

Atmosphereconditions

Theatmospheremustbefreeof
Corrosiveandexplosivegases

Powersupply
Thepowersourcemusthavethefollowingcharacteristics.
Phase
singlephase
B.C

Voltage
100,110,120V10%,0.2
Max.
200,220,230,240V10%,
0.8max.

Capacity
3.5kVAor
larger

Frequency
50or60Hz

Selectoneofthevoltagesindicatedaboveatthetimeof
installation.

26.

Page27

Donotusethepowersourcethatdoesnotconformtothe
Rulesabove.
Theuseofanenergysourcethatisnotincompliancewiththe
Requirementsspecifiedabovecancauseserious
Involvingfire,smokegenerationandexplosion.

EarthRequirements
Theearthresistancemustbelessthan100ohm.

https://translate.googleusercontent.com/translate_f

26/135

11/28/2016

OPERATIONMANUAL

Usealeakagecurrentbreakerthat
Harmonicoscillations.

1.3.2Storage

Evenwhenthefollowingconditionsaremet,
Makesurethattherearenosuddenchangesin
Temperatureorhumidity.
Thisleadstocondensation,whichcancauseproblems.

Condition

Stipulation

Temperature

10Cto40C

Relativehumidity

30%to85%(noncondensing)

Atmosphericpressure

700hPaat1060hPa

Atmosphereconditions

Theatmospheremustbefreeof
Corrosiveandexplosivegases

27

Page28

1.4ClassificationofDevices
1.4.1ClassificationaccordingtotheShockProtectionSystem
Electric.
ClassIDevices

ClassIdevicesarethosethatdependexclusivelyon
Basicinsulationforprotectionagainstelectricshock,inorderto
Toensurethatmetalpartsthatmaycomeintocontactwith
Ifthisbasicinsulationfails,donotcauseanelectricshock,
Throughthegroundwireinisprovidedasameasureof
Additionalsecurity.

https://translate.googleusercontent.com/translate_f

27/135

11/28/2016

OPERATIONMANUAL

1.4.2ClassificationaccordingtoDegreeofProtectionagainstShocks
Electric.
TypeBDevices

TypeBdevicesaredevicesthatprovidealevelof
Particularprotectionagainstelectricshock,anditapplies
Especiallyintheallowableandreliableleakagecurrentandin
Protectiveearthconnections.

28

Page29

1.4.3ClassificationaccordingtoDegreeofProtectionagainstEntry
HarmfulofWater.
StandardDevices

Usetheequipmentinaplacewherethereisnodangerof
Waterinlet.
Ifwatergetsinsidetheequipment,
Causeadangerouselectricshock.Donotallowwaterto
Reachthesurfaceoftheequipmentormoveinsideit.
Ifwatergetsintotheequipmentforanyreason,turnitoff
ImmediatelyandcalltheAuthorizedServiceCenter.

https://translate.googleusercontent.com/translate_f

28/135

11/28/2016

OPERATIONMANUAL

1.4.4ClassificationaccordingtoDegreeofProtectionagainstShocks
Electric.
Equipmentnotsuitableforuseinaninpresenceenvironment
Ofaflammableanestheticmixturewithairoranestheticmixture
Flammablewithoxygenornitrousoxide.

Useinanenvironmentcontainingflammableanestheticgas
Maycauseanexplosion.

1.4.5ClassificationaccordingtotheModeofOperation.
Continuousoperationwithintermittentload
(ContinuousoperationdependsonthespecificationoftheXraytube
mounted)

29

Page30

1.5SymbolsandNameplates
Manufacturer
1.5.1Symbols
Thesymbolsusedontheequipmentareindicatedbelow.
Symbol

Location
IdentificationPlate
Affixedtothehousing
Onthe
Force,wherethe
Protectionunit
Earth.
IdentificationPlate
Affixedtothehousing

CautionLabel/
Warning
https://translate.googleusercontent.com/translate_f

Meaning
Alternatingcurrent
B.C

EarthProtection

Classificationof
Safety:
TypeBdevice
Referenceto
Manualof
Instructions
29/135

11/28/2016

OPERATIONMANUAL

driverof
PotentialEqualization

driverof
Equalization
Potential

30

Page31

1.5.2Tags

Manufacturer'sIdentificationLabels
Manufacturer'sIdentificationLabelsareaffixedtotheindicatedlocations
below.

https://translate.googleusercontent.com/translate_f

30/135

11/28/2016

OPERATIONMANUAL

Figure1.2LabelPosition

31

Page32

https://translate.googleusercontent.com/translate_f

31/135

11/28/2016

OPERATIONMANUAL

Figure1.3ManufacturerIdentificationLabels

32

Page33

Others
LabelsCertifiedbytheEuropeanUnionareaffixedtotheindicatedplace
Asnumber1infigure1.2.

TUVApprovedSticker

CEMarking

EuropeanRepresentative

https://translate.googleusercontent.com/translate_f

32/135

11/28/2016

OPERATIONMANUAL

Figure1.4EuropeanUnionCertifiedLabels

33

Page34

1.6OperatorProfile
Item
Age
Sex
Nationality
Education

Knowledge

Language
Experience

https://translate.googleusercontent.com/translate_f

Details
AgethattheindividualcanobtainthelicenseofTechnician
Radiologyoranequivalentlicense.
withoutlimitation
withoutlimitation
Required
Eachoperatorneedstoconduct
Operatetheequipmentbeforeusingit.
Required
Eachoperatorneedstoconduct
Operatetheequipmentbeforeusingit.
withoutlimitation
Required
Eachoperatorneedstoconduct
Operatetheequipmentbeforeusingit.

33/135

11/28/2016

OPERATIONMANUAL

34

Page35

Chapter2

Configurationofthe

System

https://translate.googleusercontent.com/translate_f

34/135

11/28/2016

OPERATIONMANUAL

Thischapterpresentsthesystemconfiguration.

35

Page36

2.1StandardConfiguration
ThetwolargestcomponentsofthesystemaretheCarmsupportcar
Andstandwithmonitors.
2.1.1SupportCarwithArmC
TheCArmSupportCarcanbeoperatedtoreachposition
Moresuitableforfluoroscopyorradiography.
TheCArmSupportCarisconnectedtothemonitorcartusing
thesystem.

https://translate.googleusercontent.com/translate_f

35/135

11/28/2016

OPERATIONMANUAL

Figure2.1CArmCart
36

Page37

(1)IIControlKnob(optional)
Thisisthehandleusedbythesurgeontooperatethesystem.Thehandleisprovided
WithaCarmlockswitch.
(2)Xrayimageintensifierunit
TheabbreviationforimageintensifierisII
Xraysareconvertedtoavisibleimageonascreen
Thisfluorescentimageisconvertedtophotoelectronby
Photocathode,andthephotoelectronsareacceleratedandconvertedbyalens
Electrostatic,directingthemtothefluorescentoutputscreen.
(3)BeamLimitingDevice(Collimator)
ControlstherangeofXrayradiationwhenperformingfluoroscopyor
CaptureanXray.

(4)XrayGenerator
Thisisthexraygeneratingunit.Thexraytubeisoftheanodetype
fixed.

(5)SidePanel
Throughthispanelitispossibletolock/releasetheelectromagneticlocksofthe
ArmC,rotatethecamera/image,inverttheimage(up/downand
Right/left),open/closethecollimatorandactivatethememoryfunction.
https://translate.googleusercontent.com/translate_f

36/135

11/28/2016

OPERATIONMANUAL

Figure2.2SidePanel
(6)ArmC
TheCarmiscounterbalancedandoperatedmanually.O
Iselectromagnetictype,andcanbeswitchedonoroffwitha

37

Page38

Touch.Thearmcaneasilybepositioned.
(7)Mainpanel
Usedtodisplayandadjusttheradiologicaltechniques,writedatain
memory,andselectimagesdisplayedmemory.

Figure2.3aMainPanelACTIVE

https://translate.googleusercontent.com/translate_f

37/135

11/28/2016

OPERATIONMANUAL

Figure2.3bMainPanelFULL

(8)Handlesformaneuver
UsedtomovethestandarmC

38

Page39

(9)Parkingbrake
ThisistheparkingbrakemanualforthestandarmC

Figure2.4Parkingbrake

(10)Connector
Connectthemonitorstandstandout

https://translate.googleusercontent.com/translate_f

38/135

11/28/2016

OPERATIONMANUAL

Figure2.5Connector

39

Page40

(11)Pedal
Selectthefluoromode.
Leftfootswitch:continuousfluoroscopy,pulsedfluoroscopy
Switchrightpedal:Radiography

Figure2.6Pedal
Subtitle
1LeftPedalcontinuousfluoroscopy,pulsedorHighQuality
2RightPedalXray

(12)UnitXrayControl

https://translate.googleusercontent.com/translate_f

39/135

11/28/2016

OPERATIONMANUAL

40

Page41

2.1.2CartMonitor
Thisisthestandofthemonitor,andisusedconnectedtothestandarmC,
andcontrolsthepowersupplytoallWHA200system.

Figure2.7MonitorStand

Onthedisplaymonitor
Dependingonthedisplayyoucanviewconditions
verysmalldotsorirregularities,butdoesnotindicatea
failure.
https://translate.googleusercontent.com/translate_f

40/135

11/28/2016

OPERATIONMANUAL

Somepixelsmayhavedisplaydefects(Itcan
notlightoralwaysstaylit).
Incoldplaces,theimagemayappearwithadashorthe
poorimage,butthisdoesnotindicatefailure,theequipmentwillreturn
tonormaloncethetemperaturerises.
Donotleaveastillimageforalongtime,thiscan
causedamagetothedisplay.
Theimagemayappearindifferentshadesofcolorbetweenthe
MonitorandtheReferenceMonitor.

