Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
QDR 4000
X-RAY BONE DENSITOMETER
TECHNICAL MANUAL
Hologic Inc.
35 Crosby Drive
Bedford, MA 01730
Document # QDR-4000-TM
Revision C
REV
ECO
DESCRIPTION
APPROVED BY
DATE CHECKED BY
DATE
4109
PRODUCTION RELEASE
G. CORRIDORI
4/23/98 F. ANDERSON
4/23/98
4134
REVISED
G. CORRIDORI
5/7/98 F. ANDERSON
5/8/98
4163
REVISED
SEE BELOW
SEE BELOW
NOTICE
The information contained in this manual is confidential and proprietary to Hologic, Inc. This
information is provided only to authorized representatives of Hologics customers solely for the
purpose of facilitating the use of Hologics products. No information contained herein may be
disclosed to any unauthorized person for any purpose whatsoever without the prior written
consent of Hologic, Inc.
Hologic QDR 4000 and the Hologic logo are registered trademarks of Hologic, Inc. All other
products and company names, used in this manual, are trademarks and registered trademarks of
other manufacturers.
The procedures described in this document are intended solely for use by direct employees of
Hologic, Inc. or authorized Hologic Equipment Resellers and their trained technicians. Any
unauthorized or untrained persons performing these procedures may affect the warranty of the
Hologic Model QDR 4000.
Exercise proper caution when servicing the system. There are dangerous and potentially lethal
voltages accessible within the QDR 4000 system. To avoid exposure to shock hazards, the
Main circuit breaker should be switched off, and the power cord removed, before working inside
any part of the system.
The QDR 4000 X-ray Bone Densitometer produces ionizing radiation in the form of X-rays. It
may be dangerous to the patient, operator or service technician unless safe exposure factors and
operating instructions are observed. To avoid unsafe exposure, do not attempt to service this
equipment unless you are a Hologic, Inc. certified service technician. Exercise proper caution
when servicing the system. A dosimeter (film badge) should always be worn while on site. Dose
and scatter radiation measurements must be taken after each service call to ensure that these
parameters are still within specifications.
Hologic, Inc. has made all reasonable effort to ensure that the information in this manual is
accurate and complete. Hologic, Inc. shall not, however, be liable for any technical or editorial
errors or omissions contained herein, or for incidental, special or consequential damages in
connection with the furnishing or use of this material. The information contained in this manual
is subject to change without notice.
DRAFTED BY
P. BENOIT
ii
DATE APPROVED BY
DATE CHECKED BY
12/22/98
DATE
Construction Instructions:
1. Print black from artwork supplied 8.5x11, double-sided, white, 20# bond, 3-hole
punched paper using Docutec printer or equivalent at 1200 d.p.i. or greater.
2. Insert all copies into white, 3-ring binder with clear sheaths for cover and spine
(WB Mason pn ESS-022112SPW, or equiv).
3. Create 5th cut tabs for sections 1-11 using white tab stock. Print tab text in
black, all caps, using Times New Roman or similar font (see table). Use clear
lamination over tabs.
4. Insert tabs before each section appropriately.
5. Reproduce cover and spine from artwork supplied in color onto 90# card stock.
Print at 600 dpi or greater. Cover is 8.5 x 11 and spine is 1.125x11.
6. Insert cover and spine into appropriate sheathes on the 3-ring binder.
TAB POSITION
1
2
3
4
5
6
7
8
9
10
11
TITLE
INTRODUCTION
FUNCTIONAL DESCRIPTION
INSTALLATION
ALIGNMENT AND CALIBRATION
SYSTEM COMPONENT LOCATION
REMOVE AND REPLACE PROCEDURES
FAULT ISOLATION
PREVENTIVE MAINTENANCE
COMPONENT SUMMARY INFORMATION
SOFTWARE
ERROR MESSAGES
iii
TABLE OF CONTENTS
1.
INTRODUCTION ..............................................................................................................1-1
1.1...SYSTEM OVERVIEW ...................................................................................................1-1
1.2...X-RAY SCANNING THEORY .......................................................................................1-2
1.3...X-RAY BEAM GEOMETRY...........................................................................................1-4
1.4...PRODUCT SPECIFICATIONS .....................................................................................1-5
1.4.1 GENERAL................................................................................................................1-5
1.4.2 PHYSICAL ...............................................................................................................1-5
1.4.3 TEMPERATURE AND HUMIDITY...........................................................................1-5
1.4.4 HEAT LOAD.............................................................................................................1-5
1.4.5 SUBJECT WEIGHT .................................................................................................1-5
1.4.6 POWER REQUIREMENTS .....................................................................................1-6
1.4.7 BMD SCAN TIME, PRECISION, DOSE ..................................................................1-6
1.4.8 RESOLUTION (APPROXIMATE) ............................................................................1-6
1.4.9 DUTY CYCLE ..........................................................................................................1-7
1.4.10 LEAKAGE CURRENT...........................................................................................1-7
1.5...LEAKAGE TECHNIQUE FACTORS .............................................................................1-7
1.6...ATTENUATION EQUIVALENT OF THE PATIENT SUPPORT ....................................1-7
1.7...MEASUREMENT CRITERIA FOR TECHNIQUE FACTORS .......................................1-8
2.
FUNCTIONAL DESCRIPTION.........................................................................................2-1
2.1...THE C-ARM COMPONENTS .......................................................................................2-1
2.1.1 THE OPERATOR PANEL........................................................................................2-1
2.1.2 THE LASER ASSEMBLY.........................................................................................2-1
2.1.3 THE DETECTOR PMT POWER SUPPLY ..............................................................2-1
2.1.4 THE MAIN DETECTOR ASSEMBLY.......................................................................2-2
2.1.5 THE A/D CONVERTER ...........................................................................................2-2
2.2...THE COMPUTER COMPONENTS...............................................................................2-2
2.2.1 TIMING AND CONTROL BOARD ...........................................................................2-2
2.2.2 THE STEPPER MOTOR CONTROLLER BOARD..................................................2-3
2.3...THE X-CARRIAGE COMPONENTS.............................................................................2-3
2.3.1 THE FILTER WHEEL ..............................................................................................2-3
2.3.2 THE X-RAY CONTROLLER ....................................................................................2-3
2.3.3 THE HIGH VOLTAGE POWER SUPPLY/SOURCE ASSEMBLY ...........................2-4
2.4...THE BASE ASSEMBLY COMPONENTS .....................................................................2-4
2.4.1 THE STEPPER TRANSLATOR...............................................................................2-4
2.4.2 SIGNAL DISTRIBUTION BOARD ...........................................................................2-4
2.4.3 THE OPTICAL ENCODER ......................................................................................2-4
3.
INSTALLATION ...............................................................................................................3-1
3.1...INTRODUCTION...........................................................................................................3-1
3.1.1 TOOLS AND SUPPLIES NEEDED..........................................................................3-1
3.1.2 ROOM SIZE.............................................................................................................3-1
3.2...INITIAL INSTALLATION ...............................................................................................3-2
3.2.1 INSPECTING THE SYSTEM ...................................................................................3-2
3.2.2 PREPARING THE SYSTEM FOR INSTALLATION ................................................3-2
3.2.3 UNPACKING THE SYSTEM....................................................................................3-3
3.2.4 MOVING SYSTEM TO ASSIGNED AREA ..............................................................3-4
3.2.5 REMOVING THE TABLE.........................................................................................3-5
3.2.6 REMOVING THE FRONT PANEL...........................................................................3-5
3.2.7 REMOVING THE TANK ASSEMBLY ......................................................................3-5
3.2.8 RAISING THE SCANNER TO THE VERTICAL POSITION ....................................3-5
iv
5.
vi
8.
9.
vii
viii
TABLE OF FIGURES
Figure 1-1. Q-Value Histogram......................................................................................................1-3
Figure 1-2. QDR 4000 Pencil Beam Geometry .............................................................................1-4
Figure 2-1. The Operator Control Panel ........................................................................................2-1
Figure 3-1. Minimum Room and Door Size ...................................................................................3-2
Figure 3-2. System Shipping Hardware.........................................................................................3-3
Figure 3-3. Minimum Clearance Needed to Unskid System .........................................................3-4
Figure 3-4. Scanner Clearances....................................................................................................3-4
Figure 3-5. Preparing the Scanner for Vertical Position ................................................................3-5
Figure 3-6. Positioning the Wheel Chocks (for lifting to vertical)...................................................3-6
Figure 3-7. Lifting the Scanner to the Vertical Position .................................................................3-6
Figure 3-8. The System Dimensions and Clearances ...................................................................3-8
Figure 3-9. Installing the C-Arm...................................................................................................3-10
Figure 3-10. Connecting the C-Arm Cables ................................................................................3-11
Figure 3-11. Connecting the Operator Console...........................................................................3-12
Figure 3-12. Rear View of Computer...........................................................................................3-13
Figure 3-13. Cabling the Scanner to the Console .......................................................................3-13
Figure 3-14. Checking the Source Voltage Jumpers...................................................................3-14
Figure 3-15. Installing the Lexan Panel .......................................................................................3-14
Figure 4-1. X-Ray Controller Test Points.......................................................................................4-2
Figure 4-2. Filter Wheel Pickoff Pulse Waveform .........................................................................4-3
Figure 4-3. X-Ray Tube Voltage ....................................................................................................4-5
Figure 4-4. X-Ray Tube Current ....................................................................................................4-7
Figure 4-5. X-Ray Beam Alignment Test Point..............................................................................4-8
Figure 4-6. Adjusting the Main Detector ........................................................................................4-8
Figure 4-7. Laser Alignment--Placing Tool ..................................................................................4-10
Figure 4-8. Scan of Pliers as Seen on Screen ............................................................................4-11
Figure 4-9. Locating the Laser Diode ..........................................................................................4-12
Figure 4-10. Laser DiodeSide View from Right........................................................................4-12
Figure 4-11. Checking for Tank Assembly Radiation Leakage ...................................................4-13
Figure 5-1. Base Assembly FRUs .................................................................................................5-5
Figure 5-2. X-Carriage FRUs.........................................................................................................5-6
Figure 5-3. X/Y-Drive FRUs...........................................................................................................5-6
Figure 5-4. Front and Left Side FRUs ...........................................................................................5-7
Figure 5-5. Operator Console FRUs..............................................................................................5-7
Figure 5-6. C-Arm FRUs................................................................................................................5-8
Figure 5-7. Tank Assembly FRUs .................................................................................................5-8
Figure 5-8. Computer Assembly and Components .......................................................................5-9
Figure 6-1. System Power Control.................................................................................................6-1
Figure 6-2. Removing the Table and Front Panel .........................................................................6-3
Figure 6-3. Removing the C-Arm (side view) ................................................................................6-5
Figure 6-4. Removing the C-Arm (top view) ..................................................................................6-6
Figure 6-5. Removing the Tank Assy.Step I ..............................................................................6-8
Figure 6-6. Removing the Tank Assy.Step II .............................................................................6-8
Figure 6-7. Removing the Bellows...............................................................................................6-10
ix
Section 1
1. INTRODUCTION
1.1 SYSTEM OVERVIEW
The Hologic QDR 4000 is a follow-on to the QDR-1000. The QDR 4000 incorporates many
features of the QDR-1000, such as the dual-energy pencil beam X-ray technology. The
QDR 4000 uses many of the same mechanical and electronic assemblies.
The QDR 4000 is smaller and lighter than the 1000. Many of the mechanical and electrical
assemblies have been redesigned or replaced. The computer has been upgraded and the system
now operates under Microsoft Windows95.
The QDR 4000 estimates the bone mineral content (BMC) and bone mineral density (BMD) of
selected areas of the body. It does this using X-rays of two different energy levels. This dualenergy scheme allows soft tissue within the selected area to be subtracted out, leaving only bone
to be scanned and measured.
