The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study
Megan E. Sullivan, B.S., R.T.(T)., Patrick A. Melby, B.S. Ashley Hunzeker, M.S., CMD,
Nishele Lenards, M.S., CMD, R.T. (R)(T), FAAMD
Medical Dosimetry Program at the University of Wisconsin - La Crosse, WI
Abstract:
Introduction: Right-sided breast cancer procedures may not incorporate gating techniques
because of the extra set-up complications. This study aims to evaluate the effects of deep
inspiration breath hold (DIBH) on the dose received by the liver.
Case Description: Seven, right-sided breast patients were retrospectively evaluated for dose to
the liver. These patients had both DIBH and free-breathing (FB) gated scans taken before
treatment. Plans were created for both scans and were evaluated for volume within the field,
maximum dose, mean dose, and the volume of liver irradiated at specific dose levels (Vn).
Conclusion: The use of a DIBH gated scan reduced the volume of liver being treated during
right-sided breast treatments. The average volume in the FB scans was 39.7 cc and the DIBH
scans reduced the irradiated volume to 3.9 cc. Deep inspiration breath holds showed benefits to
the liver for all patients within this study. This included lower maximum and mean doses along
with reducing the amount of liver exposed to 5, 10, 20, and 30 Gy.
Introduction
Radiation therapy has been used to treat breast cancer for over 40 years. The results have
shown to be very beneficial in terms of cancer treatment, along with the total survival rate.
During treatments, cancer tissues and normal tissues within the treatment field are irradiated.
With increasing technology, techniques have been developed to increase the accuracy of
treatments and reduce the exposure to surrounding areas. A relatively new technique includes
the use of a four-dimensional computed tomography (4DCT), using a DIBH protocol.
The DIBH technique is one method of treatment using a gated scan. Gated scans consist
of capturing CT images during specific times of a patients breathing cycle. This technique
creates a consistent breathing interval to conduct a treatment that has far less motion compared to
treating without it. Without a method of gating, the chest can fluctuate 0.5 to 0.7 cm laterally
throughout the treatment.1During DIBH treatments, the gate at which the CT scan is acquired is
set to when the patient is at full inspiration.
Aside from motion reduction during treatment, the DIBH technique offers benefits for the
patient anatomically. As the patient holds a deep inspiration, the lung volume is increased. With
this increase, the electron density within the lung is decreased, reducing the exposure to the lung

while increasing the distance between the treatment fields from organs at risk (OR) within the
area.2 This separation has been shown to be beneficial for left-sided breast treatments because
the heart is moved away from the treatment field. This reduces unnecessary exposure to the
coronary arteries. Although the heart is usually not a concern during right-sided breast
treatments, the right-sided DIBH treatment technique can be advantageous to other organs.
Right-sided DIBH treatments are less common. Other gating techniques such as FB or no
gating are typically used for right-sided treatments. The reason for using these techniques is
because DIBH treatments can be difficult to set-up and execute if the patient has a difficult time
holding their breath. The benefit of sparing the heart has been the main focus for DIBH
treatments, but the advantage of moving the heart away from the irradiated field does not
translate for the treatment of right-sided breast cancer cases. However, if DIBH techniques are
used for right-sided breast treatments, other OR, specifically the liver, are moved away from the
radiation field. Butler et al1 reported that during normal respiration, the liver moved 1.5 to 2 cm
in the superior-inferior direction and during DIBH the distance was further increased to 3.9 cm.
The extra movement in the superior-inferior direction during DIBH may force the liver away
from the radiation field to significantly reduce additional radiation being deposited into the liver.
The suggested dose tolerances for the liver is a mean dose of less than 30 Gy, which
provides a 5% chance of radiation induced liver disease (RILD).3 The tolerance is intended for
healthy livers, which means patients with preexisting liver disease such as hepatitis or poor liver
function may have a greater risk of side effects.4 The concern for treating patients with
preexisting liver disease grows as comorbidity becomes increasingly prevalent. It is estimated
that 1 in 3 adults and 1 in 10 children in America have nonalcoholic fatty liver disease
(NAFLD).5 This occurrence has increased significantly due to the increase in obesity, type II
diabetes mellitus, and physical inactivity. The excess fat within the liver may cause hepatitis,
which may then form scar tissue within the liver, causing a decrease in liver function. With a
greater possibility for future patients presenting with decreased liver function, it is of best
practice to spare this organ from unnecessary radiation during right-sided breast treatments.
Another reason to reduce the excess radiation dose to the liver is because patients are
being treated at a younger age. The advances in screening have allowed the detection and
treatment of breast cancer earlier. Since patients are being treated at a younger age, there is more
time to develop a secondary cancer in areas that received a low dose of radiation. The purpose of
this study was to determine if DIBH would decrease the amount of radiation to the liver for