41

Page42

(1)Monitor
Displaysthefluoroscopicimagesduringfluoroscopyimages
Radiographic,maintaininglastimage(LIH)andrevisedimages.

(2)ReferenceMonitor
Displaysreferenceimages.
Itallowsrecordingpatientinformationandmakevariousadjustmentsto
imaging(fluoroscopy,radiography),postprocessing,etc.
Inadditiontooperationswiththemouseandkeyboard,youcanalsoperform
operationswiththetouchpanel(optional).

(3)MonitorPositionAdjustment
Usedtomoveupordowntoadjustthepositionof
monitor.

(4)Keyboard
Locationofthekeyboard,mouse,etc.

(5)SystemSwitch
Turnonoroffthesystem.

(6)IndicatorIrradiation
TheindicatedisilluminatedduringradiationbyXrays.

(7)ArmSupportAssistant
Fixedthedisplaypositionintheverticaldirection.Toadjustthisposition,
WemustfirstreleasethelockSupportFlyArm.

(8)PowerSwitch
Turnonoroffthesystem.

https://translate.googleusercontent.com/translate_f

41/135

11/28/2016

OPERATIONMANUAL

42

Page43

2.2MainPanel
2.2.1OpescopeActivefunctionofeachkey

Figure2.8MainPanelOpescopeActive

Legendand

1Releasingtheexposurerestrictionofxrayspresstheswitch
exposuretoxraysorthepedaltoreleasetherestrictionofxraysto
fluoroscopyandradiography.
2Exposuretoxray(fluoroscopy)pressthiskeytostart
exposuretofluoroscopy.
3filmradiographymodepressthiskeytomakearecord
photoonthefilm.TheLEDwilllightup.
4Optionaldigitalsubtractionsystem
5FilterSelectionpressthiskeytoselectthefilter.Thefilter
selectedwillchangeeverytimethekeyispressed..
https://translate.googleusercontent.com/translate_f

42/135

11/28/2016

OPERATIONMANUAL

Filter1(fine):LeftLEDlights
Filter2(medium):TherightLEDlights
Filter3(thick):ThetwoLEDswilllight
Noadditionalfilter:BothLEDsareoff
6Optionaldigitalsubtractionsystem

43

Page44

7Optionaldigitalsubtractionsystem
8Optionaldigitalsubtractionsystem
9APRThecorrespondingLEDwilllighttheAPR.
10IBSpressthiskeytoturnon/offtheIBSfunction(setfunction
Automaticbrightness)
11SelectingthesizeofthemagnificationIIpressthiskeyto
selectthesizeoftheIILEDremainslitwhenmagnification
the6inchfieldisselected,andturnsoffwhenthe9field
inchisselected.
12LaserPointerLEDwillremainlitwhilethelaserpointer
Itisbeingused.
13continuousfluoroscopyratedisplay/PulsedHighQuality
14/15SelectionFluoromodetopressthesekeystoselectthe
Fluoromode.TheLEDslightupasfollows:
Highquality(highqualitymode):
"HQ"light.
Lowdoses(lowdosemode)
"LD"willlightup.
Normal(normalmode)
AllLEDswilllightup.
16TemperaturegeneratorxrayIndicatesthetemperatureofthegenerator
Xray.asthetemperaturebecomeshigh,thelampcolorchangesfromgreento
orangeandred.Whentheindicationisorangeorred,itmeans
thatfluoroscopyisdisabledassoonasthetemperatureofthexraygenerator
istall.
17NumberDisplayindicatesthecurrentmemoryimagenumber.if
noimageisdisplayedwillnotshowanynumber.
18MemoryAddsanimageinmemory
19SelectionPresstochangetheimagedisplayedmemory.
https://translate.googleusercontent.com/translate_f

43/135

11/28/2016

OPERATIONMANUAL

20PrinteroutputPrintsimages(onlyeffectivewhena
printerisinstalled)
21Exchange/displaysthetubevoltageIndicatesthetubevoltage.Press[]/[+]
tomanuallychangethetensionadjustment.However,notethatyoucannot

44

Page45

changingthetubevoltagewhentheIBSfunction(Auto)ison
22CurrentDisplayofthexraytubeToedit/displaythecurrent
InFluoromode:
Thetubecurrentisautomaticallydisplayedinaccordancewiththefluoromode.THE
Currentcurrentcannotbechangedmanually.
Intheradiographicfilmmode:
Indicatestheproductofthetubecurrenttime.Press[]/[+]tochangethis
time.
23CumulativefluoroscopyTime/displayradiographytime:
InFluoromode:
Itshowsthecumulativetimefluoroscopy.theminutesandsecondsaredisplayed.
Whenthetimeexceeds99minutesand59seconds,oneminuteisdisplayed.
Inradiographicfilmmode:
Beforeradiography,displaystheadjustedtimeradiographic
AftertheXray,radiographicdisplaysthemeasuredtime
24Displayoffluoroscopywristband,Camera,collimatorangle,
Errordisplay:
Normallydisplaystherangeofpulsedfluoroscopy(fpslights)
Duringcontinuousfluoroscopy,nothingisdisplayed(fpsoff)
Whenthecameraoperationandthecollimator(ie,theswitchesonthepanel
sidebeingoperated),thecurrentangleisdisplayedflashing.
Ifanerroroccurs,anerrorcodeisdisplayed.
25Resetpressthiskeytoresetadisplayederror.Holding
thiskeyforonesecondthefollowinginformationwillbereset:
cumulativefluoroscopytime
Registrationinformationinthekey
MemoryImage
26APRRegistrationpressthiskeytoregistertheAPR.
27FluoroModeDisplaywilllightthewayFluoro
28ErrorDisplaystayonafteranerrorisactive
https://translate.googleusercontent.com/translate_f

44/135

11/28/2016

OPERATIONMANUAL

29Standbydisplaylightwhiletheirradiationofxraysispossible
30DisplayexposureXraysduringirradiationilluminatexray
31AlarmResetTimerfluoroscopyFortimerfluoroscopy

45

Page46

2.2.2OpescopeFullFunctionofeachkey

Figure2.9MainPanelOpescopeFull

Subtitle:
1releasebuttonlockdisplayxrays:pressto
releasethelockexposurexrays.
2Exposurekeyxray:presstostartfluoroscopy.
3Alarmresetbuttonoffluoroscopictimer:fortheaudiblealarm
timerfluoroscopy.
4Imageexchangekey(optional):Thiskeyisusedtochangethepicture
oroperationscreen.
5Automaticadjustmentkeywindow/level:automaticallyadjuststhecontrastand
thebrightnessoftheradiographicimage.
6windowadjustmentbutton:adjuststhecontrastoftheimage.
7leveladjustmentbutton:adjuststhebrightnessoftheimage.
8depthadjustmentbutton:adjuststhedepthoftheimage.
9examcompletionkey:presstoendtheexam.
10MENUbutton:pressthisbuttontoadjustthemenu.
11Multipurposekey:thiskeyisusedinconjunctionwiththedisplayedbutton
https://translate.googleusercontent.com/translate_f

45/135

11/28/2016

OPERATIONMANUAL

onthescreeninimageprocessing.

46

Page47

2.2.2.1FullOpescopeLCDDisplay
Atouchscreenisbuiltintotheunit,allowing
operationdirectlybypressingthebuttonsonthescreen.
Thebuttonsdisplayeddifferaccordingtothescreen.Thebasicscreenappears
below:

1DisplayoffluoroscopymodeLitinfluoroscopymode.
2IBS(imagebrightnessstabilizationfunction)Pressthiskey
toenableIBSfunction.ThekeyislitwhentheIBSfunctionison.
3voltageoftheXraytubeIndicatesthetubevoltage.Theadjustmentvariesfrom40to
110kV.Press[+]/[]tochangethisvaluemanually.
4currentoftheXraytubeInthefluoroscopymode,thetubecurrent
Itisautomaticallyselected.Thisvaluecannotbechangedmanually.
5SystemStatusIndicatorDisplaysthestatusofthesystem.

https://translate.googleusercontent.com/translate_f

46/135

11/28/2016

OPERATIONMANUAL

Whenfluoroscopicradiographyispossible.

47

Page48

WhenthereisthepresenceofXrayradiation.
WhenthetubeXrayisinhightemperatureorhighhumidity.
6FluoroscopyAccumulatedTimeInorderfluoroscopytime
accumulatedfluoroscopyisindicatedhere.Whenthetimeexceeds99minutes59
seconds,thedisplaychangestominuteunitsonly.
7FilterUsedtoselectthefilter.Eachtimethiskeyis
Pressthefilterchanges.Thefiltersequenceis1Filter<Filter2<Filter3.
8MoviemodeRadiographicPressthiskeywhenyouwanttoperform
rayperfilm.Thescreenradiographappears.
9XraytubetemperatureIndicatesthetemperatureoftheXraytube.
whenthetemperatureincreases,thedisplaybarextends.
10DisplaymodeDisplaysthedisplaymodedialog.Thisallowsmenu
selectthetypeofdisplay,thumbnaildisplay,performing
Zoomoperations,reversethepolarityoftheimage,reverseimage
vertically,orremasktheimage.
11SendREFPresstosendtheimagedisplayedonthemonitortothe
referencemonitor.
12SavefluoroscopicimageSavethefluoroscopicimagedisplayedon
monitor(includingloops)ontheharddrive.
13VideoPrintingPresstoprinttheimagedisplayedonthescreen
monitorinanimageprinter.(Onlyappearswhentheprinter
specifiedisconnected.)
14SelectionAPRDisplaystheAPRscreen.
15IntensifiedimagefieldsizeselectionPressto
select16cmor23cmasintensifierfieldsize
Image.
16LoopFluoroscopyUsedtoregisterafluoroscopicimage
inmemory(Maximum100imagesbeforefinalizingfluoroscopy).Stay
litwhenthefunctionison.
17LaserPointeronoroffthefunction.
18ProductDosebyAreaIndicatesthedoseperareadetectedbyasensor.

https://translate.googleusercontent.com/translate_f

47/135

11/28/2016

OPERATIONMANUAL

48

Page49

2.3Sidepanel
2.3.1lockswitcharmC
UsedtolockorreleasethearmC.