The patient lays face up on the table. With the aid of a pointing laser, the operator positions the
scanning arm over the region of interest. After entering patient data and selecting the type and
size of scan desired, the operator initiates the scan with a single keystroke. Scanning, analysis,
and presentation of the scan and analysis results are automatic, as is the printing of a hard-copy
record. The operator needs only to indicate agreement or disagreement with the regions of
interest determined by the computer.
The operator does not have to select technique factors, as tube current and voltage are
preselected and fixed. Since scanning is performed by raster-scanning a small spot, rather than by
flooding the area as in conventional radiography, the exposure time is a function of the
dimensions of the area to be measured, the desired resolution and the desired precision.
BMC results are expressed in grams of calcium hydroxyapatite, and BMD is reported in
grams/cm2 of the same compound.
The maximum patient entrance dose is less than 180 Gy (18 mrad), approximately one third the
exposure of a standard chest X-ray. Leakage radiation at one meter is less than 10 Gy/hr
(1.0 mrad/hr). The patient, operator or room require no additional shielding (unless required by
local regulations. You can place the QDR 4000 in any convenient non-shielded examination
room. (State regulations may vary.)
The QDR 4000 employs a patented Automatic Internal Reference System, which continuously
calibrates the machine to eliminate the effects of variations in temperature, tube flux, etc. There
is no daily calibration required. Hologic does require the daily scanning of a quality control
phantom to provide assurance that the system is functioning correctly, and to aid in the detection
of any long-term drift.
The X-ray scans produced by the QDR 4000 and displayed on the monitor are intended only to
locate anatomical sites for measurement, and to assure the operator that the machine is operating
properly. They are not intended as a substitute for conventional film-based diagnostic images.
1-1
1-2
Section 1 - Introduction
Tissue
Bone
Number of
points with
a specific
value of Q
Q
Figure 1-1. Q-Value Histogram
6. Uses the "non-bone" points to calculate a baseline value for each transverse scan line.
Using these points, forms a new histogram and repeats steps 4 and 5 until the results
converge.
7. Smoothes the segment boundaries to eliminate isolated noise-generated "bone" points.
8. Displays the "bone" and "non-bone" points for operator approval.
9. Determines the constant of proportionality d0 that relates the Q values to actual BMC
(grams/cm2). That constant is determined by measuring how much Q shifts when boneequivalent material is interposed by the filter wheel.
10. Calculates the total bone mineral values by adding up the Q values for all "bone" points
in each region of interest (e.g., each vertebra), and multiplying by d0.
11. Determines the bone areas by counting the number of "bone" points in each region of
interest.
12. Calculates bone mineral density using the formula:
BMD = BMC / area
13. Displays the calculated results and prints the report.
1-3
E
Collimator
F
B
X-Ray
Generator
The
distance from the X-ray source
to the detector
distance from the source to the
collimator
beam width at the detector
detector size
collimator opening
X-ray focal spot size
Is
79.7 cm
31.375 in.
23.5 cm
9.250 in.
10.4 mm
22.5 mm
3.2 mm
0.3 x 1.0 mm
The detector used for the QDR 4000 is a single CdWO4 scintillator.
1-4
Section 1 - Introduction
GENERAL
SCANNING METHOD
X-RAY SYSTEM
SCANNING SITES
SCAN REGION
SCATTER DOSE
LEAKAGE RADIATION
EXTERNAL SHIELDING
REQUIREMENT
MEASUREMENT
ACCURACY
CALIBRATION
PATIENT POSITIONING
1.4.2
PHYSICAL
WEIGHT (approx.)
Shipping
Installed
FOOTPRINT
Scanner
TABLE HEIGHT
1.4.3
OPERATING
1.4.4
HEAT LOAD
1.4.5
Humidity
20 - 80% relative humidity,
noncondensing
SUBJECT WEIGHT
1-5
1.4.6
POWER REQUIREMENTS
FREQUENCY
(Hz)
CURRENT
(Amperes)
100
50/60
12.0
0.32
120
50/60
11.5
0.32
230
50/60
5.7
1.28
1.4.7
Exam
Default scan lengths
Scan Time
(sec)
in vivo
Precision
Gy
(mrad)
AP Spine
Fast
206
1.0%
19 44
1.9 - 4.4
Performance
394
1.0%
38 88
3.8 - 8.8
263
1.0%
20 49
2.0 - 4.9
Fast
355
1.0%
28 66
2.8 - 6.6
Performance
471
1.0%
38 88
3.8 - 8.8
587
2.5%
76 176
7.6 - 17.6
Reg Hi Resolution
523
1.4%
779
.66%
Forearm
Proximal Femur (Hip)
1.4.8
RESOLUTION (Approximate)
X
mm
in.
mm
in.
AP Spine
4.3
0.17
4.3
0.17
4.3
0.17
4.3
0.17
6.1
0.24
5.2
0.20
Forearm
4.3
0.17
4.3
0.17
Reg Hi Resolution
0.127
0.005
0.254
0.01
4.3
0.17
4.3
0.17
1-6
Section 1 - Introduction
1.4.9
DUTY CYCLE
8 hours/day
24 hours/day
AP Spine
100%
50%
100%
50%
100%
50%
Forearm
100%
50%
Reg Hi Resolution
100%
50%
100%
50%
<75"A
SINGLE FAULT
<400"A
Peak Current
3.0 mA (2.0 mA avg.)
Duty Cycle
66%
70 kVp
Lexan panel
4.7 mm Al equivalent
1.9 mm Al equivalent
Pad
0.3 mm Al equivalent
0.1 mm Al equivalent
1-7
1-8
Section 2
2. FUNCTIONAL DESCRIPTION
2.1 THE C-ARM COMPONENTS
2.1.1
The operator panel located at the top front of the C-arm allows the user to implement several
manual functions. It also indicates the activity of other scanner functions through LED
indicators.
The following illustrates the location of the various operator panel switches and indicators:
X-Ray On LED
Indicator
HOLOGIC
X-Ray On
Laser
Motion Enabled
Laser On
Motion Enabled/Laser On
LED Indicator
The pointing laser is a solid-state Class II unit with a power output of less than 1.0 mW. There is
a manual shutter below to inhibit the laser beam if necessary.
2.1.3
The Detector PMT (Photo-Multiplier Tube) Power Supply is located in the sloping section of the
C-arm (see Figure 6-9). The power supply is always on if both the main circuit breaker and the
instrument power are both on. This is done to keep the photomultiplier tube conditioned. The
detector assembly takes one hour to stabilize to proper operating conditions from room
temperature.
The PMT power supply receives 15 VDC from the low voltage power supply in the base
assembly. It outputs -700 VDC to drive the PMT.
2-1
2.1.4
The detector assembly is located inside the top portion of the C-arm (see Figure 5-6). It consists
of a scintillator crystal cemented to the face of a photomultiplier tube (PMT) and a circuit board
containing a bias network and a current-to-voltage converter/amplifier. The crystal/PMT and PC
board are enclosed in a light-tight container into which the crystal and PMT are fitted. X-rays
enter through a radiotransparent tube at one end and impinge on the crystal.
WARNING: Be careful when working around the detector. It operates with a supply voltage of
approximately -700 VDC which is present whenever the main circuit breaker and the
instrument power switch are on (see Figure 6-1).
Problems can be diagnosed using an oscilloscope and/or SURVEY (Section 10.3.4).
2.1.5
The Analog-to-Digital Converter board (A/D) (see Figure 5-6) converts the voltage output from
the detector assembly into a digital number representing the logarithm of that voltage.
The board handles inputs extending over a range of 100 dB from 100 V to 10 V. It integrates
the input signal during each X-ray pulse (every line half-cycle), executes a log conversion on the
integrated signal, then performs a 12-bit analog-to-digital conversion. It then transmits the data to
the computer through the T&C board.
Errors caused by PMT dark current and amplifier offset voltages are effectively eliminated by
circuitry and software which measures the total of these errors when X-rays are off, then injects a
compensating current into the log amp input when X-rays are on.
The A/D converter board can be tested using the SURVEY utility (Section 10.3.4).
The Timing & Control (T&C) board is located in the computer (see Figure 5-8) and is the
computers interface to the A/D converter, the X-ray controller, the filter wheel, laser, status
indicators, limit switches, and the manual controls. It also provides the computer with a line sync
signal.
The T&C board buffers data flow from and to the A/D converter and supplies it with convert and
integration commands.
Responding to commands from the computer, the T&C board generates line-synchronized beamon and beam-off signals for the X-ray controller board (it does this every half cycle), and
2-2
2.2.2
This board is located in the computer (see Figure 5-8). It is the interface between the computer
and the stepper translators which in turn drive the X- and Y-axis carriages.
Problems can be diagnosed using an oscilloscope and/or MOTORX and MOTORY
(Section 10.3.3).
The Filter Wheel assembly is a rotating wheel mounted on the top of the X-ray Tank directly
below the collimator opening. The rotation of this wheel is synchronized to the firing of the Xray source. The wheel is comprised of 6 equal segments of different reference materials:
Sequence (CW)
1
2
3
4
5
6
Material
Air
Aluminum, Bone Standard, Brass
Bone Standard
Aluminum, Tissue Standard, Brass
Tissue Standard
Aluminum, Brass
A timing sensor (filter wheel pickoff) senses once per revolution the rotational position of the
wheel relative to the tank opening and is used for synchronizing. The pickoff must also be
adjusted to provide proper phasing between the filter wheel and the X-ray pulses.
2.3.2
The X-Ray Controller controls the beam excitation voltage and the filament voltage applied to
the X-ray tube to maintain a regulated X-ray energy output.
The XRC is a multi-layer PCB separated into two separate regions, Control Logic and Power
Switching. The Control Logic communicates to the PC Control Logic by Optically Isolated
Digital I/O. The Power Switching section regulates the applied power to the Tank Module. Both
the Beam Excitation Voltage and the Filament Voltage are regulated by Closed Loop Controllers.
Independent Fault Detection and Safety Circuits monitor system operation. Safety fault
conditions inhibit operation of the X-Ray Tube.
The XRC performs the following functions:
2-3
2.3.3
This unit, also called the Tank (see Figure 5-7), is a sealed unit containing the X-ray tube, the
high voltage transformer, and other circuitry needed to generate 70 kVp and 140 kVp pulses. The
unit is filled with oil and weighs approximately 90 kg (200 lb). The tank produces potentially
lethal voltages and should NEVER be opened in the field. It is not field repairable. Externally
mounted assemblies such as the filter wheel, X-ray controller, etc. should be removed before
returning it to the factory.
The HVPS/Source can be tested as described in Section 4.3.
The stepper translator (see Figure 5-1) is a multiphase signal generator and special-purpose
power amplifier. Under control of low-power direction and pulse signals delivered by the Stepper
Motor Controller board to its optically-decoupled inputs, the translator generates the multiphase
signals and the power required by the stepper motor windings.
Note: Stepper translators are encapsulated units, and are therefore not repairable in either the
field or the factory.
2.4.2
The Signal Distribution board distributes the I/O signals from the Timing & Control and Stepper
Motor Controller boards located in the computer to the appropriate circuits in the scanner. It also
conditions the outputs from the Emergency Stop relay, Auxiliary X-Ray relay and E-Stop switch
to Timing & Control board.
The Signal Distribution board requires +5 VDC and 15 VDC.
2.4.3
This device (see Figure 5-3) turns when the corresponding X or Y-drive lead screws rotate. It
then generates a square wave and sends it back to the signal distribution board. On the signal
distribution board, the optical encoder square wave increments an eight-bit counter. There is one
counter for the X-drive and another counter for the Y-drive. If the X or Y position changes, the
counter sends the new count to the computer. The software can then determine if the motion
command actually did move the respective carriage assembly.
The optical encoder can be tested using the MOTORX or MOTORY utilities (Section 10.3.3).
