Sei sulla pagina 1di 1

Olympus NDT Canada

505, boul. du Parc-Technologique


Qubec (Qubec) G1P 4S9
Canada

Tel. : (418) 872-1155


Fax. : (418) 872-5431
Web site : www.olympusndt.com

EDDY CURRENT: ECT EXTENDED ASME CALIBRATION TUBE


RECOMMENDED CALIBRATION TUBES
The CT02 ECT calibration tube below provides every defect that is required to perform a good ECT calibration and
build strong phase-to-depth curves. In fact, this tube is not only conforming to the ASME standards. In addition, it
also provides wall thinning defects in order to help users to perform a more accurate calibration.

DIFF (pits)
SP01-XXX
GR10
(internal) GR20

H100a

H80

H60

H40

H20

H100b

GR40

H100b

GR40

GR60

ABS (wall thinning)

GR10

GR20

H100a

H80

H60

H40

H20

GR60

ARTIFICIAL DEFECT PARAMETERS IN FUNCTION OF TUBE WT AND OD


Defect

Depth (%WT)

Diameter

Defect

Depth (%WT)

Diameter

GR10

10 % (internal)

H80

80 %

0.078

GR20

20 %

H60

60%

0.125

GR40

40 %

H40

40%

0.188

GR60

60 %

H20

20%

0.188 (4 holes)

Thru

If OD>0.75, 0.125
Else: 0.078

H100a

If OD>0.75, 0.067
Else: 0.052

Thru

H100b

VARIABLE CHART IN FUNCTION OF TUBE OD


Value

Value

Variable

Value

Variable

Variable

in

mm

in

mm

in

mm

1.25

32mm

3.00

76mm

TL

20.50

521mm

1.50

38mm

GLa

1/16

1.6mm

SP_OD

OD + 3/4"

OD + 19mm

3.50

89mm

GLb

1/2"

13mm

SP_TCK

3/8

10mm