Figure2.10lockswitcharmC

1slipmovementofthearmC
LEDon:ThelockarmCisreleased
LEDoff:ThearmCislocked
2RotationalmovementofthearmC
3lateralandposteriorslidingmovementofthearmC
4longitudinalmovementofthearmC

2.3.2ControlKeys

Figure2.11ControlKeys

https://translate.googleusercontent.com/translate_f

48/135

11/28/2016

OPERATIONMANUAL

49

Page50

1/2MemoryKeyusedtostoretheconditionsoffluoroscopyand
thecollimatorsettings
3Open/Closethecollimatoroctagonal
4Open/Closetheparallelcollimator
5Rotatestheparallelcollimator

Figure2.12Collimatorsoctagonalandparallel
6FastFoodholdingthiskeywhilepressingthe
rotationoftheparallelcollimatororcamerarotationspeedinthese
operationsincreases
7Rotationcamerausedtorotatethecamera
8Monitorimagereversalusedtoreversetheimagedisplayedon
themonitor.

https://translate.googleusercontent.com/translate_f

49/135

11/28/2016

OPERATIONMANUAL

50

Page51

Chapter3

Basicoperation

Thischapterexplainsthebasicproceduresforusingthesystem.

51
https://translate.googleusercontent.com/translate_f

50/135

11/28/2016

OPERATIONMANUAL

Page52

3.1BASICOPERATIONS

PoweringOn/Offpower
Toavoidproblemswiththesystemwhileexaminingapatient,
alwaysmakeaninspectiontocheckthateverythingisinaccordancewiththe
specifications.

3.1.1ConnectingtheUnits
1Unrollthesignalcableonthebackofthecarsupport
monitors.
2Connectthesignalcabletothemainunittothispositiontheconnectorand
rotateitclockwiseturnasfig.below.

Figure3.1Connector

3Connectthepowercordatthebackofthecarsupport
monitorsthelocaloutlet,confirmingthattheunitvoltageiscompatiblewith
makingselected.

52

https://translate.googleusercontent.com/translate_f

51/135

11/28/2016

OPERATIONMANUAL

Page53

3.1.2)Connecttheequipment

1Setthepowerswitchonthebackofthemonitorstandto"|"
[switchedon].

Figure3.2On/offswitch
3
2Pressthefrontofthekeymonitorstand.

AllLEDsonthemainpanelandsidepanelwilllightupfor
about1second.

Initializationofthecollimator,camerarotation,etc.,willcontinuefor
approximately30seconds.Duringthistime,thekey"init"
lightsinthepulseratedisplayedonthemainpanel.

Whenyouturnofftheequipmentandneedtoconnectagain,waita
10minuteinterval.

53

https://translate.googleusercontent.com/translate_f

52/135

11/28/2016

OPERATIONMANUAL

Page54

3.2Turningthepoweroff

3.2.1ActiveOpescope
Afteryoufinishusingthemachineperformthefinalinspectionandfilethe
results.

1MakesurethearmCisinthehome.
Theinitialpositionisoneatallscalesrecordzero.

Figure3.3StartPositionArmCOpescopeActive
3
2Pressthekeyto[OFF].
3Settheswitchonthebackofthemonitorstandto[off].
4Disconnectthemainpowercable.

54

Page55
https://translate.googleusercontent.com/translate_f

53/135

11/28/2016

OPERATIONMANUAL

3.2.2FullOpescope

strictlyfollowtheprocedurebelowtoturnoffthe
equipment.
OnlypresstheOFFbuttononthebackofthemonitorstand(s)or
unplugthepowercordafterthecompletionoftheprocess
systemshutdown.Butthesystemandthedataislost.
Also,iftheOFFkeyispressedduringthedatatransfer
imagetoamedia(CDR),theimagedatawillbelost.

1MakesurethearmCisinthehome.
Theinitialpositionisoneatallscalesrecordzero.

Figure3.4StartPositionArmCFullOpescope
2Proceedwiththeshutdownoperationoftheequipmentthroughthemonitor
ofreference.
Pressthefunctionkey
Choosefunction

todisplaythefunctionsonthemonitor.
.

Thepatient'sfilewillbeclosedandaconfirmationdialogbox
askingifyouwanttoexittheapplicationisdisplayed,confirm

55

Page56

https://translate.googleusercontent.com/translate_f

54/135

11/28/2016

OPERATIONMANUAL

clicking"Yes".
3Makesurethegreenlightontheswitchfromthefrontofthecart
monitor(s)off.
4Pressthepowerbutton(OFF)onthebackofthecart
monitor(s)

Figure3.5PowerButtonFullOpescope
5Unplugthepowercord

3.3POSITIONING(ARMCOPERATION)
PlacetheArmCintherequiredpositionfluoroscopy/radiography.

Figure3.6Locationofthepointwherethefingerscangetcaught

56

Page57

https://translate.googleusercontent.com/translate_f

55/135

11/28/2016

OPERATIONMANUAL

Beawareofthedangerofshockorholdanypartofthebodyto
equipment.
WhenoperatingthearmC,carefulnottobumptheequipmentdevices
peripherals.ManipulatethearmCbythehandles,itmayhappenaccidents
howtoholdyourfingerstotheequipment.

3
Donotplaceperipheraldeviceswithintherangeofmovement
system.
Ifperipheraldevicessuchasthemonitorstandisplacedwithinthearea
showninFigure15,thereisadangerofcollisionorcontact.

Figure3.7DriveRange

57

Page58

https://translate.googleusercontent.com/translate_f

56/135

11/28/2016

OPERATIONMANUAL

1PressthearmCofthebeaminthedirectionyouwanttomove.
TheLEDwilllightandthecorrespondingbrakeisreleased.

2Holdingthehandles,movethearmCfortheposition.

3WhenyoufinishpositioningthearmC,engagethelock.
TheLEDwillturnoffandthearmCislocked.

4Makesurethebrakeisapplied.
Evenwhenthebrakeisactivated,youcanmovetheframein
oppositiontothebrakebyapplyingaforcegreaterthan10kgf.Thisisameasure
toenabletheoperationofallpartsofthesystemevenin
criticalsituationasoverhead.

Ifmorethanoneminutepassedwithoutanyoperationintheunlockedstate
(LED),thesystemautomaticallyreturnstothelockedstate(LED
wipedout).

3.3.1VerticalmovementofarmC

Figure3.8VerticalmovementofarmC
58

Page59

https://translate.googleusercontent.com/translate_f

57/135

11/28/2016

OPERATIONMANUAL

3.3.2longitudinalmovementandlateralarmC

Figure3.9longitudinalandlateralmovementsofthearmC

59

Page60

3.3.3axialrotationalmovementofthearmC

https://translate.googleusercontent.com/translate_f

58/135

11/28/2016

OPERATIONMANUAL

Figure3.10axialrotationalmovementofthearmC

3.3.4rotationalmovementofthearmC

Figure3:11rotationalmovementofthearmC

60

Page61

3.3.5scales
Everymovementfeaturesascalethatyoucanseeasaguide
https://translate.googleusercontent.com/translate_f

59/135

11/28/2016

OPERATIONMANUAL

duringthemove.Notethattherangeofcolormatchesthecolorkey
armClock.
ThezerooneachscaleistheinitialpositionofthearmC.

Figure3.12Scale

61

Page62

3.3.6GauntletofIIControl(Optional)

TherearetwokeysonthehandleII
https://translate.googleusercontent.com/translate_f

60/135

11/28/2016

OPERATIONMANUAL

Whilethiskeyispressed,thelockonthereturnslip
sideandarmClongitudinalmovementarereleasedandtheLEDsbelow
Theywillbeilluminated.

Figure3.13lockkeysonthehandleII

62

Page63

3.4SELECTINGAPR

3.4.1OPESCOPEASSETS
https://translate.googleusercontent.com/translate_f

61/135

11/28/2016

OPERATIONMANUAL

1Pressthe[APR]key.TheLEDwilllightup.
Thereare6typesofRPAaccordingtotheobservationposition.Theconditions
greatfluoroscopyandcamerarotationpositionsaresetintheAPR.
TheAPRsaresetinthesystematthefactory,butyoucanchangeto
Accordingtooptimizeitsuse.

Figure3.14APRKey(OpescopeActive)
T
LegendPositionNote:
1HeadHead
2AbdomenAbdomen
3Upperarm,handjoints,fingers,shoulderjoints,etc.upperarm,gaskets
hand,fingers,jointsofshoulders,etc.
4LegsLegs
5/6UsersettingsUsersettings

63

Page64

3.4.2OPESCOPEFULL

AfterregisteringthepatientorpresstheAPRselectionbuttononthe
mainpanel,itwillbedisplayed.
https://translate.googleusercontent.com/translate_f

62/135

11/28/2016

OPERATIONMANUAL

1Check[APR]
TochangetheAPR,presstheselectkey.
TheselectedAPRwillappearhighlightedonthemonitor.

2Press[OK]
Theprimaryscreenisdisplayed.