2-4
Section 3
3. INSTALLATION
3.1 INTRODUCTION
The QDR 4000 is partially disassembled, packaged and shipped in one crate. The C-arm and the
computer peripheral cables are not installed. You must install the C-arm and the computer
peripheral cables after the system is placed at the installation site.
You must assemble the operator console and unpack the computer peripherals. After you
assemble the console, you must place the printer, keyboard, mouse and monitor in their places on
the console. You must then connect the cables and power cords.
The tank assembly is installed when the system is shipped. You may have to remove the tank
assembly if you need to stand the scanner vertically to move it through a doorway. Removing the
tank assembly makes the scanner considerably lighter.
WARNING: The QDR 4000 weighs 410 kg (900 lb) when shipped and 341 kg (750 lb) when
installed. With the C-arm and table removed, the weight is 295 kg (650 lb). It is
imperative that you do not attempt to lift this equipment alone.
3.1.1
A flat-head screwdriver
A Phillips-head screwdriver
An adjustable wrench*
Radiation survey meter (Victoreen 450P or equivalent)
Oscilloscope
Leakage measuring tool (099-0566)
Measuring tape or equivalent
Calibration Utilities disk (099-0560)
* A nut driver and/or socket wrench may be more useful in some cases but an adjustable wrench
is acceptable.
3.1.2
ROOM SIZE
The room must be at least 2.4 x 2.4 m (8 x 8 ft). The door must be at least 0.76 m (30 in.) wide.
If the scanner is moved to the room in the vertical position, you need 2.49 x 0.77 m
(98.00 x 30.25 in.) of space to lower the unit from vertical to horizontal . Make sure there is no
interference from built-in cabinets and/or other fixtures that can not be moved.
Note: If the customer site does not meet these minimum dimensions, you may have to
completely disassemble the unit down to the base assembly by removing:
1.
2.
3.
4.
the Table
the Tank Assembly
the Computer
the Frame
!
3-1
When the system arrives at the customer site, visually inspect the exterior of the crate for obvious
damage. Report any problems to the customers shipping department, if applicable, and Hologic
before proceeding.
3.2.2
The QDR 4000 system may be too large to move directly to the installation site from the loading
dock (or place it was delivered). If this is the case, you must:
1.
2.
3.
4.
5.
3-2
Section 3 - Installation
3.2.3
Sleeve
Carriage Brackets (2x)
(one on each side)
Anchor Brackets (2x)
Located under
machine:
Foam blocks
Wheel chocks
Ramps
Tank dolly
Unloading Ramp
(2x, mounted in
back)
3-3
To determine if the system should be moved vertically or horizontally, use Figure 3-4 and the
following table to check for minimum clearances:
D
E
A
B
C
3-4
Section 3 - Installation
Dimension
A
B
C
D
E
F
3.2.5
Inches
41
41
40
34
30
70
Centimeters
104
103
100
85
77
177
Comments
Table installed
Table removed and front panel installed
Table and front panel removed
Caster bars installed
Caster bars removed
Vertical clearance
If you must remove the table before preceding, refer to Section 6.3.1.
3.2.6
3.2.7
3.2.8
CAUTION: Do not tilt the scanner vertically onto a movers dolly. Doing this may scratch or
dent the side of the frame which cannot be repaired in the field. The caster bars
have integral padding and were designed for this purpose.
If the scanner unit needs to be set to a vertical position to move it to where it will be installed,
refer to Figure 3-5 and Figure 3-7 and.
1. Locate the two additional caster bars shipped with the system.
2. Mount the caster bars vertically into the holes provided on the lower edge of the right
side. Use the 1/2" bolts provided (located in the accessories box).
Vertical Caster
Rail (2x)
Mounting
Hardware (4x)
3-5
Wheel chocks in
place.
Position of vertical wheels after scanner is
lifted to vertical position.
Figure 3-6. Positioning the Wheel Chocks (for lifting to vertical)
3. While holding the vertical wheels upward, put the foam wheel lock in place.
WARNING: Step 3 is very important. If you fail to do this, the wheels do not drop into the
chocks properly and you may drop the scanner.
4. Position the horizontal caster wheels toward the center.
5. Place the two wheel chocks (shipped with system) in position (see Figure 3-6 and Figure
3-7).
6. Using help, tilt the scanner to the vertical position.
7. Remove foam blocks (or slide back).
You can now move the scanner to the room where it will be installed.
125 lb goes to
230 lb when
weight shifts to
other caster bar.
U Foam Caster
Wheel Lock (2x)
Section 3 - Installation
1. Position the lower horizontal caster wheels so they are facing up and secure them using
the wheel locks provided which are taped to the caster bar (Figure 3-6).
WARNING: Step 1 is very important. If you fail to do this, the wheels do not drop into the
chocks properly and you may drop the scanner.
2. Position the wheel chocks as detailed in Figure 3-6 and Figure 3-7.
WARNING: You need two people to perform step 3.
3. Carefully lower the scanner.
Figure 3-8 and the tables following show the QDR 4000 dimensions and clearances. Use this
information to plan the layout of the room and the positioning of the scanner and the operator
console.
Some things to remember while planning the layout are:
The power cord and the computer cables are approximately 15 ft (4.8 m) long
respectively and exit the scanner from the left side only.
You must set the front of the scanner table at least 54 in. (137 cm) from the wall.
Make sure there is clearance for the C-arm to move in the Y direction (left to right).
There may be pipes or wall cabinets in the way at some installations.
3-7
D
F
H
J
K
I
3.3.2
Feet
15.0
Inches
28.3
25.3
54.0
30.0
16.0
50.3
72.3
29.0
64.0
Meters
4.6
Centimeters
71.9
64.3
137.2
76.2
40.6
127.8
183.6
73.7
162.6
Note: Refer to Figure 3-1 to determine clearances needed to lower the unit.
1. Remove the two vertical caster bars on the side (if installed).
3-8
Section 3 - Installation
2. Place the unit where it will be used but leave room behind it to install the C-arm.
3.3.3
Remove the 3 bolts (one in bottom/rear) and washers holding the left caster bar.
Lift the left side of the scanner.
Remove the caster bar.
Set the left side of the scanner down.
CAUTION: Do not drop the scanner. You may cause damage to the computer components.
5.
6.
7.
8.
9.
10.
11.
Note: The four (4) caster bars must be returned to Hologic in the box provided (include brackets
and chocks).
3.3.4
3.3.5
Refer to Figure 3-2 and remove the (2) Carriage Shipping Brackets. There is one on the near right
and another on the far left.
Note: Return these brackets to Hologic.
3.3.6
Note: The mounting hardware for the C-Arm may already be installed when the system is
shipped. Refer to Figure 3-9 and remove this hardware before proceeding.
Refer to Figure 3-9 and:
1. Remove the back cover (4 screws).
2. Place the C-arm into the X-carriage assembly so that the mounting studs pass through
the mounting holes.
CAUTION: Be sure to clear the cables as they pass through the cutout.
Note: Do NOT attach washers and nuts at this time.
3. Attach the mounting screws and washers. Do not tighten.
4. Fasten the brace to the X-Carriage Assembly. Do not tighten any of the six
mounting screws.
!
3-9
3-10
Section 3 - Installation
Red Stripe
Operator Panel
Cable
Mate Pin 1
with Red Stripe
Cable Clamp
Ground Strap
Power Cable
The Stand
The Keyboard
The Monitor
The Mouse
The Printer
3-11
Operator Console
Power Source
3-12
Section 3 - Installation
Pentium 200
Pentium 333
Power Receptacle
Voltage Select Switch
Serial Port #1 (not used)
Mouse Port
Keyboard Port
Printer Port
Serial Port #2 (not used)
Video Card
SCSI Card (JAZ Drive)
Open Slot
Motor Stepper Card
Timing and Control Card
Network Card
Modem Card (optional)
Wall Jack
Phone Jack
Printer
Mouse
Computer
Monitor
Main Power
Panel
Keyboard
Power Strip
The system should be properly set to the voltage available at the customer site before it is
shipped. If set improperly, it could damage the system. Refer to Figure 3-14 to check for proper
voltage jumpers on the isolation transformer. (The location of the isolation transformer is shown
in Figure 5-1.)
3-13
BLK (1)
PRI
SEC
120
120
PRI
SEC
120
120
110
110
110
110
100
100
100
100
100
0
120
120
0
120
120
120
120
110
110
110
110
110
110
100
100
100
100
100
100
BLK (1)
BLK (2)
BLK (2)
100VAC
BLK (1)
PRI
SEC
120
120
110
110
100
BLK (2)
120VAC
230VAC
3.3.9
Cutout in table
Section 3 - Installation
4. Measure the distance from the top of the Lexan panel to the bottom of the C-arm.
5. Maneuver the C-arm vertically to achieve a distance of 19.5 in. (49.5 cm) without
twisting the arm.
6. Tighten all the mounting screws.
7. Recheck the C-arm distance. If out of specification, repeat this procedure.
SOFTWARE
Hologic installs the Operating System and all the optional software the customer ordered on the
computer before the system is shipped from the factory. If, by chance, you do need to install
additional software, use the installation notes that come with it.
3.4.2
When you boot the system for the first time, it asks you for the license agreement number. The
licensing information is shipped with the system. You must enter this information or you can not
proceed further. Follow the instructions given after you boot the system.
3.4.3
THE COMPUTER
When Windows 95 (or later) starts after you power on the computer, it checks the operation of
memory, the disks and other internal options.
If you suspect any problems, you should check the following computer options and peripherals
for proper operation:
Device
3.5 Floppy Drive
Hard Disk (C Drive)
JAZ Drive (if installed)
M/O Drive (if installed)
Printer
3.4.4
Go to Section 7.3
THE SCANNER
Device
X-Rays On/Off
Laser Beam On/Off
Table Motion X/Y Directions
Emergency Stop Switch
Operation
Go to Section 10.3.4
Go to Section 10.3.4
Go to Section 10.3.3
Go to Section 2.2.1
3-15
3.4.5
Now that you have made a preinstallation cursory check of the computer, software, and the basic
functions on the scanner, you must verify the scanner operation in detail and make all necessary
adjustments.
Use the following table as a guide to which checks and adjustments to do and the order in which
to do them.
Step
1
2
3
4
5
6
7
8
9
10
11
3.4.6
Check/Adjustment
Go to
Section 4.1
Section 4.2
Section 4.3
Section 4.5
Section 4.6
Section 4.7
Section 4.8
Section 4.9
Section 4.9
Section 4.10
Section 4.11
Radiation Measurement
Report Entry?
(Section 3.4.8)
Yes
No
Yes
No
No
No
Yes
Yes
Yes
No
Yes
SYSTEM VERIFICATION
There are a few more things you must do BEFORE signing the system over to the customer:
Step
1
2
3
4
5
3.4.7
Procedure
Insert the Spine Phantom number into the patient data base.
Run AutoScan and do 20 scans.
Analyze 20 scans.
Add the 20 scans to the QC database.
Print QC plots and other important information.
Go to
4000 Users Guide
Section 10.3
4000 Users Guide
4000 Users Guide
4000 Users Guide
REPORTS TO BE COMPLETED
Complete the following reports and return to them to Hologic when you finish the installation.
Form
Part No.