right-sided breast cancer patients and thus reducing the risk of radiation induced liver
complications.
Case Description
Patient Selection and Setup
Seven, previously treated patients with breast cancer of the right-breast and regional
lymph nodes were selected. All of these patients received radiation therapy with the DIBH
technique to minimize lung dose. During CT simulation, patients were scanned using both the
FB and DIBH techniques separately. Patients were positioned supine on an in-house, inclined
breast board with both arms raised and the heads turned towards the unaffected breast. Radioopaque wires were placed by the physician to delineate superior, inferior, medial, and lateral field
borders as well as breast tissue and lumpectomy scars. A Philips Brilliance Big Bore CT scanner
was used to acquire the FB and DIBH scans with a 3.0 mm slice thickness. The FB scan was
acquired first. For the DIBH scans, the patients were gated with Real-time Position Management
(RPM) system and were coached with breathing instructions. Before the DIBH scan began, the
patients were expected to reproduce a consistent breath hold in order to ensure reproducibility for
treatment. The expected time for patients to hold their breath for this scan was about 20 seconds.
Upon acquisition of the scans, the CT data sets were transferred to the Eclipse Treatment
Planning System (TPS) via the Digital Imaging and Communications in Medicine (DICOM)
network.
Target Delineation
The target volumes including the breast tissue, lumpectomy bed, axillary I, II, and III
lymph nodes, supraclavicular (SC) lymph nodes, and in some cases, internal mammary (IM)
nodes were contoured by the physician and outlined on both the FB and DIBH scans. The breast
tissue was cropped 5 mm away from the skin surface. Organs at risk were defined by a certified
medical dosimetrist and included the following: right and left lungs, heart, spinal cord,
esophagus, and liver.
Treatment Planning
The prescription was written for 50.4 Gy in 1.8 Gy per fraction to be delivered daily.
The target volumes included the breast tissue, lumpectomy bed, axillary I, II, and III lymph
nodes, SC lymph nodes, and in some cases, IM nodes. The planning goals for the breast tissue,
lumpectomy bed, axillary I, II, and III lymph nodes were to achieve at least 95% of the volume
to be covered by at least 95% of the prescribed dose. In cases involving the IM nodes, the goal
was to achieve 90% of the volume to be covered with 90% of the prescribed dose. A

monoisocentric technique was used for all patients. A Varian 21EX linear accelerator equipped
with 6 and 23 MV energies was available to treat DIBH breast plans. Left anterior oblique
(LAO) and right posterior oblique (RPO) treatment angles and multi-leaf collimator (MLC)
blocks for the SC fields were created based on obtaining coverage of the nodal chains and the
need to block the spinal cord, esophagus, and humeral head. An enhanced dynamic wedge was
used as needed and when field size limitations allowed. For tangent breast treatment fields, LAO
and RPO angles were used. Fields were shaped with MLC blocks based on the borders defined in
CT simulation and the ability to block normal structures such as heart and liver without
compromising coverage of the breast tissue. When helpful, a field-in-field technique was used to
block out unnecessary areas receiving a dose greater than the prescription dose. Physical wedges
were used on tangent fields as needed. The most common use was on an RPO breast field with a
30 degree right physical wedge. Figure 1 displays the image registration of the FB and DIBH
scans and Figure 2 displays a beam's eye view (BEV) for a patient. The plans used to treat the
patients were created on the DIBH scans by certified medical dosimetrists.
Plan Analysis & Evaluation
For the actual treatments, plans were generated on the DIBH scans by certified medical
dosimetrists. For research purposes, retrospective planning was performed on the FB scans and
compared against the plans created on the DIBH scans to assess the dose to the liver for each
patient. The DIBH plan was copied onto the FB scan and adjustments were made as necessary to
generate a plan tailored to the FB scan while simultaneously keeping the plan as similar to the
DIBH plan as possible. Plans were normalized appropriately to obtain the same target coverage
and evaluate comparisons fairly. Dose volume histograms (DVH) were compared and evaluated
between the FB and DIBH scans as shown in Figure 3. All planning goals remained consistent
for each patient regardless of the structure set.
One of the first areas that was analyzed between the FB and DIBH plans was the
volume of liver within the target field of treatment. The quantities were determined by
contouring the overlap of the field edge with the liver as seen in the axial plane in each plan. The
average volume of liver within the treatment field for the FB plans was 39.7 cc, whereas the
average for the DIBH plans was 3.9 cc. This comparison shows that DIBH decreased the
volume of liver being exposed by primary radiation by 90.2%. For 4 of the patients, DIBH plans
completely eliminated any primary radiation delivered to the liver as seen in the Table 1. Visual
differences of liver position between a FB scan and a DIBH scan can be seen Figures 1 and 2.

