III-236
SCOPE
Digital image acquisition, display, and storage can be applied to radiography and radioscopy. Once the analog image is converted to digital format, the data can be
displayed, processed, quantified, stored, retrieved, and
converted back to the original analog format, for example,
film or video presentation.
Digital imaging of all radiographic and radioscopic examination test results shall be performed in accordance
with the modified provisions to Article 2 as indicated
herein.
III-250
III-255
III-220
III-221
PROCEDURE REQUIREMENTS
III-258
III-231
III-260
SYSTEM CONFIGURATION
CALIBRATION
III-263
III-234
AREA OF INTEREST
III-230
GENERAL REQUIREMENTS
III-222
VIEWING CONSIDERATIONS
III-280
EVALUATION
III-286
ARTICLE 2
III-287
SYSTEM-INDUCED ARTIFACTS
III-292
The digital images shall be free of system-induced artifacts in the area of interest that could mask or be confused
with the image of any discontinuity in the original analog
image.
III-290
III-291
ARTICLE 2
EVALUATION BY MANUFACTURER
DOCUMENTATION
DIGITAL IMAGING TECHNIQUE
INFORMATION
To aid in proper interpretation of the digital examination data, details of the technique used shall accompany
the data. As a minimum, the information shall include
items specified in T-291 and II-221 when applicable,
III-221, III-222, and the following:
(a) operator identification
(b) system performance test data
21
2013 SECTION V