FDA
FDA 2579
CSD-0024-01
FS-SERVRPT
CSD-0026-02
3.4.8
3-16
Section 3 - Installation
QDR 1000/1500/2000/4000
Radiation Measurement Report
Customer Name
Customer Address
Customer Contact
Hologic Model Number
Serial Number
Date Measurement Taken
(100V)
(#90V, "110V)
(70 kVp)
(#63kVp , "77kVp)
(120V)
(#108V, "132V)
(140 kVp)
(#126kVp, "154kVp)
(230V)
(#207V, "253V)
Reproducibility Test:
SD Hi Air
("14)
SD LoAir
("14)
Tube Current:
(#2.5mA, "3.5mA )
(3mA)
Radiation Tests
Leakage Test using Victoreen Model #450P (or equivalent)
Dosage Test
Scatter Test
S/N
Pencil Beam
mR/h
Pencil Beam
mR
Pencil Beam
mR/h
F/E Signature________________________
Date______________
CSD-0026-02 Rev A
3-17
Section 4
4. ALIGNMENT AND CALIBRATION
This section shows the procedures for the following:
Checking reproducibility
Normal
2 ms
Both and Chop
Internal, Channel 2, Positive Slope,
2 Volts/Div
50 Volts/Div
4-1
L1
J2
F1
TP2
F2
J1
L2
C12
TP12
TP1
J8
C18
TP3
J4
TP11
TP4
PTC1
J6
R38
D20
D17
J5
R39
TP5
TP6
C60
S1
D24
U14
D29
D30
J7
TP7
TP8
D23
U15
PTC2
Q11
TP9
J3
TP10
D33
Chassis
Ground
4-2
Chan 1,
Filter Wheel
Pickoff Pulse
Chan 2,
120 VAC
2 mS/div
4-3
4.3.1
4-4
2 V/div
14v
7v
2 mS/div
Figure 4-3. X-Ray Tube Voltage
Note: Record the results in the Radiation Measurement Report, Section 3.4.7.
4.3.2
The following procedure details the steps to check and adjust the X-ray pulse energies. These
energies must be 702 kVp and 1402 kVp. To check and adjust them:
1. Set up scope and probes as described in the previous section.
4-5
4.3.3
WARNING: Keep hands and other body parts out of the X-ray beam path (from the top front of
the tank to the front of the C-arm. Do not place any objects in the beam path.
7. Adjust the scope to display a waveform as pictured in Figure 4-4.
Note: If the scope does not display the proper current wave forms, the XRC or the tank
assembly may have to be replaced. Before doing this, it is advised that you contact
Hologic field support personnel to discuss the situation. Double-check the readings and
make sure the scope is calibrated properly. Replacing the tank assembly is a costly and
time-consuming task.
8. Divide the peak signal on the scope by 3.33 to get the peak current in mA.
9. Press <F2> to turn off the X-rays.
Note: Record the results in the Radiation Measurement Report, Section 3.4.7.
4-6
2 V/div
8.5 - 11.5 V
2 mS/div
6.0 - 8.0 ms
Auto, Negative
2 ms
Channel 1
Line or Channel 1
1V
4. Remove the C-arm top cover and connect the channel 1 probe to the Test Point Lead
(U3-1, see Figure 4-5) on the ADC board.
4-7
FRONT
J3
J5
U3
Top View
Test Point
Lead
Clamp Screws
(top and bottom)
Slide tool back to
this point.
C-Arm
(top view)
Main Detector
Assembly
4-8
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
regardless of the direction in which the arm is flexed. If this is the case, skip to step 19.
If not, proceed to step 12.
Turn the X-ray beam off by pressing <F2>.
Gain access to the detector assembly (see Figure 4-6).
Loosen the screws on the top and bottom of the holding bracket. (The top screw allows
the detector to move forward while the bottom screw allows the detector to move from
side to side.)
Turn the X-ray beam on by pressing <F1>.
Move the detector left-right and forward-back to determine the location at which the
pulse amplitudes are maximized.
When the maximum-signal position has been located, turn the beam off and tighten the
detector mounting screws.
Turn the beam back on and verify that the maximum signal amplitude is still present. If
not, repeat steps 13-16.
Turn the X-rays off by pressing <F2>.
Remove the tool if installed.
Reinstall the C-arm top cover.
4-9
The laser beam alignment check and adjustment are done using a pointed metal tool (in this case,
we use a pair of needle-nose pliers).
WARNING: Do not use an object with a reflective surface.
1. Press the Laser button on the Operator Control Panel.
2. Place the needle-nose pliers on the table as pictured in Figure 4-7 so that the laser beam
just touches the pointed tip of the pliers.
FRONT
C-Arm
Left Side
4-10
Scan at least
this amount.
4.7.2
To adjust the laser beam, you must gain access to the laser diode assembly.
Do the following:
1. Remove the top cover of the C-arm (if not already removed).
2. Referring to Figure 4-9, locate the laser diode assembly.
4-11
4-12
Time
3.3 seconds
4.5 seconds
This procedure explains how to check for leakage radiation around the Tank Assembly:
X-Ray Beam
Belt Cutout
Tank
Aperture
Leakage Test
Tool installed
over aperture.
1.
2.
3.
4.
5.
6.
7.
4-13
4.8.2
1. Position the survey meter one meter (approx. 40) from the tank and move it all around
the outside.
The reading must be 1.0 mR/h or less.
2. Record the highest reading in the service report.
3. Turn off the X-rays (press <F2>).
4. Replace the front panel.
Note: Record the results in the Radiation Measurement Report, Section 3.4.7.
Make sure all the panels are on the scanner and the pad is on the table.
You must have an ion chamber radiation meter to measure patient dose and scatter.
WARNING: Do not place your hands or other body parts in the path of the X-ray beam.
4.9.1
1.
2.
3.
4.
5.
6.
PATIENT DOSE
Place an ion chamber radiation meter in the center of the scan area, positioned so that
the entire meter will be scanned.
Turn the meter on, set it to INTEGRATE, and reset it to zero.
Position the C-arm so that the laser is just to the left of the meter without touching it.
Scan the meter.
To do this, use AP Spine, Performance mode and reset the scan area to length = 9" and
width = 5".
Verify that the entire meter has been scanned. That is, the entire meter can be seen on
the monitor.
Record the meter reading.
4.9.2
SCATTER
4-14
4-15
4-16
Then
the coefficient of variation for absolute radiation
is less than 4%.
4-17
Section 5
5. SYSTEM COMPONENTS & LOCATION
5.1 FIELD REPLACEABLE UNITS (FRU)
The following table lists the recommended spare parts and Field Replaceable Units (FRUs) for
the QDR 4000.
5.1.1
SCANNER PARTS
PART NUMBER
010-0127
010-0158
010-0954
010-0955
010-0956
010-0957
010-0962
010-0963
010-0964
010-0972
010-0978
010-0983
010-0985
010-0986
010-0992
020-0115
030-0007-01
030-0109
030-1796
030-1884
030-2358
110-0002
140-0001
140-0092
140-0093
160-0001
180-0391
200-0281
235-0031
DESCRIPTION
LOCATION
R&R1
Figure 5-6
Figure 5-1
Figure 5-1
Figure 5-1
Figure 5-4
Figure 5-4
Figure 5-3
Figure 5-3
6.11
Figure 5-6
Figure 5-7
Figure 5-7
Figure 5-7
Figure 5-6
Figure 5-7
Figure 5-1
Figure 5-3
Figure 3-15
Figure 5-3
Not Shown
Not Shown
Figure 5-6
Figure 5-6
Figure 5-7
Figure 5-1
Figure 5-1
010-0983
Figure 5-3
Figure 5-3
6.18
6.17
6.8
6.10
6.7
6.6
6.12
3.3.10
6.14
6.13
6.6
6.16
5-1
PART NUMBER
DESCRIPTION
LOCATION
Figure 5-3
Figure 5-3
Figure 5-1
Figure 5-2
Figure 5-1
Figure 5-2
Figure 5-7
Figure 5-3
Figure 5-3
Figure 5-3
Figure 5-3
Figure 5-3
010-0957
320-0003
320-0004
430-0016
465-0001
465-0005
465-0038
485-0043
250-0001
250-0002
253-0035
253-0038
255-0002
265-0002
265-0007
285-0001
295-0020
295-0041
310-0001
310-0006
5.1.2
Figure 5-7
Figure 5-3
Figure 5-1
010-0956
Figure 5-1
Figure 5-2
010-0972
010-0956
CABLES ASSEMBLIES
PART
NUMBER
180-0006
180-0346
180-0347
180-0349
180-0350
180-0351
180-0352
180-0353
180-0354
180-0355
180-0356
180-0360
5-2
R&R1
LOCATION
DESCRIPTION
From:
To:
PMT A/D
Translator Stepper Drive
28 V Supply AC In
DC Supply AC In
LVP Supply DC Out
Power Switch
Secondary Side Xfmr
Computer/AUX X-Ray
Out
Line Filter
P Side Isolation Xfmr
Emg&Aux Relay Power
T&C Limit Sw
DIN Rail
Main Circuit Breaker
Distribution Bd P17
Distribution Board, P11
PART
NUMBER
180-0361
180-0362
180-0363
180-0364
180-0365
180-0366
180-0367
180-0368
180-0370
180-0371
180-0372
180-0373
180-0374
180-0375
180-0376
180-0377
180-0378
180-0379
180-0380
180-0383
180-0386
180-0391
180-0394
180-0404
180-0406
LOCATION
DESCRIPTION
From:
To:
Y Motion Direction
X Motion Direction
Motor Drive Speed
T&C Power
T&C Control
Motor Drive Direction
T&C Data
Y Stepper Drive
L/R Switches
A/D Data
X Stepper Drive
Fwd/Back Sw. Mounts
Fwd/Back Switches
X-Ray Controller Signal
X-Ray Cntrl. Pwr (Base)
X-Ray Controller Power
Control Panel Pwr
Emg Stop/Laser Pwr
A/D Power
Int. Computr Ac Power
X Stepper Signals
Filter Wheel P/O
Tank Power And Signal
Filter Wheel Ext.
Encoder Interface
5-3
5.1.3
COMPUTER PARTS
PART NUMBER
DESCRIPTION
LOCATION
010-0990
Computer Assembly
Figure 5-4
Figure 5-8
Figure 5-8
Figure 5-8
Figure 5-5
Figure 5-8
010-0990
Figure 5-8
010-0990
Figure 5-8
Figure 5-5
010-0990
010-0990
010-0990
010-0990
Figure 5-5
Figure 5-5
Figure 5-5
120-0049
130-0001
140-0025
300-0351
C-PRINTDLX-400
LASERJET-400
5-4
R&R
6.15
Y-Drive Assembly
(see Figure 3)
X-Carriage FRUs
(see Figure 2)
Signal Distribution Board
(140-0093)
FRONT
5-5
X-Drive Assy.
(see Figure 3)
FRONT
See Note 2.
Motor Coupling
(265-0007)
See Note 2.
Lead Screw
Assy.
See Note 1.
Optical Encoder
(285-0001)
Note 1.
X-Drive - 010-0962
Y-Drive - 010-0963
Note 2.
Replace original set
screws with 200-0281
5-6
Shaft Spacer
(235-0031)
Blue Grommet,
1/16 (2x)
(295-0041)
Computer Assembly
(010-0990)
Monitor
Keyboard
Power Outlet
(mounted on back)
Mouse
5-7
Main Detector
(010-0127)
Laser Diode
(010-0986
Top View
A/D
Converter
(140-0001)
Control Panel
(010-0972)
PMT Power
Supply Assy.
(110-0002)
FRONT
Filter Wheel
(010-0983)
Fan Assy.
(010-0992)
Filter Wheel
Motor
(320-0003)
Universal Tank Assy.
(010-0978)
5-8
JAZ Drive
2
3
3
4
5
6
7
LEGEND
Video Card
JAZ SCSI
Interface Board
Open
Stepper Motor
Controller Board
Timing and
Control Board
1
2
6
5
4
7
5-9
Section 6
6. REMOVE AND REPLACE PROCEDURES
6.1 POWERING DOWN THE SYSTEM
WARNING: In most cases it is imperative that you power down the system before gaining
access to the inside of the scanner. The scanner (instrument) contains dangerous
voltages, mechanics and heat that may cause serious bodily harm.
Refer to Figure 6-1 and follow these procedures:
1. Make sure that the system is not in use.
Note: It is good practice to ask the operator for permission to shut the system down.