Upon determining that DIBH helped to decrease the amount of liver being exposed to
radiation, the amount of liver exposed to specific dose levels was then investigated. The dose at
each of these volumes decreased from the FB to the DIBH plans. This was expected since the
overall volume of liver irradiated was also decreased. The average V30, V20, V10, and V5 for the
FB plans were 37.5 cc, 46.7 cc, 62.6 cc, and 105.4 cc, respectively (Table 1). The average V30,
V20, V10, and V5 for the DIBH plans were 2.2 cc, 3.9 cc, 7.0 cc, and 18.8 cc, respectively (Table
1). The FB scan had exposed almost twice the amount of liver to 30 Gy than the DIBH plan
exposed the liver to 5 Gy.
The last aspect evaluated between the two plans was the maximum and mean doses to the
liver. The maximum dose for the FB scans had a range between 47.3 Gy and 52.24 Gy. The
DIBH had maximum doses that ranged between 7.33 Gy and 48.8 Gy. The mean dose also
followed a similar trend with a range of 1.4 Gy to 5.57 Gy for the FB plans and 0.66 Gy to 2.00
Gy for the DIBH plans.
The factors that were evaluated in this study, including the liver volume within the
treatment field, dose volumes, mean dose, and maximum dose, were decreased in all of the
patients by the DIBH plans. The DIBH plan had a greater effect in some patients than in others.
Some variables contributing to the DIBH radiation reduction include target location, liver
location, or the effectiveness of the patients breathing.
Conclusion
For right-sided breast patients with compromised liver function, it may be beneficial to
take additional measures like DIBH to avoid treating the liver with unnecessary radiation. The
additional setup difficulties should be considered when deciding whether to use DIBH or not.
By evaluating the amount of liver being treated, along with the doses at specific volumes of the
liver, the effects of DIBH compared to FB were able to be evaluated.
The results of the current study demonstrate that DIBH techniques can reduce the
amount of radiation received by the liver. This is one of the first studies that has taken an indepth look at liver dose using DIBH techniques for multiple patients. Rice et al4 reported on an
evaluation of a single case study that calculated the dose received by the liver for both the DIBH
and FB plans and yielded similar results. For a single case, the DIBH technique reduced the
mean and maximum dose by 46 and 3.5%, respectively. The V10, V20, and V30 were all decreased by
over 50% to show an overall benefit to DIBH.

The reduction of unnecessary dose to the liver may help to prevent cases of RILD. One
case of RILD has been documented to be caused by a treatment of the right chest wall.6 The
RILD may progress to liver fibrosis and liver failure. This shows that a reduction of liver dose
may be of importance in certain patients.
The limitations of this study are primarily found in the patient sample size as it is
reflective of one institution within a one year time span. The scans that were available for this
retrospective study were from patients treated with the DIBH technique in efforts to reduce dose
to the lung. As reduction in lung dose was the original motive for the implementation of the
DIBH technique, only patients with nodal involvement were treated using DIBH. For this reason,
two-field breast plans were not evaluated in this study. Another limitation of this study that
should be noted is in reference to the resources needed to implement DIBH. The DIBH
technique is a more labor intensive process than the FB technique for both the patient and the
institution. When compared to a FB technique, DIBH plans require an additional CT scan, some
form of gating system, trained staff, ability of patients to hold breath for at least twenty seconds
at a time, and additional time on the treatment machine. Additional research should be
conducted on the extent of damage done to unhealthy livers by radiation. Overall, there was an
improvement in liver dose reduction with the DIBH technique, however, it was evident that some
patients benefited from the DIBH technique more than others. For this reason, the labor intensive
process may not offer a significant improvement in liver dose metrics for every scenario and thus
should be evaluated case by case.