Figure3.15APRKey(OpescopeFull)

64

Page65

3.4.2.1ChangingtheAPRscreen
Theywillbeshownsixbuttons[APR]simultaneouslyonthescreen
APR.Press[next]togotothenextpage.It'spossible
https://translate.googleusercontent.com/translate_f

63/135

11/28/2016

OPERATIONMANUAL

viewupto12APRs.

Figure3.16APRSelectionScreen

65

Page66

3.4.2.2ChangingthecombinationoffluoroscopyandacquisitionAPRAPR

1Presstheacquisition/fluoroscopyAPR.
Theselectionoptionsaredisplayed.
https://translate.googleusercontent.com/translate_f

64/135

11/28/2016

OPERATIONMANUAL

Figure3.17Screenselectionofacquisition/fluoroscopyAPR

2ChangetheselectionoffluoroscopyAPRandAPRacquisitionandpress[closes].
APRselectionscreenwillberedisplayed.
FluoroscopyAPRandAPRacquisitionwillbeindicatedonthedisplay.

66

Page67

https://translate.googleusercontent.com/translate_f

65/135

11/28/2016

OPERATIONMANUAL

Figure3.18Screentoclosethewindow

3Press[OK]
Theprimaryscreenisdisplayed.

67

Page68

3.5FLUOROSCOPY
3.5.1OPESCOPEASSETS

https://translate.googleusercontent.com/translate_f

66/135

11/28/2016

OPERATIONMANUAL

Figure3.19resetbutton

Whenfluoroscopyoptionequipmentistestedbeforestart
cycletests,youneedtoreset(delete)thefluoroscopytime
totalized.
Pressthe[reset]buttonforabout1second.

1Press[releasingthexrayexposurerestriction].

TheLEDlights[releasingthexrayexposurerestriction].

68

Page69

https://translate.googleusercontent.com/translate_f

67/135

11/28/2016

OPERATIONMANUAL

Figure3.20Freeingexposurerestrictionofxrays

Thexrayradiationisonlypossiblewhenthekey
ordertopreventfaultyoperation,theLEDisoffwhenthe
poweristurnedon.WhiletheLEDisoff,youcannot
initiatingradiation.

ison.In

3
2Checktheconditionoffluoroscopy.Thevariablesareset
automaticallyaccordingtotheselectedAPR,butyoucanchangethe
manuallyadjustments.

Figure3:21Checkingtheconditionsoffluoroscopy

69

Page70

3Pressthepedalorpress[exposurexray(fluoroscopy)].
Xraysareirradiatedwhilethepedalispressedorwhile
keyispressed.
Theradiationindicatorlights.
Thefluoroscopicimageisdisplayedonthemonitor.

https://translate.googleusercontent.com/translate_f

68/135

11/28/2016

OPERATIONMANUAL

Figure3.22Startfluoroscopy(exposurekeyxray)

Whenthesettingsoffluoroscopyconditions,orcollimator,orrotation
camera,areexchangedmanually,theLEDwillflashAPR.
3

70

Page71

changingthe
soFluoro

HQmodetheLEDwilllight.
InLDmode,theLEDwilllight
Inordinarywordstate,pressthebuttonthatislitto
turnofftheLED
HQ:HighQuality(highqualitymode)
Theareaisirradiatedwithahigherdoseofxraystogenerate
Highqualityimagesinpatientsdifficulttopenetratethe
Xray.
LD:LowDose(lowdosemode)
Thelowdosexraysisusedtoreducetheamount
irradiatedxrays.
Normal(normalmode)
Theirradiateddoseisstandard.

Swapping
inbetween
fluoroscopy
pulsedand

Pressthekeys
toswitchbetweenfluoroscopy
continuousorpulsedfluoroscopy.Forpulsedfluoroscopy,the
Pulserateisdisplayedwithanumericalvalue.
Tobecontinued

https://translate.googleusercontent.com/translate_f

69/135

11/28/2016

OPERATIONMANUAL

fluoroscopy
tobecontinued

Press
pulsed

Calling/
disconnectingthe
functionIBS
(Function
stabilization
Shining
image)

TheLEDwillremainlitwhiletheIBSfunction
isoff.
WhenIBSfunctionisconnected,thedoseofxraysis
automaticallyadjustedaccordingtothearea.

Changingthe
tubevoltage
Increasesvoltage
Decreasestension
OnlyvalidwhileIBSisOff
Changebypressingthekeys[+]/[].
Exchanging
filter

Thefilterischangedeverytimethekey
Filter1(fine):LeftLEDlights.
Filter2(medium):TherightLEDlightsup.
Filter3(thick):BothLEDslightup.
Noadditionalfilter:BothLEDsareoff

ispressed.

71

Page72

Selecting
thesizeofthe
magnification
II

Press
toselectthesizeof
magnificationII
Whenthefield16cm(6inches)isused,theLED
remainon.Whenthepitchof23cm(9inches)is
used,theLEDisoff(Validfor9Activemodels"
andFull.

ForcontinuousfluoroscopyinHQmode,thehighdoseisused.Forthisreason,a
warningtoneissoundedcontinuously.

https://translate.googleusercontent.com/translate_f

70/135

11/28/2016

OPERATIONMANUAL

Ifthefluoroscopytimeisexceeded,thealarmwillsound.Thealarmcanbe
stoppedbypressing

Youcancontinuewithfluoroscopyevenwhenthefluoroscopytime
setisexceededandthealarmsounding.However,thelimitforfluoroscopy
Continuousis10minutes.Asthislimitisreached,theexposureofxraysis
automaticallystop.Tocontinueafterthispointfluoroscopy
releasetheleftpedalandpressitagain(orreleasethebutton
[Exposurexray(fluoroscopy)]andpressitagain).

72

Page73

3.5.2OPESCOPEFULL

1Checkthefluoroscopicconditions.
Thefluoroscopicconditionsareautomaticallyadjustedaccordingtothe
APRselection.
Youcanchangetheseconditionsmanually.

https://translate.googleusercontent.com/translate_f

71/135

11/28/2016

OPERATIONMANUAL

Figure3:23Mainscreenoffluoroscopicconditions

2Ifthekeyisoff,pressit.
Tomaketheirradiationbuttonshouldbeon.

Figure3.24exposurelockreleasebuttonxray

73

Page74

3Presstheleftpedal.
Irradiationforfluoroscopystarts.
Duringirradiation,thekeywillremainonthemainpanel.
Whilexraysareirradiated,theirradiationindicatorislitin
Standmonitor.Thefluoroscopicimageisdisplayedonthemonitor.

https://translate.googleusercontent.com/translate_f

72/135

11/28/2016

OPERATIONMANUAL

Figure3.25Exposurebuttonxray(fluoroscopy)

3.5.2.1Changingthefluoroscopicconditions1(fluoroscopyprogram)

1Press[fluoroscopyProgram].
Fluoroscopyprogramselectionscreenisdisplayed.

Figure3.26fluoroscopyProgram
2Selectthefluoroscopyprogram.
Theselectedbuttonishighlightedandthewindowcloses.

74

Page75

Figure3.27fluoroscopyProgram

[Next]fluoroscopyprogramselectionscreenshowsfour
fluoroscopyprogramsonapage.Press[next]todisplaythe
nextpagefluoroscopyprogramselectionscreen.
3.5.2.2Changingconditionsfluoroscopy2(fluoroscopymode)

https://translate.googleusercontent.com/translate_f

73/135

11/28/2016

OPERATIONMANUAL

1Press[fluoroscopymode].
Theselectedbuttonwillappearprominently.

[HQ]HighQuality(highqualitymode)Theareaisirradiatedwithhigher
dosexraystogeneratehighqualityimagesinpatientsdifficult
penetrationofXrays.
[N]:Normal(normalmode)thestandarddoseisirradiated.

[LD]Lowdose(lowdosemode)lowdosexraysisradiated.

Figure3.28fluoroscopymode

75

Page76

3.5.2.3Changingthefluoroscopicconditions3(pulserate)

1Press[pulserate]
Pulserateselectionscreenisdisplayed.

Figure3:29Pulserate

https://translate.googleusercontent.com/translate_f

74/135

11/28/2016

OPERATIONMANUAL

2Selectthefluoroscopicpulserate.
Select[CONTINUOUS]forcontinuousfluoroscopy.Whenselectingany
anotherbutton,pulsedfluoroscopyisperformedandthepulseratewillbe
changesinrelationtotheselectedvalue.
Ifyouprefernottochange,press[closes].

Figure3.30PulseRateSelection

76

Page77

3.6CAMERAROTATION,REVERSEIMAGE,
COLLIMATOROPERATION

Figure3.31Viewofthecontrolbuttons
3.6.1CameraRotationKey
camerarotation/image
Pressingthisbuttonthecamera/imagerotates.Duringfluoroscopy,theimage
displayedtheexposedareaalsorotatesaccordingtotherotationofthecamera.ITIS
possibletoincreasethespeedofrotationofthecameraholdingthekey
fastfoodwhilepressingthecamerakeyrotation.
https://translate.googleusercontent.com/translate_f

75/135

11/28/2016

OPERATIONMANUAL

Whileholdingthecamerarotationkey,therotationangle
thecameraisdisplayedinthemainpanel.(OpescopeActive)

Figure3.32Screenrotationangle(OpescopeActive)

77

Page78

3.6.2ImagebuttonReverse
reverseimage

Pressingthisbuttonfluoroscopicimagereversed.
1.ReverseH(horizontalinversion)

2.InversionV(verticalflip)

https://translate.googleusercontent.com/translate_f

76/135

11/28/2016

OPERATIONMANUAL

78

Page79

3.6.3OperationKeyofCollimator
Pressthesebuttonstoopen/closeandrotatethecollimatorandoctagonal
parallelcollimator.
Whilepressingthebutton,theangleofrotationisindicated
onthescreeninthemainpanel.Theparallelcollimatorallowsrotations
repeated.

