TERUMO TE331 - TE332 -Terufusion Syringe Pump - E2PROM

INITIALIZE and AGING


TERUMO TE-331 TE332 Terufusion Syringe Pump. How to initialize E2PROM.
Medical (Electronic) Equipment.

1.
2.

3.
4.
5.
6.

7.
8.
9.
1.
2.

E2PROM INITIALIZING
RAM data is erased and E2PROM is refreshed by ROM data [default data at
factory]. This operation resets all the settings to the default. Readjust required.
Set the internal DIP switches 1 and 2 to ON and OFF respectively.
While holding down the [STOP] switch, press the On/Off switch. The
display first shows 888.8, then FACt, and a short beep follows. Release the
switches and check the display has changed to dIAg; which stands for
Diagnosis. Note: If you press the On/Off switch without pressing the Stop switch
Er91 appears on the 7 segment display. Turn off the pump, and repeat the
procedure above. If you do not release the switches at the beep, the display
changes to Er 8. Once Er 8 appears, you have to turn off the pump, and repeat
the procedure above.
When dIAg is displayed, press the Display Select switch 3 times. Check
the display shows CLEA which stands for clear.
Press Stop switch, and the display shows E or J. E is for Europe, and J is
for Japan.
Press and hold 'PURGE switch about 1.5 seconds to choose E for Europe.
Initialization of E2PROM starts. The display changes to Eur. As the
initialization proceeds, first 0
, then 00 ,and 000 , lastly 0000
appears on the display.the switch for Japan is the Start. If the Start switch is
pressed, JP appears on the display.
When successfully completed, can hear a short beep and see the display of
good.
In case of failure, bAd is displayed. Then the display returns to CLEA.
Press On/Off switch to Exit the initializing E2PROM mode.
Aging.
Set the DIP switches 1 and 2 to On and Off, respectively.
While holding the Stop switch, press On/Off switch. The display first
shows 888.8, then FACt and a short beep follows. Release the switches and
check the display has changed to dLAg which stands for Diagnosis. Note: If you
press the On/Off switch without pressing the Stop switch Er91 appears on the 7

segment display. Turn off the pump, and repeat the procedure above. If you do
not release the switches at the beep, the display changes to Er 8. Once Er 8
appears, you have to turn Off the pump and repeat the procedure above.
3.
When dIAG is displayed, press the Display Select switch 4 times. Check
the display shows Aging, which stands for Aging.
4.
Press Start switch, and Aging starts.
5.
When completed, press the On/Off switch to Exit the Aging Mode.
The following actions should be recognized while aging is performed.
* Motor rotates [slider moves approximately 9mm/h].
* All LED elements connected flash at 1 Hz / 50% duty.
* While the motor is rotating the minimum purge sound comes out.
Attention: You can stop aging at any time by pressing the Stop switch, and start
again by pressing the Start switch.
* The motor stops if occlusion is detected with the load on the slider at approx.
4606N=4.7kgf.