References
1. Butler LE, Forster KM, Stevens CW, et al. Dosimetric benefits of respiratory gating: a
preliminary study. J Appl Clin Med Phys. 2004;5(1):16-24. https://doi:10.1120/5.1.03.
2. Bruzzaniti V, Abate A, Pinnar P, et al. Dosimetric and clinical advantages of deep inspiration
breath-hold (DIBH) during radiotherapy of breast cancer. J Exp Clin Cancer Res. 2013;32(1):17. https://doi:10.1186/1756-9966-32-88.
3. Emami B, Lyman J, Brown A, et al. Tolerance of normal tissue to therapeutic irradiation. Int J
Rad Oncol Biol Phys. 1991;21(1):109-122. http://doi:10.1016/0360-3016(91)90171-y.
4. Rice L, Harris S, Green M, Price P. Deep inspiration breath-hold (DIBH) technique applied in
right breast radiotherapy to minimize liver radiation. BJRcase reports. 2015;1(2):1-4.
https://doi:10.1259/bjrcr.20150038.
5. Lpez-Velzquez JA, Silva-Vidal KV, Ponciano-Rodrguez G, et al. The prevalence of
nonalcoholic fatty liver disease in the Americas. Ann Hepatol. 2014;13(2):166-178.
http://www.annalsofhepatology.com/revista/numeros/2014/03_142_v13n2_2014_PrevalenceNon
alcoholic.pdf
6. Khozouz R, Huq S, Perry M. Radiation-Induced Liver Disease. J Clin Oncol.
2008;26(29):4844-4845. http://doi:10.1200/jco.2008.18.2931.

Figures

Figure 1. Image registration of FB and DIBH scans with treatment fields displayed on a patient
in the coronal plane (left) and in the sagittal plane (right).

Figure 2. Beams eye view of medial tangent field on FB scan (left) and DIBH scan (right) for a
patient. The breast contour is shown in green and liver contour is shown in brown.

10

Figure 3. Dose volume histogram in absolute dose and absolute volume for a patient. Breast
coverage is shown in magenta and liver coverage shown in brown.

11

Table 1. Liver dose metrics comparing FB and DIBH techniques for all patients (Pt).

Pt 1
Pt 2
Pt 3
Pt 4
Pt 5
Pt 6
Pt 7
AVG

Volume in
field (cc)
DIB
FB
H
16.
0
3
17.
0
3
85.
12.4
7
71.
6.9
8

Mean Dose
(Gy)
DIB
FB
H
2.8
1.43
7
1.8
0.66
6
5.5
2.0
7
5.0
2.0
9

2.3

1.4

0.69

8.3

2.5
6

1.3

3.3

1.53

3.9

3.2
4

1.38

49.
1
35.
3
39.
7

Max Dose
(Gy)
DIB
FB
H
49.1

10.88

49.1

7.33

50.2

48.8

43.6
4

41.48

47.3

6.97

52.2
4
51.3
1
48.9
8

47.1
37.32
28.56

V30 (cc)
FB
15.
8
11.
4
85.
0
65.
5
3.7
2
44.
5
36.
3
37.
5

V20 (cc)

V10(cc)

V5 (cc)

DIB
H

FB

DIB
H

FB

DIB
H

FB

DIB
H

21.3

30.2

59.9

7.4

16.4

24.0

38.9

6.2

102.
5

11

130.
6

19.1

206.
4

40.8

3.14

79

5.6

99.1

10.5

123

19.8

7.1

13.4

35.3

1.1

6.16

55.0

10

72.8

16.8

46.9

0.5

69.2

2.87

2.2

46.7

3.9

62.6

7.0

124.
4
150.
0
105.
4

31.5
31.2
18.8

Potrebbero piacerti anche