Figure3:33Collimatoroctagonalandparallel
1.Open/Closethecollimatoroctagonal
Pressingthisbuttonopensthecollimatoroctagonal.

Pressingthisbuttontheoctagonalcollimatorcloses.
https://translate.googleusercontent.com/translate_f

77/135

11/28/2016

OPERATIONMANUAL

2.Open/Closeparallelcollimator
Pressingthisbuttonopensthecollimatorparallel.

Pressingthisbuttonclosestheparallelcollimator.
3.Rotationoftheparallelcollimator
Pressingthisbuttontheparallelcollimatorrotatesin
schedule.
Pressingthisbuttontheparallelcollimatorrotatescounterclockwise
schedule.
4.Speed
Pressingthiskeytogetherwiththekeyrotation
theparallelcollimatorspeedisincreased

79

Page80

3.7FUNCTIONOFMEMORYFRAME

OPESCOPEASSETS
Youcantemporarilystoreupto8imagescurrentlydisplayed.
Settingonamonitor(standard)
Memoryimagesframearedisplayedonthemonitor.Note,however,
thatduringthefluoroscopicimagefluoroscopyappears,butthen
fluoroscopytheLIHimageisdisplayed.Youcanswitchtodisplaythe
framememory,performingtheoperationsoftheframememory.
Configurationoftwomonitors(usedonlyasthesecond
monitoroptionalisselected)
Theframememoryoftheimageisalwaysdisplayedonthereferencemonitor.
storingimages
Press

[Framememory].

Thecurrentimageisdisplayedonthemonitorisstored,andthenumberof
https://translate.googleusercontent.com/translate_f

78/135

11/28/2016

OPERATIONMANUAL

memoryisdisplayed.

Imageselection
Press

[Selectionframe].

Pressingthisbutton

theframenumberincreases.

Pressingthisbutton

framenumberdecreases.

Theexchangeimagedisplayedwhenthememorynumberischanged.
Memorynumber1:latestimageinmemory
MemoryNumber8:olderpictureinmemory

80

Page81

Whenthenumberofpayrollimagestotheframememoriesexceeds
8,thenewimageisstoredasnumber[1]andtheimage
storedinthenumber[8]iserased.

Whenthesystemisturnedoff,thestoredimageisdeleted.

https://translate.googleusercontent.com/translate_f

79/135

11/28/2016

OPERATIONMANUAL

81

Page82

3.8RADIOGRAPHY(CRRadiographicorMovie)

OPESCOPEASSETS

WhenRadiographicMoviemodeisselected,thecollimatoroctagonal
Itwillautomaticallybefullyopened.Ifnecessaryadjustthe
collimatorparallel,makethesettingonthesidepanel.

Setthetapeandtaketheradiograph.
[FilmradiographymodeorCR].

1Press

2SetthekVandmAsvaluesbypressing[+]/[].

3Releasetherightpedalorpressthehandbutton(optional).
StartpreparationsforXray.
The[ofthescreen]keywilldeletetemporarily.
https://translate.googleusercontent.com/translate_f

80/135

11/28/2016

OPERATIONMANUAL

Whenpreparationsforexposurearecompleted,the
keywillreturn

4Whenusingthepedal,holdthepedal.
Whenusingthehandcontrol(optional),pressthecontrolforthesecond
level.
Thexrayirradiationstarts.
Duringirradiation,theindicatorlightirradiationwillbe
on.

82

Page83

Keepthepedalorhandcontrol(optional)Pressed
FullXrayradiation.Ifthecommandsarereleased,the
radiationXrayswillendprematurely.

Ifthesystemhasthehandcontrol(optional),youcannottake
Xrayswiththerightpedal.

https://translate.googleusercontent.com/translate_f

81/135

11/28/2016

OPERATIONMANUAL

Figure3:34FilmRadiography(OpescopeActive)

83

Page84

DependingontheXrayparameters,onemaybenecessary
waitingtimeafterradiography.Inthiscase,theword"COOL"willbe
displayed,alongwiththewaitingtime.Duringthistimewillnot
possiblexrayfluoroscopybutyes.

https://translate.googleusercontent.com/translate_f

82/135

11/28/2016

OPERATIONMANUAL

Figure3.35DisplayCOOLandDisplayTimeout

84

Page85

OPESCOPEFULL
1Placethesupporttape
2Press[radiographymodefilm].
Thekeywilllight.

Figure3:36radiographymodeinmovie

3Press[+]/[]toadjustthetubevoltage(kV)andtheproductof
timextubecurrent(mAs).

https://translate.googleusercontent.com/translate_f

83/135

11/28/2016

OPERATIONMANUAL

Figure3:37radiographymodeinmovie

85

Page86

4Releasetherightpedal.Alternatively,pressthehandcontrol
(Optional)tothefirstlevel.

Figure3:38RadiographyinFilm
5Continuereleasingtherightpedal,orpressthehandcontrolto
thesecondlevel.
Thexraysareirradiated.Duringirradiation,thecorresponding
indicator(onthemonitorstand)willremainon.

Keepthepedalorhandcontrol(optional)Pressed
FullXrayradiation.Ifthecommandsarereleased,the
radiationXrayswillendprematurely.

6Press[radiographymodefilm]
https://translate.googleusercontent.com/translate_f

84/135

11/28/2016

OPERATIONMANUAL

ThekeygoesoutandtheXrayiscompleted.Theprimaryscreenisre
displayed.

86

Page87

3.9USEFULFUNCTIONS
3.9.1MemoryKey
Storetherotationalpositionofthecamera,collimatorposition,thefactor
loadingandotherinformation.

Thisoperationisnotpossibleduringfluoroscopy.
Informationstoredwiththisfunctioncanbeerased
pressingthebutton.

Thefollowingparameterscanbestored.
Camerarotationalposition
Positionopen/closeandrotationalangleofthecollimators
ImageInversion
FluoroMode
Pulserate
Filter
Switchingon/offIBS(imagebrightnessstabilizationfunction)
Tubevoltage(IBSisconnectedwhenthetubevoltagestarts)
magnificationsizeII

https://translate.googleusercontent.com/translate_f

85/135

11/28/2016

OPERATIONMANUAL

Figure3:39MemoryKey

87

Page88

1Holdthebutton(forabout1second).
Thecurrentconditionsarestored.
WhenthestorageendtheLED.

Ifthestoredconditionschange,theLEDwillflash,indicatinga
incompatibilitybetweenthecurrentconditionsandstoredconditions
therelevantmemorykey.
2Pressthememorykeythatisflashing.
Theconditionsstoredinthememorykeywillbecalled.
Atthesametime,theLEDstopsblinkingandstayson
continuously.
differentusesforthetwomemorykeys
1HoldthememorybuttonAforabout1second.
ThecurrentconditionswillbestoredinmemorykeyA.
WhenthestoreisfulltheLEDwillflash.
2Modifytheconditionsasthecamerarotationalpositionmode
Fluoro.
MemorykeyAflashes.
3PresstheBbuttonbriefly.
ThemodifiedconditionswillbesavedinthememorykeyB.
Whenstorageiscomplete,theLEDwillflash.
TheLEDAoff.
4PressthememorykeyA.
TheAbuttonwilllight,andthesystemwillreturntothesavedconditionsin
https://translate.googleusercontent.com/translate_f

86/135

11/28/2016

OPERATIONMANUAL

A.memorykey
TheLEDBoff.
5PressthememorykeyB.
TheLEDBlightsupandthesystemwilladopttheconditions
savedinthememorykeyB.
TheLEDAoff.

88

Page89

3.9.2ChangingtheAPRadjustment(OpescopeActive)
YoucanchangetheAPRsettings.

1ChangetheconditionsasFluoromodesettingsandconditions
Xray.
Thefollowinginformationmaybelogged.
Camerarotationalposition
Positionopens/closesthecollimatorandrotationangle
ImageInversion
fluoroscopymode
Pulserate
Filter
On/OffIBS(imagebrightnessstabilizationfunction).
Tubevoltage(asIBSisconnected,thetubevoltage
starts)
magnificationsizeII

2Press[APR]andthekeyatthesametime.
TheadjustmentintheAPRwhosebuttonsarepressedwillbewritten.
Whenregistrationiscomplete,analarmwillsound.

https://translate.googleusercontent.com/translate_f

87/135

11/28/2016

OPERATIONMANUAL

Figure3:40RegisteringthesettingsmodifiedinAPR

89

Page90

3.9.3Changingthepresettimerinfluoroscopy
Youcanadjustthetimetothepresettimerinfluoroscopy.

Screenwithpresetvalueofthetimer

Figure3:41Changingthepresettimerfluoroscopy

1InFluoromode,pressthe[changingthetimerpre
adjustedfluoroscopy].
Thevalueofthetimersetwillbedisplayedflashing.
2Eachtimeyoupressthebuttonwhilethedisplayisflashing,the
timervaluewillbeincreasedby30seconds.
Youcanadjustthevalueofthetimerasfollows:

3thecourseof3secondsaftertheoperationinstep2,thescreenreturns
thecumulativedisplaytimefluoroscopy.
https://translate.googleusercontent.com/translate_f

88/135

11/28/2016

OPERATIONMANUAL

90

Page91

3:10DIGITALRADIOGRAPHY
OPESCOPEFULL

Thedigitalradiographyconditionsareautomaticallyadjusted
basedonfluoroscopicconditions.Forthisreason,beforetakingthe
Xray,youmustconnecttheIBSfunctionandperformfluoroscopyto
thefluoroscopicimageisstabilized.
1Checktheradiographicconditions.