2. Make sure the computer is in DOS at the C:\MENU> prompt by pressing and holding
down <Alt><F1> at the selections menu.
3. Disable the X-rays by turning the X-Ray Enable Key counter-clockwise from 3 oclock
to 12 oclock. The green LED should go off.
4. Remove scanner power by switching the Instrument Power switch to the 0 position.
5. Remove operator console power by switching the Computer Power switch to the 0
position.
6. Remove power to the entire system by setting the Main Circuit Breaker to the 0
position.
7. Remove power from the system by removing the power cord from the facility power
outlet.
Power Switch Panel
X-Ray Enable Key
Switch
Computer Power
Instrument Power
I
t
w
i n
1P
w2
r0
m
- R
C e
P
no
t m
l e
Main Circuit
Breaker (CB1)
6-1
REMOVAL
Eight sets of screws and washers hold the table to the frame. To remove the table, refer to Figure
6-2 and:
1. Remove the table pad (if present).
2. Remove the eight screws and their washers.
3. Remove the table.
6.3.2
REPLACEMENT
REMOVAL
To remove the front panel, face the scanner from the front and:
1. Disengage the front panel screw by turning it turn (Figure 6-2).
6-2
Table Securing
Hardware (8x)
Table (removed)
Upper Holding
Brackets (2x)
Front Panel
Computer
Lower Holding
Brackets (2x)
6.4.2
REPLACEMENT
To install the front panel onto the scanner unit, refer to Figure 6-2 and:
1. Place the front panel onto the frame via the lower holding clips.
You must tilt the top of the panel forward to align the brackets properly.
2. Reconnect the ground strap.
3. Tilt the top of the front panel so it touches the top of the scanner frame and slide to the
right.
4. Engage the front panel screw turn.
6-3
6.5.1
REMOVAL
WARNING: Power down the system before performing this procedure (Section 6.1).
For this procedure, refer to
To remove the C-arm, refer to Figure 6-3 and
Figure 6-4 and:
1.
2.
3.
4.
6-4
Top Cover
Cables
Lower Carriage
Upper Carriage
6-5
Red Stripe
Operator Panel
Cable
Mate Pin 1
with Red Stripe
Cable Clamp
Ground Strap
Power Cable
6.5.2
REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
Refer to Figure 6-3 and:
1. Remove the back cover (4 screws).
2. Place the C-arm into the X-carriage assembly so that the mounting studs pass through
the mounting holes. Be sure to clear the cables as they pass through the cutout.
Note: Do NOT attach washers and nuts at this time.
3.
4.
5.
6.
7.
8.
6-6
6.6.1
REMOVAL
CAUTION: The X-Ray Controller Assembly (XRC, Figure 6-5), is static sensitive. All antistatic precautions must be taken before handling the XRC, such as grounded wrist straps and
static mats.
You must gain access to the inside of the scanner (Section 6.2). Refer to Figure 6-5 and:
1. Remove all the cables from the connectors and the ground strap from the XRC.
2. Remove the four XRC fastening screws and washers.
3. Remove the XRC and set aside in a static-free area.
The XRC Board can now be removed from the assembly.
6.6.2
REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
Refer to Figure 6-5 and:
1. Place the XRC assembly onto the tank assembly as pictured in Figure 6-5.
2. Install all securing nuts and screws.
3. Connect all cables and the ground strap. (Figure 9-10).
REMOVAL
WARNING: Power down the system before performing this procedure (Section 6.1).
1. Remove the front panel (Section 6.4.1).
2. Remove the XRC assembly (Section 6.6.1).
3. Remove the front tank cover (Figure 6-5).
WARNING: The tank assembly weighs approximately 200 lb (90 kg) and you must not
attempt to remove it alone.
6-7
Front Cable
Jacks (5x)
Rear Cable
Jacks (3x)
X-Drive
Step Motor
Tank Anchor
Bolts (4x)
FRONT
MOTION
X-Drive Motor
Clearance
Tank Dolly
6-8
6.7.2
REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
WARNING: The tank assembly weighs approximately 200 lb (90 kg) and you must not
attempt to replace it alone.
CAUTION: When replacing the tank assembly, take care not to damage the X-drive stepper
motor.
1. Position the lower carriage at the center of the Y-drive (left to right).
2. Position the upper carriage so that it is in the center of the scanner.
3. Referring to Figure 6-6, slide the tank assembly through the front chassis taking care to
clear the X-drive stepper motor.
4. Set the tank assembly onto the upper-carriage assembly and position it over the four
mounting holes.
5. Fasten the tank assembly using the four anchor bolts and washers.
6. Place the XRC in position and fasten with the four screws and washers (Figure 6-5).
7. Plug all the cables into the jacks on the XRC in the front and in the back.
8. Replace the front tank cover.
6-9
Top Rail
Support
Bracket
Bellows
Flathead
Screw (2x)
Washer Plate
Bellows Screw &
Washers (4x)
REMOVAL
The bellows assembly is located on the back of the scanner and is used to enclose the internal
components.
The bellows assembly consists of:
Support brackets, one on each side of the tank assembly between the tank and the
bellows.
Securing hardware
6.8.2
REPLACEMENT
To install the bellows assembly to the back of the scanner, refer to Figure 6-7 and:
6-10
REMOVAL
WARNING: Power down the system before performing this procedure (Section 6.1).
To remove the frame, you must first remove:
1.
2.
3.
4.
5.
6.
6-11
6.9.2
REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
It is unlikely you will have to replace the frame but you may have to reinstall it after removing it
while installing the scanner.
To reinstall the frame, refer to Figure 6-8 and:
1. Place the frame over the four inserts in the base assembly.
2. Install the mounting hardware.
3. Replace all the hardware and cables listed in the previous section.
6.10.1 REMOVAL
The Control Panel is located at the front-most end of the C-arm (Figure 6-9). To remove it:
1.
2.
3.
4.
6-12
Control Panel
Fastening
Screws (4x)
Main Detector
Mounting Block
B
C
Main Detector
Assy.
A/D Converter
F
PMT Power
Supply Assy.
6-13
6.11.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
Make sure that the main detector, the mounting bracket and mounting screw are assembled.
Locate the position of the main detector in the C-arm (refer to Figure 6-9) and:
1. Set the main detector in place so that the mounting screw passes through the mounting
hole in the C-arm.
2. Secure using the nut and washer fasteners.
3. Make connections G, and D. (There may be one more connector not shown in Figure
6-9.)
Note: After replacing the Main Detector, you must do the Main Detector Alignment
(Section 4.5).
6-14
12 11
Emergency
Stop Switch
Connection C.
Clamp Nut
& Washer
Place the Laser Diode Assembly into its proper position in the C-arm (Figure 6-9).
Secure the Laser Diode Assembly using the Clamp Nut and Washer.
Make Connection C.
Make the connections to pins 11 and 12 on the Emergency Stop Switch.
Note: After you replace the Laser Diode Assembly, you must perform the Laser Diode
Alignment (Section 4.7).
6.13.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
Locate the A/D Converter Assembly in Figure 6-9 and:
1. Place the A/D Converter Assembly in its proper place in the top on the C-arm.
6-15
6.14.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
Locate the PMT Power Supply Assembly in Figure 6-9 and:
1. Set the PMT Power Supply into its proper position and secure with three mounting
screws and washers.
2. Make connection J.
Note: There may be one more connection that is not shown.
6-16
6.15.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
Before installing the computer, make sure that it is configured properly, that is, control boards
properly installed and addressed; internal peripherals installed; etc. If any internal changes need
to be done to the computer components, you must remove the computer first. (See Section 9 for
board configuration data.)
Refer to Figure 6-2 to find the computer location. You must gain access to inside of the scanner.
Do this by:
a)
b)
c)
d)
To clear the computer area for better access after getting inside, you can slide the X-carriage
assembly (Figure 5-1) to the right by pushing it by hand.
CAUTION: Do not move the carriage assemblies faster than one inch per second in order to
protect the stepper motor translators.
Then:
1. Slide the computer into its place under the hold-down bracket but leave room in the
back to install the cables.
2. Install the cables coming from the Timing & Control Board and the Stepper Motor
Board into the Signal Distribution Board (Figure 5-1 and Figure 9-4) as follows:
J1, J9, J10, J11, J13, J14, J15, J16
3. Install all the cables into the back of the computer that go through the access hole on the
left side of the scanner to the operator console (the mouse, keyboard, printer, etc.).
4. Slide the computer back and to the left and secure with the hold-down bracket.
Note: The distance from the front of the computer to the front of the frame should be 5/8.
5. Reassemble the scanner.
6-17
6.16.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
To replace the AC line filter:
1.
2.
3.
4.
Place the AC line filter into its place in the base of the scanner .
Attach the proper wires.
Install the computer pedestal.
Install the computer (Section 6.15.2).
6.17.1 REMOVAL
WARNING: Power down the system before performing this procedure (Section 6.1).
To remove the Main Power Panel:
1.
2.
3.
4.
Remove the external power chord connected to the Main Power Panel.
Remove the four (4) lock nuts holding the Main Power Panel to the frame.
Remove the Main Power Panel.
Remove all the wire connections.
6.17.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
To replace the Main Power Panel:
1.
2.
3.
4.
5.
6.
6-18
6.18.1 REMOVAL
WARNING: Power down the system before performing this procedure (Section 6.1).
To remove the Power Switch Panel:
1. Remove the front panel (Section 6.4.1).
2. Remove the four (4) lock nuts holding the Power Switch Panel to the frame.
3. Slide the Power Switch Panel back along the threaded studs toward the center of the
scanner until it becomes free.
4. Remove all the wire connections.
6.18.2 REPLACEMENT
WARNING: Power down the system before performing this procedure (Section 6.1).
To replace the Power Switch Panel:
1.
2.
3.
4.
5.
6-19
Section 7
7. FAULT ISOLATION
If an error message appears, refer to Section 11 for an explanation of the message and its possible
cause.
Tabulated below are the symptoms, possible causes and recommended solutions of the problems
you may have with the QDR 4000. The problems are divided into six categories (Scanning/XRay Generation, Precision, Motion, Noise, Data Storage & Retrieval and Miscellaneous. Possible
causes are listed in order of decreasing probability. For any given symptom, investigate potential
causes in the order listed.
Symptoms
Possible Causes
Recommended Action
Go To
realign source/detector
replace the tank
repair/replace cables
repair/replace cables
3.3.11, 4.4
6.7
5.1.2
5.1.2
misaligned source/detector
oil in precollimator cup
bad A/D-to-T&C cables
bad T&C-to-X-Ray Controller
cables
bad Distribution Board
low PMT supply voltage
bad A/D board
bad X-ray controller board
filter wheel phase error
loose filter wheel segment
bad detector
bad HVPS/Source
replace board
adjust or replace supply
replace board
replace board
adjust filter wheel phasing
replace filter wheel assy.
replace detector
replace tank assembly
Figure 5-1
4.6
6.13
6.6
4.2
Scans have
wavy edges
("French fries")
or lines
Scans have
spots or streaks
4.2
Figure 5-7,
4.2
4.2
Figure 5-8
replace board
Figure 5-8
repair/replace cables
replace board
repair/replace cables
replace board
5.1.2
Figure 5-1
5.1.2
6.13
replace board
6.6
6.11
6.7
Page 10-7
7-1
repair/replace cables
5.1.2
replace fuse
Figure 4-1
repair/replace cables
5.1.2
6.7
no display of any
kind on monitor
tighten/repair/replace cable
replace monitor
replace board
5.1.3
5.1.3
5.1.3
display jumps on
monitor
defective monitor
defective display board
replace monitor
replace board
5.1.3
5.1.3
monitor display
is dim
increase brightness or
replace monitor
5.1.3
desired area is
not fully scanned
operator error
laser misaligned
retry
adjust laser
7.1.2
6.12
PRECISION PROBLEMS
Symptoms
Possible Causes
Recommended
Action
Go To
display has
spots
repair/replace cables
adjust or replace supply
replace board
replace detector
replace board
replace board
replace tank assembly
5.1.2
4.6
Figure 5-1
6.11
6.13
6.6
6.7
phantom BMD
drift or shift
misaligned source/detector
oil in precollimator cup
bad detector
align source/detector
replace tank assembly
replace detector
4.5
6.7
6.11
7.1.3
MOTION PROBLEMS
Note: First differentiate between mechanical and electrical carriage motion/leadscrew problems
by turning the leadscrew by hand (during Main Menu display, not while in Utility Local).
