Procedia Engineering 25 (2011) 71 75

Proc. Eurosensors XXV, September 4-7, 2011, Athens, Greece

Advanced photonic biosensors for point-of-care diagnostics


A.B. Gonzlez-Guerreroa, S. Dantea, Daphn Duvala, J. Osmondb, Laura M. Lechugaa*
a

Nanobiosensors and Bioanalytical Application Group, Research Center on Nanoscience and Nanotechnology (CIN2)
CSIC& CIBER-BBN, Campus UAB, 08193 Bellaterra, Spain
b
Institute of Photonic Sciences (ICFO), 08860 Castelldefels, Spain

Abstract
The application of portable, easy-to-use and highly sensitive biosensor devices for real-time diagnosis could offer
significant advantages over current analytical methods. Silicon photonic biosensors based on evanescent wave
detection have revealed themselves as the most promise candidates for achieving truly point-of-care devices.
Advantages as miniaturization, extreme sensitivity, robustness, reliability, potential for multiplexing and mass
production at low cost can be offered. We present our work towards the assembly of a point-of-care device by using
integrated silicon interferometers sensors, polymer microfluidics and diffractive grating couplers for light incoupling. The micro/nanointerferometric devices have shown sensitivity close to 10-7 in refractive index units, which
means an ability to discern, in a label-free scheme, concentrations of biological molecules at pM level.

2011 Published by Elsevier Ltd. Open access under CC BY-NC-ND license.


Keywords: optical biosensor; lab-on-a-chip; interferometers; grating coupler.

1. Introduction
Nomenclature
LOC

Lab-on-a-chip

MZI

Mach-Zehnder interferometer

BiMW

Bimodal optical waveguides

DGC

Diffractive Grating Couplers

Most diagnostics techniques are based on time-consuming, expensive, and specialised techniques
performed by trained technicians at laboratory level. These techniques typically require labelling of the
samples or reagents with fluorescent or radioactive markers. It is clear that application of a portable, easy-

1877-7058 2011 Published by Elsevier Ltd. Open access under CC BY-NC-ND license.
doi:10.1016/j.proeng.2011.12.018

72

A.B. Gonzlez-Guerrero et al. / Procedia Engineering 25 (2011) 71 75

to-use and highly sensitive lab-on-a-chip (LOC) platform for real-time diagnosis could offer significant
advantages over current methods. Main applications fields of this technology will be clinical diagnostics,
environmental monitoring, chemical and biological warfare surveillance, food industry and veterinary and
industrial process control, among others. But despite remarkable progress towards LOC clinical assay
systems, very few complete working prototypes have emerged [1]. Most of the LOC technology does not
incorporate on-chip detection and the read-out must be done with complex scanners in laboratory settings.
Incorporation of the on-chip detection by using biosensors is a new technology that shows great
promise.
Grating couplers

Surface biofunctionalization
&hdhZKhd>KK<

Integrated Microfluidics

Optical readout

Array of
Nanophotonic sensors

Electronics and
data processing

Fig. 1. Scheme of our envisioned lab-on-a-chip device

Due to the scalable fabrication and the significant sensitivities (pM-fM) for biomolecular sensing, chipintegrated photonic array structures are excellent candidates as detectors for lab-on-a-chip applications
(LOC). Integrated optical biosensors as grating-couplers, interferometers, photonic crystal, microring
resonators or slot waveguides have been extensively studied in the last years [2-7] but few of them are
commercially available yet. Interferometric biosensors have a broader dynamic range and show the best
sensitivity levels achieved so far for label-free measurements (detection at the picomolar level, with
variations of 10-7 in the refractive index-equivalent to 0.1pg/mm2) [6]. But the main drawback in the
possible commercialization of the integrated interferometric devices is the complexity of the design,
fabrication, optical adjustments and their apparently inherent instability, which is a consequence of their
high sensitivity.
In order to solve the above problems and for offering a reliable diagnostic tool, we are working in the
development of LOC devices using interferometric micro/nanosensors based on silicon technology (see
Figure 1). In the case of using integrated Mach-Zehnder Interferometer (MZI), a first Y-junction splits the
guided light into two arms, the sensing arm and the reference arm which compensates for refractive index
fluctuations and unspecific adsorption [6]. After a certain distance, the two signals are recombined into an
output optical waveguide via a second Y-junction, producing the interference of both beams (Figure 1a).
Moreover, MZI transducers present the possibility of developing biosensor microarrays, allowing the
parallel analysis of a large amount of different samples on a single chip.
Another innovative strategy which we have recently developed, is the bimodal waveguide (BiMW)
interferometer [7], a single channel waveguide interferometer which mechanism is based on the
interference of two waveguide modes of the same polarization (figure 1b). First, the light is coupled into a
rib waveguide that supports a single transversal mode. After some distance, this mode is coupled into
another waveguide which supports two transversal modes, the fundamental and the first order modes.
Both modes are propagating until the output of the chip where they create an interference pattern. As the
fundamental and first order modes have different intensity distribution at the core-cladding interface, the
interference pattern is a function of the refractive index of the sensor area. The simplicity of the design of

A.B. Gonzlez-Guerrero et al. / Procedia Engineering 25 (2011) 71 75

these bimodal waveguide (BiMW) interferometers is attractive for mass-fabrication as there is no further
need for using Y-shape splitters, which are the most complex component of the MZI devices.