Figure3:42ScreenRadiography

https://translate.googleusercontent.com/translate_f

89/135

11/28/2016

OPERATIONMANUAL

Thesefunctionsareonlydisplayedwhenatechniquethatinvolves
serialradiography,suchasDSAorserialisselected.

91

Page92

2Releasetheleftpedal.
Irradiationstartsandtheradiographicimageappears
onthemonitor.

Figure3:43Gettingradiography

https://translate.googleusercontent.com/translate_f

90/135

11/28/2016

OPERATIONMANUAL

92

Page93

Changingtheradiographicprogram(acquisitionprogram)

Changesmadethroughthemainpanelaretemporary.

1Press[selectionacquisitionprogram]
Theacquisitionscreenappears.

Figure3:44Selectionacquisitionprogram

https://translate.googleusercontent.com/translate_f

91/135

11/28/2016

OPERATIONMANUAL

93

Page94

2Selecttheacquisitionprogram.
Thebuttonishighlightedandthewindowcloses.

Figure3:45Selectionacquisitionprogram

[Next]Theselectionwindowacquisitionprogramshows
fouracquisitionprogramsonapage.Press[next]
toviewthenextprogramselectionpage
acquisition.

Ifyoudonotwanttochangetheselectionoftheacquisitionprogram,performthe
operationsdescribedbelowtoreturntothebasicscreenwithoutchangingthe
acquisitionprogram.
Press[Close]
StartAcquisition
https://translate.googleusercontent.com/translate_f

92/135

11/28/2016

OPERATIONMANUAL

94

Page95

Changingtheacquisitionrate
1Press[acquisitionrate]
Thewindowwiththeselectionoftheacquisitionrateisdisplayed.

Figure3:46ChangeofAcquisitionRate1

[Acquisitionrate]isonlydisplayedwhenatechniqueinvolving
continuousradiographyas"serial"or"DSA"isused.

Ifaprogramsequencehasanacquisitionrateexceeding5
fps,themodeishighspeedandtubevoltageandthequalityof
imagecanbechanged.

https://translate.googleusercontent.com/translate_f

93/135

11/28/2016

OPERATIONMANUAL

95

Page96

2Selecttheacquisitionrate
Thebuttonwillbehighlightedandtheselectionwindowoftheacquisitionrate
Itwillbeclosed.
Ifyouwanttokeeptherateofpurchase,pressthe[close]

Figure3:47ChangeofAcquisitionRate2

Figure3:48ChangeofAcquisitionRate3

https://translate.googleusercontent.com/translate_f

94/135

11/28/2016

OPERATIONMANUAL

96

Page97

Changingtheprogramsequence

Press[sequence]intheselectionwindowoftheacquisitionrateandmake
selection.Youwillneedtoadjusttheprogramsequenceonthemonitor.if
thereisnosetsequence,the[detailsdetails]not
appears.

1Press[details]
Theprogramsequenceselectionwindowappears.

Figure3:49acquisitionsequencedetails
2Press[]/[]toselecttheprogramsequence.
Theselectedprogramsequenceisdisplayed.

Figure3:50Selectionofprogramsequence

97
https://translate.googleusercontent.com/translate_f

95/135

11/28/2016

OPERATIONMANUAL

Page98

3Press[closes]
Theprogramsequenceselectionwindowisclosed.

ProgramSequenceDetails

Figure3:51ProgramSequencedetails.
Thefollowingdetailsscreenshowsdetailsoftheprogram
sequenceofselectedprograms.Itdisplays5levelsrecorded
acquisitionrateandtime.

98

https://translate.googleusercontent.com/translate_f

96/135

11/28/2016

OPERATIONMANUAL

Page99

3:11Tweaking
FullOpescope
Youcancalibratethetouchpanel,adjustthesound,setthetimer
Fluoroscopy,checkthesettingsstoredinmemory.

Makesurethedisplayisturnedoffbeforetheoperation.
Thechangesthataremadetotheunitwillremainthesameeffective
themachineisrestarted.

1Pression[Menu]
TheSystemSetupmenuscreenappears.

Figure3:52SystemSetupMenuScreen

99

https://translate.googleusercontent.com/translate_f

97/135

11/28/2016

OPERATIONMANUAL

Page100

Chapter4

Troubleshooting

Thischapterexplainsthecorrectiveactionstobetakeniftheequipment
presentproblem.

100

Page101
https://translate.googleusercontent.com/translate_f

98/135

11/28/2016

OPERATIONMANUAL

4.1Respondingtoproblems
Whenaproblemoccurswiththesystem,followtheinstructions
flowchartfollowingappears:

Ifyouhaveanyquestionsonprocedure,contacttechnicalassistance
authorized.

Ifyounoticeanyproblemswiththesystem,stopuseimmediately.

Repairsandmaintenanceshouldbecarriedoutonlybytechnicaland/or
authorizedservice.

101

Page102
https://translate.googleusercontent.com/translate_f

99/135

11/28/2016

OPERATIONMANUAL

FluoxogramaprocedureConcealer

Occurrenceofproblem

Ifamessage
appears,
seethelistof
errors.

No

Got
solvethe
problem?

Yes

END
Telephoneassistance
authorizedservice.

102

Page103

https://translate.googleusercontent.com/translate_f

100/135

11/28/2016

OPERATIONMANUAL

4.2Listoferrors
Ifanerroroccurs,thekeywilllightupdependingonthenatureof
errorandacodewillbedisplayedontheerrorscreen.
Code
Er1to9
Er1019
Er20to29
Er30to39
Er40to49
Er50to59
Er60to99
HEAT

Message
CPUerror
sourceoferrorofinternalenergy
controlcircuiterror
Errorinthehighvoltagegenerationcontrolsystem
Errorsassociatedwithtemperaturedetectionand
communications
Errorinmechanicalcontrolsystemcollimator
Errorsassociatedwiththesystem
Hightemperaturethroughoutthexraytube

Whilethebuttonisflashing,youcanperformfluoroscopy
radiograph.
Whilethekeyislitcontinuously,youcannot
performfluoroscopyandradiography.
Whenthexraytubeassemblyattainsahightemperature,
HEATcodeappears,andyoucannotperformradiographyand
fluoroscopy.Whenthetemperaturelowers,themessageis
automaticallyerasedandthenyoucanperformfluoroscopy
radiograph.

Figure4.1AssetOpescopeerrorscreen

103

Page104

https://translate.googleusercontent.com/translate_f

101/135

11/28/2016

OPERATIONMANUAL

Figure4.2OpescopeFullErrorScreen

4.3OperatingChange

Intheeventofoperatingchange,turnoffthemachineandenter
contactyourdealer.

ShimadzuofBrazilLtda
Address:MarquesAvenueStVincent,1771SP
Tel.:(11)21341688/Fax:(11)36112209
email:webmaster@shimadzu.com.br
website:www.shimadzu.com.br

104

Page105

https://translate.googleusercontent.com/translate_f

102/135

11/28/2016

OPERATIONMANUAL

Chapter5

MaintenanceandInspection

Thischapterexplainsthedailyandperiodicinspections.Inspectionsshouldbe
heldastomaintainequipmentperformanceandensure
safety.

105

Page106

https://translate.googleusercontent.com/translate_f

103/135

11/28/2016

OPERATIONMANUAL

5.1MaintenanceandInspections
Ourcompletequalitycontrolandinspectionprocedures
ensurethatthesystemistransportedinoptimalconditions.For
maintainoptimalconditions,weneedtoimplementandmaintain
inspection.
Themaintenanceandinspectionsforthesystemareclassifiedas
followbelow:

Typeofmaintenance/
inspection
startingthe
inspections inspections
daily
finalizingthe
inspections
Cleaninganddisinfection
periodicinspections

implementedby
Operatorormanager
equipment
Operatorormanager
equipment
Operatorormanager
equipment
Technicalassistance
authorized

Interval
Daily
Daily
Whennecessary
Every6months

5.2Dailyinspections
Dailychecksincludethebeginningandendofinspections.
Dailyinspectionsarenecessarytomaintainthesystemforalong
period.
Checkallinspectionsandcorrectiveactionshavebeencompletedbefore
tousethesystem.

106

Page107

https://translate.googleusercontent.com/translate_f

104/135

11/28/2016

OPERATIONMANUAL

ReferringtoFig.
5.3Cleaninganddisinfection

Youwillneedtocleananddisinfectthesystem.
Cleaninganddisinfectionareveryimportanttoensurethatthe
systemcanbeusedsafely.Youwillneedtofollowthemethods
indicatedbythemanufacturer

1Makesurethatthesystem'smainswitchisturnedoffanddisconnectthe
maincable.

Unplugthemaincablebeforeyoustartcleaning.
Youmustdisconnectthesystemfrommainsbeforestarting
cleaninganddisinfection.
Turnoffthesystembeforedisconnectingthemaincable.
Otherwisethereisariskofelectricshockandleakagecurrent.

2Performcleaninganddisinfection.
Cleantheoutersurfaceofthesystemwithadampcloth
glutaraldehydeina2%solution.

Donotusewateronthesystem.
Cleanthesystemwithacleanclothsoakedwithdisinfectant.

107

Page108

https://translate.googleusercontent.com/translate_f

105/135

11/28/2016

OPERATIONMANUAL

Donotspraydisinfectantdirectlyonthesystem.
Ifthedisinfectantintothesystemmaycauseproblemsor
accidents.

Donotusethebelowlistedproducts
Disinfectantschlorinebase
Disinfectantsthatcorrodemetals,plastics,rubbers,etc.
Disinfectantsatomized
volatileDisinfectants

Usetheminimumamountofdisinfectant
Overtime,thedisinfectionmaycausedamageto
paintdiscolorationordeteriorationofresinssuchasplasticand
rubbers.