If it turns freely, the problem may be electrical. If it turns very stiffly or not at all, the
problem is probably mechanical.
At the same time, check for axial play in leadscrew--there should be none.
7-2
Possible Causes
Recommended Action
Go To
Figure 2-1
turn key to ON
Figure 6-1
repair/replace cable
5.1.2
replace board
Figure 5-8
replace translator
Figure 5-1
replace motor
Figure 5-1
Figure 5-2
repair/replace cable
5.1.2
replace board
Figure 5-8
reconnect bracket
Figure 5-3
replace bearings
Figure 5-3
tighten setscrews
Figure 5-3
Electrical
no axis motion
occurs when
expected, or the
machine stalls
Mechanical
Note: Insert the Calibration Utilities disk (099-0560) into the floppy drive.
Note: Differentiate between mechanical and electrical causes of the "No Motion Detected While
X-rays ON" message by going to DOS, typing A:MOTION OFF<Enter>, rebooting and
running a scan. If the scan runs successfully, the error message source is probably
electrical. If the machine seizes up, the problem is probably mechanical.
WARNING: After making this determination, return to DOS, type
A:MOTION ON<Enter> and reboot to restore normal operation. In the
MOTION OFF state, the machine will ignore motion failure, thus putting the
patient at risk if motion failure occurs during a scan!
7-3
Symptoms
Recommended Action
7.1.4
Symptoms
noisy X- or
Y- axis motion
noisy Y-axis
motion
noisy filter
wheel
noisy fans
7.1.5
Possible Causes
Recommended Action
Go To
bellows binding
adjust bellows
Figure 5-7
clean fan
replace fan
Figure 5-7
Possible Causes
Recommended Action
Go To
closed shutter
bad
open shutter
replace laser
weak laser
repair/replace cable
tighten tank gasket screws
4.2
Figure 5-7
MISCELLANEOUS PROBLEMS
Symptoms
no laser spot
visible
6.12
see
Caution
below
9.10
Note: If an error message appears, refer to Section 11 for an explanation of the message and its
possible cause.
CAUTION: If cleaning laser aperture, do not look into the beam. Turn off laser before
inspecting.
7-4
Symptoms
Possible Causes
Recommended Action
cant boot
7.2.2
replace drive
replace computer
Go To
6.15
Symptoms
Possible Causes
Recommended Action
intermittent
problems
reseat boards
cant format
cant archive to
floppy disk
cant archeive
to JAZ drive
can't mount
optical disk
cartridge
can't archive to
optical disk
Go To
replace drive
replace cartridge
replace drive
reinsert cartridge
bad cartridge
replace cartridge
replace board
replace drive
write-enable cartridge
cartridge is full
bad cartridge
replace board
replace drive
replace battery
7-5
Symptoms
Possible Causes
Recommended Action
unable to view
scan
unable to view
scan graphics
(text OK)
replace board
7.2.3
Go To
When the computer boots at power-up or reset, Windows 95 (or later) scans the memory and
peripherals and performs a cursory check on them. If there is a problem, the monitor displays the
device causing the error. If the computer finds a faulty device, here are some things you can do to
find and correct the problem:
3 Double-click on
Troubleshooting
1 Click on
2 Click on
Start
4 Select desired
Help
topic.
printing
starting Windows
using Dial-Up
Networking
using a PC
card(PCMCIA)
7-6
7.2.4
If you can not boot the computer, that is Windows does not appear on the monitor, you can boot
in a restricted mode by doing the following:
1. Boot the computer
Windows displays the following:
Starting Windows 95. . .
2. While this message is displayed on the screen, press <F8>.
3. From the menu displayed, select the boot mode desired by highlighting using the
<Arrows> or typing in the corresponding number and pressing <Enter>.
Windows allows you to boot in eight different modes in order to isolate problems. The following
table describes the different modes:
No.
1
2
Command
Normal
Logged (\BOOTLOG.TXT)
Safe mode
Step-by-step confirmation
Function
Normal boot
Normal boot but creates a log file called bootlog.txt
located in the root directory (C:\) on the C drive. This
file contains information about which files loaded
correctly.
Starts Windows using a basic configuration. The
system may not operate as usual but you can make
changes and reboot.
Starts Windows using a basic configuration but
includes network capabilities.
Starts the computer normally but prompts you at
every step before continuing.
Boots as normal but does not bring up Windows. You
can start Windows by typing C:\WIN<Enter>.
Starts the computer with a basic configuration and
does not start Windows.
Starts MS-DOS using files that were backed up
during Windows Setup. The version of MS-DOS
depends on what version you were running before
the current version of Windows was installed.
7-7
7.3.1
The Buttons
Definition
Cascade or view
layered windows
Minimize window
and place in taskbar
Maximize size of
window
Exit and close
window
The Window
7.3.2
Cursor
Text Cursor
Left Mouse
Button
7.3.3
7-8
1 Select Start
2 Select Help
7.3.5
C:\WINDOWS\Setup.txt
other trouble
C:\WINDOWS\Readme.txt
7.3.6
1 Select Help
2 Select Help Topics
Note: You can also
press <F1>.
7-9
Select the help facility you are looking for from this window.
7.3.7
If a dialog box appears, you can get help on any field or tab by doing the following:
the printer
the monitor
the keyboard
the mouse
Internal
1. the floppy drive
2. the C-drive
7-10
7.4.2
(Explorer).
Floppy Drive
Hard Drive
JAZ or M/O Drive
CD-ROM
You should see all the drives currently on the computer. If not, there is a problem.
4. Double click on Printers.
The following window appears:
7-11
Printer currently
installed on the
computer.
You should see all the printers (usually just one) currently on the computer. If not, there is a
problem.
Checking the Disks
1. At the Windows Desktop, press <Alt><F1> to go to the DOS prompt.
2. At the DOS prompt, type CD ..<Enter> to go to the root directory.
3. Then type SCANDISK<Enter>.
The following window appears:
SCANDISK is a Windows utility that scans the disk for proper directory structure, lost files, disk
errors, etc. You can run SCANDISK on the three disks listed in the dialog box above.
4. Insert a scratch disk into the floppy drive or into the removable disk (magnetooptical or JAZ, if installed) and press Start. This gives you a confidence check on the
selected disk.
Note: When you boot the computer, Windows scans the C-drive according to the Hologic
startup file installed in the QDR 4000. So you may want to limit disk scanning to the
removable drives and leave the C-drive for the automatic scan. If the C-drive had a
problem, you would not be at this point anyway.
Returning to the Hologic Main Menu
If you are in Windows, to get to the Hologic Main Menu, the best way is to reboot the computer.
This way, all the system parameters are set up properly.
7-12
Section 8
8. PREVENTIVE MAINTENANCE
8.1 CUSTOMER PREVENTIVE MAINTENANCE
In the QDR 4000 Users Guide, Hologic requires the customer to run a daily scan of the spine
phantom supplied with the QDR 4000, add that scan to the QC database, and to apprise Hologic
Field Service if the CV of the database exceeds 0.5%. Other than daily cleaning of the unit, no
other regular maintenance activity by the customer is required or recommended.
8-1
Section 9
9. COMPONENT SUMMARY INFORMATION
9.1 SCSI INTERFACE BOARD
This board is required only if a JAZ drive option is present.
To function properly, the board's EPROM (in position U4) must be removed, and all jumpers
(W1 through W5) must be removed as well.
The SCSI board is not field-repairable.
40
20
40
38
18
38
SW1
E3
JP4
E4
SW2
SW3
ASP2
1
2
3
4
Closed
Closed
Open
Open
SW2
1
2
3
4
Closed
Closed
Closed
Open
SW3
1
2
3
4
Closed
Closed
Closed
Closed
Ext. Rtn.
2
3
4
Ext. 5 VDC
Line Rtn.
Line
JP4
1-4
9-1
21
U20
1
40
1
20
15 16
16 15
21
U21
J2
2
15 16
J3
20
1
2
J4
2
1
SW4
SW3
SW2
SW1
15 16
15 16
1
2
3
4
Closed
Closed
Closed
Closed
SW2
1
2
3
4
Closed
Closed
Open
Open
SW3
1
2
3
4
Closed
Closed
Closed
Closed
SW4
9-2
Pin
1
2
3
4
5
6
Signal
XSM1
NC
XSM3
XSM4
NC
XSM6
1
2
3
4
Closed
Closed
Closed
Closed
T S 2
7
1
Connection
D1
D2
D3
D4
D5
D6
D7
D8
X Translator
To D2
To D1
X COM
NC
X AWO
NC
X DIR
X PULSE
To
*
*
*
*
Y Translator
To D2
To D1
Y COM
NC
Y AWO
NC
Y DIR
Y PULSE
To
*
*
*
*
* Distribution board
TS2
Connection
1
M1
M2
P1
M3
M4
P2
X Translator
NC
XSM1
XSM2
+28 VDC
XSM3
XSM4
28 V RET
To
*
*
P.S.
*
*
P.S.
Y Translator
NC
YSM1
YSM2
+28 VDC
YSM3
YSM4
28 V RET
To
**
**
P.S.
**
**
P.S.
9-3
A/D Power
J4
J5
J6
J7
J10
J15
Power Supply
(15VDC, +5VDC,
timing pulse input)
J16
J17
J18
Stepper
Translator
Controls
J19
J20
J24
Motor Controller
(Y motion count)
Motor Controller
(X motion count)
J14
A/D Signals
Front Panel
J22
Motor Controller
(X&Y motor direction)
J11
J12
J8
J9
Motor Controller
(X&Y motor speed)
F/B Limit
Switch
J2
J1
J13
L/R Limits
J21
X-Motor
X-Motor
Power (In) Power (Out)
9-4
Signal Name
Description
PRIMARY+
PRIMARY
FILCS
FILAMENT_NEUTRAL
FILAMENT_LINE
FILCS+
HV_SENS_
HV_SENS_+
Safety Ground
Torque Specifications
Set wrench to
Lexan Cup
10-12 in-lb
11 in-lb
Bladder Gasket
10-12 in-lb
11 in-lb
Transformer Seal
70 in-lb
70 in-lb
36-40 in-lb
38 in-lb
Lexan Cup
Bladder Gasket
Tank Cover
Gasket
Xformer (hidden)
Seal
Front
9-5
1
3
2
Figure 9-6. Lexan Cup
9.10.3 TANK BLADDER GASKET TIGHTENING SEQUENCE
Using the torque settings in the table above and referring to the figure below, tighten the Bladder
Gasket screws as follows:
1. Tighten screw #1, then #2, then #3, then #4.
2. From the screw after #4, tighten every other screw in a clockwise direction, returning back to
#4.