Fig 2. Scheme of a) a MZI and b) a BiMW interferometer

2. Implementation of the lab-on-a-chip


The lab-on-a-chip is being assembled by integrating the following parts: (i) micro/nano sensors
fabricated with standard silicon technology, (ii) a polymer microfluidic and the flow delivery system (iii)
diffractive grating coupler for the in coupling of the light in all the photonic channels for multisensing
(iv) reliable immobilisation and regeneration protocols for the biological receptor (iv) photodetectors,
electronic & software control (v) final integration and packaging.
2.1. The sensor chip and the microfuidics
We employ dielectric waveguides (Si3N4) as core waveguides for both the MZI and the BiMW devices.
Before fabrication, modeling and numerical simulation was done using Finite Element Beam Propagation
methods for defining the proper parameters of the waveguides and layout with optimised power
amplitude. The final layout design included dozens of chip-integrated waveguides into a single wafer
substrate using silicon waveguide technology (Si/SiO2/Si3N4/SiO2). For the MZI, the waveguide core is
200 nm of Si3N4 (n=2.00). For the BiMW, the single mode part is 150 nm of Si3N4 (n=2.00) thick and the
bimodal part is 350 nm thick. The rid height is only 2 nm for both devices. Both devices are designed for
operation at visible wavelengths with monochromatic and polarized (TE or TM) light.

Fig. 3. (a) Picture of one chip with MZI devices; (b) Picture of one chip with 16 BiMW devices and sealed with the polymer
microfluidics.

The flow system consists of a flow cell with four independent channels fabricated in PDMS. The
integrated microfluidic/sensor platform is supplied with tube connections to the exterior. The flow
delivery system includes a syringe pump and injection valves. Fig. 3b) shows a photo of one BiMW chip
inside the flow cell.

73

74

A.B. Gonzlez-Guerrero et al. / Procedia Engineering 25 (2011) 71 75

2.2 Light incoupling


Diffractive grating couplers allow an efficient light coupling into the waveguides with high coupling
efficiency. Due to the dimensions of the interferometers and to the operating wavelength (visible range),
the diffractive grating length cannot exceed 100 m with a sub-micronic period. The grating has been
directly written onto the Si3N4 waveguide by electron-beam lithography. Fig. 4 shows the layout of a
MXI chip and the position of the grating. The grating period is 400 nm, with a duty cycle of 0.5 and a
depth of 40 nm (total dimensions: 20 m x 100 m). Fig. 4 also shows photos of the grating couplers and
light incoupling into a MZI device.
3.2 cm

D0
W0
I1
D1
W1
I2
I3
I4
I5

1 cm
W0
D0
W1
D1
W2
I2
I3
I4

Z t'

I5

Fig. 4. Layout of a MZI chip. Photos of the grating couplers and in-coupling of the light in a MZI device.

3. Sensitivity and biosensing evaluation


The sensitivity of MZI and BiMW devices was checked by injecting decreasing concentrations of HCl.
Recording phase variation versus refractive index variation for each concentration gives a detection limit
of 'nmin=1.110-7 RIU and 'nmin=2.510-7 RIU for MZI and BiMW, respectively. These values indicate
that the interferometric devices are one of the most sensitive sensors and could allow anultrasensitivity
label-free detection of biomolecular interactions. For biosensing evaluation, proteins, bacteria and DNA
have been tested and the devices have shown an excellent performance for label-free detection with the
immunosensing of hormones at 1 pM level and bacteria detection at 10 cfu/ml level.

We are working on the realization of a sensitive, affordable, hand-held and portable device for point of
care diagnosis. The lab-on-chip device includes ultrasensitive interferometric biosensors based on
integrated silicon-based optical waveguides which have already shown their ability to detect minute
YDULDWLRQRIWKHUHIUDFWLYHLQGH[ Q min =1.10-7) and label-free biosensing detection at pM level. Light in
coupling into the sensor waveguides is achieved by diffraction gratings couplers fabricated by electron
beam on top of the rib waveguides. Hermetic sealed microfluidics is achieved by using a polymer
cartridge. Multiplexed excitation and read-out of all the sensors within a single chip is in progress.
Acknowledgements
Authors acknowledge the financial support from M. Botn Foundation, SENA project (TSI-020301-200811) from the Spanish Ministry of Industry and LEUKEM-CHIP project from CIBER-BBN. This work
was partially developed using the micro and nano fabrication capabilities of the ICTS/clean room from
the IMB-CNM (CSIC) under GISERV

A.B. Gonzlez-Guerrero et al. / Procedia Engineering 25 (2011) 71 75

References
[1] Ligler, F. S. Perspective on Optical Biosensors and Integrated Sensor Systems. Anal. Chem. 2008; 81:519526.
[2] X. Fan, I. M. White, S. I. Shopoua, H. Zhu, J. D. Suter and Y. Sun, Sensitive optical biosensors for unlabeled targets: A
review . Anal. Chim. Acta, 2008; 620:826.
[3] D. Erickson, S. Mandal, A. H. J. Yang and B. Cordovez, Nanobiosensors: optofluidic, electrical and mechanical approaches
to biomolecular detection at the nanoscale. Microfluid. Nanofluid., 2008;4: 3352.
[4] F. Vollmer and S. Arnold,. Whisperry gallery mode biosensing: label-free detection down to single molecules. Nat. Methods,
2008; 5:591596.
[5] A.L. Wasburn and R.C. Bailey. Photonics-on-a-chip: recent advances in integrated waveguides as enabling detection
elements for real-world, lab-on-a-chip biosensing applications. Analyist. 2011; 136: 227-236.
[6] M.C. Estevez, M. lvarez and L. M. Lechuga. Integrated Optical devices for lab-on-a-chip biosensing applications. Laser &
Photonics Reviews, 2011 (In Press)
[7] K. E. Zinoviev, A.B. Gonzlez-Guerrero, C. Domnguez and L. M. Lechuga. Integrated Bimodal Waveguide Interferometric
Biosensor for Label-free Analysis. J. Lightwave Tech. 2011; 29 (13): 1926-1930.

75