3Aftercompletingthework,checkthefollowingpoints:
Thereshouldbenowaterordisinfectantjoinedthesystem.
Thetoolsusedtocleananddisinfectthesystemshouldbe
cleanafteruse.

108

Page109

https://translate.googleusercontent.com/translate_f

106/135

11/28/2016

OPERATIONMANUAL

Chapter6

optional

Thischapterexplainshowtooperatetheoptional(operation,maintenance,
setting,etc.).

109

Page110

6.1LaserIndicator(Laserpointer)
ThelaserconsistsofaprojectorunitintheXraytube,plate
withcentralidentificationintheimageintensifierandtheON/OFF
https://translate.googleusercontent.com/translate_f

107/135

11/28/2016

OPERATIONMANUAL

ofthelaserbeamonthetopside.

1Press"LaserPointer"onthemainpanelinitiatestheissueof
laserbeam.(Lamplightsuponthedashboard)
Alaserbeamisemittedwhileitappearsa
markingacrossonthemonitorscreenimage.
Thebuttonisilluminatedwhenactive.
Whenyoupressthebuttonagain,thelaserbeamis
interrupted.

2Pressingthesamekeyagainthelaserbeamisturnedoff.
Markingimagecrossdisappearsfromthemonitor.
Thelaserbeamcontinuestobeoutputwhilethebuttonis
on.

Figure6.1LaserPointer(IndicatorLaser)

110

Page111

Neverlookdirectlyintothelaserbeam.

https://translate.googleusercontent.com/translate_f

108/135

11/28/2016

OPERATIONMANUAL

laserradiationinfo:
Power:1.0mWmax.
Powerconsumption:5V200mA
CW:635nm
Class2(IEC8251)

Thisoptionwasdevelopedtomarktheareatobeexposedto
xraysmoreaccurately,soyoucanoptimizetheamountof
XraysirradiatedwhileavoidingdoseofXraysunnecessaryforthepatientand
medicalteam.

AdditionalInformation:

Thevaluesandindicationscanbeobservedinthemachinescreenor
onthemonitors.

111

Page112

6.2CassetteSupport
Youcanmountacassetteholderintheintensifierunit
Image.

https://translate.googleusercontent.com/translate_f

109/135

11/28/2016

OPERATIONMANUAL

Installthecassetteholdercorrectlyforitnot
presentlaterfailures.
Adjustingthecassetteholder
1Unscrewthescrew(1).
2Lettherunner(2)contactastheflangemounted(3),push
thecassetteholderforwardandlockthefixedguide(4)mountedontheflange
(3).
3Tightenthescrew(1)untilthetipcontactface
rearoftheblock(5).
Inthispositionthecassetteholdercanbemountedanditwill
rotatefreelyintheappendixoftheintensifierunit
image.
4Adjustthecassetteholderpositionaccordingtotheposition
Radiographic.

Figure6.2CassetteSupport

112

Page113

6.3HandTrigger
Thison/offcontroltheradiationXrays.
1Pressthehandshutterbuttontothefirstlevel(level
Radiographicpreparation).
Startspreparationforradiography.
https://translate.googleusercontent.com/translate_f

110/135

11/28/2016

OPERATIONMANUAL

TheStandbyindicatorlampequipmentwillturnoff
temporarily.

Whentheradiographicpreparationiscomplete,thelamp
standbyindicatorwilllightupagain.
2Pressthehandcontroltothesecondlevel(levelofpreparation
RadiographicandexposuretoXrays).
TheXrayradiationwillstart.

DuringexposuretoXrays,theradiationmonitorindicator
willlight.

113

Page114

https://translate.googleusercontent.com/translate_f

111/135

11/28/2016

OPERATIONMANUAL

Chapter7

Technicaldata

114

Page115

7.1TechnicalSpecifications

7.1.1)CararmCsupport
Item
FocusII
verticaloffset
horizontalscroll
axialrotationofthearc
Arcrotation
lateralmovement
https://translate.googleusercontent.com/translate_f

OpescopeActive
900mm
450mm
200mm
120th
420
12.5

FullOpescope
900mm
450mm
200mm
120th
420
12.5
112/135

11/28/2016

OPERATIONMANUAL

freeareabetweentube/II
DepthofArc

700mm
780mm(9II")

700mm
780mm(9II")

7.1.2)Unitcontroller/generatorxray
Item
Ratedpower
voltagerange(RAD/Fluoro)
mAsrange(RAD/Fluoro)
Maximumcurrent(RAD)
Maximumcurrent(FLUORO)
pulsedfluoroscopy
fluoroscopycontinues
FluoroscopyHighPulsed
Quality(BOOST)
fluoroscopytime
tobecontinued
Inverterfrequency

OpescopeActive
FullOpescope
2kW
2kW
40to110kVin1kVstep40to110kVin1kVstep
0.5to27insteps200mAs
0.5to27insteps200mAs
20mA
20mA
16mA
16mA
0.5/1/2/3.75/7.5and15fps0.5/1/2/3.75/7.5and15fps
Yes
Yes
Yes

Yes

45minutesormore

45minutesormore

50kHz

50kHz

7.1.3)tubexray
Item
caloriccapacityoftheanode
dissipationcapacity
Callusanode
caloriccapacity
set
SpotSize
AnodeAngle
anodematerial
Typeanode
feeding
inherentfiltration

OpescopeActive
100.000HU

FullOpescope
100.000HU

550W

550W

680.000HU

680.000HU

0.6mm2
8th
Tungsten
Fixed
Highfrequency
2,9mmAl

0.6mm2
8th
Tungsten
Fixed
Highfrequency
2,9mmAl

115

Page116

7.1.4)Collimator
Item
collimatoroctagonal
collimatedparallel
compensatingfilters

OpescopeActive
Motorizedwithbladesof
Lead(Pb)
Motorizedwithbladesof
Copper(Cu)andsystem
rotation.
Motor.Coppermade
(Cu)andthicknesses
0.1mm,0.2mmand0.3mm

FullOpescope
Motorizedwithbladesof
Lead(Pb)
Motorizedwithbladesof
Copper(Cu)andsystem
rotation.
Motor.Coppermade
(Cu)andthicknesses
0.1mm,0.2mmand0.3mm

7.1.5)Intensifierimages(II)
https://translate.googleusercontent.com/translate_f

113/135

11/28/2016

OPERATIONMANUAL

Item
Intensifier6Image"/4"
Intensifier7Image"
Intensifier9Image"/6"
Item
Sizeoffields
centralResolution
maxim

OpescopeActive
Yes
Optional
Optional
II6"/4"
6"/4"(16cm/10cm)

FullOpescope

Yes
II7"
7"(18cm)

54/64

50

II9"/6"
9"/6"(23cm/16cm)
50

7.1.6)ImageProcessingUnit
Item
Cameratype
acquisitionmatrix
Resolution
fluoroscopyacquisitionrate
serialacquisitionrate

Imageprocessing

OpescopeActive
CCD400kP
768x494(EIC)or752x582
(CCIR)
8bits
30frames/second
5frames/second

FullOpescope
CCD1MP
1024x1024

12bits
30frames/second
15frames/sec
windowsettings/level,brightness
andcontrast,gamma,highlight
windowsettings/level,brightnessedges,zoom/pan2xzoom,
andcontrast,gamma,highlight displayingmultipleimages
edges,zoom/pan2xzoom,
anascreenannotation,measurements,
displayingmultipleimages
SubtractionAngiography
anascreenannotation,measurements,
Digital(DSA)inrealtime,
SubtractionAngiography
remascara,displacement
Digital(DSA)inrealtime,
pixel(pixelshift),landmarking
remascara,displacement
(Anatomicalreference)
pixel(pixelshift),landmarking
positive/negativeinversion,
(Anatomicalreference)
right/leftinversion,
positive/negativeinversion,
reversingup/downmode
right/leftinversion,
roadmappingcineandRSM
reversingup/downmode
DSA(angiography
cineandroadmapping.
DigitalSubtractionintime
realtimewithmask
real).

116

Page117

AcquisitionofDSA
Capacityof
imagestorage

DICOM3.0protocol

0.5to5frames/second

0.5to7.5frames/second
Upto30,000imagesonthedisc
Upto16,770imagesonthedisc
hard
hard
Upto100imagesinmemory
Upto68imagesinmemory
RAM
RAM
Upto300images/CDR
DICOM3.0format
DICOMPrint
DICOMStorage(send/receive)

DICOMWorklist
DICOMMediaStorage

7.1.7)Monitor
Item
LCDMonitor18"
LCDmonitor18"touchscreen
https://translate.googleusercontent.com/translate_f

OpescopeActive
Optional
Optional

FullOpescope
Optional
Optional
114/135

11/28/2016

OPERATIONMANUAL

LCDMonitor15"
CRTMonitor17"
LCDMonitor19"
Monochrome
LCDmonitor19"touchscreen
LCDMonitor19"color

Optional
Yes

Optional
Optional

Optional

Yes

Optional
Optional

Optional
Optional

7.1.7)antidifusoraGrade
Item
Gridreason
griddensity
Interespaador

OpescopeActive
10:1or8:1
40lines/cm
Aluminum(Al)orfiber

FullOpescope
10:1or8:1
40lines/cm
Aluminum(Al)orfiber

Note:Thetechnicalinformationsuchascircuitdiagrams,lists
components,descriptions,instructionsforcalibrationandtestingareprovided
bythemanufacturertostaffand/orqualifiedandauthorizedtechnicalassistance
Brazil.