3. In a clockwise direction, tighten all the remaining screws.
9-6
FRONT
MOST CRITICAL
9-7
L1
J2
F1
TP2
F2
J1
L2
C12
TP12
TP1
J8
C18
TP3
J4
TP4
PTC1
TP11
J6
R38
D17
J5
R39
TP5
TP6
D20
J7
TP7
TP8
D23
PTC2
D24
D29
D30
TP9
J3
TP10
D33
D17
D20
D23
D24
D29
D30
D33
Warm Up
+5 V OK
Trim Enable
X-Ray Enabled 1
X-Ray Enabled 2
Zero Cross
+12 V OK
FUSES
F1
F2
R38
TP1
TP2
TP3
TP4
TP5
TP6
XRC Line 1
XRC Line 2
HVRef+340 VDC Return
+340 VDC
Prim +
Prim H/V Sense H/V Sense +
POTENTIOMETERS
TEST POINTS
9-8
JACKS
J1
J2
J3
J4
J5
J6
J7
J8
To X-ray tank
XRC Line Input
Filament AC
Filter Wheel AC
{Many Signals}
Filter Wheel Pickoff Signal
Filament Sense and H/V
Test Points
Fan AC
R39
TP7
TP8
TP9
TP10
TP11
TP12
HVRef +
Filament Current +
Filament Current Neutral
Filament Line
Filament Neutral
Filter Wheel Pickoff Signal
XRC Line 1 (120 VAC)
J3
1
19
37
20
J1
25
13
14
J2
U3
9-9
Section 10
10. SOFTWARE
10.1 SOFTWARE INSTALLATION AND UPGRADES
10.1.1 PROTECTION OF CUSTOMER'S FILES
Always archive the customer's patient database and QC files before performing any
work that involves writing to the hard disk. Use a separate set of new disks, not the
customer's regular database archive diskettes.
Never format or partition the hard disk without first archiving the patient database, the
QC files, and all of the scans on the hard disk.
Never power down or reboot during a scan, or while the computer is accessing the hard
disk. Powering down and rebooting may be safely accomplished only by selecting
sHutdown and the Hologic main menu.
3.
4.
5.
6.
7.
10-1
SYSTEM.PRX
STANDARD.TBL
SCONFIG.TXT
VALIDOPT.TXT
LINST*.*
AWKI.EXE
WREBOOT.EXE
9. Exit the MS-DOS window (C:\>EXIT<Enter>).
10. Reboot the computer.
10-2
At the Windows Desktop, select [Start], then [Programs], then [MS-DOS Prompt].
At the C:\>WINDOWS> prompt, type CD\<Enter>.
At the C:\> prompt, type SET ASKMACHINE=1<Enter>.
Insert V9.5 diskette one (1) into drive A:.
At the C:\> prompt, type COPY A:\*.EXE<Enter>.
At the C:\> prompt, type LINSTALL<Enter>.
Follow the instructions on the screen to install the system software. When prompted,
indicate the model type for this installation.
When you are instructed to, reboot the system.
Answer "yes" to initialize patient and QC databases.
At the Hologic Main Menu, select [Utility], then [Service], then [dbManagement] then
[dbRecover].
Insert the last (only if there is just one disk) into drive A: and press <Enter>.
Press <Enter> again to do a complete database recovery.
Insert the remaining dBarchive disks (if any) as prompted by the recovery program.
Reinstall Hologic V9.5 System Software after the dbRecover operation through
Utility/Upgrade.
Section 10 - Software
15. If applicable, install the Version 9 System Software Patch disk through Utility/Upgrade.
16. Install the Option disks (U.S. Reference Database, NHANES, Forearm, Low Density
Spine, etc.) using Utility/Upgrade.
17. Install OnePage.
18. Install Background Printing.
19. Verify that the port and spool settings in Windows Printer Properties are correct.
20. Perform five (5) QC scans and inspect BMD, BMC, and Area plots.
21. Archive a phantom scan and restore the scan to the hard drive.
22. Perform dBarchive.
.EXE
.MSG,.NDX
.BAT
\DOS
\UTIL
Explanation
This contains the AUTOEXEC.BAT and CONFIG.SYS files, which are read by
DOS at boot up.
This folder contains most of the Hologic executable programs. It contains all
the programs that can be executed from the Hologic main menu. During
standard operation of the software, this is always the "current folder". Filetypes
stored here can be identified by extension:
Executable programs
Help files. These come in pairs, with the .NDX being an index into the .MSG
file. On newer systems, these files are stored in the language directories
(FLANGENG, etc.)
Batch programs. Examples are COUNTRY.BAT and ENVIRON.BAT which are
invoked at bootup.
This folder always contains MS-DOS executable files. It may also contain
Hologic executable tools. These Hologic programs are ordinarily invoked from
the DOS prompt, by service personnel, and are not used in day-to-day
operations.
On the most recent Hologic software releases, the Hologic executable tools
have been moved from \DOS to this folder.
10-3
\SCAN\50HZ
\FLANGENG
Explanation
This folder contains ASCII protocol files that control the scanning operation.
There is one of these files for each scan protocol available on the system.
These files all end with the extension .PRO. In addition, the ASCII file
RATE.DAT describes various stepper motor characteristics at 60 Hz operation.
This folder contains the same files as \SCAN\60Hz, except for 50 Hz operation.
On System Software versions 6.01 and higher, this folder contains text, help,
menu and message files that are specific to English language operation. It also
contains the executable file SETQPRM.EXE, which loads the QDR Parameter
area with English-language values. Other files are identifiable by extension:
This folder contains text, help, menu and message files that are specific to
French language operation.
\FLANGDEU
This folder contains text, help, menu and message files that are specific to
German language operation.
\XCDATA
This folder contains data files, master menu files and database-definition files.
Data for each scan is kept in two or three files, with names in the form:
P*.P* Processed data contains analysis results and the presentation
frame.
P*.R* Raw scan data.
where the name portion of the filename is identical for the files of a single scan.
The patient-scan database contains the patient biographies and the results of
all the scans that the system has performed (or that have been restored to the
hard disk of this system). It is kept in files:
PATSCAN2.H
PATSCAN2.DBD
PATBIOG2.DAT
PATBIOG2.KEY
PATBIOG2.KE1
PATSCAN2.DAT
PATSCAN2.KEY
PATSCAN3.H
PATSCAN3.DBD
PATSCAN4.H
PATSCAN4.DBD
XPATSCN2.DAT
OPTICAL.DAT
10-4
Section 10 - Software
Folder,
Extension or
Program
Explanation
OPTICAL.KEY
The normals curve database contains the reference curves provided in the
original software and added to the system by the user:
NCURV3.H
NCURV3.DBD
NCURV2.DAT
NCURV2.KEY
The QC database contains stored scan results (generally of phantoms) and QC
setup information:
QCDB4.DBD
QCDB4.H
QCDB5.DAT
QCDB5.KEY
QCSCALE2.DAT
QCSCALE2.KEY
The file XCSYSTEM.DAT contains the currently selected scan.
The file ONDISK contains the list of scans currently on the hard drive.
.DAT files are menu files that contain all the text that the user sees during
menu selections. Not all menus are supported this way. Those that are include:
ANALYZE.DAT Menu selection within ANALYZE.
MAINP.DAT
Hologic Main Menu
SCAN.DAT
Scan selection
UTILITYP.DAT
Utility Menu
A note on Workstations
AC-1000 workstations contain no scanning hardware. Therefore the directories \SCAN\60Hz and
\SCAN\50Hz do not appear. Neither does the Scan selection on the Hologic Main Menu, nor the
Local selection on the Utility Menu. The file COUNTRY.BAT is not invoked at bootup, so no
attempt is made to determine line frequency. In addition, various executable files that control the
hardware are erased at the factory during software installation.
10-5
Description
AUTOCOMP.BAT
AUTOSCAN.EXE
BACKLASH.EXE
FSCAL.BAT
LAVRCV.EXE
MOTION.EXE
MOTORX.EXE
MOTORY.EXE
RECALYZE.BAT
SURVEY.EXE
Note: All the programs contained on the utility disk are not listed aboveonly the ones needed
by the service engineer. Some of the ones not listed are used in manufacturing and some
are called by other programs.
Description
AUTOSCAN
-En
-Hn
-Lyy
-M
-Nnn
-O [file] log acquired scan filenames for use with AUTOANALYZE (for factory use
only). Default filename is AUTOSCAN.LOG.
10-6
Section 10 - Software
Program
Description
-Tt
-Wxx
-Xxx
-Yxx
AVRSCAN
BACKLASH
10-7
Description
displays the following information about the disk drive:
- the serial number of the volume
- of total disk space (in bytes)
- available on disk (in bytes)
- the size of each allocation unit (in bytes)
- number of allocation units on the disk
- total memory (in bytes)
- total free memory (in bytes)
Note: To check the disk for errors and repair them, it is better to use
SCANDISK.
COUNTRY
DATE
DATES
DBCHECK
DBMAN
CAUTION: Always run DBarchive before running INITDB. Failure to do so will result in
the permanent loss of data.
Program
Description
KEYBUILD
MAINMENU
FSCAL
MOTION ON
disables the motion detection safety system, which normally monitors X- and Yaxis carriage motion during a scan and shuts down the X-ray source if motion
failure is detected on either axis.
Note: Do not leave the system for use with patients without invoking
MOTION OFF and restoring safety system operation.
10-8
Section 10 - Software
Program
Description
MOTION
OFF
MOTORX,
MOTORY
control and monitor the X-axis and Y-axis drive stepper motors. Written primarily
for use by Engineering, they serve in the field as means for isolating the cause of
motor stalling (by changing the stepping rate), and for monitoring the manual
motion request switches, limit switches and Motion Counter board outputs.
The MOTORX display, which is typical, is shown below.
Note: See Section 10.3.3 for more information on MOTORX and MOTORY.
ONDISK
checks the hard disk for patient files not currently in the database, and adds any
that it finds. Safe to run anytime.
PAD
allows laser on/off control and control of table and arm motion from the keyboard.
Pressing any key 0-4 or 6-9 on the numeric keypad initiates the indicated action.
Pressing <DEL> stops it.
Motor Control Pad
8 IN
PgUp
9(HomeX)
(X)
4LEFT
6RIGHT
(Y)
(Y)
2 OUT
PgDn
3(HomeY)
(Y)
0POSITION LIGHT
10-9
SCANDISK
MODSCREC
SET
SURVEY
Description
run on older scans whenever the bone correction factor (BCF) is changed in the
ENVIRON.BAT file. Adjusts the BMC of older scans to be within 0.5% of the
BMC value printed on the phantom label (Section 4.10.2).
Example: C:\RECALYZE<Enter> after the desired scans have been selected.
checks and repairs disks under Windows.
when scans are copied via DOS, running SCANTODB will add them to the
database.
Example:
MODSCREC \XCDATA\P*.P* <Enter>
displays the current DOS path, line frequency, Q-numbers, machine model
number (QDR 4000) and serial number. Performance Series machine serial
numbers end in "P", others are numeric only.
allows monitoring of the A/D converter output, the status of the HVPS, the
system line frequency and the instantaneous HI/LO signal values for bone, tissue
and air. It also affords on/off control of the X-ray beam and the filter wheel motor,
and selection of the collimator.
Note: See Section 10.3.4 for more information about the SURVEY
program.
TIME
UNITS
displays the current time in the computer's internal clock, and allows that time to
be changed if desired.
allows selection of English or metric units for displays.
1
1
1
1
1
1
1
1
0
0
0
0
1
1
1
1
1
0
0
0
0
X-Motor Control
Command
G
ASCII/BIN
Number of Steps 05279 Reset
Slope
001 DoWhite
Rate
050 Wait
First Rate
001 Abort
Slope Divisor
0
Cy525
command code
1
1
1
1
1
(ASCII).
0000
10-10
Section 10 - Software
PARAMETER
Number Of Steps
Slope
Rate
First Rate
Slope Divisor
MOTORX
2000
180
127
003
MOTORY
2000
180
50
003
001
001
After all five parameters have been entered, highlight the Command line, press <Del> to clear
the line, then type + or - to determine desired direction of motion (see table below) and press
<Enter>.
COMMAND
+
-
MOTORX
arm to rear
arm to front
MOTORY
arm left
arm right
Again press <Del> to clear the Command line, then type G and press <Enter> to initiate motion.