117

Page118

https://translate.googleusercontent.com/translate_f

115/135

11/28/2016

OPERATIONMANUAL

Figure7.1xraytube

Subtitle:
Referenceaxistheaxisofreference
Focalspotfocalpoint
UnitUnit

118

Page119

https://translate.googleusercontent.com/translate_f

116/135

11/28/2016

OPERATIONMANUAL

Figure7.2Heatingandcoolingcurvesofthexraytubeanodeand
Subtitle:
XRaytubeassemblyheatingandcoolingcurveheatingandcoolingcurve
xraytube
AnodeHeatingandCoolingCurveheatingandcoolingcurveoftheanode

119

Page120

https://translate.googleusercontent.com/translate_f

117/135

11/28/2016

OPERATIONMANUAL

Figure7.3Loadevaluationboardsandcharacteristicsofemissionandfilament
Subtitle:
SingleLoadRatingChartsApplicationAssessmentFrameworksinasinglecharge
AndFilamentEmissionCharacteristicsemissioncharacteristicsandfilament
XrayTubecurrentTubecurrentxray
TimeTime
FilamentCurrentFilamentCurrent
FilamentVoltageFilamentvoltage

120

Page121

Monitor
Kind
LCDMonitor18"

LCDmonitor18"touchscreen

LCDMonitor15"

CRTMonitor17"

https://translate.googleusercontent.com/translate_f

Specification
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFT
ScreenSize:18"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFTandtouch
ScreenSize:18"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFT
ScreenSize:15"(diagonal)
ScreenMatrix:1280x1024
ScreenType:CRT
ScreenSize:17"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
118/135

11/28/2016

OPERATIONMANUAL

LCDMonitor19"Monochrome

LCDmonitor19"touchscreen

LCD19"ColorMonitor

TFT
ScreenSize:19"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFTandtouch
ScreenSize:19"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFT
ScreenSize:19"(diagonal)

Videotape
Videosignal:NTSCstandardcolorsignal
Power:AC120V,60Hz,0.6A
Videosignal:
Input:BNC1.0Vpp,75ohms.Sin(4P)Y:1.0Vpp,75ohms.W:
0.286Vpp,75ohms.
Output:BNC1.0Vpp,75ohms.Sout(4P)Y:1.0Vpp,75ohms.W:
0.286Vpp,75ohms.
Audiosignal:
Input:LINEINCH1/CH2(PINRCA)10dBu,47kohms.Microphone

121

Page122

NormalaudioCH1(RCAJackQ)to600ohmmic60dBu,3kohms.
Output:AUDIOMONITOR(RCAPIN)5dBu,47kohm.LINEOUTCH
1/CH2(RCAPIN)10dBu,47kohm.
tapeformat/speed:SVHS/VHS/33.35mm/sec

VideoColorPrinter
Resolution:403DPI
PrintedImagesize:(NTSC)3"x27/8"/5"x35/8"
Printspeed(NTSC)1929seconds
Powersupply:100120volts,1.8A220240volt,1.0A

VideoPrinterBlackandWhite
Resolution:1280DOT
Effectivepixels:508linesx1280dots
PrintSpeed:3.07.0onthe2nd
https://translate.googleusercontent.com/translate_f

119/135

11/28/2016

OPERATIONMANUAL

Powersupply:100120volts,1.8A220240volt,1.0A

optional
TapeSupport
IndicatorLaser
SterilisationCase
HandControl
LCDMonitor
VideoSonyPrinter
DVDrecorderSony
ExtensionDistanceSourceSkin

OptionalOPESCOPEFULL
LCDMonitorTouch
DICOM3.0Kit(Print(IncludesCDROMwithsoftware)/Storage(send/receive)
(IncludesCDROMwithsoftware)andWorklist(IncludesCDROMwithsoftwareandPC,
consistsofMonitor,CPU,mouseandkeyboard))

122

Page123

FluoroimageinSupportCarDisplaywithCArm
Dosemeter(DAP)

OptionalOPESCOPEASSETS
VariableField(9"/6")toII
Fibergrid
HandleControlII(ImageIntensifier)
TypeMemory"LOOP"
LCDMonitor6.4"
DigitalImageProcessingUnitwithSubtractionAngiography
Digital(DSA)
MODRecorder
MODMedia

WeightCarHolderwithArmC:250kg
WeightCartMonitors:145kg
https://translate.googleusercontent.com/translate_f

120/135

11/28/2016

OPERATIONMANUAL

Thecontrollerunitofxrayconstitutesthejunctionhousing(outsideand
xraygenerator)

TheCarchsupport,supportxraytube,xraytube,collimator,
highvoltagegenerator,imageintensifier,powersupplyand
controllerunitxraysareofexclusiveuseoftheequipment.

Thelaserpoint,monitors,standandsupportDICOMkitaremanufacturing
Shimadzu,buttheymaybeusedinotherShimadzuequipment.

ThevideoprinterandvideocassettearenotmanufacturedbyShimadzu,
gettingthepurchasethecustomer'sdiscretion.(Seeminimumspecifications)

graphicillustrations

123

Page124

Figure7.4xraytube

https://translate.googleusercontent.com/translate_f

121/135

11/28/2016

OPERATIONMANUAL

Figure7.5Collimator

124

Page125

Figure7.6highvoltagegeneratorphotos

https://translate.googleusercontent.com/translate_f

122/135

11/28/2016

OPERATIONMANUAL

Figure7.7imageintensifierPhoto

125

Page126

Figure7.8Monitor

https://translate.googleusercontent.com/translate_f

123/135

11/28/2016

OPERATIONMANUAL

Figure7.9Figurepowersupplyboard
Subtitle:
2Baseboard
3Limiter
4Label
5Spacers

7Limiter
8,9,10,11and12Powersupply
13Filter
14Label

126

Page127

Figure7.10Laserpointillustration

https://translate.googleusercontent.com/translate_f

124/135

11/28/2016

OPERATIONMANUAL

Figure7.11StandformonitorPhoto

127

Page128

https://translate.googleusercontent.com/translate_f

125/135

11/28/2016

OPERATIONMANUAL

Figure7.13SupportCarBoomCwith9II"

128

Page129

https://translate.googleusercontent.com/translate_f

126/135

11/28/2016

OPERATIONMANUAL

Figure7.14SupportCarBoomCwith7II"

129

Page130

Figure7.15SupportCarBoomCIIwith6"/4"

https://translate.googleusercontent.com/translate_f

127/135

11/28/2016

OPERATIONMANUAL

130

Page131

7.2ConditionsforInstallationandUse
1.EnvironmentalConditions:
Temperature

10Cto40C

Relativehumidity

30%to85%(noncondensing)

Atmosphericpressure

700hPato1060hPa

Atmosphereconditions

Theatmosphereshouldbegasfree
corrosiveandexplosive

2.SourceofEnergy:
Equipment

SimplephaseAC

Frequency

50or60Hz

Capacity

3.5kVA

VoltageandImpedance

100,110,120V10%0.2Max.
200,220,230,240V10%0.8Max.

Earth

typeD

3.PowerConsumption
Inuse:

35A(100,110,120V),17,5A(200.220,230,240V)

standby:

15A(100,110,120V),10A(200.220,230,240V)

Installation
Installationoftheequipmentshouldbeperformedonlybypersonnel
and/ordealer.

https://translate.googleusercontent.com/translate_f

128/135

11/28/2016

OPERATIONMANUAL

AuthorizedTechnicalAssistance
ShimadzuofBrazilLtda
Address:MarquesAvenueStVincent,1771SP
Tel.:(11)21341688/Fax:(11)36112209
email:webmaster@shimadzu.com.br
website:www.shimadzu.com.br

131

Page132

7.3ConditionsforStorage
Temperature

10Cto40C

Relativehumidity

30%to85%(noncondensing)

Atmosphericpressure

700hPato1060hPa

Atmosphereconditions

Theatmosphereshouldbegasfree
corrosiveandexplosive

7.4DoseInformation

7.4.1LocationofSignificantZonesofOccupation

ArcCposition
TheforwardverticalpositionseeFigure7:16
AnteroposteriorverticalpositionseeFigure7:17
HorizontalpositionseeFigure7:18
SignificantzoneofoccupationandmeasurementpointsseeFigures
7:16,7:17and7:18

Fantom
Externaldimensions:35x35x15(cm)

https://translate.googleusercontent.com/translate_f

129/135

11/28/2016

OPERATIONMANUAL

132

Page133

https://translate.googleusercontent.com/translate_f

130/135

11/28/2016

OPERATIONMANUAL

Figure7.16PositionposteroanteriorVertical(measuredinmm)

133

Page134

https://translate.googleusercontent.com/translate_f

131/135

11/28/2016

OPERATIONMANUAL

Figure7:17PositionAnteroPosteriorVertical(measuredinmm)

134

Page135

https://translate.googleusercontent.com/translate_f

132/135

11/28/2016

OPERATIONMANUAL

Figure7.18Horizontalposition(measuredinmm)

135

Page136

Heightfromtheground

Figure7.19posterioranteriorpositionvertical(medicalcm)

https://translate.googleusercontent.com/translate_f

133/135

11/28/2016

OPERATIONMANUAL

Heightfromtheground

Figure7.20anteroposteriorverticalposition(measuredincm)

136

Page137

Heightfromtheground

Figure7.21Horizontalposition(measuredincm)

RepresentativeandTechnicalAssistanceinBrazil
ShimadzuofBrazilLtda
AvenueMarquesdeSoVicente,1771BarraFunda
SoPauloSPCEP:01139003
Tel.:(11)21341688
Fax:(11)36112209
email:marketing@shimadzu.com.br

https://translate.googleusercontent.com/translate_f

134/135

11/28/2016

OPERATIONMANUAL

______________________
Technicalmanager
ConstantineLuizDiPIPI
CREA/SP0682526854

________________________
Legalresponsible
AkihikoKato
CEO

137