Observe the Motion line at the bottom of the -Motor Status box (except in MOTORC): it should
change as the arm moves, verifying proper Motion Counter operation.
Repeatedly pressing <Enter> should cause the arm to reach a motion limit and the CW Limit or
CCW Limit status in the -Motor Status box to change from 0 to 1, indicating proper limit switch
operation.
Throwing the arm manual control switches should change 0 s to 1 s in the -Motor Status box
(thereby verifying proper operation) thusly:
SWITCH
arm FORWARD/BACK switch UP
arm FORWARD/BACK switch DOWN
arm LEFT/RIGHT switch LEFT
arm LEFT/RIGHT switch RIGHT
MOTORX
CW SWITCH
CCW SWITCH
-
MOTORY
CW SWITCH
CCW SWITCH
10-11
16-bit A/D
Binary Data
X-Ray Status
Readback
Registers
Settings
Registers
Error
Counters
Filter
Wheel
Signals
Function
Keys
Legend
Function
turns X-rays on
turns X-rays off
turns calibration on
Key
<F4>
<F5>
<F6>
Function
turns calibration off
turns filter wheel motor on
turns filter wheel motor off
CAUTION: Turning the beam on with <F1> bypasses the system's normal X-ray safety features.
Be especially careful to keep your hands, head and other body parts out of the
beam.
REGISTER
Chopper
DISPLAY
0 or 1
DESCRIPTION
Filter wheel pickoff pulses are detected.
0=no pickoff pulse detected.
1=pickoff pulse detected.
Hi/Lo
0 or 1
Interlock
ON or
OFF
10-12
Section 10 - Software
REGISTER
Emergency
DISPLAY
ON or
OFF
DESCRIPTION
ON when the red STOP button is pushed in or the key switch is off
(emergency status enabled).
OFF when red STOP button is pulled out and the key switch is
turned to the ON position (normal operation).
HiVoltage
ON or
OFF
A/D
Readback
16
Individual
Bits, each
0 or 1
The output of the Analog to Digital (A/D) Converter board. The bits
are displayed in two colors:
RED = The bit has not yet changed state since SURVEY has been
started or reset by the <HOME> key.
GREEN = The bit has changed state at least once since SURVEY
was started or was reset by the <HOME> key.
Pressing the <HOME> key resets all the bits back to red. This can
be helpful in finding any "stuck" bits.
The following six registers display the detected amplitude of the dual energy X-ray pulses after
passing through the three filter wheel segments. The information is displayed as A/D numbers (04095). The higher the number, the fewer X-rays that are hitting the detector. Readings that reach
either extreme (0 or 4095) indicate something wrong with the detector or A/D converter.
REGISTER
Hi Bone
Lo Bone
Hi Tissue
Lo Tissue
Hi Air
Lo Air
DISPLAY
|
|
0-6535
|
|
|
DESCRIPTION
140 kVp X-ray pulse through bone segment.
70 kVp X-ray pulse through bone segment.
140 kVp X-ray pulse through tissue segment.
70 kVp X-ray pulse through tissue segment.
140 kVp X-ray pulse through air segment.
70 kVp X-ray pulse through air segment.
Note: The A/D "raw" number displayed by SURVEY is inverted relative to the detector output
voltage level (4095=0 Volts and 00000=10.0 Volts).
The following registers indicate the results of operator commands to the HVPS\Source and filter
wheel motor.
REGISTER
WarmUp
DISPLAY
ON or OFF
DESCRIPTION
Normal state is OFF and should only be ON for approximately 3-5
seconds while the X-rays are coming on or going off.
HiVoltage
ON or OFF
X-ray
Enable
ON or OFF
10-13
DISPLAY
ON or OFF
DESCRIPTION
Defaults to OFF. A better name would be "PMT dark current offset".
When all readback numbers are 3762, the PMT dark current has
been cancelled out.
S/W Debug
ON or OFF
Filter
Wheel
ON or OFF
Frequency
50 Hz or 60
Hz
Note: SURVEY also has three counters which increment when there is a filter wheel pickoff
error or a Hi/Low phasing fault.
All three of the following counters must be at zero before X-rays can be produced. The
counters can be reset only by exiting SURVEY (press <Esc>) and then restarting the
program.
REGISTER
Pickoff Out
Of Phase
DISPLAY
0 or a
decimal
number
DESCRIPTION
Increments if the filter wheel pickoff is out of phase with the AC line.
i.e., if the pickoff pulse occurs during the negative half cycle of the
AC line.
If the Pickoff Out of Phase counter is incrementing, the filter wheel
can be phased manually by alternately pressing <F5> and <F6>
until the wheel is in phase and the counter stops incrementing.
Long filter wheel motor off times improve the chances of the wheel
synchronizing with the positive half-cycle of the AC line when
restarted.
No Filter
Wheel
Pickoff
0 or a
decimal
number
Hi-Lo Out
Of
Sequence
0 or a
decimal
number
10-14
Section 10 - Software
3. Type b<Enter>.
The following screen appears:
5. Enter the correct information in the screen above and press <F10>.
The following screen appears if you have just pressed <F10> or typed y<Enter> when asked if
the information is correct:
10-15
6. Select the 20 scans just completed. (Follow the directions on the screen.)
The following screen appears:
This screen displays the analysis of the scan values and recommends whether to recalibrate or
not.
7. Select y or n, depending on the recommendations and your decision.
If you select y, the following screen appears:
10-16
Section 10 - Software
10-17
Section 11
11. ERROR MESSAGES
The QDR 4000 system software reports four types of error messages. They are:
Informational:
User Error:
The user made a mistake, such as entering invalid data, inserting the
wrong diskette or optical cartridge, or positioning a region of interest
incorrectly.
Hardware Error:
Software Error:
The QDR software contains numerous "sanity checks" that are used
during software development to indicate that something is wrong.
During normal operation, these messages should never appear. If they
do, Hologic should be notified.
This document covers only hardware and software errors. Informational and user errors are
self-explanatory.
The following list specifies, for each error message:
The error messages are in alphabetical order, with the exception of messages generated by SCAN
in the form:
device: Message Text
These messages are listed in order by "Message Text".
Analysis aborted. d0= ___ below acceptable limit.
Hardware or
Software Error:
Action:
CALIB: Timeout
Hardware Error:
Action:
11-1
For some reason, the system was unable to obtain the results of a scan
that you selected for normals plotting. This message should be
reported to Hologic.
Action:
The system was unable to write a reference curve that you entered or
edited.
Action:
Check that the hard drive is not full. That is, check the message
"Room for __ scans" in the status window and ensure that it does not
say zero. If it is full, then archive and delete one or more scans and
try again.
If the hard drive is not full, then you have either a hard drive failure
or a corrupt reference curve database.
11-2
The device driver which performs the X-ray scan acquired data faster
than the scan program could write it to disk. This may indicate a hard
drive problem or a problem with the Timing & Control board.
Action:
Verify that there was space on the target drive. Check the "Room for
... " message in the status window. Run appropriate diagnostics.
Action:
11-3
Action:
Action:
Check for disk full. Get out of the Hologic menu (Press <Alt>
<F1>), go to the \WORM directory and run the program WCHKDSK
c:, where "c:" is the drive letter from the message.
If the disk is not full, then run the WDIAGS diagnostics.
11-4
Software Error:
The optical file that you are attempting to restore scans from seems to
be clobbered.
Action:
The data points did not correctly alternate between high and low
energy.
Action:
Check the Timing & Control board, replace if defective. Check the
A-C input reference voltage to the Timing & Control board.
11-5
Internal Error:..
or
Invalid ...
Software Error:
Any message that begins with the word "Invalid" should be reported
to Hologic. Please include as much information as possible about
what you were doing at the time.
There was an error reading or writing a disk drive. The message should
identify the drive and say something about the nature of the error.
Action:
Install the tissue bar initialization software that comes with the tissue
bar.
11-6
Hardware Error:
Action:
Action:
Action:
11-7
out of memory
Software Error:
Action:
Ensure that there is adequate space on the hard drive (Check the
"Room for ..." message in the status window. You should always
leave room for 1 or 2 scans). There may be a disk problem;
appropriate diagnostics should be run. If you are unable to find a
hardware problem, the message should be reported to Hologic.
Include as much information as possible about what you were doing
at the time.
PGLINE: ...
Software Error:
During a scan, the filter wheel phase at the start of every even line
must be the same. It must also be the same at the start of every odd
line. This condition was not met, indicating the line frequency may
not be 50 or 60 Hz This error usually occurs when running with a
motor-generator.
This may also indicate a software error involving the RATE.DAT
file.
11-8
One of the stepper motor controller chips did not respond correctly to
an attempt to read it. The message identifies which controller it is.
Action:
To diagnose problems with stepper motors, try going into DOS and
invoking the appropriate motor control program (MOTORX or
MOTORY). These allow direct control over the CY525 chip.
Replace the CY525 chip or the Stepper MotorController board if
defective.
Action:
Action:
Ensure that there is adequate space on the hard drive (Check the
"Room for ..." message in the status window. You should always
leave room for 1 or 2 scans). There may be a disk problem;
appropriate diagnostics should be run. If you are unable to find a
hardware problem, the message should be reported to Hologic.
Include as much information as possible about what you were doing
at the time.
11-9
If you are encountering problems with line frequency (see the description
of "Unable to determine Line Frequency" error message), then you can
get this message.
Otherwise, this message should be reported to Hologic.
During a scan, the starting chopper wheel phase for each line must be
one of the values: 0, 2, 4. This condition was not met.
Action:
Check the stepper motor controller. Ensure that you have the latest
chopper pickoff synchronization fix installed (Part number 061000195-01. This is included in all software releases 5.11.04 and later,
and 6.00.01 and later).
the analyze data file either did not exist or did not
contain any valid entries
Software Error:
Action:
Action:
11-10
When restoring files from optical ("Optical" on main menu), you selected
an archive that contains no scans. This situation should not occur in
normal operation with current software (versions 5.20 and later).
Action:
Action:
Action:
Verify that the Chopper Pickoff Synchronization fix (Part no. 061000195-01) has been installed. This fix is included in software
releases 5.11.04 and later and 6.00.01 and later).
Check filter wheel pickoff for residual noise, add shielded cable A/R.
Check Timing & Control board and reseat, modify or replace as
required.
One of the stepper motor controller chips did not respond correctly to
an attempt to write to it. The message identifies which controller it
is.
Action:
To diagnose problems with stepper motors, try going into DOS and
invoking the appropriate motor control program (MOTORX or
MOTORY). These allow direct control over the CY525 chip.
Replace the CY525 chip or the Stepper Motor Controller board if
defective.
This message indicates a problem with the scan protocol file (that
instructs the scan program how to perform a scan). If this happens on
one of the officially-released Hologic scan protocols, Hologic should
be notified.
11-11
Action:
Ensure that there is adequate space on the hard drive (Check the
"Room for ..." message in the status window. You should always
leave room for 1 or 2 scans). There may be a disk problem;
appropriate diagnostics should be run. If you are unable to find a
hardware problem, the message should be reported to Hologic.
Include as much information as possible about what you were doing
at the time.
Action:
11-12
User or Hardware
Error:
This usually means that you attempted to format a diskette that was
defective.
Action:
11-13
Action:
Ensure that there is adequate space on the hard drive (Check the
"Room for ..." message in the status window. You should always
leave room for 1 or 2 scans). There may be a disk problem;
appropriate diagnostics should be run. If you are unable to find a
hardware problem, the message should be reported to Hologic.
Include as much information as possible about what you were doing
at the time.
Action:
This usually means that the hard drive is full. Check the "Room for
"message in the status window. If the disk is NOT full, then this
indicates a hard drive problem.
11-14
The filter wheel belt may be too tight. Check the tension, adjust as
required.
The software was not properly installed. The device driver SM.SYS
(loaded by command in CONFIG.SYS at system startup) does not
appear to be there.
11-15